K260726 · Institut Straumann AG · KCT · Jun 3, 2026 · General Hospital
Device Facts
Record ID
K260726
Device Name
Straumann® ProClean™ Cassette (041.800 ; 041.801)
Applicant
Institut Straumann AG
Product Code
KCT · General Hospital
Decision Date
Jun 3, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The ProClean™ Cassette is indicated to store and organize Straumann® surgical instruments required during dental implant treatment of the Straumann iExcel™ and Straumann iGuide™ surgical procedures. The ProClean™ Cassette is indicated to hold and organize the Straumann® surgical instruments during automated cleaning/thermal disinfection, and sterilization in healthcare facilities by healthcare professionals. The ProClean™ Cassette is only indicated for automated cleaning and thermal disinfection in a washer-disinfector. The ProClean™ Cassette is not indicated to maintain sterility and must be used in conjunction with a legally marketed, validated sterilization pouch or a sterilization container. The Straumann ProClean™ Cassette has been validated for the following maximum load (including instruments and cassette): Max. 1223 grams. Recommended sterilization time (exposure time at the sterilization temperature) and drying time: Method: Moist Heat (Autoclave) Fractionated vacuum; Conditions: 132 °C (270 °F) for 4 min; Drying time: 30 min
Device Story
Reusable stainless steel cassette system; stores/organizes surgical/prosthetic instruments for Straumann dental implant procedures; facilitates automated cleaning, thermal disinfection, and sterilization. Device features color-coded grommets and instrument pictograms for positioning. Used in healthcare facilities by healthcare professionals. Requires use with external FDA-cleared sterilization pouch or container to maintain sterility post-process. Validated for moist heat (autoclave) sterilization (132°C, 4 min) with 30-minute drying time. Maximum load capacity 1223g. Benefits include organized instrument management and standardized reprocessing workflows.
Clinical Evidence
No clinical data. Bench testing only. Validated per FDA guidance 'Reprocessing Medical Devices in Health Care Settings' (2015). Cleaning validated per ISO 17664:2021 and AAMI ST98:2022. Sterilization validated per ISO 17665:2024 to SAL 10⁻⁶. Biocompatibility evaluated per ISO 10993-1:2021; cytotoxicity testing (ISO 10993-5) showed no cytotoxic response after 100x reprocessing cycles.
Technological Characteristics
Materials: Stainless steel (1.4301, 1.4305, 1.4310) and silicone grommets. Perforated design for steam penetration. Reusable. Sterilization: Moist heat (fractionated vacuum autoclave, 132°C/4 min). Dimensions/Weight: Max load 1223g. No software or electronics.
Indications for Use
Indicated for healthcare professionals in healthcare facilities to store, organize, and facilitate automated cleaning, thermal disinfection, and sterilization of Straumann® surgical instruments used in iExcel™ and iGuide™ dental implant procedures.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Nobel Biocare N1 PureSet Tray, Nobel Biocare N1 PureSet Plate, Prosthetic PureSet Tray, Prosthetic PureSet Plate (K191475)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 3, 2026
Institut Straumann AG
% Jennifer Jackson
Senior Director of Regulatory Affairs and Quality
Straumann USA, LLC
60 Minuteman Rd.
Andover, Massachusetts 01810
Re: K260726
Trade/Device Name: Straumann® ProClean™ Cassette (041.800 ; 041.801)
Regulation Number: 21 CFR 880.6850
Regulation Name: Sterilization Wrap
Regulatory Class: Class II
Product Code: KCT
Dated: March 5, 2026
Received: March 5, 2026
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260726 - Jennifer Jackson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260726 - Jennifer Jackson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**STEPHEN A. ANISKO -S** Digitally signed by
STEPHEN A. ANISKO -S
Date: 2026.06.03
15:20:53 -04'00'
Stephen Anisko
Acting Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260726
Device Name
Straumann® ProClean™ Cassette (041.800 ; 041.801)
Indications for Use (Describe)
The ProClean™ Cassette is indicated to store and organize Straumann® surgical instruments required during dental implant treatment of the Straumann iExcel™ and Straumann iGuide™ surgical procedures.
The ProClean™ Cassette is indicated to hold and organize the Straumann® surgical instruments during automated cleaning/thermal disinfection, and sterilization in healthcare facilities by healthcare professionals.
The ProClean™ Cassette is only indicated for automated cleaning and thermal disinfection in a washer-disinfector. The ProClean™ Cassette is not indicated to maintain sterility and must be used in conjunction with a legally marketed, validated sterilization pouch or a sterilization container.
The Straumann ProClean™ Cassette has been validated for the following maximum load (including instruments and cassette): Max. 1223 grams.
Recommended sterilization time (exposure time at the sterilization temperature) and drying time:
| Method | Conditions | Drying time |
| --- | --- | --- |
| Moist Heat (Autoclave) Fractionated vacuum | 132 °C (270 °F) for 4 min | 30 min |
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{4}
# Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
## 510(k) Summary
### Submitter's Contact Information
Submitter: Straumann USA, LLC
60 Minuteman Road
Andover, MA 01810
Registration No.: 1222315 Owner/Operator No.: 9005052
On behalf of:
Institut Straumann AG
Peter Merian weg 12
4052 Basel, Switzerland
Contact Person: Jennifer M. Jackson, MS
Sr. Director of Regulatory Affairs & Quality
Phone Number: +1-978-747-2509
Fax Number: +1-978-747-0023
Prepared By: Mohammed Ghaleb
International Regulatory Affairs Senior Manager
Institut Straumann AG
Phone number: +41619651667
Date Prepared: June 3, 2026
### Name of the Device
Trade Names: Straumann® ProClean™ Cassette (041.800 ; 041.801)
Common Name: Sterilization wrap containers, trays, cassettes & other accessories
Classification Name: Sterilization wrap
Regulation Number: 880.6850
Device Classification: II
Product Code: KCT
Institut Straumann AG
Page 1 of 7
{5}
# Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
## Predicate Device(s)
- K191475 - Nobel Biocare N1 PureSet Tray, Nobel Biocare N1 PureSet Plate, Prosthetic PureSet Tray, Prosthetic PureSet Plate
## Reference Device:
- K191522 - Straumann Modular Cassette
## Device Description:
The Straumann ProClean™ Cassette is designed as a reusable wash and sterilization cassette system and for the storage of surgical and prosthetic instruments of the Straumann® Dental Implant System. In addition to its core functionality, the trays of the Straumann ProClean™ Cassette features distinct color-coded grommets and instrument pictograms to facilitate the surgical procedure and the correct use and positioning of the instruments.
## Indications for Use:
The ProClean™ Cassette is indicated to store and organize Straumann® surgical instruments require during dental implant treatment of the Straumann iExcel™ and Straumann iGuide™ surgical procedures.
The ProClean™ Cassette is indicated to hold and organize the Straumann® surgical instruments during automated cleaning/thermal disinfection, and sterilization in healthcare facilities by healthcare professionals.
The ProClean™ Cassette is only indicated for automated cleaning and thermal disinfection in a washerdisinfector. The ProClean™ Cassette is not indicated to maintain sterility and must be used in conjunction with a legally marketed, validated sterilization pouch or a sterilization container.
The Straumann ProClean™ Cassette has been validated for the following maximum load (including instruments and cassette): Max. 1223 grams.
Recommended sterilization time (exposure time at the sterilization temperature) and drying time:
| Method | Conditions | Drying time |
| --- | --- | --- |
| Moist Heat (Autoclave) Fractionated vacuum | 132 °C (270 °F) for 4 min | 30 min |
Institut Straumann AG
Page 2 of 7
{6}
## Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
### Technological Characteristics
The technological characteristics of the subject devices are compared to the predicate device in the following table:
| FEATURE | Subject Device K260726 | Predicate Device K191475 | Reference Device K191522 | COMPARISON DISCUSSION |
| --- | --- | --- | --- | --- |
| **Indications for Use** | The ProClean™ Cassette is indicated to store and organize Straumann® surgical instruments require during dental implant treatment of the Straumann iExcel™ and Straumann iGuide™ surgical procedures. | The Nobel Biocare PureSet Trays are used in healthcare facilities to store and organize Nobel Biocare surgical/prosthetic instruments and components during cleaning/sterilization and during implant/prosthetic treatment. The Nobel Biocare PureSet Trays are not intended on their own to maintain sterility; they are intended to be used in conjunction with a legally marketed, validated, FDA-cleared sterilization pouch or sterilization wrap. | Straumann® Modular Cassette is used in healthcare facilities to organize, enclose, sterilize, transport and store Straumann instruments between surgical uses. | Identical |
| | The ProClean™ Cassette is indicated to hold and organize the Straumann® surgical instruments during automated cleaning/thermal disinfection, and sterilization in healthcare facilities by healthcare professionals. | Sterilization validations for the worst- case PureSet Tray included surgical/prosthetic instruments such as torque wrenches, implant drivers, direction indicators, drills, screw taps, screw driver and irrigation needles. The PureSet Trays were validated for a | Straumann® Modular Cassette is not intended to maintain sterility on its own, but is intended to be used in conjunction with a legally marketed, validated sterilization double pouch to maintain the sterility of the enclosed devices. | |
| | The ProClean™ Cassette is only indicated for automated cleaning and thermal disinfection in a washerdisinfector. | | The Straumann® Modular Cassette has been validated for the following maximum loads: | |
| | The ProClean™ Cassette is not indicated to maintain sterility and must be used in conjunction with a legally | | - Module A 400g - BCC Maximum permissible stack BCC 611g The A module is intended to be sterilized individually, without stacking with other modules. | |
Institut Straumann AG
Page 3 of 7
{7}
## Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
| | marketed, validated sterilization pouch or a sterilization container. ProClean™ Cassette sterilization parameters: The Straumann ProClean™ Cassette has been validated for the following maximum load (including instruments and cassette): Max. 1223 grams. Recommended sterilization time (exposure time at the sterilization temperature) and drying time: For the United States and Canada: Moist Heat (Autoclave) Fractionated vacuum 132°C (270°F) for 4 min 30 min drying time | maximum load of 1635 grams (Trefoil PureSet Tray), 1082 grams (NobelActive/ NobelPar allel CC PureSet Tray), 945 grams (NobelReplace CC PureSet Tray), 454 grams (Nobel Biocare N1™ PureSet) and 486 grams (Prosthetic PureSet). | The B and C module are intended to be sterilized individually, or by stacking the B module on top of C module bases. The maximum permissible stack for sterilization is one B module on top of two C module bases. The B module lid and C module lid could be used to enclose an ultrasonic mat (Art. No. 041.774) for ultrasonic bath cleaning used instruments. Only use the following sterilization parameters: • Fractionated vacuum: 132°C (270 °F) for 4 minutes with 30 minutes drying time | |
| --- | --- | --- | --- | --- |
| **Product Code** | KCT | KCT | KCT | Identical |
| **Design** | Stainless Steel cassette, Silicone grommets | Single level tray with grommets and a basket for holding tooling in specific locations and covering lid (without handle). | Plastic modules, trays and lids | Same material |
| **Materials** | ProClean Cassette: - Stainless steel (1.4301, 1.4305, 1.4310) - Silicone (grommets) | PureSet Tray - Stainless steel construction (1.4301,1.4303, 1.4310, 1.4024) | Polyphenylsulfone Silicone Stainless steel (lid hinge) | Identical |
Institut Straumann AG
Page 4 of 7
{8}
## Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
| | | - PEEK grommets and storage sleeves - PEEK mini tray locks - Silicone elastomer feet PureSet Tray Plate Aluminum construction with anodization | | |
| --- | --- | --- | --- | --- |
| **Materials compatible with sterilization method** | Yes | Yes | Yes | Identical |
| **Perforated** | Yes; allows moist heat (steam) penetration to achieve sterilization | Yes; allows moist heat (steam) penetration to achieve sterilization | Yes; allows moist heat (steam) penetration to achieve sterilization | Identical |
| **Reusable** | Yes | Yes | Yes | Identical |
| **Sterilization method** | Moist heat (steam) | Moist heat (steam) | Moist heat (steam) | Identical |
| **Cycles and parameters** | Fractionated vacuum 132°C (270 °F) for 4 minutes with 30 minutes drying time | Fractionated vacuum 132°C (270 °F) for 4 minutes with 20 minutes drying time Gravity Displacement 132°C (270 °F) for 15 minutes 30 minutes drying time | Fractionated vacuum 132°C (270 °F) for 4 minutes with 30 minutes drying time | Identical drying time identical to Reference Device |
| **Sterile barrier** | FDA cleared sterilization pouch or metal sterilization container | FDA cleared sterilization wrap/pouch | FDA cleared sterilization wrap/pouch | Equivalent; A validation was performed with both sterile barriers. |
Institut Straumann AG
Page 5 of 7
{9}
# Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
The subject and predicate device share the following characteristics:
- indications for use
- design
- identical sterilization method and similar sterilization parameters
- reusable
The subject device is technologically different from the predicate device as follows:
- drying time
- weight
- Size
Nevertheless, the related performance is proven through testing.
## Summary of Non-Clinical Testing
The performance during multiple reprocessing steps for the ProClean Cassette, as recommended in the labeling, was validated according to applicable recommendations in the FDA guidance document “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling, issued on March 17, 2015”.
The test results demonstrate that the ProClean Cassette is similar to the predicate device.
### Cleaning and Sterilization Validation
The subject device, i.e. Straumann® ProClean Cassettes system, is intended to store and organize Institut Straumann AG surgical instruments and components during cleaning, disinfection, sterilization, and implant treatment. The cassette system is a multiple-use device and must be cleaned, disinfected, and sterilized prior to use at the patient. The sterilization procedures reflect the FDA guidance document, “Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling Guidance for Industry and Food and Drug Administration Staff, Document issued on: March 17, 2015”
Cleaning validation was done under worst case conditions in accordance with ISO 17664:2021 and AAMI ST98:2022.
The sterilization parameters have been validated in accordance with ISO 17665:2024, to a sterility assurance level (SAL) of 10⁻⁶ using the biological indicator (BI) overkill method. In addition to the SAL validation, dry time was validated using full cycle parameters.
### Biocompatibility
The biological assessment of the Straumann® ProClean Cassettes was performed according to ISO 10993-1:2021 “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process” and the biocompatibility evaluation flow chart according to the FDA Guidance document “Use of International Standard ISO 10993-1:2020, ‘Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process’, Guidance for Industry and Food and Drug Administration Staff.
Institut Straumann AG
Page 6 of 7
{10}
# Traditional 510(k) Submission
Straumann ProClean™
K260726 510(k) Summary
The results show that none of the test items after 1x, 50x and 100x reprocessing elicited a cytotoxic response. All results are above the limit of 70% viability of cells (see ISO 10993-5 and Table 1). This means that the test items do not release substances in cytotoxic concentrations. In this study under the given conditions no leachable substances were released in cytotoxic concentrations from the test item.
Therefore, the Straumann ProClean Cassette is considered biologically safe for its intended use according to ISO 10993-1 and FDA's guidance document on ISO 10993-1.
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the subject device is as safe, as effective, and performs as well as the legally marketed predicate device (K191475).
Institut Straumann AG
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.