stentFIX System Set (100.50n)

K260487 · Ovesco Endoscopy AG · PKL · Oct 2, 2026 · Gastroenterology, Urology

Device Facts

Record IDK260487
Device NamestentFIX System Set (100.50n)
ApplicantOvesco Endoscopy AG
Product CodePKL · Gastroenterology, Urology
Decision DateOct 2, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Instrument for flexible endoscopy for fixation of metal stents in the gastrointestinal tract for use in adults only.

Device Story

Endoscopic instrument for metal stent fixation in gastrointestinal tract; used in adults. System components: application cap with mounted Nitinol clip, application ring, thread, thread retriever, hand wheel. Operation: cap mounted to diagnostic gastroscope tip; hand wheel tightens thread to pull clip toward distal cap end for release. Clip exerts constant force to secure tissue and stent position. Device allows passage through stents (≥18 mm) for proximal or distal fixation. Used by physicians during flexible endoscopy. Benefits: secure stent positioning; long-term implantable; biocompatible; MRI conditional.

Clinical Evidence

No clinical testing submitted. Substantial equivalence demonstrated via performance bench testing, including functional capability, robustness, endoscope compatibility, clip application force, and application ring strength, compared against predicate and established precedents for cap-based endoscopic clip application.

Technological Characteristics

Over-the-scope clip system; Nitinol (superelastic memory-shape alloy) clip; application cap, ring, and thread assembly. Compatible with flexible endoscopes. Biocompatible materials. MRI conditional. Mechanical actuation via hand wheel-driven thread release.

Indications for Use

Indicated for fixation of metal stents in the gastrointestinal tract in adult patients.

Regulatory Classification

Identification

A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.

Special Controls

*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{1} ![LOGO](FDA U.S. FOOD & DRUG ADMINISTRATION) October 2, 2026 Ovesco Endoscopy AG % Maria Kiel Project Manager Novineon Cro Friedrich-Miescher-Strasse 9 Tuebingen, BW 72076 Germany Re: K260487 Trade/Device Name: stentFIX System Set (100.50n) Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligation Regulatory Class: Class II Product Code: PKL Dated: September 3, 2026 Received: September 3, 2026 Dear Maria Kiel: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {2} K260487 - Maria Kiel Page 2 Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the {3} K260487 - Maria Kiel Page 3 Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, SIVAKAMI VENKATACHALAM -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {4} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260487 | ? | | Please provide the device trade name(s). | | ? | | stentFIX System Set (100.50n) | | | | Please provide your Indications for Use below. | | ? | | Instrument for flexible endoscopy for fixation of metal stents in the gastrointestinal tract for use in adults only. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {5} K260487 Page 1 of 2 | 510(k) #: K260487 | 510(k) Summary | Prepared on: 2026-10-01 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Ovesco Endoscopy AG | | | Applicant Address | Friedrich-Miescher-Strasse 9 Tuebingen BW 72076 Germany | | | Applicant Contact Telephone | +49 160 2294081 | | | Applicant Contact | Ms. Nikol Gross | | | Applicant Contact Email | nikol.gross@ovesco.com | | | Correspondent Name | novineon CRO | | | Correspondent Address | Friedrich-Miescher-Strasse 9 Tuebingen BW 72076 Germany | | | Correspondent Contact Telephone | +49707198979134 | | | Correspondent Contact | Dr. Maria Kiel | | | Correspondent Contact Email | maria.kiel@novineon.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | stentFIX System Set (100.50n) | | | Common Name | Hemorrhoidal ligator | | | Classification Name | Hemostatic Metal Clip For The Gi Tract | | | Regulation Number | 876.4400 | | | Product Code(s) | PKL | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K183309 | stentfix OTSC System Set | PKL | | Device Description Summary | | 21 CFR 807.92(a)(4) | | The stentFIX System Set is a product for flexible endoscopy to be used in the gastrointestinal tract. It is intended for the fixation of metal stents in the digestive tract. The design of the stentFIX System allows the passage through a stent (from Ø 18 mm), therefore, fixation is possible either on the proximal or distal end. The set consists of the following components: - application cap with a mounted clip and an application ring - thread - thread retriever | | | {6} K260487 Page 2 of 2 - hand wheel for clip release. The stentFIX clip is delivered by means of an application cap mounted to the tip of diagnostic gastroscopes. The clip is released by tightening the thread via hand wheel, which pulls the clip towards the distal cap end. The clip is made of superelastic Nitinol and exerts a constant force to secure the tissue in the application area, thus securing the positioning of the stent. The stentFIX clip is biocompatible and conditionally MRI safe and can remain in the body as a long-term implant. Scientific concepts forming the basis for the device: - Established memory-shape alloy clips used for over-the-scope clip in flexible endoscopy in the gastrointestinal tract. Significant physical and performance characteristics: - Device design: Device design is guided by compatibility with flexible endoscopes and device influence on endoscopic functions, e.g., vision and maneuverability. - Material used: Materials used need to be sufficiently biocompatible considering the duration and nature of contact to the human body. - Physical properties: Dimensions need to allow insertion via flexible endoscope into the gastrointestinal tract. ## Intended Use/Indications for Use 21 CFR 807.92(a)(5) Instrument for flexible endoscopy for fixation of metal stents in the gastrointestinal tract for use in adults only. ## Indications for Use Comparison 21 CFR 807.92(a)(5) Predicate Device: stentfix OTSC System Set Submission Number: K183309 Regulation Number: 876.4400 Product Code: PKL (hemostatic metal clip for the gi tract) The proposed Predicate Device is an earlier generation of the Subject Device. The proposed Predicate Device and the Subject Device have the same intended use. Differences in technical characteristics are identified. Questions of safety and effectiveness associated with these differences are shown to equally apply to the Predicate Device. The identified differences do therefore not raise different questions of safety and effectiveness. Methods for investigating differences effects on safety and effectiveness are applied. Data from these methods support substantial equivalence. ## Technological Comparison 21 CFR 807.92(a)(6) The subject device is a further development of the predicate device. They share common technological aspects. I.e., the devices constitute over-the-scope clips intended for stent fixation in the gastrointestinal tract. ## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) The sponsor manufactures a number of devices similar to the subject device, including the predicate device. Performance bench testing settings investigated key aspects of device safety and effectiveness. The devices were tested for functional capability and robustness, endoscope compatibility, clip application force and application ring strength. The testing was conducted for the predicate device (partly using reference devices) and was repeated for the subject device, in case of relevant device changes. Not applicable - no clinical testing submitted for demonstration of substantial equivalence. The subject device was subjected to performance bench testing previously applied to other devices that constitute a cap attached to an endoscope tip for gastrointestinal clip application. The established precedent from these devices was met as evidenced by application of acceptance criteria and / or comparison to their respective performance bench testing results, thus indicating substantial equivalence.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...