Better Diagnostics Dental Assist (BDDA) Version 1.0 is a computer aided detection ("CADe") software intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. The device is also intended to identify and mark regions in relation to suspected dental findings which include marginal discrepancy, calculus, periapical radiolucency (Only IOPA), crowns, filling and root canals. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. The intended patient population of the device is patients who are 18 years or older and do not have any remaining primary teeth.
Device Story
BDDA v1.0 is a CADe software that processes bitewing and periapical dental radiographs (PNG, BMP, JPG) to assist dentists. It uses cloud-hosted computer vision models to detect dental findings (calculus, marginal discrepancy, periapical radiolucency, crowns, fillings, root canals) and measure CEJ-to-alveolar bone distance. The system integrates with dental practice management software via API. The presentation layer displays AI-generated bounding boxes, polygons, and bone-level measurements on an 'AI Results Screen'. Dentists review these annotations concurrently with original images to support clinical decision-making. The device does not provide diagnoses or treatment recommendations; it serves as an adjunct tool to improve diagnostic accuracy. It is intended for use by dentists and hygienists in clinical settings.
Clinical Evidence
Evidence includes standalone and MRMC reader studies. Standalone testing (n=1,318 images) evaluated sensitivity/specificity against expert consensus; all primary endpoints met (e.g., Restoration sensitivity 0.988-0.989, Bone Level MAE 0.203-0.235mm). MRMC study (15 dentists, 1,397-1,451 images) showed statistically significant improvement in AFROC AUC for all anomalies (e.g., Calculus BW AUC 0.809 to 0.913, p<0.05). Surface-level sensitivity/specificity also significantly improved with BDDA assistance.
Technological Characteristics
Cloud-based CADe software; processes PNG, BMP, JPG images. Uses computer vision models for object detection, segmentation, and key point detection. Integrates via API with dental practice management software. Presentation layer uses Angular.js/node.js. Complies with IEC 62304 (software lifecycle), IEC 62366-1 (usability), and ISO 14971 (risk management).
Indications for Use
Indicated for patients 18+ years old with permanent dentition requiring dental services. Used by dental professionals to aid in measuring mesial/distal bone levels and identifying marginal discrepancy, calculus, periapical radiolucency (IOPA only), crowns, fillings, and root canals on bitewing and periapical radiographs.
Regulatory Classification
Identification
Medical image analyzers, including computer-assisted/aided detection (CADe) devices for mammography breast cancer, ultrasound breast lesions, radiograph lung nodules, and radiograph dental caries detection, is a prescription device that is intended to identify, mark, highlight, or in any other manner direct the clinicians' attention to portions of a radiology image that may reveal abnormalities during interpretation of patient radiology images by the clinicians. This device incorporates pattern recognition and data analysis capabilities and operates on previously acquired medical images. This device is not intended to replace the review by a qualified radiologist, and is not intended to be used for triage, or to recommend diagnosis.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Design verification and validation must include:
(i) A detailed description of the image analysis algorithms including a description of the algorithm inputs and outputs, each major component or block, and algorithm limitations.
(ii) A detailed description of pre-specified performance testing methods and dataset(s) used to assess whether the device will improve reader performance as intended and to characterize the standalone device performance. Performance testing includes one or more standalone tests, side-by-side comparisons, or a reader study, as applicable.
(iii) Results from performance testing that demonstrate that the device improves reader performance in the intended use population when used in accordance with the instructions for use. The performance assessment must be based on appropriate diagnostic accuracy measures (
*e.g.,* receiver operator characteristic plot, sensitivity, specificity, predictive value, and diagnostic likelihood ratio). The test dataset must contain a sufficient number of cases from important cohorts (*e.g.,* subsets defined by clinically relevant confounders, effect modifiers, concomitant diseases, and subsets defined by image acquisition characteristics) such that the performance estimates and confidence intervals of the device for these individual subsets can be characterized for the intended use population and imaging equipment.(iv) Appropriate software documentation (
*e.g.,* device hazard analysis; software requirements specification document; software design specification document; traceability analysis; description of verification and validation activities including system level test protocol, pass/fail criteria, and results; and cybersecurity).(2) Labeling must include the following:
(i) A detailed description of the patient population for which the device is indicated for use.
(ii) A detailed description of the intended reading protocol.
(iii) A detailed description of the intended user and user training that addresses appropriate reading protocols for the device.
(iv) A detailed description of the device inputs and outputs.
(v) A detailed description of compatible imaging hardware and imaging protocols.
(vi) Discussion of warnings, precautions, and limitations must include situations in which the device may fail or may not operate at its expected performance level (
*e.g.,* poor image quality or for certain subpopulations), as applicable.(vii) Device operating instructions.
(viii) A detailed summary of the performance testing, including: test methods, dataset characteristics, results, and a summary of sub-analyses on case distributions stratified by relevant confounders, such as lesion and organ characteristics, disease stages, and imaging equipment.
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 15, 2026
Better Diagnostics AI Corp.
% Raj Zaveri
Founder & CEO
LucidMind Consulting
B202, Park Ave., Mg Rd.
Kandivli W.
Mumbai, Maharashtra 400067
INDIA
Re: K260851
Trade/Device Name: Better Diagnostics Dental Assist (BDDA) Version 1.0
Regulation Number: 21 CFR 892.2070
Regulation Name: Medical Image Analyzer
Regulatory Class: Class II
Product Code: MYN
Dated: March 16, 2026
Received: March 16, 2026
Dear Raj Zaveri:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260851 - Raj Zaveri
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260851 - Raj Zaveri
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang, Ph.D.
Assistant Director
DHT8B: Division of Radiologic Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260851
Device Name
Better Diagnostics Dental Assist (BDDA) Version 1.0
Indications for Use (Describe)
Better Diagnostics Dental Assist (BDDA) Version 1.0 is a computer aided detection ("CADe") software intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. The device is also intended to identify and mark regions in relation to suspected dental findings which include marginal discrepancy, calculus, periapical radiolucency (Only IOPA), crowns, filling and root canals. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. The intended patient population of the device is patients who are 18 years or older and do not have any remaining primary teeth.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
## 510(k) Summary
K260851
1. General Information
| 510(k) Owner | Better Diagnostics AI Corp. |
| --- | --- |
| Manufacturer Address | 29 Alcott Way Avon CT, 06001USA |
| Correspondence person | Raj Zaveri, Founder & CEOLucidMind Consulting |
| Correspondent address | B202, Park Ave., Mg Rd.Kandivli W., Mumbai,Maharashtra 400067 IND |
| Contact information | Email: rajzaveri10@gmail.comPhone: +91 8369438297 |
| Date prepared | March, 2026 |
### 2. Proposed device
Proprietary Name of Device: Better Diagnostics Dental Assist (BDDA) (Version 1.0)
Regulation Name: Medical Image Analyzer
Classification Panel: Radiology
Regulation Number: 21 CFR § 892.2070
Page 1 of 23
{5}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
Regulation Class: Class II
Product Code: MYN
### 3. Predicate Device
Proprietary Name of Device: Second Opinion
Premarket Notification: K210365
Regulation Name: Medical Image Analyzer
Classification Panel: Radiology
Regulation Number: 21 CFR § 892.2070
Regulation Class: Class II
Product Code: MYN
### 4. Predicate Device
Proprietary Name of Device: Overjet Dental Assist
Premarket Notification: K210187
Regulation Name: Medical image management and processing system.
Classification Panel: Radiology
Regulation Number: 21 CFR § 892.2050
Regulation Class: Class II
Product Code: LLZ
### 5. Device Description
Better Diagnostics Dental Assist (BDDA) Version 1.0 is a computer-aided detection (CADe) software designed for the automated measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. This software measures and displays the distance between the CEJ (Cemento-Enamel Junction) and the tip of the alveolar bone (in mm). The device is also intended for the automated detection of suspected dental findings which include marginal discrepancy, calculus, periapical radiolucency, crowns, filling, and root canal in bitewings and periapical dental radiographs. This software offers supplementary information to assist dentists in their diagnosis of these findings. It is important to note that BDDA
Page 2 of 23
{6}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
v1.0 is not meant to replace a comprehensive clinical evaluation by a dentist, which should consider other pertinent information from the image, the patient's medical history and clinical examination. This software is intended for use in patients who are 18 years or older.
BDDA v1.0 does not make treatment recommendations or provide a diagnosis. Dentists should review images annotated by BDDA v1.0 concurrently with original, unannotated images before making the final diagnosis on a case. BDDA v1.0 is an adjunct tool and does not replace the role of the dentist. The CAD generated output should not be used as the primary interpretation by the dentists. BDDA v1.0 is not designed to detect conditions and findings other than the ones mentioned above.
BDDA v1.0 comprises four main components:
- Presentation Layer: This component includes “AI Results Screen” a web-based interface (user interface) that allows users to view AI marked annotations. This is a custom code provided by Better Diagnostics AI Corp to Dental PMS (Practice Management Software) customers and imaging firms for utilization of BDDA v1.0 software. User Interface uses Angular.js and node.js technology to show images on the “AI Results Screen”. The system can process PNG, BMP, JPG format images. All images are converted into JPEG format for processing. Computer Vision Models return AI annotations and co-ordinates to the business layer. The business layer sends coordinates to the UI layer; bounding boxes and/or polygons are drawn on the image using custom code written in Angular.js and node.js. Dentists can view, accept or reject the annotations (bounding boxes and polygons) based on their evaluation.
Further, based on the coordinates shared by the business layer, lines are drawn (in mm) on the image that denotes the distance between the CEJ (Cemento-Enamel Junction) and the tip of the alveolar bone. This is also based on custom code written in Angular.js and node.js.
- Application Programming Interface (API): APIs are a set of definitions and protocols for building and integrating application software. It's sometimes referred to as a contract between an information provider and an information user. BDDA v1.0 APIs connect the Dental PMS with the business layer. API takes images input from Dental PMS and passes it to the business layer. It also receives annotations and co-ordinates from the business layer and passes it back to the presentation layer hosted by Dental PMS.
- Business Layer: Receives request from the API Gateway and passes it to the computer vision models. It also receives the bounding boxes and
Page 3 of 23
{7}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
polygon coordinates and retrieves images from the cloud storage. It sends annotations and co-ordinates to the “AI Results screen”.
- Computer Vision Models (CV Models): These models are hosted on a cloud computing platform and are responsible for image processing. For calculus, the model provides a binary indication to determine the presence or absence of calculus. If calculus and marginal discrepancy is detected, the software will output the coordinates of the bounding boxes. For Periapical Radiolucency and restorations, the models will output the coordinates of the polygon. For Bone level, the model measures and displays the distance between the CEJ (Cemento-Enamel Junction) and RBL (Remaining alveolar Bone Levels) (in mm).
AI models have three parts:
○ Pre-processing module: Standardization of image to specific height and width to maintain consistency for AI models. Pre-processing also includes classifications AI model to determine the type of image i.e IOPA, Bitewings.
- Core Module: Object detection, Segmentation and Key point detection AI models provide findings, annotations and co-ordinates to draw bounding boxes, polygons and lines (in mm).
- Post processing: includes cleanup process to remove outliers/incorrect annotations from the images.
## 6. Intended Use/Indications for Use
Better Diagnostics Dental Assist (BDDA) Version 1.0 is a computer aided detection ("CADe") software intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. The device is also intended to identify and mark regions in relation to suspected dental findings which include marginal discrepancy, calculus, periapical radiolucency (Only IOPA), crowns, filling and root canals. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. The intended patient population of the device is patients who are 18 years or older and do not have any remaining primary teeth.
Page 4 of 23
{8}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
7. Substantial Equivalence
| | Proposed Device | Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| 510(k) Number | K260851 | K210365 | K210187 |
| Applicant | Better Diagnostics AI Corp | Pearl, Inc. | Overjet, Inc. |
| Device Name | Better Diagnostics Dental Assist | Second Opinion | Overjet Dental Assist |
| Classification Regulation | 21 CFR 892.2070 | 21 CFR 892.2070 | 21 CFR 892.2050 |
| Product Code | MYN | MYN | LLZ |
| Indications for use | Better Diagnostics Dental Assist (BDDA) Version 1.0 is a computer aided detection ("CADe") software intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. The device is also intended to identify and mark regions | Second Opinion® is a computer aided detection ("CADe") software to identify and mark regions in relation to suspected dental findings which include Caries, Discrepancy at the margin of an existing restoration, Calculus, Periapical radiolucency, Crown (metal, including zirconia & non-metal), Filling (metal & non-metal), Root canal, Bridge, and Implants. It is designed to aid dental | Overjet Dental assist is a radiological semi-automated image processing software device intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The |
Page 5 of 23
{9}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| | Proposed Device | Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| | in relation to suspected dental findings which include marginal discrepancy, calculus, periapical radiolucency (Only IOPA), crowns, filling and root canals. It should not be used in-lieu of full patient evaluation or solely relied upon to make or confirm a diagnosis. The system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. The intended patient population of the device is patients who are 18 years or older and do not have any remaining primary teeth. | health professionals to review bitewing and periapical radiographs of permanent teeth in patients 12 years of age or older as a second reader. | system is to be used by trained professionals including, but not limited to, dentists and dental hygienists. The intended patient population of the device is patients living in the US, who are 22 years or older and that do not have any remaining primary teeth. Overjet has not evaluated the performance of the device on primary dentition. |
| Image Modality | Radiograph | Radiograph | Radiograph |
| Study Type | Bitewing and periapical Images | Bitewing and periapical Images | Bitewing and periapical Images |
| Clinical Output | Bounding boxes for calculus and | Bounding boxes /Fixed | Line |
Page 6 of 23
{10}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| | Proposed Device | Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| | marginal discrepanciesPolygon for Periapical Radiolucency and restorationsLines (in mm) for distance between CEJ and RBL | | |
| Patient Population | Patients requiring dental services, all sexes, 18 years of age or older with permanent teeth | Patients requiring dental services, all sexes, 12 years of age or older with permanent teeth | Patients requiring dental services, all sexes, 22 years of age or older with permanent teeth |
| Hardware requirements | WINDOWS 11 or higher | WINDOWS 7 or higher | Any |
| Intended User | Dental clinicians | Dental clinicians | Dental clinicians |
| Image Format | Accepts image formats JPG, and PNG, BMP | Accepts image formats from RVG, DICOM, JPEG, TIFF, and PNG and converts to JPEG. | Accepts image formats from JPG, PNG, JFIF, EOP, ETP, JIF |
| Processing Architecture | 1. APIs - Dental PMS/Imaging software send x-ray image to BDDA2. Computer vision models process the x-ray image and | Utilizes computer vision neural network algorithms, developed from open-source models using supervised machine learning techniques. | - |
Page 7 of 23
{11}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| | Proposed Device | Predicate Device | Predicate Device |
| --- | --- | --- | --- |
| | send results presentation layer 3. Presentation layers is plots rectangle, polygon and lines shape annotations on the x-ray image and it is displayed to dentist using “AI Results screen” | | |
| Type of device | CADe | CADe | CADe |
Better Diagnostics Dental Assist (BDDA) Software (Version 1.0) and Second Opinion are both class II devices with the same product classification code - MYN. Both devices are computer aided detection ("CADe") software intended to identify and mark regions in relation to similar suspected dental findings on bitewing and periapical radiographs.
Further, Better Diagnostics Dental Assist (BDDA) Software (Version 1.0) and Overjet Dental Assist are both class II devices and are CADe software intended to aid dental professionals in the measurements of mesial and distal bone levels associated with each tooth from bitewing and periapical radiographs.
Better Diagnostics Dental Assist (BDDA) Software (Version 1.0) has minor differences compared to the predicate devices in terms of the following:
- BDDA v1.0 is intended to identify and mark regions in relation to suspected dental findings which include marginal discrepancy, calculus, periapical radiolucency, crowns, filling, and root canal whereas Second opinion also includes caries in addition to the aforementioned findings. This is an additional feature offered by the predicate device and has no impact whatsoever on the outcome of BDDA v1.0.
- BDDA v1.0 output includes bounding boxes for calculus and marginal discrepancies, polygon for Periapical Radiolucency and restorations; and lines in mm for distance between the CEJ (Cemento-Enamel Junction) and the tip of the alveolar bone whereas Second Opinion and Overjet Dental Assist output includes bounding boxes and lines respectively. The additional functionality of
Page 8 of 23
{12}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
the BDDA v1.0 has no impact whatsoever on the outcome of BDDA v1.0. Further, the software verification and validation performed as per IEC 62304 validates the performance of BDDA v1.0.
- BDDA v1.0 is intended for patients aged 18 years or above, whereas Second Opinion identifies and marks similar suspected dental findings on radiographs acquired from patients aged 12 years or above and Overjet Dental Assist is intended for patients aged 22 years and above. This additional functionality (12 to 18 years) on the Second Opinion's part has no impact whatsoever on the outcome of BDDA v1.0. The results of the standalone and MRMC study discussed in the subsequent section along with the software verification and validation performed as per IEC 62304 validate the performance of BDDA v1.0.
- BDDA v1.0 supports BMP, JPG, and PNG, Second opinion supports RVG, DICOM, JPEG, TIFF, and PNG and converts to JPEG whereas Overjet Dental Assist supports JPG, PNG, JFIF, EOP, ETP, JIF. The additional formats supported by Second opinion and Overjet Dental Assist have no impact whatsoever on the outcome of BDDA v1.0. The software verification and validation performed as per IEC 62304 validates this claim.
# 8. Performance testing
# a. Summary of Non-Clinical Performance data
Safety and performance of BDDA v1.0 has been evaluated and verified in accordance with software specifications and the following applicable performance standards through software verification and validation, identification and mitigation of device-related hazards via cybersecurity and risk management, labeling validation, human factors testing, standalone and clinical performance testing :
- IEC 62304 Edition 1.1 2015-06 Medical device software – Software life cycle processes
- IEC 62366-1:2015: Medical devices-Part 1: Application of usability engineering to medical devices.
- ISO 14971 Third Edition 2019-12 Medical Devices - Application of risk management to medical devices.
- ISO 15223-1:2021: Medical devices Symbols to be used with information to be supplied by the manufacturer.
Additionally, the software validation activities were performed in accordance with the FDA Guidance documents, “Guidance for the Content of Premarket
Page 9 of 23
{13}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
Submissions for Software Contained in Medical Devices” and “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”.
Performance data from standalone and clinical performance assessments are summarized below.
### b. Summary of Clinical Performance data
#### i. Standalone Testing
A comprehensive standalone performance assessment was conducted to evaluate the diagnostic accuracy of the BDDA Software (version 1.0) across five intended dental indications: Existing Restoration, Margin Discrepancy, Calculus, Periapical Radiolucency, and Bone Level. All evaluations were performed independently of dentist interaction and were assessed against expert-established ground truth defined by consensus agreement from at least two of three licensed dentists with more than ten years of clinical experience.
#### Primary Performance
For Existing Restoration, Margin Discrepancy, Calculus, and Periapical Radiolucency, primary performance was evaluated at the dental surface level using sensitivity and specificity as predefined endpoints. Analyses were conducted separately for BW and IOPA radiographs, with Periapical Radiolucency assessed exclusively on IOPA images in accordance with clinical diagnostic relevance. Across all four radiographic anomalies and applicable imaging modalities, the BDDA Software consistently met all predefined performance goals. All lower bounds of the adjusted 95% confidence intervals for both sensitivity and specificity exceeded the respective acceptance thresholds (0.75 for sensitivity and 0.95 for specificity).
The evaluation was supported by a large and representative dataset. For Existing Restoration, a total of 7,174 dental surfaces from 895 BW images and 4,573 dental surfaces from 897 IOPA images were analyzed. For Margin Discrepancy, analysis included 9,773 dental surfaces from 700 BW images and 4,716 dental surfaces from 699 IOPA images. For Calculus, 8,218 dental surfaces from 715 BW images and 4,558 dental surfaces from 737 IOPA images were evaluated. For Periapical Radiolucency, which was assessed on IOPA images only, a total of 2,935 dental surfaces from 821 IOPA images were included.
- For Existing Restoration, the software achieved exceptionally high sensitivity, specificity, PPV, and NPV on both BW and IOPA images. On BW images, sensitivity was 0.989 (95% CI: 0.984–0.994), specificity was 0.990 (95% CI: 0.987–0.993), PPV was 0.979 (95% CI: 0.972–0.985), and NPV was 0.995 (95% CI: 0.992–0.997). On IOPA images, sensitivity reached 0.988 (95% CI: 0.979–0.994), specificity was 0.989 (95% CI: 0.985–0.993), PPV
Page 10 of 23
{14}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
was 0.970 (95% CI: 0.958–0.980), and NPV was 0.996 (95% CI: 0.992–0.998). In all cases, the lower bounds of the confidence intervals for sensitivity and specificity were well above the predefined acceptance criteria of 0.75 and 0.95, respectively.
- For Margin Discrepancy, BDDA demonstrated high sensitivity, near-perfect specificity, PPV, and NPV on both BW and IOPA images. On BW images, sensitivity was 0.917 (95% CI: 0.891–0.940), specificity was 0.997 (95% CI: 0.996–0.998), PPV was 0.957 (95% CI: 0.936–0.974), and NPV was 0.994 (95% CI: 0.992–0.996). On IOPA images, sensitivity was 0.908 (95% CI: 0.877–0.935), specificity was 0.998 (95% CI: 0.996–0.999), PPV was 0.981 (95% CI: 0.962–0.996), and NPV was 0.989 (95% CI: 0.985–0.992). All confidence interval lower bounds for sensitivity and specificity exceeded predefined thresholds.
- For Calculus, robust performance was observed across both imaging modalities. On BW images, sensitivity was 0.916 (95% CI: 0.893–0.936), specificity was 0.980 (95% CI: 0.976–0.983), PPV was 0.835 (95% CI: 0.810–0.860), and NPV was 0.991 (95% CI: 0.988–0.993). On IOPA images, sensitivity was 0.884 (95% CI: 0.857–0.908), specificity was 0.974 (95% CI: 0.969–0.979), PPV was 0.855 (95% CI: 0.826–0.882), and NPV was 0.980 (95% CI: 0.974–0.985). All confidence interval lower bounds exceeded predefined sensitivity and specificity acceptance criteria, with consistently high specificity across modalities.
- For Periapical Radiolucency, which was evaluated exclusively on IOPA images, BDDA met all required performance goals. Sensitivity was 0.809 (95% CI: 0.769–0.847), specificity was 0.973 (95% CI: 0.965–0.981), PPV was 0.847 (95% CI: 0.808–0.882), and NPV was 0.966 (95% CI: 0.957–0.974). The lower bounds of the confidence intervals for sensitivity and specificity exceeded the predefined acceptance thresholds.
For the Bone Level indication, primary performance metrics included sensitivity for detection of pathologic bone-level reduction, specificity for identification of normal bone levels, and quantitative agreement assessed by MAE. Performance was evaluated against predefined acceptance criteria of sensitivity \( \geq 0.82 \) , specificity \( \geq 0.81 \) , and MAE \( \leq 1.5 \) mm.
- Performance for the Bone Level indication was evaluated using both diagnostic classification and quantitative measurement accuracy. Across a total of 1,318 images (681 BW and 637 IOPA) and 12,031 interproximal line measurements, BDDA demonstrated high diagnostic accuracy for identifying abnormal bone levels, with sensitivity and specificity exceeding predefined acceptance thresholds across both imaging modalities. The lower bounds of
Page 11 of 23
{15}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
the 95% confidence intervals surpassed the required criteria for all evaluated subgroups.
- Quantitative accuracy was assessed using mean absolute error (MAE) relative to expert-established reference measurements. For BW images, the overall MAE was 0.203 mm (95% CI: 0.197–0.208), and for IOPA images, the overall MAE was 0.235 mm (95% CI: 0.226–0.243), both well below the predefined acceptance threshold of 1.5 mm. These results indicate precise and reliable measurement of interproximal alveolar bone levels.
Table 1: Standalone Study - Overall Performance Summary - Primary Endpoints (Sensitivity and Specificity)
| Radiographic Anomaly | Primary Endpoints | Act. Surface Perf. | Act. Surface Perf. (Adj. CI) | PDT | Meet PDT |
| --- | --- | --- | --- | --- | --- |
| (1) Restoration | BW Sensitivity | 0.989 | [0.984, 0.994] | 0.75 | Yes |
| | BW Specificity | 0.990 | [0.987, 0.993] | 0.95 | Yes |
| | IOPA Sensitivity | 0.988 | [0.979, 0.994] | 0.75 | Yes |
| | IOPA Specificity | 0.989 | [0.985, 0.993] | 0.95 | Yes |
| (2) Margin Discrepancy | BW Sensitivity | 0.917 | [0.891, 0.940] | 0.75 | Yes |
| | BW Specificity | 0.997 | [0.996, 0.998] | 0.95 | Yes |
| | IOPA Sensitivity | 0.908 | [0.877, 0.935] | 0.75 | Yes |
| | IOPA Specificity | 0.998 | [0.996, 0.999] | 0.95 | Yes |
| (3) Calculus | BW Sensitivity | 0.916 | [0.893, 0.936] | 0.75 | Yes |
Page 12 of 23
{16}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Primary Endpoints | Act. Surface Perf. | Act. Surface Perf. (Adj. CI) | PDT | Meet PDT |
| --- | --- | --- | --- | --- | --- |
| | BW Specificity | 0.980 | [0.976, 0.983] | 0.95 | Yes |
| | IOPA Sensitivity | 0.884 | [0.857, 0.908] | 0.75 | Yes |
| | IOPA Specificity | 0.974 | [0.969, 0.979] | 0.95 | Yes |
| (4) Periapical Radiolucency | IOPA Sensitivity | 0.809 | [0.769, 0.847] | 0.75 | Yes |
| | IOPA Specificity | 0.973 | [0.965, 0.981] | 0.95 | Yes |
| (5) Bone Level | BW Sensitivity | 0.866 | [0.858, 0.875] | 0.82 | Yes |
| | BW Specificity | 0.913 | [0.900, 0.925] | 0.81 | Yes |
| | BW MAE | 0.203 | [0.197, 0.208] | 1.5 mm | Yes |
| | IOPA Sensitivity | 0.889 | [0.876, 0.900] | 0.82 | Yes |
| | IOPA Specificity | 0.915 | [0.897, 0.931] | 0.81 | Yes |
| | IOPA MAE | 0.235 | [0.226, 0.243] | 1.5 mm | Yes |
Act. Surface Perf. = Actual Surface Performance; Act. Surface Perf. (Adj. CI) = Actual Surface Performance (Adjusted 95% Confidence Interval); PDT = Predefined Threshold; Se = Sensitivity; Sp = Specificity
Page 13 of 23
{17}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
Table 2: Standalone Study - Overall Performance Summary - Primary Endpoints (PPV and NPV)
| Radiographic Anomaly | Primary Endpoints | Act. Surface Perf. | Act. Surface Perf. (Adj. CI) |
| --- | --- | --- | --- |
| (1) Restoration | BW Surface Level PPV | 0.979 | [0.972, 0.985] |
| | BW Surface Level NPV | 0.995 | [0.992, 0.997] |
| | IOPA Surface Level PPV | 0.970 | [0.958, 0.980] |
| | IOPA Surface Level NPV | 0.996 | [0.992, 0.998] |
| (2) Margin Discrepancy | BW Surface Level PPV | 0.957 | [0.936, 0.974] |
| | BW Surface Level NPV | 0.994 | [0.992, 0.996] |
| | IOPA Surface Level PPV | 0.981 | [0.962, 0.996] |
| | IOPA Surface Level NPV | 0.989 | [0.985, 0.992] |
| (3) Calculus | BW Surface Level PPV | 0.835 | [0.810, 0.860] |
| | BW Surface Level NPV | 0.991 | [0.988, 0.993] |
| | IOPA Surface Level PPV | 0.855 | [0.826, 0.882] |
| | IOPA Surface Level NPV | 0.980 | [0.974, 0.985] |
| (4) Periapical Radiolucency | IOPA Surface Level PPV | 0.847 | [0.808, 0.882] |
| | IOPA Surface Level NPV | 0.966 | [0.957, 0.974] |
Page 14 of 23
{18}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Primary Endpoints | Act. Surface Perf. | Act. Surface Perf. (Adj. CI) |
| --- | --- | --- | --- |
| (5) Bone Level | BW PPV | 0.966 | [0.961, 0.970] |
| | BW NPV | 0.708 | [0.690, 0.725] |
| | IOPA PPV | 0.964 | [0.956, 0.971] |
| | IOPA NPV | 0.763 | [0.739, 0.785] |
Act. Surface Perf. = Actual Surface Performance; Act. Surface Perf. (Adj. CI) = Actual Surface Performance (Adjusted 95% Confidence Interval); PPV = Positive Predictive Value; NPV = Negative Predictive Value
### Secondary Performance
In addition to the primary surface-level evaluation, a secondary assessment was conducted to evaluate the diagnostic performance of the BDDA Software (version 1.0) at the image level for four radiographic anomalies: Existing Restoration, Margin Discrepancy, Calculus, and Periapical Radiolucency. Image-level analysis was performed to determine whether the software could reliably identify radiographs containing relevant radiographic anomalies, complementing the surface-level assessment.
Image-level sensitivity and specificity were evaluated separately for bitewing (BW) and intraoral periapical (IOPA) radiographs, consistent with clinical practice. Performance was assessed under both Conservative and Optimistic classification definitions, representing strict and permissive criteria for true positive classification, respectively.
Across all four evaluated radiographic anomalies, the BDDA Software met or exceeded all predefined image-level performance criteria. For all imaging modalities and evaluative definitions, the lower bounds of the adjusted 95% confidence intervals for both sensitivity and specificity exceeded the predefined acceptance threshold of 0.75. Overall, the secondary image-level results were consistent with the strong performance observed in the primary surface-level analysis. The BDDA Software demonstrated reliable and stable diagnostic performance across radiographic anomalies, imaging modalities, and evaluative frameworks.
Table 3: Standalone Study - Overall Performance Summary - Secondary Endpoints (Sensitivity and Specificity)
Page 15 of 23
{19}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Secondary Endpoints | Act. Surface Perf. | Act. Surface Perf. (CI) | PDT | Meet PDT |
| --- | --- | --- | --- | --- | --- |
| (1) Restoration | BW Image Level Se (Conservative) | 0.961 | [0.948, 0.975] | 0.75 | Yes |
| | BW Image Level Se (Optimistic) | 0.988 | [0.976, 0.995] | 0.75 | Yes |
| | BW Image Level Sp | 1.000 | [0.988, 1.000] | 0.75 | Yes |
| | IOPA Image Level Se (Conservative) | 0.973 | [0.955, 0.985] | 0.75 | Yes |
| | IOPA Image Level Se (Optimistic) | 0.984 | [0.969, 0.992] | 0.75 | Yes |
| | IOPA Image Level Sp | 1.000 | [0.990, 1.000] | 0.75 | Yes |
| (2) Margin Discrepancy | BW Image Level Se (Conservative) | 0.848 | [0.806, 0.884] | 0.75 | Yes |
| | BW Image Level Se (Optimistic) | 0.966 | [0.942, 0.982] | 0.75 | Yes |
| | BW Image Level Sp | 0.994 | [0.979, 0.999] | 0.75 | Yes |
| | IOPA Image Level Se (Conservative) | 0.872 | [0.833, 0.906] | 0.75 | Yes |
| | IOPA Image Level Se (Optimistic) | 0.954 | [0.926, 0.973] | 0.75 | Yes |
| | IOPA Image Level Sp | 0.997 | [0.984, 0.999] | 0.75 | Yes |
| (3) Calculus | BW Image Level Se (Conservative) | 0.829 | [0.783, 0.869] | 0.75 | Yes |
| | BW Image Level Se (Optimistic) | 0.949 | [0.919, 0.971] | 0.75 | Yes |
| | BW Image Level Sp | 0.870 | [0.833, 0.901] | 0.75 | Yes |
Page 16 of 23
{20}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Secondary Endpoints | Act. Surface Perf. | Act. Surface Perf. (CI) | PDT | Meet PDT |
| --- | --- | --- | --- | --- | --- |
| | IOPA Image Level Se (Conservative) | 0.806 | [0.759, 0.848] | 0.75 | Yes |
| | IOPA Image Level Se (Optimistic) | 0.948 | [0.917, 0.969] | 0.75 | Yes |
| | IOPA Image Level Sp | 0.879 | [0.843, 0.907] | 0.75 | Yes |
| (4) Periapical Radiolucency | IOPA Image Level Se (Conservative) | 0.815 | [0.771, 0.853] | 0.75 | Yes |
| | IOPA Image Level Se (Optimistic) | 0.858 | [0.818, 0.891] | 0.75 | Yes |
| | IOPA Image Level Sp | 0.929 | [0.901, 0.951] | 0.75 | Yes |
Act. Surface Perf. = Actual Surface Performance; Act. Surface Perf. (CI) = Actual Surface Performance (95% Confidence Interval); PDT = Predefined Threshold; Se = Sensitivity; Sp = Specificity
Table 4: Standalone Study - Overall Performance Summary - Secondary Endpoints (PPV and NPV)
| Radiographic Anomaly | Definition Type | Secondary Endpoints | Act. Surface Perf. | Act. Surface Perf. (CI) |
| --- | --- | --- | --- | --- |
| (1) Restoration | Conservative | BW Image Level PPV | 1.000 | [0.994, 1.000] |
| | | BW Image Level NPV | 0.929 | [0.895, 0.954] |
| | | IOPA Image Level PPV | 1.000 | [0.993, 1.000] |
| | | IOPA Image Level NPV | 0.959 | [0.933, 0.977] |
Page 17 of 23
{21}
[NO TEXT]
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Definition Type | Secondary Endpoints | Act. Surface Perf. | Act. Surface Perf. (CI) |
| --- | --- | --- | --- | --- |
| | Optimistic | BW Image Level PPV | 1.000 | [0.994, 1.000] |
| | | BW Image Level NPV | 0.977 | [0.953, 0.991] |
| | | IOPA Image Level PPV | 1.000 | [0.993, 1.000] |
| | | IOPA Image Level NPV | 0.975 | [0.953, 0.989] |
| (2) Margin Discrepancy | Conservative | BW Image Level PPV | 0.993 | [0.976, 0.999] |
| | | BW Image Level NPV | 0.864 | [0.826, 0.896] |
| | | IOPA Image Level PPV | 0.997 | [0.982, 0.999] |
| | | IOPA Image Level NPV | 0.889 | [0.854, 0.918] |
| | Optimistic | BW Image Level PPV | 0.994 | [0.979, 0.999] |
| | | BW Image Level NPV | 0.966 | [0.942, 0.982] |
| | | IOPA Image Level PPV | 0.997 | [0.983, 0.999] |
| | | IOPA Image Level NPV | 0.957 | [0.931, 0.975] |
| (3) Calculus | Conservative | BW Image Level PPV | 0.834 | [0.789, 0.874] |
| | | BW Image Level NPV | 0.865 | [0.828, 0.897] |
| | | IOPA Image Level PPV | 0.840 | [0.794, 0.879] |
| | | IOPA Image Level NPV | 0.852 | [0.814, 0.884] |
Page 18 of 23
{22}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Definition Type | Secondary Endpoints | Act. Surface Perf. | Act. Surface Perf. (CI) |
| --- | --- | --- | --- | --- |
| | Optimistic | BW Image Level PPV | 0.852 | [0.811, 0.888] |
| | | BW Image Level NPV | 0.956 | [0.929, 0.975] |
| | | IOPA Image Level PPV | 0.860 | [0.820, 0.895] |
| | | IOPA Image Level NPV | 0.955 | [0.929, 0.974] |
| (4) Periapical Radiolucency | Conservative | IOPA Image Level PPV | 0.904 | [0.868, 0.934] |
| | | IOPA Image Level NPV | 0.858 | [0.824, 0.888] |
| | Optimistic | IOPA Image Level PPV | 0.909 | [0.874, 0.937] |
| | | IOPA Image Level NPV | 0.887 | [0.855, 0.914] |
Act. Surface Perf. = Actual Surface Performance; Act. Surface Perf. (CI) = Actual Surface Performance (95% Confidence Interval); PPV = Positive Predictive Value; NPV = Negative Predictive Value
### Generalizability Performance
To evaluate the generalizability of the BDDA Software's performance across diverse patient demographics and imaging conditions, subgroup analyses were conducted based on age, sex, sensor type, and restoration type (for Existing Restoration only). Each subgroup was analyzed independently using the same performance metrics applied in the primary and secondary analyses.
Across all eligible subgroups, the BDDA Software demonstrated generally consistent diagnostic performance, with sensitivity and specificity estimates in most subgroups remaining aligned with the overall study results. While minor variations were observed in a few smaller subgroups, none showed patterns indicating a meaningful
Page 19 of 23
{23}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
loss of diagnostic accuracy or systematic performance degradation, almost all subgroups demonstrated a clinically meaningful improvement in performance.
The standalone performance assessment demonstrated that the BDDA Software (version 1.0) consistently meets all predefined clinical performance objectives across its five intended indications. High sensitivity and specificity were observed for all diagnostic indications, supported by narrow confidence intervals and consistent performance across imaging modalities and evaluative definitions. For the Bone Level indication, the combination of strong diagnostic accuracy and precise quantitative measurement further supports clinical reliability.
### ii. Clinical Evaluation-Reader Improvement
In addition to standalone performance evaluations, a multi-reader multi-case (MRMC) reader study was conducted to assess the clinical impact of BDDA Software (version 1.0) assistance on dentist diagnostic performance for three radiographic anomalies: Margin Discrepancy, Calculus, and Periapical Radiolucency (IOPA images only). The primary objective of the study was to evaluate whether BDDA assistance improves diagnostic accuracy compared with unaided interpretation.
The MRMC study employed a fully crossed, random-reader, random-case design. A total of 15 U.S.-licensed dentists independently interpreted the same set of radiographs under both unaided and BDDA-assisted conditions. Diagnostic accuracy was measured using the area under the AFROC AUC, and analyses were performed separately for BW and IOPA radiographs, consistent with clinical use. The MRMC datasets included 1,397 images for Margin Discrepancy (699 BW and 698 IOPA) with 13,795 tooth surfaces evaluated (9,193 BW and 4,602 IOPA); 1,451 images for Calculus (714 BW and 737 IOPA) with 12,803 tooth surfaces evaluated (8,225 BW and 4,578 IOPA); and 821 IOPA images for Periapical Radiolucency with 2,920 IOPA tooth surfaces evaluated.
### Primary Performance
Across all evaluated radiographic anomalies and image types, BDDA assistance resulted in a statistically significant improvement in AFROC AUC compared with unaided interpretation. For every endpoint, the mean AUC difference (Aided – Unaided) was positive, the corresponding 95% confidence interval excluded zero, and the hypothesis test demonstrated statistical significance.
- For Margin Discrepancy, BDDA assistance improved AUC from 0.808 to 0.905 on BW radiographs (difference: 0.097, 95% CI: [0.087, 0.120]) and from 0.831 to 0.917 on IOPA radiographs (difference: 0.086, 95% CI: [0.083, 0.112]).
Page 20 of 23
{24}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
- For Calculus, aided performance increased from 0.809 to 0.913 on BW radiographs (difference: 0.104, 95% CI: [0.078, 0.137]) and from 0.793 to 0.910 on IOPA radiographs (difference: 0.117, 95% CI: [0.098, 0.147]).
- For Periapical Radiolucency (IOPA only), BDDA assistance improved AUC from 0.847 to 0.939 (difference: 0.092, 95% CI: [0.075, 0.110]).
Overall, the findings establish that BDDA Software (version 1.0) provides a meaningful diagnostic benefit across all MRMC-evaluated radiographic anomalies and imaging modalities, thereby meeting the predefined primary performance objectives of the study.
Table 5: MRMC Study - Overall Performance Summary - Primary Endpoints AFROC AUC
| Radiographic Anomaly | Primary Endpoints | Act. Aided Perf. | Act. Unaided Perf. | Diff. [Aided-Unaided] | Diff. (95% CI) | Sig. |
| --- | --- | --- | --- | --- | --- | --- |
| (1) Margin Discrepancy | BW AFROC | 0.905 | 0.808 | 0.097 | [0.087, 0.120] | Yes |
| | IOPA AFROC | 0.917 | 0.831 | 0.086 | [0.083, 0.112] | Yes |
| (2) Calculus | BW AFROC | 0.913 | 0.809 | 0.104 | [0.078, 0.137] | Yes |
| | IOPA AFROC | 0.910 | 0.793 | 0.117 | [0.098, 0.147] | Yes |
| (3) Periapical Radiolucency | IOPA AFROC | 0.939 | 0.847 | 0.092 | [0.075, 0.110] | Yes |
Act. Aided Perf. = Actual Aided Performance; Act. Unaided Perf. = Actual Unaided Performance; Diff. = Difference; Diff. (95% CI) = Difference with 95% Confidence Interval; Sig. = Statistically Significant
### Key Secondary Performance
Beyond the primary AFROC AUC analysis, a key secondary assessment was conducted to evaluate surface-level diagnostic performance by comparing reader sensitivity and specificity with and without BDDA Software (version 1.0) assistance for each imaging modality. This analysis was performed for three radiographic
Page 21 of 23
{25}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
anomalies: Margin Discrepancy, Calculus, and Periapical Radiolucency (IOPA images only).
Across all evaluated radiographic anomalies and applicable imaging modalities, BDDA assistance resulted in statistically significant improvements in both sensitivity and specificity at the dental surface level. For all secondary endpoints, the aided–unaided differences were positive, and the corresponding 95% confidence intervals excluded zero, indicating statistically meaningful improvements in diagnostic accuracy. These findings confirm that BDDA Software improves reader diagnostic performance not only at the image level but also at the clinically actionable dental surface level, supporting its intended use as a computer-assisted decision-support tool.
Table 6: MRMC Study - Overall Performance Summary - Key Secondary Endpoints Sensitivity and Specificity at the Surface Level
| Radiographic Anomaly | Key Secondary Endpoints | Act. Aided Perf. | Act. Unaided Perf. | Diff. [Aided-Unaided] | Diff. (95% CI) |
| --- | --- | --- | --- | --- | --- |
| (1) Margin Discrepancy | BW Surface Se | 0.914 | 0.788 | 0.126 | [0.105, 0.147] |
| | BW Surface Sp | 0.988 | 0.980 | 0.008 | [0.005, 0.011] |
| | IOPA Surface Se | 0.908 | 0.813 | 0.095 | [0.074, 0.116] |
| | IOPA Surface Sp | 0.988 | 0.971 | 0.017 | [0.013, 0.021] |
| (2) Calculus | BW Surface Se | 0.907 | 0.716 | 0.191 | [0.137, 0.246] |
| | BW Surface Sp | 0.988 | 0.983 | 0.006 | [0.004, 0.008] |
| | IOPA Surface Se | 0.908 | 0.709 | 0.199 | [0.153, 0.246] |
| | IOPA Surface Sp | 0.980 | 0.966 | 0.013 | [0.009, 0.018] |
Page 22 of 23
{26}
BETTERDIAGNOSTICS.AI
Better Diagnostics AI Corp.
29 Alcott Way Avon CT, 06001
USA
| Radiographic Anomaly | Key Secondary Endpoints | Act. Aided Perf. | Act. Unaided Perf. | Diff. [Aided-Unaided] | Diff. (95% CI) |
| --- | --- | --- | --- | --- | --- |
| (3) Periapical Radiolucency | IOPA Surface Se | 0.909 | 0.747 | 0.162 | [0.129, 0.196] |
| | IOPA Surface Sp | 0.988 | 0.979 | 0.009 | [0.005, 0.012] |
Act. Aided Perf. = Actual Aided Performance; Act. Unaided Perf. = Actual Unaided Performance; Diff. = Difference; Diff. (95% CI) = Difference with 95% Confidence Interval; Se = Sensitivity; Sp = Specificity
### Generalizability Performance
Generalizability of BDDA Software (version 1.0) performance was evaluated through additional image-level assessments, exploratory ROC analyses, and predefined subgroup analyses within the MRMC study. Across these analyses, BDDA assistance demonstrated stable diagnostic performance that was consistent with the primary and key secondary MRMC findings. Improvements in image-level sensitivity and specificity, as well as ROC AUC at both the surface and image levels, supported the robustness of aided interpretation across multiple analytic perspectives.
Subgroup analyses were conducted across reader experience levels, patient age groups, sex, and a range of imaging sensor types. Across the majority of evaluated subgroups, BDDA assistance yielded diagnostic performance patterns comparable to the overall study results. These findings indicate that the diagnostic benefit of BDDA is maintained across diverse reader backgrounds, patient demographics, and imaging acquisition conditions.
## 9. Conclusion
Based on the information submitted in this premarket notification, and the indications for use, technological characteristics and performance testing, the Better Diagnostics Dental Assist (BDDA) Software (Version 1.0) raises no new questions of safety and effectiveness and is substantially equivalent to the predicate devices Second Opinion (K210365) and Overjet Dental Assist (K210187) in terms of safety, effectiveness, and performance.
Page 23 of 23
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.