← Product Code [DXN](/productcode/DXN) · K260670

# YUWELL® Electronic Blood Pressure Monitor (K260670)

_Jiangsu Yuyue Medical Equipment & Supply Co., Ltd. · DXN · Oct 28, 2026 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K260670

## Device Facts

- **Applicant:** Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Oct 28, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This device is intended to measure blood pressure and pulse rate in patients over 12 years old with a wrist circumference ranging from 4.92 in to 8.46 in (12.5 cm to 21.5 cm) in households or medical centers. It is not suitable for neonates, pregnant individuals, or patients with pre-eclampsia.

## Device Story

Battery-powered, automatic, noninvasive wrist blood pressure monitor; uses oscillometric method. Device inflates cuff via piezoelectric pump; monitors cuff pressure; captures pulse waveform during inflation. Software analyzes waveform to determine systolic/diastolic blood pressure and pulse rate. Used in households or medical centers by lay persons or clinicians. Features include irregular heartbeat (IHB) detection, movement error indication, cuff wrapping detection, and optional Bluetooth/voice broadcast. Output displayed on LCD; provides BP classification per 2025 AHA/ACC guidelines. IHB feature acts as signal quality indicator; not for arrhythmia diagnosis. Benefits include convenient, noninvasive monitoring of BP and pulse rate for hypertension management.

## Clinical Evidence

Clinical study per ISO 81060-2:2018+AMD1:2020 (n=85) validated BP accuracy; mean difference ± SD was 1.1 ± 6.7/0.4 ± 7.2 mmHg (systolic/diastolic). Pulse rate accuracy validated against ECG reference (n=35, 105 datasets); mean error -0.38 bpm, SD 1.76 bpm, RMSE 1.79 bpm. All results met pre-defined acceptance criteria.

## Technological Characteristics

Oscillometric measurement; semiconductor pressure sensor; piezoelectric pump; LCD display; 2 AAA batteries; IP22 rated; Type BF applied part. Bluetooth (optional). Complies with ISO 81060-2, IEC 80601-2-30, IEC 60601-1, IEC 60601-1-2, ISO 10993.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Wrist Blood Pressure Monitor Model BP4350 ([K182166](/device/K182166.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

FDA U.S. FOOD & DRUG
ADMINISTRATION

October 28, 2026

Jiangsu Yuyue Medical Equipment & Supply Co., Ltd.

Zhang Fang (Fawn)

Director of Regulatory and Compliance

Jiangsu Yuyue Medical Equipment& Supply Co., Ltd.

#1 Baisheng Rd. Development Zone, Danyang, Jiangsu 212300 China

Danyang, Jiangsu 212300

China

Re: K260670

Trade/Device Name: YUWELL® Electronic Blood Pressure Monitor

Regulation Number: 21 CFR 870.1130

Regulation Name: Noninvasive Blood Pressure Measurement System

Regulatory Class: Class II

Product Code: DXN

Dated: September 30, 2026

Received: September 30, 2026

Dear Zhang Fang (Fawn):

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food & Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993

www.fda.gov

{2}

K260670 - Zhang Fang (Fawn)

Page 2

Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the

{3}

K260670 - Zhang Fang (Fawn)

Page 3

Office of Industry and Consumer Education (OICE) to ask a question about a specific regulatory topic. See the OICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-office-industry-and-consumer-education-oice) for more information or contact OICE by email (AskOICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# STEPHEN C. BROWNING -S

CDR Stephen Browning

Assistant Director

Division of Cardiac Electrophysiology,

Diagnostics, and Monitoring Devices

Office of Cardiovascular Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{4}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K260670

Device Name
YUWELL® Electronic Blood Pressure Monitor

Indications for Use (Describe)

This device is intended to measure blood pressure and pulse rate in patients over 12 years old with a wrist circumference ranging from 4.92 in to 8.46 in (12.5 cm to 21.5 cm) in households or medical centers. It is not suitable for neonates, pregnant individuals, or patients with pre-eclampsia.

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)

☑ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{5}

# 510 (k) Summary

## Applicant/Sponsor Information

|  **Applicant** | : JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO., LTD.  |
| --- | --- |
|  **Address** | : No.1 Baisheng Road Development Zone, Danyang, Jiangsu 212300 CHINA  |
|  **Manufacturing Facility** | : No.1 Baisheng Road Development Zone, Danyang, Jiangsu 212300 CHINA  |
|  **Contact of Correspondent** | : Fang (Fawn) Zhang  |
|  **Phone** | : (206) 639-1311  |
|  **Email** | : fawn.zhang@yuwell.com  |

## Trade Name, Common Name, Classification

|  **Trade Name** | : YUWELL® Electronic Blood Pressure Monitor  |
| --- | --- |
|  **Common Name** | : YUWELL® Electronic Blood Pressure Monitor  |
|  **Classification Name** | : System, Measurement, Blood-pressure, Non-invasive  |
|  **Product Code** | : DXN  |
|  **Regulation Number** | : 21 CFR 870.1130  |
|  **Device Class** | : 2  |

## Identification of Predicate Device

The identification of predicates within this submission is as follows:

|  **Manufacturer** | : Omron Healthcare, Inc.  |
| --- | --- |
|  **Trade Name** | : Wrist Blood Pressure Monitor Model BP4350  |
|  **Common Name** | : Wrist Blood Pressure Monitor Model BP4350  |
|  **Classification Name** | : System, Measurement, Blood-pressure, Non-invasive  |
|  **Product Code** | : DXN  |
|  **Regulation Number** | : 21 CFR 870.1130  |
|  **Device Class** | : 2  |
|  **FDA 510 (k) #** | : K182166  |

## Description of the Device

YUWELL® Electronic Blood Pressure Monitor is a battery-powered, automatic, noninvasive, wrist blood pressure measuring system intended for over-the-counter (OTC) use. It is designed for wrist circumference ranging from 4.92 in to 8.46 in (12.5 cm to 21.5 cm). The systolic and diastolic blood pressures are measured using the oscillometric method, where the cuff is

K260670

Page 1 of 10

{6}

inflated with an air pump and it deflates via an exhaust valve. During inflation, the cuff pressure is monitored, and the pulse waveform is captured. The pulse waveform is then further analyzed by a software, which determines the pulse rate, as well as the systolic and diastolic blood pressures.

### **Proposed Indications for Use**

This device is intended to measure blood pressure and pulse rate in patients over 12 years old with a wrist circumference ranging from 4.92 in to 8.46 in (12.5 cm to 21.5 cm) in households or medical centers. It is not suitable for neonates, pregnant individuals, or patients with pre-eclampsia.

### **Technological Characteristics**

The YUWELL® Electronic Blood Pressure Monitors are intended for medical professionals and lay persons to use at healthcare facilities or at home. The device applies the oscillometric method for blood pressure and pulse rate measurements. The device has multiple auxiliary functions for accurate measurement and convenient operation, such as indication for movement error, indication for irregular heartbeat (IHB), detection for cuff wearing, recorded measurement results check, display unit setting, display time and date setting, voice broadcast (optional), Bluetooth transmission (optional), etc.

Here, irregular heartbeat (IHB) means more than two pulse waves are identified where the instantaneous heart rate exceeds the average heart rate by 25% during blood pressure measurement. When an IHB detected, it indicates that the measurement results may be inaccurate. The device's IHB feature serves only as a signal quality indicator for potentially inaccurate measurements and the device is not indicated for patients with Arrhythmias.

### **Substantial Equivalence**

Comparison of technological characteristics:

|  Elements of Comparison | Subject Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Manufacture | JIANGSU YUYUE MEDICAL EQUIPMENT & SUPPLY CO., LTD. | Omron Healthcare, Inc. | /  |
|  Device Name | YUWELL® Electronic Blood Pressure Monitor | Wrist Blood Pressure Monitor Model BP4350 | /  |
|  Model | YE8100B | BP4350 | /  |
|  510(k) | - | K182166 | /  |
|  Regulation | 21 CFR 870.1130 | 21 CFR 870.1130 | Same  |
|  Product Code | DXN | DXN | Same  |

K260670

Page 2 of 10

{7}

|  Elements of Comparison | Subject Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Classification | Class II | Class II | Same  |
|  Intended Use/Indication for Use | This device is intended to measure blood pressure and pulse rate in patients over 12 years old with a wrist circumference ranging from 4.92 in to 8.46 in (12.5 cm to 21.5 cm) in households or medical centers. It is not suitable for neonates, pregnant individuals, or patients with pre-eclampsia. | The device is a digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 5.3 inches to 8.5 inches (13.5 cm to 21.5 cm). The device detects the appearance of irregular heartbeats during measurement and gives a warning signal with readings. | Note No.1  |
|  Patient Population | Patients who are more than 12 years old. The device is not suitable for neonates, pregnant individuals, or patients with pre-eclampsia. | Adults | Note No.2  |
|  Environment of Use | Household or medical center | Home | Note No.3  |
|  **Performance**  |   |   |   |
|  Principle of Operation | Oscillometric method | Cuff oscillation method | Same  |
|  Measurement Site | Wrist | Wrist | Same  |
|  Measurement Range | • Pressure range: 0~300 mmHg • Pulse rate: 40~200 beats/min • Diastolic: 40-210 mmHg • Systolic: 60~260 mmHg | • Pressure range: 0~299 mmHg • Pulse rate: 40~180 beats/min • Diastolic: 40-215 mmHg • Systolic: 60~260 mmHg | Note No.4  |
|  Pressure Sensor | Semiconductor pressure | Semiconductor pressure | Same  |

K260670

Page 3 of 10

{8}

|  Elements of Comparison | Subject Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|   | sensor | sensor |   |
|  Applicable Cuff (Wrist Circumference) | 12.5~21.5 cm | 13.5~21.5 cm | Note No.5  |
|  Accuracy of Pressure Indicator | Within ±3 mmHg | Within ±3 mmHg | Same  |
|  Accuracy of Pulse Rate | Within ±5% of reading | Within ±5% of reading | Same  |
|  Inflation Method | Automatic inflation with piezoelectric pump | Automatic inflation with piezoelectric pump | Same  |
|  Deflation Method | Automatic rapid deflation value | Automatic rapid deflation value | Same  |
|  Display | LCD digital display | LCD digital display | Same  |
|  Power source | 2 “AAA” batteries | 2 “AAA” batteries | Same  |
|  Operating Conditions | •+5°C to +40°C •10% to 90% RH , non-condensing •70 kPa to 106 kPa | • +10°C ~ +40°C •15% ~ 90% RH (non-condensing) • 80 kPa ~ 106 kPa | Note No.6  |
|  Storage Conditions | •-25°C to +60°C •10% to 90% RH, non-condensing •50 kPa to 106 kPa | • -20°C ~ +60°C •10% ~ 90% RH (non-condensing) | Note No.7  |
|  Dimensions (mm) | Approx. 85mm x 58mm x 21mm | 91 mm x 63 mm x 13 mm (not including the wrist cuff) | Note No.8  |
|  Weight (g) | About 93g (Without batteries) | Approximately 91 g (not including batteries) | Note No.9  |
|  Operation Mode | Continuous operation | Continuous operation | Same  |
|  IP Classification | IP22 | IP22 | Same  |
|  Electric Classification | Internally powered, type BF applied part (cuff is applied part) | Internally powered ME equipment, Type BF (wrist cuff) | Same  |
|  Service Life | 5 years | 5 years | Same  |
|  **Functions**  |   |   |   |
|  Time and Date | Yes | Yes | Same  |

K260670

Page 4 of 10

{9}

|  Elements of Comparison | Subject Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Setting |  |  |   |
|  Voice Broadcast | Optional | No | Note No.10  |
|  Voice Volume Setting | Optional | No | Note No.11  |
|  Voice On/Off Key | Optional | No | Note No.12  |
|  Unit Setting | YES Note: Display measurement results by “mmHg” or “kPa”. | No Note: Display measurement results by “mmHg” only. | Note No.13  |
|  Movement Error Indication | Yes | Yes | Same  |
|  Cuff Wrapping Detection | Yes | Yes | Same  |
|  Indications for Irregular Heartbeats | Yes | Yes | Same  |
|  Memory Function | Yes | Yes | Same  |
|  Memory Capacity | Up to 299 sets of data | Up to 100 sets of data per user | Note No.14  |
|  Bluetooth Transmission | Optional | Yes | Note No.15  |
|  Indication for Low Battery | Yes | Yes | Same  |
|  Automatic Shutdown | Yes | Yes | Same  |
|  View Average Value | View average value of latest three times measurement | - Viewing the average of the latest 2 or 3 readings taken within the most recent 10 minutes timeframe - View the weekly average readings taken in the morning over the past 4 weeks | Note No.16  |

K260670

Page 5 of 10

{10}

|  Elements of Comparison | Subject Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  Blood Pressure Classification | The feature displays a user's blood pressure classification. “■” corresponds to the BP classification in 2025 AHA/ACC Hypertension Guideline. | Symbol(**High**): appears if the systolic blood pressure is 130 mmHg or above and/or if the diastolic blood pressure is 80 mmHg or above. Refer to classification of BP (ACC/AHA 2017 Hypertension guidelines) | Note No.17  |
|  **Safety and Performance**  |   |   |   |
|  Performance | Comply ANSI/AAMI/ISO 81060-2 and IEC 80601-2-30 | Comply ANSI/AAMI/ISO 81060-2 and IEC 80601-2-30 | Same  |
|  Electrical Safety | Comply IEC 60601-1, IEC 60601-1-6 and IEC 60601-1-11 | Comply IEC 60601-1, IEC 60601-1-6 and IEC 60601-1-11 | Same  |
|  EMC | Comply IEC 60601-1-2 | Comply IEC 60601-1-2 | Same  |
|  Biocompatibility | Comply to ISO 10993 and FDA Guidance (Tested items including Cytotoxicity, Sensitization and Intracutaneous Reactivity) | Comply to ISO 10993 and FDA Guidance (Tested items including Cytotoxicity, Sensitization and Intracutaneous Reactivity) | Same  |
|  FCC | Comply with Part 15 of federal communications commission (FCC) rules | Comply with Part 15 of federal communications commission (FCC) rules | Same  |

### **Justification for the Differences**

#### **Note No.1, Note No.2, Note 3 and Note No.5:**

1) According to clause 5.1.3 of ISO 81060-2:2018+AMD1:2020, the clinical investigation requirements for adults and adolescents over age 12 are essentially the same. Object evidence supports that the subject device has met the requirements in ISO 81060-2:2018+AMD1:2020 as that of the predicate device. Therefore, the subject and the predicate devices are substantially equivalent as both have met the requirements in ISO 81060-2:2018+AMD1:2020.

2) While the subject device is designed for household and medical center use and the predicate device is for home use, the difference does not affect the substantial equivalence of the devices

K260670

Page 6 of 10

{11}

because healthcare professionals at the medical center are trained in medical care and they are better educated to properly use the device compared to a lay person. Thus, the addition of medical center use does not introduce new risks and does not compromise the effectiveness of the subject device relative to the predicate device.

3) The intended wrist circumference of the subject device is longer than that of the predicate device; however, the subject device's wrist circumference complies with ISO 81060-2:2018+AMD1:2020, so the subject and the predicate devices are substantially equivalent despite of their difference in the intended patient population with different wrist circumference.

4) The subject device can detect irregular heartbeats during measurements and gives signal quality indicator to alert users to potentially inaccurate blood pressure measurements. Therefore, the subject and the predicate devices are substantially equivalent in this regard.

# **Note No.4:**

1) The subject device's cuff pressure range is 0 to 300 mmHg, whereas that of the predictive device is 0 to 299 mmHg. The slight difference does not present additional risks.

2) The subject device's pulse rate range is wider than that of the predicate device, but both contain the pulse range of the majority of the population. Therefore, the differences do not lead to additional risks.

3) The subject device's diastolic pressure range is marginally narrower than that of the predicate device, whereas the systolic pressure range is identical for both. The blood pressure measurement range of subject device complies with IEC 80601-2-30. Therefore, the differences do not lead to additional risks.

# **Note No.6 and Note No.7:**

The operating conditions and storage conditions of the subject device and the predicate device differ slightly. While the subject devices operating and storage environment is slightly broader than that of the predicate device, the subject device's operating and storage environment meets the requirements of IEC 60601-1, IEC 60601-1-11 and IEC 80601-2-30. So, the differences do not raise any concerns regarding substantial equivalence.

# **Note No.8 and Note No.9:**

The dimensions and weight of the subject and the predicate devices are quite similar. These parameters do not affect the substantial equivalence of the subject device. Thus, the subject device is substantially equivalent to the predicate device despite their differences in these parameters.

# **Note No.10, Note No.11 and Note No.12:**

The subject device provides an optional voice broadcast function, whereas the predicate device does not. As this function is independent of the blood pressure algorithm, it does not present

K260670

Page 7 of 10

{12}

additional risks and does not affect the substantial equivalence of the subject device to the predicate device. Thus, the subject device is substantially equivalent to the predicate device despite of this additional function.

# Note No.13:

The subject device can display measurement results by “mmHg” or “kPa”, while the predicate device can display by “mmHg” only. The additional display unit does not affect the substantial equivalence of the subject device. In addition, this feature is validated according to IEC 80601-2-30, so the subject device is as safe and effective as the predicate device despite of this additional feature.

# Note No.14:

The predicate device’s memory values are divided into User 1 and User 2, with each user able to store up to 100 sets, whereas the subject device does not distinguish between users and stores up to 299 sets. Although the memory capacity of the subject device differs slightly from that of the predicate device, this feature is for user convenience, and not the intended use of the device. In addition, both devices’ memory capacity is sufficient for users to view recent measurements, thus the difference in memory capacity between the subject device and the predicate device does not raise any concerns regarding substantial equivalence.

# Note No.15:

The predicate device has a Bluetooth feature whereas the subject device offers that as an option. Thus, the difference is some of the subject devices may not have the Bluetooth feature, which enables users to view measurements remotely on a mobile APP. Thus, the Bluetooth function is a nice-to-have feature and not an absolute necessity for the device to perform its intended use. This difference between the subject and the predicate device does not raise concerns regarding substantial equivalence.

# Note No.16:

The predicate device offers more average readings options than the subject device. This is a nice-to-have feature and does not affect the substantial equivalence of the intended use of these devices.

# Note No.17:

The subject device displays a user’s BP classification per 2025 AHA/ACC Hypertension Guideline. In comparison, the predicate device does not provide BP classification, but alerts a user when systolic blood pressure is/exceeds 130 mmHg and/or diastolic blood pressure is/exceeds 80 mmHg (high BP per 2017 AHA/ACC Hypertension Guideline). Blood pressure classifications are the same for 2017 and 2025 AHA/ACC Hypertension Guidelines. As clearly stated in the User’s Manual, the BP classification displayed on this device is not for clinical diagnosis, and the alert is for the patient to repeat testing and/or seek doctor’s attention. Thus,

K260670

Page 8 of 10

{13}

this difference between the subject and predicate devices does not raise new concerns regarding substantial equivalence.

# **Summary of Testing**

# **1) Nonclinical Testing Summary**

The design and manufacturing of YUWELL® Electronic Blood Pressure Monitor are subject to verification and validation testing in conformance with regulatory guidance and recognized consensus standards.

- Electrical safety test according to ANSI/AAMI/IEC 60601-1, IEC 60601-1-8, IEC 60601-1-11 and IEC 80601-2-30 standards
- Electromagnetic compatibility test according to IEC 60601-1-2 and IEC TS 60601-4-2 standards
- Usability test according to IEC 60601-1-6 and IEC 62366-1 standards
- Performance test according to IEC 80601-2-30 standard
- The irregular heartbeat (IHB) feature of the subject device serves as a signal quality indicator to alert users to potentially inaccurate blood pressure measurements. This feature has been validated through testing using a blood pressure simulator. To set up the test, the blood pressure simulator was set to the arrhythmia mode. The arrhythmia mode mentioned herein refers solely to the preset mechanical simulation waveforms of the blood pressure simulator, which are used to test the IHB signal quality indication function. First, the simulator's output waveforms were verified to meet the definition of an irregular pulse wave and the preset values were recorded. Second, the subject device's readings were compared against the simulator's preset values. Third, ten sets of measurements were compiled in each mode, and the mean absolute deviation between the subject device readings and the simulator's preset values was calculated. The mean absolute deviation for both systolic and diastolic pressures exceeded \(5\mathrm{mmHg}\), it demonstrated that the blood pressure measurement was inaccurate when the IHB symbol was present.
Biocompatibility test according to ISO 10993 and the FDA guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"
- Software verification and validation per FDA guidance "Content of Premarket Submissions for Device Software Functions" and "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" and IEC 62304
- Bluetooth transmission complies with Part 15 of federal communications commission (FCC) rules

# **2) Clinical Testing summary**

i) Blood Pressure Clinical Testing

K260670

Page 9 of 10

{14}

The accuracy of YUWELL® Electronic Blood Pressure Monitor has been demonstrated through a clinical study per ISO 81060-2:2018+AMD1:2020 in the intended patient population, which is individuals over 12 years old. The study included Eighty-five (85) subjects, who met the study inclusion criteria for wrist circumference and systolic and diastolic BP. Validation and data analysis were performed per protocol. In the ISO 81060-2:2018+AMD1:2020 validation procedure (criterion 1), the mean ± SD of the differences between the test device and the reference BP was 1.1 ± 6.7/0.4 ± 7.2 mmHg (systolic/diastolic). The mean differences between the two observers and the test device readings were 1.1 ± 5.8 mmHg for systolic BP and 0.4 ± 6.6 mmHg for diastolic BP (criterion 2). The clinical study results have met both ISO criteria, and thus validated the accuracy of YUWELL® Electronic Blood Pressure Monitor.

# ii) Pulse Rate Clinical Testing

A clinical validation study was conducted to evaluate the pulse rate (PR) measurement accuracy of the subject device by comparison with manually annotated electrocardiogram (ECG) recordings. The study had a total of thirty-five (35) subjects enrolled, and all their age were above 12 years old. Their validated pulse rate ranged from 42 to 183 bpm with no subjects excluded from the analysis. The study method is as such: Following a 5-minute seated rest, the subject's pulse rate was measured simultaneously by the subject device and a reference ECG. There were three (3) repeated measurements per subject, yielding 35x3=105 data sets that resulted in a mean error of -0.38 bpm, a standard deviation of 1.76 bpm, and an overall root mean square error of 1.79 bpm. Bland-Altman analysis revealed all the observed differences were within the pre-defined acceptance criterion of ±5%. These results demonstrated good agreement between the device PR outputs and the ECG reference values across the validated range.

# Conclusion

Compared to the predicate device, the subject device has similar intended use and technical characteristics as the predicate device. All testing results have demonstrated that the subject device is as safe and effective as the predicate device. Therefore, they are substantially equivalent.

K260670

Page 10 of 10

---

**Source:** [https://fda.innolitics.com/device/K260670](https://fda.innolitics.com/device/K260670)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).