When used with Anteralign™ LS Spinal System, the Anteralign™ LS coverplate and cage is only intended to be used according to the indications of Medtronic's Anteralign™ LS Spinal System. Anteralign™ LS Spinal System Indications for Use The Anteralign™ LS Spinal System is intended to be used in spinal fusion procedures on skeletally mature patients with symptomatic Degenerative Disc Disease (DDD, defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies), degenerative spondylolisthesis, and/or spinal stenosis, at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion. These patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. These patients should have had six months of nonoperative treatment prior to treatment with this device. Additionally, the Anteralign™ LS Spinal System and optional Anteralign™ LS Coverplate can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. The Anteralign™ LS interbody cage may be used as a stand-alone device or in conjunction with supplemental fixation. The Anteralign LS™ interbody fusion Device e may be inserted via minimally invasive or open anterior or oblique approach at one or two contiguous levels from L2 to S1. These approaches include anterior and oblique. When used as a stand-alone device, the Anteralign™ LS cage must be used with 3 screws with devices that have standard lordosis (≤16°). If the physician chooses to use less than 3 screws or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation (e.g. posterior fixation) must be used. Additionally, cages with lordosis angles 16° or greater are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation).
Device Story
Anteralign LS coverplate system; includes coverplate, locking screw, and integrated locking mechanism; used exclusively with Medtronic Anteralign LS Spinal System. Implanted via minimally invasive or open anterior/oblique approach by surgeons in clinical settings. Provides supplemental stability to interbody fusion cages; prevents screw back-out. Benefits patients by maintaining construct integrity during spinal fusion procedures. Not a required element for the base spinal system; optional component.
Clinical Evidence
No clinical testing was performed. Evidence is based on non-clinical bench testing, including clinical simulation validation, MR testing, and mechanical performance testing (ASTM F2077, ASTM F2267, and DRAFT ASTM Z8423Z).
Technological Characteristics
Titanium alloy (ASTM F136). Components: coverplate, locking screw, integrated locking mechanism. Dimensions/form factor: designed for lumbar interbody fusion. Connectivity: N/A. Sterilization: Not specified. Software: N/A.
Indications for Use
Indicated for skeletally mature patients with symptomatic Degenerative Disc Disease (DDD), degenerative spondylolisthesis, and/or spinal stenosis at one or two contiguous levels from L2 to S1 requiring interbody fusion. Includes patients with up to Grade 1 Spondylolisthesis or retrolisthesis. Requires six months of prior nonoperative treatment. Also indicated as an adjunct to fusion for multilevel degenerative scoliosis and sagittal deformity.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology (K214010)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
June 10, 2026
Tyber Medical, LLC
Rajan Kaur
Senior Regulatory Affairs Specialist
83 South Commerce Way, Suite 310
Bethlehem, Pennsylvania 18017
Re: K260810
Trade/Device Name: Anteralign LS coverplate
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: OVD
Dated: March 12, 2026
Received: March 12, 2026
Dear Rajan Kaur:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260810 - Rajan Kaur
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260810 - Rajan Kaur
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**BRENT SHOWALTER -S**
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Anteralign LS coverplate | | |
| Please provide your Indications for Use below. | | ? |
| When used with Anteralign™ LS Spinal System, the Anteralign™ LS coverplate and cage is only intended to be used according to the indications of Medtronic's Anteralign™ LS Spinal System. | | |
| Anteralign™ LS Spinal System Indications for Use The Anteralign™ LS Spinal System is intended to be used in spinal fusion procedures on skeletally mature patients with symptomatic Degenerative Disc Disease (DDD, defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies), degenerative spondylolisthesis, and/or spinal stenosis, at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion. These patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. These patients should have had six months of nonoperative treatment prior to treatment with this device. Additionally, the Anteralign™ LS Spinal System and optional Anteralign™ LS Coverplate can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity. | | |
| The Anteralign™ LS interbody cage may be used as a stand-alone device or in conjunction with supplemental fixation. The Anteralign LS™ interbody fusion Device e may be inserted via minimally invasive or open anterior or oblique approach at one or two contiguous levels from L2 to S1. These approaches include anterior and oblique. When used as a stand-alone device, the Anteralign™ LS cage must be used with 3 screws with devices that have standard lordosis (≤16°). If the physician chooses to use less than 3 screws or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation (e.g. posterior fixation) must be used. Additionally, cages with lordosis angles 16° or greater are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation). | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
510(k) #:
# 510(k) Summary
Prepared on: 2026-06-10
## Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Tyber Medical LLC |
| --- | --- |
| Applicant Address | 83 South Commerce Way, Suite 310 Bethlehem PA 18017 United States |
| Applicant Contact Telephone | 856-558-0632 |
| Applicant Contact | Mrs. Rajan Kaur |
| Applicant Contact Email | rkaur@exalta.com |
## Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Anteralign LS coverplate |
| --- | --- |
| Common Name | Intervertebral body fusion device |
| Classification Name | Intervertebral Fusion Device With Bone Graft, Lumbar |
| Regulation Number | 888.3080 |
| Product Code(s) | MAX, OVD |
## Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K214010 | Anteralign™ Spinal System with Titan nanoLOCK™ Surface Technology | MAX |
| K222465 | Tyber Medical Anatomical Plating System | HRS |
## Device Description Summary
21 CFR 807.92(a)(4)
The Anteralign™ LS coverplate System consists of a coverplate with a locking screw and an integrated locking mechanism. All implant components are manufactured from titanium alloy conforming to ASTM F136. The Anteralign™ LS coverplate may only be used with Medtronic's Anteralign™ LS Spinal System. The coverplate is not a required element for use of Medtronic's Anteralign™ LS Spinal System.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
When used with Anteralign™ LS Spinal System, the Anteralign™ LS coverplate and cage is only intended to be used according to the indications of Medtronic's Anteralign™ LS Spinal System.
Anteralign™ LS Spinal System Indications for Use
The Anteralign™ LS Spinal System is intended to be used in spinal fusion procedures on skeletally mature patients with symptomatic Degenerative Disc Disease (DDD, defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies), degenerative spondylolisthesis, and/or spinal stenosis, at one or two contiguous levels from L2 to S1 whose condition requires the use of interbody fusion. These patients may also have up to Grade 1 Spondylolisthesis or retrolisthesis at the involved levels. These patients should have had six months of nonoperative treatment prior to treatment with this device. Additionally, the Anteralign™ LS Spinal System and optional Anteralign™ LS Coverplate can be used as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity.
The Anteralign™ LS interbody cage may be used as a stand-alone device or in conjunction with supplemental fixation. The Anteralign
{5}
LS™ interbody fusion Device e may be inserted via minimally invasive or open anterior or oblique approach at one or two contiguous levels from L2 to S1. These approaches include anterior and oblique. When used as a stand-alone device, the Anteralign™ LS cage must be used with 3 screws with devices that have standard lordosis (≤16°). If the physician chooses to use less than 3 screws or none of the provided screws, additional supplemental fixation in the lumbar spine must be used to augment stability. When used in patients as an adjunct to fusion in patients diagnosed with multilevel degenerative scoliosis and sagittal deformity conditions, additional supplemental fixation (e.g. posterior fixation) must be used. Additionally, cages with lordosis angles 16° or greater are intended to be used with supplemental fixation (e.g. facet screws or posterior fixation).
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The subject device has the same indications for use as the predicate device.
## Technological Comparison
21 CFR 807.92(a)(6)
A comparison of the subject device and the predicate device demonstrated that the Anteralign LS coverplate is substantially equivalent to the previously cleared Anteralign Spinal System (K214010) in regards to the technological characteristic.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
The following Non-clinical Testing was performed:
1) Clinical Simulation Validation
2) MR Testing
3) Mechanical Testing
Mechanical performance was assessed against the predicate device against the following FDA Guidance and test standards:
1) "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Intervertebral Body Fusion Device",
2) ASTM F2077 "Standard Test Methods for Intervertebral Body Fusion Devices",
3) ASTM F2267 "Standard Test Method for Measuring Load-Induced Subsidence of Intervertebral Body Fusion Device Under Static Axial Compression",
4) DRAFT ASTM Z8423Z Static Push-Out Test Method for Intervertebral Body Fusion
The conclusions drawn from the nonclinical and clinical tests demonstrate that the subject device is as safe and effective, and performs as well as the legally marketed predicate device.
No clinical testing was performed.
Nonclinical testing concluded that the Anteralign LS Coverplate System is substantially equivalent to the Anteralign LS system currently on market. The coverplate does not introduce a new worst case construct in any of the test conditions, as provided in MEMO-X02-05-02A and TR-202535-X02-05.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.