Smart Electric Wheelchair (R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK)
K260937 · Shanghai Waylong Intelligent Technology Co.,Ltd · ITI · Jun 24, 2026 · Physical Medicine
Device Facts
Record ID
K260937
Device Name
Smart Electric Wheelchair (R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK)
Applicant
Shanghai Waylong Intelligent Technology Co.,Ltd
Product Code
ITI · Physical Medicine
Decision Date
Jun 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3860
Device Class
Class 2
Indications for Use
The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
Device Story
Battery-powered, four-wheeled electric wheelchair; foldable frame; models R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK. User operates via joystick-controlled electronic controller; system includes brushless DC motors and automatic electromagnetic braking. Provides mobility for disabled or elderly individuals; intended for indoor/outdoor use. Features include speed selection, horn, lighting, and USB port. Subject device utilizes magnesium alloy frame; PET cushions; lithium-ion battery. Safety verified via ISO 7176 series testing, including stability, braking, and obstacle climbing. Output is physical mobility; device benefits patient by providing independent transport.
Clinical Evidence
No clinical data or animal studies were conducted. Substantial equivalence was demonstrated through bench testing, including the full ISO 7176 series, ISO 16840-10, ISO 7176-31, and IEC 60601-1-2/IEC 60601-4-2 EMC testing.
Technological Characteristics
Magnesium alloy frame; PET back/seat cushions; brushless DC motors (24VDC, 220W); lithium-ion battery (24V, 9.9Ah or 19.8Ah); joystick-based electronic controller; automatic electromagnetic braking; non-pneumatic/rubber tires; foldable form factor; ISO 7176 series compliance.
Indications for Use
Indicated for disabled or elderly adults (22+ years) requiring mobility assistance while in a seated position for indoor and outdoor use.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 24, 2026
Shanghai Waylong Intelligent Technology Co.,Ltd
% Andrew Wang
Consultant
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Bldg., 1500# Century Ave.
Shanghai,
China
Re: K260937
Trade/Device Name: Smart Electric Wheelchair (R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK)
Regulation Number: 21 CFR 890.3860
Regulation Name: Powered Wheelchair
Regulatory Class: Class II
Product Code: ITI
Dated: May 28, 2026
Received: May 29, 2026
Dear Andrew Wang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260937 - Andrew Wang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260937 - Andrew Wang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

for Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | K260937 | ? |
| --- | --- | --- | --- |
| Please provide the device trade name(s). | | | ? |
| Smart Electric Wheelchair (R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK) | | | |
| Please provide your Indications for Use below. | | | ? |
| The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | | ? |
{4}
# 510(k) Summary
*Document Prepared Date: 2026-03-18*
## 1. Applicant Information
Shanghai Waylong Intelligent Technology Co.,Ltd
Address: 2-3/F, Building 3, No. 238 Huida Road, Zhangyan Town, Jinshan District, Shanghai
Contact: Mrs. Jin Chunxiang
Tel: +86-13564145173
Email: jin_chunxiang@launchdesign.com.cn
Submission Correspondent:
Primary Contact: Andrew Wang
Shanghai SUNGO Management Consulting Co., Ltd.
14th Floor, Dongfang Building, 1500# Central Ave., Shanghai 200122, China
Tel: 0086-21-58817802
Email: yuanju.wang@sungoglobal.com
## 2. Subject Device
Smart Electric Wheelchair, K260937
Model: R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK
Classification regulations: 21 CFR 890.3860
Product code: ITI
Classification: Class II
## 3. Substantial Equivalence Information
Anhui Longway Medical Technology Co., LTD.
Electric Wheelchair (LW01301A07), K250366
Classification regulations: 21 CFR 890.3860
Product code: ITI
Classification: Class II
## 4. Device Description
The Electric Wheelchair is a battery powered four wheeled vehicle.
The wheelchair can easily fold and unfold for transportation or storage.
The max loading of the device is 125 kg. Only for one person sit.
### Model: R1-Std-BK
It consists one Lithium battery with an off-board battery charger, frame, controller, motors, seat, back support, control device (including the battery power indicator, power button, horn button, mute button, speed indicator, speed select button '-', speed select button '+', joystick, light, lighting button, USB port, Battery charger
{5}
socket), arm supports, two rear wheels, two casters(front wheels), foot support, anti-tip devices, push handle, safety belt, storage basket.
# Model: R1-Pro-BK
It consists one Lithium battery with an off- board battery charger, frame, controller, motors, seat, back support, control device (including the battery power indicator, power button, horn button, mute button, speed indicator, speed select button "-", speed select button "+", joystick, light, lighting button, USB port, Battery charger socket), arm supports, two rear wheels, two casters(front wheels), foot support, anti-tip devices, push handle, safety belt, storage basket, mobile phone stand, arm support bag.
# Models: R1-Ultra-GK, R1-Ultra-WK
It consists one Lithium battery with an off-board battery charger, frame, controller, motors, seat, back support, control device (including the battery power indicator, power button, horn button, mute button, speed indicator, speed select button "-", speed select button "+", joystick, light, lighting button, USB port, Battery charger socket), arm supports, two rear wheels, two casters(front wheels), foot support, anti-tip devices, push handle, safety belt, storage basket, head support, mobile phone stand, arm support bag, back support backpack.
# 5. Indication for Use
The device is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position.
# 6. Product parameter
| Elements of Comparison | Subject Device (K260937) | Predicate Device (K250366) | Remark |
| --- | --- | --- | --- |
| Manufacturer | Shanghai Waylong Intelligent Technology Co.,Ltd | Anhui Longway Medical Technology Co., LTD. | -- |
| Common or Usual name | Smart Electric Wheelchair | Electric Wheelchair | -- |
| Model(s) | R1-Std-BK, R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK | LW01301A07 | -- |
| Classification | Class II | Class II | S.E. |
| Classification regulation | 21 CFR890.3860 | 21 CFR890.3860 | S.E. |
| Product code | ITI | ITI | S.E. |
| Indications for use | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | It is a motor driven, indoor and outdoor transportation vehicle with the intended use to provide mobility to a disabled or elderly person limited to a seated position. | S.E. |
{6}
| Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark |
| --- | --- | --- | --- |
| Intended user | disabled or elderly person limited to a seated position. | disabled or elderly person limited to a seated position. | S.E. |
| Use condition | indoor and outdoor use | indoor and outdoor use | S.E. |
| Number of wheels | 4, including two front wheels and two rear wheels | 4, including two front wheels and two rear wheels | S.E. |
| Function of wheels | Front wheels: driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | Front wheels: driven wheels suitable for rotation, acceleration, retrograde Rear wheels: driving wheels to control the speed and direction | S.E. |
| Movement control method | By Joystick control | By Joystick control | S.E. |
| Driving system | Direct drive on the rear wheels | Direct drive on the rear wheels | S.E. |
| Brake system | Automatic electromagnetic brake system | Automatic electromagnetic brake system | S.E. |
| Braking distance | 1.4 m | 1.2 m | Minor difference on braking distance will not cause different performance. Shorter distance for braking will be more safety. |
| Maximum safe operational incline degree | 8.4° (Seat plane angle) / Max slope tested >10.6° | 6° | The subject device supports a slightly higher safe operational incline. It has passed ISO 7176-1, 2, and 3 stability and braking tests at the specified inclines, demonstrating equivalent or better safety. |
| Armrest | PUR | PUR | S.E. |
{7}
| Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark |
| --- | --- | --- | --- |
| Main frame material | Magnesium Alloy | Aluminum Alloy | Although the materials differ, both are lightweight, high-strength alloys. The subject device has fully passed ISO 7176-8 static, impact, and fatigue strength tests with a 125kg test dummy, ensuring equivalent structural safety. |
| Back cushion | PET (Polyethylene terephthalate) | Nylon | Different synthetic polymers are used. However, the PET material of the subject device has passed ISO 16840-10 flammability testing. A Biocompatibility Justification is provided, demonstrating no new biological risks. Thus, they are S.E. |
| Seat cushion | PET (Polyethylene terephthalate) | Nylon | |
| Overall Dimension (length*width*height) | 1040*650*1022mm | 1120*640*1000mm | Minor difference on wheelchair dimension will not cause different performance. All safety and performance have been validated with the maximum rated weight dummy. |
| Folded Dimension (length*width*height) | 780mm*650mm*430mm | 760*640*460mm | |
| Front wheel size/type | R1-Std-BK: 200*50mm (Approx 8"), PU Solid Tire R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK: 200*50mm (Approx 8"), Rubber, Non-Pneumatic hollow Tire | 8" x 2"/PU Solid tire | Minor material differences are mitigated. Both are non-inflatable puncture-proof designs. Safety is verified through ISO 7176 stability and obstacle testing. S.E. |
{8}
| Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark |
| --- | --- | --- | --- |
| Rear wheel size/type | R1-Std-BK: 12.5*2.25inch (≈317.5*57.15mm), Rubber, Inflatable Tire R1-Pro-BK, R1-Ultra-GK, R1-Ultra-WK: 12.5*2.25inch (≈317.5*57.15mm), Rubber, Non-Pneumatic hollow Tire | 12"x 2"/ PU Pneumatic tire | Minor material differences are mitigated. Both are non-inflatable puncture-proof designs. Safety is verified through ISO 7176 stability and obstacle testing. S.E. |
| Max speed forward | Up to 2.3 m/s (8.28 km/h) | Up to 6.84 km/h (1.9 m/s), adjustable | Different max speed forward will not affect safety and performance of the subject device as all related stability tests are performed according to standard ISO 7176 series. |
| Max Speed backward | 0.7 m/s (2.52 km/h) | Less than 3.96 km/h (1.1 m/s) | Different max speed backward will not affect safety and performance of the subject device as all related stability tests are performed according to standard ISO 7176 series. |
| Max loading weight | 125 kg | 120 kg | Difference on loading weight will not cause different performance. All safety and performance have been validated with the maximum loading weight. |
{9}
| Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark |
| --- | --- | --- | --- |
| Maximum distance of travel on the fully charged battery | R1-Std-BK, R1-Pro-BK: 10.3km R1-Ultra-GK, R1-Ultra-WK: 21.3km | 17.6 km | The subject device complies with ISO 7176-4: 2008 Wheelchairs - Part 4: Energy consumption of electric wheelchairs and scooters for determination of theoretical distance range, these differences do not affect safety and effectiveness. |
| Battery | R1-Std-BK, R1-Pro-BK: Li-ion battery pack; 24V, 9.9Ah R1-Ultra-GK, R1-Ultra-WK: Li-ion battery pack; 24V, 19.8Ah | li-ion battery pack; rechargeable, 24 VDC 13Ah | Minor difference on battery capacity will not cause different performance. Larger battery capacity will contribute to the maximum distance. |
| Motor | Brushless DC motor; 24VDC, 220W; 2pcs | Brushless DC motor; 24VDC; 250W; 2pcs | The subject device utilizes slightly lower power motors. However, it successfully passed all ISO 7176 performance requirements, including maximum speed, obstacle climbing, and incline tests, proving the motor power is sufficient and safe. S.E. |
| Electronic controller | Brushless motor controller | Brushless motor controller | S.E. |
| Turning Radius | 925 mm | 958 mm | The turning capabilities are highly comparable. The subject device's turning space meets all requirements for indoor/outdoor maneuverability as verified by ISO 7176-5. S.E. |
{10}
| Elements of Comparison | Subject Device | Predicate Device (K250366) | Remark |
| --- | --- | --- | --- |
| Maximum obstacle climbing | 30 mm | 30 mm | S.E. |
# **Substantial Equivalence Discussion:**
The proposed device and predicate device are complying to the same ISO standards, ISO 7176-1, ISO 7176-2, ISO 7176-3, ISO 7176-4, ISO 7176-5, ISO 7176-6, ISO 7176-7, ISO 7176-8, ISO 7176-9, ISO 7176-10, ISO 7176-11, ISO 7176-13, ISO 7176-14, ISO 7176-15, ISO 16840-10, ISO 7176-21, ISO 7176-25, ISO 7176-31 and FDA guidance Submission for Power Wheelchair.
The proposed device performs in a similar manner to the predicate device. All these tests have corresponding requirements/ control criteria following above mentioned standards. And the test results show that the subject product is substantially equivalent to the predicate device in performance.
The performance testing demonstrates that the subject device is substantially equivalent to the predicate devices regarding Static ability (tipping angle), The Dynamic stability (Safe Gradient Maximum Gradient), Brake performance, Theoretical distance range, Dimension and weight, Maximum speed, Dimension of wheel, Static, impact and fatigue strengths, Climatic tests, Obstacle-climbing ability, Dummy, friction of test surfaces, Power and control systems, Documentation and labeling, Resistance to ignition, Electromagnetic Compatibility and Electrical Safety, Batteries and chargers.
The non-clinical laboratory data support the safety and performance of the subject device and demonstrate that the subject device should perform as intended in the specified use conditions.
# 7. Product Performance
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| Biocompatibility | Comply with ISO 10993-1, FDA Guidance | Comply with ISO 10993-1, FDA Guidance | S.E. |
| EMC | ISO 7176-21, IEC 60601-1-2, IEC 60601-4-2 | ISO 7176-21, IEC 60601-1-2, IEC 60601-4-2 | S.E. |
| Performance | ISO 7176 series | ISO 7176 series | S.E. |
| Label and labeling | Conforms to FDA Regulatory | Conforms to FDA Regulatory | S.E. |
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| ISO 7176-1 | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | The Static stability has been determined after the testing according to the ISO 7176-1, and test results meet its design specification. | S.E. |
| ISO 7176-2 | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | The dynamic stability has been determined after the testing according to the ISO 7176-2, and test results meet its design specification. | S.E. |
| ISO 7176-3 | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | The effectiveness of brakes has been determined after the testing according to the ISO 7176-3, and test results meet its design specification. | S.E. |
| ISO 7176-4 | The theoretical distance | The theoretical distance | S.E. |
{11}
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| | range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification. | range has been determined after the testing according to the ISO 7176-4, and test results meet its design specification. | |
| ISO 7176-5 | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | The dimensions, mass has been determined after the testing according to the ISO 7176-5. | S.E. |
| ISO 7176-6 | The dimensions, mass has been determined after the testing according to the ISO 7176-6. | The dimensions, mass has been determined after the testing according to the ISO 7176-6. | S.E. |
| ISO 7176-7 | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7. | The seating and wheel dimensions has been determined after the testing according to the ISO 7176-7. | S.E. |
| ISO 7176-8 | All test results meet the requirements in Clause 4 of ISO 7176-8. | All test results meet the requirements in Clause 4 of ISO 7176-8. | S.E. |
| ISO 7176-9 | The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9. | The test results shown that the device under tests could continue to function according to manufacturer's specification after being subjected to each of the tests specified in Clause 8 of ISO 7176-9. | S.E. |
| ISO 7176-10 | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | The obstacle-climbing ability of device has been determined after the testing according to the ISO 7176-10. | S.E. |
| ISO 7176-11 | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11. | The test dummies used in the testing of ISO 7176 series are meet the requirements of ISO 7176-11. | S.E. |
| ISO 7176-13 | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved. | The coefficient of friction of test surfaces has been determined, which could be used in other 7176 series tests involved. | S.E. |
| ISO 7176-14 | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14. | All test results meet the requirements in Clause 7, 8, 9, 10, 11, 12, 13, 14, 15, 17 of ISO 7176-14. | S.E. |
| ISO 7176-15 | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15. | The test results shown that information disclosure, documentation and labelling of device meet the requirements of ISO 7176-15. | S.E. |
| ISO 16840-10 | The performance of resistance to ignition meet the requirements of ISO 16840-10:2021. | The performance of resistance to ignition meet the requirements of ISO 16840-10:2021. | S.E. |
| ISO 7176-21 | The EMC performance results meet the requirements of ISO 7176-21. | The EMC performance results meet the requirements of ISO 7176-21. | S.E. |
{12}
| Item | Proposed Device | Predicate Device | Results |
| --- | --- | --- | --- |
| ISO 7176-25 | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25. | The performance of batteries and charger of device meet the Requirements in Clause 5 and 6 of ISO 7176-25. | S.E. |
| ISO 7176-31 | The lithium-ion battery systems and chargers have been tested and meet the requirements of ISO 7176-31:2023. | N/A | S.E. (The subject device demonstrates compliance with the latest FDA-recognized consensus standard for lithium-ion battery safety, exceeding or matching the safety profile of the predicate). |
### 8. Summary of clinical testing
No animal study and clinical studies are available for our device. Clinical testing was not required to demonstrate the substantial equivalence of the power wheelchair to its predicate device.
### 9. Conclusion
While there are minor differences between the subject device and the predicate device—specifically regarding the frame material (Magnesium vs. Aluminum alloy), cushion material (PET vs. Nylon), slightly higher maximum speed (2.3 m/s vs. 1.9 m/s), and higher maximum loading weight (125 kg vs. 120 kg)—these differences do not raise any new questions of safety and effectiveness. The subject device has been comprehensively evaluated through bench performance testing (including the complete ISO 7176 series, ISO 16840-10, ISO 7176-31, and IEC 60601-1-2) at its maximum specifications (e.g., tested with a 125kg dummy at maximum speed). The successful test results demonstrate that the subject device manages these structural and operational differences effectively and safely.
Based on the comparison and analysis above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate devices, K250366 Electric Wheelchair from Anhui Longway Medical Technology Co., LTD.
Signature
Jin chunxiang
Typed Name Jin Chunxiang
Date (YYYY-MM-DD) 2026-03-20
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.