SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)
K261221 · Olympus Surgical Technologies America · GEX · Jul 24, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261221
Device Name
SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)
Applicant
Olympus Surgical Technologies America
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
Device Story
Thulium fiber laser system; produces 1940 nm pulsed coherent near-infrared light; directed via optical fiber to treatment zone; controlled by integrated LED touch screen and footswitch. Used in OR/clinic settings by physicians for soft tissue surgery and lithotripsy. System includes laser console, sterile fibers (150-940 microns), and accessories. Laser energy absorbed by target tissues to perform incision, excision, resection, ablation, coagulation, hemostasis, and vaporization. Benefits include precise tissue management and stone fragmentation. Subject device features minor material changes to fiber strain relief and protective cap; increased shelf-life.
Clinical Evidence
Bench testing only. Verification testing included mechanical and functional performance comparisons (dust cover function, flammability, strain relief pull testing) and shelf-life validation. All tests passed.
Technological Characteristics
Thulium fiber laser; 1940 nm wavelength; 550 nm aiming beam. Max power 60W (Premium) or 35W (Pro). Fiber core diameters 150-940 microns. Materials: Quartz, Silicone, Tefzel. Sterilization: Ethylene Oxide. Standards: ISO 14971, IEC 60601-1, IEC 60825-1, IEC 62304, ISO 10993 series, ASTM F1980 (shelf-life).
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{1}

July 24, 2026
Olympus Surgical Technologies America
Dolan Mills
Sr. Program Manager, Regulatory Affairs
800 West Park Drive
Westborough, MA 01581
Re: K261221
Trade/Device Name: SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument for use In General and Plastic Surgery and In Dermatology
Regulatory Class: II
Product Code: GEX
Dated: June 19, 2026
Received: June 22, 2026
Dear Dolan Mills:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261221 - Dolan Mills
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{3}
K261221 - Dolan Mills
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MARK J. ANTONINO -S
Mark J. Antonino, M.S.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261221 | ? |
| Please provide the device trade name(s). | | ? |
| SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE Laser Fibers, and Accessories) | | |
| Please provide your Indications for Use below. | | ? |
| The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. | | |
| Urology • Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH] • Laser Resection of the Prostate (LRP)• Laser Enucleation of the Prostate (LEP)• Laser Ablation of the Prostate (LAP) • Transurethral Incision of the Prostate (TUIP) • Condylomas • Urethral/ureteral strictures • Lesions of external genitalia • Bladder neck incisions (BNI) • Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors• Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi • Treatment of distal impacted fragments remaining in the ureters following lithotripsy. | | |
| Lithotripsy and Percutaneous Urinary Lithotripsy Indications • Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones• Endoscopic fragmentation of renal calculi• Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed. | | |
| Gastroenterology Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including: • Appendectomy • Polyps • Biopsy • Gall Bladder calculi • Biliary/Bile duct calculi • Ulcers • Gastric ulcers • Duodenal ulcers • Non Bleeding Ulcers • Pancreatitis • Haemorrhoids • Cholecystectomy • Benign and Malignant Neoplasm • Angiodysplasia • Colorectal cancer • Telangiectasias • Telangiectasias of the Osler-Weber-Renu disease • Vascular Malformation • Gastritis • Esophagitis • Esophageal ulcers • Varices • Colitis • Mallory-Weiss tear • Gastric Erosions | | |
| Gynecology Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☐ Adults (22 years old and greater) | ? |
{5}
SOLTIVE Laser System
Gyrus ACMI, Inc.
K261221
Pg 1 of 6
# 510(k) Summary of Safety and Effectiveness
Gyrus ACMI, Inc. SOLTIVE™ LASER SYSTEM
(SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™
Laser Fibers, and Accessories)
# I. Submitter
510(k) Submitter:
Gyrus ACMI, Inc.
9600 Louisiana Ave. North
Brooklyn Park, MN 55445
USA
Establishment Registration Number:
3011050570
Contract Manufacturer:
IPG Medical Corporation
225 Cedar Hill Street
Marlborough, MA 01752, USA
Contact Person:
Dolan Mills
Sr. Program Manager, Regulatory Affairs
Phone: 1-901-355-0007
Email: dolan.mills@olympus.com
Date Prepared:
June 19, 2026
# II. Device
Trade Name:
SOLTIVE™ Laser System (SOLTIVE™ Pro
SuperPulsed Laser, SOLTIVE™ Premium
SuperPulsed Laser, SOLTIVE™ Laser Fibers,
and Accessories)
Regulation Name:
Laser surgical instrument for use in general and
plastic surgery and in dermatology
Classification Name:
Powered Laser Surgical Instrument
21 CFR 878.4810
Regulation/CFR Citation Number:
Product Code:
GEX
Classification:
Class II
Review Panel:
General Surgery Devices
# III. Predicate Device
SOLTIVE Laser Systems
K242191
1
{6}
SOLTIVE Laser System
Gyrus ACMI, Inc.
K261221
Pg 2 of 6
This predicate has been subject to design-related recalls which are related to this submission.
#### IV. Device Description:
The SOLTIVE™ Laser System is a thulium fiber laser, producing a pulsed beam of coherent near- infrared light (1940 nm) upon activation by a footswitch. The beam is then directed to the treatment zone by means of an optical fiber coupled to a handpiece. An integrated LED touch screen gives the user control over the necessary laser system parameters. The SOLTIVE Laser System is equipped with a 550 nm aiming beam.
The SOLTIVE Laser System is comprised of a choice of two models - Premium Laser System (PLS) and Pro Laser System (SLS), laser fibers and accessories. The Premium laser has a maximum power output of 60 Watts and a maximum Frequency output of 2400 Hz and a secondary, foldable screen is provided. The Pro laser can operate at a maximum power of 35 Watts and Frequency output limited to 100Hz. Both systems are compact, designed to fit onto an optional laser cart and are operated with a footswitch (wired or wireless). The Soltive Laser System must be operated with a Soltive Laser Fiber, offered in quartz core diameters from 150 – 940 microns.
The system includes:
- Laser console
- Sterile laser fibers – single-use and reusable options in sizes 150um, 200um, 365um, 550um, and 940um
- Footswitch, wireless or wired options.
- Accessories – power cord, HDMI cable, safety goggles/glasses, fiber strippers, fiber cleaver, blast shields, footswitch (wired and wireless), cart.
The devices do not incorporate medicinal substances, tissues, or blood products. The Soltive laser fibers are provided in single-use and reusable options, and both single-use and reusable fibers are initially sterilized by Ethylene Oxide.
#### V. Indications For Use
The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery.
2
{7}
SOLTIVE Laser System
Gyrus ACMI, Inc.
K261221
Pg 3 of 6
## Urology
- Ablation of Benign Prostatic Hyperplasia (Hypertrophy) [BPH]
- Laser Resection of the Prostate (LRP)
- Laser Enucleation of the Prostate (LEP)
- Laser Ablation of the Prostate (LAP)
- Transurethral Incision of the Prostate (TUIP)
- Condylomas
- Urethral / ureteral strictures
- Lesions of external genitalia
- Bladder neck incisions (BNI)
- Ablation and resection of bladder tumors, urethral tumors, and ureteral tumors
- Endoscopic fragmentation of urethral, ureteral, bladder, and renal calculi
- Treatment of distal impacted fragments remaining in the ureters following lithotripsy
## Lithotripsy and Percutaneous Urinary Lithotripsy Indications
- Endoscopic fragmentation of urethral, ureteral, bladder and renal calculi including cystine, calcium oxalate, monohydrate and calcium oxalate dehydrate stones
- Endoscopic fragmentation of calculi
- Treatment of distal impacted fragments of steinstrasse when guide wire cannot be passed
## Gastroenterology
Open and endoscopic gastroenterology surgery (incision, excision, resection, ablation, vaporization, coagulation and haemostasis) including:
- Appendectomy
- Angiodysplasia
- Polyps
- Colorectal cancer
- Biopsy
- Telangiectasias
- Gall Bladder calculi
- Telangiectasias of the Osler-Weber-Renu disease
- Biliary/Bile duct calculi
- Vascular Malformation
- Ulcers
- Gastritis
- Gastric ulcers
- Esophagitis
3
{8}
SOLTIVE Laser System
Gyrus ACMI, Inc.
K261221
Pg 4 of 6
- Duodenal ulcers
- Esophageal ulcers
- Non Bleeding Ulcers
- Varices
- Pancreatitis
- Colitis
- Haemorrhoids
- Mallory-Weiss tear
- Cholecystectomy
- Gastric Erosions
- Benign and Malignant Neoplasm
### Gynecology
Open, endoscopic (including hysteroscopic) and laparoscopic gynecological surgery (incision, excision, resection, ablation, vaporization, coagulation and hemostasis) of soft tissue.
### VI. Comparison of technological characteristics with the predicate device
| Gyrus ACMI SOLTIVE Laser System | | | |
| --- | --- | --- | --- |
| Design Feature | Subject | K242191 Predicate | Comparison |
| Indications for Use comparison | The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. | The SOLTIVE™ Laser System (SOLTIVE™ Pro SuperPulsed Laser, SOLTIVE™ Premium SuperPulsed Laser, SOLTIVE™ Laser Fibers, and Accessories) is intended for incision, excision, resection, ablation, coagulation, hemostasis, and vaporization of soft tissue, with or without an endoscope, in the following indications: urology, lithotripsy, gastroenterological surgery and gynecological surgery. | Same |
| Function of Device | Thulium Fiber Laser generator with fiber optic delivery | Thulium Fiber Laser generator with fiber optic delivery | Same |
| Mechanics of Action | Light/energy absorption by target tissues | Light/energy absorption by target tissues | Same |
4
{9}
SOLTIVE Laser System
Gyrus ACMI, Inc.
K261221
Pg 5 of 6
| Gyrus ACMI SOLTIVE Laser System | | | |
| --- | --- | --- | --- |
| Design Feature | Subject | K242191 Predicate | Comparison |
| Laser Source | Thulium | Thulium | Same |
| Accessories | Surgical fibers (reusable and single use) available in following nominal core diameters (150, 200, 365, 550, 940 μm) | Surgical fibers (reusable and single use) available in following nominal core diameters (150, 200, 365, 550, 940 μm) | Same |
| Major Software version | 2.3 | 2.2 | Similar |
| Configuration file version | V6.0 | V5.0 | Similar |
| Patient Contacting Materials (Fibers) | Quartz Silicone Tefzel | Quartz Silicone Tefzel | Same |
| IFU, Soltive Laser System and Fibers | PN0015551, revision AK PN0015555, revision AE PN0014679, revision AG | PN0015551, revision AJ PN0015555, revision AC PN0014679, revision AF | Similar |
The subject device has identical technological characteristics (i.e., design, material, chemical composition, principle of operation, energy source, etc.) as the predicate device.
The difference between subject and predicate is a minor change to the Fiber strain relief material and protective cap material at the proximal end of the device, which is not patient contacting, and an increase of the Fiber shelf-life from 1 to 3 years. These differences between the subject and predicate devices do not raise different questions of safety and effectiveness.
### VII. Performance Data
Non-clinical bench testing was performed to verify that the changed material and shelf-life meets all applicable product requirements. The results support the claim of substantial equivalence of the subject device to the predicate device.
Mechanical and Functional: Verification and comparison bench tests were conducted to evaluate the mechanical and functional performance as compared to the predicate. Tests included dust cover function, flammability testing, and strain relief pull testing.
Result - All testing passed, meeting the acceptance criteria of the Design Verification testing. This demonstrates the fibers are equivalent to the predicate.
### Compliance to Standards
The following standards have been used during the design and testing of the subject device system in total.
Applied standards:
5
{10}
SOLTIVE Laser System
Gyrus ACMI, Inc.
K261221
Pg 6 of 6
- ISO 14971:2019
- ISO 15223-1:2021+AMD1:2025
- IEC 60601-1:2005+AMD1:2012+AMD2:2020
- IEC 60601-1-2:2020
- IEC 60825-1:2014
- IEC 62366-1:2015+AMD1:2020
- IEC 60601-2-22:2019
- IEC 60601-6: 2010+A1, A2:2020
- IEC 62304:2006 + A1 2015
- ANSI/AAMI ISO 11135: 2014+A1:2018
- ISO 11607-1:2019
- ASTM F1980-16:2021
- ASTM F1886/F1886M-09:2016
- ASTM F88/F88M-21
- ASTM D4169:2022
- ASTM F2096:2019
- ISO 17665:2024
- ISO 17664-1:2021
- ISO 10993-1:2018
- ISO 10993-5:2009
- ISO 10993-7:2019
- ISO 10993-10:2021
- ISO 10993-11:2017
- ISO 10993-17:2013
- ISO 10993-18:2022
- ISO 10993-23:2025
- ISO 14937:2009
### VIII. Conclusion:
Based on the bench testing, the modified device was found to be substantially equivalent to the predicate device.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.