Polyethylene;Polyethylene Fusion;Polyethylene OrthoTape™

K260775 · Teleflex Medical, LLC · GAT · May 6, 2026 · General, Plastic Surgery

Device Facts

Record IDK260775
Device NamePolyethylene;Polyethylene Fusion;Polyethylene OrthoTape™
ApplicantTeleflex Medical, LLC
Product CodeGAT · General, Plastic Surgery
Decision DateMay 6, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.5000
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Polyethylene Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Polyethylene Fusion Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Polyethylene OrthoTape™ Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries.

Device Story

Braided multifilament surgical sutures composed of Ultra High Molecular Weight Polyethylene (UHMWPE); available in round, coreless, or flat tape configurations; used for soft tissue approximation and ligation in orthopedic procedures. Variants include standard Polyethylene Suture, Polyethylene Fusion (transitioning between round and tape), and Polyethylene OrthoTape™. Devices are provided sterile for single-use; may include attached stainless steel needles. Sutures (without needles) are MR safe. Operation involves manual handling by surgeons during surgical procedures to secure tissue. Output is physical tissue approximation. Benefits include high tensile strength, low knot profile, and improved visibility via colorants (D&C Violet No. 2, Pigment Green 7).

Clinical Evidence

Bench testing only. Performance testing included USP requirements for diameter, needle attachment, and tensile strength, as well as dimensional requirements (height, width, length). MR safety evaluation was conducted for the suture material (excluding needles). Biocompatibility was supported by leveraging data from predicate K181774.

Technological Characteristics

Materials: Ultra High Molecular Weight Polyethylene (UHMWPE) with D&C Violet No. 2 (≤0.2% w/w) or Pigment Green 7 (≤0.5% w/w). Construction: Braided multifilament (round, coreless, or flat tape). Sterilization: Sterile, single-use. MR status: MR safe (without needles). Compliance: USP tensile strength and needle attachment requirements.

Indications for Use

Indicated for approximation and/or ligation of soft tissues, including allograft tissue for orthopedic surgeries.

Regulatory Classification

Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1} FDA U.S. FOOD & DRUG ADMINISTRATION May 06, 2026 Teleflex Medical, LLC Rebecca Walker Sr. Regulatory Affairs Specialist Contact Address Re: K260775 Trade/Device Name: Polyethylene; Polyethylene Fusion; Polyethylene OrthoTape™ Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable Poly(Ethylene Terephthalate) Surgical Suture Regulatory Class: Class II Product Code: GAT Dated: March 11, 2026 Received: March 10, 2026 Dear Rebecca Walker: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K260775 - Rebecca Walker Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {3} K260775 - Rebecca Walker Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, TEK N. LAMICHHANE -S Tek N. Lamichhane, Ph.D. Assistant Director DHT4B: Division of Plastic and Reconstructive Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K260775 | | | Device Name Polyethylene; Polyethylene Fusion; Polyethylene OrthoTape™ | | | Indications for Use (Describe) Polyethylene Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Polyethylene Fusion Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Polyethylene OrthoTape™ Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {5} Traditional 510(k) 510(k) Summary Teleflex Polyethylene Sutures # 510(k) Summary # Teleflex Polyethylene Sutures # K260775 Prepared on: April 16, 2026 Applicant Name: Teleflex Medical, LLC 3015 Carrington Mill Blvd. Applicant Contact: Laura Medlin Laura.Medlin@teleflex.com 919-544-8000 Correspondent: Rebecca Walker Rebecca.Walker@teleflex.com 919-544-8000 # 1. Device Name Device Trade Name: Teleflex Polyethylene Suture Teleflex Polyethylene Fusion Suture Teleflex Polyethylene OrthoTape™ Suture Common Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Classification Name: Suture, Nonabsorbable, Synthetic, Polyethylene Regulation Number: 21 CFR 878.5000 Product Code(s): GAT # 2. Legally Marketed Predicate Devices | Predicate 510(k) # | Predicate Trade Name (Primary Predicate is listed first) | Predicate Product Code | | --- | --- | --- | | K181774 | Teleflex Surgical Sutures (Force Fiber® Suture, Force Fiber® OrthoTape® Suture) | GAT | | K172016 | Force Fiber® Fusion Suture | GAT | | K231163 | Riverpoint HS Suture | GAT | | K162310 | Stryker XBraid TT Suture | GAT | # 3. Device Description Summary Teleflex Polyethylene Sutures (Polyethylene Suture, Polyethylene Fusion Suture, and Polyethylene OrthoTape™ Suture) are braided multifilament made from up to 100% Ultra High Molecular Weight Polyethylene (UHMWPE). The Polyethylene Sutures are available in multiple colors and braid patterns which are used for multi-strand procedures, better visibility and suture differentiation during surgery. They are available multicolored or co-braided with nylon, polypropylene, or polyester. Teleflex Polyethylene Sutures consist of the following variants, differentiated by the braid type: Teleflex Medical, LLC {6} Traditional 510(k) 510(k) Summary Teleflex Polyethylene Sutures - Polyethylene Suture Polyethylene Sutures feature a round and coreless braid that will flatten upon itself, during the tying process, to provide a snug, low knot profile that allows precise knot placement and a smooth tie down. Polyethylene sutures meet USP with the exception of diameter (oversized). - Polyethylene Fusion Suture Polyethylene Fusion Sutures are braided to transition from round coreless suture to tape suture and back to round suture within the same braid construction. The tape section(s) are flat in shape and differ from USP requirements. The round section(s) exceed USP for diameter. The suture meets USP tensile strength requirements and USP needle attachment requirements. - Polyethylene OrthoTape™ Suture Polyethylene OrthoTape™ Sutures are flat in shape and differ from USP requirements. Polyethylene OrthoTape™ is designed to meet the Teleflex internally developed specifications for width, height, length, tensile strength, and needle attachment. All Polyethylene OrthoTape™ Suture sizes meet USP tensile strength requirements and USP needle attachment requirements for USP size #2 suture. All three variants are available in a variety of lengths, with and without attached stainless steel needles. The finished sutures are provided sterile, for single use only and are packaged in cartons as single packs or multi-packs. Teleflex Polyethylene sutures, with the needles removed, are MR safe (i.e., an item that poses no known hazards in all MR environments). The purpose of this submission is to add the following colors to the Teleflex Polyethylene Suture family: - UHMWPE containing D&C Violet No. 2, ≤0.2% w/w - UHMWPE containing Pigment Green 7 (phthalocyanine), ≤0.5% w/w The violet and green UHMWPE colorants are offered in the following configurations for all variants: - Solids (100% UHMWPE of violet, green) - Violet or green co-braids with white polyester, green polyester, black nylon, or blue polypropylene - Violet or green multicolor braids with blue, black or white UHMWPE The Teleflex Polyethylene Suture, Polyethylene Fusion Suture, and Polyethylene OrthoTape™ Suture were formerly marketed by Teleflex as Force Fiber®, Force Fiber® Fusion Suture, and Force Fiber® OrthoTape™ Suture. This submission is removing private label brand names from labeling. 4. Intended Use/Indications for Use Polyethylene Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Polyethylene Fusion Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Polyethylene OrthoTape™ Suture is indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. Teleflex Medical, LLC Page 2 of 4 {7} Traditional 510(k) 510(k) Summary Teleflex Polyethylene Sutures # 5. Substantial Equivalence ## Indications for Use Comparison The Teleflex Polyethylene Suture line extension has identical indications for use as the predicate device(s). Like the predicate Force Fiber and Force Fiber OrthoTape™ Sutures (cleared under Teleflex Surgical Sutures (K181774)) and Force Fiber Fusion Suture (K172016), the subject Polyethylene, Polyethylene Fusion and Polyethylene OrthoTape™ Sutures are all indicated for use in approximation and/or ligation of soft tissues, including use of allograft tissue for orthopedic surgeries. ## Technological Comparison The Teleflex Polyethylene Suture line extension has the same principles of operation, and similar technological characteristics as the predicate Force Fiber and Force Fiber OrthoTape™ sutures, cleared under K181774 and Force Fiber Fusion cleared under K172016. Both the subject and the predicate devices have the same USP performance requirements, are sterilized using the same processes, and are comprised of the same materials of construction. The difference in technological characteristics are limited to the addition of two new color additives on the UHMWPE Fibers, D&C Violet No. 2 and Pigment Green 7 (phthalocyanine), which has been added to all three braid variants within the Teleflex Polyethylene Sutures family. The table below provides cross-reference to the most recent 510(k) clearance for each Teleflex Polyethylene Suture variant, the colorants being added, and the reference devices in which the colorants have received 510(k) clearance. | Subject Device (Predicate 510(k)) | Material/ Color | Colorant | % w/w | Reference Device for Colorant | | --- | --- | --- | --- | --- | | Polyethylene Suture (K181774, K191268) | UHMWPE, Violet | D&C Violet No. 2 | 0.2% | Stryker XBraid TT (K162310) Note^{1} | | Polyethylene Fusion (K172016) | | | | | | Polyethylene OrthoTape™ (K181774) | UHMWPE, Green | Pigment Green 7 (phthalocyanine) | ≤0.5% | Riverpoint Medical, HS Fiber Suture (K231163) | ¹D&C Violet No. 2 at the same w/w has also been previously cleared for Teleflex Bondek™ Sutures (K181774) and Teleflex Monodek™ Sutures (K181774). These are absorbable sutures and have been included as reference. The addition of the violet and green UHMWPE to the Teleflex Polyethylene Suture family does not raise new questions on safety or effectiveness as demonstrated through predicate and reference device comparison and performance testing. Biological evaluation submitted for the predicate under K181774 has been leveraged to support biocompatibility. Additionally, the new colorants used have been cleared for UHMWPE sutures within the same indications for use. # 6. Non-Clinical Testing The Teleflex Polyethylene Sutures have been tested for conformance to the following internal and performance requirements as stated in the applicable special controls for surgical sutures: - Sutures - Diameter <861> - Sutures - Needle Attachment <871> Teleflex Medical, LLC {8} Traditional 510(k) 510(k) Summary Teleflex Polyethylene Sutures - Tensile Strength <881> - Dimensional Requirements (Height, Width, Length) Additionally, an evaluation of safety in the Magnetic Resonance (MR) environment to support the designation of MR Safe (without needles) was conducted. The results of design verification and stability testing completed on the subject device confirm the Teleflex Polyethylene Sutures (Polyethylene Suture, Polyethylene Fusion Suture and Polyethylene OrthoTape™ Suture) have met all required performance requirements and are substantially equivalent to the predicate device. # 7. Conclusions The subject device does not raise new questions on safety or effectiveness as demonstrated through predicate comparison and performance testing presented in this 510(k). Teleflex Polyethylene Sutures (Polyethylene Suture, Polyethylene Fusion Suture and Polyethylene OrthoTape™ Suture) are substantially equivalent to the predicate device. Teleflex Medical, LLC Page 4 of 4
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...