K260817 · Virtual Incision · SAB · Aug 20, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260817
Device Name
MIRA Surgical System
Applicant
Virtual Incision
Product Code
SAB · General, Plastic Surgery
Decision Date
Aug 20, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4962
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MIRA Surgical System is a miniaturized robotic assisted surgery device intended to assist in the visualization of tissues and provide accurate and precise control of surgical instruments to grasp, retract, dissect, and suture while maintaining hemostasis with electrocautery during manipulation of tissues. The device is indicated for mobilization of the colon during minimally invasive colectomy in adults at least 5'0" (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg). The MIRA Surgical System is also indicated for minimally invasive benign (sub-) total hysterectomy with or without salpingectomy in adults with a uterine size not exceeding 12 cm longitudinally on imaging, at least 5'0" (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg), and a BMI less than 40kg/m2. It is for prescription use only and is to be used by trained physicians in any operating room environment in accordance with the representative surgical procedures listed in the MIRA Surgical System User Manual.
Device Story
MIRA Surgical System is a miniaturized robotic-assisted platform for soft tissue surgery. System components: Surgeon Control Console (main processor, display, hand/pedal inputs), Companion Cart (Interface Pod, ESU), reusable Surgical Minibot (two-armed, inserted via port), articulating Surgical Camera, Support Arm Assembly, and single-use sterile instruments (graspers, dissectors, needle drivers). Surgeon operates Minibot and Camera from console; real-time video feed guides tissue manipulation. Instruments perform grasping, retraction, dissection, suturing, and electrocautery-based hemostasis. Used in operating rooms by trained physicians. System provides precise instrument control, enabling minimally invasive procedures; benefits include reduced invasiveness compared to traditional surgery. Hardware modifications and new needle driver instruments added to predicate design; software updates resolve anomalies and improve functionality.
Clinical Evidence
Prospective, multi-center, non-randomized clinical study (n=35) for benign hysterectomy. Primary endpoint: procedural success. Results: 34/35 (97.1%) successful completion without conversion. Mean uterine weight 136.1g. No intraoperative or postoperative device/procedure-related SAEs. Adverse events (10 non-serious) were within expected levels for minimally invasive surgery. Bench, cadaver, and animal studies also performed.
Technological Characteristics
Miniaturized electromechanical robotic system. Components: Surgeon Console, Companion Cart, Minibot, Camera, Support Arm. Energy: Bipolar/monopolar via ESU. Materials: Biocompatible patient-contacting materials. Connectivity: Integrated system. Software: Embedded control system. Sterilization: Reusable components with sterilization trays; single-use sterile instruments.
Indications for Use
Indicated for mobilization of the colon during minimally invasive colectomy in adults >= 5'0" (1.52m) and >= 100 lbs (45.36kg). Also indicated for minimally invasive benign (sub-) total hysterectomy with/without salpingectomy in adults >= 5'0" (1.52m), >= 100 lbs (45.36kg), BMI < 40kg/m2, and uterine size <= 12 cm longitudinally.
Regulatory Classification
Identification
The MIRA Surgical System is a miniaturized robotic assisted surgery device indicated for the mobilization of the colon in adults at least 5'0" (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg) who are undergoing minimally invasive colectomy procedures. It is intended to assist in the visualization of tissues and provide accurate and precise control of surgical instruments to grasp, retract, and dissect while maintaining hemostasis with electrocautery during manipulation of tissues.
Special Controls
In combination with the general controls of the FD&C Act, the table mounted miniaturized electromechanical surgical system is subject to the following special controls:
(1) Data obtained from premarket clinical performance validation testing and postmarket surveillance acquired under anticipated conditions of use must demonstrate that the device performs as intended in the intended patient population, unless FDA determines based on the totality of the information provided for premarket review that data from postmarket surveillance is not required.
(i) Data provided from (1) must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in a variety of disease etiologies relevant to the device intended use. The test data set must include data acquired from a patient population that is representative of the intended patient population.
(ii) Objective performance measures (i.e., rate and number of conversions to open surgery, rate of device related adverse events and their severity, cause, and outcomes) must be reported with relevant descriptive or developmental performance measures.
(2) Animal performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must evaluate:
(i) All surgical tasks pertinent to the device’s intended use, demonstrating the device can accurately and precisely control attached surgical instruments without excess of adverse events, device and not-device related.
(ii) Acute and chronic histopathology and gross examination of the affected organs and surrounding tissue.
(3) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
(4) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(5) Human factors validation testing must be performed and must demonstrate that the device/user interfaces of the system support safe use in all use environments.
(6) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness of the device has not been evaluated for outcomes related to the treatment or prevention of cancer, including but not limited to risk reduction, overall survival, disease-free survival and local recurrence, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A detailed summary of the post-market surveillance data collected under paragraph (1) of this section and any necessary modifications to the labeling to accurately reflect outcomes based upon the post-market surveillance data collected under paragraph (1) of this section.
(7) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and precision;
(ii) System testing;
A. Instrument reliability;
B. Thermal effects on tissue;
C. User-device interface performance;
D. Workspace access testing; and
E. Performance testing with compatible devices.
(8) Software verification, validation, and hazard analysis must be performed.
(9) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(10) Performance data must demonstrate the sterility of all patient-contacting device components.
(11) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(12) Performance data must validate the reprocessing instructions for the reusable components of the device.
(13) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(14) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(15) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device-specific use training program, which did not require submission of a premarket notification during the reporting period.
{1}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 20, 2026
Virtual Incision
Florence Beck
VP Clinical and Regulatory Affairs
1501 Old Cheney Rd.
Lincoln, Nebraska 68512
Re: K260817
Trade/Device Name: MIRA Surgical System
Regulation Number: 21 CFR 878.4962
Regulation Name: Table Mounted Miniaturized Electromechanical Surgical System
Regulatory Class: Class II
Product Code: SAB
Dated: July 20, 2026
Received: July 20, 2026
Dear Florence Beck:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260817 - Florence Beck
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260817 - Florence Beck
Page 3
Sincerely,
MARK
Digitally signed by
MARK TRUMBORE -S
TRUMBORE -S
Date: 2026.08.20
18:32:26 -04'00'
Mark Trumbore, Ph.D.
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
# INDICATIONS FOR USE
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260817
Please provide the device trade name(s).
MIRA Surgical System
Please provide your Indications for Use below.
The MIRA Surgical System is a miniaturized robotic assisted surgery device intended to assist in the visualization of tissues and provide accurate and precise control of surgical instruments to grasp, retract, dissect, and suture while maintaining hemostasis with electrocautery during manipulation of tissues.
The device is indicated for mobilization of the colon during minimally invasive colectomy in adults at least 5'0" (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg).
The MIRA Surgical System is also indicated for minimally invasive benign (sub-) total hysterectomy with or without salpingectomy in adults with a uterine size not exceeding 12 cm longitudinally on imaging, at least 5'0" (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg), and a BMI less than 40kg/m2.
It is for prescription use only and is to be used by trained physicians in any operating room environment in accordance with the representative surgical procedures listed in the MIRA Surgical System User Manual.
Please select the types of uses (select one or both, as applicable).
Prescription Use (Part 21 CFR 801 Subpart D) ☑
Over-The-Counter Use (21 CFR 801 Subpart C) ☐
{5}
VIRTUALINCISION
# 510(k) Summary
## I. SUBMITTER INFORMATION
**Submitter:** Virtual Incision Corporation
1501 Old Chaney Rd
Lincoln, NE 68512**Contact:** Florence Beck
VP Clinical and Regulatory Affairs
Phone Number: 352-275-7267
Email: florence.beck@virtualincision.com**Date Summary Prepared:** August 20, 2026
## II. DEVICE INFORMATION
**Trade/Device Name:** MIRA Surgical System**Common Name:** Table mounted miniaturized electromechanical surgical system**Regulation Number:** 21 CFR 878.4962**Regulatory Class:** Class II**Product Code:** SAB**Submission Type:** Traditional 510(k)
## III. PEDICATE DEVICE INFORMATION
**Predicate Device:** MIRA Surgical System (DEN230025)
## IV. DEVICE DESCRIPTION
The MIRA Surgical System is a miniaturized robotic assisted surgery device consisting of an integrated package designed to enable soft tissue surgery using a minimally invasive approach. The MIRA Surgical System refers to the system and all of its components. The system consists of a Surgeon Control Console, Companion Cart, Surgical Minibot with Sterilization Tray, Surgical Camera with Sterilization Tray, Support Arm Assembly with Sterilization Tray, and single use MIRA Surgical Instruments and Accessories. These are described below:
Surgeon Control Console: The Surgeon Control Console contains a main processor that performs system-level control functions and system monitoring. The Console includes a main display showing the real-time video feed from the Surgical Camera, hand input devices, pedal inputs, and an interactive touchscreen. The Minibot and Surgical Camera are controlled by the surgeon seated at the Surgeon Control Console.
Companion Cart: The Companion Cart is a mechanical cart used to hold the Interface Pod, Electrical Surgical Unit (ESU) and serves as a staging shelf for the Minibot and Surgical Camera.
510(k) Summary
K260817
Page 1 of 7
{6}
VIRTUALINCISION
The Interface Pod is a box that serves as a single point connection for the Minibot, Surgical Camera, and ESU to the Surgeon Control Console. The ESU is bolted to the companion cart and has a unique connector that can only interface with the Minibot device.
Surgical Minibot with Sterilization Tray: The Surgical Minibot is a reusable device component with two arms that is inserted into the insufflated abdominal cavity through a disposable insertion port to perform surgery. Single-use sterile MIRA Surgical Instruments are installed in the Minibot arms. The Minibot is robotically controlled by the surgeon.
Surgical Camera with Sterilization Tray: The Camera is a reusable articulating endoscope used to visualize the surgical target. The Camera mechanically interfaces with the Minibot and can be robotically articulated by the surgeon.
Support Arm Assembly with Sterilization Tray: The Support Arm Assembly is a reusable device component used to hold the Minibot in place. The device is easily adjustable to many positions and orientations. The proximal end of the Support Arm mounts to the surgical table rail using a rail clamp. The distal end attaches to the Minibot Clamp to hold the Minibot.
Surgeon Chair: An off-the-shelf adjustable surgeon chair is provided with the MIRA Surgical System. The following single-use sterile disposables are used with the MIRA Surgical System:
MIRA Surgical Instruments: MIRA Instruments come in various configurations and are customized to connect into the Minibot arms to grasp, retract, dissect, and suture while maintaining hemostasis with electrocautery during manipulation of tissues.
Camera Nest and Lumen Plug Kit: The Lumen Plug is placed onto the Minibot to prevent lubricant and debris from entering the Minibot lumen during insertion. The Nest is placed on the Minibot to connect to the Camera.
## V. INTENTED/INDICATION FOR USE
The MIRA Surgical System is a miniaturized robotic assisted surgery device intended to assist in the visualization of tissues and provide accurate and precise control of surgical instruments to grasp, retract, dissect, and suture while maintaining hemostasis with electrocautery during manipulation of tissues.
The device is indicated for mobilization of the colon during minimally invasive colectomy in adults at least 5'0' (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg).
The MIRA Surgical System is also indicated for minimally invasive benign (sub-) total hysterectomy with or without salpingectomy in adults with a uterine size not exceeding 12 cm longitudinally on imaging, at least 5'0' (1.52 m) tall and having a weight of at least 100 lbs (45.36 kg), and a BMI less than 40kg/m2.
510(k) Summary
K260817
Page 2 of 7
{7}
VIRTUALINCISION
It is for prescription use only and is to be used by trained physicians in any operating room environment in accordance with the representative surgical procedures listed in the MIRA Surgical System User Manual.
# VI. COMPARISONS OF INTENTED USE/INDICATIONS FOR USE AND TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICE
The intended use of the MIRA Surgical System remains unchanged and continues to fall under the general product code SAB: table mounted, miniaturized electromechanical surgical system. Devices classified under this code are intended to enable a qualified user to perform robotic assisted, minimally invasive surgical procedures including the mobilization of tissue within the intra-abdominal cavity and along the colon.
The proposed indication expands upon the currently cleared use by adding minimally invasive benign hysterectomy as an additional specialty and anatomic location within the abdomen and to add suturing as a task provided by the surgical instruments. As part of this expansion, clinical performance testing evaluated the system in the umbrella surgical procedure benign hysterectomy. Representative surgical procedures under this umbrella have been added to the MIRA Surgical System User Manual.
Clinical Evidence generated under an IDE clinical trial supports that the inclusion of these gynecological procedures does not introduce new or increased procedural risks beyond those within the predicate intended use and remains comparable to those expected with minimally invasive surgical techniques. Like the existing indication, the MIRA Surgical System performs as intended in the intended patient population with an acceptable safety profile compared to hysterectomy procedures performed with traditional laparoscopic surgery.
# VII. TECHNOLOGICAL COMPARISON
Both the subject MIRA Surgical System and predicate MIRA Surgical System (DEN230025) have the same technological characteristics and intended use. All components and accessories included in the predicate are included in the subject device.
The subject device expands MIRA Surgical Instrument Family to include single use sterile Needle Drivers to allow for suturing. Like the predicate MIRA Surgical Instruments, the subject device is intended to assist in the visualization of tissues and provide accurate and precise control of surgical instruments. They are installed onto the Minibot and serve as the hands of the system, functioning in a similar fashion to manual laparoscopic instruments. Like the predicate, they have similar dimensional profile, are single use, disposable, and made of biocompatible materials. Packaging, sterilization, shelf-life, installation, use, and removal of the Needle Drivers are the same as the Bipolar Grasper except that there are no electrocautery contacts within the instrument nor electrocautery function. All performance testing passed and did not produce any unexpected issues of safety or effectiveness. The introduction of Needle Drivers does not pose any additional risk and does not affect safety or effectiveness or raise
510(k) Summary
K260817
Page 3 of 7
{8}
VIRTUALINCISION
different types of questions of safety and effectiveness as demonstrated by bench, animal, and clinical performance testing. The Needle Driver Instruments are substantially equivalent to the predicate.
This submission is also intended to bring the agency up to date on all incremental non-significant changes made to the predicate device and implemented through internal design control process since the initial De Novo authorization to the current subject device in this submission. There are no modifications to the intended use, fundamental scientific technology, or principles of operation. Completed testing demonstrated that the subject device continues to perform in accordance with its specifications. The subject device complies with the same standards and special controls as the predicate device. Below is a comparison summary of the technological characteristics.
**Design:** The subject device and the predicate device have the same design. Both are a miniaturized robotic assisted surgery device consisting of an integrated package designed to enable soft tissue surgery using a minimally invasive approach. Both consist of the same components and function. Some engineering changes to hardware of the predicate device components were replaced with similar parts on the subject device due to part obsolescence and/or to improve reliability and manufacturability of the device. A risk-based assessment of the hardware changes did not identify new risks or significantly modified existing risks. Minor software updates to the predicate device were implemented to resolve software anomalies and improve functionality but did not impact the intended use or the usability by the user. Design verification and/or validation activities did not produce any unexpected issues of safety or effectiveness.
**Materials:** Both the subject device and predicate device consist of biocompatible patient contacting materials. Some patient contacting materials on the surgical instruments and Minibot of the predicate device were replaced with similar biocompatible materials on the subject device. A risk assessment did not identify any new or increased biocompatibility concerns. Design verification and/or validation activities did not produce any unexpected issues of safety or effectiveness.
**Principle of Operation:** Both the subject device and predicate device have the same principles of operation listed in the device description.
**Energy Source:** Both the subject device and predicate device use bipolar and monopolar energy via single use surgical instruments from the same ESU located on the Companion Cart.
**Conclusion:** The predicate and proposed MIRA Surgical Systems have the same intended use and similar technological characteristics and principles of operation. In addition to nonsignificant design changes, a needle driver instrument has been added to aid in surgical suturing. The labeling has been changed to add suturing and the expanded indication statement and listed as new representative surgical procedures in the User Manual. Completed performance testing (Bench, Animal, and Clinical) and continued adherence to Special
510(k) Summary
K260817
Page 4 of 7
{9}
VIRTUALINCISION
Controls supports a determination of substantial equivalence to the predicate MIRA Surgical System (DEN230025).
### VIII. NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS
Non-clinical performance testing demonstrated that the subject device continues to meet the performance specifications and Special Controls as intended under anticipated conditions of use. This is supported by anatomical studies, human factors validation, Software testing, System-wide, and System-component verification and validation testing.
Cadaver and Synthetic Cadaver Studies- MIRA Surgical System was extensively evaluated in the cadaveric and synthetic models to support the proposed indication. This anatomic data demonstrates that the device met User Needs and performed as intended under anticipated conditions of use without incidents that would indicate the potential of device related adverse events.
Animal Performance Testing- Acute and Chronic GLP testing demonstrates that the device performed as intended under anticipated conditions of use. Like the predicate, all surgical tasks pertinent to the hysterectomy procedure were completed successfully. The subject device was able to accurately and precisely control the MIRA surgical instruments without excessive adverse events. This was confirmed by histopathology and gross examination of the affected organs and surrounding tissues.
Human Factors- Human Factors validation testing for the subject device was completed with the expanded surgeon user group (minimally invasive gynecological surgeons). Like the predicate, the device/user interfaces of the system support safe use in all use environments. Clinical Performance: Clinical performance testing for the expanded indication of the MIRA Surgical System was completed in the US under Investigational Device Exemptions (IDE).
A prospective multi-center, multi-operator, non-randomized, consecutively enrolled clinical evaluation study was conducted in 35 patients presenting with an indication for benign hysterectomy. The mean age was 46.1 years, with a range of 25 to 74 years. The mean body mass index (BMI) was 29.2 kg/m², ranging from 21.5 to 41.5 kg/m². The clinical profile of enrolled subjects were representative of the intended patient population. Patients were followed for at least 42 days post-surgical procedure.
# Outcomes:
The results of the study demonstrated that the MIRA Surgical System performed as intended in a clinical use environment. In 34 of 35 (97.1%) clinical cases, the procedures were successfully
510(k) Summary
K260817
Page 5 of 7
{10}
VIRTUALINCISION
completed with no conversions to alternative surgical techniques and with full vaginal cuff healing as expected at the last follow-up visit. The specimen weight ranged from 31.2 g to 384 g, with a mean uterine weight of 136.1 g and a median of 117.0 g. The MIRA Surgical System was able to adequately and primarily dissect, expose the uterus and adnexal tissues, perform the colpotomy, and close the vaginal cuff with the MIRA surgical instruments. Hemostasis could be maintained with the MIRA monopolar scissors and bipolar graspers alone in all 34 subjects. There were no malignancies reported in the pathology reports.
Safety Assessments:
The MIRA Surgical System demonstrated an acceptable safety profile.
- There were no intraoperative device or procedure related serious adverse events (SAE) reported.
- There were no postoperative device or procedure related SAEs reported.
- There was one SAE reported deemed unrelated to the device or unrelated to the procedure.
- Incidence of clinically relevant adverse events were within expected levels when compared to the published literature.
There were no deaths or unanticipated adverse events. One reported serious adverse event unrelated to both the device and the procedure was anemia secondary to pre-operative menorrhea, only diagnosed following surgery. Nine subjects (25.7%) experienced ten non-serious, clinically relevant adverse events (two intraoperative Grade I related to the device, five postoperative Grade I related to the procedure, and three postoperative Grade II related to the procedure; per Clavien-Dindo grading). All events are resolved without sequelae. These adverse events were within expected levels.
# IX. CONCLUSIONS
In 35 U.S. subjects undergoing total hysterectomy for benign indications, the MIRA Surgical System met its primary effectiveness endpoint and demonstrated a favorable safety profile. The device achieved high procedural completion rates, reliable hemostasis with minimal blood loss, short length of stay, complete cuff healing, and complication rates consistent with standard minimally invasive hysterectomy. These findings support the safety and effectiveness of the system in the intended patient population.
The clinical evidence generated demonstrates that the MIRA Surgical System performs as intended in the intended patient population with an acceptable safety profile compared to procedures performed with traditional laparoscopic surgery. The expansion of this indication does not introduce new or increased procedural risks beyond those within the predicate
510(k) Summary
K260817
Page 6 of 7
{11}
VIRTUALINCISION
intended use and remains comparable to those expected with minimally invasive surgical techniques.
Nonclinical testing demonstrated that the subject device performs as intended in the intended conditions of use in accordance with performance specifications and confirm the implemented component modifications and addition of the needle drivers do not present different questions of safety or effectiveness than the predicate device The subject device complies with the same standards and special controls as the predicate device. Thus, the device is substantially equivalent to the predicate.
510(k) Summary
K260817
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.