K261225 · Cascination AG · JAK · Sep 29, 2026 · Radiology
Device Facts
Record ID
K261225
Device Name
CAS-One IR
Applicant
Cascination AG
Product Code
JAK · Radiology
Decision Date
Sep 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
AI/ML, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K261225 · Sep 29, 2026
CAS-One IR
Cascination AG
Representative clinical datasets from multiple clinical sites
Clinical datasets were used to validate the performance of the device's organ and lesion segmentation algorithms by comparing automated outputs against clinical ground truth.
Patients undergoing CT-guided procedures; Number of Sites: Multiple clinical sites
Not applicable for this study
Dice similarity coefficients for organ and lesion segmentation
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Organ segmentation
Machine learning algorithm
Dice >= 0.60 to 0.90 depending on structure
Dice 0.67 to 0.97 depending on structure
Representative clinical datasets acquired across multiple clinical sites
—
—
—
Lesion segmentation
Machine learning algorithm
Dice >= 0.80
Dice 0.86
Representative clinical datasets acquired across multiple clinical sites
—
—
—
Effective treatment volume assessment
Machine learning algorithm
Dice >= 0.80
Dice 0.86
Representative clinical datasets acquired across multiple clinical sites
—
—
—
Indications for Use
CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification of instrument position and performance during Computed Tomography (CT) guided procedures. In planning, the desired needle configuration and performance is defined relative to the target anatomy. In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored. In verification, the achieved instrument configuration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data. CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures. CAS-One IR is intended to be used for patients older than 18 years and eligible for CT-guided percutaneous interventions.
Device Story
CAS-One IR is a stereotactic image-guided system for CT-guided percutaneous interventions. It consists of the CAS-One IR Software (parent device) and hardware accessories (platform, baseplate, cushion, aiming device, guide arm, pointer, tray). The system takes CT or MRI image datasets as input. During planning, the user defines needle configuration relative to target anatomy. During navigation, the system displays instrument position relative to the patient, provides alignment guidance, and monitors respiratory levels using stereophotogrammetric optical tracking. During verification, the system overlays pre- and post-treatment images to confirm instrument placement and performance. The device is used in clinical settings by trained physicians. It aids in precise needle placement, potentially improving procedural accuracy and patient outcomes. The system includes automated segmentation of various organs and lesions to assist in planning and assessment. Output is visualized on dual touch screens. The platform supports data management via USB/LAN and utilizes a dedicated tracking camera for real-time instrument pose determination.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing: software verification and validation, segmentation algorithm validation (Dice coefficients reported for 12 structures), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and summative usability testing with interventional oncologists. Navigation accuracy was validated using a phantom model, demonstrating non-inferiority of the new tracking camera compared to the predicate.
Technological Characteristics
System comprises software running on a mobile platform with stereophotogrammetric optical tracking (NDI Polaris Lyra). Materials include stainless steel, titanium, PEEK, carbon fiber, and polymethacrylimide foam. Connectivity via Ethernet/USB. Software uses machine learning for organ/lesion segmentation. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Platform runs on Windows 10/11 IoT Enterprise LTSC.
Indications for Use
Indicated for patients >18 years old eligible for CT-guided percutaneous interventions using rigid straight instruments (needles/probes). Used by physicians trained for CT procedures.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
September 29, 2026
Cascination AG
% Philip Frost
Regulatory and Clinical Affairs Specialist
Steigerhubelstrasse 3
BERN, 3008
SWITZERLAND
Re: K261225
Trade/Device Name: CAS-One IR
Regulation Number: 21 CFR 892.1750
Regulation Name: Computed Tomography X-Ray System
Regulatory Class: Class II
Product Code: JAK
Dated: August 28, 2026
Received: August 28, 2026
Dear Philip Frost:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K261225 - Philip Frost
Page 2
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K261225 - Philip Frost
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
Diagnostic X-ray Systems Team
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261225 | ? |
| Please provide the device trade name(s). | | ? |
| CAS-One IR | | |
| Please provide your Indications for Use below. | | ? |
| CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification of instrument position and performance during Computed Tomography (CT) guided procedures. In planning, the desired needle configuration and performance is defined relative to the target anatomy. In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored. In verification, the achieved instrument configuration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data. CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures. CAS-One IR is intended to be used for patients older than 18 years and eligible for CT-guided percutaneous interventions. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
CASCINATION+
# 510(K) SUMMARY – K261225
This summary of 510(k) substantial equivalence information is submitted in accordance with the requirements of 21 CFR §807.92.
## I. SUBMITTER
Manufacturer: CASCINATION AG
Steigerhubelstrasse 3
3008 Bern
Switzerland
Tel: +41 31 306 26 26
Contact Person: Philip Frost
Regulatory and Clinical Affairs Specialist
Date Prepared: September 25, 2026
{5}
510(k) Summary – K261225
Page: 2/18
## II. SUBJECT DEVICE
| Device Name: | CAS-One IR |
| --- | --- |
| Parent Device: | CAS-One IR Software |
| Accessories: | CAS-One IR Platform A, CAS-One IR Baseplate A, CAS-One IR Cushion A, CAS-One IR Accessory Tray A, CAS-One IR Aiming Device A, CAS-One IR Guide Arm A, CAS-One IR Pointer A, CAS-One IR Consumable Set General |
| Common Name: | System, X-Ray, Tomography, Computed |
| Classification Name: | Computed Tomography X-Ray System |
| Regulation: | 892.1750 |
| Regulatory Class: | Class II |
| Product Code: | JAK |
CAS-One IR is an updated version of the Predicate Device (K232022) and is an image-guided system for CT-guided percutaneous interventions. The medical functionality (planning, guidance, and verification) is provided by the Parent Device, CAS-One IR Software. The Accessories support the Parent Device in achieving this functionality. The CAS-One IR Software and the Accessories below are separate finished devices, each with their own labeling (see Device Description).
| Name | REF |
| --- | --- |
| CAS-One IR Software | 20016 |
| CAS-One IR Platform A | 10006 |
| CAS-One IR Baseplate A | 30238 |
| CAS-One IR Cushion A | 30128 |
| CAS-One IR Accessory Tray A | 30007 |
| CAS-One IR Aiming Device A | 30085 |
| CAS-One IR Guide Arm A | 30026 |
| CAS-One IR Pointer A | 30027 |
| CAS-One IR Consumable Set General | 50002 |
The overall intended use, functionality, use context, and clinical workflow remain unchanged relative to the Predicate Device.
CASCINATION AG
{6}
510(k) Summary – K261225
Page: 3/18
### III. PREDICATE DEVICE
Primary Predicate Device
| Company: | CASCINATION AG Steigerhubelstrasse 3 CH-3008 Bern Switzerland |
| --- | --- |
| Device Name: | CAS-One IR |
| 510(k) number: | K232022 |
| Classification Name: | Computed tomography x-ray system |
| Regulation: | 892.1750 |
| Common Name: | System, X-Ray, Tomography, Computed |
| Regulatory Class: | Class II |
| Product Code: | JAK |
### IV. INDICATIONS FOR USE
CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification of instrument position and performance during Computed Tomography (CT) guided procedures.
In planning, the desired needle configuration and performance is defined relative to the target anatomy.
In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored.
In verification, the achieved instrument configuration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data.
CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures.
CAS-One IR is intended to be used for patients older than 18 years and eligible for CT-guided percutaneous interventions.
### V. DEVICE DESCRIPTION
CAS-One IR comprises the CAS-One IR Software (Parent Device), and eight Accessories. All Accessories and components are intended to be used together as part of the CAS-One IR procedure, and none performs a standalone medical function independent of the CAS-One IR Software.
The CAS-One IR Software is a medical device software from which the medical purpose of the CAS-One IR procedure is derived, while the hardware Accessories enable tracking, positioning, instrument alignment, and patient stabilization, and consumable Accessories support the procedural workflow.
CASCINATION AG
{7}
510(k) Summary – K261225
Page: 4/18
Each Accessory listed in Table 1 is a finished device, individually labeled, intended to support, supplement, and/or augment the performance of the Parent Device, the CAS-One IR Software, during image-guided, percutaneous interventions.
None of the Accessories has a medical purpose independent of the CAS-One IR Software.
Table 1
Accessories for the CAS-One IR Software
| Accessory Name | Description | Reference Image |
| --- | --- | --- |
| CAS-One IR Platform A: | The CAS-One IR Platform A provides computing, data storage and exchange, display, user input, and stereophotogrammetric (optical) position sensing; the CAS-One IR Software runs on the Platform. |  |
| CAS-One IR Baseplate A: | The CAS-One IR Baseplate A provides interfaces for patient immobilization and for attachment of instruments. |  |
| CAS-One IR Cushion A: | The CAS-One IR Cushion A immobilizes the patient. |  |
CASCINATION AG
{8}
510(k) Summary – K261225
Page: 5/18
| Accessory Name | Description | Reference Image |
| --- | --- | --- |
| CAS-One IR Accessory Tray A: | The CAS-One IR Accessory Tray contains and holds CASCINATION instruments during cleaning, disinfection, steam sterilization, storage, and aseptic presentation. |  |
| CAS-One IR Aiming Device A: | The CAS-One IR Aiming Device A adjusts the orientation of an instrument's trajectory and provides physical guidance for instruments. |  |
| CAS-One IR Guide Arm A: | The CAS-One IR Guide Arm A holds and stabilizes instruments. |  |
| CAS-One IR Pointer A: | The CAS-One IR Pointer A is an optically tracked tool that enables determination of its pose for alignment with the planned trajectory. |  |
| CAS-One IR Consumable Set General: | The CAS-One IR Consumable Set General contains consumable items supporting the CAS-One IR procedure workflow. |  |
CASCINATION AG
{9}
510(k) Summary – K261225
Page: 6/18
## VI. SUBSTANTIAL EQUIVALENCE
The following characteristics were compared between the Subject Device and the Predicate Device in order to demonstrate substantial equivalence (Table 2).
Table 2
Summary of the Substantial Equivalence (SE) Comparison to Predicate Device
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| **Intended Use** | | | |
| Indications for Use Statement | CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification of instrument position and performance during Computed Tomography (CT) guided procedures. In planning, the desired needle configuration and performance is defined relative to the target anatomy. In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored. In verification, the achieved instrument configuration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data. CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures. CAS-One IR is intended to be used for patients older than 18 years and eligible for CT-guided percutaneous interventions. | CAS-One IR is a user controlled, stereotactic accessory intended to assist in planning, navigation and manual advancement of one or more instruments, as well as in verification of instrument position and performance during Computed Tomography (CT) guided procedures. In planning, the desired needle configuration and performance is defined relative to the target anatomy. In navigation, the instrument position is displayed relative to the patient and guidance for needle alignment is provided while respiratory levels are monitored. In verification, the achieved instrument configuration and performance are displayed relative to the previously defined plan through an overlay of the pre- and post- treatment image data. CAS-One IR is indicated for use with rigid straight instruments such as needles and probes used in CT guided interventional procedures performed by physicians trained for CT procedures. CAS-One IR is intended to be used for patients older than 18 years and eligible for CT-guided percutaneous interventions. | **Same** No changes. |
CASCINATION AG
{10}
510(k) Summary – K261225
Page: 7/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| User | Physicians trained for interventional CT procedures | Physicians trained for interventional CT procedures | **Same** No changes. |
| **Device Structure** | | | |
| Device Structure | CAS-One IR, consisting of the CAS-One IR Software, its hardware components and consumables used exclusively together for the CAS-One IR procedure. | CAS-One IR, consisting of the CAS-One IR Software (Parent Device) and eight separately labeled and packaged Accessories (CAS-One IR Platform A, Baseplate A, Cushion A, Accessory Tray A, Aiming Device A, Guide Arm A, Pointer A, and Consumable Set General) used exclusively together for the CAS-One IR procedure. | **Substantially equivalent** Same software, hardware, and consumables used together for the same procedure. Only the regulatory structure changed: hardware and consumables are now separately labeled Accessories to the CAS- One IR Software, with no independent medical purpose. No impact on intended use, functionality, safety, or performance. |
| **Technical Characteristics** | | | |
| **CAS-One IR Software** | | | |
| Compatible Operating System(s) | Windows 10 IoT Enterprise LTSC. | Windows 10 IoT Enterprise LTSC, or Windows 11 IoT Enterprise LTSC. | **Substantially equivalent** Upgrade within the Microsoft Windows family. The change does not affect the intended use or fundamental functionality. Cybersecurity of the subject operating system was re- assessed and verified for Windows 11 IoT Enterprise LTSC (see Performance Data, Cybersecurity). |
| Image datasets | DICOM file format: • CT or MRI image modality | DICOM file format: • CT or MRI image modality | **Same** No changes. |
| Data Management | Functionality to import image datasets from USB and LAN. | Functionality to import image datasets from USB and LAN. | **Same** No changes. |
CASCINATION AG
{11}
510(k) Summary – K261225
Page: 8/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| DICOM Viewer | ○ Functionality to view, pan, zoom, adjust brightness and intensity, and carry out measurements in image data and to overlay image data ○ Standard perspectives (axial, coronal, sagittal) ○ Oblique perspectives aligned with the planned trajectory ○ Toggle (show/hide) of individual annotations and trajectories | ○ Functionality to view, pan, zoom, adjust brightness and intensity, and carry out measurements in image data and to overlay image data ○ Standard perspectives (axial, coronal, sagittal) ○ Oblique perspectives aligned with the planned trajectory ○ Toggle (show/hide) of individual annotations and trajectories ○ Adjustable oblique MPR perspectives ○ Toggle (show/hide) of all annotations and trajectories | **Substantially equivalent** The adjustable oblique MPR perspectives extend the oblique viewing already available in the Predicate Device to the MPR views. The show/hide-all toggle combines the existing individual toggles into a single action. Both are established DICOM-viewing functions that display the same image data without modification. They introduce no new clinical output or interpretive task and have no impact on core functionality, safety, or performance. |
| Planning and Fusion | ○ Trajectory Planning ○ Single / Multiple needles ○ Display of manufacturer ablation parameters ○ Point-based image fusion CT to CT or MR ○ Real time visualization of patient markers | ○ Trajectory Planning ○ Single / Multiple needles ○ Display of manufacturer ablation parameters ○ Point-based image fusion CT to CT or MR ○ Real-time and static visualization of patient markers ○ Patient-Marker Assessment Support | **Substantially equivalent** Minor visualization and workflow refinements; no impact on core functionality, safety, or performance. |
| Organ segmentation | Segmentation of: ○ liver tissue, ○ portal vein, ○ hepatic vein, ○ kidney tissue, ○ lung tissue, ○ lung airway, | Segmentation of: ○ liver tissue, ○ portal vein, ○ hepatic vein, ○ kidney tissue, ○ lung tissue, ○ lung airway, ○ gallbladder, ○ heart, ○ spleen, ○ digestive system, ○ bone and cartilage | **Substantially equivalent** The Subject Device uses the same segmentation algorithm and was validated using the same testing method as the Predicate Device. The technological change is that additional organs are being segmented as an output. |
CASCINATION AG
{12}
510(k) Summary – K261225
Page: 9/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| Lesion segmentation | Segmentation of liver lesions | Segmentation of liver lesions Sub-capsular lesion support for liver | **Substantially equivalent** Minor segmentation improvement using the same machine learning algorithm and development process - no new functionality nor modification of the nature of the output. Sub-capsular lesion support refines margin representation using existing segmentation outputs without modifying underlying algorithms. |
| Placement | Instrument placement support. | Instrument placement support. | **Same** No changes. |
| Verification | Instrument Detection. | Instrument Detection. | **Same** No changes. |
| Assessment | Manual review of planning and validation images (toggle visibility) AblaSure® Assessment: ○ Effective treatment volume for the liver ○ Tumor coverage for the liver ○ Margin coverage for the liver | Manual review of planning and validation images (toggle visibility) AblaSure® Assessment: ○ Effective treatment volume for the liver ○ Tumor coverage for the liver ○ Margin coverage for the liver Re-ablation workflow for the liver | **Substantially equivalent** Minor segmentation improvement using the same machine learning algorithm and development process - no new functionality nor modification of the nature of the output. Re-ablation for liver introduces no new capability. The same result was already achievable with the Predicate Device by starting a new case from the existing image data and planning a new trajectory. The re-ablation workflow only carries the existing data and plan forward directly, removing the manual re-setup steps. It produces no new output and does not change the intended use, functionality, safety, or performance. |
CASCINATION AG
{13}
510(k) Summary – K261225
Page: 10/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| **Accessories** | | | |
| **CAS-One IR Platform** | | | |
| General | **Dimensions:** - Height: 2.0 m - Width: 0.7 m - Depth: 1.2 m**Weight:** 85 kg **Type:** Mobile equipment **Operating mode:** Continuous use **Insulation class:** 1 **Ingress protection:** IP20 | **Dimensions:** - Height: 2.0 m - Width: 0.7 m - Depth: 1.2 m**Weight:** 85 kg **Type:** Mobile equipment **Operating mode:** Continuous use **Insulation class:** 1 **Ingress protection:** IP20 | **Same** No changes. |
| Installed Operating system | Windows 10 IoT Enterprise LTSC | Windows 11 IoT Enterprise LTSC | **Substantially equivalent** - Same OS family and edition: Windows NT, IoT Enterprise LTSC. - Same architecture: x64 (64-bit). - Same application binary interface (ABI): Win32 API, PE executable format and x64 calling convention. The CAS-One IR 4.3.0 binaries run unchanged on both. - Same servicing model: LTSC, with long-term, security-only updates and no feature updates. - Differences limited to: a newer OS build, updated drivers, and equal or stronger security defaults. |
| Power supply | **Power:** 600 W **Input voltage:** 85-264 **Input frequency:** 47-440 Hz **Certifications:** - IEC60601-1 3rd edition, **MTBF:** > 320,307 hours | **Power:** 600 W **Input voltage:** 85-264 **Input frequency:** 47-440 Hz **Certifications:** - IEC60601-1 3rd edition **MTBF:** > 400,000 hours | **Substantially equivalent** Integration of new AC/DC module led to new IEC 60601-1 and IEC 60601-1-2 testing (according CB Scheme by external accredited lab) No impact on core functionality otherwise. |
| I/O Ports | 1 x Ethernet 1Gbit/s 1 x USB-A(USB3.0) | 1 x Ethernet 1Gbit/s 1 x USB-A(USB3.2) | **Substantially equivalent** USB 3.2 is backwards compatible, Ethernet speed is the same. |
CASCINATION AG
{14}
510(k) Summary – K261225
Page: 11/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| Camera | **Type:** Stereophotogrammetric (optical) position sensing **Model:** NDI Polaris Vicra **Volumetric accuracy (RMS / 95% CI):** 0.25 mm / 0.5 mm **Max. frame rate:** 20 Hz **Measurement volume:** depth 1420 mm, width 938 mm, height 887 mm **Interface / power:** USB (serial API) **Weight:** 0.8 kg | **Type:** Stereophotogrammetric (optical) position sensing **Model:** NDI Polaris Lyra Position Sensor NB **Volumetric accuracy (RMS / 95% CI):** 0.20 mm / 0.40 mm **Max. frame rate:** 60 Hz **Measurement volume:** depth 1420 mm, width 1190 mm, height 1080 mm **Interface / power:** Ethernet with PoE (internal PoE injector) **Weight:** 0.8 kg | **Substantially equivalent** Different tracking-camera model with higher frame rate, larger measurement volume and improved volumetric accuracy. Navigation accuracy was re-assessed for the subject camera and shown non-inferior to the predicate camera. |
| Touch screens | **Model:** 2 × ACL OR-MD 24 **Resolution:** 1920 × 1200 | **Model:** 2 × ACL OR-MD 24 **Resolution:** 1920 × 1200 | **Substantially equivalent** Touch screens upgraded to current-generation models with the same display and touch-input functions. Platform re-tested to IEC 60601-1 and IEC 60601-1-2 under the CB Scheme (see Performance Data). No impact on intended use, functionality, safety, or performance. |
| Control box | **Mainboard platform:** Mini-ITX, Intel Q87, LGA1150 **Memory support:** up to 16 GB DDR3/DDR3L 1333/1600 **TPM:** TPM 1.2 (optional) **CPU cores / threads:** 4 / 8 **CPU clock:** 3.1 GHz base / 3.9 GHz turbo **CPU cache / TDP:** 8 MB L3 / 65 W **GPU CUDA cores / memory:** 1280 / 5 GB GDDR5X **GPU bandwidth / interface:** up to 200 GB/s / PCIe 3.0 x16 **GPU max power:** 75 W **RAM:** 16 GB DDR3 **Storage:** 1 TB mSATA SSD (SATA III 6 Gb/s) | **Mainboard platform:** Mini-ITX, LGA1700, R680E/Q670E/H610E chipset **Memory support:** up to 64 GB DDR4-3200, ECC or non-ECC **TPM:** TPM 2.0 (onboard) **CPU cores / threads:** 12 (8P + 4E) / 20 **CPU clock:** P-cores 1.4-4.6 GHz, E-cores 1.0-3.4 GHz **CPU cache / TDP:** 25 MB Smart Cache / 35 W **GPU CUDA cores / memory:** 2816 / 16 GB GDDR6 ECC **GPU bandwidth / interface:** 224 GB/s / PCIe 4.0 x8 **GPU max power:** 70 W **RAM:** 32 GB DDR4 **Storage:** 256 GB (OS) + 2 TB RAID 1 (data and application) | **Substantially equivalent** Computing components upgraded to current-generation models with equal or higher performance. Platform re-tested to IEC 60601-1 and IEC 60601-1-2 under the CB Scheme (see Performance Data). No impact on intended use, functionality, safety, or performance. |
CASCINATION AG
{15}
510(k) Summary – K261225
Page: 12/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| Frame and housing | Powder Coated Aluminium Frame 4x Caster with brakes 2x Mechanical monitor arms 1x Mechanical camera arm | Powder Coated Aluminium Frame 4x Caster with brakes 2x Mechanical monitor arms 1x Mechanical camera arm | **Substantially equivalent** Housing color change only; no impact on intended use, functionality, safety, or performance. |
| **CAS-One IR Baseplate A** | | | |
| CAS-One IR Baseplate A | Provides interfaces for patient immobilization and for attachment of instruments. **Dimensions:** ~1500 × 500 × 20 mm **Weight:** 7 kg **Material:** Polymethacrylimide foam, carbon fiber, glass fiber reinforced polyamide | Provides interfaces for patient immobilization and for attachment of instruments. **Dimensions:** ~1500 × 500 × 20 mm **Weight:** 7 kg **Material:** Polymethacrylimide foam, carbon fiber, glass fiber reinforced polyamide | **Substantially equivalent** Label change. No impact on intended use, functionality, safety, or performance. The device has no patient contact. |
| Adapter | **Material:** Stainless steel | **Material:** Stainless steel | **Same** No change. |
| **CAS-One IR Cushion A** | | | |
| CAS-One IR Cushion A | Cushion that conforms to the patient and, once air is evacuated, maintains a rigid shape to immobilize the patient. **Dimensions:** ~2000 × 700 × 70mm **Weight:** ~4.5 kg **Materials:** sealed airtight bag with plastic pellets; valve; loop tape | Cushion that conforms to the patient and, once air is evacuated, maintains a rigid shape to immobilize the patient. **Dimensions:** ~2000 × 700 × 70mm **Weight:** ~4.5 kg **Materials:** sealed airtight bag with plastic pellets; valve; loop tape | **Substantially equivalent** Label change from system level to device level. No impact on intended use, functionality, safety, or performance. The device has no patient contact. |
| **CAS-One IR Accessory Tray A** | | | |
| CAS-One IR Accessory Tray A | Reusable tray that holds CASCINATION instruments in place during cleaning, disinfection, steam sterilization, storage, and aseptic presentation. **Dimensions:** 406 × 253 × 70mm **Materials:** stainless steel, silicone, polypropylene, PPSU | Reusable tray that holds CASCINATION instruments in place during cleaning, disinfection, steam sterilization, storage, and aseptic presentation. **Dimensions:** 390 × 230 × 93mm **Materials:** stainless steel, silicone, polypropylene, PPSU | **Substantially equivalent** No impact on intended use, functionality, safety, or performance. The device has no patient contact. |
CASCINATION AG
{16}
510(k) Summary – K261225
Page: 13/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| **CAS-One IR Aiming Device A** | | | |
| CAS-One IR Aiming Device A | Mechanical guide that holds the instrument and allows fine adjustment of its trajectory **Arm Connection:** Rigid connection to the guide arm **Material:** Stainless steel, titanium alloy, PEEK **Adjustment axes:** Two rotational axes with zero-position markings | Mechanical guide that holds the instrument and allows fine adjustment of its trajectory **Arm Connection:** Rigid connection to the guide arm **Material:** Stainless steel, titanium alloy, PEEK **Adjustment axes:** Two rotational axes with zero-position markings | **Substantially equivalent** Label change from system level to device level. No impact on intended use, functionality, safety, or performance. |
| **CAS-One IR Guide Arm A** | | | |
| CAS-One IR Guide Arm A | Articulated arm with an optional extender for additional reach that, once its joints are locked, holds a stable position. **Dimensions (Arm / Extender):** ~436 × 35 × 119 mm / ~183 × 25 × 10 mm **Weight (Arm / Extender):** ~2 kg / ~0.15 kg **Joints:** 3, articulated, lockable **Materials:** stainless steel, aluminium / stainless steel | Articulated arm with an optional extender for additional reach that, once its joints are locked, holds a stable position. **Dimensions (Arm / Extender):** ~460 × 63 × 90 mm / ~185 × 25 × 28 mm **Weight (Arm / Extender):** ~1.4 kg / ~0.16 kg **Joints:** 3, articulated, lockable **Materials:** stainless steel, aluminium, PEEK / stainless steel, PEEK | **Substantially equivalent** Same operating principle, 3 lockable joints, central locking handle, interfaces, and extender length as the Predicate Device. Minor design changes only: slightly longer arm, redesigned handle, improved screw retention, and added PEEK guiding sleeves. No impact on intended use, functionality, safety, or performance. The device has no patient contact. |
| **CAS-One IR Pointer A** | | | |
| CAS-One IR Pointer A | Two-component tracked pointer whose pose is determined for alignment with the planned trajectory. **Dimension:** ~116 × 84 × 25mm **Raw Material:** titanium/ stainless steel | Single-component tracked pointer whose pose is determined for alignment with the planned trajectory. **Dimension:** ~116 × 84 × 25mm **Raw Material:** stainless steel | **Substantially equivalent** Same dimensions and marker geometry. Changes are limited to a single-component design instead of two components and a marker array material change from titanium to stainless steel. No impact on intended use, functionality, safety, or performance. |
| **CAS-One IR Consumable Set General** | | | |
| CAS-One IR Consumable Set General | Consumable components supporting the procedure | Consumable components supporting the procedure | **Same** No changes. |
CASCINATION AG
{17}
510(k) Summary – K261225
Page: 14/18
| Item | Predicate Device (K232022) | Subject Device (K261225) | SE Conclusion |
| --- | --- | --- | --- |
| **Performance Testing** | | | |
| Performance Testing | ○ Software verification and validation (“moderate” level of concern) ○ Internal Test Standards ○ Segmentation algorithm verification and validation ○ Electrical safety (IEC 60601-1) ○ Electromagnetic compatibility (IEC 60601-1-2) ○ Usability validation | ○ Software verification and validation (“moderate” level of concern) ○ Internal Test Standards ○ Segmentation algorithm verification and validation ○ Electrical safety (IEC 60601-1) ○ Electromagnetic compatibility (IEC 60601-1-2) ○ Usability validation ○ Hardware verification and validation (including compatibility, mechanical, environmental, and transport testing) | **Substantially equivalent** Verification and validation activities were performed in accordance with applicable standards and design control processes and demonstrate safety and effectiveness as compared to predicate. |
# **Substantial Equivalence Discussion:**
# Intended Use
The Subject Device (CAS-One IR) has the same intended use as the Predicate Device. The Subject Device, together with its Accessories, is an updated version of the Predicate Device and delivers the same clinical functionality and medical purpose. Differences between the Subject and Predicate Device are:
Hardware changes consist only of minor design and material modifications, consolidation of predicate components into individually labeled Accessories, and incremental upgrades to computing or camera components intended to maintain state-of-the-art performance and cybersecurity. These changes do not affect the intended purpose, functionality, safety, or performance of the Subject Device.
Changes to the software consist of user interface enhancements, visualization improvements, workflow restructurings that build directly upon existing capabilities, and an extension of the organ segmentation outputs to additional anatomical structures. The software relies on the same core programming languages, logic, and machine learning algorithms as the Predicate Device. The extended segmentation outputs constitute a difference in technological characteristics, which does not raise different questions of safety and effectiveness, as demonstrated by segmentation validation.
# Conclusion
The Subject Device is substantially equivalent to the Predicate Device (K232022) with respect to intended use and technological characteristics. This conclusion is based upon a comparison of intended use, technological characteristics, and non-clinical performance testing.
CASCINATION AG
{18}
510(k) Summary – K261225
Page: 15/18
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### i. Software Verification and Validation Testing
Software verification and validation testing were conducted to demonstrate safety and effectiveness of the Subject Device as compared to the predicate. Software documentation was prepared according to the “*Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff*” (June 14, 2023).
All planned software verification and validation testing were successfully completed, and the testing confirmed that the Subject Device performs as intended and meets all predefined acceptance criteria, thereby demonstrating its safety and effectiveness as compared to the predicate.
#### ○ Internal Test Standards
Internal test protocols were executed and documented in test reports to demonstrate navigation accuracy performance. Navigation accuracy was evaluated using a carbon phantom representing patient size, with ground truth geometrical data including trajectory entry points, direction vectors, and single marker positions. Testing was performed under worst-case conditions, comprising a reduced marker configuration (4 of 6 markers visible), a long instrument insertion depth of 225 mm to simulate large patient geometry, and camera tracking positions most distant to the tracking camera.
Testing was conducted for both supported optical tracking cameras (NDI Polaris Vicra and Lyra), with 112 measurements per camera across multiple trajectories. Lateral deviation at the target was defined as the metric of interest, as it is the clinically relevant measure of targeting accuracy, and was assessed against a predefined acceptance threshold.
All acceptance criteria were met for both tracking cameras. Mean lateral deviation was 1.00 mm (SD 0.63 mm) for Lyra and 0.93 mm (SD 0.63 mm) for Vicra. In addition, a paired non-inferiority comparison of Lyra versus Vicra was performed with both cameras mounted co-axially, tracking the same 16 trajectories from the same camera position with 4 of 6 markers visible. The mean paired difference in alignment accuracy was 0.126 mm (SD 0.254 mm), and the upper bound of the one-sided 97.5% confidence interval (0.262 mm) was below the predefined non-inferiority margin of 0.5 mm. The Lyra camera is therefore non-inferior to the Vicra camera.
CASCINATION AG
{19}
510(k) Summary – K261225
Page: 16/18
# ○ Segmentation Validation
Segmentation performance was evaluated using representative clinical datasets acquired across multiple clinical sites, covering multiple anatomical structures. Performance was assessed using Dice similarity coefficients and statistical confidence limits. Predefined acceptance criteria for mean Dice and lower confidence limits were met for all evaluated structures, demonstrating accurate and reliable segmentation performance (Table 3). The segmentation algorithms are based on the same underlying methodological approach as the Predicate Device and exhibit equivalent performance characteristics.
Table 3
Summary – Segmentation Performance
| Segmentation Target | Acceptance Criterion (Dice) | Mean Dice |
| --- | --- | --- |
| Bone and Cartilage | ≥ 0.85 | 0.91 |
| Gallbladder | ≥ 0.85 | 0.89 |
| Heart | ≥ 0.85 | 0.93 |
| Digestive System | ≥ 0.85 | 0.88 |
| Spleen | ≥ 0.85 | 0.97 |
| Lesion | ≥ 0.80 | 0.86 |
| Effective Treatment Volume | ≥ 0.80 | 0.86 |
| Liver | ≥ 0.90 | 0.95 |
| Liver Vessel (Portal Vein) | ≥ 0.60 | 0.67 |
| Liver Vessel (Hepatic Vein) | ≥ 0.60 | 0.72 |
| Lung (incl. airways, excl. vessels) | ≥ 0.90 | 0.97 |
| Kidney | ≥ 0.85 | 0.88 |
# ii. Hardware Verification and Validation
Verification and validation activities were conducted in accordance with established design control procedures and covered mechanical, functional, environmental, and transport testing representative of distribution conditions. Design verification demonstrated that the device conforms to defined design input requirements through mechanical and functional testing. Device validation demonstrated that the device fulfills user needs and intended use under representative conditions.
CASCINATION AG
{20}
510(k) Summary – K261225
Page: 17/18
# - ○ **Electrical safety (IEC 60601-1)**
The CAS-One IR Platform was tested under the CB Scheme according to IEC 60601-1:2005 + A1:2012 + A2:2020. All applicable clauses of the standard were verified through laboratory testing, including mechanical strength, electrical insulation, protection against electric shock, leakage currents, dielectric strength, and marking and labeling requirements. All tests were conducted by an accredited laboratory using calibrated equipment and documented procedures. The overall result of the CB Scheme type testing was pass.
# - ○ **Electromagnetic compatibility (IEC 60601-1-2)**
The CAS-One IR Platform was tested according to IEC 60601-1-2:2014 + A1:2020 under the CB Scheme. All applicable EMC requirements were evaluated, including emissions (radiated and conducted), immunity to electrostatic discharge, radiated RF fields, electrical fast transients, surges, conducted disturbances, magnetic fields at mains frequency, and voltage dips and interruptions.
Testing was performed by an accredited laboratory using calibrated equipment and standardized setups according to the normative test procedures. The system met the acceptance criteria in all tests. No degradation of performance, functional loss, or unsafe conditions were observed during or after exposure to electromagnetic disturbances. The overall result of the CB Scheme EMC testing according to IEC 60601-1-2:2014 + A1:2020 was pass.
### iii. Human Factors and Usability Validation
The usability engineering process was performed in accordance with the international standard “*AAMI / ANSI / IEC 62366-1:2015 + A1:2020, Medical Devices - Part 1: Application of Usability Engineering to Medical Devices*”. Usability impact assessments, comparing the Subject Device to the Predicate Device, were carried out to identify the hazard-related use scenarios affected by the changes. Summative evaluations of these scenarios were performed with five interventional oncologists in Europe; English was the language used for these evaluations, including for the user interface and instructions for use. Use of the device requires mandatory manufacturer training, which is identical worldwide. The Subject Device guides users through the same tasks in each workflow module. All participants completed the evaluated tasks, and no use errors were observed. The scope of the summative evaluations were strictly limited to the changes introduced for each respective version. The evaluations were not performed with 15 residents of the United States, and therefore, do not constitute complete human factors validation testing as described in the FDA guidance “*Applying Human Factors and Usability Engineering to Medical Devices – Guidance for Industry and Food and Drug Administration Staff (2016-02)*”.
The CAS-One IR Platform A was evaluated according to IEC 60601-1-6:2010 + A1:2013 + A2:2020 under the CB Scheme, in conjunction with IEC 62366-1:2015 + A1:2020. The usability engineering process was implemented in compliance with the applicable requirements, including risk assessment according to ISO 14971:2019. The evaluation fulfilled the usability requirements of IEC 60601-1-6 under the CB Scheme.
CASCINATION AG
{21}
510(k) Summary – K261225
Page: 18/18
# iv. Clinical Studies
Clinical testing was not required to demonstrate the safety and effectiveness of the Subject Device as compared to the Predicate. This conclusion is based upon a comparison of intended use, technological characteristics, and non-clinical performance data (software verification and validation testing (including internal test standards and segmentation validation), hardware verification and validation (including electrical safety and electromagnetic compatibility), and human factors and usability validation).
# v. Cybersecurity
Cybersecurity was addressed in accordance with section 524B of the FD&C Act and FDA's guidance Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions. The documentation includes a security architecture description, threat modeling, a cybersecurity risk assessment, a Software Bill of Materials with a vulnerability assessment of third-party components, and a plan for postmarket monitoring and management of vulnerabilities. An independent third party performed security testing on the Subject Device configuration, CAS-One IR Software version 4.3 on Windows 11 IoT Enterprise LTSC. The testing included manual penetration testing of the application (thick client) and the platform, vulnerability scanning, port scanning, and CIS benchmark assessment of the operating system image. No critical or high-severity vulnerabilities were identified. Medium- and low-severity findings were identified, including the use of third-party software components with known vulnerabilities.
# VIII. CONCLUSIONS
The Subject Device is substantially equivalent to the Predicate Device with regard to device performance and safety. This conclusion is based upon a comparison of the intended use, technological characteristics, and non-clinical performance testing. An impact assessment on usability, comparing the Subject Device to the Predicate Device, and summative usability testing of the changes were carried out and demonstrated that the Subject Device, as compared to the Predicate, is safe and effective for the intended users, uses, and use environments. Performance testing demonstrates that the Subject Device is safe and effective as compared to the Predicate. Therefore, it can be concluded that the Subject Device is substantially equivalent to the Predicate Device.
CASCINATION AG
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.