The device is indicated to treat androgenetic alopecia for men and women. It is designed to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV.
Device Story
Dome-shaped low-level laser therapy (LLLT) helmet; treats androgenetic alopecia. Device contains 88-400 diode lasers (650±20nm, 5mW each); total output 440-2000mW. Operated by patient at home; fixed treatment durations (20-30 minutes). Device exposes scalp to red light to promote hair growth. No operator-adjustable emission levels. Powered by internal lithium battery via 5V/2A adapter. Safety features include compliance with IEC 60601-1, 60601-1-11, 60601-1-2, and 60825-1 (Class 3R).
Clinical Evidence
Bench testing only. No clinical data provided. Evidence includes biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -11), EMC (IEC 60601-1-2), and laser safety (IEC 60825-1).
Technological Characteristics
ABS plastic helmet; 88-400 diode lasers (650±20nm); 5mW per diode; Class 3R laser (IEC 60825-1). Powered by internal lithium battery. Biocompatibility per ISO 10993-5, 10993-10, 10993-23. Fixed treatment parameters.
Indications for Use
Indicated for men and women with androgenetic alopecia; males with Norwood-Hamilton IIa-V; females with Ludwig-Savin I-II; Fitzpatrick Skin Phototypes I-IV.
Regulatory Classification
Identification
An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.
Special Controls
*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 16, 2026
Zdeer Technology (Henan) Co., Ltd.
Wenlin Xiang
Manager
East Of Wei Wu Rd. And S. Of Qin Gong E. Rd.,
Toubao Village Committee, Laotown Town
Changge, Xuchang City, Henan 461500
China
Re: K260899
Trade/Device Name: Laser Hair Growth Device (F21151, F21152, F21153, F21154, F21155, F21156, F21157, F21158, F21159)
Regulation Number: 21 CFR 890.5500
Regulation Name: Infrared Lamp
Regulatory Class: Class II
Product Code: OAP
Dated: March 18, 2026
Received: March 18, 2026
Dear Wenlin Xiang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260899 - Wenlin Xiang
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260899 - Wenlin Xiang
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S Digitally signed by TANISHA L. HITHE -S Date: 2026.06.16 19:12:44 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K260899
Device Name
Laser Hair Growth Device (F21151, F21152, F21153, F21154, F21155, F21156, F21157, F21158, F21159)
Indications for Use (Describe)
The device is indicated to treat androgenetic alopecia for men and women. It is designed to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
510(k)s –510(k) Summary---
# K260899 - 510(k) Summary
'510(k) Summary' as required by 21 CFR Part 807.92.
**Date: 2026-05-21**
## I. Submitter
| **Manufacturer:** | Zdeer Technology (Henan) Co., Ltd East of Wei Wu Road and south of Qin Gong East Road, Toubao Village Committee, Laotown Town, Changge, Xuchang City, Henan Province, P.R. China +86-15637477318 |
| --- | --- |
| **Direct contact:** | Wenlin Xiang (Manager) Tel: +86-15527272371 Email: 1132075798@qq.com |
## II. Correspondent Consultant
Zdeer Technology (Henan) Co., Ltd
East of Wei Wu Road and south of Qin Gong East Road, Toubao Village Committee, Laotown Town, Changge, Xuchang City, Henan Province, P.R. China
Wenlin Xiang (Manager)
Tel: +86-15527272371
Email: 1132075798@qq.com
## III. Device
| **Trade/Proprietary Name:** | Laser hair growth device |
| --- | --- |
| **Model:** | F21151, F21152, F21153, F21154, F21155, F21156, F21157, F21158, F21159 |
| **Common Name:** | Laser hair growth device |
| **Classification Name:** | Laser, Comb, Hair |
| **Classification:** | Regulation Number: 21 CFR 890.5500 |
| **Product Code1:** | OAP |
1
{5}
510(k)s–510(k) Summary
| Review Panel: | Physical Medicine |
| --- | --- |
| Regulatory Class | II |
#### IV. Predicate Device
| Applicant | Predicate Device | 510(k) Number |
| --- | --- | --- |
| Freedom Laser Therapy, Inc. | ID-510 iRestore Elite | K222081 |
| Shenzhen CosBeauty Technology Co., Ltd. | Hair Care 88 | K213447 |
| Shenzhen Kaiyan Medical Equipment Co., Ltd. | M-130,M-300;M-180A | K253669 |
#### V. Device Description
The device is a low-level laser system operating at 650±20nm. The physical configuration is that of a helmet containing an inner and outer liner, stabilized with a silicone pad and a pair of headbands. The device helmet is constructed from ABS plastic. An internal lithium battery should be charged by a 5V, 2A adapter which complies with safety standards (not provided by zdeer). The helmet contains 88, 120, 160, 200, 240, 280, 320, 360 and 400 units of 5 mW diode lasers that emit red light. This system delivers fixed laser emission levels measured to be a total of 440, 600, 800, 1000, 1200, 1400, 1600, 1800, 2000 mW, which cannot be altered by the operator. The treatment durations are also fixed by design: 30 minutes for models F21151~5 and 20 minutes for models F21156~9 respectively
Laser hair growth device is a dome-shaped low-level laser therapy (LLLT) device designed to promote hair growth in women and men by exposing the entire scalp to the light therapy of visible red light-emitting diodes at 650nm and 5 mW each. Laser hair growth device includes 9 models, F21151, F21152, F21153, F21154, F21155, F21156, F21157, F21158, F21159. F21159 is the typical model which has maximum output. All models are the same in structure, intended use, performance and operation. The difference is the number of diodes and laser emission type.
#### VI. Indications for Use
The device is indicated to treat androgenetic alopecia for men and women. It is designed to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V, and in females with androgenetic alopecia who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV.
2
{6}
510(k)s –510(k) Summary
# VII. Comparison of Technological Characteristics With the Predicate Device
Laser hair growth device raises no safety or efficacy concerns when compared to the predicate devices.
A technical comparison to the predicate is provided below:
| Comparison Elements | Subject Device | Predicate Device I | Predicate Device II | Predicate Device III | Comparison |
| --- | --- | --- | --- | --- | --- |
| K Number | Pending | K222081 | K213447 | K253669 | / |
| Trade name | Laser hair growth device | ID-510 iRestore Elite | LLLT Laser Hair Growth Cap, Model: Hair Care 88, Hair Care 135, Hair Care 210 | Aduro Hair Growth; Trudermal Halo Hair growth system | / |
| Model | F21151, F21152, F21153, F21154, F21155, F21156, F21157, F21158, F21159 | ID-510 iRestore Elite | Hair Care 88 | M-130,M-300;M-180A | / |
| Classification name | Laser, Comb, Hair | Laser, Comb, Hair | Laser, Comb, Hair | Laser, Comb, Hair | / |
| Product code | OAP | OAP | OAP | OAP | / |
| Intended purpose/Indications for Use | Laser hair growth device is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V or females with androgenic alopecia who have Ludwig-Savin Classifications of I-II and | The ID-510 iRestore Elite is indicated to promote hair growth in males who have Norwood-Hamilton Classifications of IIa-V, and in females with androgenetic alopecia | LLLT Laser Hair Growth Cap is indicated to promote hair growth in males with androgenic alopecia who have Norwood-Hamilton classifications of IIa-V or females with | The device are indicated to treat Androgenetic Alopecia and promote hair growth in males who have Norwood- Hamilton Classifications of IIa - V patterns of hair loss and to treat Androgenetic Alopecia and promote hair growth in females | Same |
3
{7}
510(k)s –510(k) Summary
| | both with Skin Phototypes I-IV. | who have Ludwig-Savin Classifications of I-II, and in both, Fitzpatrick Classification of Skin Phototypes I to IV. | androgenic alopecia who have Ludwig-Savin Classifications of I-II and both with Fitzpatrick Skin Phototypes I-IV. | who have Ludwig-Savin Scale I-1 to I-4, II-1, II-2 or frontal patterns of hair loss; both with Fitzpatrick Skin Types I - IV. | |
| --- | --- | --- | --- | --- | --- |
| Location for use | Head | Head | Head | Head scalp | Same |
| OTC or Rxonly | OTC | OTC | OTC | OTC | Same |
| Type of laser | Laser diodes | Laser diodes | Laser diodes | Laser diodes | Same |
| Wavelength of laser and LLLT | 650±20nm | 680±10nm 625±10 nm and 655±10nm | 650 nm±10 nm | Red+laser 650 nm +/-5nm Red 660nm+/-10nm | Same red laser |
| Amount of laser diodes | 88,120 160, 200, 240,280, 320, 360, 400 | 300 units of 5 mw diode lasers and 200 units of 5 mw LED | 88 | M-130: 130 Red LEDS M-180A: 180 Red LEDS M-300: 200 Red LEDS +100 Laser | Energy emitted is within the range of the Predicate Device |
| Irradiance (mW/cm²) | <5mW/cm² | / | / | <5 mW/cm² | Minor different Note1 |
| Total dose(J/cm²) | F21151~5:7.80J/cm², 7.81J/cm², 7.68J/cm², 7.75J/cm², 7.66J/cm², F21156~9: 3.09J/cm², | / | / | M-130,M-180A:3J/cm² M-300: 7.5J/cm² | Minor different Note 1 |
4
{8}
510(k)s –510(k) Summary
| | 3.18J/cm², 3.23J/cm² 3.12J/cm², | | | | |
| --- | --- | --- | --- | --- | --- |
| Energy of per laser diode | ≤5mW | 5mW | 5mW ± 20% | 5mW | Same |
| Classification according to IEC 60825-1 | Class 3R | Class 3R | Class 3R | Class 3R | Same |
| Treatment time | 16 weeks, for 30/12-minute treatment times | 16 weeks, for 12-minute treatment times | 16 weeks, for 30-minute treatment times | for 16 weeks M-130,M-180A:10 minutes M-300: 25 minutes | Does not raise any new questions of safety or effectiveness as compared to the predicate device. |
| Treatment frequency | 3 times per week spaced out every other day | 3 times per week spaced out every other day | 3 times per week spaced out every other day | every other day | Same |
| Applicable people | Norwood-Hamilton IIa–V (males) Ludwig-Savin I–II (females) | Norwood-Hamilton IIa–V (males) Ludwig-Savin I–II (females) | Norwood-Hamilton IIa–V (males) Ludwig-Savin I–II (females) | Norwood-Hamilton IIa–V (males) Ludwig-Savin I–II (females) | Same |
| Applicable skin | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Fitzpatrick Skin Phototypes I-IV | Same |
| Shape design | Helmet | Helmet | Helmet | Helmet | Same |
| Safety feature | Complied with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 60825-1 | Complied with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 60825-1 | Complied with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 and IEC 60825-1 | Complied with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2 | Same |
5
{9}
510(k)s –510(k) Summary
| Biocompatibility | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10, ISO 10993-23. | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10, ISO 10993-23. | All body-contacting materials are complied with ISO 10993-5 and ISO 10993-10, ISO 10993-23. | All body-contacting materials are complied with ISO 10993-5 and ISO 10993-10, ISO 10993-23. | Same |
| --- | --- | --- | --- | --- | --- |
Note1: Although the irradiance and total dose values were not obtained from Predicate Devices I and II, they fall within the established range of Predicate Device III for models F21156~9. For models F21151~5, the irradiance and total dose are slightly higher than those of Predicate Device III; however, these minor variances between the subject device and the predicate devices do not raise new questions of safety or effectiveness.
6
{10}
510(k)s –510(k) Summary
### VIII. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of Laser hair growth device was conducted in accordance with the "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices -Part 1: Evaluation and Testing Within a Risk Management Process, Document Issued on September 4, 2020', as recommended by FDA. The testing was performed to, and passed, including:
- ISO 10993-5:2009/(R)2014, Biological Evaluation of Medical Devices –Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2021, Biological Evaluation of Medical Devices –Part 10: Tests for Skin Sensitization
- ISO 10993-23:2021 First edition 2021-01, Biological evaluation of medical devices - Part 23: Tests for irritation
#### 2) Electrical and EMC Safety
Electrical safety and EMC safety testing was performed to, and passed, the following standards:
- IEC 60601-1:2020 Medical electrical equipment –Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-11:2020 Medical electrical equipment –Part 1-11: General requirements for basic safety and essential performance –Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-1-2 :2020 Medical electrical equipment –Part 1-2: General requirements for basic safety and essential performance –Collateral standard: electromagnetic compatibility – Requirements and tests
In addition, testing to IEC 60825-1:2007 certifies the laser system to classification 3R, which is the same as the predicate devices.
#### 3) Software verification and validation
Software documentation consistent with Basic Documentation was provided. Verification and validation test were conducted according to the FDA “Guidance for Pre-Market Submissions and for Software Contained in Medical Devices”
#### Summary
Based on the above performance as documented in this application, the Laser hair growth device was found to have a safety and effectiveness profile that is the same as the predicate device.
### VIII. Conclusions
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the comparison of intended use, design, materials and performance, the laser hair growth device is to be concluded substantially equivalent with its predicate devices.
7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.