AMSCO 700 Steam Sterilizer

K261055 · STERIS Corporation · FLE · Jun 24, 2026 · General Hospital

Device Facts

Record IDK261055
Device NameAMSCO 700 Steam Sterilizer
ApplicantSTERIS Corporation
Product CodeFLE · General Hospital
Decision DateJun 24, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6880
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The AMSCO 700 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles: [Table 1 data omitted for brevity, but included in full in source]. The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 700 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual loading. The ALUS can start a cycle automatically when equipped with the optional bar code reader.

Device Story

Steam sterilizer using saturated steam to process heat-stable healthcare products; removes air via gravity displacement or mechanical vacuum; exposes load to steam at defined time/temperature; dries load using latent heat and vacuum. Operated by technicians in central sterile processing or surgical suites. Includes Automated Load and Unload System (ALUS) with optional barcode reader for cycle initiation. Provides output via electronic control and printer. Benefits include standardized, validated sterilization of surgical instruments and fabric packs to ensure sterility assurance level (SAL) of 10^-6.

Clinical Evidence

No clinical data. Bench testing only. Performance validated against ANSI/AAMI ST8:2013(R)2018, ASME Boiler and Pressure Vessel Code, IEC 61010-1, IEC 61010-2-040, and IEC 60601-1-2. Software validated per IEC 62304.

Technological Characteristics

Chamber material: 316L stainless steel. Sensing: Dual element RTDs in chamber/jacket drains, pressure transducer. Energy: Steam, water, electricity, air. Connectivity: Embedded controller, printer, optional barcode reader. Sterilization: Saturated steam. Form factor: 26"x26" chamber in various depths. Software: Embedded controller.

Indications for Use

Indicated for sterilization of heat and moisture-stable materials in healthcare facilities, including wrapped instrument trays, fabric packs, and immediate use steam sterilization (IUSS) for single unwrapped trays.

Regulatory Classification

Identification

A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION June 24, 2026 STERIS Corporation Karly Eitzman Regulatory Affairs Specialist 5960 Heisley Rd. Mentor, Ohio 44060 Re: K261055 Trade/Device Name: AMSCO 700 Steam Sterilizer Regulation Number: 21 CFR 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: Class II Product Code: FLE, PEC Dated: March 31, 2026 Received: March 31, 2026 Dear Karly Eitzman: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K261055 - Karly Eitzman Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K261055 - Karly Eitzman Page 3 Sincerely, **STEPHEN A. ANISKO -S** Digitally signed by STEPHEN A. ANISKO -S Date: 2026.06.24 19:55:49 -04'00' Stephen Anisko Acting Assistant Director DHT4C: Division of Infection Control Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. 510(k) Number (if known) K261055 Device Name AMSCO 700 Steam Sterilizer Indications for Use (Describe) The AMSCO 700 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles: Table 1. AMSCO 700 Steam Sterilizer sterilization cycles and cycle values | Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Maximum Recommended Load | | --- | --- | --- | --- | --- | | Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. Refer to Table 2 for recommended quantities. | | Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each and Fabric Packs. See Table 2 for maximum quantities. | | Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack. | | Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. Refer to Table 2 for recommended quantities. | | Prevac | 270°F (132°C) | 10 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. Refer to Table 2 for recommended quantities. | | Prevac-IUSS | 270°F (132°C) | 4 minutes | 1 minute | Immediate use – single unwrapped tray | | Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. Refer to Table 2 for recommended quantities. | | Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. Refer to Table 2 for recommended quantities. | | Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A | | DART | 270°F (132°C) | 3.5 minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack | | Leak Test | N/A | N/A | N/A | N/A | The maximum quantities of wrapped instrument trays or fabric packs for each proposed model are as follows: Table 2. AMSCO 700 Steam Sterilizer maximum load per sterilizer size | Sterilizer Size | Wrapped Instrument Trays | Fabric Packs | | --- | --- | --- | | 26" x 26" x 39" | 9 | 12 | | 26" x 26" x 51" | 12 | 16 | | 26" x 26" x 63" | 15 | 20 | The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 700 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual loading. The ALUS can start a cycle automatically when equipped with the optional bar code reader. Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) CONTINUE ON A SEPARATE PAGE IF NEEDED. FORM FDA 3881 (8/23) Page 1 of 2 PSC Publishing Services (301) 443-6740 EF {4} This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/23) Page 2 of 2 {5} STERIS # **510(k) Summary** **For** **AMSCO 700 Steam Sterilizer** **K261055** STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No.: (440) 357-9198 Contact: Karly Eitzman Regulatory Affairs Specialist Phone: (440) 514-4512 Email: Karly_Eitzman@steris.com Summary Date: March 31, 2026 5960 Heisley Road • Mentor, OH 44060-1834 • USA • 440-354-2600 March 31, 2026 Page 1 of 5 {6} # **STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION AMSCO 700 Steam Sterilizer**--- # **1. Device Name** | Trade Name: | AMSCO 700 Steam Sterilizer | | --- | --- | | Device Class: | Class II | | Common/Usual Name: | Steam Sterilizer | | Classification Name: | Sterilizer, Steam Sterilizer Automated Loading System | | Classification Number: | 21 CFR 880.6880 | | Product Code: | FLE, PEC | # **2. Predicate Device** K232485 AMSCO 600 Steam Sterilizer # **3. Description of Device** The AMSCO 700 Steam Sterilizer uses saturated steam, generated from a house steam utility (e.g. boiler system) or from a steam generator, to sterilize heat-stable health care products. The sterilizer accomplishes this by removing the air in the chamber, exposing the load to saturated steam for a defined combination of time and temperature, and drying the load. Removal of air from the chamber occurs using either of two methods, gravity displacement or mechanical vacuum. Once the air removal phase is completed, the sterilizer progresses to the steam exposure phase. During the steam exposure phase, every surface of the load is exposed to saturated steam for a defined combination of time and temperature. Once the steam exposure phase is completed, steam is removed from the chamber, and the load is dried using the latent heat in the load and the vacuum pump. The sterilizers are generally operated by technicians in a central service or sterile processing department of healthcare facilities. Sterilizers may also be located in a surgical suite to allow for Immediate Use Steam Sterilization (IUSS) for instances where an instrument is needed immediately for a procedure (e.g. after an instrument has been dropped and there is no replacement readily available). Standard practices for use of sterilizers in health care facilities are provided by various organizations (e.g. ANSI/AAMI ST79). The Automated Load and Unload System (ALUS) is used with the AMSCO 700 Steam Sterilizer's existing transfer carriages and loading carts. It consists of a conveyor system which attaches to the load and/or unload ends of the steam sterilizer. It has a series of barcode labels which correspond to pre-programmed cycles and an optional scanner, which when fitted to the system will communicate to the sterilizer which cycle to initiate. ---March 31, 2026 Page 2 of 5 {7} # **STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION AMSCO 700 Steam Sterilizer** # **4. Intended Use/Indications for Use** The AMSCO 700 Steam Sterilizers are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are equipped with the following factory-programmed cycles: **Table 1. AMSCO 700 Steam Sterilizer sterilization cycles and cycle values** | Cycles | Sterilize Temperature | Sterilize Time | Dry Time | Maximum Recommended Load | | --- | --- | --- | --- | --- | | Prevac | 270°F (132°C) | 4 minutes | 20 minutes | Fabric Packs. *Refer to Table 2 for recommended quantities.* | | Prevac | 270°F (132°C) | 4 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each and Fabric Packs. *See Table 2 for maximum quantities.* | | Prevac | 270°F (132°C) | 4 minutes | 5 minutes | Single Fabric Pack. | | Prevac | 275°F (135°C) | 3 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. *Refer to Table 2 for recommended quantities.* | | Prevac | 270°F (132°C) | 10 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. *Refer to Table 2 for recommended quantities.* | | Prevac-IUSS | 270°F (132°C) | 4 minutes | 1 minute | Immediate use – single unwrapped tray | | Gravity | 270°F (132°C) | 15 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. *Refer to Table 2 for recommended quantities.* | | Gravity | 250°F (121°C) | 30 minutes | 30 minutes | Double wrapped instrument trays, maximum weight 25 lbs. (11.3 kg) each. *Refer to Table 2 for recommended quantities.* | | Warm-Up | 270°F (132°C) | 3 minutes | 1 minute | N/A | | DART | 270°F (132°C) | 3.5 minutes | 1 minute | Bowie-Dick Test Pack, DART Test Pack | | Leak Test | N/A | N/A | N/A | N/A | The maximum quantities of wrapped instrument trays or fabric packs for each proposed model are as follows: **Table 2. AMSCO 700 Steam Sterilizer maximum load per sterilizer size** | Sterilizer Size | Wrapped Instrument Trays | Fabric Packs | | --- | --- | --- | | 26' x 26' x 39' | 9 | 12 | | 26' x 26' x 51' | 12 | 16 | | 26' x 26' x 63' | 15 | 20 | The Automated Load and Unload System (ALUS) provides semi-automated loading and unloading from an AMSCO 700 steam sterilizer when a cycle is complete. Alternatively, the ALUS may also be used to provide automatic unloading only in combination with manual loading. The ALUS can start a cycle automatically when equipped with the optional bar code reader. March 31, 2026 Page 3 of 5 {8} # **STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION AMSCO 700 Steam Sterilizer** # **5. Technological Characteristics Comparison** **Table 3. Device Comparison Table for the AMSCO 700 Steam Sterilizer and the predicate device, the AMSCO 600 Steam Sterilizer** | Feature | Predicate Device (K232485) AMSCO 600 Steam Sterilizer | Proposed Device AMSCO 700 Steam Sterilizer | Comparison | | --- | --- | --- | --- | | **Intended Use** | The AMSCO 600 Steam Sterilizer is designed for sterilization of heat and moisture-stable materials used in healthcare facilities. | The AMSCO 700 Steam Sterilizer is designed for sterilization of heat and moisture-stable materials used in healthcare facilities. | Same, except for model number | | **Critical Process Parameters** | - Time - Chamber Temperature - Pressure | - Time - Chamber Temperature - Pressure | Same | | **Control** | Embedded Controller | Embedded Controller | Same | | **SAL** | 10^{-6} | 10^{-6} | Same | | **Sterilant** | Saturated Steam | Saturated Steam | Same | | **Utilities** | Steam, Water, Electricity, Air | Steam, Water, Electricity, Air | Same | | **Chamber Material** | 316L Stainless Steel | 316L Stainless Steel | Same | | **Nominal Chamber Size** | - 26' w x 26' h x 39' d - 26' w x 26' h x 51' d - 26' w x 26' h x 63' d | - 26' w x 26' h x 39' d - 26' w x 26' h x 51' d - 26' w x 26' h x 63' d | Same | | **Door** | 304L Stainless Steel 26' x 26' Power vertical sliding | 316L Stainless Steel 26' x 26' Power vertical sliding, assisted by counterweights | Similar, Grade of stainless steel and door control assist differs | | **Chamber Pressure Rating** | 45 psig, 300°F | 45 psig, 300°F | Same | | **Door Seal** | Steam activated door seal | Steam activated door seal | Same | | **External Process Monitors** | - Electronic Control - Printer | - Electronic Control - Printer | Same | | **Internal Process Monitors** | **Temperature** - Dual element RTD located in chamber drain - RTD located in the jacket drain - RTD located in mixing tank **Pressure** - Pressure transducer in chamber | **Temperature** - Dual element RTD located in chamber drain - RTD located in the jacket drain - RTD located in mixing tank **Pressure** - Pressure transducer in chamber | Same | | **Performance** | Meets ANSI/AAMI ST8:2013 | Meets ANSI/AAMI ST8:2013(R)2018 | Same, standard was reaffirmed in 2018 | | **Accessories** | BI, CI, Pouches, Trays, Wraps, Tape, Containers, Shelves, Loading Equipment, automated loading system | BI, CI, Pouches, Trays, Wraps, Tape, Containers, Shelves, Loading Equipment, automated loading system | Same | | **Test Cycles** | Warm Up, Leak Test, DART (Bowie Dick) Test | Warm Up, Leak Test, DART (Bowie Dick) Test | Same | | **Cycles** | 270°F, Prevac, 4' Full fabric pack | 270°F, Prevac, 4' Full fabric pack | Same, | March 31, 2026 Page 4 of 5 {9} # **STERIS TRADITIONAL 510(k) PREMARKET NOTIFICATION AMSCO 700 Steam Sterilizer** | Feature | Predicate Device (K232485) AMSCO 600 Steam Sterilizer | Proposed Device AMSCO 700 Steam Sterilizer | Comparison | | --- | --- | --- | --- | | | 270°F, Prevac, 4' Full tray 270°F, Prevac, 4' One fabric pack 270°F, Prevac, 4' IUSS 275°F, Prevac, 3' Full fabric 250°F, Gravity, 30' Full tray 270°F, Prevac, 10' Full tray | 270°F, Prevac, 4' Full tray 270°F, Prevac, 4' One fabric pack 270°F, Prevac, 4' IUSS 275°F, Prevac, 3' Full fabric 270°F, Gravity, 15' Full tray 250°F, Gravity, 30' Full tray 270°F, Prevac, 10' Full tray | except for additional Gravity cycle at 270°F | | **Full Loads** | - 39': 9, 25-lb double wrapped trays or 12, fabric packs - 51': 12, 25-lb double wrapped trays or 16, fabric packs - 63': 15, 25-lb double wrapped trays or 20, fabric packs | - 39': 9, 25-lb double wrapped trays or 12, fabric packs - 51': 12, 25-lb double wrapped trays or 16, fabric packs - 63': 15, 25-lb double wrapped trays or 20, fabric packs | Same | The proposed device has the same intended use, conditions of use, and principles of operation as the predicate. Although there are minor differences in technological characteristics, they do not raise different concerns of safety or effectiveness, and the provided performance test methods and performance data demonstrate the performance of the proposed device to achieve the intended use is equivalent to the predicate. # **6. Summary of Nonclinical Tests** **Table 4. Summary of Nonclinical Tests** | Test | Acceptance Criteria | Conclusion | | --- | --- | --- | | Sterilizer Performance | Meets requirements of ANSI/AAMI ST8 | Pass | | Pressure Vessel Safety | Meets requirements of ASME Boiler and Pressure Vessel Code, VIII – Div. 1 | Pass | | General Electrical Safety | Meets requirements of IEC 61010-1 | Pass | | Sterilizer Electrical Safety | Meets requirements of IEC 61010-2-040 | Pass | | Electromagnetic Compatibility | Meets requirements of IEC 60601-1-2 with IEC TS 60601-4-2 | Pass | | Software Validation | Meets requirements of IEC 62304 | Pass | # **7. Conclusion** Based on the intended uses, technological characteristics and non-clinical performance data, the AMSCO 700 Steam Sterilizer is as safe, as effective and performs as well or better than the legally marketed predicate device, the AMSCO 600 Steam Sterilizer (K232485), Class II (21 CFR 880.6860), product code FLE. March 31, 2026 Page 5 of 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...