Navigated LLIF Impactable Tracker

K261060 · Tyber Medical, LLC · OLO · May 29, 2026 · Neurology

Device Facts

Record IDK261060
Device NameNavigated LLIF Impactable Tracker
ApplicantTyber Medical, LLC
Product CodeOLO · Neurology
Decision DateMay 29, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2

Indications for Use

The Navigated Instruments are intended to be used during preparation and placement of Medtronic Implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Navigated Instruments are specifically designed for use with the Medtronic StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.

Device Story

Navigated LLIF Impactable Tracker; surgical instrument for spinal procedures; used with Medtronic StealthStation™ System for intraoperative navigation. Device incorporates Medtronic single-use sterile spheres to enable tracking of instrument position within surgical field relative to patient anatomy (CT/MR models, fluoroscopy, or digitized landmarks). Operated by surgeons in clinical settings (OR) to assist in precise placement of spinal implants. Provides real-time spatial tracking data to surgeon; facilitates navigation during open or minimally invasive spinal surgery; improves accuracy of instrument positioning; reduces reliance on repeated imaging.

Clinical Evidence

Bench testing and cadaveric simulation study conducted. Surgeons evaluated performance and usability of the Navigation Tracker under simulated clinical conditions. Study confirmed device performed as intended, supported successful completion of surgical workflow, and met all predefined acceptance criteria.

Technological Characteristics

Surgical grade stainless steel (ASTM F899); optional trays/cases include aluminum (ASTM B209) and silicone. Passive navigation via attachment of Medtronic single-use sterile spheres for tracking by Medtronic StealthStation™ System. Non-sterile, reusable instrument. No internal software or electronics; relies on external computer-assisted surgery system for navigation functionality.

Indications for Use

Indicated for use during spinal surgery for preparation and placement of Medtronic implants; assists surgeons in locating anatomical structures in open or minimally invasive procedures; intended for use with Medtronic StealthStation™ System in conditions where stereotactic surgery is appropriate.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION May 29, 2026 Priscila Saraiva Senior Regulatory Affairs Specialist 83 S. Commerce Way Suite 310 Bethlehem, Pennsylvania 18017 Re: K261060 Trade/Device Name: Navigated LLIF Impactable Tracker Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: March 31, 2026 Received: March 31, 2026 Dear Priscila Saraiva: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1} K261060 - Priscila Saraiva Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K261060 - Priscila Saraiva Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (303) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K261060 | | | Device Name Navigated LLIF Impactable Tracker | | | Indications for Use (Describe) The Navigated Instruments are intended to be used during preparation and placement of Medtronic Implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Navigated Instruments are specifically designed for use with the Medtronic StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K261060 # 510(k) Summary Device Trade Name: Navigated LLIF Impactable Tracker Applicant Name: Tyber Medical, LLC Contact: Ms. Priscila Saraiva Date Prepared: May 26, 2026 Common Name: Stereotaxic instrument Classification Name: Orthopedic Stereotaxic Instrument Product Code: OLO Predicate Device: K252742 Navigated Instruments Device Description: The Navigated Instruments are spine preparation instruments made of surgical grade stainless steel per ASTM F899. These instruments are specifically designed for use in procedures where the use of stereotactic surgery may be appropriate. When used with the appropriate Medtronic single-use sterile spheres the subject devices can be used as navigated surgical instruments with the Medtronic Stealth™ Systems (Medtronic computer-assisted surgery system) to track the instruments in the surgical field. Specialized cases and trays are available for optional use with the Navigated Instruments. They are manufactured from stainless steel per ASTM F899, aluminum per ASTM B209, and silicone. Ancillary instrumentation is available for optional use with the Navigated Instruments. These instruments are not specifically intended to be used with the subject device. The Navigated Instruments are provided non-sterile. Indications for Use: The Navigated Instruments are intended to be used during preparation and placement of Medtronic Implants during spinal surgery to assist the surgeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Navigated Instruments are specifically designed for use with the Medtronic StealthStation™ System, which is indicated for any medical condition in which the use of stereotactic surgery may be Page 1 of 2 {5} K261060 appropriate and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy. **Substantial Equivalence:** The subject device indications for use/intended use are the same the predicate device. **Performance Testing:** The subject Navigated LLIF Impactable Tracker is made of the same materials, have the same intended use, utilize the same principles of operation, and are similar in design to the predicate devices cleared under K252742 and K124004. Simulation testing was conducted with surgeons using cadaveric specimens to evaluate the performance and usability of the Navigation Tracker under simulated clinical conditions. The test demonstrated that the Navigation Tracker performed as intended and supported successful completion of the intended surgical workflow. All predefined acceptance criteria were met during the evaluation, confirming the device's ability to function safely and effectively for its intended use. **Conclusion:** Based on the information contained in this submission, the Navigated LLIF Impactable Tracker is substantially equivalent to the predicate device. Page 2 of 2
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...