K261039 · Abbott Point of Care, Inc. · CGL · Jun 26, 2026 · Clinical Chemistry
Device Facts
Record ID
K261039
Device Name
i-STAT Crea cartridge with the i-STAT 1 System
Applicant
Abbott Point of Care, Inc.
Product Code
CGL · Clinical Chemistry
Decision Date
Jun 26, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1225
Device Class
Class 2
Indications for Use
The i-STAT Crea cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of creatinine in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Device Story
i-STAT Crea cartridge used with i-STAT 1 handheld analyzer; measures creatinine in arterial, venous, or capillary whole blood. System uses microfabricated electrochemical sensors; creatinine hydrolyzed to creatine via creatinine amidohydrolase; creatine hydrolyzed to sarcosine via creatine amidinohydrolase; sarcosine oxidation produces H2O2; H2O2 oxidized at platinum electrode to generate current proportional to creatinine concentration. Used in point-of-care or clinical laboratory settings by trained medical professionals. User applies 65 µL blood to cartridge; inserts into analyzer. Analyzer automates fluid movement, timing, and heating; displays quantitative result. Provides rapid renal function assessment to support clinical decision-making for renal disease diagnosis and dialysis monitoring.
Clinical Evidence
Bench testing only. Precision evaluated via multi-site, multi-day studies (n=90 per level) and whole blood precision studies. Linearity confirmed over 0.2–20.0 mg/dL range (R²=0.999). Interference testing performed per CLSI EP07-ED3. Method comparison (n=364 arterial, 168 venous, 168 capillary) against comparative methods showed high correlation (r=1.00). Matrix comparison (n=354) confirmed equivalence between non-anticoagulated and heparinized samples.
Indicated for patients requiring quantitative creatinine measurement in arterial, venous, or capillary whole blood for diagnosis/treatment of renal disease or monitoring renal dialysis. No specific age or gender restrictions noted.
Regulatory Classification
Identification
A creatinine test system is a device intended to measure creatinine levels in plasma and urine. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
Predicate Devices
Nova StatSensor Creatinine Hospital Meter System (K171059)
Submission Summary (Full Text)
{0}
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
# DECISION SUMMARY
ASSAY ONLY
# I Background Information:
A 510(k) Number
K261039
B Applicant
Abbott Point of Care, Inc.
C Proprietary and Established Names
i-STAT Crea cartridge with the i-STAT 1 System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CGL | Class II | 21 CFR 862.1225 - Creatinine Test System | CH - Clinical Chemistry |
# II Submission/Device Overview:
A Purpose for Submission:
Modification to a previously cleared device
B Measurand:
Creatinine
C Type of Test:
Quantitative amperometric assay
# III Intended Use/Indications for Use:
A Intended Use(s):
See Indications for Use below.
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
# **B Indication(s) for Use:**
The i-STAT Crea cartridge with the i-STAT 1 System is intended for use in the in vitro quantification of creatinine in arterial, venous or capillary whole blood in point of care or clinical laboratory settings.
Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes.
# **C Special Conditions for Use Statement(s):**
Rx - For Prescription Use Only
For point-of-care or clinical laboratory settings
# **D Special Instrument Requirements:**
i-STAT 1 Analyzer
# **IV Device/System Characteristics:**
# **A Device Description:**
The single-use, white disposable i-STAT Crea cartridge is used with the i-STAT 1 analyzer as part of the i-STAT 1 System to measure creatinine in arterial, venous, or capillary whole blood. The i-STAT Crea cartridges contain the required sensors, a fluid pack (calibrant pouch), a sample entry well and closure, fluid channels, waste chamber, and the necessary mechanical features for controlled fluid movement within the cartridge. The i-STAT 1 analyzer interacts with the i-STAT Crea cartridge to move fluid across the sensors and generate a quantitative result, provided as a direct readout of numerical values. Cartridges require two to three drops of whole blood applied to the cartridge.
Reagents
Each i-STAT Crea cartridge contains a ground electrode and amperometric sensor for the measurement of specific analytes. It also contains a buffered aqueous calibrant solution with known concentrations of analytes. A list of reactive ingredients relevant for the i-STAT Crea cartridge is shown below:
| Sensor | Reactive Ingredient | Biological Source | Minimum Quantity |
| --- | --- | --- | --- |
| Crea | Creatinine | NA | 158.4 umol/L |
| | Creatine Amidinohydrolase | microbial | 0.01 IU |
| | Creatinine Amidohydrolase | microbial | 0.02 IU |
| | Sarcosine Oxidase | microbial | 0.001 IU |
# **B Principle of Operation:**
The enabling technology for the i-STAT 1 System is in the microfabricated electrochemical sensors located in the single-use disposable i-STAT cartridges. Creatinine on the i-STAT cartridge is measured amperometrically. Creatinine in the sample is hydrolyzed to creatine in a reaction catalyzed by the enzyme creatinine amidohydrolase. Creatine is then hydrolyzed to
K261039 - Page 2 of 14
{2}
sarcosine in a reaction catalyzed by the enzyme creatine amidinohydrolase. The oxidation of sarcosine, catalyzed by the enzyme sarcosine oxidase, produces hydrogen peroxide (H₂O₂). The liberated hydrogen peroxide is oxidized at the platinum electrode to produce a current which is proportional to the sample creatinine concentration.
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
Nova StatSensor Creatinine Hospital Meter System
# B Predicate 510(k) Number(s):
K171059
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K261039 | K171059 |
| --- | --- | --- |
| Device Trade Name | Creatinine test in the i-STAT Crea cartridge with the i-STAT 1 System | Nova StatSensor Creatinine Hospital Meter System |
| **General Device Characteristic Similarities** | | |
| Intended Use/Indications For Use | In vitro quantification of creatinine in arterial, venous, or capillary whole blood in point of care or clinical laboratory settings. | Same |
| Principle of Measurement | Amperometric measurement of oxidized hydrogen peroxide | Same |
| Traceability | Traceable to NIST SRM967 | Same |
| **General Device Characteristic Differences** | | |
| Reportable Range | 0.2 – 20.0 mg/dL | 0.3 – 12.0 mg/dL |
| Sample Volume | 65 μL | 1.2 μL |
# VI Standards/Guidance Documents Referenced:
CLSI EP05-A3: Evaluation of Precision of Quantitative Measurement Procedures; Approved Guideline – Third Edition.
K261039 - Page 3 of 14
{3}
CLSI EP06-Ed2: Evaluation of Linearity of Quantitative Measurement Procedures – Second Edition.
CLSI EP07-ED3: Interference Testing in Clinical Chemistry – Third Edition.
CLSI EP09c: Measurement Procedure Comparison and Bias Estimation Using Patient Samples. 3rd Edition.
CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition.
CLSI EP35: Assessment of Equivalence or Suitability of Specimen Types for Medical Laboratory Measurement Procedures. 2nd Edition.
CLSI EP37: Supplemental Tables for Interference Testing in Clinical Chemistry. 2nd Edition.
CLSI EP39: A Hierarchical Approach to Selecting Surrogate Samples for the Evaluation of In Vitro Medical Laboratory Tests. 1st Edition.
## VII Performance Characteristics (if/when applicable):
### A Analytical Performance:
#### 1. Precision/Reproducibility:
Precision studies were conducted following the recommendations in CLSI EP05-A3
##### *Internal Precision Study (Aqueous Control Materials)*
The precision of the i-STAT Creatinine test in the *i-STAT Crea* cartridge with the *i-STAT 1 System* was evaluated using five (5) levels of aqueous control materials. The study was conducted using thirteen (13) analyzers and one (1) test cartridge lot over at least 20 days at one (1) site. Repeatability, between-run, between-day, and within-laboratory precision were estimated for each level. The results are shown below.
| 20-day precision of the i-STAT Crea cartridge on the i-STAT 1 analyzer | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Test (units) | Fluid Level | N | Mean | Repeatability | | Between-Run | | Between-Day | | Within-Laboratory | |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Crea (mg/dL) | L1 | 82 | 16.46 | 0.363 | 2.20 | 0.005 | 0.03 | 0.245 | 1.49 | 0.438 | 2.66 |
| | L2 | 83 | 4.38 | 0.101 | 2.31 | 0.033 | 0.74 | 0.039 | 0.88 | 0.113 | 2.58 |
| | L3 | 83 | 1.67 | 0.034 | 2.06 | 0.000 | 0.00 | 0.015 | 0.88 | 0.037 | 2.24 |
| | L4 | 82 | 0.46 | 0.019 | 4.22 | 0.004 | 0.89 | 0.007 | 1.49 | 0.021 | 4.57 |
| | L5 | 81 | 0.23 | 0.020 | 8.31 | 0.009 | 3.80 | 0.003 | 1.46 | 0.022 | 9.25 |
Levels tested for creatinine have more than 80 replicates due to additional cartridges run.
K261039 - Page 4 of 14
{4}
# Lot-to-Lot Precision
The precision of the i-STAT Creatinine test in the i-STAT Crea cartridge with the i-STAT 1 System across multiple cartridge lots was evaluated at one site. The study was conducted using three (3) lots of i-STAT Crea cartridges, each lot tested in five (5) replicates over five (5) non-consecutive days with aqueous control materials using 15 i-STAT 1 analyzers. The results of the lot-to-lot precision study for the i-STAT Crea cartridge on the i-STAT 1 System are shown below.
| Lot-to-Lot Precision of i-STAT Crea Cartridge on the i-STAT 1 Analyzer | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Test (units) | Level | N | Mean | Repeatability | | Between-Lot | | Between-Day | | Within-Laboratory | |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Crea (mg/dL) | L1 | 75 | 16.74 | 0.285 | 1.7 | 0.217 | 1.3 | 0.148 | 0.9 | 0.388 | 2.3 |
| | L2 | 75 | 4.44 | 0.067 | 1.5 | 0.061 | 1.4 | 0.023 | 0.5 | 0.094 | 2.1 |
| | L3 | 75 | 1.66 | 0.028 | 1.7 | 0.030 | 1.8 | 0.006 | 0.4 | 0.042 | 2.5 |
| | L4 | 75 | 0.44 | 0.019 | 4.3 | 0.013 | 2.9 | 0.004 | 0.9 | 0.023 | 5.3 |
# Instrument-to-Instrument Precision
The precision of the i-STAT Creatinine test in the i-STAT Crea cartridge with the i-STAT 1 System across multiple i-STAT 1 analyzers was evaluated. The study was conducted using three (3) i-STAT 1 analyzers and one (1) lot of i-STAT Crea cartridges tested in five (5) replicates over five (5) non-consecutive days with aqueous control materials. The results of the instrument-to-instrument precision study for the i-STAT Crea cartridge on the i-STAT 1 System are shown below.
| Instrument-to-Instrument Precision of i-STAT Crea Cartridge on the i-STAT 1 Analyzer | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Test (units) | Level | N† | Mean | Repeatability | | Between-Instrument | | Between-day | | Within-Laboratory | |
| | | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| Crea (mg/dL) | L1 | 77 | 16.39 | 0.370 | 2.26 | 0.000 | 0.0 | 0.198 | 1.21 | 0.420 | 2.56 |
| | L2 | 75 | 4.37 | 0.070 | 1.6 | 0.017 | 0.4 | 0.038 | 0.9 | 0.081 | 1.90 |
| | L3 | 75 | 1.67 | 0.030 | 1.78 | 0.000 | 0.0 | 0.015 | 0.92 | 0.033 | 2.01 |
| | L4 | 75 | 0.46 | 0.023 | 5.0 | 0.007 | 1.5 | 0.000 | 0.0 | 0.024 | 5.22 |
| | L5 | 76 | 0.23 | 0.026 | 11.6 | 0.003 | 1.3 | 0.003 | 1.3 | 0.027 | 11.70 |
†The N for some levels tested have more than 75 replicates due to additional cartridges run.
# Multi-site, multi-day precision (aqueous materials)
Multi-site, multi-day precision testing was performed at three (3) point of care sites using a panel of aqueous material containing five levels of creatinine. At each site, each level was tested once per day by two (2) operators for five (5) days on six (6) i-STAT 1 analyzers (per site) using one (1) lot of i-STAT Crea cartridges. Repeatability, between-day, between-operator, within-site, between-site variance components, and reproducibility were calculated for all sites combined and are provided below.
K261039 - Page 5 of 14
{5}
| Multi-site multi-day precision of the i-STAT Crea cartridge on the i-STAT 1 analyzer | | | | | | | | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Level | N | Mean | Repeatability | | Between-Day | | Between-Operator | | Within-Site | | Between-Site | | Reproducibility | |
| | | | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| L1 | 90 | 16.33 | 0.415 | 2.5 | 0.263 | 1.6 | 0.000 | 0.0 | 0.492 | 3.0 | 0.520 | 3.2 | 0.715 | 4.4 |
| L2 | 90 | 3.71 | 0.081 | 2.2 | 0.045 | 1.2 | 0.029 | 0.8 | 0.097 | 2.6 | 0.084 | 2.3 | 0.128 | 3.5 |
| L3 | 90 | 1.02 | 0.037 | 3.6 | 0.023 | 2.2 | 0.000 | 0.0 | 0.043 | 4.2 | 0.020 | 2.0 | 0.047 | 4.6 |
| L4 | 90 | 0.49 | 0.025 | 5.1 | 0.014 | 2.9 | 0.000 | 0.0 | 0.029 | 5.8 | 0.003 | 0.6 | 0.029 | 5.9 |
| L5 | 90 | 0.37 | 0.043 | 11.6 | 0.007 | 1.9 | 0.000 | 0.0 | 0.043 | 11.7 | 0.021 | 5.7 | 0.048 | 13.0 |
# Precision (whole blood) at Point of Care Sites
The precision of the i-STAT Creatinine test in the i-STAT Crea cartridge on the i-STAT 1 System was evaluated using whole blood specimens in point-of-care settings at multiple clinical sites. At each site, anticoagulated arterial, venous, or capillary whole blood specimens were tested in duplicate. For each sample type, the specimens were grouped into subintervals of the creatinine test reportable range for analysis. The results are summarized below.
| Whole blood precision of venous, arterial, and capillary whole blood for the i-STAT Crea cartridge on the i-STAT 1 analyzer | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Test (units) | Sample Type | Sample Range | N | Mean | SD | %CV |
| Crea (mg/dL) | Venous Whole Blood | 0.2-1.0 | 66 | 0.76 | 0.040 | 5.23 |
| | | >1.0-2.0 | 25 | 1.38 | 0.040 | 2.89 |
| | | >2.0-7.5 | 32 | 3.92 | 0.070 | 1.78 |
| | | >7.5-12.0 | 16 | 9.68 | 0.108 | 1.11 |
| | | >12.0-20.0 | 27 | 15.76 | 0.292 | 1.85 |
| | Arterial Whole Blood | 0.2-1.0 | 146 | 0.73 | 0.037 | 5.00 |
| | | >1.0-2.0 | 105 | 1.35 | 0.045 | 3.31 |
| | | >2.0-7.5 | 82 | 3.68 | 0.067 | 1.83 |
| | | >7.5-12.0 | 14 | 9.25 | 0.154 | 1.66 |
| | | >12.0-20.0 | 14 | 15.07 | 0.280 | 1.86 |
| | Capillary Whole Blood | 0.2-1.0 | 94 | 0.72 | 0.048 | 6.69 |
| | | >1.0-2.0 | 24 | 1.32 | 0.054 | 4.10 |
| | | >2.0-7.5 | 26 | 4.07 | 0.105 | 2.57 |
| | | >7.5-12.0 | 10 | 9.26 | 0.297 | 3.20 |
| | | >12.0-20.0 | 7 | 15.30 | 0.245 | 1.60 |
# 2. Linearity:
The linearity study was conducted following the recommendations in CLSI EP06-Ed2. The linearity of the i-STAT Creatinine test in the i-STAT Crea cartridge with the i-STAT 1 System was evaluated using five lots of i-STAT CREA cartridges, 30 i-STAT 1 analyzers, and venous whole blood collected with lithium heparin from one healthy donor, altered to 11 creatinine levels ranging from <0.2 mg/dL to > 20.0 mg/dL. Each sample level was tested in replicates of three for each of the five i-STAT CREA cartridge lots (for a total of 15 results per level) on i-STAT 1 analyzers. The data was analyzed using weighted least square linear regression. The observed non-linearity within the reportable range did not exceed ±7.23%.
K261039 - Page 6 of 14
{6}
| Regression summary for the creatinine test in the i-STAT Crea cartridge on the i-STAT 1 analyzer | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Test | Units | Reportable Range | Range Tested | Slope | Intercept | R² |
| Creatinine | mg/dL | 0.2 – 20.0 | 0.14 – 22.92 | 0.999 | 0.024 | 0.999 |
The results support the linearity of the i-STAT Creatinine test in the i-STAT Crea cartridge on the i-STAT 1 System over the reportable range of 0.2 – 20.0 mg/dL.
### 3. Analytical Specificity/Interference:
#### Interference
The analytical specificity of the i-STAT Creatinine test in the i-STAT Crea cartridge with the i-STAT 1 System was evaluated by conducting interference testing following the recommendations in CLSI EP07 ED3. Interference from potentially interfering substances was assessed using lithium heparin venous whole blood samples spiked to two concentrations of low (0.5 – 1.1 mg/dL) and high (1.5 – 2.5 mg/dL) creatinine. Each low and high sample was further divided into two aliquots: control (with no added interferent) and test (with added interferent). Each sample was measured in replicates of approximately 10 using four lots of the i-STAT Crea cartridge. A substance was identified as an interferent if the difference in the mean between the control and test sample was outside of the predefined allowable error, for Creatinine: ± greater of 0.2 mg/dL, or 10% of the control mean or median (mg/dL).
The following table lists the concentrations of each substance at which no significant interference was found.
| Substance¹ | Highest concentration at which no interference was observed |
| --- | --- |
| Acetaldehyde² | 0.2 mg/dL |
| Acetaminophen | 15.6 mg/dL |
| N-Acetyl-L-Cysteine | 15.0 mg/dL |
| L-Ascorbic Acid | 5.25 mg/dL |
| β-Hydroxybutyric Acid² | 62.46 mg/dL |
| Bicarbonate² | 294 mg/dL |
| Bilirubin, Conjugated | 40 mg/dL |
| Bilirubin, Unconjugated | 40 mg/dL |
| Calcium Chloride | 20 mg/dL |
| Cefoxitin | 697 mg/dL |
| Creatine² | 5.01 mg/dL |
| Dopamine Hydrochloride | 0.0621 mg/dL |
| Formaldehyde | 0.399 mg/dL |
| Intralipid 20% | 3527 mg/dL |
| Lactate (Lithium Lactate) | 90 mg/dL |
| Methyldopa | 2.25 mg/dL |
| pH | 8.0 pH units |
| Pyruvate (Lithium Pyruvate) | 5 mg/dL |
| Salicylate (Lithium Salicylate) | 2.86 mg/dL |
| Triglyceride | 1500 mg/dL |
K261039 - Page 7 of 14
{7}
| Substance^{1} | Highest concentration at which no interference was observed |
| --- | --- |
| Uric Acid | 23.5 mg/dL |
$^{1}$ The compound tested to evaluate the interfering substance is presented in parenthesis.
$^{2}$ The test concentration for this substance is not included in CLSI guideline EP37 1$^{st}$ edition
| Interfering Substance | Interferent Testing Concentration mmol/L | Interferent Testing Concentration mg/dL | Interference | Comment |
| --- | --- | --- | --- | --- |
| Bromide (Lithium Bromide) | 2.5 | 21.7 | No | Refer to comment below. |
| | 37.5 | 325.7 | Yes | Use another method. Refer to comment below. |
Labeling includes the following comment:
- Bromide has been tested at two levels: a toxic level of 37.5 mmol/L and a therapeutic plasma concentration level of 2.5 mmol/L. The latter is the peak plasma concentration associated with halothane anesthesia, in which bromide is released. Abbott has not identified a therapeutic condition that would lead to levels consistent with 37.5 mmol/L. Bromide increases variability in the i-STAT Crea cartridge which leads to inconsistent results, Abbott recommends using another method for bromide concentration greater than 2.5 mmol/L.
For those substances that on initial screening were found to interfere, dose response testing was conducted to establish the concentration limit below which no significant interference is expected. The results are summarized in the table below:
| Interfering Substance | Creatinine Level | Interferent Testing Concentration | Highest Concentration Tested without Interference | Interference Results |
| --- | --- | --- | --- | --- |
| Glycolic Acid | High | 10.0 mmol/L | 4.9 mmol/L | Decreased results at > 4.9 mmol/L |
| Hemoglobin^{3} | High | 10 g/L | 8 g/L | Decreased results at > 8 g/L |
| Hydroxyurea | High | 0.405 mmol/L | 0.02 mmol/L | Increased results > 0.02 mmol/L |
| | Low | | 0.02 mmol/L | |
| Nithiodote (Sodium Thiosulfate) | Low | 16.7 mmol/L | 14.5 mmol/L | Increased results at > 14.5 mmol/L |
$^{3}$ The test concentration denotes the quantity of free hemoglobin added to the whole blood test sample.
Labeling includes the following comments:
- Glycolic acid may be present in patient samples as the primary toxic metabolite of ethylene glycol ingestion. Glycolic acid concentrations of approximately 3–10
K261039 - Page 8 of 14
{8}
mmol/L are consistent with clinically significant ethylene glycol exposure and may be associated with evolving metabolic acidosis. In a study of 35 patients who had ingested ethylene glycol, initial glycolic acid concentrations of 0 to 38 mmol/L corresponded to ethylene glycol levels of 0.97 – 130.6 mmol/L¹
¹ Porter, W. H., Crellin, M., Rutter, P. W., and Oeltgen, P. (2000) Interference by glycolic acid in the Beckman synchron method for lactate: a useful clue for unsuspected ethylene glycol intoxication. Clin Chem. 46, 874-875
- Hydroxyurea is a DNA synthesis inhibitor used in the treatment of sickle cell anemia, HIV infection, and various types of cancer. The malignancies that it is used to treat include melanoma, metastatic ovarian cancer, and chronic myelogenous leukemia. It is also used in the treatment of polycythemia vera, thrombocythemia, and psoriasis. At typical doses ranging from 500 mg to 2 g/day, concentrations of hydroxyurea in a patient's blood may be sustained at approximately 100 to 500 μmol/L. Higher concentrations may be observed soon after dosing or at higher therapeutic doses.
- Nithiodote (Sodium Thiosulfate) is indicated for the treatment of acute cyanide poisoning. The journal article titled "Falsely increased chloride and missed anion gap elevation during treatment with sodium thiosulfate" indicated that sodium thiosulfate could be used in the treatment of calciphylaxis indicating that "the highest concentration likely to be seen in plasma [is] after infusion of a 12.5 g dose of sodium thiosulfate pentahydrate. Assuming that the 12.5 g dose of sodium thiosulfate pentahydrate is distributed in a typical blood volume of 5 L with a hematocrit of 40%, the peak sodium thiosulfate plasma concentration expected is 16.7 mmol/L.¹⁵
¹⁵ Wendroth, S. M., Heady, T. N., Haverstick, D. M., Bachmann, L. M., Scott, M. G., Boyd, J. C., and Bruns, D. E. (2014) Falsely increased chloride and missed anion gap elevation during treatment with sodium thiosulfate. Clin Chim Acta. 431, 77-79
- The sponsor provided information to support the labeling limitation that pCO2 is a factor that may affect creatinine results. The labeling states;
"The following manual calculation should be applied for creatinine results >2.0 mg/dL if the pCO2 value is suspected to be abnormal (i.e. <20 or >60 mmHg): Creatinine corrected = creatinine * (1 + 0.0025 * (PCO2 – 40)), where pCO2 is measured in mmHg.
No correction is required for creatinine results ≤2.0 mg/dL."
### Altitude Study
Studies were conducted to evaluate the effect of altitude on the i-STAT Creatinine test in the i-STAT Crea cartridge. Evaluation of the study data showed the predicted bias observed at the medical decision levels (0.6, 1.6, and 6.0 mg/dL) between the creatinine results obtained from the i-STAT Crea cartridges tested at a simulated altitude of 10,000 feet above sea level and at sea level were between 0.5 and 3.5%. The results of the study support the claim of equivalent performance between sea level (control condition) and at an altitude of approximately 10,000 feet (ft) above sea level (test condition) when using the i-STAT Creatinine test.
K261039 - Page 9 of 14
{9}
# Oxygen Sensitivity
Studies were conducted to evaluate the effect of oxygen on the i-STAT Creatinine test in the i-STAT Crea cartridge. Evaluation of the study data showed the difference in creatinine results at the oxygen levels evaluated were within 0.2 mg/dL or 10%. The results of the study support the claim that no impact on performance was found at oxygen levels within 24 - 605 mmHg when using the i-STAT Creatinine test.
# 4. Assay Reportable Range:
The claimed reportable range for the i-STAT Crea cartridge is 0.2 – 20.0 mg/dL.
# 5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
# Traceability
The creatinine assay on the i-STAT Crea cartridge with the i-STAT 1 System and creatinine values assigned to i-STAT controls and calibration verification materials are traceable to the U.S. National Institute of Standards and Technology (NIST) standard reference material NIST SRM 967.
# Sample Stability
The sponsor has provided information to support the following sample stability claims in their labeling:
- Balanced heparin anticoagulant or lithium heparin anticoagulant whole blood: 30 minutes
- Non-anticoagulant whole blood: 3 minutes at room temperature (18-30°C).
# 6. Detection Limit:
Detection limit studies were based on CLSI EP17-A2.
# Limit of Blank and Detection (LoB/LoD)
The LoB and LoD of the i-STAT Creatinine test in the i-STAT Crea cartridge were evaluated on the i-STAT 1 analyzer. The study was conducted across four days using two lots of i-STAT CREA cartridges and whole blood samples altered to achieve a blank creatinine level for LoB testing and two low creatinine levels for LoD testing. As whole blood is not stable over multiple days, it was not possible to run the same low-level sample across multiple test events. Consequently, fresh samples were prepared for each day of testing, and the analysis was performed after all days of testing were completed.
The LoB (Cr0) sample was tested using 20 i-STAT CREA cartridges from each of the two cartridge lots on i-STAT 1 analyzers for a total of 40 test results per test event. The higher of the two LoB values from each lot is reported in the labeling as the LoB.
The LoD (Cr1, Cr2) samples were tested using 20 i-STAT CREA cartridges from each of the two cartridge lots on i-STAT 1 analyzers for a total of 40 test results per sample level per test event. The LoD was calculated using a parametric data analysis for each of the two lots. The higher of the two LoD values from each lot is reported in the labeling as the LoD.
K261039 - Page 10 of 14
{10}
| **Summary of LoB and LoD results for the creatinine test in the i-STAT Crea cartridge on the i-STAT 1 analyzer** | | | |
| --- | --- | --- | --- |
| **Test** | **Units** | **LoB** | **LoD** |
| Creatinine | mg/dL | 0.07 | 0.12 |
*Limit of Quantitation (LoQ)*
The LoQ of the i-STAT Creatinine test in the i-STAT Crea cartridge was evaluated on the i-STAT 1 analyzer using two cartridge lots and whole blood that was altered to four low creatinine levels (< 0.2 mg/dL). The study was conducted across four days. Whole blood samples from four healthy donors were altered to a low creatinine level. As whole blood is not stable over multiple days, it was not possible to run the same low-level sample across multiple test events. Consequently, fresh samples were prepared for each day of testing, and the analysis was performed after all days of testing were completed. The TE estimates for each sample and cartridge lot were calculated, and the highest result obtained was 0.0725 mg/dL. The lowest level of creatinine that met the TE goal (set as 0.3 mg/dL) for each individual cartridge lot is the LoQ value for each individual lot. The determined LoQ is the highest LoQ value from the two lots.
The LoQ for the i-STAT Creatinine test was determined to be 0.13 mg/dL, which is below the lower limit of the reportable range for the i-STAT Creatinine test as shown in the table below.
| **Summary of LoQ result for the creatinine test in the i-STAT Crea cartridge on the i-STAT 1 analyzer** | | | |
| --- | --- | --- | --- |
| **Test** | **Units** | **Lower limit of the reportable range** | **LoQ** |
| Creatinine | mg/dL | 0.2 | 0.13 |
7. Assay Cut-Off:
Not applicable.
**B Comparison Studies:**
1. Method Comparison
The method comparison study for arterial, venous, and capillary whole blood specimens on the i-STAT Crea cartridge with the i-STAT 1 System was conducted following the recommendations in CLSI EP09c-ED3. Three method comparison studies were conducted under multiple clinical study protocols.
- Heparinized arterial and venous whole blood specimens collected across multiple point of care sites and collection locations were evaluated using i-STAT Crea cartridges on the i-STAT 1 analyzer against whole blood specimens tested on a comparative method. A Passing-Bablok linear regression analysis was performed using the first replicate result from the i-STAT 1 analyzer versus the mean result from the comparative method.
- Two (2) capillary specimens collected from skin puncture with balanced heparin capillary tubes from subjects 18 years or older across multiple point of care sites were
K261039 - Page 11 of 14
{11}
evaluated and analyzed in singlicate on the i-STAT Crea cartridge on the i-STAT 1 analyzer against a venous whole blood specimen from the same subject on the comparative method.
- Two (2) capillary specimens collected from heelstick puncture with balanced heparin capillary tubes from neonate subjects across multiple point of care sites were evaluated and analyzed in singlicate on the i-STAT Crea cartridge on the i-STAT 1 analyzer and on the comparative method.
A Passing-Bablok linear regression analysis was performed using the results of each sample type from the i-STAT Crea cartridge versus the comparative method results. Method comparison results for 364 arterial, 168 venous, and 168 capillary whole blood specimens are shown below.
| Method comparison results for arterial, venous, and capillary specimens | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Test | Sample Type | Comparative Method | N | Slope | Intercept | r |
| Creatinine (mg/dL) | Arterial | i-STAT CHEM8+ | 364 | 0.98 | 0.01 | 1.00 |
| | Venous | i-STAT CHEM8+ | 168 | 0.99 | 0.01 | 1.00 |
| | Capillary | epoc Blood Analysis System (capillary) or i-STAT CHEM8+ (venous) | 168 | 1.00 | 0.00 | 1.00 |
The predicted bias at the medical decision levels, 0.6, 1.6, and 6.0 mg/dL, was reviewed and found to be acceptable.
## 2. Matrix Comparison:
The matrix comparison studies were conducted following the recommendations in CLSI EP35.
To evaluate the performance of the i-STAT Creatinine test in the i-STAT Crea cartridge on the i-STAT 1 System comparing whole blood specimens collected without anticoagulant (candidate specimen) to samples collected with balanced heparin or lithium heparin anticoagulant (primary specimen), a total of 354 matched whole blood samples were tested which included native arterial (n=182) and native venous (n=155) specimens.
Two (2) native whole blood specimens (venous or arterial) were collected from each subject using one of the following sets of collection devices:
Venous or arterial specimens were collected in one (1) non-anticoagulated syringe (candidate specimen collection device) and one (1) heparin syringe (primary specimen collection device).
For some subjects, venous specimens were collected in one (1) non-anticoagulated tube (candidate specimen collection device) and one (1) evacuated heparin tube (primary specimen collection device).
Non-anticoagulated whole blood specimens were collected from 17 subjects to prepare contrived specimen pairs (candidate and primary specimens) for testing. Specimens were contrived to achieve adequate distribution of creatinine levels across the reportable range of 0.2–20.0 mg/dL. Specimens with high creatinine
K261039 - Page 12 of 14
{12}
concentrations were contrived by adding an appropriate amount of creatinine spike solution (720 mg/dL in deionized water) to the plasma layer of the non-anticoagulated blood samples. No specimens were contrived to represent low creatinine concentrations. The contrived whole blood specimens were aliquoted into two (2) collection devices as follows: A non-anticoagulated syringe was used as candidate collection device. A heparin syringe was used as primary collection device.
Candidate specimens were tested in duplicate using two (2) i-STAT Crea cartridges on two (2) i-STAT 1 analyzers within three (3) minutes of sample collection. Primary specimens were tested in duplicate using two (2) i-STAT Crea cartridges on two (2) i-STAT 1 analyzers within 30 minutes of sample collection.
A Passing-Bablok linear regression analysis was performed using the first replicate result from the candidate (y-axis) versus the first replicate result from the primary specimen (x-axis). The regression analysis results are summarized below.
| Matrix comparison results | | Arterial samples All sites | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Test (units) | N* | Candidate Specimen Range (mg/dL) | Primary Specimen Range (mg/dL) | r | Slope | Intercept |
| Creatinine (mg/dL) | 191 | 0.3-18.2 | 0.3-17.9 | 1.00 | 1.00 | 0.00 |
*Includes nine (9) contrived arterial specimens
| Matrix comparison results | | Venous samples All sites | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Test (units) | N* | Candidate Specimen Range (mg/dL) | Primary Specimen Range (mg/dL) | r | Slope | Intercept |
| Creatinine (mg/dL) | 163 | 0.4-19.2 | 0.4-18.6 | 1.00 | 1.00 | 0.00 |
*Includes eight (8) contrived venous specimens
The results of non-anticoagulated venous and arterial whole blood specimens when tested using the i-STAT Crea cartridge on the i-STAT 1 analyzer supports the claim that the performance is equivalent to anticoagulated (balanced heparin or lithium heparin) venous and arterial whole blood specimens.
### C Clinical Evidence:
#### 1. Clinical Sensitivity:
Not applicable.
#### 2. Clinical Specificity:
Not applicable.
K261039 - Page 13 of 14
{13}
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
E Expected Values/Reference Range:
The reference range for the i-STAT Creatinine test in the i-STAT Crea cartridge is based on literature reference and is listed below.
| Reference range | | | | |
| --- | --- | --- | --- | --- |
| Test | Units | Reportable Range | Reference Range | |
| | | | Arterial | Venous |
| Creatinine | mg/dL | 0.2 – 20.0 | 0.6 – 1.3^{1} | |
| | μmol/L | 18-1768 | 53-115 | |
$^{1}$ Burtis, C. A., Ashwood, E. R., and Bruns, D. E. (2006) Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, 4th Ed. St. Louis, MO: Elsevier Saunders.
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K261039 - Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.