K260474 · TELA Bio, Inc. · FTL · Sep 28, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260474
Device Name
Liora Monofilament Scaffold
Applicant
TELA Bio, Inc.
Product Code
FTL · General, Plastic Surgery
Decision Date
Sep 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Liora is indicated for soft tissue reinforcement where weakness exists. It is indicated specifically in muscle flap transfer procedures as a temporary support to assist in securing muscle flaps to defect sites.
Device Story
Liora Monofilament Scaffold; sterile, resorbable, monofilament mesh composed of 100% PLGA; used for soft tissue reinforcement and temporary support during muscle flap transfer procedures. Device provided in various shapes/sizes; trimmable by surgeon to fit defect. Implanted by surgeons in clinical settings. Functions as mechanical support to secure muscle flaps to defect sites; degrades via bulk hydrolysis. Provides temporary reinforcement; strength profile supports tissue during healing. Benefits include stabilization of muscle flaps during reconstruction. No software or AI components.
Clinical Evidence
No clinical data. Evidence consists of benchtop performance testing (tensile strength, stiffness, suture pullout, burst strength, tear resistance), biocompatibility testing (ISO 10993-1), chemical characterization (ISO 10993-18), and in vitro degradation testing (ASTM F1635-24). In vivo implantation data also provided.
Technological Characteristics
100% PLGA resorbable monofilament scaffold. Sterilized via Ethylene Oxide (SAL 10^-6). Degradation via bulk hydrolysis. Mechanical testing per ASTM D638-22 (tensile), ASTM D1388-23 (stiffness), ASTM D6797-24 (burst), and ISO 9073-4:2021 (tear). Biocompatibility per ISO 10993-1:2018.
Indications for Use
Indicated for soft tissue reinforcement where weakness exists, specifically as temporary support for securing muscle flaps to defect sites in muscle flap transfer procedures.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 28, 2026
TELA Bio, Inc.
John Urtz
Director - Regulatory and Quality
1 Great Valley Pkwy.
Suite 24
Malvern, Pennsylvania 19355
Re: K260474
Trade/Device Name: Liora Monofilament Scaffold
Regulation Number: 21 CFR 878.3300
Regulation Name: Surgical Mesh
Regulatory Class: Class II
Product Code: FTL
Dated: August 26, 2026
Received: August 26, 2026
Dear John Urtz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260474 - John Urtz
Page 2
may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K260474 - John Urtz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TEM
LAMICHHANE -S
Tek N. Lamichhane, Ph.D.
Assistant Director
DHT4B: Division of Plastic and
Reconstructive Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260474 | ? |
| Please provide the device trade name(s). | | ? |
| Liora Monofilament Scaffold | | |
| Please provide your Indications for Use below. | | ? |
| Liora is indicated for soft tissue reinforcement where weakness exists. It is indicated specifically in muscle flap transfer procedures as a temporary support to assist in securing muscle flaps to defect sites. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
510(k) Summary – Liora Monofilament Scaffold-K260474
# 510(k) Summary
| **Device trade name:** | Liora Monofilament Scaffold |
| --- | --- |
| **Device common name:** | Surgical Mesh |
| **Classification name:** | Mesh, surgical, polymeric |
| **Regulation number:** | 21 CFR § 878.3300 |
| **Product code:** | FTL |
| **Submitter:** | TELA Bio, Inc. 1 Great Valley Parkway Suite 24 Malvern, PA 19355 |
| **Contact person:** | John Urtz Director – Regulatory and Quality TELA Bio, Inc. jurtz@telabio.com; 484-320-2884 |
| **Date prepared:** | September 23, 2026 |
| **Predicate device:** | Polydioxanone Surgical Scaffold |
| --- | --- |
| **510(k) number:** | K181094 |
| **Regulation number:** | 21 CFR § 878.3300 |
| **Product code:** | FTL |
| **Purpose of predicate device:** | Primary predicate device utilized for comparison of intended use and technological characteristics including head-to-head preclinical and *in vitro* testing. |
| **Reference device:** | OviTex PRS (Long-Term Resorbable) |
| --- | --- |
| **510(k) number:** | K243595 |
| **Regulation number:** | 21 CFR § 878.3300 |
| **Product code:** | FTM , FTL |
| **Purpose of reference device:** | Reference device utilized for comparison of preclinical bench testing results. |
| **Reference device:** | TIGR Matrix |
| --- | --- |
| **510(k) number:** | K191749 |
| **Regulation number:** | 21CFR878.3300 |
| **Product code:** | FTL, OWT, OOD |
| **Purpose of reference device:** | Reference device utilized for purposes of head-to-head *in vivo* implantation testing. |
Page 1 of 5
{6}
510(k) Summary – Liora Monofilament Scaffold-K260474
### Device Description
Liora Monofilament Scaffold ("Liora") is a sterile resorbable monofilament scaffold composed of 100% Poly(lactic co-glycolic acid) ("PLGA"). Liora is provided in various shapes and sizes to suit surgeon preference and the nature of the soft tissue repair. The device may be trimmed to further accommodate an individual patient's requirements.
### Intended Use / Indications for Use
Liora is indicated for soft tissue reinforcement where weakness exists. It is indicated specifically in muscle flap transfer procedures as a temporary support to assist in securing muscle flaps to defect sites.
### Indications for Use Comparison
Liora has the same intended use as the predicate device, Polydioxanone Surgical Scaffold as both devices are indicated for soft tissue reinforcement where weakness exists. Liora is further specified for use in muscle flap procedures as a temporary support to assist in securing muscle flaps to defect sites.
### Summary Comparing the Technological Characteristics of the Subject and Predicate / Reference Devices
Liora has similar technological characteristics as the predicate device as both are 100% resorbable surgical meshes made from monofilament materials. Differences between the subject and predicate devices include materials of construction and physical properties such as thickness, areal density, and pore size. Liora is offered in shapes and sizes similar to the predicate device and reference devices.
Although differences exist between Liora and the predicate and reference devices, these differences do not raise different questions of safety or effectiveness. Any measured differences between the subject and predicate device are not clinically significant with respect to Liora's indications for use and are supported by preclinical bench testing performed on both Liora and the reference device(s).
A comparison of the technological characteristics is included in the table below.
Page 2 of 5
{7}
510(k) Summary – Liora Monofilament Scaffold-K260474
| | **Subject Device** Liora Monofilament Scaffold | **Predicate Device** Polydioxanone Surgical Scaffold (K181094) | **Reference Device** OviTex PRS LTR (K243595) | **Reference Device** TIGR Matrix (K191749) | **Comparison (Subject to Predicate)** |
| --- | --- | --- | --- | --- | --- |
| **Indications for Use** | Liora is indicated for soft tissue reinforcement where weakness exists. It is indicated specifically in muscle flap transfer procedures as a temporary support to assist in securing muscle flaps to defect sites. | Polydioxanone Surgical Scaffold is intended for use in reinforcement of soft tissue where weakness exists. | OviTex PRS is for implantation to reinforce soft tissue where weakness exists in patients requiring soft tissue repair or reinforcement in plastic and reconstructive surgery. The device is supplied sterile and is intended for one-time use. | The TIGR Matrix Surgical Mesh is indicated for use in the reinforcement of soft tissue, where weakness exists in patients undergoing plastic and reconstructive surgery, or for use in procedures involving soft tissue repair, such as for the repair of hernias or other fascial defects that require the addition of a reinforcing material to obtain the desired surgical result. | **Equivalent** |
| **Regulation Number** | 21 CFR § 878.3300 | 2 1CFR § 878.3300 | 21 CFR § 878.3300 | 21 CFR § 878.3300 | **Identical** |
| **Product Code** | FTL | FTL | FTM, FTL | FTL, OWT, OOD | **Identical** |
| **Materials of Construction** | Poly (lactic-co-glycolic Acid) ('PLGA') – 100% resorbable polymer | Polydioxanone ('PDO') – 100% resorbable polymer | Ovine derived collagen and Poly (lactic-co-glycolic Acid) (PLGA) resorbable polymer | Copolymer of glycolide / lactide / trimethyl carbonate (40% w/w) and copolymer of lactide / trimethyl carbonate (60% w/w) – 100% resorbable polymer | **Equivalent** |
| **Degradation Method** | Bulk hydrolysis | Bulk hydrolysis | Enzymatic and bulk hydrolysis | Bulk hydrolysis | **Identical** |
Page 3 of 5
{8}
510(k) Summary – Liora Monofilament Scaffold-K260474
| | Subject Device Liora Monofilament Scaffold | Predicate Device Polydioxanone Surgical Scaffold (K181094) | Reference Device OviTex PRS LTR (K243595) | Reference Device TIGR Matrix (K191749) | Comparison (Subject to Predicate) |
| --- | --- | --- | --- | --- | --- |
| Device Thickness (range) | 0.498mm – 0.537mm | 0.427mm – 0.431mm | 0.596mm – 0.878mm | 0.687mm – 0.713mm | Equivalent |
| Areal Density (range) | 43.0 – 54.0 g/m² | 74.2 – 76.0 g/m² | N/A | 147.8 – 152.3 g/m² | Equivalent |
| Mean Major Pore Size (Linear) | 2.8mm x 2.5mm | 2.0mm x 1.6mm | N/A | 1.0mm x 2.3mm | Equivalent |
| Mean Major Pore Size (Areal) | 4.84mm² | 1.83mm² | N/A | 1.71mm² | Equivalent |
| Shapes Offered | Rectangle, Contour, Circle | Rectangle | Rectangle, Contour, Circle, Oval | Rectangle | Equivalent |
| Maximum Surface Area (footprint) | 750cm² | 500cm² | 641cm² | 600cm² | Equivalent |
| Usage | Single use | Single use | Single use | Single use | Identical |
| Biocompatibility¹ | Established | Established | Established | Established | Identical |
| Sterilization method | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Identical |
| Sterility Assurance Level (SAL) | 10⁻⁶ | 10⁻⁶ | 10⁻⁶ | 10⁻⁶ | Identical |
¹ In accordance with ISO 10993-1:2018 – Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process
Page 4 of 5
{9}
510(k) Summary – Liora Monofilament Scaffold-K260474
## Non-clinical Performance Data
Nonclinical testing was conducted based on the FDA Guidance, Guidance for the Preparation of a Premarket Notification Application for a Surgical Mesh. The data submitted in this application includes preclinical and animal studies on biocompatibility, benchtop performance, packaging, sterilization, and shelf-life.
Biocompatibility evaluation was conducted per ISO 10993-1 with cytotoxicity, sensitization, irritation, acute systemic toxicity, genotoxicity, implantation, hemolysis, and material mediated pyrogenicity testing completed. Comprehensive chemical characterization was conducted per ISO 10993-18 and a toxicological risk assessment completed to address additional biocompatibility endpoints. The overall biological evaluation report assessed all completed testing and found that the device is biocompatible.
Benchtop testing included device characterization (thickness, areal density, and pore size), tensile strength per ASTM D638-22, device stiffness per ASTM D1388-23, suture pullout strength, burst strength per ASTM D6797-24, and tear resistance per ISO 9073-4:2021. Predicate and reference devices were tested head-to-head against the subject device. Simulated use testing was also conducted on the subject device. In vivo implantation data were also used to support the performance of the device.
Benchtop mechanical and in vitro testing per ASTM F1635-24 Standard Test Method for in vitro Degradation Testing of Hydrolytically Degradable Polymer Resins and Fabricated Forms for Surgical Implants has been conducted and shows that Polydioxanone Surgical Scaffold retains greater strength than Liora at the time of implantation through approximately two months. Throughout that timeframe, Liora has mechanical strength equivalent to the implantation strength of the reference device, OviTex PRS (Long-Term Resorbable). Liora retains its original strength for approximately three months based on in vitro degradation testing.
## Clinical Performance Data
Substantial equivalence was not based on an assessment of clinical performance data.
## Conclusion
All evaluations, including biocompatibility testing, animal testing, mechanical testing, and in vitro degradation testing support that the Liora device is substantially equivalent to the predicate device, Polydioxanone Surgical Scaffold (K181094).
Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.