K261154 · In2Bones USA, LLC · HWC · May 30, 2026 · Orthopedic
Device Facts
Record ID
K261154
Device Name
CoLink® & CoLag® Non-Sterile Screws
Applicant
In2Bones USA, LLC
Product Code
HWC · Orthopedic
Decision Date
May 30, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The In2Bones USA CoLink Non-Sterile screws when used in conjunction with compatible plating systems are indicated for: CoLink T8 2.7mm Screws: When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients. CoLink T15 3.0mm Screws: When used with the CoLink Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients. CoLink T15 3.5mm Screws: When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients in both pediatric and adult patients. When used with the CoLink Mfx Plating System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients. The In2Bones USA CoLag Non-sterile screws, when used in conjunction with compatible plating systems or used independently in a lag screw technique are indicated for: CoLag T8 3.0mm Screws: When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). CoLag T15 4.0mm Screws: When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). When used with the CoLink® Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients.
Device Story
Titanium alloy (Ti6Al4V ELI) bone screws; intended for internal bone fixation; used independently or with compatible plating systems (CoLink Afx, CoLink Mfx, CoLag Fracture and Correction System). Provided non-sterile; requires cleaning and steam sterilization at healthcare facility prior to surgical implantation. Operates via mechanical fixation principles; threaded design provides stability for fractures, osteotomies, and fusions. Used by surgeons in clinical settings. Benefits include stabilization of bone segments to facilitate healing.
Clinical Evidence
Bench testing only. Evidence includes biocompatibility (ISO 10993-1), steam sterilization validation (ISO 17665), cleaning validation, and mechanical performance testing. No clinical data provided.
Technological Characteristics
Material: Titanium Alloy (ASTM F136). Principle: Mechanical bone fixation via threaded screws. Dimensions: 8mm-60mm lengths. Connectivity: None. Sterilization: Steam sterilization required by user. Software: None.
Indications for Use
Indicated for pediatric and adult patients requiring internal bone fixation, stabilization of fractures/osteotomies, joint fusion, or reconstruction. Specific applications include ankle (tibia/fibula), midfoot (Lis Franc injuries), and various bones (humerus, radius, ulna, calcaneus, small bones).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 30, 2026
In2Bones USA, LLC
Tina Mornak
Manager, Regulatory Affairs
7750a Trinity Road
Cordova, Tennessee 38018
Re: K261154
Trade/Device Name: CoLink® & CoLag® Non-Sterile Screws
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC
Dated: May 12, 2026
Received: May 13, 2026
Dear Tina Mornak:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261154 - Tina Mornak
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K261154 - Tina Mornak
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261154 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| CoLink® & CoLag® Non-Sterile Screws | | |
| Please provide your Indications for Use below. | | ? |
| The In2Bones USA CoLink Non-Sterile screws when used in conjunction with compatible plating systems are indicated for: | | |
| CoLink T8 2.7mm Screws: When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients. | | |
| CoLink T15 3.0mm Screws: When used with the CoLink Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries. | | |
| CoLink T15 3.5mm Screws: When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients in both pediatric and adult patients. | | |
| When used with the CoLink Mfx Plating System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients. | | |
| The In2Bones USA CoLag Non-sterile screws, when used in conjunction with compatible plating systems or used independently in a lag screw technique are indicated for: | | |
| CoLag T8 3.0mm Screws: When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). | | |
| CoLag T15 4.0mm Screws: When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). | | |
| When used with the CoLink® Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
{4}
Please select the age group(s) for which the device(s) is to be used.
☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☑ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater)
{5}
510(k) Summary
CoLink® & CoLag® Non-Sterile Screws
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, ConMed Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for this 510(k).
I. SUBMITTER
In2Bones USA
7750A Trinity Road
Cordova, TN 38018
Company Contact: Tina Mornak
Manager, Regulatory Affairs
724-518-3191
tinamornak@conmed.com
Date Prepared: May 29, 2026
II. DEVICE NAME
Proprietary Name: CoLink® & CoLag® Non-Sterile Screws
Common Name: Screw, Fixation, Bone
Panel: 87-Orthopedic
Product Code: HWC
Device Class: II
Regulation Number: 21 CFR 888.3040
III. PREDICATE/LEGALLY MARKETED DEVICE
Primary Device Name: CoLink® Afx
Company Name: In2Bones USA
510(k): K181113
Additional Predicate Device Name: CoLink® Mfx
Company Name: In2Bones USA
510(k): K210060
Additional Predicate Device Name: CoLag®2
Company Name: In2Bones USA
510(k): K180377
IV. DEVICE DESCRIPTION
The subject CoLink and CoLag Non-Sterile screws are titanium alloy (Ti6Al4V ELI) screws intended for internal bone fixation in adult patients. They are provided non-sterile and to be cleaned and steam sterilized at the healthcare facility prior to surgical implantation. These screws may be used independently or in
K261154 | 1
{6}
conjunction with previously cleared In2Bones plating systems, consistent with their cleared clinical applications.
V. INTENDED USE / INDICATIONS FOR USE
The In2Bones USA CoLink Non-Sterile screws when used in conjunction with compatible plating systems are indicated for:
CoLink T8 2.7mm Screws:
When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients.
CoLink T15 3.0mm Screws:
When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients.
When used with the CoLink Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients.
CoLink T15 3.5mm Screws:
When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients.
When used with the CoLink Mfx Plating System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients.
The In2Bones USA CoLag Non-sterile screws, when used in conjunction with compatible plating systems or used independently in a lag screw technique are indicated for:
CoLag T8 3.0mm Screws:
When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges).
CoLag T15 4.0mm Screws:
When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges).
When used with the CoLink Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients.
K261154 | 2
{7}
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The CoLink & CoLag Non-Sterile Screws are similar to the predicate device in that the design of both include threaded titanium alloy bone screws with similar materials, geometry, drive interfaces, and mechanical fixation principles. The main difference in the subject device from the predicate device is the change from sterile to nonsterile configuration, as all other design and functional characteristics remain the same. CoLink & CoLag Non-Sterile Screws are safe, effective and substantially equivalent to the predicate as demonstrated by non-clinical performance testing.
| | Subject Device | Predicate Devices |
| --- | --- | --- |
| Device Name | CoLink & CoLag Non-Sterile Screws | CoLink AFX Plating System CoLink Mfx Plating System CoLag Fracture and Correction System |
| Manufacturer | In2Bones USA | In2Bones USA |
| Regulation Number | 21 CFR 888.3040 | 21 CFR 888.3040, 21 CFR 888.3030 |
| Product Code | HWC | HWC, HRS |
| Device Class/Name | Class II | Class II |
| Fundamental Scientific Technology | metallic bone fixation screws | metallic bone fixation screws |
| Indications for Use | The In2Bones USA CoLink and CoLink Afx Non-Sterile screws when used in conjunction with compatible plating systems are indicated for: **CoLink T8 2.7mm Screws:** • When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients. **CoLink T15 3.0mm Screws:** • When used with the CoLink Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients. **CoLink T15 3.5mm Screws:** • When used with the CoLink Afx Plating System: Stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula in both pediatric and adult patients. • When used with the CoLink Mfx Plating System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, | The In2Bones USA LLC CoLink Afx Plating System is indicated for stabilization and fixation of fractures and osteotomies in the ankle, including tibia and fibula. The In2Bones USA LLC CoLink Mfx Plating System is indicated for stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries. The In2Bones USA LLC Fracture and Correction System lag screws are intended to be used as stand-alone bone screws, or in a plate-screw system for internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). |
K261154 | 3
{8}
K261154 | 4
| | including repair of Lis Franc injuries in both pediatric and adult patients. The In2Bones USA CoLag Non-sterile screws, when used in conjunction with compatible plating systems or used independently in a lag screw technique are indicated for: **CoLag T8 3.0mm Screws:** • When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). **CoLag T15 4.0mm Screws:** • When used with the Colag Fracture and Correction System: Internal bone fixation for bone fractures, fusions, osteotomies and non-unions of various bones, including humerus, radius, ulna, tibia, calcaneus, fibula, and small bones (metacarpals, metatarsals, and phalanges). • When used with the CoLink Mfx System: Stabilization and fixation of fractures, revision procedures, joint fusion, osteotomies and reconstruction in the midfoot bones of the foot, including repair of Lis Franc injuries in both pediatric and adult patients. | |
| --- | --- | --- |
| Material | Titanium Alloy (ASTM F136) | Titanium Alloy (ASTM F136) |
| Geometry and Dimensions | 8mm-60mm length (CoLink Afx) | 8mm-60mm length (CoLink Afx) |
| | 8mm – 60mm length (CoLink Mfx) | 8mm – 60mm length (CoLink Mfx) |
| | 12mm-60mm length (CoLag) | 8mm-115mm length (CoLag) |
{9}
VII. PERFORMANCE DATA/RATIONALES
The implants subject of this submission were determined not to be worse case and were adopted into previous validations conducted:
- Biocompatibility per ISO 10993-1
- Steam Sterilization validation per ISO 17665
- Cleaning validation
- Mechanical performance
VIII. CONCLUSION
The subject CoLink, CoLink Afx & CoLag Non-Sterile Screws are substantially equivalent in design, materials, indications for use, principles of operation and technological characteristics to the predicate devices. Based upon the findings of non-clinical testing, the differences present no issues of safety and efficacy, and the subject CoLink & CoLag Non-Sterile Screws are substantially equivalent to the predicates: CoLink Afx Plating System screws (K181113), CoLink Mfx Plating System screws (K210060) and CoLag2 screws (K180377).
K261154 | 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.