TetraGraph is intended to deliver stimulations to a nerve and record, measure, analyze and report muscle electrical activity to determine muscle function.
Device Story
TetraGraph is a portable, battery- or wall-powered EMG-based neuromuscular transmission monitor used in hospital settings (perioperative, recovery, critical care). It delivers electrical stimulation to a peripheral nerve and records the evoked Muscle Action Potential (MAP) via surface electrodes. The device automatically measures, analyzes, and reports muscle electrical activity to determine the level of neuromuscular blockade. Clinicians use the displayed quantitative values to assess patient relaxation status during procedures requiring neuromuscular blocking agents (e.g., intubation, mechanical ventilation). The device supplements clinical data to aid in managing patient paralysis and recovery. It is operated by healthcare professionals in clinical environments.
Clinical Evidence
No clinical test was performed in support of this premarket notification. Substantial equivalence is supported by bench testing, including software verification and validation, cybersecurity mitigation, electrical safety, EMC, and electrode verification testing.
Technological Characteristics
Portable EMG-based monitor. Uses constant current square wave stimulation (200 or 300 μs pulse width). Stimulation current range: 10-80 mA. Components include TetraGraph monitor (SEN 2015) and single-use electrode arrays. Connectivity includes optional TetraHub. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-2-40 (nerve stimulators).
Indications for Use
Indicated for monitoring the relaxation of the patient when neuromuscular blockade has been administered. Applicable to infants (29 days to <2 years), children (2-12 years), adolescents (12-22 years), and adults (22+ years).
Regulatory Classification
Identification
An electrical peripheral nerve stimulator (neuromuscular blockade monitor) is a device used to apply an electrical current to a patient to test the level of pharmacological effect of anesthetic drugs and gases.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 22, 2026
Senzime AB
% Elisa Maldonado-Holmertz
RA/QA Consultant
Obelix Consulting, LLC
806 Jefferson St.
Bastrop, Texas 78602
Re: K261098
Trade/Device Name: TetraGraph Neuromuscular Transmission Monitor (SEN 2015)Accessories:TetraSens (SEN 2012)TetraSens Pediatric (SEN 2013)TetraSensitive (SEN 2016)TetraHub (SEN 2017)
Regulation Number: 21 CFR 868.2775
Regulation Name: Electrical Peripheral Nerve Stimulator
Regulatory Class: Class II
Product Code: KOI
Dated: April 24, 2026
Received: April 24, 2026
Dear Elisa Maldonado-Holmertz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K261098 - Elisa Maldonado-Holmertz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K261098 - Elisa Maldonado-Holmertz
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bradley Q. Quinn -S
Bradley Quinn
Assistant Director
DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K261098 | ? |
| Please provide the device trade name(s). | | ? |
| TetraGraph Neuromuscular Transmission Monitor (SEN 2015) Accessories: TetraSens (SEN 2012) TetraSens Pediatric (SEN 2013) TetraSensitive (SEN 2016) TetraHub (SEN 2017) | | |
| Please provide your Indications for Use below. | | ? |
| The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade has been administered. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
Senzime AB
Special 510(k) Premarket Submission
TetraGraph Neuromuscular Transmission Monitor
SENZIME
# 510(k) Summary
## Submission Sponsor and contact person
Sponsor Contact: Johanna Faris, Head of QA/RA/S
Submission Sponsor: Senzime AB
Address: Verkstadsgatan 8, 75323 Uppsala, Sweden
Email: johanna.faris@senzime.com
Tel number: +46 (0)705286330
Submission Correspondent: Elisa Maldonado-Holmertz
Company Name: Obelix Consulting, LLC
Address: 1806 Jefferson St, Bastrop, TX 78602
Email: elisamh@obelixconsult.com
Tel number: +1 512 431 6069
## Date Prepared
20 May 2026
## Device Identification
Table 1. Device identification
| The 510(k) number under which the legally marketed (existing) device was cleared | Traditional 510(k): K220530, cleared August 17, 2022 |
| --- | --- |
| Type of this 510(k) Submission: | Special |
| Trade or Proprietary Name: | Tetragraph Neuromuscular Transmission Monitor |
| Common or Usual Name: | Tetragraph |
| Regulation Description: | Stimulator, Nerve, Peripheral, Electric |
| Regulation Classification: | 868.2775 |
| Product Code | KOI |
| Class of Device: | Class II |
| Review Panel: | Division of Anesthesia, Respiratory, and Sleep Devices (DHT1C) |
| Reason for Submission: | New accessories: electrode size and interface |
| Prior Related Submissions: | K190795 K220530 |
| Multiple Devices: | None; this is the only device in the submission |
Legally marketed predicate device
1(6)
{6}
Senzime AB
Special 510(k) Premarket Submission
TetraGraph Neuromuscular Transmission Monitor
SENZIME
Predicate Device - K220530 Tetragraph Neuromuscular Transmission Monitor.
## Device Description
The TetraGraph Neuromuscular Transmission (NMT) Monitor is a portable, battery- or wall power operated EMG-based neuromuscular transmission monitor for use in hospital settings including perioperative and in recovery and critical care environments following or during the application of Neuromuscular block. TetraGraph is intended to deliver stimulations to a nerve and record, measure, analyze and report muscle electrical activity to determine muscle function. TetraGraph is a prescription-only medical device.
Neuromuscular Transmission (NMT) is the transfer of an electrical impulse between a motor nerve and its associated muscle. The NMT is blocked by neuromuscular blocking agents ("NMBAs") which cause transient muscle paralysis preventing the patient from moving and breathing spontaneously. NMBAs are commonly used during different clinical procedures, for example, muscle relaxation is used during general anesthesia to enable endotracheal intubation and mechanical ventilation and to provide optimal surgical conditions. Muscle relaxation may also be used in critical care during mechanical ventilation. In these circumstances, TetraGraph can be used as an objective monitor of neuromuscular transmission.
TetraGraph undertakes this function by electrical stimulation of the peripheral nerve and directly measuring the evoked response of the muscles (Muscle Action Potential (MAP)), thus providing a quantitative and automatic measurement of muscle response to a stimulus using electromyography (EMG). The TetraGraph is a prescription-only medical device and is indicated for use in hospitals. TetraGraph supplements the use of clinical information/data obtained from other monitors. The level of neuromuscular block is routinely measured by stimulating a peripheral nerve. The TetraGraph controls the level of electrical stimulation applied to the nerve and monitors the muscle response by the use of Electromyography (EMG) detected by electrodes on the muscle. The measured values that are displayed on TetraGraph provide the clinicians with an indication on the level of neuromuscular blockade in the patients.
2(6)
{7}
Senzime AB
Special 510(k) Premarket Submission
TetraGraph Neuromuscular Transmission Monitor
^{}[]
SENZIME
# TetraGraph consists of the following main components:
SEN 2015 TetraGraph Neuromuscular Transmission Monitor
Accessories:
SEN 2012 TetraSens
SEN 2013 TetraSens Pediatric
SEN 2016 TetraSensitive
Optional accessory:
SEN 2017 TetraHub
# Intended Use
TetraGraph is intended to deliver stimulations to a nerve and record, measure, analyze and report muscle electrical activity to determine muscle function.
# Indication for Use Statement
The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade has been administered.
# Intended Use and Indications for Use Comparison
Thes subject Intended Use and Indication for Use Statement is the same as predicate K220530.
## Technological Comparison
| Trade Name | TetraGraph Neuromuscular Transmission Monitor (Part number SEN 2015) | TetraGraph Neuromuscular Transmission Monitor (Part number SEN 2001) | Comment |
| --- | --- | --- | --- |
| Device | SUBJECT (K261098) | PREDICATE (K220530) | |
| Classification Panel | Anesthesiology | Anesthesiology | Same as its predicate device |
| Regulation Number | 868.2775 | 868.2775 | Same as its predicate device |
| Regulatory Class | Class II | Class II | Same as its predicate device |
{8}
Senzime AB
Special 510(k) Premarket Submission
TetraGraph Neuromuscular Transmission Monitor
SENZIME
| Trade Name | TetraGraph Neuromuscular Transmission Monitor (Part number SEN 2015) | TetraGraph Neuromuscular Transmission Monitor (Part number SEN 2001) | Comment |
| --- | --- | --- | --- |
| Device | SUBJECT (K261098) | PREDICATE (K220530) | |
| Product Code | KOI | KOI | Same as its predicate device |
| Intended use | TetraGraph is intended to deliver stimulations to a nerve and record, measure, analyze and report muscle electrical activity to determine muscle function. | TetraGraph is intended to deliver stimulations to a nerve and record, measure, analyze and report muscle electrical activity to determine muscle function. | Same as its predicate device |
| Indications for Use | The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered. | The TetraGraph Neuromuscular Transmission (NMT) Monitor is indicated for monitoring the relaxation of the patient when neuromuscular blockade is administered. | Same as its predicate device |
| Description of design | The TetraGraph Neuromuscular Transmission (NMT) Monitor (TetraGraph) is a portable EMG-based neuromuscular transmission monitor for use perioperative and in recovery and critical care environments following or during the application of Neuromuscular block. | The TetraGraph Neuromuscular Transmission (NMT) Monitor (TetraGraph) is a portable EMG-based neuromuscular transmission monitor for use perioperative and in recovery and critical care environments following or during the application of Neuromuscular block. | Same as its predicate device |
| Patient population group | All patient populations (except neonates) | all patient populations, (except neonates) | Same as its predicate device |
| Operating Principle | Electromyography (EMG) | Electromyography (EMG) | Same as its predicate device |
| Stimulation Patterns | Single Twitch (ST): pulse of 200 or 300μs duration at 10 or 5 seconds, Train-of-Four (TOF): 4 pulses of 200 or 300 μs duration at 2 Hz, repeated at user selected frequency of 15 seconds, 1 minute,5 minutes, 15 minutes or 60 minutes Post-tetanic Count (PTC): PTC consists of Tetanic Stimulation, a set of 250 | Single Twitch (ST): pulse of 200 or 300 μs duration at 10 seconds Train-of-Four (TOF): 4 pulses of 200 or 300 μs duration at 2 Hz repeated at user selected frequency of 20 seconds, 1 minute, 5 minutes, 15 minutes or 60 minutes Post-tetanic Count (PTC): PTC consists of Tetanic Stimulation, a set of 250 | Minor change in stimulation patterns. The differences do not raise new questions of safety or effectiveness. |
4(6)
{9}
Senzime AB
Special 510(k) Premarket Submission
TetraGraph Neuromuscular Transmission Monitor
^{}[]
SENZIME
| Trade Name | TetraGraph Neuromuscular Transmission Monitor (Part number SEN 2015) | TetraGraph Neuromuscular Transmission Monitor (Part number SEN 2001) | Comment |
| --- | --- | --- | --- |
| Device | SUBJECT (K261098) | PREDICATE (K220530) | |
| | pulses (pulses at 50Hz over 5 seconds stimulated according to the current setting in the monitor) followed by up to 20 ST pulses at 1 Hz | pulses (pulses at 50Hz over 5 second at 50mA, 200 μs) followed by up to 20 ST pulses at 1 Hz | |
| Stimulating electrode placement | Peripheral Nerve | Peripheral Nerve | Same as its predicate device |
| Stimulation Current Range | 10-80 mA | 10-60 mA | Minor change in the stimulation current range. The differences do not raise new questions of safety or effectiveness. |
| Stimulation Pulse Width | Square wave, constant current: 200 μs or 300 μs | Square wave, constant current: 200 μs or 300 μs | Same as its predicate device |
| Electrode Connection | Reusable Cable | Reusable Cable | Same as its predicate device |
| Electrode for Stimulation | Single use electrode array (4 electrodes on an array) | Single use electrode array (4 electrodes on an array) | Same as its predicate device |
There is no change in the technological characteristics for the subject TetraGraph monitoring system relative to the predicate TetraGraph monitoring system cleared under Senzime 510(k) K220530 supporting substantial equivalence.
## Performance Testing
The safety and performance of the subject device, TetraGraph Neuromuscular Transmission Monitor, has been evaluated and verified in accordance with its performance specification and intended use to support a determination of substantial equivalence to the predicate device.
The following verification and validation testing was performed to support the modifications described in this Special 510(k) submission. All testing was completed and met predetermined acceptance criteria.
- Software Verification and Validation, including Cybersecurity mitigation
5(6)
{10}
Senzime AB
Special 510(k) Premarket Submission
TetraGraph Neuromuscular Transmission Monitor
^{}[]
SENZIME
^{}[]
- Verification against IEC 60601-2-40
- Electrical safety testing against IEC 60601-1
- EMC testing against IEC 60601-1-2
- Electrode Verification testing
Performance testing demonstrates that the proposed device is as safe and effective, performs as well as the predicate, and is deemed substantially equivalent.
## Clinical Performance Data
No clinical test was performed in support of this premarket notification.
## Statement on Substantial Equivalence
Based on this, the proposed TetraGraph Neuromuscular Transmission Monitor is substantially equivalent to the predicate device.
6(6)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.