K260517 · Zhejiang Curaway Medical Technology Co., Ltd. · KNW · Aug 28, 2026 · Gastroenterology, Urology
Device Facts
Record ID
K260517
Device Name
Automatic Core Biopsy Instrument
Applicant
Zhejiang Curaway Medical Technology Co., Ltd.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Aug 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
Device Story
Automatic Core Biopsy Instrument; manual, spring-loaded, automatic-activation biopsy device. Components: inner stylet, sampling cannula, cutting cannula, adjustable knob, security lock, button, handle, protective sheath. Optional coaxial biopsy needle for guidance. Used by clinicians in clinical settings to obtain soft tissue samples. Operation: clinician inserts device into target tissue; triggers automatic activation to capture core sample. Output: tissue core sample for diagnostic evaluation. Benefits: enables minimally invasive tissue acquisition for pathology/diagnosis.
Clinical Evidence
No clinical data. Bench testing only: biocompatibility (ISO 10993-5, -10, -11, -23), sterilization (ISO 11135, ISO 11737-2), packaging/shelf-life (ASTM F1980-21), simulated transportation (ASTM D4169), and performance comparisons (appearance, dimensions, sampling, connection, stiffness, corrosion, puncture force).
Indicated for patients requiring soft tissue biopsies (liver, kidney, prostate, spleen, breast, lung, lymph nodes, tumors). Not for use in bone. Prescription use only.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{1}
FDA U.S. FOOD & DRUG
ADMINISTRATION
August 28, 2026
Zhejiang Curaway Medical Technology Co., Ltd.
Yin Li
Regulatory affairs specialist
Rm. 106, Bldg. 1, # 600, 21st Ave., Baiyang Sub-District, Qiantang New District
Hangzhou, Zhejiang 310018
China
Re: K260517
Trade/Device Name: Automatic Core Biopsy Instrument
Regulation Number: 21 CFR 876.1075
Regulation Name: Gastroenterology-Urology Biopsy Instrument
Regulatory Class: Class II
Product Code: KNW
Dated: July 31, 2026
Received: July 31, 2026
Dear Yin Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260517 - Yin Li
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260517 - Yin Li
Page 3
Sincerely,
# JESSICA CARR -S
Jessica Carr, PhD
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260517 | ? |
| Please provide the device trade name(s). | | ? |
| Automatic Core Biopsy Instrument | | |
| Please provide your Indications for Use below. | | ? |
| Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
Curaway
Zhejiang CuraWay Medical Technology Co., Ltd.
K260517
# 510(K) Summary
| Company Name/Owner | Zhejiang CuraWay Medical Technology Co., Ltd. |
| --- | --- |
| Contact person/Author | Yin Li |
| Date prepared | August 28, 2026 |
| Contact details Address | Room 106, Building 1, No. 600, 21^{st} Avenue, Baiyang Sub-district, Qiantang New District, 310018 Hangzhou City, Zhejiang Province, China |
| Contact phone number | 86-571-87016876 |
| Trade name | Automatic Core Biopsy Instrument |
| Common name | Core Biopsy Instrument, Biopsy Needle |
| Classification name | Instrument, biopsy |
| Review panel | Gastroenterology/Urology |
| Regulation number | 21 CFR 876.1075 |
| Product code | KNW |
| Predicate device | Corvocet Biopsy System (K180450) |
1
{6}
Curaway
Zhejiang CuraWay Medical Technology Co., Ltd.
## 1. Device description
Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues. It has one model BN-OCR-13, and this model has two specifications: BN-OCR-13/xxyy and BN-OCR-13/xxyy/I. Specification BN-OCR-13/xxyy contains one biopsy instrument, and specification BN-OCR-13/xxyy/I contains one biopsy instrument and one coaxial biopsy needle.
Biopsy instrument of Automatic Core Biopsy Instrument is consisted of Inner stylet, Sampling cannula, Cutting cannula, Adjustable knob, Security lock, Button, Handle and Protective sheath.
Coaxial biopsy needle (internal code: CBN) of Automatic Core Biopsy Instrument is consisted of Inner Stylet, Cutting Cannula, Operating Handle of Cutting Cannula, Operating Handle of Inner Stylet, Blunt Needle (optional), Depth Stopper, and Protective Sheath. It is designed as guiding needle for Automatic Core Biopsy Instrument. The Cutting Cannula of coaxial biopsy needle is one gauge-size larger than the biopsy instrument, e.g., 19 gauge coaxial biopsy needle for a 20 gauge biopsy instrument.
## 2. Indications for use
Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. It is not intended for use in bone.
## 3. Technological characteristics and product feature comparison
The following characteristics of subject device and predicate device are considered as substantially equivalent:
- Indications for use
- Operation mechanics
- Sterilization method
- Body contact sites
- Penetration depth
Comparison table
| Comparison Items | Subject Device | Predicate Device | Comment |
| --- | --- | --- | --- |
| | Automatic Core Biopsy Instrument | Corvocet Biopsy System | |
| Indications for use | Automatic Core Biopsy Instrument is intended for use in obtaining biopsies from soft tissues such | The disposable Corvocet Biopsy System is intended for use in obtaining core biopsy samples from | Same |
2
{7}
Curaway
Zhejiang CuraWay Medical Technology Co., Ltd.
| Comparison Items | Subject Device | Predicate Device | Comment |
| --- | --- | --- | --- |
| | Automatic Core Biopsy Instrument | Corvocet Biopsy System | |
| | as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. It is not intended for use in bone. | |
| Operation mechanics | Automatic activation | Automatic activation | Same |
| Body contact site | Liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. | Liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors. | Same |
| Nominal length of Cutting Cannula | 8cm, 10cm, 13cm, 16cm, 20cm, 25cm | 10cm, 15cm, 20cm, 25 cm | Different |
| Gauge size | 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G | Same |
| Penetration depth | 10~25mm | 10~25mm | Same |
| Sterilization method | EO sterilization | EO sterilization | Same |
Difference in nominal length of cutting cannula:
The nominal length range of predicate device is from 10cm to 25cm, whereas that of the subject device ranges from 8cm to 25cm. Considering different lesion depths, 8cm is provided by subject device. Besides, given the same needle gauge, a longer needle is more susceptible to fracture. A shorter length presents a lower risk of breakage. This difference does not cause risks of safety or effectiveness.
## 4. Performance data
The following design verification and validation were performed to demonstrate the substantial equivalence of the subject device to the predicate device.
### 4.1 Biocompatibility testing
Biocompatibility testing was conducted in accordance with FDA guidance document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" and
3
{8}
Curaway
Zhejiang CuraWay Medical Technology Co., Ltd.
demonstrates that the product meets biological safety requirements for externally communicating medical devices with tissue contact for less than 24 hours.
• In Vitro Cytotoxicity Test (ISO 10993-5: 2009)
- Intracutaneous Reactivity Test (ISO 10993-23: 2021)
- Skin Sensitization test (ISO 10993-10: 2021)
• Acute Systemic Toxicity Test (ISO 10993-11: 2017)
• Pyrogen Test (ISO 10993-11: 2017)
### 4.2 Sterilization validation
Sterilization validation is performed according to ISO 11135: 2014, EO and ECH are validated according to ISO 10993-7: 2008. SAL= \( 10^{-6} \) .
The sterility test is carried out according to ISO 11737-2: 2019.
### 4.3 Packaging and shelf-life testing
The primary package of the proposed device is able to maintain the sterility and performances of the product during its claimed shelf life. The testing is carried out in accordance with ASTM F1980-21.
### 4.4 Simulated transportation
Simulated transportation testing is conducted in accordance with ASTM D 4169. Test results indicate that product packaging can protect products from damage during transportation. Product packaging is sufficient enough to maintain the sterility and product performances during transportation.
### 4.5 Performance comparison with predicate devices
Comparison test was conducted between the subject device and predicate device to compare their performances, including appearance, dimensions, sampling performance, connection firmness, stiffness, toughness, resistance to corrosion, puncture force, chemical properties, etc.
## 5. Clinical data
No clinical data is provided.
## 6. Conclusion
Based on the device comparison and results of nonclinical testings, the subject device is substantially equivalent to the predicate device. The subject device incorporates comparable technological characteristics, demonstrate equivalent safety and performance, and does not raise new questions of safety or effectiveness.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.