← Product Code [KNS](/productcode/KNS) · K260962

# Sphincterotome (K260962)

_Taewoong Medical Co., Ltd. · KNS · May 21, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K260962

## Device Facts

- **Applicant:** Taewoong Medical Co., Ltd.
- **Product Code:** [KNS](/productcode/KNS.md)
- **Decision Date:** May 21, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Sphincterotome is indicated for use in the selective cannulation of the Common Bile Ducts (CBD) and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The Sphincterotome can also be used to inject contrast medium.

## Device Story

Endoscopic electrosurgical accessory; used for selective cannulation of Common Bile Duct and sphincterotomy of Papilla of Vater/Sphincter of Oddi. Device comprises three-lumen polymer sheath, cutting wire (SUS304), and handle. Operates via physician manipulation of handle to rotate wire and application of monopolar radio frequency current from external ESU to perform incision. Sheath lumens facilitate guidewire introduction and contrast agent/saline injection. Used in clinical endoscopic settings by physicians. Provides controlled tissue incision; benefits patient by enabling minimally invasive biliary access and therapeutic intervention.

## Clinical Evidence

Bench testing only. Testing included visual inspection, dimensional analysis, tensile strength, leakage, electrical resistance, endoscopy compatibility, device operability, cutting wire function, HF dielectric strength, connector retention, rotation, and thermal performance. Biocompatibility verified per ISO 10993 series. Electrical safety and EMC verified per IEC 60601 series.

## Technological Characteristics

Materials: SUS304 cutting wire, PTFE sheath, ABS handle. Energy: Monopolar radio frequency current (max 700V peak). Dimensions: 1.6mm distal tip O.D., 1950mm length. Connectivity: Compatible with standard endoscope working channels (>=2.8mm) and guidewires (0.021"-0.035"). Sterilization: Ethylene Oxide (EO), SAL 10^-6. Standards: ISO 10993-1, IEC 60601-1, IEC 60601-2-2, IEC 60601-2-18.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Autotome™ Pro RX 39 Sphincterotome/Autotome™ Pro RX 44 Sphincterotome ([K232162](/device/K232162.md))
- Autotome™ RX Model ([K013153](/device/K013153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

May 21, 2026

Taewoong Medical Co., Ltd.
% Matthew Krueger
Principal Consultant
Biologics Consulting Group, Inc.
100 Daingerfield Rd.
Suite 101
Alexandria, Virginia 22314

Re: K260962

Trade/Device Name: Sphincterotome
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic Electrosurgical Unit And Accessories
Regulatory Class: Class II
Product Code: KNS
Dated: March 23, 2026
Received: March 23, 2026

Dear Matthew Krueger:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K260962 - Matthew Krueger
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K260962 - Matthew Krueger
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

ANTHONY LEE -S

Anthony Lee, Ph.D., MBA
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260962 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  Sphincterotome  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  Sphincterotome is indicated for use in the selective cannulation of the Common Bile Ducts (CBD) and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The Sphincterotome can also be used to inject contrast medium.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

510(k) Summary
Taewoong Medical
K260962
Page 1 of 7

# 1. SUBMITTER INFORMATION

Applicant: Taewoong Medical
Contact: Yongjin Jeff Kim
Phone: +82 70 4649 1543
Email: jinjeff@stent.net
Address: 14, Gojeong-ro, Wolgot-myeon, Gimpo-si, Gyeonggi-do, 10022, Republic of Korea

# 2. CORRESPONDENT INFORMATION

Contact: Matthew Krueger
Title: Principal Consultant
Firm: Biologics Consulting Group
Phone: (667) 352-2578
Email: mkrueger@biologicsconsulting.com
Address: 100 Daingerfield Rd., Suite 101
Alexandria, VA 22314

# 3. DATE PREPARED: MARCH 20, 2026

# 4. DEVICE INFORMATION

Device Name: Sphincterotome
Regulation Number: 21 CFR 876.4300
Regulation Name: Endoscopic electrosurgical unit and accessories.
Product Code: KNS
Regulatory Class: Class II

# 5. PREDICATE DEVICE INFORMATION

Primary Predicate Device Name: Autotome™ Pro RX 39 Sphincterotome/Autotome™ Pro RX 44 Sphincterotome
510(k) Number: K232162
Manufacturer: Boston Scientific Corporation

Page 1 of 7

{5}

510(k) Summary
Taewoong Medical
K260962
Page 2 of 7

Secondary Predicate Device Name: Autotome™ RX Model
510(k) Number: K013153
Manufacturer: Boston Scientific Corporation

## 6. DEVICE DESCRIPTION

The Sphincterotome is an endoscopic electrosurgical accessory and consists of three parts: a cutting wire, a sheath, and a handle. The cutting wire is insulated with a non-conductive polymer tube and features a pre-curved and tapered tip. The sheath is made of a three-lumen polymer tube: one lumen for a cutting wire that transmits high-frequency current to the target site, another lumen for delivering contrast agents or saline, and the third lumen for introducing a guidewire. The proximal end of the sheath and cutting wire assembly terminates in a handle. The handle includes three parts: a plug that transmits high frequency electric current from the ESU to the cutting wire, an injection port, and a guidewire port. Physicians can manipulate the cutting wire by rotating the handle during the procedure.

## 7. INDICATIONS FOR USE

Sphincterotome is indicated for use in the selective cannulation of the Common Bile Ducts (CBD) and the transendoscopic sphincterotomy of the Papilla of Vater and/or the Sphincter of Oddi. The Sphincterotome can also be used to inject contrast medium.

## 8. COMPARISON OF INTENDED USE AND TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Page 2 of 7

{6}

510(k) Summary
Taewoong Medical
K260962
Page 3 of 7

Table 1 provides the technological comparison.

Page 3 of 7

{7}

510(k) Summary

Taewoong Medical

K260962

Page 4 of 7

Table 1: Technological Comparison to Predicate Devices

|   | Subject Device | Primary Predicate Device (K232162) | Secondary Predicate Device (K013153)  |
| --- | --- | --- | --- |
|  Device Name | Sphincterotome | Autotome™ Pro RX 39 Sphincterotome Autotome™ Pro RX 44 Sphincterotome | Autotome™ RX Model  |
|  Common Name | Sphincterotome | Sphincterotome | Sphincterotome  |
|  Manufacturer | Taewoong Medical | Boston Scientific | Boston Scientific  |
|  Clinical Condition | An incision of papilla | An incision of papilla | An incision of papilla  |
|  Application | Common Bile Duct, Papilla of Vater and/or Sphincter of Oddi | Common Bile Duct, Papilla of Vater and/or Sphincter of Oddi | Papilla of Vater and/or Sphincter of Oddi  |
|  Procedure | Endoscopic | Endoscopic | Endoscopic  |
|  Size Specification | - Distal tip O.D.: 1.6 mm - Tip length: 7 mm - Cutting wire length: 20/25/30 mm - Length: 1950 mm | - Distal tip O.D.: 1.47 (4.4Fr) mm - Tip length: 5 mm - Cutting wire length: 25 mm - Length: 2000 mm | - Distal tip O.D.: 1.3 (3.9Fr) / 1.47 (4.4Fr) / 1.63 (4.9Fr) mm - Tip length: 5 mm - Cutting wire length: 20 / 30 mm - Length: 2000 mm  |
|  Materials | - Cutting wire: SUS304 - Sheath: PTFE tube - Handle: ABS | - Cutting wire: Metal - Sheath: Polymer | - Cutting wire: Metal - Sheath: Polymer  |
|  Injection | Yes | Yes | Yes  |
|  Compatible endoscopy working channel | 2.8 mm or larger | 2.8 mm or larger | 2.8 mm or larger  |
|  Compatible guidewire | Traditional guidewire (4.5m) or short guidewire (2.6m) with 0.021” ~ 0.035” diameter | Traditional guidewire (4.5m) or short guidewire (2.6m) with 0.021” ~ 0.035” diameter | Traditional guidewire (4.5m) or short guidewire (2.6m) with 0.021” ~ 0.035” diameter  |
|  Energy Used | Monopolar Radio Frequency Current | Monopolar Radio Frequency Current | Monopolar Radio Frequency Current  |
|  Maximum Voltage Rating | 700 V peak (1400 V peak-to-peak) | 750 V peak (1500 V peak-to-peak) | 750 V peak (1500 V peak-to-peak)  |
|  Insulation | Yes | Yes | No  |
|  Packaging | Single-use EO sterilized pouch with one device per pouch | Single-use EO sterilized pouch with one device per pouch | Single-use EO sterilized pouch with one device per pouch  |
|  Shelf Life | 3 years | 3 years | 3 years  |

Page 4 of 7

{8}

510(k) Summary
Taewoong Medical
K260962
Page 5 of 7

|   | Subject Device | Primary Predicate Device (K232162) | Secondary Predicate Device (K013153)  |
| --- | --- | --- | --- |
|  Sterilization method / SAL | EO Sterilization, SAL of 10^{-6} | EO Sterilization, SAL of 10^{-6} | EO Sterilization, SAL of 10^{-6}  |
|  Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | Conforms with ISO 10993-1  |
|  Electromagnetic compatibility and Electrical Safety | Conforms with IEC 60601-1 | Conforms with IEC 60601-1 | Conforms with IEC 60601-1  |

Page 5 of 7

{9}

510(k) Summary
Taewoong Medical
K260962
Page 6 of 7

# 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING

## Biocompatibility Testing

Biocompatibility testing was conducted to the following standards:

- ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation
- ISO 10993-10:2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- USP 43 &lt;85&gt;, Bacterial Endotoxin Testing
- USP 43 &lt;161&gt;, Bacterial Endotoxin Testing

## Electrical Safety

The subject device was tested against the following standard for electrical safety:

- IEC 60601-1:2005/AMD1:2012/AMD2:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance

## Electromagnetic Compatibility (EMC)

The subject device was tested against the following standards for electromagnetic compatibility:

- IEC 60601-1-2:2014/AMD1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-2-2:2017/AMD1:2023, Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
- IEC 60601-2-18:2009, Medical electrical equipment - Part 2-18: Particular requirements for the basic safety and essential performance of endoscopic equipment

## Software

Not applicable as the subject device does not contain software.

## Performance Testing

The following performance testing was conducted:

- Visual inspection
- Dimension
- Tensile Strength

{10}

510(k) Summary
Taewoong Medical
K260962
Page 7 of 7

- Leakage
- Resistance
- Packaging compatibility
- Endoscopy compatibility
- Device Operability
- Cutting wire function
- HF Dielectric Strength
- Interconnection conditions for HF applications
- Detachable connector retention
- Rotation
- Thermal performance

## 10. CONCLUSION

The results of the performance testing described above demonstrate that the Sphincterotome performs equivalently as compared to the primary and secondary predicate devices and supports a determination of substantial equivalence.

Page 7 of 7

---

**Source:** [https://fda.innolitics.com/device/K260962](https://fda.innolitics.com/device/K260962)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
