Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Female (CH06 - CH16); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Straight (CH06 - CH18); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Coude (CH10 - CH18).
Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Female (CH06 - CH16); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Straight (CH06 - CH18); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Coude (CH10 - CH18).
Applicant
Convatec
Product Code
EZD · Gastroenterology, Urology
Decision Date
Aug 5, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
To provide an intermittent pathway for drainage of fluids from the bladder. The catheter is inserted through the urethra.
Device Story
Cure Aqua is a sterile, single-use, hydrophilic intermittent urethral catheter for bladder drainage. Device consists of a thermoplastic polyurethane (TPU) shaft with a hydrophilic coating, stored in a wetting solution (distilled water, PVP, NaCl) within a primary pouch. Includes an LDPE handling sleeve to minimize direct contact during insertion and a color-coded TPU funnel. Available in straight (male/female) and coudé (male) tip configurations. Used in home and clinical settings by patients, caregivers, or healthcare professionals. The coudé tip aids navigation of urethral structures; a blue stripe provides visual orientation. The device is inserted through the urethra to drain urine. Benefits include reduced friction during insertion due to the hydrophilic coating and ease of use for patients with adequate dexterity. Sterilized via E-beam irradiation.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1, -5, -10, -11, -23), sterilization validation (ISO 11137-1), packaging integrity (ISO 11607-1/2, ASTM F88, ASTM D4169), and physical performance (ISO 20696 for dimensions, tensile strength, flow rate, kink stability; ASTM D1894 for coefficient of friction). All tests met predetermined acceptance criteria.
Indicated for bladder drainage in males, females, pediatric (infant, child, adolescent), and adult patients with urinary system dysfunction. Prescription use only. Requires assessment by a healthcare provider for appropriate dexterity, cognitive understanding, and anatomical size for intermittent self-catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 5, 2026
Convatec
Kim Jones
Senior Regulatory Affairs Specialist
First Ave., Deeside Industrial Estate
Deeside
Flintshire, CH5 2NU
UNITED KINGDOM
Re: K260737
Trade/Device Name: Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter
Female (CH06 - CH16);
Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter
Male Straight (CH06 - CH18);
Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter
Male Coude (CH10 - CH18)
Regulation Number: 21 CFR 876.5130
Regulation Name: Urological Catheter and accessories
Regulatory Class: II
Product Code: EZD
Dated: July 3, 2026
Received: July 6, 2026
Dear Kim Jones:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260737 - Kim Jones
Page 2
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260737 - Kim Jones
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
ANDREW M. FALES -S
for Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260737 | ? |
| Please provide the device trade name(s). | | ? |
| Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Female (CH06 - CH16); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Straight (CH06 - CH18); Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Coude (CH10 - CH18) | | |
| Please provide your Indications for Use below. | | ? |
| The Cure Aqua Ready to Use Hydrophilic Intermittent Catheter is an intermittent urinary catheter that is inserted through the urethra and indicated for the purpose of bladder drainage. | | |
| Please select the types of uses. | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☑ Infants (29 days old to < 2 years old) ☑ Children (2 years old to < 12 years old) ☑ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
convatec
— forever caring —
K260737
Page 1 of 5
# 510(k) Summary
Date Prepared: 4-August-2026
Submitter:
Convatec,Ltd
Deeside Industrial Park
Deeside
Flintshire CH5 2NU
510K number:
K260737
Contact Person:
Kim Jones, +44 7590333564, kim.jones@convatec.com
Name of Device:
Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Female (CH06 - CH16);
Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Straight (CH06 - CH18);
Cure Aqua™ Ready to Use Hydrophilic Intermittent Catheter Male Coude (CH10 - CH18)
Classification Name:
21 CFR 876.5130 Urological Catheter and Accessories
Product Code:
EZD
Classification:
Class II
Predicate Device:
Cure Hydrophilic Catheter (K132500)
The predicate device, K132500, has not been subject to a design-related recall.
Device Description:
Cure Aqua™ Ready-to-Use Hydrophilic Intermittent Catheter (Cure Aqua) is a sterile, single use, disposable, intermittent urethral catheter. A urethral catheter is a tubular device that is introduced through the urethra into the bladder, providing a pathway for drainage. The catheter is made from thermoplastic polyurethane (TPU) with a hydrophilic coating. The coated catheter is stored in a wetting solution of distilled water, Polyvinylpyrrolidone K30 (PVP) and NaCl within its primary pack, making it ready-to-use. The device includes a low-density polyethylene (LDPE) sleeve and TPU molded funnel with different colorants according to the size designation. The catheter shaft is bonded to the funnel by an adhesive mixture of cyclohexanone and 2-Butanone.
Cure Aqua catheters come with either a straight or coudé tip option. Coudé tips are available for male catheters, and the bent tip allows the catheter to navigate the urethral structures and access the bladder. The coudé tip catheter has a blue stripe along the top portion of the entire length of the tubing for the purpose of providing visual confirmation of the coudé tip's orientation as the catheter is used.
The Cure Aqua catheter also has a color-coded funnel to indicate the French size (CH), a universal system for measuring the catheter's outer diameter. The color-coding helps users and healthcare professionals identify the correct size catheter.
{5}
[Non-Text]
convatec
- forever caring -
K260737
Page 2 of 5
Catheters are individually packed into a sealed, composite film, primary pouch along with a handling sleeve. The handling sleeve is provided to minimize touching the catheter shaft directly during insertion and retraction. The individual, packaged catheters are then placed in secondary and tertiary packaging and sterilized by E-beam irradiation.
The Cure Aqua is for both males and females. The selection of the French size (CH) size of the catheter is to be determined by healthcare professional and is based on the size of the patient. Cure Aqua is available by prescription use only.
The devices are surface devices with mucosal membrane contact with prolonged contact duration (>24 hours to 30 days) due to cumulative use.
The devices are intended to be used in both professional healthcare facilities and the home environment.
Table 1: Catheter configurations by size and tip type
| Patient Sex | Tip | Size (CH) | Effective Length (mm) |
| --- | --- | --- | --- |
| Male | Straight | 6 – 18 | 365 |
| Male | Coudé | 10 – 18 | 365 |
| Female | Straight | 6 – 16 | 155 |
### Intended Use:
To provide an intermittent pathway for drainage of fluids from the bladder. The catheter is inserted through the urethra.
Indications for Use:
| Subject Device | Predicate Device |
| --- | --- |
| Cure AquaTM Ready to Use Hydrophilic Intermittent Catheter (K260737) | Cure Catheter Hydrophilic Coated, (K132500) |
| The Cure Aqua Ready to Use Hydrophilic Intermittent Catheter is an intermittent urinary catheter that is inserted through the urethra and indicated for the purpose of bladder drainage. | The Cure CatheterTM Hydrophilic Coated is an intermittent urinary catheter that is inserted through the urethra and indicated for the purpose of bladder drainage. The urinary catheter comes in a variety of sizes, packaged sterile for single-use. |
### Intended Users:
Males, Females, Pediatric (Infant, Child, Adolescent, Transitional Adolescent) and Adult who need assistance with bladder drainage due to conditions causing dysfunction of the urinary system. Additional users are healthcare professionals including nurses, and/or caregivers, and family members who help patients with catheterization. Target users will be those who are able to self-catheterise (intermittent self-catheterisation or ISC) and have no or minimal dexterity impairment. Cure Aqua is a prescribed medical device. During consultation with a Health Care Provider, the user will be assessed for a suitable and appropriate level of dexterity, cognitive understanding and anatomical size to perform ISC with Cure Aqua.
### Technological Comparison:
The subject device, Cure Aqua, and the predicate device, Cure Hydrophilic Catheter (K132500), are hydrophilic intermittent urinary catheters and share the same intended technological principles. These devices provide a pathway for the intermittent drainage of urine from the bladder. The catheter is inserted through the urethra.
{6}
[Non-Text]
convatec
- forever caring -
K260737
Page 3 of 5
| Description | Subject Device | Primary Predicate Device | Comparison |
| --- | --- | --- | --- |
| Name | Cure AquaTM Ready to Use Hydrophilic Intermittent Catheter | Cure Catheter Hydrophilic Coated, K132500 | N/A |
| Manufacturer | Convatec | Convatec | Same |
| Product Code | EZD (catheter, straight) | EZD (catheter, straight) | Same |
| Classification Regulation | 876.5130 – Urological catheter and accessories | 876.5130 – Urological catheter and accessories | Same |
| Classification | II | II | Same |
| Use Environment | Home and healthcare (hospital/clinical) settings | Home and healthcare (hospital/clinical) settings | Same |
| Device Uses | Single patient, single use | Single patient, single use | Same |
| Size RangeEffective length and outer diameter (OD) | Female straight tipEffective Length: 155 mmOD (CH): 6, 8, 10, 12, 14, 16Male straight tipEffective Length 365 mmOD (CH): 6, 8, 10, 12, 14, 16, 18Male coudé tipEffective Length 365 mmOD (CH): 10, 12, 14, 16, 18 | Female straight tipLength (nominal): 6 in (152.5 mm)OD (CH): 6, 8, 10, 12, 14, 16, 18Male straight & coude tip: Length (nominal): 16 in (390 mm)OD (CH): 8, 10, 12, 14, 16, 18Pediatric straight tip: Length (nominal): 10 in (250 mm)OD (CH): 6, 8, 10, 12, 14. | Change in catheter size.The predicate device includes a larger outer diameter size for the female straight tip configuration (18Fr).The subject device includes a smaller outer diameter for the male straight tip (6Fr). |
| Tube Material | TPU and hydrophilic coating.Coudé tipped tubing contains a blue stripe comprised of TPU and <1% of colorant. | Polyvinyl chloride (PVC) and hydrophilic coating.Coudé tipped tubing contains a blue stripe comprised of PVC and <1% of colorant. | Change in tube material. |
| Catheter Coating and Wetting solution | Catheter coating: Hydrophilic polymer made of PU and polyacrylamide.Lubricating wetting solution: sterile water, PVP, and NaCl. | Catheter coating: Hydrophilic polymer made of PVP.Lubricating wetting solution: sterile water. | Change in catheter coating and wetting solution. |
| Catheter Funnel Material | TPU and colorantsCH06: Light GreenCH08: BlueCH10: BlackCH12: WhiteCH14: GreenCH16: OrangeCH18: Red | PVC and colorantsCH06: Light GreenCH08: BlueCH10: BlackCH12: WhiteCH14: GreenCH16: OrangeCH18: Red | Change in catheter funnel material. |
| Bonding adhesive | Cyclohexanol and butanone. | Cyclohexanol | Change in bonding adhesive |
{7}
[LOGO]
convatec
— forever caring —
K260737
Page 4 of 5
| Description | Subject Device | Primary Predicate Device | Comparison |
| --- | --- | --- | --- |
| **Catheter Tube Outer Diameter (mm)** | CH6: 2.0 mm CH8: 2.7 mm CH10: 3.3 mm CH12: 4.0 mm CH14: 4.7 mm CH16: 5.3 mm CH18: 6.0 mm. | CH6: 2.3 mm CH8: 3.0 mm CH10: 3.6 mm CH12: 4.3 mm CH14: 5.0 mm CH16: 5.6 mm CH18: 6.0 mm. | Change in catheter size. The predicate device includes a larger outer diameter size for the female straight tip configuration (18Fr). The subject device includes a smaller outer diameter for the male straight tip (6Fr). |
| **Catheter Tensile Strength** | Minimum tensile strength: CH06: ≥10N CH08: ≥10N CH10: ≥10N CH12: ≥10N CH14: ≥20N CH16: ≥20N CH18: ≥20N | Minimum tensile strength: CH06 – CH08: ≥1.0kg (9.8N) CH10 – CH14: ≥1.5kg (14.7N) CH16 – CH18: ≥2.0kg (19.6N) | Change in catheter tensile strength. |
| **No Touch Functionality** | Handling sleeve composed of LDPE. | Handling sleeve composed of LDPE. | Same |
| **Primary Packaging** | Bottom and top films: Composite film made of polyamide, aluminum, and PE. | Bottom film: Polypropylene and PE. Top film: Dialysis medical grade paper. | Change in primary catheter packaging. The subject device has different packaging to enable the device to be sterilized by e-beam rather than Ethylene Oxide. |
| **Sterilization** | E-beam SAL: 10^{-6} | Ethylene Oxide (EO) SAL: 10^{-6} | Change in sterilization method |
| **Shelf Life** | 3 years | 3 years | Same |
| **Biocompatibility** | Contact classification according to ISO 10993-1:2018: • Surface medical device, with mucosal membrane contact • Prolonged duration (>24 hours - ≤30 days) due to potential repeated cumulative contact. | Contact classification according to ISO 10993-1:2018: • Surface medical device, with mucosal membrane contact • Prolonged duration (>24 hours - ≤30 days) due to potential repeated cumulative contact. | Same |
The differences in technological characteristics do not raise different questions of safety and effectiveness.
### Non-Clinical Performance Data:
Non-clinical performance testing for Cure Aqua was conducted per applicable sections of voluntary and FDA consensus standards.
{8}
convatec
— forever caring —
K260737
Page 5 of 5
Qualification of biological safety assessment according to ISO 10993-1:2018, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process and FDA Guidance “Use of International Standard ISO 10993-1”. The following biological endpoints were addressed:
- Cytotoxicity was conducted according to ISO 10993-5:2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.
- Sensitization was conducted according to ISO 10993-10: 2021, Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- Irritation was conducted according to ISO 10993-23:2021, Biological evaluation of medical devices - Part 23: Tests for irritation.
- Acute systemic toxicity was conducted according to ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
- Subacute/subchronic toxicity was conducted according to ISO 10993-11:2017, Biological evaluation of medical devices - Part 11: Tests for systemic toxicity.
Sterilization and packaging validation met the requirements of the following FDA recognized standards:
- Sterilization validation was performed by conducting dose setting per ISO 11137-1:2015 A1: 2023.
- Sterile packaging in accordance with ISO-11607-1: 2020+A1:2023 and 11607-2: 2020+A1:2023.
- Accelerated Real Time aged shelf-life testing according to ASTM F1980-21.
- Performance testing of Shipper Containers and Systems according to ASTM D4169-22.
- Package integrity testing according to ASTM F88/F88M.
The physical performance properties of the Cure Aqua Catheter met the following standards:
- Dimensional analysis was conducted in accordance with ISO 20696:2018.
- Tensile testing performed in accordance with ISO 20696:2018.
- Flow Rate testing performed in accordance with ISO 20696:2018.
- Kink Stability testing performed in accordance with ISO 20696:2018.
- Coefficient of friction according to ASTM D1894:24.
All tests met the predetermined acceptance criteria. The results of the testing show substantial equivalence in performance of the subject device when compared to the predicate device (K132500).
# Conclusion:
The data and information provided in this submission support the conclusion that the Cure Aqua device (subject device) is substantially equivalent to its predicate device, Cure Hydrophilic Catheter (K132500) and the proposed subject device is as safe and effective as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.