Venus Diamond: Direct restorations of all cavity classes • Direct composite veneers • Shape corrections of teeth (i.e. diastemas, congenital defects in teeth, etc.) • Splinting of teeth • Restoration of primary teeth • Core build-up • Repairs of porcelain, composite (in combination with an adequate repair-system) • Composite attachments for aligner treatment • Fixation of retainers. Venus Diamond Flow: Enlarged fissure sealing • Cavity lining – as the first layer for Class I and II cavities • Class I, II, III, IV and V direct restorations • Small repairs of direct and indirect restorations combined with a suitable bonding agent • Splinting of teeth • Composite attachments for aligner treatment • Fixation of retainers. Venus Pearl: Direct restorations of all cavity classes • Direct composite veneers • Shape corrections of teeth (i.e. diastemas, congenital defects in teeth, etc.) • Splinting of teeth • Restoration of primary teeth • Core build-up • Repairs of porcelain, composite (in combination with an adequate repair-system) • Composite attachments for aligner treatment • Fixation of retainers. Venus Bulk Flow One: Class I, II, III, IV and V direct restorations • First-layer lining for class I and II cavities • Repair of direct and indirect restorations in combination with a suitable adhesive • Splinting of teeth • Extended fissure sealing • Composite attachments for aligner treatment • Fixation of retainers.
Device Story
Light-curing, radiopaque, nano-hybrid dental composites; supplied in syringes or PLT units. Used by dental professionals in clinical settings for restorative procedures and orthodontic applications. Input: composite material applied to tooth structure or aligner templates. Transformation: polymerization via dental curing lights (450–480 nm). Output: hardened composite restoration, attachment, or fixation. Healthcare providers apply material, shape, and cure; output provides structural restoration or orthodontic anchorage. Benefits: tooth repair, aesthetic correction, and orthodontic treatment support.
Clinical Evidence
No clinical data submitted. Substantial equivalence established via bench testing, including flexural strength, radiopacity, depth of cure, wear resistance, bond strength, thermal cycling, and biocompatibility (ISO 10993).
Indicated for dental patients requiring direct restorations (all classes), fissure sealing, cavity lining, tooth shape correction, splinting, core build-up, repair of existing restorations, composite attachments for aligner therapy, and retainer fixation.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 11, 2026
Kulzer, LLC
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120a
Madison, New Jersey 07940
Re: K260783
Trade/Device Name: Venus Diamond; Venus Diamond Flow; Venus Pearl; Venus Bulk Flow One
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth shade resin material
Regulatory Class: Class II
Product Code: EBF, EBC, LBH
Dated: March 9, 2026
Received: March 10, 2026
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260783 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{3}
K260783 - Dave Yungvirt
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260783 | |
| Device Name Venus Bulk Flow One | |
| Indications for Use (Describe) • Class I, II, III, IV and V direct restorations • First-layer lining for class I and II cavities • Repair of direct and indirect restorations in combination with a suitable adhesive • Splinting of teeth • Extended fissure sealing • Composite attachments for aligner treatment • Fixation of retainers | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{5}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260783 | |
| Device Name Venus Diamond Flow | |
| Indications for Use (Describe) • Enlarged fissure sealing • Cavity lining – as the first layer for Class I and II cavities • Class I, II, III, IV and V direct restorations • Small repairs of direct and indirect restorations combined with a suitable bonding agent • Splinting of teeth • Composite attachments for aligner treatment • Fixation of retainers | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{6}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260783 | |
| Device Name Venus Diamond | |
| Indications for Use (Describe) • Direct restorations of all cavity classes • Direct composite veneers • Shape corrections of teeth (i.e. diastemas, congenital defects in teeth, etc.) • Splinting of teeth • Restoration of primary teeth • Core build-up • Repairs of porcelain, composite (in combination with an adequate repair-system) • Composite attachments for aligner treatment • Fixation of retainers | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{7}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260783 | |
| Device Name Venus Pearl | |
| Indications for Use (Describe) • Direct restorations of all cavity classes • Direct composite veneers • Shape corrections of teeth (i.e. diastemas, congenital defects in teeth, etc.) • Splinting of teeth • Restoration of primary teeth • Core build-up • Repairs of porcelain, composite (in combination with an adequate repair-system) • Composite attachments for aligner treatment • Fixation of retainers | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{8}
K260783
KULZER
MITSUI CHEMICALS GROUP
510(k) Summary – Traditional 510(k)
Date Prepared: 03/09/2026
## Contact Details
Applicant Name: Kulzer LLC
Applicant Address: 4315 S Lafayette Blvd South Bend, IN 46614 United States
Applicant Contact Telephone: (574) 299-5467
Applicant Contact: Mr. Aiden Lane
Applicant Contact Email: aiden.lane@kulzer-dental.com
## Device Name and Classification (Bundled Submission):
### Device Trade Name:
- Venus Diamond
- Venus Diamond Flow
- Venus Pearl
- Venus Bulk Flow One
Common Name: Tooth shade resin material
Classification Name: Material, Tooth Shade, Resin
Regulation Number: 872.3690
Product Code(s): EBF
## Bundling Rationale:
This 510(k) submission includes four polymer-based, tooth-shade resin materials regulated under 21 CFR 872.3690 (product code EBF). Because the devices share similar technology, intended use, testing methodologies, and risk profiles, bundling them into a single submission is appropriate and consistent with FDA's “Bundling Multiple Devices or Multiple Indications in a Single Submission” policy.
{9}
KULZER
MITSUI CHEMICALS GROUP
# Legally Marketed Predicate Devices
For the purpose of substantial equivalence, the following primary predicates apply:
- Venus Diamond -> Neun (K073554)
- Venus Diamond Flow -> Venus Diamond Flow (K091635)
- Venus Pearl -> Venus Pearl PLT A1 (K112501)
- Venus Bulk Flow ONE -> Venus Bulk Flow ONE (K220605)
# Reference Devices
Used to support methods/benchmarks for aligner-related indications:
- AlignerFlow LC (K231817) – DYH, 21 CFR 872.3750
- GC Ortho Connect / GC Ortho Etching Gel (Aligner Connect) (K150128) – DYH, 21 CFR 872.3750
# Device Description
All four devices are light-curing, radiopaque, polymer-based dental composites (nano/nanohybrid) supplied in syringes and PLT units for professional use. They are applied in thin layers and polymerized using dental curing lights (typ. 450–480 nm; ~600–1550 mW/cm²). Packaging sizes and handling characteristics (viscosity/flow for the “Flow” and “Bulk Flow” variants; modeling characteristics for “Diamond” and “Pearl”) align with their intended use in direct restorations and related clinical procedures, including aligner attachment fabrication and retainer fixation.
Representative compositions include inorganic fillers (e.g., barium alumino-borosilicate glass, silica, ytterbium fluoride), dimethacrylate resins (e.g., UDMA, TEGDMA, TCD-urethaneacrylate, E4BADMA), initiators (e.g., camphorquinone), stabilizers (e.g., BHT), and pigments/fluorescent agents. Radiopacity meets ISO 4049 acceptance criteria and is within the enamel/dentin-relevant ranges reported for each product.
# Indications for Use (Per Device)
Venus Diamond:
Direct restorations of all cavity classes • Direct composite veneers • Shape corrections of teeth (i.e. diastemas, congenital defects in teeth, etc.) • Splinting of teeth • Restoration of primary teeth • Core build-up • Repairs of porcelain, composite (in combination with an adequate repair-system) • Composite attachments for aligner treatment • Fixation of retainers
Venus Diamond Flow:
- Enlarged fissure sealing • Cavity lining – as the first layer for Class I and II cavities • Class I, II, III, IV and V direct restorations • Small repairs of direct and indirect restorations combined with a suitable bonding agent • Splinting of teeth • Composite attachments for aligner treatment • Fixation of retainers
{10}
KULZER
MITSUI CHEMICALS GROUP
Venus Pearl:
- Direct restorations of all cavity classes • Direct composite veneers • Shape corrections of teeth (i.e. diastemas, congenital defects in teeth, etc.) • Splinting of teeth • Restoration of primary teeth • Core build-up • Repairs of porcelain, composite (in combination with an adequate repair-system) • Composite attachments for aligner treatment • Fixation of retainers
Venus Bulk Flow One:
- Class I, II, III, IV and V direct restorations • First-layer lining for class I and II cavities • Repair of direct and indirect restorations in combination with a suitable adhesive • Splinting of teeth • Extended fissure sealing • Composite attachments for aligner treatment • Fixation of retainers
Indications for Use Comparison – comparison to predicates & justification
All core restorative indications are the same as those of the respective primary predicates within 21 CFR 872.3690 (EBF). The added aligner-related indications (composite attachments; retainer fixation) fall within the same intended use (restorative/composite materials for intraoral application by dental professionals) and do not alter the devices' general purpose or mechanism of action. Performance aspects relevant to these uses (e.g., bond strength, depth of cure, wear, radiopacity, handling) are addressed by bench testing using methods and acceptance criteria aligned with polymer-based restorative materials, and by reference device experience for aligner applications; no new questions of safety or effectiveness are raised.
Differences between the subject devices' indications and those of the predicates do not alter the intended therapeutic use and do not raise new questions of safety or effectiveness. The added aligner-related uses fall within the same general intended use of polymer-based restorative composite systems as described in 21 CFR 807.92(a)(5).
{11}
4
KULZER
MITSUI CHEMICALS GROUP
# Technological Comparison
## Venus Bulk Flow One
Venus Bulk Flow One is a flowable, light curing, radio-opaque nano-hybrid composite. Venus Bulk Flow One's primary packaging consists of a 2.0 g syringe with a plunger and closing cap, intended for use at room temperature and 30–70% relative humidity. A prefilled 0.2g PLT and dispensing tips are used for dispensing the product. The composite is applied in thin layers (maximum 4 mm) to the designated area. Each layer is then individually polymerized using a suitable dental light-curing unit (e.g., Kulzer Translux®) with the following operating parameters: wavelength peak of 450–480 nm and light output of 1550–600 mW/cm².
The predicate device Venus Bulk Flow One (K220605) is also a flowable, light curing, radio-opaque nano-hybrid composite with an identical composition to the applicant device. The primary packaging is also a 2.0 g syringe with a plunger and closing cap, intended for use at room temperature and 30–70% relative humidity, a prefilled 0.2g PLT and dispensing tips are used for dispensing the product. The composite is applied in the same way as the application device, in thin layers (max. 4mm) to the designated area that are polymerized individually using a suitable dental light-curing unit. Immediately after polymerization, both the applicant and predicate devices are indicated for rough finishing using diamond burs and final shaping and contouring using a fine-grained diamond stone, a multi-blade tungsten carbide bur, polishing strips, finishing diamonds, flexible discs, silicone polishers, and polishing brushes.
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Trade Name | Venus Bulk Flow ONE | Venus Bulk Flow ONE | Aligner Flow LC | GC Ortho Connect, GC Ortho Etching Gel (Aligner Connect) |
| 510(k) | N/A | K220605 | K231817 | K150128 |
| Regulation Name | Tooth Shade Resin Material | Tooth Shade Resin Material | Bracket Adhesive Resin And Tooth Conditioner | Bracket Adhesive Resin And Tooth Conditioner |
| Regulation Number | 21 CFR 872.3690 | 21 CFR 872.3690 | 21 CFR 872.3750 | 21 CFR 872.3750 |
| Regulatory Class | Class II | Class II | Class II | Class II |
| Product Code | EBF | EBF | DYH | DYH |
| Prescription/OTC | Prescription | Prescription | Prescription | Prescription |
| Manufacturer | Kulzer GmbH | Kulzer GmbH | VOCO GmbH | GC Orthodontics America Inc. |
| Physical Properties | All physical properties of the applicant device are identical to previous predicate device "Venus Bulk Flow ONE" as no composition changes have been made.
Depth of cure: min. 4.0 ± 0.5 mm
Flexural strength: min. 100 MPa
Wear Resistance: 117 ± 17 μm
Dimension stability: 3.5 ± 0.1 %-vol
Fluorescence: 118% | Depth of cure: min. 4.0 ± 0.5 mm
Flexural strength: min. 100 MPa
Wear Resistance: 117 ± 17 μm
Dimension stability: 3.5 ± 0.1 %-vol
Fluorescence: 118%
Compatibility to aligner materials: 0.19 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: 2.3 mm
Flexural strength: 128 MPa
Wear Resistance: 143 ± 8 μm
Dimension stability: 3.2 ± 0.1 %-vol
Fluorescence: 134%
Compatibility to aligner materials: 0.18 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: 3.5 mm
Flexural strength: 115 MPa
Wear Resistance: 138 ± 20 μm
Dimension stability: 3.7 ± 0.1 %-vol
Fluorescence: 112%
Compatibility to aligner materials: 0.20 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage |
{12}
5
KULZER
MITSUI CHEMICALS GROUP
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Application Characteristics | Competibility to aligner materials: 0.19 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage
The following steps represent the standard clinical procedure for direct composite restorations using Venus Bulk Flow ONE. These are consistent with generally accepted practices in restorative dentistry.
1. Tooth Preparation
2. Isolation
3. Matrix and Wedge Placement
4. Adhesive Application
5. Hygienic Sleeve Usage
6. Composite Placement
7. Alternative Usage
8. Light Curing
9. Layering
10. Optional Topcoat
11. Finishing and Polishing
12. Remove Protective Sleeve
13. Dispose of Cannula and Reseal
14. Do Not Reuse Contaminated Syringes
15. Dispose of Used PLT
Note: Detailed instructions and specifications are provided in the product's official IFU. In addition to the steps listed above, further guidance has been included regarding aligner therapy, as follows
Application of attachments for clear aligner therapy
1. fill the composite into the attachment's recesses of the template. Prepare the enamel to be treated as described under I. Filling therapy. Place the template on the prepared row of teeth and cure the | The following steps represent the standard clinical procedure for direct composite restorations using Venus Bulk Flow ONE. These are consistent with generally accepted practices in restorative dentistry.
1. Tooth Preparation
2. Isolation
3. Matrix and Wedge Placement
4. Adhesive Application
5. Hygienic Sleeve Usage
6. Composite Placement
7. Alternative Usage
8. Light Curing
9. Layering
10. Optional Topcoat
11. Finishing and Polishing
12. Remove Protective Sleeve
13. Dispose of Cannula and Reseal
14. Do Not Reuse Contaminated Syringes
15. Dispose of Used PLT
Note: Detailed instructions and specifications are provided in the product's official IFU. Always refer to the IFU for comprehensive guidance. | The composite is used in the adhesive technique with a dentine enamel bond. Any light-curing bonding material can be used. The extruded composite is applied into the aligner template, placed on the prepared teeth and each attachment is light cured through the transparent template for the fabrication of attachments. Following removal of the attachment template, finish the attachments in accordance with generally applicable standards if and as necessary.
Once the aligner treatment is completed, the attachments can be removed using a grinding/finishing diamond burr. Subsequent polishing of the tooth surface is then recommended. Excess material can be made visible on the tooth surface with a UV lamp (360-395 nm) and removed accordingly. | For bonding GC Aligner Connect to enamel and/or dentin, use a light-cured bonding system (e.g. G-Premio BOND). Follow bonding agent manufacturer's instructions.
Place dispensing tip as close as possible to the transfer tray, and slowly push the plunger to extrude material.
Set transfer tray in the oral cavity. Light cure GC Aligner Connect on each attachment area with an LED light curing device.
Remove transfer tray from the oral cavity.
If necessary, finish and polish using standard techniques.
When orthodontic treatment no longer requires the attachment, remove attachment with diamond bur or silicon point. |
5
{13}
KULZER
MITSUI CHEMICALS GROUP
| Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- |
| attachments according to the table.
2. After completion of the clear aligner therapy, remove the composite attachments with rotating instruments and polish the tooth surface, e.g. Venus Supra.
3. UV-A -light may be used for detection in order to differentiate between composite and tooth substance. | | | |
# Venus Diamond Flow
Venus Diamond Flow is a flowable, light curing, radio-opaque nano-hybrid composite. Venus Diamond Flow's primary packaging is a 1.8g syringe with a plunger and a closing cap, intended for use at room temperature and 30–70% relative humidity. A prefilled 0.2g PLT and flow cannulas are used for dispensing the product. The composite is applied in thin layers (max. 2mm, baseliner max. 1mm) to the designated area. Each layer is then individually polymerized using a suitable dental light-curing unit (e.g., Kulzer Translux®) with the following operating parameters: wavelength peak of 450–480 nm and light output of 1550–600 mW/cm².
The predicate device Venus Diamond Flow (K091635) is a flowable, light curing, radio-opaque nano-hybrid composite with an identical composition to the applicant device. The primary packaging is also a 2.0g syringe with a plunger and a closing cap, intended for use at room temperature and 30–70% relative humidity, and a prefilled 1.8g PLT and flow cannulas are used for dispensing the product. The composite is applied in the same way as the application device, in thin layers (max. 2mm, baseliner max. 1mm) to the designated area that are polymerized individually using a suitable dental light-curing unit. Immediately after polymerization, both the applicant and predicate devices are indicated to be prepared and polished by finishing diamonds, flexible discs, silicone polishers and polishing brushes.
| Trade Name | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Venus Diamond Flow | Venus Diamond Flow | Aligner Flow LC | GC Ortho Connect, GC Ortho Etching Gel
(Aligner Connect) | |
| 510(k) | N/A | K091635 | K231817 | K150128 |
| Regulation Name | Tooth Shade Resin Material | Sealant Pit and Fissure and Conditioner | Bracket Adhesive Resin and Tooth Conditioner | Bracket Adhesive Resin and Tooth Conditioner |
| Regulation Number | 21 CFR 872.3690 | 21 CFR 872.3690 | 21 CFR 872.3750 | 21 CFR 872.3750 |
| Regulatory Class | Class II | Class II | Class II | Class II |
| Product Code | EBF | EBF | DYH | DYH |
| Prescription/OTC | Prescription | Prescription | Prescription | Prescription |
{14}
7
KULZER
MITSUI CHEMICALS GROUP
| Manufacturer | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Physical Properties | All physical properties of the applicant device are identical to previous predicate device “Venus Diamond Flow” as no composition changes have been made.
Depth of cure: min. 2.0 ± 0.5 mm
Flexural strength: min. 100 MPa
Wear Resistance: 117 ± 17 μm
Dimension stability: 3.5 ± 0.1 %-vol
Fluorescence: 118%
Compatibility to aligner materials: 0.15 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: min. 2.0 ± 0.5 mm
Flexural strength: min. 100 MPa
Wear Resistance: 117 ± 17 μm
Dimension stability: 3.5 ± 0.1 %-vol
Fluorescence: 118%
Compatibility to aligner materials: 0.15 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: 2.3 mm
Flexural strength: 128 MPa
Wear Resistance: 143 ± 8 μm
Dimension stability: 3.2 ± 0.1 %-vol
Fluorescence: 134%
Compatibility to aligner materials: 0.18 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: 3.5 mm
Flexural strength: 115 MPa
Wear Resistance: 138 ± 20 μm
Dimension stability:3.7 ± 0.1 %-vol
Fluorescence: 112%
Compatibility to aligner materials: 0.20 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage |
| Application Characteristics | The following steps represent the standard clinical procedure for direct composite restorations using Venus Diamond Flow These are consistent with generally accepted practices in restorative dentistry.
1. Rubber Dam Placement
2. Cavity Preparation
3. Adhesive Application
4. Composite Placement
5. Light Curing
6. Layer
7. Finishing and Polishing
Note: Detailed instructions and specifications are provided in the product's official IFU. In addition to the steps listed above, further guidance has been included regarding aligner therapy, as follows
Application of attachments for clear aligner therapy
1. fill the composite into the attachment's recesses of the template. Prepare the enamel to be treated as described under I. Filling therapy. Place the template on the prepared row of teeth and cure the | The following steps represent the standard clinical procedure for direct composite restorations using Venus Diamond Flow. These are consistent with generally accepted practices in restorative dentistry.
1. Rubber Dam Placement
2. Cavity Preparation
3. Adhesive Application
4. Composite Placement
5. Light Curing
6. Layer
7. Finishing and Polishing
Note: Detailed instructions and specifications are provided in the product's official IFU. Always refer to the IFU for comprehensive guidance. | The composite is used in the adhesive technique with a dentine enamel bond. Any light-curing bonding material can be used. The extruded composite is applied into the aligner template, placed on the prepared teeth and each attachment is light cured through the transparent template for the fabrication of attachments. Following removal of the attachment template, finish the attachments in accordance with generally applicable standards if and as necessary.
Once the aligner treatment is completed, the attachments can be removed using a grinding/finishing diamond burr. Subsequent polishing of the tooth surface is then recommended.
Excess material can be made visible on the tooth surface with a UV lamp (360-395 nm) and removed accordingly. | For bonding GC Aligner Connect to enamel and/or dentin, use a light-cured bonding system (e.g. G-Premio BOND). Follow bonding agent manufacturer's instructions.
Place dispensing tip as close as possible to the transfer tray, and slowly push the plunger to extrude material.
Set transfer tray in the oral cavity.
Light cure GC Aligner Connect on each attachment area with an LED light curing device.
Remove transfer tray from the oral cavity.
If necessary, finish and polish using standard techniques.
When orthodontic treatment no longer requires the attachment, remove attachment with diamond bur or silicon point. |
7
{15}
8
KULZER
MITSUI CHEMICALS GROUP
| Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- |
| attachments according to the table.
2. After completion of the clear aligner therapy, remove the composite attachments with rotating instruments and polish the tooth surface, e.g. Venus Supra.
3. UV-A -light may be used for detection in order to differentiate between composite and tooth substance. | | | |
8
{16}
9
KULZER
MITSUI CHEMICALS GROUP
Venus Diamond
Venus Diamond is a light-curing, radiopaque nano-composite. Its primary packaging consists of a 4.0 g syringe with a plunger and closing cap. The product should be used above room temperature, within 30–70% relative humidity, and may be heated for use up to a maximum of 129 °F. The individual application quantity of the composite for the treatment is usually taken from the syringe and presented in a tray using a sterile dental instrument. Otherwise a prefilled 0.25g PLT is used for dispensing the product. The composite is applied in thin layers (max. 1mm-2mm, depending on shade) to the designated area. Each layer is then individually polymerized using a suitable dental light-curing unit (e.g., Kulzer Translux®) with the following operating parameters: wavelength peak of 450–480 nm and light output of 1550–600 mW/cm².
The predicate device Neun (K073554) is also a light curing, radio-opaque nano composite with an identical composition to the applicant device. The primary packaging is also a 4.0g syringe with a plunger and a closing cap and a prefilled 0.25g PLT. The composite is applied in the same way as the application device, in thin layers (max. 1mm-2mm, depending on shade) to the designated area that are polymerized individually using a suitable dental light-curing unit. After final build up and polymerization is completed, both the applicant and predicate devices are indicated to be finished and polished by diamond burs, fine-grained diamond stones, multi-fluted carbide burs, silicone rubber polishing points, cups, discs, or similar tools.
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Trade Name | Venus Diamond | Neun | Aligner Flow LC | GC Ortho Connect, GC Ortho Etching Gel (Aligner Connect) |
| 510(k) | N/A | K073554 | K231817 | K150128 |
| Regulation Name | Tooth Shade Resin Material | Tooth Shade Resin Material | Bracket Adhesive Resin And Tooth Conditioner | Bracket Adhesive Resin And Tooth Conditioner |
| Regulation Number | 21 CFR 872.3690 | 21 CFR 872.3690 | 21 CFR 872.3750 | 21 CFR 872.3750 |
| Regulatory Class | Class II | Class II | Class II | Class II |
| Product Code | EBF | EBF | DYH | DYH |
| Prescription/OTC | Prescription | Prescription | Prescription | Prescription |
| Manufacturer | Kulzer GmbH | Kulzer GmbH | VOCO GmbH | GC Orthodontics America Inc. |
| Physical Properties | All physical properties of the applicant device are identical to previous predicate device “Neun” as no composition changes have been made.
Depth of cure: min. 2.0 ± 0.5 mm
Flexural strength: min. 130 MPa
Wear Resistance: 153 ± 9 μm
Dimension stability: 1.7 ± 0.1 %-vol
Fluorescence: 118%
Compatibility to aligner materials: 0.15 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: min. 2.0 ± 0.5 mm
Flexural strength: min. 130 MPa
Wear Resistance: 153 ± 9 μm
Dimension stability: 1.7 ± 0.1 %-vol
Fluorescence: 118%
Compatibility to aligner materials: 0.15 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: 2.3 mm
Flexural strength: 128 MPa
Wear Resistance: 143 ± 8 μm
Dimension stability: 3.2 ± 0.1 %-vol
Fluorescence: 134%
Compatibility to aligner materials: 0.18 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Depth of cure: 3.5 mm
Flexural strength: 115 MPa
Wear Resistance: 138 ± 20 μm
Dimension stability:3.7 ± 0.1 %-vol
Fluorescence: 112%
Compatibility to aligner materials: 0.20 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage |
{17}
KULZER
MITSUI CHEMICALS GROUP
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Application Characteristics | Microleakage: Low level of microleakage | | | |
| | The following steps represent the standard clinical procedure for direct composite restorations using Venus Diamond. These are consistent with generally accepted practices in restorative dentistry. | The following steps represent the standard clinical procedure for direct composite restorations using Venus Diamond. These are consistent with generally accepted practices in restorative dentistry. | The composite is used in the adhesive technique with a dentine enamel bond. Any light-curing bonding material can be used. The extruded composite is applied into the aligner template, placed on the prepared teeth and each attachment is light cured through the transparent template for the fabrication of attachments. Following removal of the attachment template, finish the attachments in accordance with generally applicable standards if and as necessary. Once the aligner treatment is completed, the attachments can be removed using a grinding/finishing diamond burr. Subsequent polishing of the tooth surface is then recommended. Excess material can be made visible on the tooth surface with a UV lamp (360-395 nm) and removed accordingly. | For bonding GC Aligner Connect to enamel and/or dentin, use a light-cured bonding system (e.g. G-Premio BOND). Follow bonding agent manufacturer’s instructions. Place dispensing tip as close as possible to the transfer tray, and slowly push the plunger to extrude material. Set transfer tray in the oral cavity. Light cure GC Aligner Connect on each attachment area with an LED light curing device. Remove transfer tray from the oral cavity. If necessary, finish and polish using standard techniques. When orthodontic treatment no longer requires the attachment, remove attachment with diamond bur or silicon point. |
| | 8. Rubber Dam Placement | 1. Rubber Dam Placement | | |
| | 9. Cavity Preparation | 2. Cavity Preparation | | |
| | 10. Adhesive Application | 3. Adhesive Application | | |
| | 11. Composite Placement | 4. Composite Placement | | |
| | 12. Light Curing | 5. Light Curing | | |
| | 13. Layer | 6. Layer | | |
| | 14. Finishing and Polishing | 7. Finishing and Polishing | | |
| | Note: Detailed instructions and specifications are provided in the product’s official IFU. In addition to the steps listed above, further guidance has been included regarding aligner therapy, as follows | Note: Detailed instructions and specifications are provided in the product’s official IFU. Always refer to the IFU for comprehensive guidance. | | |
| Application of attachments for clear aligner therapy | | | | |
| 1. Fill the composite into the attachment’s recesses of the template. Prepare the enamel to be treated as described under I. Filling therapy. Place the template on the prepared row of teeth and cure the attachments according to the table. | | | | |
| 2. After completion of the clear aligner therapy, remove the composite attachments with rotating instruments and polish the tooth surface, e.g. Venus Supra. | | | | |
| 3. UV-A -light may be used for detection in order to differentiate between composite and tooth substance. | | | | |
10
{18}
11
KULZER
MITSUI CHEMICALS GROUP
The tables differ mainly in indications for use, physical properties, and application steps, with applicant devices adding aligner-related functions not found in predicates, while reference devices are aligner-only. Regulation numbers, trade names, and 510(k) listings also vary across devices.
Venus Pearl
Venus Pearl is a light curing, radio-opaque nanocomposite. Its primary packaging is a 3.0g syringe with a plunger and a closing cap. The product should be used above room temperature, within 30–70% relative humidity, and may be heated for use up to a maximum of 129 °F. The individual application quantity of the composite for the treatment is usually taken from the syringe and presented in a tray using a sterile dental instrument. Otherwise a prefilled 0.2g PLT is used for dispensing the product. The composite is applied in thin layers (max. 1mm-3mm, depending on shade) to the designated area. Each layer is then individually polymerized using a suitable dental light-curing unit (e.g., Kulzer Translux®) with the following operating parameters: wavelength peak of 450–480 nm and light output of 1550–600 mW/cm².
The predicate device Venus Pearl PLT Refill A1 (K112501) is also a light curing, radio-opaque nano composite with an identical composition to the applicant device. The primary packaging is also a 3.0g syringe with a plunger and a closing cap and a prefilled 0.2g PLT. The composite is applied in the same way as the application device, in thin layers (max. 1mm-3mm, depending on shade) to the designated area that are polymerized individually using a suitable dental light-curing unit. After final build up and polymerization is completed, both the applicant and predicate devices are indicated for gross finishing using diamond burs and final shaping and finishing using fine-grained diamond stones, multifluted carbide burs, abrasive strips, flexible silicone rubber polishing points, cups, discs, or similar tools.
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| Trade Name | Venus Pearl | VENUS PEARL PLT REFILL A1 | Aligner Flow LC | GC Ortho Connect, GC Ortho Etching Gel (Aligner Connect) |
| 510(k) | N/A | K112501 | K231817 | K150128 |
| Regulation Name | Tooth Shade Resin Material | Tooth Shade Resin Material | Bracket Adhesive Resin And Tooth Conditioner | Bracket Adhesive Resin And Tooth Conditioner |
| Regulation Number | 21 CFR 872.3690 | 21 CFR 872.3690 | 21 CFR 872.3750 | 21 CFR 872.3750 |
| Regulatory Class | Class II | Class II | Class II | Class II |
| Product Code | EBF | EBF | DYH | DYH |
| Prescription/OTC | Prescription | Prescription | Prescription | Prescription |
| Manufacturer | Kulzer GmbH | Kulzer GmbH | VOCO GmbH | GC Orthodontics America Inc. |
| Physical Properties | All physical properties of the applicant device are identical to previous predicate device “Venus Pearl” as no composition changes have been made.
Depth of cure: min. 2.0 ± 0.5 mm
Flexural strength: min. 125 MPa
Wear Resistance: 131 ± 6 μm
Dimension stability: 1.9 ± 0.1 %-vol | Depth of cure: min. 2.0 ± 0.5 mm
Flexural strength: min. 125 MPa
Wear Resistance: 131 ± 6 μm
Dimension stability: 1.9 ± 0.1 %-vol
Fluorescence: 116%
Compatibility to aligner materials: 0.14 mm size reduction after 1000 cycles
Bond Strength: Pass | Depth of cure: 2.3 mm
Flexural strength: 128 MPa
Wear Resistance: 143 ± 8 μm
Dimension stability: 3.2 ± 0.1 %-vol
Fluorescence: 134%
Compatibility to aligner materials: 0.18 mm size reduction after 1000 cycles
Bond Strength: Pass | Depth of cure: 3.5 mm
Flexural strength: 115 MPa
Wear Resistance: 138 ± 20 μm
Dimension stability:3.7 ± 0.1 %-vol
Fluorescence: 112%
Compatibility to aligner materials: 0.20 mm size reduction after 1000 cycles
Bond Strength: Pass |
{19}
KULZER
MITSUI CHEMICALS GROUP
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| | Fluorescence: 116%
Compatibility to aligner materials: 0.14 mm size reduction after 1000 cycles
Bond Strength: Pass
Microleakage: Low level of microleakage | Microleakage: Low level of microleakage | Microleakage: Low level of microleakage | Microleakage: Low level of microleakage |
| Application Characteristics | The following steps represent the standard clinical procedure for direct composite restorations using Venus Pearl. These are consistent with generally accepted practices in restorative dentistry.
1. Rubber Dam Placement
2. Cavity Preparation
3. Adhesive Application
4. Composite Placement
5. Light Curing
6. Layer
7. Finishing and Polishing
Note: Detailed instructions and specifications are provided in the product's official IFU. In addition to the steps listed above, further guidance has been included regarding aligner therapy, as follows
Application of attachments for clear aligner therapy
1. Fill the composite into the attachment's recesses of the template. Prepare the enamel to be treated as described under I. Filling therapy. Place the template on the prepared row of teeth and cure the attachments according to the table.
2. After completion of the clear aligner therapy, remove the composite attachments with rotating instruments and polish the tooth surface, e.g. Venus Supra.
3. UV-A -light may be used for detection in order to differentiate between | The following steps represent the standard clinical procedure for direct composite restorations using Venus Pearl. These are consistent with generally accepted practices in restorative dentistry.
1. Rubber Dam Placement
2. Cavity Preparation
3. Adhesive Application
4. Composite Placement
5. Light Curing
6. Layer
7. Finishing and Polishing
Note: Detailed instructions and specifications are provided in the product's official IFU. Always refer to the IFU for comprehensive guidance. | The composite is used in the adhesive technique with a dentine enamel bond. Any light-curing bonding material can be used. The extruded composite is applied into the aligner template, placed on the prepared teeth and each attachment is light cured through the transparent template for the fabrication of attachments. Following removal of the attachment template, finish the attachments in accordance with generally applicable standards if and as necessary. Once the aligner treatment is completed, the attachments can be removed using a grinding/finishing diamond burr. Subsequent polishing of the tooth surface is then recommended. Excess material can be made visible on the tooth surface with a UV lamp (360-395 nm) and removed accordingly. | For bonding GC Aligner Connect to enamel and/or dentin, use a light-cured bonding system (e.g. G-Premio BOND). Follow bonding agent manufacturer's instructions. Place dispensing tip as close as possible to the transfer tray, and slowly push the plunger to extrude material. Set transfer tray in the oral cavity. Light cure GC Aligner Connect on each attachment area with an LED light curing device. Remove transfer tray from the oral cavity. If necessary, finish and polish using standard techniques. When orthodontic treatment no longer requires the attachment, remove attachment with diamond bur or silicon point. |
12
{20}
13
KULZER
MITSUI CHEMICALS GROUP
| | Applicant Device | Predicate Device | Reference Device | Reference Device |
| --- | --- | --- | --- | --- |
| | composite and tooth substance. | | | |
Each table above includes a different applicant and corresponding predicate device; however, each applicant device aligns appropriately with its designated predicate. Each applicant device also demonstrates different material performance values based on its specific requirements. While the application instructions share a similar overall structure and content, the individual steps and level of detail differ. Despite these variations, none introduce new questions of safety or effectiveness relative to the primary predicate devices.
13
{21}
KULZER MITSUI CHEMICALS GROUP
# Non-Clinical and/or Clinical Tests Summary & Conclusions
Testing was performed on representative shades/viscosities of the four devices using recognized consensus standards and validated internal methods, as applicable. All tests met prespecified acceptance criteria and/or applicable standard requirements.
# Performance Testing Summary (as required under 807.92(b)(1)):
Flexural Strength (ISO 4049): All devices exceeded $\geq 100$ MPa requirement; comparable to predicates.
- Radiopacity (ISO 4049): All devices met $\geq 1$ mm/100% Al threshold; results 2.0–3.5 mm Al.
- Depth of Cure: Achieved claimed depths (e.g., 4 mm for Bulk Flow).
- Wear Resistance: Comparable to predicates under simulated masticatory cycling.
- Bond Strength: Acceptable enamel/dentin adhesion comparable to reference aligner materials.
- Thermal Cycling: No degradation affecting performance.
- Biocompatibility: Cytotoxicity, sensitization, irritation, and systemic assessments passed.
All tests support substantial equivalence.
The following FDA-recognized consensus standards were applied, as required under 21 CFR 807.92(b)(1):
- ISO 4049: Dentistry — Polymer-based restorative materials
- ISO 10993-5 / -10 / -11: Biological evaluation of medical devices
Internal validated methods for depth of cure, wear, shrinkage, surface polishability, thermal testing, and handling properties were used.
All testing met or exceeded acceptance criteria.
Both the applicant and the predicate/reference devices use the same testing methods which conclude through biocompatibility, bench, and thermal testing that the applicant devices perform as well or better than the selected predicate/reference devices.
# Clinical Data Statement (807.92(b)(2)):
No clinical data was submitted, referenced, or relied upon to support this submission. Non-clinical performance data was sufficient to establish substantial equivalence.
# Substantial Equivalence Conclusion:
The non-clinical data demonstrate that Venus Diamond, Venus Diamond Flow, Venus Pearl, and Venus Bulk Flow ONE are as safe, as effective, and perform as well as their respective primary predicates for the stated indications. Differences in indications limited to aligner attachments and retainer fixation do not change the intended use or raise new questions of safety or effectiveness; performance expectations are addressed by bench testing and informed by reference device experience for these specific applications. The devices therefore meet the requirements for substantial equivalence under 21 CFR 807.92(b)(3).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.