CONTOUR® PLUS BLUE Blood Glucose Monitoring System; CONTOUR® NEXT GEN Blood Glucose Monitoring System; CONTOUR® NEXT ONE Blood Glucose Monitoring System
K260545 · Ascensia Diabetes Care · NBW · Aug 27, 2026 · Clinical Chemistry
Device Facts
Record ID
K260545
Device Name
CONTOUR® PLUS BLUE Blood Glucose Monitoring System; CONTOUR® NEXT GEN Blood Glucose Monitoring System; CONTOUR® NEXT ONE Blood Glucose Monitoring System
Applicant
Ascensia Diabetes Care
Product Code
NBW · Clinical Chemistry
Decision Date
Aug 27, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CONTOUR® PLUS BLUE blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® PLUS BLUE blood glucose monitoring system is intended to be used by a single person and should not be shared. The CONTOUR® PLUS BLUE blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program. The CONTOUR® PLUS BLUE blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes or for neonatal use. The CONTOUR® PLUS blood glucose test strips are for use with the CONTOUR® PLUS BLUE meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips. The system is intended for in vitro diagnostic use only. The CONTOUR® NEXT GEN Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® NEXT GEN Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® NEXT GEN Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program. The CONTOUR® NEXT GEN Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. The system is intended for in vitro diagnostic use only. The CONTOUR® NEXT ONE Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips or palm. The CONTOUR® NEXT ONE Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® NEXT ONE Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program. The CONTOUR®NEXT ONE Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing (palm) should be done only during steady state times (when glucose is not changing rapidly). The CONTOUR® NEXT test strips are for use with the CONTOUR® NEXT ONE blood glucose meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips or palm. The system is intended for in vitro diagnostic use only.
Device Story
Systems consist of blood glucose meter, test strips, and CONTOUR™ DIABETES App; utilize amperometric biosensor technology to measure glucose in capillary whole blood; results displayed in 5-7 seconds. Meters feature smartCOLOR® or smartLIGHT® indicators for target range visualization; store test results and trends. Bluetooth Low Energy (BLE) enables wireless data transmission to mobile devices running the CONTOUR™ DIABETES App. Used by patients at home for self-testing; aids in monitoring diabetes control program effectiveness. Healthcare providers review app-synced data to inform clinical decision-making. Benefits include improved patient self-management through trend tracking and target range alerts.
Clinical Evidence
No new clinical or analytical performance testing was required; previous testing was leveraged. Usability testing was conducted per FDA human factors guidance to ensure the modified CONTOUR™ DIABETES App is intuitive for typical users. Testing confirmed that app modifications do not impact meter performance.
Technological Characteristics
Amperometric glucose measurement using FAD-dependent glucose dehydrogenase. Connectivity via Bluetooth Low Energy (BLE). Software-based data management via cloud-enabled mobile app (iOS/Android). Meters are automatically coded. Single-use test strips. Compliant with IEC 62304 (software lifecycle) and IEC 82304 (health software).
Indications for Use
Indicated for people with diabetes for self-testing of glucose in fresh capillary whole blood (fingertip or palm) to monitor diabetes control program effectiveness. Not for diagnosis/screening of diabetes, neonatal use, or use in healthcare/assisted-use settings. Contraindicated for critically ill patients, those with reduced peripheral blood flow, severe hypotension, dehydration, shock, or hypoxia.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
CONTOUR® PLUS BLUE Blood Glucose Monitoring System (K241787)
Submission Summary (Full Text)
{0}
[LOGO]
FDA
U.S. FOOD & DRUG
ADMINISTRATION
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
## DECISION SUMMARY
ASSAY AND INSTRUMENT
### I Background Information:
A 510(k) Number
K260545
B Applicant
Ascensia Diabetes Care
C Proprietary and Established Names
CONTOUR® PLUS BLUE Blood Glucose Monitoring System
CONTOUR® NEXT GEN Blood Glucose Monitoring System
CONTOUR® NEXT ONE Blood Glucose Monitoring System
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| NBW | Class II | 21 CFR 862.1345 - Glucose Test System | CH - Clinical Chemistry |
### II Submission/Device Overview:
A Purpose for Submission:
Modification to the CONTOUR™ DIABETES App, a component of the CONTOUR® PLUS BLUE, CONTOUR® NEXT GEN and CONTOUR® NEXT ONE Blood Glucose Monitoring Systems, to include changes to the user interface, data management, and cloud integration.
B Measurand:
Glucose in fresh capillary whole blood
C Type of Test:
Quantitative amperometric assay (glucose dehydrogenase-FAD)
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
{1}
### III Intended Use/Indications for Use:
#### A Intended Use(s):
See Indications for Use below.
#### B Indication(s) for Use:
# **CONTOUR® PLUS BLUE Blood Glucose Monitoring System:**
The CONTOUR® PLUS BLUE blood glucose monitoring system consists of the CONTOUR® PLUS BLUE meter, the CONTOUR® PLUS blood glucose test strips, and the CONTOUR™ DIABETES App.
The CONTOUR® PLUS BLUE blood glucose monitoring system is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® PLUS BLUE blood glucose monitoring system is intended to be used by a single person and should not be shared. The CONTOUR® PLUS BLUE blood glucose monitoring system is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program.
The CONTOUR® PLUS BLUE blood glucose monitoring system should not be used for the diagnosis of or screening for diabetes or for neonatal use. The CONTOUR® PLUS blood glucose test strips are for use with the CONTOUR® PLUS BLUE meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips.
The system is intended for in vitro diagnostic use only.
# **CONTOUR® NEXT GEN Blood Glucose Monitoring System:**
The CONTOUR® NEXT GEN Blood Glucose Monitoring System consists of the CONTOUR® NEXT GEN meter, CONTOUR® NEXT blood glucose test strips and the CONTOUR™ DIABETES App.
The CONTOUR® NEXT GEN Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips. The CONTOUR® NEXT GEN Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® NEXT GEN Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program.
The CONTOUR® NEXT GEN Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use.
The system is intended for in vitro diagnostic use only.
K260545 - Page 2 of 10
{2}
# **CONTOUR® NEXT ONE Blood Glucose Monitoring System:**
The CONTOUR® NEXT ONE Blood Glucose Monitoring System consists of the CONTOUR® NEXT ONE meter, CONTOUR® NEXT blood glucose test strips and the CONTOUR™ DIABETES App.
The CONTOUR® NEXT ONE Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose in fresh capillary whole blood drawn from the fingertips or palm. The CONTOUR® NEXT ONE Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. The CONTOUR® NEXT ONE Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program.
The CONTOUR®NEXT ONE Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes or for neonatal use. Alternative site testing (palm) should be done only during steady state times (when glucose is not changing rapidly). The CONTOUR® NEXT test strips are for use with the CONTOUR® NEXT ONE blood glucose meter to quantitatively measure glucose in fresh capillary whole blood drawn from the fingertips or palm.
The system is intended for in vitro diagnostic use only.
# **C Special Conditions for Use Statement(s):**
OTC - Over The Counter
The following Special Conditions for Use Statements apply to the CONTOUR® PLUS BLUE, CONTOUR® NEXT GEN, and CONTOUR® NEXT ONE blood glucose monitoring systems:
- This device is not intended for use in healthcare or assisted-use settings such as hospitals, physician offices, or long-term care facilities because it has not been cleared by FDA for use in these settings, including for routine assisted testing or as part of glycemic control procedures.
- Use of this device on multiple patients may lead to transmission of Human Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or other bloodborne pathogens.
- Not to be used for the diagnosis of or screening for diabetes
- The system should not be used to test critically ill patients and should not be used by persons with reduced peripheral blood flow.
- Inaccurate results may occur in severely hypotensive individuals, those dehydrated or patients in shock.
- Inaccurate low results may occur for individuals experiencing a hypoxia state, or a hyperglycemic-hyperosmolar state, with or without ketosis.
- Not indicated for neonatal use.
- This system has not been tested at altitudes higher than 20,674 feet (6,301 meters).
- For single-patient use only.
The following Special Conditions for Use Statements apply to both the CONTOUR® PLUS BLUE and CONTOUR® NEXT GEN blood glucose monitoring systems:
K260545 - Page 3 of 10
{3}
- Not intended for use with Alternative Site Testing (AST)
The following Special Conditions for Use Statements apply to the CONTOUR® NEXT ONE blood glucose monitoring system:
- Alternative site testing (palm) should be done only during steady state times (when glucose is not changing rapidly). Ask your health care professional if Alternative Site Testing (AST) is right for you. Do not calibrate a continuous glucose monitoring device from an AST result. Do not calculate an insulin dose based on an AST result.
# D Special Instrument Requirements:
CONTOUR® PLUS BLUE Blood Glucose Meter
CONTOUR® NEXT GEN Blood Glucose Meter
CONTOUR® NEXT ONE Blood Glucose Meter
# IV Device/System Characteristics:
# A Device Description:
The CONTOUR® PLUS BLUE, CONTOUR® NEXT GEN, and CONTOUR® NEXT ONE Blood Glucose Monitoring Systems consist of a meter (CONTOUR® PLUS BLUE blood glucose meter, CONTOUR® NEXT GEN blood glucose meter, and CONTOUR® NEXT ONE blood glucose meter, respectively) and blood glucose test strips (CONTOUR® PLUS blood glucose test strips, CONTOUR® NEXT blood glucose test strips, and CONTOUR® NEXT blood glucose test strips, respectively).
Control solutions (Levels 1 and 2 of CONTOUR® PLUS or CONTOUR® NEXT control solutions), MICROLET® NEXT lancing device with MICROLET® lancets (K220633), and the CONTOUR™ DIABETES App are to be used with the systems and are provided separately.
The CONTOUR™ DIABETES App is a cloud-enabled mobile application that operates on portable devices such as a smart phone or tablet running either the Apple iOS or Android operating system. The CONTOUR™ DIABETES App is used as an optional display for the results obtained with the subject device and may be used to change the CONTOUR® PLUS BLUE, CONTOUR® NEXT GEN, and CONTOUR® NEXT ONE meter settings.
# B Principle of Operation:
The CONTOUR® PLUS BLUE, CONTOUR® NEXT GEN, and CONTOUR® NEXT ONE Blood Glucose Monitoring Systems quantitatively measure the amount of glucose in fresh capillary whole blood. The glucose measurement is achieved using an amperometric detection method by the reaction of FAD dependent glucose dehydrogenase and an electron mediator in the test strip with glucose in the sample producing an electrical current which is proportional to the amount of glucose in the sample. The meter measures the current and converts it to the
K260545 - Page 4 of 10
{4}
corresponding glucose concentration, which is displayed by the meter's display screen in plasma equivalents.
# C Instrument Description Information:
1. Instrument Name:
CONTOUR® PLUS BLUE Blood Glucose Meter
CONTOUR® NEXT GEN Blood Glucose Meter
CONTOUR® NEXT ONE Blood Glucose Meter
2. Specimen Identification:
There is no sample identification function with these devices. Samples are applied directly to the test strip as they are collected.
3. Specimen Sampling and Handling:
The systems are intended to be used with capillary whole blood from the fingertip. The whole blood sample is applied directly to the test strip by capillary action. Samples are tested immediately upon collection.
4. Calibration:
The meters are automatically coded and do not require calibration or coding by the user.
5. Quality Control:
Two levels of glucose control solutions are available for use with each of the blood glucose monitoring system and are used as a quality control check to ensure that the systems are working properly. Recommendations on when to test with control solutions are provided in the labeling. Acceptable ranges for each level of control solution are printed on the test strip vial label and foil packet. If controls solutions fall outside the printed ranges the labeling instructs the customer to not use the system until the issue has been resolved and provides customer service contact details for support, as needed. The three meters recognize the control solutions automatically and differentiate the control solution results from patient results. The control test results are not included in the patient glucose averaging features such as blood glucose averages.
This medical device product has functions subject to FDA premarket review as well as functions that are not subject to FDA premarket review. For this application, if the product has functions that are not subject to FDA review, FDA assessed those functions only to the extent that they either could adversely impact the safety and effectiveness of the functions subject to FDA premarket review or they are included as a labeled positive impact that was considered in the assessment of the functions subject to FDA premarket review.
K260545 - Page 5 of 10
{5}
# V Substantial Equivalence Information:
# A Predicate Device Name(s):
CONTOUR® PLUS BLUE Blood Glucose Monitoring System
# B Predicate 510(k) Number(s):
K241787
# C Comparison with Predicate(s):
| Device & Predicate Device(s): | K260545 | K260545 | K260545 | K241787 |
| --- | --- | --- | --- | --- |
| Device Trade Name | CONTOUR® NEXT ONE | CONTOUR® PLUS BLUE | CONTOUR® NEXT GEN | CONTOUR® PLUS BLUE |
| General Device Characteristic Similarities | | | | |
| Intended Use | Quantitative measurement of glucose in fresh capillary whole blood intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid in monitoring the effectiveness of a diabetes control program. | Same | Same | Same |
| Wireless transmission | BLE | Same | Same | Same |
| General Device Characteristic Differences | | | | |
| Database Storage | Removal of data storage in the app. | Removal of data storage in the app. | Removal of data storage in the app. | Data storage in the app with periodic |
K260545 - Page 6 of 10
{6}
| | Replaced with real time cloud-based data storage | Replaced with real time cloud-based data storage | Replaced with real time cloud-based data storage | backup to the cloud |
| --- | --- | --- | --- | --- |
| Third-Party Integrations | Data Sharing via Third-Party integrations unavailable | Data Sharing via Third-Party integrations unavailable | Data Sharing via Third-Party integrations unavailable | Data Sharing via Third-Party integrations available |
## VI Standards/Guidance Documents Referenced:
IEC 82304 Edition 1.0 2016-10: Health software – Part 1: General requirements for product safety
IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION: Medical device software - Software life cycle processes
IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION: Medical devices - Part 1: Application of usability engineering to medical devices
FDA Guidance, “Self-Monitoring Blood Glucose Test Systems for Over-the-Counter Use”, September 29, 2020
FDA Guidance, “Applying Human Factors and Usability Engineering to Medical Devices”, February 3, 2016
FDA Guidance, “Cybersecurity in Medical Devices: Quality Management System Considerations and Content of Premarket Submissions”, February 3, 2026
## VII Performance Characteristics (if/when applicable):
The device modifications described in this submission do not impact the glucose measurement functions of the system, therefore the performance data was leveraged from previously cleared devices (K160682 for CONTOUR® NEXT ONE, K193407 for CONTOUR® NEXT GEN, and K231679 for CONTOUR® PLUS BLUE).
### A Analytical Performance:
1. Precision/Reproducibility:
Previously established in K160682, K193407, and K231679.
K260545 - Page 7 of 10
{7}
2. Linearity:
Previously established in K160682, K193407, and K231679.
3. Analytical Specificity/Interference:
Previously established in K160682, K193407, and K231679.
The sponsor includes the following information in the labeling:
Xylose: Do not use during or soon after xylose absorption testing. Xylose in the blood will cause an interference.
4. Assay Reportable Range:
As previously established in K193407, K231679 and K160682, the assay reportable range is 20-600 mg/dL.
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability to the National Institute of Standards and Technology (NIST) SRM 917 glucose reference material, as previously established in K160682, K193407, and K231679.
Test strip stability protocols and acceptance criteria were previously reviewed in K160682, K193407, and K231679 and found to be acceptable to support the open-vial and shelf-life stability claims.
6. Detection Limit:
Previously established in K160682, K193407, and K231679.
7. Assay Cut-Off:
Not applicable.
8. Accuracy (Instrument):
Not applicable.
9. Carry-Over:
Not applicable. The device uses single-use test strips.
# B Comparison Studies:
1. Method Comparison with Predicate Device:
Not applicable.
K260545 - Page 8 of 10
{8}
2. Matrix Comparison:
Not applicable. These devices are only intended for use with fresh capillary whole blood.
# C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Lay-user Performance study: Previously established in K160682, K193407, and K231679.
# D Clinical Cut-Off:
Not applicable.
# E Expected Values/Reference Range:
The sponsor states the following in the labeling:
Blood glucose values will vary depending on food intake, medication dosages, health, stress, or activity. Nondiabetic plasma glucose concentrations should be less than 100 mg/dL in the fasting state and less than 140 mg/dL in the postprandial state (after a meal). You should consult with your health care professional for glucose values specific to your needs.
American Diabetes Association. 2. Classification and diagnosis of diabetes: Standards of medical care in diabetes—2020. Diabetes Care. 2020; 43(supplement 1): S14–S31.
# F Other Supportive Instrument Performance Characteristics Data:
# Electrical safety and EMC
The sponsor did not change physical components of the device systems. The electrical and electromagnetic compatibility (EMC) testing performed in K160682 and K241787 are applicable to their respective device systems and devices systems were found to be compliant.
# Flex Studies
The sponsor did not change physical components of the device systems. The flex studies performed in K160682, K193407, K231679 and K241787 are applicable to their respective device systems and demonstrate the devices are robust under the specified use conditions.
K260545 - Page 9 of 10
{9}
# Software Verification and Validation
The sponsor provided detailed information on software changes made to support the modifications to the App. The verification and validation of the software functions were reviewed and found acceptable.
# Cybersecurity
The sponsor demonstrated that cybersecurity risks associated with the modifications were adequately addressed for both systems.
# Usability
Usability testing was conducted in accordance with FDA Guidance, Applying Human Factors and Usability Engineering to Medical Devices, Feb 3, 2016; to ensure that the CONTOUR™ DIABETES App was easy to use by typical customers.
# VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
# IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K260545 - Page 10 of 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.