The Elevation Spine Saber-C System is a cervical interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) and is for use in one or two adjacent spinal levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The SABER-C Spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone. The SABER-C Spacer is intended to be used with supplemental fixation, such as anterior cervical plates or posterior cervical screw fixation. The SABER-C Spacer is available in a solid titanium and a porous 3D printed titanium offering. When the SABER-C Spacer is used with the SABER-C Anterior Cervical Plate and screws, the plate-spacer-screw assembly can be used as a stand-alone device with the SABER-C Anterior Cervical Plate acting as the supplemental fixation. When the SABER-C Spacer is used with the SABER-C Anterior Cervical Plate and spikes, the plate-spacer-spike assembly should be used with additional supplemental fixation such as posterior cervical screw fixation. The SABER-C Anterior Cervical Plate, when used without the spacer component and with screws, is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis. The Anterior Cervical Plate is not to be used with spikes when used without the Spacer component.
Device Story
SABER-C System is a cervical interbody fusion device and anterior cervical plate system; provides structural stability following discectomy. Components include interbody spacers (solid or porous 3D-printed titanium) and anterior cervical plates with screw or spike fixation. Inserted via anterior cervical approach by surgeons. Spacers filled with autograft or allogenic bone graft to promote fusion. Plate-spacer-screw assembly functions as stand-alone device; plate-spacer-spike assembly requires additional posterior fixation. Plate used alone for fixation in trauma, tumor, or deformity cases. Benefits include spinal stabilization and fusion support in skeletally mature patients.
Clinical Evidence
Bench testing only. Performed static and dynamic axial compression, compression shear, and torsion per ASTM F2077; expulsion and subsidence per ASTM F2267.
Technological Characteristics
Materials: Titanium alloy (ASTM F136, ASTM F1295), PEEK (ASTM F2026), Tantalum (ASTM F560). Manufacturing: Additive manufacturing (3D printing) and traditional machining. Form factor: Cervical interbody spacers and anterior cervical plates. Sterilization: Not specified. Connectivity: N/A.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) at one or two adjacent levels. Also indicated for anterior cervical plate fixation (C2-C7) for DDD, trauma, tumors, deformity, pseudoarthrosis, failed fusion, spondylolisthesis, and spinal stenosis. Contraindications include use of spikes without spacer component.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 27, 2026
Elevation Spine
% Christine Scifert
Partner
MRC Global, LLC
9160 Hwy 64, Suite 12
P.O. Box 330
Lakeland, Tennessee 38002
Re: K260660
Trade/Device Name: SABER-C System
Regulation Number: 21 CFR 888.3080
Regulation Name: Intervertebral Body Fusion Device
Regulatory Class: Class II
Product Code: OVE, KWQ, ODP
Dated: February 27, 2026
Received: March 2, 2026
Dear Christine Scifert:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260660 - Christine Scifert
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260660 - Christine Scifert
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter -S
Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K260660
Page 1 of 1 | ? |
| Please provide the device trade name(s). | | ? |
| SABER-C System | | |
| Please provide your Indications for Use below. | | ? |
| The Elevation Spine Saber-C System is a cervical interbody fusion device intended for use in skeletally
mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) and is for use in one or
two adjacent spinal levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by
history and radiographic studies. These patients should be skeletally mature and have had at least six (6)
weeks of non-operative treatment. The SABER-C Spacer is to be filled with autograft bone and/or allogenic
bone graft composed of cancellous and/or corticocancellous bone.
The SABER-C Spacer is intended to be used with supplemental fixation, such as anterior cervical plates or
posterior cervical screw fixation. The SABER-C Spacer is available in a solid titanium and a porous 3D
printed titanium offering. When the SABER-C Spacer is used with the SABER-C Anterior Cervical Plate and
screws, the plate-spacer-screw assembly can be used as a stand-alone device with the SABER-C Anterior
Cervical Plate acting as the supplemental fixation. When the SABER-C Spacer is used with the SABER-C
Anterior Cervical Plate and spikes, the plate-spacer-spike assembly should be used with additional
supplemental fixation such as posterior cervical screw fixation.
The SABER-C Anterior Cervical Plate, when used without the spacer component and with screws, is
intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative
disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by
patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as
kyphosis, lordosis, or scoliosis), pseudoarthrosis, failed previous fusion, spondylolisthesis, and spinal
stenosis. The Anterior Cervical Plate is not to be used with spikes when used without the Spacer
component. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is
to be used. | ☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☑ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater) | ? |
{4}
K260660
Page 1 of 3
510(k) Summary
SABER-C System
February 27, 2026
Company: Elevation Spine, Inc.
2511 Garden Rd, Suite B125
Monteray, CA 93940
Primary Contact: Christine Scifert – Partner
MRC Global
9160 Hwy 64, Ste 12
PO Box 330
Phone: (901) 831-8053
Email: christine.scifert@AskMRCGlobal.com
Company/Secondary Contact: John Kirwan
VP R&D
Elevation Spine, Inc
jkirwan@elevationspine.com
413-427-6155
Trade Name: SABER-C System
Common Name: Intervertebral Fusion Device With Integrated Fixation, Cervical
Classification: Class II
Regulation: 21 CFR 888.3080 (Intervertebral Fusion Device With Integrated Fixation, Cervical)
Panel: Orthopedic
Product Code: OVE, KWQ, ODP
Primary Predicate: Elevation Spine SABER-C System – K250540
## Device Description:
The SABER-C System is a cervical interbody fusion device used to provide structural stability in skeletally mature individuals following discectomy. The SABER-C System is inserted through an anterior cervical approach and is available in various geometric and material options to fit the clinical needs of the patient. The SABER-C System Anterior Cervical Plate is an anterior cervical fixation device this is available in various geometric and fixation options and can be used with various types of interbody spacers.
{5}
K260660
Page 2 of 3
The previously cleared SABER-C interbody devices are offered manufactured from titanium alloy per ASTM F136 or titanium-coated radiolucent polyether ether ketone (PEEK), with titanium alloy or tantalum markers, as specified in ASTM F2026, ASTM F136, ASTM F1295, and ASTM F560. The SABER-C Anterior Cervical Plate and two fixation options, spikes or screws, are manufactured from titanium alloy, as specified in ASTM F136 and ASTM F1295. The plate is not to be used with spikes when used without the Spacer component.
The subject submission seeks to add additively manufactured components, expand size offerings, and gain clearance for minor design modifications to the previously cleared devices.
## Indications for Use:
The Elevation Spine Saber-C System is a cervical interbody fusion device intended for use in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine (C2-T1) and is for use in one or two adjacent spinal levels. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) weeks of non-operative treatment. The SABER-C Spacer is to be filled with autograft bone and/or allogenic bone graft composed of cancellous and/or corticocancellous bone.
The SABER-C Spacer is intended to be used with supplemental fixation, such as anterior cervical plates or posterior cervical screw fixation. The SABER-C Spacer is available in a solid titanium and a porous 3D printed titanium offering. When the SABER-C Spacer is used with the SABER-C Anterior Cervical Plate and screws, the plate-spacer-screw assembly can be used as a stand-alone device with the SABER-C Anterior Cervical Plate acting as the supplemental fixation. When the SABER-C Spacer is used with the SABER-C Anterior Cervical Plate and spikes, the plate-spacer-spike assembly should be used with additional supplemental fixation such as posterior cervical screw fixation.
The SABER-C Anterior Cervical Plate, when used without the spacer component and with screws, is intended for anterior screw fixation to the cervical spine (C2-C7) for the following indications: degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), trauma (including fractures), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudoarthrosis, failed previous fusion, spondylolisthesis, and spinal stenosis. The Anterior Cervical Plate is not to be used with spikes when used without the Spacer component.
## Substantial Equivalence:
The subject Saber-C components are substantially equivalent to the following predicate devices:
**Primary Predicate:**
Elevation Spine SABER-C System (K250540)
**Additional Predicates:**
Globus Medical, Hedron IC (K191243)
{6}
K260660
Page 3 of 3
The subject components are identical in indications to the primary predicate device. Device sizing, geometry, and technological characteristics, and materials are similar to the predicate Hedron IC (K191243). Materials, manufacturing, sterilization, and packaging are identical to those of the primary predicate, Saber-C (K250540).
## Performance Testing:
The following bench performance testing has been performed on the subject additively manufactured interbody devices Static and Dynamic Axial Compression, Static and Dynamic Compression Shear, Static and Dynamic Torsion per ASTM F2077, Expulsion, and Subsidence per ASTM F2267
## Conclusion:
Based on the performance analysis and the comparison to the predicate devices, the subject device is determined to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.