K261232 · Jochu Technology Co., Ltd. · INI · Jul 14, 2026 · Physical Medicine
Device Facts
Record ID
K261232
Device Name
IERO Mobility Scooter (IERO-01)
Applicant
Jochu Technology Co., Ltd.
Product Code
INI · Physical Medicine
Decision Date
Jul 14, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3800
Device Class
Class 2
Indications for Use
This device is intended for medical purposes to provide mobility to persons restricted to a seated position.
Device Story
IERO Mobility Scooter (IERO-01) is a motorized three-wheeled vehicle for adults restricted to a seated position. Device converts electrical energy from lead-acid batteries into mechanical motion via a motor and drive mechanism. User operates device using hand-controlled throttle and steering column. Features include seat belts, cushions, and onboard charger. Device is intended for independent operation by the user. Performance is verified through bench testing of speed, braking, stability, and energy consumption. Benefits include restored mobility for individuals with physical limitations.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via ISO 7176 series (stability, braking, energy consumption, dimensions, speed, obstacle climbing, fatigue strength, climatic tests) and ISO 16840-10 (ignition resistance). Biocompatibility confirmed via ISO 10993-5, -10, and -23. Electrical safety and EMC confirmed via ISO 7176-14 and -21.
Technological Characteristics
Steel structure with powder coating; knitted fabric/PU foam seating. Powered by 2x12V/50Ah lead-acid batteries. Motorized three-wheeled vehicle; dimensions 1360x625x1066mm; weight 110kg. Max speed 11.5 km/h; max load 150kg. IPX4 water resistance. Non-sterile. Software: PG, S-Drive S120A motor controller firmware.
Indications for Use
Indicated for adults restricted to a seated position who possess normal eyesight and hearing, and are physically and mentally capable of operating an electric mobility device.
Regulatory Classification
Identification
A motorized three-wheeled vehicle is a gasoline-fueled or battery-powered device intended for medical purposes that is used for outside transportation by disabled persons.
{1}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
July 14, 2026
Jochu Technology Co., Ltd.
Jimmy(Chihming) Wei
President
# 42, Kuangfu Rd., Hsin Chu Industrial Park,
Hukou Hsin Chu, Taiwan (R.O.C.)
Hsin Chu,
Taiwan
Re: K261232
Trade/Device Name: IERO Mobility Scooter (IERO-01)
Regulation Number: 21 CFR 890.3800
Regulation Name: Motorized Three-Wheeled Vehicle
Regulatory Class: Class II
Product Code: INI
Dated: April 14, 2026
Received: April 15, 2026
Dear Jimmy(Chihming) Wei:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261232 - Jimmy(Chihming) Wei
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K261232 - Jimmy(Chihming) Wei
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Digitally signed by
MARY S. KESZLER -S
Date: 2026.07.14
11:43:19 -04'00'
for Tushar Bansal, PhD
Acting Assistant Director
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261232 | ? |
| Please provide the device trade name(s). | | ? |
| IERO Mobility Scooter (IERO-01) | | |
| Please provide your Indications for Use below. | | ? |
| This device is intended for medical purposes to provide mobility to persons restricted to a seated position. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
# 510(k) Summary - K261232
## I. SUBMITTER
Device Submitter:
**JOCHU technology CO., LTD.**
**No. 42, Guangfu Rd., Hukou Township, Hsinchu County 303036, Taiwan (R.O.C.)**
Phone: 886 – 3 – 5981919 - 1600
Fax: 886 – 3 – 5983600
Contact Person: JIMMY(CHIHMING) WEI
Date Prepared: December 22, 2025
## II. DEVICE
Name of Device: IERO Mobility Scooter
Common or Usual Name: Electrical scooter
Classification Name: Motorized three-wheeled vehicle.
Regulatory Class: II
Product Code: INI
## III. PREDICATE DEVICE
KARMA Scooter KS-747.2, K041677
This predicate has not been subject to a design-related recall.
No reference devices were used in this submission.
## IV. DEVICE DESCRIPTION
Device design and function : The IERO Mobility scooter does not have any novel features since it is a mobility device, designed for adults who are restricted to a seated position but still have normal eyesight and hearing, and are physically and mentally capable of operating an electric mobility device and has similar specifications to other similar devices on the market, such as appearance and performance.
material used : Most structure made by steel with powder coating. Seat fabric, back fabric used by Knitted Fabrics.
physical properties : L1360mm/W625mm/H1066 mm, Total weight 110KG, Batteries (lead-acid), Speed 11.5 km/h, Min. braking distance 2.5m, Rated slope 12°, Max. climbable obstacle height 80 mm.
The associated accessories include:
**JOCHU technology CO., LTD. - K261232**
{6}
- Seat belts
- Cushion
- Back Cushion
- Batteries and chargers
# V. INDICATIONS FOR USE
This device is intended for medical purposes to provide mobility to persons restricted to a seated position.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Mobility Scooter is intended for medical purposes to provide mobility to persons restricted to a seated position. The subject and predicate devices are based on the following same technological elements:
- Principle of operation
- Charging Device
- Batteries
- Duration of contact
- Sterilization method
The following technological differences exist between the subject and predicate devices:
- overall dimension
- weight
- Speed
- Drive range
- water resistant
The major differences existing of the two Mobility Scooters are the different overall dimension, weight and Speed between the two devices. The differences are not safety aspect. So the new device is substantially equivalent to the predicate device in this aspect.
JOCHU technology CO., LTD. - K261232
{7}
# - Substantial Equivalence Comparison Table
| | New device | Predicate device |
| --- | --- | --- |
| **K number** | K261232 | K041677 |
| **Manufacturer** | Jochu Technology Co., Ltd. | Karma Medical Products CO., LTD. |
| **Brand and Product Items** | IERO Mobility scooter IERO-01 | Karma Scooter KS-747.2 |
| **Classification** | Class II | Class II |
| **Device picture** | | |
| **1. Overall description of the device design (Technical Characteristics)** | | **Comparison (Same/Similar/Different)** |
| **Indications for Use** | This device is intended for medical purposes to provide mobility to persons restricted to a seated position. | This device is intended for medical purposes to provide mobility to persons restricted to a seated position. |
| | | Same |
JOCHU technology CO., LTD. - K261232
{8}
| **Principle of operation** | Revolves around converting electrical energy from rechargeable batteries into mechanical motion through a motor and drive mechanism. The user controls the scooter's speed and direction using throttle controls and steering mechanisms, while safety features, braking systems, and battery management contribute to a secure and efficient mobility experience. | The movement of the Scooter is controlled by the rider who uses hand controls locals at the top of the steering column. The device can be disassembled for transport and is provided with an onboard battery charger. | Same |
| --- | --- | --- | --- |
| **Speed** | 11.5 km/h | 15 km/h | Similar to the predicate device. Minor deviation due to commercial consideration which would not affect the safety and performance (supported by testing conducted in accordance with **ISO 7176-4,6**). |
JOCHU technology CO., LTD. - K261232
{9}
| **Drive range** | 11.5 km/h: 26 km | 15 km/h: 42 km | Similar to the predicate device. Minor deviation due to commercial consideration which would not affect the safety and performance (supported by testing conducted in accordance with **ISO 7176-4**). |
| --- | --- | --- | --- |
| **Maximum Load** | Maximum payload: 150 kg | Maximum payload: 135 kg | Similar to the predicate device. Minor deviation due to commercial consideration which would not affect the safety and performance (supported by testing conducted in accordance with **ISO 7176-5**). |
| **Seat Dimensions** | Width: 460 mm Depth: 457.2 mm Height: 660 mm | Width: 430 mm Depth: 435 mm Height: 610 ~ 635 mm | Similar to the predicate device. Minor deviation due to commercial consideration which would not affect the safety and performance (supported by testing conducted in accordance with **ISO 7176-5 and ISO 7176-7**). |
JOCHU technology CO., LTD. - K261232
{10}
| **Minimum Braking distance** | Speed (11.5 km/h): 2600 mm | Speed (15 km/h): 3500 mm | Similar to the predicate device. Minor deviation due to commercial consideration which would not affect the safety and performance (supported by testing conducted in accordance with **ISO 7176-3 and ISO 7176-6**). |
| --- | --- | --- | --- |
| The new device and the predicate device have minor differences in speed, drive range, maximum load, seat dimensions, and minimum braking distance. However, these differences do not affect the performance and safety of the new device. The performance of the new device has been verified through safety and performance testing, the details are as follows: -Speed, drive range, maximum load, seat dimensions, and minimum braking distance: please refer to the testing conducted in accordance with **ISO 7176-4, ISO 7176-5, ISO 7176-6, and ISO 7176-7**. | | | Substantially Equivalent: **Yes** |
| **2. Materials (Biology Characteristics)** | | | **Comparison (Same/Similar/ Different)** |
| **For similar kind and duration of contact** | Contact the surface of body / long term (> 30 days) | Contact the surface of body / long term (> 30 days) | Same |
| **Sterilization method** | Non-sterile | Non-sterile | Same |
| **Single use/ reusable** | Reusable | Reusable | Same |
| **Cleanability/ Infection Control** | IPX4 | This product is not guaranteed to be water resistant. | Different. This difference would not affect the safety and |
JOCHU technology CO., LTD. - K261232
{11}
| | | | performance (supported by testing conducted in accordance with **ISO 7176-9**). |
| --- | --- | --- | --- |
| **Biocompatibility and chemical characteristics** | Conduct biocompatibility evaluation according to ISO 10993-1:2018, ISO 10993-5: 2009, ISO 10993-10: 2021, and ISO 10993-23:2021. | N/A | Predicate devices do not reveal their biocompatibility information. The safety regarding biocompatibility and chemical characteristics is supported by the biocompatibility evaluation conducted in accordance with **ISO 10993-1, ISO 10993-5, ISO 10993-10, and ISO 10993-23**. |
| The predicate device does not provide biocompatibility information. However, the subject device is intended for use with clothing and involves long-term contact. Safety with respect to biocompatibility and chemical characteristics is supported by the biocompatibility evaluation reports conducted in accordance with ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-10:2021, and ISO 10993-23:2021. Based on these evaluations, the subject device is considered as safe and as effective as the predicate device. | | | Substantially Equivalent: **Yes** |
| **3.Energy sources** | | | **Comparison** (Same/Similar/Different) |
| **Charging Device** | Output current :6 A ± 8 % Output voltage: 24 V nominal | Output current :6 A Output voltage: 24 VDC | Same (supported by testing conducted in accordance with **IEC 60601-1-2 and ISO 7176-14**). |
| **Batteries** | 2 x 12 V/50 Ah leak proof/AGM | 2 x 12 V/50 Ah leak proof/AGM | Same (supported by testing conducted in accordance with **ISO 7176-14**). |
JOCHU technology CO., LTD. - K261232
{12}
| The energy sources of the new device and the predicate device are the same. The electrical safety and electromagnetic compatibility of the new device can be supported by the testing, the details are as follows: - Charging and batteries: please refer to the testing conducted in accordance with IEC 60601-1-2. | | | Substantially Equivalent: **Yes** |
| --- | --- | --- | --- |
| **4. Other key technological features** | | | **Comparison** (Same/Similar/ Different) |
| **software features** | PG,S-Drive S120A motor controller firmware | N/A | Predicate devices do not reveal their software information. The safety regarding software are supported by the evaluation of EN62304 Gap Analysis |
| **clinical purpose** | This device is intended for medical purposes to provide mobility to persons restricted to a seated position. | This device is intended for medical purposes to provide mobility to persons restricted to a seated position. | Same |
| **Software verification and validation Report and EN62304 Gap Analysis.** The clinical purpose of the predicate device - 'Karma Scooter KS-747.2' is consistent with the clinical purpose of the new device. According to MDCG 2020-5 Clinical Evaluation - Equivalence published by the European Commission, similar relevant critical performance in view of the expected clinical effect for a specific intended purpose is acceptable. | | | Substantially Equivalent: **Yes** |
| **Summary** Scientific justification of similar but not the same features in the safety and clinical performance of the devices are as below: 1. Charging and batteries: please refer to the testing conducted in accordance with ISO 7176-14,21. 2. Speed, drive range, maximum load, and seat dimensions: please refer to the testing conducted in accordance with ISO 7176-4, 5, 6, 7. | | | |
JOCHU technology CO., LTD. - K261232
{13}
3. Biocompatibility and chemical characteristics: please refer to biocompatibility evaluation report of 10993-1:2018, ISO 10993-5: 2009 and ISO 10993-10: 2021, biocompatibility evaluation report.
4. Clinical purpose:
- Karma Scooter KS-747.2 Instruction for use:
https://www.karmamedical.com/featured_item/ks-747-2/
- IERO Mobility scooter Instruction for use:
The differences between the new device and the predicate device are minor deviations that would not affect the safety and performance of the device. For some technical parameters, and speed and the biocompatibility and chemical characteristic consideration, the predicate device did not reveal the relevant information. However, according to the performance test report and biocompatibility evaluation report of ISO 10993-1:2018, 10993-5: 2009, ISO 10993-10: 2021 and ISO 10993-23:2021, the subject device is believed to be as safe and as effective as the predicate device under its clinical condition. Therefore, there are no significant differences in the safety and performance between the subject device and the predicate device.
JOCHU technology CO., LTD. - K261232
{14}
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Non-Clinical Testing Summary Table
| # | Test Standard | Test standard | Result |
| --- | --- | --- | --- |
| Biocompatibility testing | | | |
| 1 | In vitro cytotoxicity | ISO 10993-5:2009 | Pass |
| 2 | Intracutaneous Irritation Study in White Rabbits | ISO 10993-23:2021/Amd 1:2025 | Pass |
| 3 | Guinea Pig Skin Sensitization Study | ISO 10993-10:2021 | Pass |
| Electrical safety and electromagnetic compatibility | | | |
| 1 | Wheelchairs – Part 14: Power and control systems for electrically powered wheelchairs and scooters — Requirements and test methods | ISO 7176-14:2008 | Pass |
| 2 | Requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and motorized scooters | ISO 7176-21:2009 | Pass |
| Software Verification and Validation Testing | | | |
| 1 | Medical device software — Software life cycle processes | IEC 62304:2006 | Pass |
| Safety and Performance Testing | | | |
| 1 | Wheelchairs – Part 1: Determination of static stability | ISO 7176-1:2014 | Pass |
| 2 | Wheelchairs – Part 2: Determination of dynamic stability of electrically powered wheelchairs | ISO 7176-2:2017 | Pass |
| 3 | Wheelchairs – Part 3: Determination of effectiveness of brakes | ISO 7176-3:2012 | Pass |
| 4 | Wheelchairs – Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range | ISO 7176-4:2008 | Pass |
| 5 | Wheelchairs – Part 5: Determination of dimensions, mass, and maneuvering space | ISO 7176-5:2008 | Pass |
| 6 | Wheelchairs – Part 6: Determination of maximum speed of electrically powered wheelchairs | ISO 7176-6:2018 | Pass |
| 7 | Wheelchairs – Part 7: Measurement of seating and wheel dimensions | ISO 7176-7:1998 | Pass |
| 8 | Wheelchairs – Part 8: Requirements and test methods for static, impact and | ISO 7176-8:2014 | Pass |
JOCHU technology CO., LTD. - K261232
{15}
| | fatigue strengths | | |
| --- | --- | --- | --- |
| 9 | Wheelchairs – Part 9: Climatic tests for electric wheelchairs | ISO 7176-9:2009 | Pass |
| 10 | Wheelchairs – Part 10: Determination of obstacle-climbing ability of electrically powered wheelchairs | ISO 7176-10:2008 | Pass |
| 11 | Wheelchairs – Part 11: Test dummies | ISO 7176-11:2012 | Pass |
| 12 | Wheelchairs – Part 13: Determination of coefficient of friction of test surfaces | ISO 7176-13:1989 | Pass |
| 13 | Wheelchairs – Part 15: Requirements for information disclosure, documentation, and labelling | ISO 7176-15:1996 | Pass |
| 14 | ISO 16840-10:2021 Wheelchair seating —Part 10: Resistance to ignition of postural support devices — Requirements and test method | ISO 16840-10:2021 | Pass |
### A. Biocompatibility testing
The safety regarding biocompatibility and chemical characteristics are supported by the biocompatibility evaluation report of ISO 10993-1:2018, ISO 10993-5: 2009, ISO 10993-10: 2021, and ISO 10993-23:2021.
Biocompatibility testing:
•ISO 10993-1:2018
•ISO 10993-5:2009
Knitted Fabrics,
PU Foam,
•ISO 10993-10:2021
Knitted Fabrics,
PU Foam, Report
•ISO 10993-23:2021
Knitted Fabrics,
PU Foam,
### B. Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the IERO Mobility Scooter. The system complies with ISO 7176-14:2008 and ISO 7176-21:2009 for EMC.
### C. Software Verification and Validation Testing
Software verification and validation testing were conducted per IEC 62304:2006 and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "Minor" level of concern, the risk of this software has been assessed as level A, so there is no risk that is not within the scope under normal circumstances.
### D. Safety and Performance Testing
JOCHU technology CO., LTD. - K261232
{16}
The safety and performance of the device per ISO 7176-1:2014, ISO 7176-2:2017, ISO 7176-3:2012, ISO 7176-4:2008, ISO 7176-5:2008, ISO 7176-6:2018, ISO 7176-7:1998, ISO 7176-8:2014, ISO 7176-9:2009, ISO 7176-10:2008, ISO 7176-11:2012, ISO 7176-13:1989, and ISO 7176-15:1996, ISO 16840-10:2021. The device has passed all the criteria.
#### **E. Medical Device Recalls**
There are no significant serious adverse events that occurred.
### **VIII. CONCLUSIONS**
The differences between the new device and the predicate device are minor deviations that would not affect the safety and performance of the device. For some technical parameters, and speed and the biocompatibility and chemical characteristic consideration, the predicate device did not reveal the relevant information. However, according to the performance test report and biocompatibility evaluation report of ISO 10993-1:2018, 10993-5: 2009, ISO 10993-10: 2021, and ISO 10993-23:2021 the subject device is believed to be as safe and as effective as the predicate device under its clinical condition. Therefore, there are no significant differences in the safety and performance between the subject device and the predicate device.
JOCHU technology CO., LTD. - K261232
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.