Bone Conduction Hearing Aid (XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3)
Applicant
Shenzhen Magnet Technology Co., Ltd.
Product Code
LXB · Ear, Nose, Throat
Decision Date
Jun 11, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3302
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Bone Conduction Hearing Aid (models: XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3) is wearable sound-amplifying device intended to compensate impairments in personal hearing. The fundamental operating principle is to receive, amplify, and transfer sound via the skin and the bone of the skull to the inner ear of a hearing impaired person. The amplification suits the needs of a mild to a moderate hearing loss. They require individual fitting in performance executed by a hearing aid professional. The device is intended for adults and children (≥12 years).
Device Story
Wearable bone conduction hearing aid; compensates for mild-to-moderate hearing loss. Input: ambient sound via internal transducer; signal conditioning unit processes audio. Output: mechanical vibrations transferred through skin/skull to inner ear. Operated by patient in home/public environments; requires professional fitting. Features Bluetooth 5.3 connectivity for mobile devices. Three scenario modes (noisy, conventional, quiet) selectable by user. Benefits: improved hearing clarity; reduced noise floor compared to predicate. Output limitation circuitry prevents damage to remaining hearing.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1, -1-2, -1-11), electromagnetic immunity (IEC 60118-13), performance characteristics (IEC 60118-9), battery safety (IEC 62133-2), biocompatibility (ISO 10993-5, -10, -23), and cybersecurity/wireless coexistence (UL 2900-1, IEEE C63.27, AAMI TIR69).
Technological Characteristics
Materials: ABS, PC, Silicone. Sensing: Input transducer, signal conditioning unit, bone conduction output transducer. Energy: 210mAh rechargeable Li-ion polymer battery. Form factor: Headband-mounted. Connectivity: Bluetooth 5.3. Software: Digital programmable, 2-channel, volume/mode control, output limiting. Standards: IEC 60118-9, IEC 60601-1, IEC 62304.
Indications for Use
Indicated for adults and children (≥12 years) with mild to moderate hearing loss requiring sound amplification via bone conduction.
Regulatory Classification
Identification
A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.
{1}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 11, 2026
Shenzhen Magnet Technology Co., Ltd.
% Riley Chen
RA specialist
Feiying Drug & Medical Consulting Technical Service Group
Rm.2401 Zhenye International Business Center, # 3101-90, Qianhai Rd.
Shenzhen, Guangdong 518052
China
Re: K261256
Trade/Device Name: Bone Conduction Hearing Aid (XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3)
Regulation Number: 21 CFR 874.3302
Regulation Name: Bone-Conduction Hearing Aid
Regulatory Class: Class II
Product Code: LXB
Dated: April 16, 2026
Received: April 16, 2026
Dear Riley Chen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261256 - Riley Chen
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K261256 - Riley Chen
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
SHUCHEN PENG -S
Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261256 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Bone Conduction Hearing Aid (XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3) | | |
| Please provide your Indications for Use below. | | ? |
| Bone Conduction Hearing Aid (models: XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3) is wearable sound-amplifying device intended to compensate impairments in personal hearing. The fundamental operating principle is to receive, amplify, and transfer sound via the skin and the bone of the skull to the inner ear of a hearing impaired person. The amplification suits the needs of a mild to a moderate hearing loss. They require individual fitting in performance executed by a hearing aid professional. The device is intended for adults and children (≥12 years). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
Shenzhen Magnet Technology Co., Ltd
510(k)s – 510(k) Summary---
# 510(k) Summary-K261256
**'510(k) Summary' as required by 21 CFR Part 807.92.**
## I. Submitter
Shenzhen Magnet Technology Co., Ltd
No.901, Building 9 of Zhongyuntai Industrial Park, Songbai Rd., Tangtou Community, Shiyan St.,
Bao'an Dist., Shenzhen, China
Post code: 518108
Fax: +86 0755-82598747
Xiangshun Tian
Title: Engineer
Tel.: +86 - 13699780460
Email: 51683858@qq.com
Date of preparation: 2026-6-02
## II. Device Information
Name of Device: Bone Conduction Hearing Aid
Models: XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3
Common or Usual Name: Hearing aid, bone conduction
Regulation Description: Bone-conduction hearing aid
Regulatory Class: II
Product Code: LXB
Regulation Number: 21 CFR 874.3302
## III. Predicate Device
| Predicate Device | Manufacturer | 510(k) Number | Product code |
| --- | --- | --- | --- |
| Bone conduction hearing aid | BHM-Tech | K121793 | LXB |
## IV. Device Description
The bone conduction hearing aid is intended as hearing compensation of bone conduction hearing loss patient. The intended operator is the patient (adults and children $\geq 12$ years) with mild to a moderate hearing loss. No special skill, training or knowledge is required; the operator needs to read and follow the instruction manual. The device is intended to be used in home healthcare environment and public environment such as office. Hearing aid is made up of a bone conduction (output) transducer, control box (including an input transducer and a signal conditioning unit), battery (an internal rechargeable lithium-ion polymer battery), and headband (with built-in wiring).
The hearing aid can connect with the mobile phone through Bluetooth, the Bluetooth version is 5.3. Mobile devices used by users must support Bluetooth 4.2 or above.
Page 1 of 6
{6}
Shenzhen Magnet Technology Co., Ltd
510(k)s – 510(k) Summary
## V. Indications for Use
Bone Conduction Hearing Aid (models: XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3) is wearable sound-amplifying device intended to compensate impairments in personal hearing. The fundamental operating principle is to receive, amplify, and transfer sound via the skin and the bone of the skull to the inner ear of a hearing impaired person. The amplification suits the needs of a mild to a moderate hearing loss. They require individual fitting in performance executed by a hearing aid professional. The device is intended for adults and children (≥12 years).
## VI. Comparison of Technological Characteristics With the Predicate Device
The Bone Conduction Hearing Aid (models: XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3) has the same intended use as the predicate device. The technological characteristics, features, specifications, design are similar to the predicate device. Any minor differences between the subject device and the listed predicate device do not raise any issues of safety or efficacy. Performance data supports that the device is substantially equivalent to the predicate device for its intended use.
| Comparison Elements | Subject Device | Predicate Device | Remark |
| --- | --- | --- | --- |
| **K number** | **K261256** | **K121793** | / |
| Trade name | Bone Conduction Hearing Aid | Bone conduction hearing aid | / |
| Model | XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3 | An Evo1 | / |
| Regulation number | 874.3302 | 874.3302 | Same |
| Device class | Class II | Class II | Same |
| Product code | LXB | LXB | Same |
| Indication for use/ Intended use | Bone Conduction Hearing Aid (models: XTS-AISW-D1, XTS-AISW-D2, XTS-AISW-D3) is wearable sound-amplifying device intended to compensate impairments in personal hearing. The fundamental operating principle is to receive, amplify, and transfer sound via the skin and the bone of the skull to the inner ear of a hearing impaired person. The amplification suits the needs of a mild to a moderate hearing loss. They require individual fitting in performance executed by a hearing aid professional. The device is intended for adults and children (≥12 years). | Bone conduction hearing aids by BHM-Tech are wearable soundamplifying devices intended to compensate impairments in personal hearing. The fundamental operating principle is to receive, amplify, and transfer sound via the skin and the bone of the skull to the inner ear of a hearing impaired person. The amplification suits the needs of a mild to a moderate hearing loss. They require individual fitting in performance executed by a hearing aid professional. The target populations for the devices are as follows: AN-Evo1: adults and children (≥12 years) | Same |
| OTC or Prescription | Prescription use | Prescription use | Same |
Page 2 of 6
{7}
Shenzhen Magnet Technology Co., Ltd
510(k)s – 510(k) Summary
| **Comparison Elements** | **Subject Device** | **Predicate Device** | **Remark** |
| --- | --- | --- | --- |
| **K number** | **K261256** | **K121793** | / |
| Software/Firmware/ Microprocessor Control? | Yes | Yes | Same |
| Power Source | Polymer lithium battery (210mAh) | Battery type 675 | Different **Note 1** |
| Design | Position transducers of the hearing aid over the temporal bones, close to ears and place the headband on the the back side of head. | Bone Conduction Aid which can be mounted with special Extension tips on different Eyeglasses. | Different **Note 2** |
| Target population | Adults and children (≥12 years) | Adults and children (≥12 years) | Same |
| Materials | ABS, PC, Silicone | Medical Grade plastics | Different, but solved by biocompatibility test |
| Operation mechanism | Circuit type: Digital Programmable: Yes Volume Control Channels: Two Volume control: Yes Mode switch: Yes Direct Audio Input: No Induction Coil: No Low Battery Indication: Yes Trimmer: Yes Program Switch Tones: Yes Output-Limitation: Yes | Circuit type: Digital Programmable: Yes Channels: Two Volume control: Yes Mode switch: Yes Direct Audio Input: No Induction Coil: Yes Low Battery Indication: Yes Trimmer: Yes Program Switch Tones: Yes Output-Limitation: Yes, MPO (Maximum Power Output) | Similar **Note 3** |
| **Reporting of data** | **According to IEC 60118-9:2019** | **According to DIN IEC 118-9:1987** | / |
| Maximum OFL90 (dB) | XTS-AISW-D1: 106 XTS-AISW-D2: 105 XTS-AISW-D3: 104 | 117dBOFL | Similar **Note 4** |
| High-frequency average OFL90 (dB) | XTS-AISW-D1: 101 XTS-AISW-D2: 99 XTS-AISW-D3: 97 | 113dBOFL | Similar **Note 4** |
| Maximum Full-on Gain (dB) | XTS-AISW-D1: 29.7 XTS-AISW-D2: 27.8 XTS-AISW-D3: 29.9 | 48dB | Different **Note 4** |
| Full-on Gain (High-Frequency Average) (dB) | XTS-AISW-D1: 26.5 XTS-AISW-D2: 25.0 XTS-AISW-D3: 23.5 | Not publicly available | Different **Note 4** |
Page 3 of 6
{8}
Shenzhen Magnet Technology Co., Ltd
510(k)s – 510(k) Summary---
| **Comparison Elements** | **Subject Device** | **Predicate Device** | **Remark** |
| --- | --- | --- | --- |
| **K number** | **K261256** | **K121793** | / |
| Equivalent Input Noise Level (dB) | XTS-AISW-D1: 11 XTS-AISW-D2: 13 XTS-AISW-D3: 15 | 26dB | Different **Note 5** |
| Total Harmonic Distortion (THD) | XTS-AISW-D1: 1.5% XTS-AISW-D2: 1.5% XTS-AISW-D3: 1.5% | Not publicly available | Different **Note 6** |
### **Comparison in Detail(s):**
#### **Note 1:**
Though the power source is different from the predicate device, they are both powered by internal rechargeable battery. The lithium battery of the subject device complies with the IEC 62133-2 standard, and the device is complies with IEC 60601-1 and IEC 60601-1-2 requirements, so this difference will not raise any safety or effectiveness issues.
#### **Note 2:**
Though the design of the subject device is a little different from the predicate device, the use method and work principle of the subject device and the predicate device is the same, their transducer of the device is held against the head and is driven electrically by the amplifier to transmit the amplified sound as vibrations to the underneath bones of the skull, therefore, this difference will not raise any safety or effectiveness issues.
#### **Note 3:**
Though the subject device is not equipped with an induction coil, the device offers users with three scenario modes (Mode 1: Noisy environment, Mode 2: Conventional environment, mode 3: Quiet environment) which can meet users' daily needs, and the device complies with IEC 60118-9 and IEC 60118-13 requirements, so this difference will not raise any safety or effectiveness issues.
#### **Note 4:**
Though the 'Maximum OFL90', 'High-frequency average OFL90', 'Maximum Full-on Gain' and 'Full-on Gain (High-Frequency Average)' are a slightly different from the predicate device, the subject device complies with the IEC 60118-9 standard and the parameters specification do not exceed the output tolerance range specified by the standard, which ensures that the device can precisely meet the hearing compensation needs of patients. Since the parameter settings depend on the degree of the user's hearing loss, the core of a hearing aid lies in precisely limiting the parameter output within a safe range to prevent any sound damage to the patient's remaining hearing. Therefore, such minor differences will not raise any safety or effectiveness issues.
#### **Note 5:**
Equivalent Input Noise Level reflects the noise level generated by the internal circuits of the hearing aid.
Page 4 of 6
{9}
Shenzhen Magnet Technology Co., Ltd
510(k)s – 510(k) Summary
A lower equivalent input noise indicates that the device will not produce noticeable hissing sounds in a quiet environment, thereby providing a more comfortable listening experience. The equivalent input noise level of the subject device is smaller than that of the predicate device. Since a smaller equivalent input noise level means lower noise generated by the hearing aid, the performance of the subject device is superior to the predicate device, and the subject device also complies with the IEC 60118-9 requirements. Therefore, this difference will not cause any safety or effectiveness issues.
**Note 6:**
Lower total harmonic distortion indicates clearer and more accurate audio output, and the subject device complies with the IEC 60118-9 requirements, therefore, this difference will not cause any safety or effectiveness issues.
**VII. Non-Clinical Testing**
The following performance data were provided in support of the substantial equivalence determination.
1) Electrical Safety
- IEC 60601-1: 2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC TS 60601-4-2: 2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-1-11:2020 Medical Electrical Equipment –Part 1-11: General Requirements for Basic Safety and Essential Performance –Collateral Standard: Requirements for Medical Electrical Equipment and Medical Electrical Systems Used in the Home Healthcare Environment
- IEC 60118-13:2019 Electroacoustics - Hearing aids - Part 13: Requirements and methods of measurement for electromagnetic immunity to mobile digital wireless devices
- IEC 60118-9: 2019 Electroacoustics - Hearing aids - Part 9: Methods of measurement of the performance characteristics of bone conduction hearing aids
- IEC 62133-2: 2021 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
2) Biocompatibility Testing
- ISO 10993-5: 2009, Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23: 2021, Biological evaluation of medical devices - Part 23: Tests for irritation
3) Software Verification and Validation
Page 5 of 6
{10}
Shenzhen Magnet Technology Co., Ltd
510(k)s – 510(k) Summary
Software documentation consistent with *Basic Documentation* of concern was submitted in this 510(k). System validation testing presented in this 510(k) demonstrated that all software requirement specifications are met and all software hazards have been mitigated to acceptable risk levels.
- IEC 62304: 2015 Medical device software - Software life cycle processes
4) Cybersecurity & Wireless Coexistence
- UL ANSI 2900-1 First Edition 2017 Standard for Safety, Standard for Software Cybersecurity Network-Connectable Products, Part 1: General Requirements
- IEEE ANSI USEMCSC C63.27-2021 American National Standard for Evaluation of Wireless Coexistence
- AAMI TIR69:2017/(R2020) Technical Information Report Risk management of radio-frequency wireless coexistence for medical devices and systems.
**VIII. Clinical Testing**
Not applicable.
**IX. Conclusions**
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is substantially equivalent to the legally marketed predicate device.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.