Pocket Ultrasound System (C10MTpro,C10MXpro,C10MSpro,C10MBpro,C10MRpro)
Applicant
Beijing Konted Medical Technology Co.,Ltd
Product Code
IYN · Radiology
Decision Date
Jul 23, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Pocket Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B +M), and Pulsed Wave modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications depending on different configuration of array probes: - Convex array probe - Fetal/Obstetrics, Gynecological, Abdominal, Urology - Linear array probe - Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal, Pediatric The Pocket Ultrasound System is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
Device Story
Portable, software-controlled diagnostic ultrasound system; acquires/displays real-time ultrasound data. Comprises wireless transducers (convex/linear arrays) and MY USG Android app on COTS smartphone/tablet. Transducers use inverse piezoelectric effect to emit/receive ultrasonic waves; echoes converted to electrical signals; processing unit calculates intensity/timing to form images. Supports B-mode, B+M, Color Doppler, and Pulsed Wave Doppler. Used by healthcare professionals in hospitals/clinics. Output displayed on mobile device touch screen; aids clinical decision-making via real-time visualization of tissues, organs, and blood flow. Wireless communication via direct Wi-Fi. Battery-powered.
Clinical Evidence
Bench testing only. Performance evaluated via electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and ultrasonic specific testing (IEC 60601-2-37:2024). Image performance assessment conducted to verify accuracy of frequency and depth measurements for new probe models.
Technological Characteristics
Wireless transducers (convex/linear arrays); frequencies 3.5MHz–19MHz. Principle: ultrasonic echo imaging via inverse piezoelectric effect. Connectivity: direct Wi-Fi to COTS Android device. Power: rechargeable Li-ion battery (DC 5V). Software: MY USG Android app. Patient-contacting materials compliant with ISO 10993-5/10. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring fetal/obstetrics, gynecological, abdominal, urological, small parts (breast, thyroid), carotid, peripheral vessel, muscular-skeletal, or pediatric examinations. Intended for use by qualified healthcare professionals in clinical environments.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 23, 2026
Beijing Konted Medical Technology Co.,Ltd
% Xuexia Ren
Official Correspondent
Chengdu Lizhu Technology Co., Ltd.
#1179c, 1st Floor, # 2 Niusha Heng St., Jinjiang District
Chengdu, Sichuan Province 610065
CHINA
Re: K261081
Trade/Device Name: Pocket Ultrasound System
(C10MTpro,C10MXpro,C10MSpro,C10MBpro,C10MRpro)
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX
Dated: June 22, 2026
Received: June 23, 2026
Dear Xuexia Ren:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261081 - Xuexia Ren
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261081 - Xuexia Ren
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D.
O'hara -S
Date: 2026.07.23 14:50:32 -04'00'
For
Yanna Kang, Ph.D.
Assistant Director
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261081
Device Name
Pocket Ultrasound System (C10MTpro,C10MXpro,C10MSpro,C10MBpro,C10MRpro)
Indications for Use (Describe)
The Pocket Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B +M), and Pulsed Wave modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications depending on different configuration of array probes:
- Convex array probe - Fetal/Obstetrics, Gynecological, Abdominal, Urology
- Linear array probe - Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal, Pediatric
The Pocket Ultrasound System is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
# 510(k) Summary
K261081
# I. SUBMITTER
Name: Beijing Konted Medical Technology Co., Ltd.
Address: Room 111,1F, Building 3, No. 27, Yong wang Road, Daxing Biological Pharmaceutical Industry Base, Daxing District, Beijing, 102629 PEOPLE'S REPUBLIC OF CHINA
Name of contact person: Deyi Zhu
Tel: +86 10-60219113
Fax: +86 10-60219213
E-Mail: deyiz@konted.cn
Submission date: 2026-06-23
# II. Subject Device
Device trade name: Pocket Ultrasound System
Model: C10MTpro,C10MXpro,C10MSpro,C10MBpro,C10MRpro
Regulation number: 21 CFR 892.1550
Regulation Name: Ultrasonic pulsed doppler imaging system
Regulation class: II
Panel: Radiology
Product code: IYN, IYO, ITX
# III. Predicative/Reference Device
Predicative Device:
Submission Number: K241979
Device Name: Pocket Ultrasound System
Device Model: C10, C10TX, C10RL, C10RLpro
Applicant: Beijing Konted Medical Technology Co., Ltd
Regulation number: 21 CFR 892.1550
Regulation Name: Ultrasonic pulsed doppler imaging system
Regulation class: II
Panel: Radiology
Product code: IYN, IYO, ITX
Reference Device:
Submission Number: K253448
Page 1 of 10
{5}
Device Name: Sonosite MTUltrasound System
Applicant: FUJIFILM SonoSite, Inc.
Regulation number: 21 CFR 892.1550
Regulation Name: Ultrasonic pulsed doppler imaging system
Regulation class: II
Panel: Radiology
Product code: IYN, IYO, ITX
# IV. Device Description
The Pocket Ultrasound System is designed to be a portable, general-purpose, software-controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through an off-the-shelf (OTS) Android device. the system mainly includes the ultrasound equipment and software application components. The ultrasound equipment is used to complete the ultrasonic imaging function and the software application components are used to complete the image display, measurement, patient information management as well as image storage etc. The Pocket Ultrasound System comprises a series of wireless transducers employing Wi-Fi-based technology to communicate with smartphone devices via direct Wi-Fi. One of the software application components is a software APP (MY USG), which is deployed on Android devices which supported Wi-Fi wireless communication. The other one is the software contained in the ultrasound equipment to control operation of the equipment and transfer images via Wi-Fi to the smartphone devices. This allows the user to export ultrasound images and display them across a range of portable personal devices. The communication protocol between the software APP and the other software contained in the equipment is specially defined for unique connection.
The system is only intended for use by qualified healthcare professionals who are appropriately trained and qualified in the use of ultrasound imaging technology and the associated clinical applications. The Pocket Ultrasound System comprises the following:
A commercial off-the-shelf (COTS) Android smartphone/tablet device.
the software application (MY USG) running as an App on the COTS device with Android system.
The ultrasound equipment/Transducers: C10MTpro/C10MXpro/C10MSpro/C10MBpro/C10MRpro
- Accessory: Wireless Battery Charger & Power Adapter
The Pocket Ultrasound System is designed with four ultrasound modes of operation, including B-mode, Combination Mode (B+M mode), Color Doppler mode and Pulsed Wave Doppler mode.
Real time image and dynamic changes of tissues, organs or vessels will be showed in the image under B-mode, while coloring blood flow can be seen under Color Doppler mode and Pulsed Wave Doppler mode. Combination mode B+M-mode is the combination of the B-mode and M-mode of operation, superimposed on the display. M-mode is defined as time motion display of the ultrasound wave along a chosen ultrasound line, so it's usually applied in diagnosis of cardiovascular diseases.
The system provides real-time images for tissues, organs or blood flow via touch screen of the COTS Android smartphone/tablet device, which, in combination with the embedded software system contained in
Page 2 of 10
{6}
the ultrasound device, provides users a friendly and convenient graphical interface and makes users be able to easily control the whole system.
The following are the parameters of the newly added models:
| Items | C10MBpro | C10MRpro | C10MSpro | C10MTpro | C10MXpro |
| --- | --- | --- | --- | --- | --- |
| Transducer type | Linear array | Linear array | Linear array | Convex array | Linear array |
| Nominal frequency | 15MHz | 19MHz | 10MHz | 3.5 MHz | 7.5 MHz |
| Frequency class | High frequency | High frequency | High frequency | Low frequency | High frequency |
| Array length (L) or radius of curvature (R) | L=19.2mm | L=12.8mm | L=25.6mm | R=60mm | L=38.4mm |
| Scan Depth | 20~40mm | 0~25mm | 0~80mm | ≥160mm | 20~100mm |
| Clinical application | Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal and Pediatric. | Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal and Pediatric. | Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal and Pediatric. | Fetal/Obstetrics, Gynecological, Abdominal, Urology | Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal and Pediatric. |
# V. Indication For Use
The Pocket Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+M), and Pulsed Wave modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications depending on different configuration of array probes:
Convex array probe - Fetal/Obstetrics, Gynecological, Abdominal, Urology
Linear array probe - Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal, Pediatric
The Pocket Ultrasound System is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
# VI. Comparison to predicate device and conclusion
Page 3 of 10
{7}
The intended use, main materials, functional design, safety and performance characteristics of the subject device is substantially equivalent to the predicate/reference devices quoted above.
The differences between the subject device and predicate/reference devices do not raise new issues of safety or effectiveness.
| Elements of Comparison | Subject Device | Predicate Device | Reference Device | Discussion |
| --- | --- | --- | --- | --- |
| Device Name | Pocket Ultrasound System | Pocket Ultrasound System (C10) | Sonosite MT Ultrasound System | -- |
| 510(k) Number | K261081 | K241979 | K253448 | -- |
| Model | C10MTpro,C10MXpro,C10M Spro,C10MBpro,C10MRpro | C10, C10TX, C10RL, C10RLpro | -- | -- |
| Manufacture | Beijing Konted Medical Technology Co., Ltd | Beijing Konted Medical Technology Co., Ltd | FUJIFILM SonoSite, Inc. | -- |
| Classification Name: | Ultrasonic pulsed doppler imaging system | Ultrasonic pulsed doppler imaging system | Ultrasonic Pulsed Doppler Imaging System | SE |
| Device Classification: | II | II | II | SE |
| Clinical characteristics | | | | |
| Indications for Use | The Pocket Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+M), and Pulsed Wave modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications depending on different configuration of array probes: Convex array probe - Fetal/Obstetrics, Gynecological, Abdominal, | The Pocket Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+M), and Pulsed Wave modes. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications depending on different configuration of array probes: Convex array probe - Fetal/Obstetrics, Gynecological, Abdominal, | The Sonosite MT Ultrasound System is a general-purpose ultrasound system intended for use by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Abdominal Adult Cephalic Cardiac Adult Cardiac Pediatric Fetal-OB/GYN Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) | Different, see Note 1 |
Page 4 of 10
{8}
| | Urology Linear array probe - Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal, Pediatric The Pocket Ultrasound System is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals. | Urology Linear array probe - Small Parts (breast, thyroid), Carotid, Peripheral Vessel, Muscular-Skeletal, Pediatric Phase array probe - Cardiology, Cardiac Fetal Echo. The Pocket Ultrasound System is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals. | Ophthalmic Pediatric Peripheral Vessel Small Organ (breast, thyroid, testicles, prostate) Transvaginal Needle Guidance Modes of operation include: B Mode (B), M-Mode (M) (including simultaneous M-mode and anatomical M Mode), PW Doppler (PWD) (including High Pulse Repetition Frequency (HPRF) and simultaneous PWD for certain exam types), Tissue Doppler Imaging (TDI), Continuous Wave Doppler (CWD), Color Power Doppler, Velocity Color Doppler, Color Variance, Tissue Harmonic Imaging (THI), Multi-beam imaging, Steep Needle Profiling, Trapezoid, and combined modes, including duplex and triplex imaging: B+M, B +PWD, B+CWD, B+C, (B+C)+PWD, (B+C)+CWD. This device is indicated for Prescription Use Only. The Sonosite MT Ultrasound System is intended to be used in medical practices, clinical environments, including Healthcare facilities, Hospitals, Clinics and clinical point-of-care for diagnosis of patients. The system is used with a transducer attached and is powered either by battery or by AC electrical power. The clinician is positioned next to the patient and places the transducer onto the patient's | |
| --- | --- | --- | --- | --- |
Page 5 of 10
{9}
| | | | body where needed to obtain the desired ultrasound image | |
| --- | --- | --- | --- | --- |
| Intended patient population | For use in all patients | For use in all patients | For use in all patients | SE |
| Intended use environment | Hospital, clinics, for use by healthcare professionals | Hospital, clinics, home for use by healthcare professionals | Hospital, clinics, home for use by healthcare professionals | Different, see Note 5 |
| General technological characteristics | | | | |
| Working principle | Principle of operation of the Pocket Ultrasound System is based on ultrasonic echo imaging, which is a common medical imaging technology to use the transmission speed of ultrasound in different tissues and reflection degree differences, to get images through computer processing for diagnosis and treatment. The transducers inside the probe are responsible to emitting and receiving ultrasonic waves. When a pulse with a certain frequency and magnitude onto the transducers, then ultrasonic waves are generated through the inverse piezoelectric effect and transmitted to the human body tissue through the probe. When the ultrasonic wave meets the tissue interface, it will generate echoes, which are scattered in different tissues or blood flow and received by the | Principle of operation of the Pocket Ultrasound System is based on ultrasonic echo imaging, which is a common medical imaging technology to use the transmission speed of ultrasound in different tissues and reflection degree differences, to get images through computer processing for diagnosis and treatment. The transducers inside the probe are responsible to emitting and receiving ultrasonic waves. When a pulse with a certain frequency and magnitude onto the transducers, then ultrasonic waves are generated through the inverse piezoelectric effect and transmitted to the human body tissue through the probe. When the ultrasonic wave meets the tissue interface, it will generate echoes, which are scattered in different tissues or blood flow and received by the | -- | SE with the predicated device |
Page 6 of 10
{10}
| | probe and converted into electrical signals. The receiver amplifies and processes these electrical signals before transmitting them to the processing unit. Based on the signals received, the processing unit will calculate the intensity and timing of the echoes to form an image. | probe and converted into electrical signals. The receiver amplifies and processes these electrical signals before transmitting them to the processing unit. Based on the signals received, the processing unit will calculate the intensity and timing of the echoes to form an image. | | |
| --- | --- | --- | --- | --- |
| Design | Autocorrelation for color processing and FFT for pulse Doppler processing. Supporting Linear and Convex array probes. Cine play back capability Image file archive. | Autocorrelation for color processing and FFT for pulse Doppler processing. Supporting Linear, Convex and Phase array probes. Cine play back capability Image file archive. | -- | SE with the predicated device |
| Operating Controls | Convex: ≥160mm Linear: 0~100mm | Convex: 90~305mm Linear: 20~100mm | -- | Different, see Note 2 |
| | Gain | Gain | -- | SE with the predicated device |
| | Focus | Focus | -- | SE with the predicated device |
| | Color box size/positive can be adjust | Color box size/positive can be adjust | -- | SE with the predicated device |
| | Baseline | Baseline | -- | SE with the predicated device |
Page 7 of 10
{11}
| | Cline control: Toggling freeze key | Cline control: Toggling freeze key | -- | SE with the predicated device |
| --- | --- | --- | --- | --- |
| Operation Mode | B mode B/M mode Color Doppler mode Pulse Wave mode | B mode B/M mode Color Doppler mode Pulse Wave mode | -- | SE with the predicated device |
| Transducers | Convex Array; Linear Array | Convex Array; Phase Array; Linear Array | Linear Array; Curved Linear Array; Phased Array; Intracavitary | Different, see Note 3 |
| Frequency | C10MTpro: Convex Array 3.5MHz C10MBpro: Linear Array 15MHz C10MRpro: Linear Array 19MHz C10MSpro: Linear Array 10MHz C10MXpro: Linear Array 7.5MHz | Convex Array: 3.2MHz Phase Array: 2.5MHz Linear Array: 7.5MHz | 1.0-19.0-18MHz | Different, see Note 4 |
| Probe connection to display unit | Wireless connection via Wi - Fi | Wireless connection via Wi - Fi | -- | SE with the predicated device |
| Power source and power requirement | Rechargeable battery(Li-ion) DC 5V | Rechargeable battery(Li-ion) DC 5V | -- | SE with the predicated device |
| Device and Operating system for the software APP | A commercial off-the-shelf Android device (COTS) | A commercial off-the-shelf Android device (COTS) | -- | SE with the predicated device |
Page 8 of 10
{12}
| Acoustic output levels | Track 3 Ispta.3≤720W/cm2 MI≤1.9 TI≤6.0 | Track 3 Ispta.3≤720W/cm2 MI≤1.9 TI≤6.0 | Track 3 Ispta.3: 621 mW/cm2 MI:1.72 TI Value: 4.18 | SE |
| --- | --- | --- | --- | --- |
| Safety and Effectiveness | | | | |
| Patient contacting material | Per ISO 10993-5 and ISO 10993-10 | Per ISO 10993-5 and ISO 10993-10 | ISO 10993-1 | SE |
| Electrical Safety | Evaluated according to IEC 60601-1 | Evaluated according to IEC 60601-1 | IEC 60601-1 | SE |
| EMC | Evaluated according to IEC 60601-1-2 | Evaluated according to IEC 60601-1-2 | IEC 60601-1-2 | SE |
| Performance Safety | Evaluated according to IEC 60601-2-37 | Evaluated according to IEC 60601-2-37 | IEC 60601-2-37 | SE |
# Comparison in Detail(s):
# Note 1:
The subject device does not have phase array probe. The Convex array probe and Linear array probe have the same indications for use with the predicated device. And the indications are included in the subset of indications cleared on reference device. The subject device does not have Intracavitary.
So the differences of "Indications for use" in the table above will not raise any safety or effectiveness issue.
# Note 2:
The Depth of subject device and predicate device looks different. The probe of the subject device have higher frequency, which is to obtain clearer scan images. The higher frequency leads to a change in the scan depth. The safety of the subject device were tested through IEC 60601-2-37 and IEC 62359, the bench testing ensured the accuracy of the frequency and depth measurements.
So the differences of Depth in the table above will not raise any safety or effectiveness issue.
# Note 3:
The "Transducers" of subject device and predicate device looks different. Predicate device includes Convex Array; Phase Array; Linear Array. Subject device adds new models to predicate device, the Transducers range is covered by the predicate device. No new types of legendary equipment have been added.
Page 9 of 10
{13}
So the differences of "Transducers" in the table above will not raise any safety or effectiveness issue.
# Note 4:
C10MTpro has the similar frequency with predicated device on the same transducers. C10MXpro has the same frequency with predicated device on the same transducers. The frequency of C10MBpro, C10MRpro and C10MSpro are higher than that of predicated device, but they are within the range available on the reference device. Performance of the Subject Device will be confirmed through electrical safety, EMC and performance testing.
The Frequency of subject device and predicate device looks different. The probe of the subject device have higher frequency, which is to obtain clearer scan images. The safety of the subject device were tested through IEC 60601-2-37 and IEC 62359, the testing ensured the accuracy of the frequency and depth measurements.
So the differences of Frequency in the table above will not raise any safety or effectiveness issue.
# Note 5:
The intended use environment of subject device does not include home environment. The subject device is not designed for home use. The range of environments in which the subject device can be used is included within the use environment of predicted device.
The electrical safety test of the subject device complies with the requirements of IEC 60601-1.
So the differences of intended use environment in the table above will not raise any safety or effectiveness issue.
# VII. Summary of Testing
The original models have been successfully evaluated and cleared under K241979. 5 models are added in this submission. The new models features single probe, and the functions can be covered by the original ones. They operate at higher frequency, enabling clearer images to be obtained. The hardware design, materials, and battery used in the new models are all the same as those in the original models.
Therefore, the models submitted this time underwent EMC testing (IEC 60601-1-2:2014) and ultrasonic specific testing (IEC 60601-2-37:2024 and image performance assessment).
# IX. Conclusions
In conclusion, the technological characteristics, features, specifications, materials, mode of operation, and intended use of the subject device substantially equivalent to the predicate/reference device quoted above. The differences between the subject device and its predicate/reference device do not introduce a new intended use and do not raise different questions of safety and effectiveness. Verification and validation testing demonstrated that no adverse effects have been introduced by these differences and that the device performs as intended. It can be concluded that the subject device is substantially equivalent to the legally marketed predicate device.
Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.