The ResolutionMD solution is an enterprise imaging viewer that is used for the secure display, multiplanar reformatting and 3D visualization of medical image data and reports. It also provides for mobile access and real-time collaboration of clinical information.The software is intended for use as a review and analysis tool by authorized healthcare professionals to drive clinical management. When interpreted by a physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.ResolutionMD displays both lossless and lossy compressed images. For lossy images, the healthcare professional must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory clearances, technical specifications, and quality control requirements for their use and maintenance. For mammography, mobile devices do not comply with the minimum display requirements; therefore mobile device usage for mammography is for reference and referral only.Healthcare institutions may want to provide patients access to their clinical information. This access is for non-diagnostic informational purposes only.
Device Story
ResolutionMD is an enterprise imaging viewer software; enables secure display, multiplanar reformatting (MPR), and 3D visualization of medical images/reports. Operates via client-server architecture; server-side processing on Linux/Windows; accessed via HTML5 web browsers on desktops or mobile devices. Integrates with PACS/VNA to retrieve DICOM/non-DICOM data. Features include window/level, rotation, measurement, markup, and real-time collaboration. Used by healthcare professionals in clinical settings to aid diagnosis and management. No local client software installation required. Benefits include centralized access to clinical information and remote diagnostic capabilities on validated hardware. Patient access for non-diagnostic informational purposes is supported.
Clinical Evidence
No clinical studies performed. Evidence based on bench testing (unit, integration, system levels), regression testing, and validation by clinical experts/end users. Testing confirmed DICOM compliance, image rendering, spatial resolution, and quality for mammography/DBT. Compliance with IEC 62304, IEC 62366-1, NEMA PS 3.1-3.20, ISO 10918-1, and IEC 62563-1.
Technological Characteristics
SaMD; client-server architecture; HTML5-based. Supports DICOM (CT, MR, CR, DX, MG, etc.) and non-DICOM (JPG, TIFF, AVI, MP4). Server runs on Windows/Linux. Connectivity via web browser. Standards: IEC 62304, IEC 62366-1, NEMA PS 3.1-3.20 (DICOM), ISO 10918-1 (JPEG), IEC 62563-1 (Handheld Display). No AI/ML.
Indications for Use
Indicated for authorized healthcare professionals to review, analyze, and visualize medical image data (including mammography and digital breast tomosynthesis) to aid in clinical management and diagnosis. Mobile device usage for mammography is restricted to reference and referral only. Not intended to replace full radiology workstations.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
July 21, 2026
ResolutionMD, Inc.
Sharon Cholowsky
Director Regulatory & Compliance
Suite 314, 1212 31 Ave. NE
Calgary, AB T2E 7S8
Canada
Re: K260500
Trade/Device Name: ResolutionMD
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: June 19, 2026
Received: June 22, 2026
Dear Sharon Cholowsky:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260500 - Sharon Cholowsky
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K260500 - Sharon Cholowsky
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, Ph.D.
Assistant Director, Imaging Software Team
DHT8B: Division of Radiological Imaging
Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260500 | ? |
| Please provide the device trade name(s). | | ? |
| ResolutionMD | | |
| Please provide your Indications for Use below. | | ? |
| The ResolutionMD solution is an enterprise imaging viewer that is used for the secure display, multiplanar reformatting and 3D visualization of medical image data and reports. It also provides for mobile access and real-time collaboration of clinical information.The software is intended for use as a review and analysis tool by authorized healthcare professionals to drive clinical management. When interpreted by a physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.ResolutionMD displays both lossless and lossy compressed images. For lossy images, the healthcare professional must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory clearances, technical specifications, and quality control requirements for their use and maintenance. For mammography, mobile devices do not comply with the minimum display requirements; therefore mobile device usage for mammography is for reference and referral only.Healthcare institutions may want to provide patients access to their clinical information. This access is for non-diagnostic informational purposes only. | | |
| Please select the types of uses (select one or both, as applicable). | Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
ResolutionMD
K260500
# 510(k) Summary
For 510(k) #: K260500
As per 21 CFR 807.92
Prepared on June 20, 2026
## Contact Details
| Applicant Name: | ResolutionMD, Inc. |
| --- | --- |
| Applicant Address: | Suite 314, 1212 31 Avenue NE Calgary AB T2E 7S8 Canada |
| Applicant Contact Telephone: | 1-587-430-4591 |
| Applicant Contact: | Sharon Cholowsky |
| Applicant Contact Email: | regulatory@resolutionmd.com |
## Device Name
| Device Trade Name: | ResolutionMD |
| --- | --- |
| Common Name: | Medical image management and processing system |
| Classification Name: | System, Image Processing, Radiological |
| Regulation Number: | 892.2050 |
| Product Code: | LLZ |
## Predicate Devices
| 510(k) #: K161130 | Device Name: ResolutionMD | Product Code: LLZ |
| --- | --- | --- |
| 510(k) #: K172490 | Device Name: eUnity | Product Code: LLZ |
## Device Description Summary
The ResolutionMD product is an enterprise imaging viewer software solution which enables users to access a centralized application running on a Linux or Windows server from an HTML5 web browser or HTML5 browser on a mobile device. The ResolutionMD application server interacts with a PACS or vendor-neutral archive (VNA) to enable users to search and view DICOM and non-DICOM clinical information.
ResolutionMD is a software application system which runs on general purpose computing hardware and operating systems software. It uses a client-server architecture whereby all processing takes place on the server. There is no local client software to install or download; the software is installed on a server.
## Indications for Use
The ResolutionMD solution is an enterprise imaging viewer that is used for the secure display, multiplanar reformatting and 3D visualization of medical image data and reports. It also provides for mobile access and real-time collaboration of clinical information.
The software is intended for use as a review and analysis tool by authorized healthcare professionals to drive clinical management. When interpreted by a physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.
510(k) Summary - ResolutionMD
Page 1 of 5
{5}
Resolution
MD
ResolutionMD displays both lossless and lossy compressed images. For lossy images, the healthcare professional must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory clearances, technical specifications, and quality control requirements for their use and maintenance. For mammography, mobile devices do not comply with the minimum display requirements; therefore mobile device usage for mammography is for reference and referral only.
Healthcare institutions may want to provide patients access to their clinical information. This access is for non-diagnostic informational purposes only.
## Indications for Use Comparison
The subject device indications for use (IFU) statement is similar to the predicate device statement. Phrases relating to mammography diagnostic use were modified, thereby adding support for diagnostic use of mammographic images, including digital breast tomography images. Additionally, some existing phrases were slightly modified for further clarity or standardization of terms.
The differences of the predicate and reference IFU statements and the subject device IFU statement do not present any new safety and effectiveness concerns because the new wording only adds further clarity and information to aid the healthcare professional and does not alter the clinical workflow of the device or how it will be used.
| | Subject Device | Predicate Device | Reference Device for Mammo |
| --- | --- | --- | --- |
| Feature | ResolutionMD (This submission) | ResolutionMD (K161130) | eUnity (K172490) |
| Model # | Model #: RESMD-D | Model #: RESMD-C | N/A |
| Class, Product Code | Class 2, LLZ | Same | Same |
| Intended Use | An enterprise imaging viewer that is used for the secure display, multiplanar reformatting and 3D visualization of medical image data and reports. It also provides for mobile access and real-time collaboration of clinical information. It is to be used as a review and analysis tool by authorized healthcare professionals to drive clinical management. | Same | Same |
| Pres. use or OTC | Prescription use | Same | Same |
| Intended Users | Healthcare professionals | Same | Same |
| Mammography support | Support for diagnostic use of mammographic (incl digital breast tomography) images | Support for non-diagnostic use of mammographic (incl digital breast tomography) images. | Same as subject device |
| Mammo mobile use | Mobile usage for mammography is for reference and referral only. | Same | Same |
| Indications for Use Statement | The ResolutionMD solution is an enterprise imaging viewer that is used for the secure display, multiplanar reformatting and 3D visualization of medical image data and reports. It also provides for mobile access and real-time collaboration of clinical information. | ResolutionMD software is an enterprise medical image viewer used with off-the-shelf servers, web browsers, and specific mobile devices for the 2D display, Multi-planar reformatting and 3D visualization of medical image data | eUnity is a software application that displays medical image data and associated clinical reports to aid in diagnosis for healthcare professionals. It performs operations relating to the transfer, storage, display, and |
510(k) Summary - ResolutionMD
Page 2 of 5
{6}
Resolution
MD
| | The software is intended for use as a review and analysis tool by authorized healthcare professionals to drive clinical management. When interpreted by a physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.ResolutionMD displays both lossless and lossy compressed images. For lossy images, the healthcare professional must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory clearances, technical specifications, and quality control requirements for their use and maintenance. For mammography, mobile devices do not comply with the minimum display requirements; therefore mobile device usage for mammography is for reference and referral only.Healthcare institutions may want to provide patients access to their clinical information. This access is for non-diagnostic informational purposes only. | and reports. It provides collaboration and integrated secure audio-video communication, and displays DICOM and non-DICOM medical images and reports.ResolutionMD is intended for use as a diagnostic, review, and analysis tool by trained healthcare professionals to drive clinical management. When interpreted by a trained physician, reviewed images may be used to aid in diagnosis. When used on a mobile device, ResolutionMD is not intended to replace full radiology workstations.ResolutionMD is not to be used for primary mammography diagnoses. | measurement of image data.eUnity allows users to perform image manipulations, including window/level, rotation, measurement and markup. eUnity provides 2D display, Multi-Planar Reformatting and 3D visualization of medical image data, and mobile access to images.eUnity displays both lossless and lossy compressed images. For lossy images, the medical professional user must determine if the level of loss is acceptable for their purposes. Display monitors or mobile devices used for reading medical images for diagnostic purposes must comply with applicable regulatory approvals and with quality control requirements for their use and maintenance. For Mobile diagnostic usage only when a full workstation is not available.Mobile usage for mammography is for reference and referral only. |
| --- | --- | --- | --- |
### Technological Comparison
The subject device, ResolutionMD (model # RESMD-D), is a modification of the predicate device, ResolutionMD (model # RESMD-C), with improvements to various existing features, such as usability enhancements, performance improvements, integration improvements, updated software coding standards and supported software, etc. There were no changes to the client-server architecture, image display technology, reconstruction/projection algorithms, clinical workflow, mobile diagnostic use, etc.
These improvements do not alter the fundamental technology, do not introduce any new risks, and do not affect the safety/ effectiveness or the intended use of the device. No additional testing (outside of the standard verification and validation procedures) was needed to confirm the safety and effectiveness of these changes.
| | Subject Device | Predicate Device |
| --- | --- | --- |
| Feature | ResolutionMD (This submission) | ResolutionMD (K161130) |
| Model # | Model #: RESMD-D | Model #: RESMD-C |
| Product type | Stand-alone software device (SaMD). Does not include AI/ML. | Same |
510(k) Summary - ResolutionMD
Page 3 of 5
{7}
Resolution
MD
| DICOM modalities support | CT, MR, CR, DX, ES, KO, NM, OP, PT, SC, US, XA, RF, MG IO, XC, SR, RTIMAGE, OPT, ECG | CT, MR, CR, DX, ES, KO, NM, OP, PT, SC, US, XA, RF, MG |
| --- | --- | --- |
| Non-DICOM support | JPG, TIFF, AVI, MP4 and other non-DICOM data | Same |
| Software Architecture | Client/server. No user installation of client. | Same |
| Server platform | Microsoft Windows OS or Red Hat Linux OS. General purpose computing hardware. | Same |
| Client platform - Desktop/laptop | HTML5 web browser on desktop operating systems. | HTML5 or Flex web browser on desktop operating systems. |
| Client platform - Mobile device | HTML5 browser on iOS and Android mobile devices. Optionally using a shell iOS/Android app that uses the HTML5 client. | Native iOS app and native Android app. HTML5 browser on iOS mobile devices. |
| Mobile diagnostic use | Diagnostic use on validated iOS and Android mobile devices. Mobile usage for mammography is for reference/referral only. | Same |
| Secure data access | Data maintained on server-side; no data stored on client. Encryption, authentication, audit logs, etc. | Same |
| Compression | Lossless and lossy JPEG image compression. Image is initially displayed with no compression (100% quality). Interactive slider provides ability to increase/decrease quality. | Lossless and lossy JPEG image compression. Image is initially displayed at 80% quality. Interactive slider provides ability to increase/decrease quality. |
| Search interface | Ability to search for patient studies in PACS and archives through a variety of configurable methods, based on several criteria. Also can visually see the timeline of patient imaging. | Ability to search for patient studies in PACS and archives through a variety of configurable methods, based on several criteria. |
| 2D viewing | 2D image viewer displaying DICOM images, reports, and non-DICOM data. | Same |
| MIP, MPR, cMPR viewing | MIP, MPR, and curvedMPR reconstructions and projections | Same |
| 3D viewing | 3D volume rendering | Same |
| Metadata display | Display of patient/study information, including a more visible patient banner. Optionally toggle display off, for confidentiality. | Display of patient/study information. Optionally toggle display off, for confidentiality. |
| Interactive user controls | Window/level, rotate, pan, zoom, presets, reset, hotkeys, slice scrolling, 3D lens. | Same |
| Measurement & Markup tools | Tools include freehand, text, point, linear, ROI, volume, angle, calipers, 3D segmentation, ECG. | Tools include freehand, text, point, linear, ROI, volume, angle, calipers, 3D segmentation. |
| Multi-study layouts | Multi-study intelligent layouts with reference lines and linked scrolling. Includes multi-monitor support. | Multi-study intelligent layouts with reference lines and linked scrolling. |
| Significant findings | GSPS/CSPS, key image objects (KO), and KIN display and navigation. | Same |
| Collaboration | Real-time shared image viewing session with other invited users, including a text chat. | Real-time shared image viewing session with other invited users, including an audio-visual component. |
| Printing and sharing | Image/Report printing and downloading; DICOM secondary capture | Same |
510(k) Summary - ResolutionMD
Page 4 of 5
{8}
ResolutionMD
## Non-Clinical and Clinical Tests Summary
The ResolutionMD device was verified at the unit, integration, and system levels, including feature testing, regression testing, and deployment/upgrade testing, with both manual and automated testing across various configurations, browsers, monitors, and mobile devices. The testing included a number of different areas, including elements of safety, performance, security, and usability requirements. Validation testing was also completed by clinical experts and end users, based on various typical workflows, including mobile diagnostic use for image types other than mammography.
Testing for mammography diagnostic use demonstrated that ResolutionMD maintained appropriate DICOM compliance, image rendering behavior, spatial resolution performance, and overall image quality for mammography and DBT usage when using appropriate viewing hardware.
The software development process complies with IEC 62304 “Medical Device Software – Software Life Cycle” and IEC 62366-1 “Medical devices Usability Engineering” standards. The software design and testing complies with appropriate technical standards, such as the NEMA PS 3.1 – 3.20 DICOM standard, ISO 10918-1 JPEG standard, and Annex D Handheld Display Devices of IEC 62563-1 “Medical image display systems” standard.
Verification and validation testing confirmed the device meets performance, safety, usability and security requirements. The testing and results did not raise new questions of safety and effectiveness from those associated with the predicate device and demonstrated that the ResolutionMD device performs substantially equivalent to the predicate device.
No additional clinical studies were needed to support the changes that are part of this submission. Ongoing clinical evidence for the device is regularly gathered through post-market surveillance, including literature review and data analysis, as well as an in-depth review of information through clinical evaluation.
## Conclusions
Based on the above considerations, ResolutionMD, Inc. believes that the ResolutionMD software is substantially equivalent to the predicate device and is as safe and effective as the predicate for its intended use.
510(k) Summary - ResolutionMD
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.