Midas Restore is a light-curable polymerizable resin intended to be used for the fabrication of individual and fixed definitive full single crowns; definitive partial crowns (e.g., inlays, onlays) in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material.
Device Story
Midas Restore is a photo-polymer methacrylate resin for 3D printing dental prosthetics. Input: 3D scan data of patient oral cavity. Operation: Resin is 3D printed layer-by-layer using a compatible printer, then light-cured to form the final appliance. Used in dental offices/labs by clinicians. Output: Definitive dental prosthetics (crowns, veneers, artificial teeth). Benefits: Provides an alternative to traditional restorative materials with verified flexibility, sorption, and biocompatibility for long-term oral use.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, -3, -5, -10, -11, and -23. Bench testing performed per ISO 10477 and ISO 4049, including flexural strength, water sorption/solubility, radio opacity, print accuracy, surface finish, shear bond strength, free monomer extraction, and depth of cure. All tests passed.
Technological Characteristics
Photo-polymer methacrylate resin; light-based curing; 3D printing fabrication. Complies with ISO 10477 (flexural strength, sorption, solubility, surface finish, shear bond strength), ISO 4049 (radio opacity), and ISO 20795-1 (monomer extraction). Non-sterile delivery. Compatible with VITA classical A1-D4 shade guide.
Indications for Use
Indicated for patients requiring definitive dental prosthetics including full single crowns, partial crowns (inlays, onlays), veneers, and artificial teeth for removable dentures. For use by dental professionals in offices and laboratories.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 15, 2026
Sprintray, Inc.
Sara Moghtadernejad
Global Director of RA/QA
2710 Media Center Dr., Suite #100a
Los Angeles, California 90065
Re: K260661
Trade/Device Name: Midas Restore
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth Shade Resin Material
Regulatory Class: Class II
Product Code: EBF, ELM
Dated: February 27, 2026
Received: March 2, 2026
Dear Sara Moghtadernejad:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260661 - Sara Moghtadernejad
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260661 - Sara Moghtadernejad
Page 3
Sincerely,

Bobak
Shirmohammadi -S
For Michael E. Adjodha, M.ChE., RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260661 | |
| Device Name Midas Restore | |
| Indications for Use (Describe) Midas Restore is a light-curable polymerizable resin intended to be used for the fabrication of individual and fixed definitive full single crowns; definitive partial crowns (e.g., inlays, onlays) in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
1
510(k) SUMMARY
SprintRay's Midas Restore
Submitter: SprintRay Inc.
2710 Media Center Drive, Suite 100A
Los Angeles, CA 90065
Phone: (800) 914-8004
Contact Person: Dr. Sara Moghtadernejad
Date Prepared: February 27, 2026
Trade/Device Name: Midas Restore
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth Shade Resin Material
Common Name: Material, tooth shade, resin
Regulatory Class: Class II
Product Code: EBF
Secondar Product Code: ELM
Predicate Devices: Digital Crown (K222623)
Device Description
SprintRay Midas Restore is a photo-polymer methacrylate resin material used in conjunction with a 3D printer and a scanned 3D image in a dental office to build dental prosthetics by 3D printing layer upon layer of the composite material. SprintRay Midas Restore is offered in various shades such as Bleach, A1, A2, A3, B1, C2, and unpigmented, compatible with the VITA classical A1-D4 shade guide for color matching. Midas Restore is an alternative to traditional dental prosthesis material that is intended exclusively for professional dental work.
SprintRay Midas Restore resin is intended exclusively for professional dental work. Fabrication of dental prosthetics with SprintRay Midas Restore resin requires computer-aided design and CAD/CAM manufacturing system that
{5}
includes the following components not part of the device: Midas Restore file created in an optical impression system, 3D printer, and curing light equipment.
Midas Restore resin is designed to meet appropriate ISO standards for flexibility and sorption, to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions on curing, and cleaning the final device prior to providing it to a patient are provided in the device indication for use document.
## Intended Use / Indications for Use
Midas Restore is a light-curable polymerizable resin intended to be used for the fabrication of individual and fixed definitive full single crowns; definitive partial crowns (e.g., inlays, onlays) in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material.
## Summary of Technological Characteristics
Light-based curing of a 3D printed acrylate resin is the technological principle for both the subject and predicate devices. The Midas Restore resin used with the Midas 3D printer, which relies on scanned images of the patient's oral cavity to produce a dental appliance. At a high level, the subject and predicate devices are based on the following same technological elements:
- are acrylate resin
- are used in conjunction with 3D printers, which rely on common 3D images to define the fabricated dental appliance
- are cured prior to final trimming and cleaning
The following technological differences exist between the subject and predicate devices:
- differences in acrylate resin material composition
## Performance Data
The subject device is considered tissue contacting for a period longer than 30 Days. The following Biocompatibility testing was performed on samples of the Midas Restore resin that had been formed into dental appliances using a 3D printer in accordance with the FDA Blue Book Memorandum #G95-1 and International Standard ISO 10993-1 and ISO 7405, as recognized by FDA.
- ISO 10993-3 Genotoxicity, carcinogenicity and reproductive toxicity
{6}
- ISO 10993-5 In Vitro Cytotoxicity
- ISO 10993-11 Systemic Toxicity
- ISO 10993-10 skin sensitization
- ISO 10993-23 skin irritation
Additional following bench testing based on the test steps laid out in ISO 10477 was performed using dental appliance fabricated from Midas Restore resin.
- Ultimate Flexural Strength (ISO 10477)
- Water Sorption and Water Solubility (ISO 10477)
- Radio Opacity (ISO 4049)
- Print Accuracy and Dimension Stability
- Surface Finish (ISO 10477)
- Shear Bond Strength (ISO 10477)
- Free Monomer Extraction (ISO 20795-1)
- Tolerance
- Depth of Cure
- Viscosity
- Shade and Color Stability
- Hardware/Software Validation Testing
## Conclusions
The Midas Restore resin is as safe and effective as its predicate device. The Midas Restore resin has the same intended use and indication, and similar technological characteristics, and principles of operation as its predicate device. The minor technological differences between the Midas Restore resin and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Midas Restore resin is as safe and effective as the predicate device. Thus, the Midas Restore resin is substantially equivalent.
{7}
4
# SPRINTRAY, INC.'S
# MIDAS RESTORE RESIN
# SUBSTANTIAL EQUIVALENCE CHART
| | Digital Crown (Predicate Device) | Midas Restore (Subject Device) |
| --- | --- | --- |
| Intended Use & Indications for Use | SprintRay Digital Crown is a light-curable polymerizable resin intended to be used for the fabrication of; individual and fixed definitive full single crowns; definitive partial crowns in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures; and individual and removable monolithic full and partial dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material. | Midas Restore is a light-curable polymerizable resin intended to be used for the fabrication of individual and fixed definitive full single crowns; definitive partial crowns (e.g., inlays, onlays) in anterior and posterior area, individual and fixed single veneers; artificial teeth for dental prostheses, which are used for removable definitive full dentures in dental offices and laboratories. The material is an alternative to traditional restorative dental material. |
| User Population | Clinicians in dental offices provided permanent partial crowns and artificial teeth as indicated in the above IFU for patients. | Same |
| Material Type | Light-curable polymerizable resin | Same |
| Accessories | SprintRay Commercially available 3D printers | SprintRay Midas 3D printer |
| Volume Provided | 250g bottle | 4ml Midas Capsule |
| Shelf Life | 1.5 years | 2 years |
| Biocompatibility | Tested to ISO-10993-1, -3, -5, -10 and -11 | Tested to ISO-10993-1, -3, -5, -10, -11, and -23 |
| Performance Testing | Testing performed using methods laid out in ISO- 10477 | Same |
| Flexural Strength (>50.0 MPa) | 125-174 MPa | 120-150 MPa |
| Sorption (<40 μg/mm³) | 17.35 ± 2.56 μg/mm³ | 22-24 μg/mm³ |
| Solubility (<7.5 μg/mm³) | 2.16 ±1.30 μg/mm³ | 1.0-2.5 μg/mm³ |
| Monomer Methyl Methacrylate (<2.2S%) | Not detectable | Not detectable |
{8}
| Radio Opacity per ISO 10477 | Equivalent to aluminum | Equivalent to aluminum |
| --- | --- | --- |
| Shear Bond Strength ISO 10477 | PASS | PASS |
| Shade and Color Stability after Irradiation and Water Sorption | Shade does not change, PASS | Shade does not change, PASS |
| Surface Finish | Polishable | Polishable |
| Hardware/Software Validation Testing | Validated with SprintRay Workflow | Validated with SprintRay workflow |
| Accuracy Fitting | PASS using recommended SprintRay Workflow | PASS using recommended SprintRay Workflow |
| Depth of Cure | Between 100 to 300 microns at the 100 microns print layer time | Between 100 to 300 microns at the 100 microns print layer time |
| Shade | Pass per VITA shade guide | Pass per VITA shade guide |
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.