RadioViewAI is a web-based PACS and image management software intended for the display, transmission, storage, and review of multi-modality DICOM images. The software enables trained healthcare professionals such as physicians, radiologists, and certified medical technicians to view and assess medical images within the clinical workflow through web based PACS viewer. All images should be viewed on monitors that meet FDA-recommended technical specifications for diagnostic purposes. RadioViewAI is not intended to replace dedicated diagnostic workstations.
Device Story
RadioViewAI is a web-based, zero-footprint PACS and image management system; operates via cloud architecture or on-premise ConnectAI app; receives DICOM images from hospital PACS/modalities (CT, MR, US, etc.). Users (radiologists, physicians, technicians) access images via standard web browsers; system provides standard visualization tools (window/level, zoom, pan, measurements, annotations). Optionally integrates FDA-cleared third-party AI models to display AI-generated outputs alongside original images; does not modify AI outputs or original DICOM data. Facilitates clinical workflow by enabling study review, comparison, and diagnostic assessment. Benefits include centralized storage, remote access, and integrated visualization of AI-assisted diagnostic findings; does not perform independent diagnostic analysis.
Clinical Evidence
Bench testing only. Validation testing protocol evaluated output functions and actions; results confirmed adherence to predetermined acceptance criteria. No clinical data provided.
Technological Characteristics
Web-based PACS; zero-footprint browser interface. Cloud-hosted (HITRUST certified) or on-premise (ConnectAI app). Supports DICOM standards (NEMA PS 3.1-3.20). Role-based access control; encrypted data transmission/storage. Compliant with IEC 62304, ISO 14971, IEC 62366-1, ISO 13485.
Indications for Use
Indicated for viewing and assessing multi-modality DICOM images (DX, DR, CR, CT, MR, US, RF, XA, XR, NM, PT, 2D/3D mammography) by proficient medical experts including physicians, radiologists, and medical technicians. Lossy compressed mammographic and digitized film screen images are for reference only, not primary diagnostic interpretation.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 31, 2026
Neurocareai, Inc. (Dba Savelife.Ai)
% Junaid Siddiq Kalia
Chief Executive Officer
1740 Lonesome Dove Dr.
PROSPER, TX 75078
Re: K260936
Trade/Device Name: RadioViewAI
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: June 23, 2026
Received: June 23, 2026
Dear Junaid Siddiq Kalia:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260936 - Junaid Siddiq Kalia
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K260936 - Junaid Siddiq Kalia
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Jessica S. Lamb, Ph.D.
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260936
Device Name
RadioViewAI
Indications for Use (Describe)
RadioViewAI is a web-based PACS and image management software, used for viewing and assessing multi modality DICOM images Ex: DX, DR, CR, CT, MR, US, RF, XA, XR, NM, PT, and 2D/3D mammography. It gathers digital images and information from a variety of sources adhering to DICOM standard, including PACS systems, digital and computed radiographic equipment, CT and MR scanners, ultrasound and RF machines, PET units, secondary capture tools and imaging gateways. RadioViewAI facilitates transmission, storage, processing and visualization of DICOM images and data within the system itself or over computer networks spanning different locations.
Lossy compressed mammographic images and digitized film screen images as received from hospitals must be reviewed for reference purposes only, not for primary diagnostic interpretation on the RadioViewAI. To ensure quality, DICOM images should only be viewed on a monitor that adheres to the technical specifications outlined by the FDA.
The software shall only be used by proficient and certified medical experts, including physicians, radiologists, and medical technicians.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{5}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission

## NeuroCare.AI
### 510(k) Summary of RadioViewAI
by
### NEUROCAREAI INC (DBA SaveLife.AI)
Applicant Name: NEUROCAREAI INC. (DBA SaveLife.AI)
1740 Lonesome Dove Dr
Prosper, TX 75078 USA
Phone Number: +1 (214) 346-6083
Whatsapp: +1 (469) 954-0346
Contact Person: Junaid Kalia
Chief Executive Officer
Email: junaidkalia@neurocare.ai
Date Prepared: March 19, 2026
### Device Name and Classification
Name of Device: RadioViewAI
510k Number: K260936
Classification Name: Medical Image Management and Processing System
Common or Usual Name: System, Image Processing, Radiological
Classification Panel: Radiology
Regulation Number: 21 CFR 892.2050
Regulatory Class: Class II
Product Code: LLZ
Predicate Device:
| Manufacturer | Device Name | Product Code | Application Number |
| --- | --- | --- | --- |
| CARPL.AI Inc. | CARPL | LLZ | K232891 |
510(k) Summary
Page 1 of 9
{6}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
# Device Description:
RadioViewAI is a web-based PACS that provides comprehensive image viewing capabilities through a zero-footprint browser-based interface accessible via standard web browsers without requiring client-side software installation. The system operates on a secure cloud architecture where medical images and associated metadata are stored on cloud and accessed remotely by authorized users through encrypted web connections. The software receives medical images from Hospital PACS and other DICOM-compliant imaging modalities including DX, DR, CR, CT, MR, US, RF, XA, XR, NM, PT, and 2D/3D mammography, and provides secure, centralized storage while maintaining image integrity and associations with patient demographic and study information.
The web-based viewer provides standard image manipulation tools including window/level adjustment, zoom, pan, rotation, measurements, and annotations, enabling simultaneous display of multiple studies for comparison of current and prior examinations. The system is fully compliant with DICOM standards for image communication and storage, supporting DICOM services including retrieval, storage, transmission, and visualization of the DICOM images for radiologist's review and interpretation. The software integrates with hospital PACS systems through DICOM interfaces, providing worklist management and study routing capabilities, and supports radiologist reading workflows.
When assessing images for diagnostic purposes, it becomes the duty of the healthcare expert to ascertain whether the image quality is appropriate for clinical use. The system offers the choice to incorporate third-party AI models that have been cleared by the FDA. A regulatory compliance team ensures that only FDA-cleared third-party algorithms are made available within RadioViewAI. Each algorithm developer must prove regulatory clearances before their products can be incorporated with RadioViewAI. The solution solely aids in visualizing the outcomes of these third-party AI models without modifications. The safety and efficacy of the third-party model are governed by the regulatory approval granted to the original manufacturer of the said model.
RadioViewAI integrated FDA-cleared algorithms list is exclusively managed by NEUROCAREAI, customers do not have the ability technically or administratively to add or modify which FDA-cleared AI algorithms are integrated with RadioViewAI. Only FDA-cleared algorithms which have passed rigorous integration testing, regulatory and quality standards review by NEUROCAREAI to guarantee functionality, safety and security are made available in RadioViewAI.
The system implements role-based access control with user authentication, and employs encryption for data transmission and storage in compliance with HIPAA security and privacy requirements. The software can be deployed on cloud via PACS to PACS connection or on-premises ConnectAI app over standard TCP/IP networks with appropriate bandwidth for image transmission. User access is supported through modern web browsers (Chrome, Firefox, Safari, Edge) on desktop or laptop devices. The system maintains full compliance with DICOM standards, and supports all standard DICOM image formats and modalities for comprehensive medical imaging workflow management.
RadioViewAI simply presents the simple AI response output, and the initial anonymized image remains consistently available. The duty of evaluating the AI output, validating the results, and conducting the diagnosis lies with competent medical professionals. RadioViewAI current list of integrated FDA cleared AI Models:
510(k) Summary
Page 2 of 9
{7}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
| Device Name | FDA 510(K)-Number | Company Name |
| --- | --- | --- |
| Hyper Insight - ICH | K240353 | Sk, Inc. (PurpleAI) |
| NeuroICH | K241719 | NEUROCAREAI INC (SaveLife.AI) |
# **Intended Use:**
RadioViewAI is a web-based PACS and image management software intended for the display, transmission, storage, and review of multi-modality DICOM images. The software enables trained healthcare professionals such as physicians, radiologists, and certified medical technicians to view and assess medical images within the clinical workflow through web based PACS viewer. All images should be viewed on monitors that meet FDA-recommended technical specifications for diagnostic purposes. RadioViewAI is not intended to replace dedicated diagnostic workstations.
# **Indications for Use:**
RadioViewAI is a web-based PACS and image management software, used for viewing and assessing multi modality DICOM images Ex: DX, DR, CR, CT, MR, US, RF, XA, XR, NM, PT, and 2D/3D mammography. It gathers digital images and information from a variety of sources adhering to DICOM standard, including PACS systems, digital and computed radiographic equipment, CT and MR scanners, ultrasound and RF machines, PET units, secondary capture tools and imaging gateways. RadioViewAI facilitates transmission, storage, processing and visualization of DICOM images and data within the system itself or over computer networks spanning different locations.
Lossy compressed mammographic images and digitized film screen images as received from hospitals must be reviewed for reference purposes only, not for primary diagnostic interpretation on the RadioViewAI. To ensure quality, DICOM images should only be viewed on a monitor that adheres to the technical specifications outlined by the FDA.
The software shall only be used by proficient and certified medical experts, including physicians, radiologists, and medical technicians.
RadioViewAI incorporates the following:
1. RadioViewAI supports two integration options for receiving imaging data from hospital PACS systems:
- **Direct PACS to PACS Connection:** In this configuration, the hospital PACS or DICOM server is configured to send imaging data directly to the NEUROCAREAI Orthanc server hosted on the HITRUST certified cloud. This direct connection enables seamless transfer of DICOM studies from the hospital infrastructure to the RadioViewAI cloud environment without requiring any on-premise intermediary application.
- **ConnectAI App On-Premise Integration:** In this configuration, the ConnectAI application is installed on-premise at the hospital facility. The hospital PACS or
510(k) Summary
Page 3 of 9
{8}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
DICOM server is configured to send imaging data directly to the ConnectAI app and securely routes it to the main Orthanc server on the HITRUST certified cloud
In both integration options, the database syncs itself with the cloud Orthanc server and pushes the data as soon as received by the hospital PACS/DICOM server.
2. It has a dataset manager i.e., study list/worklist, which includes all the studies that are uploaded to the system. The worklist provides search and filter capabilities to manage patient studies efficiently.
3. It has a radiology workflow management that allows administrators to manage user roles. Radiologists can then view the study images, diagnose them, and complete their reading workflow through the RadioViewAI web viewer.
4. Radiologists use it to view the DICOM images for diagnosis and assessment. The PACS viewer application offers standard PACS functionality provided by the OHIF (Open Health Imaging Foundation) viewer framework. The viewer supports DICOM images from Digital X-Ray (DX), Digital Radiography (DR), Computerized Radiography (CR), Ultrasound (US), Computed Tomography (CT), Magnetic Resonance (MR), Nuclear Medicine (NM), Digital Mammography (MG), Positron Emission Tomography (PT), Radiographic imaging (XR), Radio Fluoroscopy (RF), and X-Ray Angiography (XA).
5. It only incorporates algorithms that have received FDA clearance. A regulatory compliance team at NEUROCAREAI ensures that only FDA-cleared third-party algorithms are made available within RadioViewAI. Each algorithm developer must demonstrate regulatory clearances before their products can be incorporated. NEUROCAREAI verifies the 510(k) clearance status against the FDA database before proceeding with integration.
6. Where enabled, the system optionally forwards eligible DICOM studies to external FDA-cleared third-party AI models. The AI Model Service transmits DICOM instances to the external AI model. The external AI model processes the received DICOM instances in accordance with its FDA-cleared intended use and returns its output to the RadioViewAI environment.
7. It can provide priority status indicators for the studies in the worklist based on the outputs provided by the third-party AI models that have been cleared for this intended use.
8. For models cleared to assist the radiologist, it displays the output of the third-party AI model in the Viewer for visualization by the Radiologist to assist in diagnosing the study. RadioViewAI does not modify AI-generated DICOM objects prior to display.
9. The OHIF Viewer integrated into the RadioViewAI platform is a zero-footprint, browser-based DICOM viewer. It supports key image visualization features including window/level adjustments, zoom, pan, rotation, stack scrolling, series comparison, and layout customization. Additional tools include distance and angle measurements, region-of-interest (ROI) annotations, overlay display controls, and metadata panels displaying patient and study information derived from DICOM headers.
10. RadioViewAI provides a list of annotation and measurement tools that the user can use to annotate a case and carry out measurements. The following tools are provided:
- Length: The user can use this to measure the distance between two points on the image, which is expressed in mm/cm. The user places the cursor over the starting point and drags to draw a segment with visible measurement.
- Annotate: Users can mark a region of interest on the image. After the user finishes the marking, a label can be added to the annotated area for documentation.
510(k) Summary
Page 4 of 9
{9}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
- Angle: Angle tool can be used when a user wants to measure the angle between two lines. The user draws the first arm of the angle and releases it, and similarly draws the second arm, after which the value is displayed in degrees.
- Bidirectional: It is a tool to measure a region of interest's length and width. The user annotates a suspected lesion; using this tool the user can obtain the length and width of the marked ROI. The values are displayed in mm.
- Ellipse: This tool helps users mark a region of interest on an image in a defined elliptical border. The user draws an ellipse by dragging the mouse cursor on the ROI and releasing it to complete the shape.
- Rectangle: Similar functions to Ellipse. The only difference is the shape of the ROI, which appears as a rectangle.
- Freehand/Close Polygon: This is used for giving a freehand shape to a region of interest. The user clicks to create nodes around the ROI and completes the shape.
- Eraser: The user can click on a marked annotation and erase it.
- Reset: This button restores the study image to its original state.
There are no significant differences with the predicate device CARPL, that raise new questions of safety or effectiveness.
| Features compared | Subject Device (RadioViewAI) | Predicate Device (CARPL) K232891 |
| --- | --- | --- |
| **Manufacturer** | NEUROCAREAI INC. | CARPL.AI Inc. |
| **Common Name** | Medical image management and processing system | Medical image management and processing system |
| **Classification Name** | System, Image Processing, Radiological | System, Image Processing, Radiological |
| **Classification** | Class II | Class II |
| **Regulation Number** | 21 CFR 892.2050 | 21 CFR 892.2050 |
| **Product Code** | LLZ | LLZ |
| **Indications For Use** | RadioViewAI is a web-based PACS and image management software, used for viewing and assessing multi modality DICOM images Ex: DX, DR, CR, CT, MR, US, RF, XA, XR, NM, PT, and 2D/3D mammography. It gathers digital images and information from a variety of sources adhering to DICOM standard, including PACS systems, digital and computed radiographic equipment, CT and MR scanners, ultrasound | CARPL, a web-based PACS and radiology workflow management device, used for viewing and assessing DICOM images Ex: DX, DR, CR, CT, MR, US, RF and 2D/3D mammography. It gathers digital images and information from a variety of sources that adhere to the DICOM standard. These sources encompass a range of devices, including digital and computed radiographic equipment, CT and MR scanners, ultrasound and RF machines, PET units, secondary capture tools, imaging gateways. |
510(k) Summary
Page 5 of 9
{10}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
| | and RF machines, PET units, secondary capture tools and imaging gateways. RadioViewAI facilitates transmission, storage, processing and visualization of DICOM images and data within the system itself or over computer networks spanning different locations.Lossy compressed mammographic images and digitized film screen images as received from hospitals must be reviewed for reference purposes only, not for primary diagnostic interpretation on the RadioViewAI. To ensure quality, DICOM images should only be viewed on a monitor that adheres to the technical specifications outlined by the FDA.The software shall only be used by proficient and certified medical experts, including physicians, radiologists, and medical technicians. | CARPL enables the storage, transmission, processing, and visualization of images and data within the system itself or over computer networks spanning different locations.Only pre-processed DICOM images specifically intended for presentation are suitable for primary image diagnosis in mammography. Lossy compressed Mammographic images and digitized film screen images must not be reviewed for primary image interpretation on the CARPL. To ensure accurate interpretation, mammographic images should only be evaluated on a monitor that adheres to the technical specifications outlined by the FDA.This system is designed exclusively for use by proficient and certified medical experts, including physicians, radiologists, and medical technicians. |
| --- | --- | --- |
| Modalities | DX, DR, CR, CT, MR, US, RF, XA, XR, NM, PT, and 2D/3D Mammography | DX, DR, CR, CT, MR, US, RF and 2D/3D Mammography |
| Web Browser Software | Google Chrome, Mozilla Firefox, Safari, and Edge | Google Chrome, Mozilla, and Edge |
| Resolution | 32 bit Color Display & 1920x1080 | 32 bit Color Display & 1920x1080 |
| Image Storage | YES | YES |
| Software Environment | ConnectAI App: Windows 11 or higher; Cloud: HITRUST certified environment; Viewer: Zero-footprint web application | OS: Windows 10 or higher |
| Main Functions | • Log In• Worklist – Search Filter• Worklist – Open image• Worklist – Study List• Worklist – Series• Viewer – View exam | • Log In• Worklist – Search Filter• Worklist – Open image• Worklist – Study List• Worklist – Report• Worklist – Series |
510(k) Summary
Page 6 of 9
{11}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
| | • Viewer – Control View window • Viewer – View mode (real resolution) • Viewer – Stacking • Viewer – Changing the layout • Viewer – Comparative study • Viewer – Preset filter • Viewer – Zoom • Viewer – Panning • Viewer – Invert Image • Viewer – Rotation • Viewer – Reference line • Viewer – Measure • Viewer – Inverting image color • Viewer – Cine • Viewer – Overlaying • Viewer – Window/Level | • Viewer – View exam • Viewer – Control View window • Viewer – View mode (real resolution) • Viewer – Highlight • Viewer – Stacking • Viewer – Changing the layout • Viewer – Comparative study • Viewer – Preset filter • Viewer – Zoom • Viewer – Panning • Viewer – Invert Image • Viewer – Rotation • MIP/MPR Reconstruction • Viewer – Reference line • Viewer – Sharpening • Viewer – Measure • Viewer – Inverting image color • Viewer – Cine • Viewer – Overlaying |
| --- | --- | --- |
| **Optional Integration of FDA-cleared 3rd party AI models** | YES - Optional integration of FDA-cleared third-party AI models (visualization only; no modification of outputs). | YES - Optional integration of FDA-cleared third-party AI models (visualization only; no modification of outputs). |
| **Device Components** | • ConnectAI app installed on-premise for DICOM routing • Cloud-based PACS storage (HITRUST certified) • Zero-footprint web based PACS viewer (OHIF) • Admin panel managed by NEUROCAREAI • Role-based access and secure encrypted communication • Optional integration layer for FDA-cleared third-party AI algorithms | • CARPL Gateway installed on-site for DICOM routing • CARPL Viewer with MPR/MIP tools • Workflow manager (study assignments, collaboration) • Optional integration layer for FDA-cleared third-party AI algorithms |
| **Operation Features** | • Web environment-based PACS • Viewing and handling DICOM medical images • Review study located on cloud • View multiple AI model output • Role-based access control | • Web environment-based PACS • Viewing and handling DICOM medical images • Review and report study located on a server • View multiple AI model output |
510(k) Summary
Page 7 of 9
{12}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
The technological principle for both the subject and predicate devices is the same in terms of prescription use, support of various modalities, resolution, image storage, use of the DICOM standard, and operation features. Most of the features, specifications, and functions of both the subject and predicate devices, like indications for use, software web browser, software intended environment, study viewer, and study worklist are similar.
RadioViewAI provides the feature of optional integration with external FDA-cleared third-party AI models like the predicate device CARPL does. The integration with the FDA-cleared third-party AI models is optional based on the user's discretion, and always in accordance with the third-party manufacturer's regulatory clearance. RadioViewAI integrated FDA-cleared algorithms list is exclusively managed by NEUROCAREAI. Customers do not have the ability technically or administratively to add or modify which FDA-cleared AI algorithms are integrated with RadioViewAI. Only FDA-cleared algorithms which have passed rigorous integration testing, regulatory and quality standards review by NEUROCAREAI to guarantee functionality, safety and security are made available in RadioViewAI. Hence, the difference in the list of offered AI algorithms does not affect the product effectiveness and safety.
Both devices provide similar basic image manipulation tools like window/level adjustment, zoom, pan, rotation, measurements, and annotations, enabling simultaneous display of multiple studies for comparison of current and prior examinations. The minor differences in the capability of tools provided does not raise any new questions of safety and effectiveness. Both devices have the capability to integrate and display outputs from external AI algorithms. Neither device performs diagnostic analysis, or alters medical images. Both devices are intended to be used by certified medical experts including physicians, radiologists and medical technicians who are experts in the independent diagnostic review and interpretation of medical images.
Any differences between the subject and predicate devices are limited to backend architectural implementation, including microservice-based synchronization workflows and cloud scalability mechanisms. These differences do not affect intended use, core technological principles, safety, or effectiveness and do not introduce new potential risks.
## **PERFORMANCE DATA**
### **Non-clinical testing:**
RadioViewAI has been tested and has passed all predetermined testing criteria. The Validation Testing protocol was designed to evaluate output functions and actions performed by NEUROCAREAI INC. and followed the process documented in the Software Validation Testing Protocol. Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met.
Based on the performance as documented in the Validation Testing, RadioViewAI was found to have a safe and effectiveness profile that is substantially equivalent to the predicate device.
The following Standards were used to develop RadioViewAI, and the device has met all the requirements listed in the Standards except for inapplicable requirements:
510(k) Summary
Page 8 of 9
{13}
NEUROCAREAI INC.
RadioViewAI 510(k) Submission
| Standard | FDA Recognition # | Title |
| --- | --- | --- |
| IEC 62304:2006+A1:2015 | 13-79 | Medical device software – Software life cycle processes |
| ISO 14971:2019 | 4-392 | Medical devices – Application of risk management to medical devices |
| ISO/TR 24971:2020 | N/A (Technical Report) | Guidance on the application of ISO 14971 |
| ISO/TR 80002-1:2009 | N/A (Technical Report) | Guidance on the application of ISO 14971 to medical device software |
| IEC 62366-1:2015+A1:2020 | 5-166 | Application of usability engineering to medical devices |
| ISO 13485:2016 | 1-69 | Quality management systems – Requirements for regulatory purposes |
| Nema PS 3.1-3.20 2024e | 12-363 | Digital Imaging and Communications in Medicine (DICOM) set |
## CONCLUSION
The information presented in the 510(k) for RadioViewAI contains adequate information, data, and nonclinical (validation) test results to demonstrate substantial equivalence to the predicate device. RadioViewAI was shown to be substantially equivalent to the predicate device CARPL (K232891) in the areas of technical characteristics, general function, application, and intended use and does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.
510(k) Summary
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.