The HERizon RD5™ Resecting Device is intended for intrauterine use by physicians trained in hysteroscopy to resect and remove tissue such as endometrial polyps.
Device Story
Battery-powered, motor-driven, manually-actuated hysteroscopic surgical instrument; used in OR, surgical centers, or physician offices. Input: manual activation via button. Transformation: motor spins internal stainless steel blade within distal cannula window; blade oscillates radially; mechanical valve opens to enable suction. Output: resected tissue evacuated through device to external suction source/tissue catch. Benefit: enables minimally invasive removal of intrauterine tissue under direct visualization. Physician-operated; provides mechanical resection of polyps.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993), sterilization validation (ISO 11135), electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and mechanical/functional performance (cutting performance, torque, tensile strength, vacuum maintenance) at baseline and after 1-year accelerated aging.
Technological Characteristics
Battery-powered (up to 6V); 5Fr (1.65 mm) outer diameter; 36.3 cm cannula length. Materials: stainless steel (distal patient-contacting components), molded plastics (handle). Sensing/Actuation: motor-driven radial blade oscillation. Connectivity: standard luer/stepped connectors for fluid management/suction. Sterilization: Ethylene Oxide (EO). Standards: ISO 10993, ISO 11135, IEC 60601-1, IEC 60601-1-2, ASTM D4169, ASTM F1980.
Indications for Use
Indicated for adults (22 years and older) requiring intrauterine resection and removal of tissue, such as endometrial polyps, by physicians trained in hysteroscopy.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 18, 2026
Axora Medical, Inc.
% Natalie J. Kennel
Regulatory Affairs Consultant
NJK & Associates, Inc.
13721 Via Tres Vista
San Diego, CA 92129
Re: K260984
Trade/Device Name: HERizon RD5™ Resecting Device
Regulation Number: 21 CFR§ 884.1690
Regulation Name: Hysteroscope and Accessories
Regulatory Class: II
Product Code: HIH
Dated: July 22, 2026
Received: July 23, 2026
Dear Natalie J. Kennel:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260984 - Natalie J. Kennel
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260984 - Natalie J. Kennel
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Reginald K. Avery -S
for Jason R. Roberts, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260984 | ? |
| Please provide the device trade name(s). | | ? |
| HERizon RD5™ Resecting Device | | |
| Please provide your Indications for Use below. | | ? |
| The HERizon RD5™ Resecting Device is intended for intrauterine use by physicians trained in hysteroscopy to resect and remove tissue such as endometrial polyps. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
AXORA MEDICAL
K260984
Page 1 of 7
# 510(k) Summary
Sponsor: Axora Medical Inc.
1 Independent Drive
Jacksonville, FL 32202
Contact Person: Ms. Natalie J. Kennel
Consultant
NJK & Associates, Inc.
13721 Via Tres Vista
San Diego, CA 92129 USA
Phone: (858) 705-0350
Fax: (858) 764-9739
Email: NKennel@njkconsulting.com
Date Prepared: August 16, 2026
## DEVICE INFORMATION:
Proprietary Name: HERizon RD5™ Resecting Device
Regulation Name: Hysteroscope and accessories
Classification: Class II
Product Codes: HIH
Regulations: 21 CFR 884.1690
Classification Panel: Obstetrics/Gynecology
## PREDICATE DEVICES:
Table 1 contains information about the predicate devices:
Table 1 Predicate Device Information
| 510(k)# | Product Name | Sponsor /510(k) holder | Clearance Date | Role |
| --- | --- | --- | --- | --- |
| K233500 | BenestaTM Tissue Removal | Caldera Medical | Nov. 30, 2023 | Primary predicate |
| K191538 | ResectrTM Device | Boston Scientific/Minerva Surgical/Axora Medical\( ^{1} \) | Jan.16, 2020 | Secondary predicate |
None of the predicate devices have been the subject of a design related recall.
\( ^{1} \) Boston Scientific Corporation obtained the clearance and later sold this clearance to Minerva Surgical. Axora Medical acquired this asset from Minerva on Jan. 30, 2026
HERizon RD5™ Resecting Device
Page 1 of 7
{5}
K260984
Page 2 of 7
# PRODUCT DESCRIPTION:
The HERizon RD5™ Resecting Device is a battery powered, motor driven, and manually actuated hysteroscopic surgical instrument used to resect and remove tissue under direct visualization. The device is supplied sterile for single use only. The device is offered in a 5Fr size for compatibility with hysteroscopes that can accommodate instruments with a 1.65 mm outer diameter or larger. The device can be used with any fluid management system made with standard luer and/or stepped connectors.
The HERizon RD5™ device consists of a distal and proximal section. The distal section is comprised of an outer cannula with an open Resecting Window at the distal end, and a cutting element that rotates within the lumen of the outer cannula to resect tissue within the Resecting Window. The proximal section, made of molded plastics, is comprised of a cylindrically shaped handle with an Activation Button. Depressing the Activation Button actuates a motor that spins a blade within the Resecting Window to resect and remove target tissue. Depressing the Activation Button also opens a mechanical valve within the handle body. When open, evacuation of resected tissue is enabled. The devices are provided assembled with a vacuum tube port for use with a suction source. As tissue is resected, it is evacuated through the length of the device and out the vacuum tube port. The distal section, the patient contacting portion, including the cannula and cutting element are made of stainless steel.
# INDICATIONS FOR USE:
The HERizon RD5™ Resecting Device is intended for intrauterine use by physicians trained in hysteroscopy to resect and remove tissue such as endometrial polyps.
# SUBSTANTIAL EQUIVALENCE
Table 2 contains the substantial equivalence table comparing the HERizon RD5™ Resecting Device to its predicates. The table shows that the subject device has the same intended use as both predicates. The subject device uses similar design and technology as one or both of the predicates. The subject device is battery powered, same as its primary predicate device. The subject device is a single use disposable device provided sterile, same as both predicates. The subject device has similar dimensions and compatibility as at least one model of both predicates. The subject device is made of the same type of materials as both predicates. The subject device has some minor differences in the way the resection is activated in the predicates but the differences do not raise different issues of safety and effectiveness.
The HERizon RD5™ Resecting Device has the same intended use and similar technology characteristics as compared to its primary and secondary predicate devices. The difference between the subject device and predicate devices do not raise different questions of safety or effectiveness.
HERizon RD5™ Resecting Device
Page 2 of 7
{6}
K260984^{}[] Page 3 of 7
**Table 2 Substantial Equivalence Table Comparing HERizon RD5™ Resecting Device to its Predicates**
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate | Comparison |
| --- | --- | --- | --- | --- |
| Trade Name | HERizon RD5™ Resecting Device | Benesta® Tissue Removal Device | Resectr™ Tissue Resection Device | NA |
| 510(k) | New | K233500 | K191538 | NA |
| Submitter/ 510(k) Holder | Axora Medical Inc. | Caldera Medical Inc. | Minerva Surgical/Axora Medical^{2} | NA |
| Device type | Hysteroscope and accessories | Hysteroscope and accessories | Hysteroscope and accessories | Same |
| Classification | Class II | Class II | Class II | Same |
| Regulation | 884.1690 | 884.1690 | 884.1690 | Same |
| Product code | HIH | HIH | HIH | Same |
| Indications for Use | The *HERizon RD5™* Resecting Device is intended for intrauterine use by physicians trained in hysteroscopy to resect and remove tissue such as: endometrial polyps | The Caldera Medical Benesta™ Tissue Removal Device is intended for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retained products of conception. | The Resectr™ Tissue Resection Device is intended for intrauterine use by physicians trained in hysteroscopy to resect and remove tissue, including focal lesions such as endometrial polyps. | Same intended use to resect and remove intrauterine tissue. The indications included are in both primary and secondary predicate |
| Usage | Single use disposable | Single use disposable | Single use disposable | Same |
| Use Environment | operating room, surgical center and physician's office environments where surgical procedures are performed | operating room, surgical center and physician's office environments where surgical procedures are performed | operating room, surgical center and physician's office environments where surgical procedures are performed | Same |
| Power source/ Operating voltage | Battery/ Up to 6 V | Battery/ 7.7 - 9.0 V | Manual | Primary predicate is battery powered with similar voltage. Slight difference in voltage does not raise different issues of safety or effectiveness. |
$^{2}$ Boston Scientific Corporation obtained the clearance and later sold this clearance to Minerva Surgical. Axora Medical acquired this asset from Minerva on Jan. 30, 2026.
HERizon RD5™ Resecting Device
Page 3 of 7
{7}
K260984^{}[] Page 4 of 7
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate | Comparison |
| --- | --- | --- | --- | --- |
| Sterilization | Provided sterile- Ethylene oxide | Provided sterile- Ethylene oxide | Provided sterile- Ethylene oxide | Same |
| Outer diameter | 5F (1.65 mm) | Benesta OD 4.0 mm (12F) | 5F (1.65 mm) and 9F (3.0mm) | Subject device same as one of two Resectr™ predicate device models (5F) |
| Cannula length | 36.3 cm | 35 cm | 35.3 cm | Subject device similar to both predicates |
| Resecting Window length | 5.0 mm | 15.0 mm | 5.0 mm (5 F) | Subject device same as Resectr™ predicate |
| Directionality of resecting blade | Resecting blade oscillates radially relative to the cannula | Resecting blade translates axially along the cannula length. During translation, resecting blade rotates | Resecting blade oscillates radially relative to the cannula | Subject device same as Resectr™ predicate |
| Resecting Window orientation | Side facing window | Side facing window | Side facing window | Same |
| Resecting Window orientation indicator | Laser markings | Laser markings | Laser markings | Same |
| Blade actuation | Upon pressing a button, a motor is activated that rotates the resection blade. The resection blade rotates three times in one direction before changing directions and rotating three times in the opposite direction. | Blade is actuated using a momentary switch to activate the motor. An additional sliding switch sets the on/off state of the device which connects or disconnects the internal battery to the rest of the internal electronics. | Each handle squeeze and release on the Resectr™ device rotates the blade bi-directionally for a total of six resection actions across the cannula window. | Subject device is similar to the predicate devices. Primary predicate blade is actuated with a switch or motor like subject device. Secondary predicate is actuated with manual squeeze. Differences are minor and do not raise different issues of safety or effectiveness |
| Blade material/ Patient contact & duration | Stainless Steel / Externally communicating, limited duration (< 24 hours) | Stainless Steel / Externally communicating, limited duration (< 24 hours) | Stainless Steel/ Externally communicating, limited duration (< 24 hours) | Subject device material, patient contact and duration is same as both predicates |
| Vacuum and tissue collection | HERizon RD5™ Resecting Device is connected to an external suction source and tissue catch. | Benesta™ device is connected to an external suction source and tissue catch. | Resectr™ Device is connected to an external suction source and tissue catch. | Same |
| Supplied accessories | None | None | None | Same |
HERizon RD5™ Resecting Device
Page 4 of 7
{8}
K260984^{}[] Page 5 of 7
| Characteristic | Subject Device | Primary Predicate | Secondary Predicate | Comparison |
| --- | --- | --- | --- | --- |
| Compatibility | Hysteroscopes with a working channel of 1.65 mm or larger | Hysteroscopes with a working channel of 4.0 mm (12F) | Hysteroscopes with a working channel of 1.65 mm or larger (5F) or 3.0 mm or larger (9F) | Subject device is same as one of two Resectr™ predicate device models. |
| Bench testing | Range of mechanical, functional, and performance testing | Unknown | Range of mechanical, functional, and performance testing | Subject device was subjected to the same or similar bench testing as Resectr™ predicate device. Bench testing of primary predicate is unknown due to 510(k) statement |
HERizon RD5™ Resecting Device
Page 5 of 7
{9}
K260984
Page 6 of 7
# NON-CLINICAL PERFORMANCE DATA
# Biocompatibility
Biocompatibility studies were performed in accordance with the 2023 FDA guidance document “Use of International Standard ISO 10993-1, Biological Evaluation of Medical Devices – Part 1: Evaluation and testing within a risk management process” and ISO 10993-1:2018 as follows for a device where the distal portion of the device has patient contact that is externally communicating with a limited duration of less than 24 hours:
Cytotoxicity (ISO 10993-5:2009)
Sensitization (ISO 10993-10:2021
Vaginal irritation (ISO 10993-23:2021)
- Acute systemic toxicity (ISO 10993-11:2017)
Material mediated pyrogenicity (ISO 10993-11:2017, USP<151>2017)
# Sterilization
The HERizon RD5™ Resecting Device is packaged for single use and terminally sterilized with Ethylene Oxide (EO). EO sterilization validation was performed per ISO 11135:2014/AMD 1:20218 ‘Sterilization of health care products – Ethylene oxide – Requirements for the development, validation, and routine control of a sterilization process for medical devices – Amendment 1: Revision of Annex E Single batch release.’ EO residuals were confirmed to be within ISO 10993-7:2008/AMD 1:2019 ‘Biological Evaluation of Medical Devices – Part 7: Ethylene Oxide Sterilization Residuals.’.
# Package Integrity
Sterile barrier/packaging validation was performed following simulated shipping and handling per ASTM D4169 -23e1 ‘Standard Practice for Performance Testing of Shipping Containers and Systems’ and related standards as follows:
- Visual seal inspection (ASTM F1886/F18886M-16 Standard Test Method for Determining Integrity of Seals for Flexible Packaging by Visual Inspection
- ASTM F1929-23 Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- Pouch Seal Peel Strength Testing (ASTM F88/F88M-21 'Standard Test Method for Seal Strength of Flexible Barrier Materials')
# Electrical Safety/EMC
The battery powered HERizon RD5™ Resecting Device was tested and found to conform to the following electrical safety and EMC standards:
Electrical safety (IEC 60601-1 Edition 3.2 2020-08 ‘Consolidated Version: Medical Electrical Equipment – Part 1: General Requirements for Basic Safety and Essential Performance’.
HERizon RD5™ Resecting Device
Page 6 of 7
{10}
K260984
Page 7 of 7
IEC 60086-5 Primary Batteries – Part 5 with Aqueous Electrolytes
EMC (IEC 60601-1-2:2014/AMD1:2020)
# Mechanical/Functional Performance/Shelf-Life Testing
Mechanical and functional performance testing was conducted on samples at baseline T = 0 and aged T = 1 year accelerated aging (per ASTM F1980). The performance tests completed to demonstrate that the HERizon RD5™ Resecting Device functions as intended include:
- Cutting/Resection Performance
- Device/Scope Compatibility
- Dimensional Measurements
- Durability
- Functionality of Activation Button
- Outer Tube (Cannula) Rotation
- Vacuum Maintenance
- Torque strength testing
- Tensile testing
All test results met the predefined acceptance criteria.
# CLINICAL STUDIES
No clinical studies were performed.
# CONCLUSION
The results of the above-described performance testing demonstrate that the HERizon RD5™ Resecting Device is as safe and effective as the predicate devices and supports a determination of substantial equivalence.
HERizon RD5™ Resecting Device
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.