← Product Code [GCJ](/productcode/GCJ) · K260414

# “Tech-Image” Video Endoscope System (K260414)

_Tech-Image CO., LTD. · GCJ · Jun 10, 2026 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K260414

## Device Facts

- **Applicant:** Tech-Image CO., LTD.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jun 10, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The “Tech-Image” Video Endoscope System is intended to be used as a single-use endoscopic video camera by qualified physicians to provide real-time illumination and visualization of internal anatomical structures during various general surgical and minimally invasive endoscopic procedures in diagnostic and therapeutic interventions.

## Device Story

Integrated digital imaging platform; consists of 13.3-inch Full-HD LCD console (CYTS-02) and TISE Series single-use video endoscopes; utilizes Chip-on-Tip CMOS technology with distal LED illumination; captures high-resolution imaging data; supports external video output via HDMI. Used in OR/clinic by physicians for real-time visualization of internal anatomy during minimally invasive procedures (laparoscopy, arthroscopy). Single-use architecture eliminates reprocessing/cross-contamination risks. Output displayed on console for surgical guidance; assists clinical decision-making by providing clear anatomical visualization.

## Clinical Evidence

Bench testing only. Evaluations included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), optical/thermal safety (IEC 60601-2-18), and biocompatibility (ISO 10993-1). Performance testing compared resolution, SNR, distortion, FOV, and DOF against predicate. Mechanical integrity tested for larger diameter models. All results met predetermined acceptance criteria.

## Technological Characteristics

Integrated digital imaging system; Chip-on-Tip CMOS sensor; distal LED illumination. Materials: SUS 304 stainless steel (ASTM A312), silicone elastomer, borosilicate glass. Dimensions: 4.7–10.0 mm OD; 218–273 mm length. Connectivity: HDMI output. Sterilization: Ethylene Oxide (EO) per ISO 11135-1. Software: IEC 62304 compliant.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- Arthrex NanoScope System ([K201134](/device/K201134.md))
- Richard Wolf PANOVIEW Ultra Telescopes ([K203226](/device/K203226.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

**FDA** **U.S. FOOD & DRUG**  
ADMINISTRATION

Tech-Image CO., LTD.  
% Yin-Jiun Tseng  
CEO & Co-Founder  
WeMED Bio-Tech, Inc.  
Contact Address

June 10, 2026

Re: K260414

Trade/Device Name: "Tech-Image" Video Endoscope System

Regulation Number: 21 CFR 876.1500

Regulation Name: Endoscope And Accessories

Regulatory Class: Class II

Product Code: GCJ, HRX

Dated: May 8, 2026

Received: May 11, 2026

Dear Yin-Jiun Tseng:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

U.S. Food & Drug Administration  
10903 New Hampshire Avenue  
Silver Spring, MD 20993  
www.fda.gov

{1}

K260414 - Yin-Jiun Tseng

Page 2

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

{2}

K260414 - Yin-Jiun Tseng

Page 3

Colin K. Chen -S
Digitally signed by
Colin K. Chen -S
Date: 2026.06.10
16:44:50 -04'00'

Colin K. Chen, Ph.D.
Acting Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

K260414

Page 1 of 1

# Indications for Use

|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260414  |
| --- | --- |
|  Please provide the device trade name(s).  |   |
|  “Tech-Image” Video Endoscope System  |   |
|  Please provide your Indications for Use below.  |   |
|  The “Tech-Image” Video Endoscope System is intended to be used as a single-use endoscopic video camera to provide visualization of the interior of the body’s internal anatomy during various diagnostic and surgical endoscopic procedures. These applications include, but are not limited to: general surgery and minimally invasive surgery (e.g., orthopedic, spine, laparoscopic, and plastic surgical procedures).  |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |

{4}

K260414

Page 1 of 8

# 510(k) Summary

Tech-Image Video Endoscope System

U.S. FOOD AND DRUG ADMINISTRATION

Center for Devices and Radiological Health (CDRH)

Premarket Notification 510(k) Summary

## SECTION I: GENERAL INFORMATION

### A. Applicant Information

*Tech-Image CO., LTD.*

*11 F., No. 37, Sec. 3, Minquan E. Rd., Zhongshan Dist., Taipei City 104478, Taiwan*

*Contact: Ms. Karen Lu, Ph.D.*

*Phone: +886 2 85113386*

*Email: karen@chiyitech.com*

*Correspondent to*

*WeMED Bio-Tech Inc.*

*Contact: Mr. Yin-Jiun Tseng, Ph.D.*

*Phone: +886 2 26974518*

*Email: courant@wemed-biotech.com*

### B. Device Information

|  Item | Details  |
| --- | --- |
|  **510(k) Submitter** | Tech-Image CO., LTD.  |
|  **Device Trade Name** | “Tech-Image” Video Endoscope System  |
|  **Device Common Name** | Endoscope and accessories  |
|  **Product Code** | GCJ (21 CFR 876.1500)  |
|  **Subsequent Product Code** | HRX (21 CFR 888.1100)  |
|  **Classification** | Class II  |
|  **Review Panel** | General & Plastic Surgery  |
|  **Regulation Medical Specialty** | Gastroenterology/Urology  |
|  **Submission Type** | Traditional 510(k)  |
|  **Predicate Device** | K201134  |
|  **Secondary Predicate** | K203226  |

{5}

K260414

Page 2 of 8

## SECTION II: INTENDED USE

### A. Intended Use Statement

The “Tech-Image” Video Endoscope System is intended to be used as a single-use endoscopic video camera by qualified physicians to provide real-time illumination and visualization of internal anatomical structures during various general surgical and minimally invasive endoscopic procedures in diagnostic and therapeutic interventions.

### B. Indications for Use

The “Tech-Image” Video Endoscope System is intended to be used as a single-use endoscopic video camera to provide visualization of the interior of the body’s internal anatomy during various diagnostic and surgical endoscopic procedures. These applications include, but are not limited to: general surgery and minimally invasive surgery (e.g., orthopedic, spine, laparoscopic, and plastic surgical procedures).

### C. Patient Population

Adult patients undergoing general or minimally invasive surgery (e.g., orthopedic, spine, laparoscopic, and plastic surgical procedures) requiring visualization of internal anatomical structures.

## SECTION III: DEVICE DESCRIPTION

### A. General Device Description

The "Tech-Image" Video Endoscope System is an integrated digital imaging platform designed for the real-time visualization of internal anatomical structures. The system consists of the CYTS-02 Video Display Device, a reusable 13.3-inch Full-HD LCD console, and the TISE Series Single-use Video Endoscopes. Utilizing advanced "Chip-on-Tip" CMOS technology and integrated LED illumination, the system captures high-resolution imaging data and supports external video output via a dedicated HDMI port. The TISE endoscopes are available in various configurations—including different diameters, viewing angles, and working lengths—to accommodate diverse clinical requirements. When used in conjunction with their corresponding sterile cannula accessories, the system facilitates precise access through surgically created openings during minimally invasive procedures, such as laparoscopy or arthroscopy. This single-use architecture effectively eliminates reprocessing risks and

{6}

K260414

Page 3 of 8

cross-contamination, ensuring consistent optical performance and patient safety.

### B. Manufacturing & Materials

|  Component | Material | Specification | Rationale  |
| --- | --- | --- | --- |
|  Shaft | SUS 304 Stainless Steel | ASTM A312 | Biocompatible, corrosion-resistant, surgical standard  |
|  Optical Window |  |  | High durability, optical clarity, scratch-resistant  |
|  Seals/Gaskets | Silicone elastomer | ISO 9002 synthetic sapphire | ISO 10993 biocompatible for prolonged contact  |
|  Internal Lenses | Borosilicate Glass | Optical-grade | Precision optics, chemically stable  |

### C. Sterilization

Sterilization Method: Ethylene Oxide (EO) gas per ISO 11135-1

- Sterilization Assurance Level (SAL): 10⁻⁶
- Shelf Life: 3 years when stored in protective packaging at room temperature

### D. Non-Clinical Testing Summary

A series of non-clinical evaluations were performed to verify the safety and effectiveness of the “Tech-Image” Video Endoscope System and to demonstrate substantial equivalence to the primary predicate device (K201134). The following tests were conducted:

- **Electrical Safety and Electromagnetic Compatibility (EMC):** The system was tested and found to be in compliance with the IEC 60601-1 standard for general electrical safety and the IEC 60601-1-2 standard for electromagnetic compatibility. The results confirmed that the device performs safely in its

{7}

K260414

Page 4 of 8

intended clinical environment.

- **Optical and Thermal Safety:** The distal LED illumination and imaging system were evaluated according to IEC 60601-2-18. Testing verified light output safety, color consistency, and thermal management. Measurements confirmed that the temperature of the endoscope tip remains within safe limits during continuous operation.
- **Performance Testing – Bench:** Side-by-side comparative testing was performed on the subject device and the predicate to evaluate key imaging parameters, including resolution (spatial frequency response), Signal-to-Noise Ratio (SNR), image distortion, Field of View (FOV), and Depth of Field (DOF). For models with larger outer diameters (up to 10.0mm), mechanical integrity and structural strength tests were conducted to ensure compatibility with standard surgical portals.
- **Biocompatibility:** In accordance with ISO 10993-1 and FDA guidance, the patient-contacting components of the single-use endoscopes were categorized as "Externally Communicating Devices" with limited contact (< 24 hours) with tissue, bone, and/or blood. Testing included Cytotoxicity, Sensitization, and Irritation/Intracutaneous Reactivity. All results met the predetermined acceptance criteria.

### SECTION IV: SUBSTANTIAL EQUIVALENCE & PREDICATE COMPARISON

A. Predicate Device

- 510(k) Number: K201134
- Device: Arthrex NanoScope System
- Product Code / Regulation: GCJ / 21 CFR 876.1500 (Primary) and HRX / 21 CFR 888.1100 (Subsequent)
- Clearance Date: September 4, 2020
- Rationale: K201134 is designated as the primary predicate for the "Tech-Image" Video Endoscope System due to its identical product codes

{8}

K260414

Page 5 of 8

(Primary: GCJ; Subsequent: HRX) and fundamental intended use in diagnostic and surgical procedures, specifically within spinal and orthopedic applications. The system architecture is substantially equivalent, featuring a reusable video processing unit and single-use imaging components.

#### **B. Secondary Predicate Device**

- ■ 510(k) Number: K203226
- ■ Device: Richard Wolf PANOVIEW Ultra Telescopes
- ■ Product Code / Regulation: GCM / 21 CFR 876.1500 (Primary) and GCJ / 21 CFR 876.1500 (Subsequent) / HET 21 CFR 884.1720
- ■ Clearance Date: May 14, 2021
- ■ Rationale: K203226 is designated as the secondary predicate for the “Tech-Image” Video Endoscope System to address a specific dimensional characteristic of the subject device — the outer diameter (OD) — that extends beyond the dimensional range of the primary predicate (K201134, OD: 1.9 mm). Both the subject device and K203226 share the same intended use (visualization of internal anatomical structures during minimally invasive endoscopic procedures), the same regulatory classification (Class II, 21 CFR 876.1500), and the same medical specialty (Gastroenterology & Urology). The OD range of the subject device (4.7–10.0 mm) falls entirely within the cleared OD of K203226 (10.0 mm). Therefore, the use of K203226 as a secondary predicate supports the conclusion that the subject device's dimensional specifications do not raise new questions of safety and effectiveness, and that the subject device is substantially equivalent to legally marketed predicates.

{9}

K260414

Page 6 of 8

# C. Substantial Equivalence Comparison Table

|  Features | Subject Device | Primary Predicate | Secondary Predicate | Result  |
| --- | --- | --- | --- | --- |
|  Device Trade Name | Tech-Image™ Video Endoscope System | Arthrex NanoScope System | Richard Wolf PANOVIEW Ultra Telescopes |   |
|  510 (k) |  | K201134 | K203226 |   |
|  Classification | Class II | Class II | Class II | Identical  |
|  Medical Specialty | Gastroenterology & Urology | Gastroenterology & Urology | Gastroenterology & Urology | Identical  |
|  Primary Product Code | GCJ | GCJ | GCM | Same as Predicate  |
|  Substantial Product Code | HRX | HRX | GCJ HET | Same as Predicate  |
|  Regulation No. | 21 CFR 876.1500 Endoscope and Accessories | 21 CFR 876.1500 Endoscope and Accessories | 21 CFR 876.1500 Endoscope and Accessories | Identical  |
|  Review Panel | General & Plastic Surgery | General & Plastic Surgery | General & Plastic Surgery | Identical  |
|  Intended Use | The 'Tech-Image' Video Endoscope System is intended to be used as a single-use endoscopic video camera by qualified physicians to provide real-time illumination and visualization of internal anatomical structures during various general surgical and minimally invasive endoscopic procedures in diagnostic and therapeutic interventions. | The Arthrex NanoScope System provides image processing and digital documentation for endoscopic procedures. The system utilizes a handpiece which provides distal LED illumination to the surgical site using a fiber optic bundle surrounding a high-resolution camera sensor. | The PANOVIEW Ultra telescopes (common name: telescopes) are rigid endoscopes which are used in invasive surgery or via natural orifices for diagnosis and therapy and serve to visualize the inside of the body. The products are used for interventions in different medical disciplines, such as surgery and laparoscopic interventions e.g. in urology and gynecology. The submitted devices can be grouped by working length (WL) and direction of view. | Similar  |
|  Indications for Use | The 'Tech-Image' Video Endoscope System is intended to be used as a single-use endoscopic video camera to provide visualization of the interior of the body's internal anatomy during various diagnostic and surgical endoscopic procedures. These applications include, but are not limited to: general surgery and minimally invasive surgery (e.g., orthopedic, spine, laparoscopic, and plastic surgical procedures). | The Arthrex NanoScope System is intended to be used as an endoscopic video camera in a variety of endoscopic diagnostic and surgical procedures, including but not limited to: orthopedic, spine, laparoscopic, urologic, sinuscopic, and plastic surgical procedures. The device is also intended to be used as an accessory for microscopic surgery. | The products are used in invasive surgery or via natural orifices for diagnosis and therapy and serve to visualize the inside of the body. The products are used for interventions in different medical disciplines, such as surgery and laparoscopic interventions e.g. in urology and gynecology. | Similar  |
|  System Configuration | Single-Use Endoscope + Reusable control unit | Single-Use Endoscope + Reusable control unit (CCU/Console) | Reusable rigid endoscope telescope + external light source + external camera | Same System Components Subject & Primary Predicate are single-use  |
|  Biocompatibility | ISO 10993-1 compliant | ISO 10993-1 compliant | ISO 10993-1 compliant | Identical  |
|  Electrical Safety | IEC 60601-1 | IEC 60601-1 | N/A (passive optical device) | Identical  |
|  EMC | IEC 60601-1-2 | IEC 60601-1-2 | N/A (passive optical device) | Identical  |
|  Thermal Safety | IEC 60601-2-18 | IEC 60601-2-18 | IEC 60601-2-18 | Equivalent  |
|  Sterilization | EO | EO | Steam sterilization (moist heat) | Subject & Primary Predicate are EO  |
|  Software Validation | IEC 62304 |  | No software | N/A  |

{10}

K260414

Page 7 of 8

|  Human Factors / Usability | IEC 62366-1 | Per IEC 60601-1-6 | N/A | Similar to Predicate  |
| --- | --- | --- | --- | --- |
|  Performance Testing | ISO 8600 | Bench Testing | ISO 8600-1/-3/-5 | Equivalent  |
|  OD(Diameter) | 4.7 - 8.6 - 10.0 mm | Scope diameter 1.9mm; utilizes accessory scope sheath of up to 3.4 mm diameter | 10.0 mm | Equivalent  |
|  Working Length | 218 - 273 mm | 125, 180, 250 mm | 305, 440 mm | Similar  |
|  FOV | 130° | 120° | 0°, 30°, 50° | Similar  |
|  Material | Stainless steel + ABS + PEEK... | - | Optical glass + stainless steel | Similar  |
|  Imaging Sensor | High-Res CMOS | CMOS | Rod lens optical system | Similar  |
|  Sensor Location | Distal Tip (Chip-on-Tip) | Distal Tip | N/A (not chip-on-tip) | Similar  |
|  Resolution | 1080p | Standard/High Res | Optical eyepiece / monitor via attached camera | Similar  |
|  Light Source | Distal LED | Distal LED | External cold light source (fiber bundle) | Similar  |
|  Working Channel | Integrated (Optional) | N/A | N/A | Similar  |
|  Energy Source | AC Mains (via CCU) | AC Mains | External light source (passive device) | Similar  |

{11}

K260414

Page 8 of 8

# **D. Substantial Equivalence Discussion**

The collective data from the identified predicate devices establish a robust basis for the substantial equivalence of the “Tech-Image” Video Endoscope System. Detailed side-by-side comparisons and supporting performance data are documented comprehensively and we affirm that the subject device is as safe and effective as the legally marketed devices, raising no new questions of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/device/K260414](https://fda.innolitics.com/device/K260414)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
