Training, validation, and test sets derived from 2400 concurrent in-lab polysomnography (PSG) recordings from more than 1100 subjects
—
Prospective, multi-center study of 475 subjects (Watch) and 469 subjects (Ring) undergoing 2-night PSG
—
Indications for Use
The Sleep Apnea Feature is an over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Wearable and Phone. This feature is intended to detect signs of moderate to severe sleep apnea in the form of significant breathing disruptions in users 18 years and older, over a two-night monitoring period. It is intended for on-demand use and requires two nights of valid data within a 10-day period. This feature is not intended for users who have previously been diagnosed with sleep apnea. Users should not use this feature to replace traditional methods of diagnosis and treatment by a qualified clinician.
Device Story
Software-only mobile medical application (SaMD) operating on Samsung Galaxy Wearables (Watch/Ring) and smartphones; utilizes PPG, accelerometer, and PPG-derived SpO2 data collected during sleep; deep learning algorithm analyzes sensor inputs to identify breathing disruptions; provides binary risk assessment (positive/negative) for moderate-to-severe sleep apnea based on two-night time-weighted average; intended for home use by patients; results exported by user to share with clinicians; aids in identifying potential sleep apnea for further clinical evaluation; benefits include early detection of sleep apnea signs in undiagnosed populations.
Clinical Evidence
Prospective multi-center study (n=475 for Watch, n=469 for Ring) using 2-night PSG as ground truth. Watch: 90.8% sensitivity, 90.3% specificity. Ring: 77.9% sensitivity, 95.9% specificity. Performance consistent across gender, age, BMI, and skin tone subgroups. 90.0% (Watch) and 89.8% (Ring) of subjects within pre-specified breathing disruption accuracy zones.
Technological Characteristics
SaMD; PPG, accelerometer, and SpO2 sensors; deep learning algorithm; runs on Android/WearOS platforms; home use; non-invasive; software-only; no hardware materials; cybersecurity controls per FDA 2023 guidance; locked algorithm post-validation.
Indications for Use
Indicated for users 18+ years old without a prior sleep apnea diagnosis to detect signs of moderate to severe sleep apnea via significant breathing disruptions over a two-night monitoring period.
Regulatory Classification
Identification
An over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Watch and Phone. It is intended to detect signs of moderate to severe obstructive sleep apnea in the form of significant breathing disruptions in adult users 22 years and older, over a two-night monitoring period. It uses software algorithms to analyze input sensor signals (PPG and actigraphy) to provide a risk assessment for sleep apnea. It is not intended to provide a standalone diagnosis, replace traditional methods of diagnosis (e.g., polysomnography), assist clinicians in diagnosing sleep disorders, or be used as an apnea monitor.
Special Controls
In combination with the general controls of the FD&C Act, the over-the-counter device to assess risk of sleep apnea is subject to the following special controls:
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 20, 2026
Samsung Electronics Co., Ltd.
% Chaoyi Kang
Lead Biomedical Engineer
Samsung Research America, Inc.
665 Clyde Ave.
Mountain View, California 94043
Re: K261011
Trade/Device Name: Sleep Apnea Feature (v2.0)
Regulation Number: 21 CFR 868.2378
Regulation Name: Over-The-Counter Device To Assess Risk Of Sleep Apnea
Regulatory Class: Class II
Product Code: QZW
Dated: June 16, 2026
Received: June 17, 2026
Dear Chaoyi Kang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261011 - Chaoyi Kang
Page 2
FDA's substantial equivalence determination also included the review and clearance of your Predetermined Change Control Plan (PCCP). Under section 515C(b)(1) of the Act, a new premarket notification is not required for a change to a device cleared under section 510(k) of the Act, if such change is consistent with an established PCCP granted pursuant to section 515C(b)(2) of the Act. Under 21 CFR 807.81(a)(3), a new premarket notification is required if there is a major change or modification in the intended use of a device, or if there is a change or modification in a device that could significantly affect the safety or effectiveness of the device, e.g., a significant change or modification in design, material, chemical composition, energy source, or manufacturing process. Accordingly, if deviations from the established PCCP result in a major change or modification in the intended use of the device, or result in a change or modification in the device that could significantly affect the safety or effectiveness of the device, then a new premarket notification would be required consistent with section 515C(b)(1) of the Act and 21 CFR 807.81(a)(3). Failure to submit such a premarket submission would constitute adulteration and misbranding under sections 501(f)(1)(B) and 502(o) of the Act, respectively.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these
{2}
K261011 - Chaoyi Kang
Page 3
requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Binoy J.
Mathews -S
Digitally signed by Binoy J.
Mathews -S
Date: 2026.07.20 16:10:09
-04'00'
For
James J. Lee, Ph.D.
Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261011
Device Name
Sleep Apnea Feature v2.0
Indications for Use (Describe)
The Sleep Apnea Feature is an over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Wearable and Phone.
This feature is intended to detect signs of moderate to severe sleep apnea in the form of significant breathing disruptions in users 18 years and older, over a two-night monitoring period. It is intended for on-demand use and requires two nights of valid data within a 10-day period.
This feature is not intended for users who have previously been diagnosed with sleep apnea. Users should not use this feature to replace traditional methods of diagnosis and treatment by a qualified clinician.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
SAMSUNG
ELECTRONICS
Sleep Apnea Feature v2.0
### Samsung Sleep Apnea Feature v2.0
### 510(k) Summary
### K261011
Applicant Information
| Manufacturer: | Samsung Electronics Co., Ltd 129, Samsung-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do 16677, Korea |
| --- | --- |
| Contact Person: | Chaoyi Kang, Ph.D. Samsung Research America 665 Clyde Avenue, Mountain View, CA 94043 1-650-305-9330 chaoyi.kang@samsung.com |
| Date Prepared: | July 20^{th}, 2026 |
Device Information
| Trade/Device Name: | Sleep Apnea Feature v2.0 |
| --- | --- |
| Classification: | - QZW - Over-the-counter device to assess risk of sleep apnea (21 CFR 868.2378) |
| Predicate Device: | - Sleep Apnea Feature (DEN230041) |
| Reference Device | - Sleep Apnea Notification Feature (SANF) (K240929) |
Device Description
The Samsung Sleep Apnea Feature v2.0 is a Software as a Medical Device (SaMD) application, requiring a pair of mobile medical apps: one on a compatible Samsung Galaxy mobile device and the other on a Samsung accessory device (either a Samsung Galaxy Watch or Samsung Galaxy Ring), which are off-the-shelf computing platforms.
Following onboarding, the feature utilizes photoplethysmography (PPG) data, accelerometer data, and PPG-derived pulse oximetry data collected from the wearable device to monitor the user's sleep periods for repetitive, significant breathing disruptions that cause significant changes in relative blood oxygenation level.
The Sleep Apnea risk assessment feature provides users with a binary classification of moderate to severe sleep apnea risk based on two nights of sufficient breathing disruption data. The algorithm calculates the time-weighted average of breathing disruption value from the two nights. If the weighted average is 15 or more breathing disruptions per hour, the user will receive a positive risk assessment along with the average breathing disruptions observed during those two nights. The user is then advised to contact their doctor to discuss this result. Additionally, the mobile application provides the user with instructions for use,
Samsung Electronics Co., Ltd.
Page 1 of 9
{5}
**SAMSUNG**^{}[] ELECTRONICS
Sleep Apnea Feature v2.0---
education and troubleshooting information, and the option to export their risk assessment result with breathing disruption data.
### Intended Use/Indications for Use
The Sleep Apnea Feature v2.0 is indicated as follows:
*The Sleep Apnea Feature is an over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Wearable and Phone.*
*This feature is intended to detect signs of moderate to severe sleep apnea in the form of significant breathing disruptions in users 18 years and older, over a two-night monitoring period. It is intended for on-demand use and requires two nights of valid data within a 10-day period.*
*This feature is not intended for users who have previously been diagnosed with sleep apnea. Users should not use this feature to replace traditional methods of diagnosis and treatment by a qualified clinician.*
### Comparison of Technological Characteristics
The subject Sleep Apnea Feature v2.0 device is substantially equivalent to the predicate (DEN23004), as they share similar intended use, technological characteristics, and principles of operation. The difference in indications does not represent a new intended use. Both the subject and predicate devices are software-only mobile medical applications intended to detect signs of moderate-to-severe sleep apnea and are not intended to provide a standalone diagnosis.
A more detailed comparison between the subject device, the predicate device (Samsung Sleep Apnea Feature, DEN23004), and the reference device (Sleep Apnea Notification Feature, K240929) is provided in **Table 1**.
Samsung Electronics Co., Ltd.
Page 2 of 9
{6}
SAMSUNG
ELECTRONICS
Sleep Apnea Feature v2.0
Table 1. Technological Comparison of Subject Device, Predicate Device, and Reference Device
| Characteristic | Subject Device: Samsung Sleep Apnea Feature v2.0 | Predicate Device: Samsung Sleep Apnea Feature DEN230041 | Reference Device: Sleep Apnea Notification Feature K240929 | Comparison |
| --- | --- | --- | --- | --- |
| Indications for Use | The Sleep Apnea Feature is an over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Wearable and Phone. This feature is intended to detect signs of moderate to severe sleep apnea in the form of significant breathing disruptions in users 18 years and older, over a two-night monitoring period. It is intended for on-demand use and requires two nights of valid data within a 10-day period. This feature is not intended for users who have previously been diagnosed with sleep apnea. Users should not use this feature to replace traditional methods of diagnosis and treatment by a qualified clinician. | The Sleep Apnea Feature is an over-the-counter (OTC) software-only, mobile medical application operating on a compatible Samsung Galaxy Watch and Phone. This feature is intended to detect signs of moderate to severe obstructive sleep apnea in the form of significant breathing disruptions in adult users 22 years and older, over a two-night monitoring period. It is intended for on demand use. This feature is not intended for users who have previously been diagnosed with sleep apnea. Users should not use this feature to replace traditional methods of diagnosis and treatment by a qualified clinician. The data provided by this device is also not intended to assist clinicians in diagnosing sleep disorders. | The Sleep Apnea Notification Feature (SANF) is a software-only mobile medical application that analyzes Apple Watch sensor data to identify patterns of breathing disturbances suggestive of moderate-to-severe sleep apnea and provides a notification to the user. This feature is intended for over-the-counter (OTC) use by adults age 18 and over who have not previously received a sleep apnea diagnosis and is not intended to diagnose, treat, or aid in the management of sleep apnea. The absence of a notification is not intended to indicate the absence of sleep apnea. | The indications for use of the subject device deviate from the predicate due to the expanded user population (adult users over 18 rather than 22), which is the same as the reference device and supported by clinical validation study. |
| Principle of Operation | The Sleep Apnea Feature uses software algorithms to analyze input PPG, accelerometer, and blood oxygen signals and provide a risk assessment for sleep apnea. | The Sleep Apnea Feature uses software algorithms to analyze input blood oxygen sensor signals and provide a risk assessment for sleep apnea. | SANF uses software algorithms to analyze input accelerometer sensor signals and provide a risk assessment for sleep apnea. | Substantially equivalent to predicate device |
| Overall Device Design | A software-only device, and uses software algorithms to analyze input sensor signals from a general-purpose computing platform and provide a risk assessment for sleep apnea. | A software-only device, and uses software algorithms to analyze input sensor signals from a general-purpose computing platform and provide a risk assessment for sleep apnea. | A software-only device, and uses software algorithms to analyze input sensor signals from a general-purpose computing platform and provide a risk assessment for sleep apnea. | Substantially equivalent to predicate device |
Samsung Electronics Co., Ltd.
Page 3 of 9
{7}
**SAMSUNG**
ELECTRONICS
Sleep Apnea Feature v2.0
| Characteristic | Subject Device: Samsung Sleep Apnea Feature v2.0 | Predicate Device: Samsung Sleep Apnea Feature DEN230041 | Reference Device: Sleep Apnea Notification Feature K240929 | Comparison |
| --- | --- | --- | --- | --- |
| | Assessments are based on sensor data collected over a 2-day period. The device is intended to provide a risk assessment for signs of moderate to severe sleep apnea. | Assessments are based on sensor data collected over a 2-day period. The device is intended to provide on demand assessments to detect signs of sleep apnea, such that a user must actively choose to initiate a monitoring period. | Assessments are based on sensor data collected over 30-day periods. The device is intended to provide opportunistic detection of sleep apnea, such that after initial enrollment no user interaction is required for the device to perform as intended. | |
| OTC/Rx | Over-the-counter | Over-the-counter | Over-the-counter | Same as predicate device. |
| Use Environment | Home | Home | Home | Same as predicate device. |
| Device Components | Software-only | Software-only | Software-only | Same as predicate device. |
| Device Input | Blood oxygen level (SpO_{2}) data, PPG data, Accelerometer Data, Sleep Data. | Blood oxygen level (SpO_{2}) data, Sleep Data | Accelerometer data | Substantially equivalent to the predicate device |
| Key Sensor | PPG, Accelerometer | PPG, Accelerometer | Accelerometer | Uses the key sensor from both the predicate and reference device |
| Clinical Performance | Watch: Sensitivity: 90.8% 95% CI [86.0%, 94.4%] Specificity: 90.3% 95% CI [86.1%, 93.6%] Ring: Sensitivity: 77.9% 95% CI [71.5%, 83.5%] Specificity: 95.9% 95% CI [92.8%, 97.9%] | Sensitivity: 82.7% 95% CI [76.7%, 87.6%] Specificity: 87.7% 95% CI [83.1%, 91.4%] | Sensitivity: 66.3% 95% CI [62.2%, 70.3%] Specificity: 98.5% 95% CI [98.0%, 99.0%] | Clinical performance of the subject device operating on Watch is improved compared to the predicate device. Subject device operating on Ring is substantially equivalent to the reference device. |
| Algorithms | PPG and accelerometer-based algorithm to detect desaturation events | PPG-based algorithm to detect desaturation events | Accelerometer-based algorithm to detect breathing changes | Substantially equivalent to the predicate and reference device |
Samsung Electronics Co., Ltd.
Page 4 of 9
{8}
**SAMSUNG**
ELECTRONICS
Sleep Apnea Feature v2.0
| Characteristic | Subject Device: Samsung Sleep Apnea Feature v2.0 | Predicate Device: Samsung Sleep Apnea Feature DEN230041 | Reference Device: Sleep Apnea Notification Feature K240929 | Comparison |
| --- | --- | --- | --- | --- |
| Platforms | Galaxy Watch6 running WearOS 5.0, Galaxy Ring running latest firmware, Galaxy mobile phone with Android 12 | Galaxy Watch4 or later running WearOS 5.0+, Galaxy mobile phone with Android 9+ | Apple Watch Series 9 or later, Apple Watch Ultra 2, iPhone with iOS 18+ | Substantially equivalent to predicate device, Galaxy Ring provides the same type of sensors as Galaxy Watch. |
Samsung Electronics Co., Ltd.
Page 5 of 9
{9}
SAMSUNG
ELECTRONICS
Sleep Apnea Feature v2.0
# Performance Data
# Algorithm Development
Samsung Sleep Apnea Feature v2.0 includes a revised deep learning algorithm that identifies breathing disruptions and classify sleep apnea using photoplethysmography and accelerometer data from Galaxy Watch or Galaxy Ring. The model was trained on wearable sensor data collected during approximately 2400 concurrent in-lab polysomnography (PSG) recordings from more than 1100 subjects with varying clinical categories for sleep apnea. The development dataset included a diverse group of subjects in terms of age, gender, ethnicity, and BMI that appropriately represented the intended users of the device. The development dataset was divided into training, validation, and test sets. This standard algorithm development method ensures no subject overlap and matches distributions of demographics. After algorithm development was complete, the model was locked prior to verification and validation.
# Non-clinical Testing
The following tests were conducted, showing that the subject SaMD demonstrates safety, effectiveness, and substantial equivalence to the predicate and reference devices. No new safety issues were found during this testing.
- Software Verification Testing following recommendations in FDA's 2023 Guidance, "Content of Premarket Submissions for Device Software Functions", at Basic Documentation Level.
- Cybersecurity testing that adheres to FDA's 2023 Guidance, "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions."
- Labeling Verification Testing
# Off-the-Shelf Commercial Platform (Hardware and Software) Assessment
Per FDA's 2020 Guidance, "Multiple Function Device Products: Policy and Considerations", testing was performed on different aspects of the commercial platform hosting the subject SaMD (i.e., representing other functions), showing they do not negatively impact safely and effectiveness of the subject SaMD. This included the following:
- SpO₂ Bench Testing
- Low Perfusion Bench Testing
- Platform Interface Software Testing (Android and WearOS)
- Platform Interface Hardware Testing
- General safety tests: Electrical safety, electromagnetic compatibility, radio frequency emissions, material safety for skin contact, thermal safety for skin contact tests
Samsung Electronics Co., Ltd.
Page 6 of 9
{10}
SAMSUNG
ELECTRONICS
Sleep Apnea Feature v2.0
Device robustness tests: Resistance to electrostatic discharge, water ingress and breakage tests.
### Human Factor Validation:
To evaluate the effectiveness of control measures designed to reduce or eliminate use-related hazards and/or potential use errors, especially those associated with moderate harm including over-relying on the Samsung Sleep Apnea Feature 2.0 in a way that leads to delayed or no medical treatment of a serious condition, a human factor validation study was conducted in accordance with the following guidance documents:
- Food and Drug Administration. (2016). Applying human factors and usability engineering to medical devices: guidance for industry and Food and Drug Administration staff
- IEC 62366-1, Medical devices – Application of usability engineering to medical devices, which is published by the International Electrotechnical Commission and has been adopted in the United States as ANSI/AAMI/IEC 62366-1:2015+AMD1:2020
- IEC TR 62366-2, Medical devices—Part 2: Guidance on the application of usability engineering to medical devices, which is published by the International Electrotechnical Commission (adopted in the United States as AAMI/IEC TIR62366-2:2016)
### Clinical Testing
The Sleep Apnea Feature was validated via a prospective, multi-center study where study participants (i.e., an enriched population of individuals aged 18 and older) underwent a 2-night polysomnography (PSG) study while simultaneously wearing a Samsung Galaxy Watch (n = 475) and Samsung Galaxy Ring (n = 469) with the Sleep Apnea Feature. The completed study included subjects across the spectrum of sleep apnea severity classifications (Normal, Mild, Moderate, and Severe), ensuring appropriate representation for each apnea-hypopnea index (AHI) range.
The Sleep Apnea feature's sensitivity and specificity was assessed when identifying subjects who exhibited moderate-to-severe sleep apnea during that 2-night PSG study for both watch and ring platforms (Table 2).
Table 2. Classification Performance Analysis
| Watch (N=475) | | Ring (N=469) | |
| --- | --- | --- | --- |
| Sensitivity | Specificity | Sensitivity | Specificity |
| 90.8%95% CI [86.0%, 94.4%] (188 of 207) | 90.3%95% CI [86.1%, 93.6%] (242 of 268) | 77.9%95% CI [71.5%, 83.5%] (155 of 199) | 95.9%95% CI [92.8%, 97.9%] (259 of 270) |
Furthermore, additional analysis demonstrated:
Samsung Electronics Co., Ltd.
Page 7 of 9
{11}
SAMSUNG
ELECTRONICS
Sleep Apnea Feature v2.0
1. The Sleep Apnea Feature was able to provide classification for 87.9% of sleep sessions for the watch and 86.0% of sleep sessions for the ring.
2. The Sleep Apnea Feature provided breathing disruption measurements within the pre-specified breathing disruption accuracy zones¹ in 90.0% of the subjects for the watch and 89.8% of the subjects for the ring.
Observed performance for relevant subgroups is shown in Table 3 showing the Sleep Apnea Feature's classification performance is reasonably consistent across the subgroups.
Table 3. Subgroup Performance Analysis
| | Watch | | | Ring | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | N | Sensitivity | Specificity | N | Sensitivity | Specificity |
| Gender | Male | 229 | 94.1% | 88.3% | 231 | 85.5% | 92.0% |
| | Female | 246 | 84.7% | 91.4% | 238 | 63.2% | 98.2% |
| Age | <40 | 128 | 86.5% | 95.6% | 126 | 66.7% | 98.9% |
| | 40-55 | 165 | 94.6% | 91.2% | 164 | 81.7% | 92.5% |
| | 55+ | 182 | 89.6% | 83.7% | 179 | 78.9% | 96.4% |
| BMI (lb/in²) | <25 | 53 | 83.3% | 91.5% | 53 | 85.7% | 97.8% |
| | >= 25 | 422 | 91.0% | 90.0% | 416 | 77.6% | 95.5% |
| Skin Tone (Monk Scale) | 1-4 | 147 | 97.0% | 90.0% | 147 | 80.9% | 100% |
| | 5-6 | 159 | 84.4% | 93.7% | 158 | 71.0% | 100% |
| | 7-10 | 169 | 90.8% | 87.1% | 164 | 81.2% | 88.4% |
### Predetermined Change Control Plan (PCCP)
The Samsung Sleep Apnea Feature v2.0 contains a Predetermined Change Control Plan (PCCP), which complies with Section 3308 of the Food and Drug Omnibus Reform Act (FDORA) of 2022. The PCCP includes a list of the allowable device software modifications and protocols describing the verification and validation activities that will support the proposed potential changes (see Table 4). The modification protocols incorporate impact assessment considerations and specify requirements for data management, including data sources, collection, storage, and sequestration, as well as documentation and data re-use practices. Specific test methods are detailed to establish substantial equivalence relative to the initial submission and include analysis methods and acceptance criteria. To ensure that the validation test dataset is representative of the intended use population, it specifies demographic requirements for age, sex, race, ethnicity, sleep apnea severity, and body mass index.
¹ Breathing disruption accuracy zones are set at ± 5 at 0 PSG apnea-hypopnea index (AHI), ± 10 at 15 AHI, ± 15 at 30 AHI.
Samsung Electronics Co., Ltd.
Page 8 of 9
{12}
SAMSUNG
ELECTRONICS
Sleep Apnea Feature v2.0
No modifications described in the PCCP allow adaptive learning of algorithms in the field. Modifications to all algorithms, including neural networks, will be verified and validated and then locked prior to release. Furthermore, all modifications have the following requirements:
- No concurrent changes to hardware sensors or sensor framework functions on the wearable platform
- No changes to the intended use of the device or its use conditions.
- Full verification and validation through pre-specified modification protocols.
Users are to be subsequently informed of changes via an update notice and revision of the labeling.
Table 4. Summary of Modifications
| Type | List of Modification | Test Method |
| --- | --- | --- |
| Modifications to the respiratory event detection and breathing disruption computation | - Modifications to the SpO_{2} signal quality check module | Comparative analysis to establish substantial equivalency of SpO_{2} performance to the original clearance |
| | - Modifications to the pre-processing and segmentation of model input data - Modification to the breathing disruptions algorithm's hyper-parameters - Re-training of the breathing disruptions calculation neural network - Modification to the breathing disruptions regression fitting | Comparative analysis to establish substantial equivalency of classification (sensitivity and specificity) and breathing disruption performance to the original clearance |
| Modifications to sleep apnea classification logic | - Modification to sleep session data requirement and merging logic - Modifications of data requirement for calculating classification result - Modifications to multi-night sleep apnea classification logic | Comparative analysis to establish substantial equivalency of classification (sensitivity and specificity) performance to the original clearance |
## Conclusion
The subject device and predicate device have the same intended use, the same key technological characteristics, and similar signal-acquisition capabilities. In clinical testing, Samsung's updated Sleep Apnea Feature v2.0 demonstrated that it has comparable sleep apnea classification performance as the predicate (DEN230041). Combined with passing bench and human factors tests, we conclude that the subject device is safe, effective, and possesses an appropriate level of specificity and sensitivity for general population obstructive sleep apnea risk assessment. Therefore, Samsung Sleep Apnea Feature v2.0 is substantially equivalent to the Samsung Sleep Apnea Feature (DEN230041).
Samsung Electronics Co., Ltd.
Page 9 of 9
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.