The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide. The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminants. The SafeTrack™ magnet is detected by magnetic sensors to provide visual guidance for tracking the position of a needle with respect to ultrasound image data.
Device Story
Kit includes needle guide, probe cover, conductive gel, patient drape with elastic bands, and needle with integrated SafeTrack™ magnet; used with Accuro® 3S ultrasound system. Magnet on needle hub emits field detected by system sensors; system displays needle location relative to ultrasound image. Used by trained healthcare professionals in hospitals/clinics to facilitate needle placement; improves accuracy and guidance during percutaneous procedures. Sterile, single-use components.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing and simulated use evaluations.
Technological Characteristics
Single-use sterile kit; includes needle guide, probe cover, conductive gel, patient drape, and needle with integrated magnet. Sterilized via ethylene oxide. Complies with ISO 10993 (biocompatibility), ISO 11135 (sterilization), and ISO 11607 (packaging). Magnetic sensing principle for needle tracking.
Indications for Use
Indicated for healthcare professionals to assist in needle tip placement relative to anatomical structures using ultrasound imaging. Used in hospital and clinic settings. No specific patient population restrictions stated.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Rivanna Medical, Inc.
F. William Mauldin
Chairman, Co-founder, CEO
2400 Hunters Way
Charlottesville, Virginia 22911
July 24, 2026
Re: K261014
Trade/Device Name: Accuro® 3S Needle Guide Kit
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: ITX
Dated: June 26, 2026
Received: June 26, 2026
Dear F. William Mauldin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261014 - F. William Mauldin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261014 - F. William Mauldin
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Michael D. O'hara -S
Date: 2026.07.24 13:17:57 -04'00'
For
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K261014
Device Name
Accuro® 3S Needle Guide Kit
Indications for Use (Describe)
The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide.
The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminants.
The SafeTrack™ magnet is detected by magnetic sensors to provide visual guidance for tracking the position of a needle with respect to ultrasound image data.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
RIVANNA®
510(k) Summary- K261014
# K261014 510(k) Summary
This summary is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
Date:
Submitter:
March 27th, 2026
Rivanna Medical, Inc.
2400 Hunters Way
Charlottesville, VA 22911
Primary Contact Person:
F. William Mauldin, Jr.
CEO
Rivanna Medical, Inc.
T: 800-645-7508
Subject Device Trade Name: Accuro® 3S Needle Guide Kit
Common Name: Transducer, Ultrasonic, Diagnostic
Device Class: Class II
Product Codes: Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Legally Marketed Predicate Device:
Primary Predicate Device: K250469 Accuro® 3S Needle Guide Kit
Common/Usual Name: Transducer, Ultrasonic, Diagnostic
Classification Name: Class II
Product Code: Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX
Reference Device: K142445 Pinpoint™ GT Safety Introducer Needle
Common/Usual Name: Catheter Introducer
Classification Name: Class II
Product Code: Introducer, Catheter, 21 CFR 870.1340, DYB
Device Description:
The Accuro 3S Needle Guide Kit is designed for use with compatible ultrasound equipment, including the Accuro 3S diagnostic ultrasound imaging system. The Accuro® 3S Needle Guide Kit consists of the following single use sterile disposable components: Accuro® 3S Needle Guide with integrated probe cover, conductive ultrasound gel, a patient drape with two integrated probe elastic bands, and a needle with the integrated SafeTrack™ magnet.
RIVANNA | rivannamedical.com | info@rivannamedical.com
T: 800-645-7508 | Fax: 434-483-5720 | 2400 HUNTERS WAY, CHARLOTTESVILLE, VA 22911
{5}
510(k) Summary- K261014
The kit components are assembled into a sterilized CSR Wrap then packaged in a sterile tray. The final kit is sterilized under ethylene oxide.
These Accuro® 3S Needle Guide Kit components may be packaged together or combined separately with other sterile FDA-cleared single-use products.
Intended Use/Indications for Use:
The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide.
The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminant.
The SafeTrack™ magnet is detected by magnetic sensors to provide visual guidance for tracking the position of a needle with respect to ultrasound image data.
Comparison to Predicate Device(s):
| | **Accuro® 3S Needle Guide Kit** | **Accuro® 3S Needle Guide Kit** [K250469] | **Pinpoint™ GT Safety Introducer Needle** [K142445] |
| --- | --- | --- | --- |
| **Trade Name** | Accuro® 3S Needle Guide Kit | Accuro® 3S Needle Guide Kit | Pinpoint™ GT Safety Introducer Needle |
| **FDA Classification Regulation** | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 870.1340 |
| **FDA Product Code** | ITX | ITX | DYB |
| **US Classification** | Class II | Class II | Class II |
| **Type of device** | Diagnostic Ultrasound Transducer | Diagnostic Ultrasound Transducer | Catheter Introducer |
| **Indications for Use** | The Accuro® 3S Needle Guide Kit is intended to be used with the Accuro® 3S diagnostic ultrasound imaging system. The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle | The Accuro® 3S Needle Guide Kit is intended to be used with the Accuro® 3S diagnostic ultrasound imaging system. The Accuro® 3S Needle Guide Kit supports alignment of a needle with the ultrasound imaging plane to assist the healthcare professional in placing the tip of a needle | The Pinpoint™ GT Safety Introducer Needle is intended for patients requiring percutaneous access to place a guidewire for subsequent placement of catheters or other medical procedures requiring introducer needle access. The Pinpoint™ GT Safety Introducer Needle may be used in any appropriate patient population. |
2
{6}
510(k) Summary- K261014
| | relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide. The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminant. The SafeTrack™ magnet is detected by magnetic sensors to provide visual guidance for tracking the position of a needle with respect to ultrasound image data. | relative to a specific anatomical structure. The elastic bands on the Patient Drape are intended to stabilize the positioning of the Accuro® 3S Needle Guide. The Accuro® 3S Probe Cover sheathes the transducer and isolates a needle insertion site from microbial and other contaminant. | |
| --- | --- | --- | --- |
| **Intended User** | Qualified and trained healthcare professional | Qualified and trained healthcare professional | Qualified and trained healthcare professional |
| **Prescription Use** | Yes | Yes | Yes |
| **Use Environment** | Hospital, clinic | Hospital, clinic | Hospital, clinic |
| **Biocompatible patient contact surface** | Yes | Yes | Yes |
| **Sterilization Method** | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| **Disposition** | Disposable | Disposable | Disposable |
| **Contents** | - Needle Guide - Conductive Ultrasound Gel - Probe Cover - Elastic Bands - Patient Drape - Needle with integrated SafeTrack™ Magnet | - Needle Guide - Conductive Ultrasound Gel - Probe Cover - Elastic Bands - Patient Drape | - Needle with integrated magnet |
Accuro® 3S Needle Guide Kit is substantially equivalent to the predicate device (K250469). The Accuro® 3S Needle Guide Kit indications for use is similar to the primary predicate device, Accuro® 3S Needle Guide Kit (K250469) and the reference device, Pinpoint™ GT Safety Introducer Needle (K142445). Both the subject device and the primary predicate device, Accuro® 3S Needle Guide Kit,
3
{7}
510(k) Summary- K261014
are indicated to cover an ultrasound probe and isolate from microbial and other contaminants, and both have needle guides to assist with placement of a needle to a specific anatomical structure. The subject device and the predicate also both have the patient drape with bands that support the needle guide positioning. The subject, primary predicate, and reference devices are prescription use and intended for the same use environments.
The key difference between the subject and primary predicate device is the inclusion of the previously 510(k) cleared needle with the SafeTrack™ magnet preassembled onto the needle. The magnet is integrated onto the needle hub and does not come in direct contact with the patient. The intended function of the SafeTrack™ magnet is to emit a magnetic field detectable by the magnetic sensors in the compatible Accuro® 3S diagnostic ultrasound imaging system probe to display the needle location. The new addition of the needle with the integrated SafeTrack™ magnet is similar to the reference device, Pinpoint™ GT Safety Introducer Needle, as both needles have a magnetic source intended to be detectable by their respective compatible ultrasound system probes, to display an indicator of the needle. The difference between the subject device and the Pinpoint™ GT Safety Introducer Needle is that the subject needle has a magnet attached onto the needle hub and the reference needle has a magnet integrated into the needle hub itself. Non-clinical testing for performance, including usability and safety has been conducted to ensure that no new risks are introduced due to the addition of the needle with the SafeTrack™ magnet. This includes verification testing and application of ISO 14971 for risk management.
# Summary of Non-Clinical Tests:
Design functionality of the subject device was tested to ensure that it meets the requirements of the intended end users. Testing included simulated use evaluations performed by representative end users in addition to benchtop testing.
The following benchtop verification and validation testing was performed following internal test protocols:
- Needle guide dimensions
- Needle insertion force and damage during needle insertion
- Needle angulation accuracy
- Needle tip depth accuracy
- Needle guide attachment / removal force
- Sterile barrier integrity
- Packaging integrity
- Elastic bands holding force
- Needle gate pinch force
- Elastic bands attachment durability
- Imaging quality following administration of the needle guide and probe cover
Probe midline needle insertion
Material dimensions
- Packaging configuration
4
{8}
510(k) Summary- K261014
Assurances of quality were further established by employing the following elements of product design and development in accordance with 21 CFR 820 and ISO 14971:2019:
- Risk analysis
- Requirements development
- Design reviews
- Verification and Validation
- Safety compliance verification
The following guidance documents and standards were used to determine appropriate methods for evaluating the performance of the device:
- AAMI / ANSI / ISO 10993-5:2009/(R)2014, Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- AAMI / ANSI ST72:2019, Bacterial Endotoxins - Test Methods, Routine Monitoring, And Alternatives To Batch Testing
- ASTM D4332-14, Standard Practice For Conditioning Containers, Packages, Or Packaging Components For Testing
- ASTM F1980-16, Standard Guide For Accelerated Aging Of Sterile Barrier Systems For Medical Devices
- ISO 10993-1:2018, Biological Evaluation of Medical Devices; Part 1 –Evaluation and Testing
- ISO 10993-7:2008, Biological Evaluation of Medical Devices Part 7: Ethylene Oxide Sterilization Residuals
- ISO 10993-10:2021, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
- ISO 10993-11:2017, Biological Evaluation Of Medical Devices - Part 11: Tests For Systemic Toxicity
- ISO 10993-12:2021, Biological Evaluation Of Medical Devices - Part 12: Sample Preparation And Reference Materials
- ISO 11135: 2014, Sterilization of healthcare products - Ethylene Oxide- Requirements For Development, Validation And Routine Control Of A Sterilization
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices – Part 1: Requirements for materials, sterile barrier systems and packaging systems, Amendment 1: Application of risk management 2023
- ISO 11607-2:2019, Packaging for terminally sterilized medical devices – Part 2: Validation requirements for forming, sealing and assembly processes, Amendment 1: Application of risk management 2023
- FDA-2013-D-0305, Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process”, Guidance for Industry and Food and Drug Administration Staff, 09/2023.
5
{9}
510(k) Summary- K261014
# Summary of Clinical Tests:
The subject of this premarket submission, Accuro® 3S Needle Guide Kit, did not require clinical studies to support the determination of substantial equivalence.
# Conclusion:
Accuro® 3S Needle Guide Kit exhibits the same characteristics and indications for use as legally marketed predicate and reference devices. Conclusions drawn from performance tests demonstrate substantial equivalence (SE). Therefore, Rivanna Medical considers Accuro® 3S Needle Guide Kit as substantially equivalent.
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.