The MONARCH™ Bronchoscope and the MONARCH™ Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
Device Story
The MONARCH™ Platform is a robotic, electromechanical system for bronchoscopy. It consists of a cart with robotic arms, a tower with a display and computer, a flexible bronchoscope with a distal camera, and a controller. The physician operator uses the controller to manipulate the bronchoscope, which provides continuous video visualization. The system integrates pre-operative CT scans with electromagnetic (EM) navigation to track the bronchoscope tip relative to patient anatomy. The platform can optionally interface with Cone Beam CT (CBCT) systems to receive real-time imaging updates, correcting for anatomical changes during procedures. AI/ML algorithms assist in bronchoscope tip localization, 3D airway segmentation, and nodule segmentation. The device is used in clinical settings to assist physicians in navigating to target anatomy for diagnostic or therapeutic interventions, potentially improving procedural accuracy and access to lung nodules.
Clinical Evidence
No clinical data was required. Substantial equivalence was demonstrated through non-clinical performance testing, including electrical verification, location accuracy tests, software functional testing, simulated use accuracy, and pre-clinical cadaver studies to assess workflow and clinical accuracy.
Technological Characteristics
Electromechanical robotic system; EM navigation; flexible bronchoscope with distal camera; pull-wire articulation. Connectivity includes Ethernet for CBCT integration. Software includes AI/ML-based functions for tip localization, airway segmentation, and nodule segmentation. System components: cart, tower, controller, and accessories. Sterilization and material specs are consistent with the predicate.
Indications for Use
Indicated for bronchoscopic visualization and access to patient airways for diagnostic and therapeutic procedures in adults (22 years and older).
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
July 25, 2026
Auris Health, Inc.
Shikha Dallas
Director, Regulatory Affairs
5490 Great America Pkwy.
Santa Clara, California 95054
Re: K260382
Trade/Device Name: MONARCH™ Platform (MON-000008)
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories
Regulatory Class: Class II
Product Code: EOQ
Dated: March 13, 2026
Received: March 13, 2026
Dear Shikha Dallas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260382 - Shikha Dallas
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260382 - Shikha Dallas
Page 3
Sincerely,
JOYCE C.
LIN -S
for Shu-Chen Peng, Ph.D.
Assistant Director
DHT1B: Division of Dental and
ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260382 | ? |
| Please provide the device trade name(s). | | ? |
| MONARCH™ Platform (MON-000008) | | |
| Please provide your Indications for Use below. | | ? |
| The MONARCH™ Bronchoscope and the Monarch Platform™ and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{5}
510(k) Summary – K260328
# [807.92(a)(1)] Submitter Information
| Applicant: | Johnson & Johnson MedTech Auris Health, Inc. 5490 Great America Parkway Santa Clara, CA 95054 USA Telephone: 1-800-434-0032 USA 1-650-264-6000 Worldwide |
| --- | --- |
| Contact Person: | Shikha Dallas Johnson & Johnson MedTech Auris Health, Inc. 5490 Great America Parkway Santa Clara, CA 95054 USA Telephone: 425-236-4604 |
| Date Summary Prepared: | July 24, 2026 |
| --- | --- |
# [807.92(a)(2)] Name of Device
| Device Trade Name: | MONARCH™ Platform |
| --- | --- |
| Device Common Name: | Bronchoscope (Flexible or Rigid) |
| Device Classification: | Class II, 21 CFR 874.4680 |
| Product Code | EOQ |
| Subsequent Product Code | JAK |
# [807.92(a)(3)] Legally Marketed Devices
| Predicate Device: | MONARCH™ Platform 510(k) #: K243219 |
| --- | --- |
# [807.92(a)(4)] Device Description
| Device Description: | The MONARCH™ Platform is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The MONARCH™ Platform enables robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician operator. |
| --- | --- |
The MONARCH™ Platform allows for precise access of the lung anatomy and continuous visualization using the bronchoscope distal tip camera. The MONARCH™ Platform consists of four major components, (1) MONARCH™ Cart, (2) MONARCH™ Tower, (3) MONARCH™ Bronchoscope, (4) MONARCH™ Controller and working channel instruments and accessories. The MONARCH™ Cart provides support for the effector arms. It includes up to three robotic arms and the electronic systems required to power and operate the robotic system. The robotic arms possess multiple degrees of freedom. The MONARCH™ Tower is the primary user's (i.e. physician) procedural display interface. It contains a monitor for user viewing and computers running the system software. The tower provides connectivity for
1
{6}
510(k) Summary – K260328
the bronchoscope camera. The user controls the system with a controller which transmits user inputs through the electromechanical system to the bronchoscope. The flexible MONARCH™ Bronchoscope has a working channel and a camera at the distal end. Additionally, the MONARCH™ Platform includes electromagnetic (EM) navigation and integrates a pre-operative computed tomography (CT) scan into an intra-operative interface, displaying the modeled bronchoscope tip location relative to the pre-operative scan anatomy. The MONARCH™ Platform is being updated to receive 3D imaging data via an ethernet connection directly from additional models of Cone Beam CT (CBCT) Systems with 3D imaging technology. This will enable real time navigation updates during a procedure with the device. This integration feature is optional for the user, but if selected, use of the MONARCH™ Window Field Generator is required. The modified device includes software updates for navigation and presentation of the anatomy for display and planning purposes, an updated USB stick, and software revisions for the purposes of cybersecurity.
# **[807.92(a)(5)] Intended Use**
# **Indications for Use:**
The MONARCH™ Bronchoscope and the MONARCH™ Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures.
# **Difference in Indications from Predicate Device:**
The intended use of the device is unchanged from the Predicate.
# **[807.92(a)(6)] Technical Characteristics**
# **Technological Characteristics:**
The substantial equivalence of the MONARCH™ Platform to the Predicates is shown by similarity in intended use, indications for use, and performance. The Subject device and the Predicate device have similar technological characteristics with respect to principles of operation, system design, and performance. The subject and Predicate device achieve the same outcomes.
The main design difference of the Subject device is the ability to directly connect to additional models Cone Beam Computed Tomography systems for the purpose of receiving imaging data during a procedure. Performance testing on the Subject device and comparison to the predicate device support that there are no new questions of safety or effectiveness.
Like the Predicate MONARCH™ Platform, the modified MONARCH™ Platform is an electromagnetic and image-guided navigation system intended for use during robotically assisted bronchoscopy procedures.
Physicians can track and display the real-time location of the
2
{7}
510(k) Summary – K260328
bronchoscope relative to pre-acquired CT images, while simultaneously receiving a live and continuous video image from the bronchoscope, and navigational displays on the device monitor. The modified MONARCH™ Platform may be optionally interfaced to Cone Beam Computed Tomography (CBCT) systems to receive near-real time imaging updates during a procedure. The Subject device of this submission has the same technological characteristics as the Predicate except for updated software and new hardware to facilitate communications with a Cone Beam Computed Tomography system.
Like the Predicate, the MONARCH™ Platform utilizes electromagnetic tracking technology for navigation, uses anatomical reference points on the patient's anatomy for intraoperative registration to the image-based model of the anatomy, and uses CT image sets as reference images for the image-based model. The instructions for use and product labels have been updated to reflect the revised device.
# Artificial Intelligence / Machine Learning Information
Like the predicate MONARCH™ Platform, K243219, the proposed device employs Machine Learning in the development of certain Device Software Functions which are considered to be Artificial Intelligence. The AI software in the MONARCH™ Platform, Bronch 4.2 uses Machine Learning to train the following static Device Software Functions.
- Bronchoscope tip localization - The Navigation software includes a vision-based AI/ML algorithm for estimating the position and orientation of the scope tip.
- 3D Airway Model Segmentation - The Pre-Operative Planning software uses an AI/ML algorithm for generating a 3D model of the patient lung airways and displays it to the user.
- Nodule Segmentation - The Pre-Operative Planning software utilizes an AI/ML algorithm to generate a 3D model (representation) and boundary surface of the nodule within the CT scan based on the user-selected location.
The remaining Device Software Functions in the proposed device do not use AI/ML functions.
The AI/ML functions in the MONARCH™ Platform are not user selectable, and do not require software inputs from other software sources. The physician user makes all decisions to the setup of the room, planning, and navigation of the patient's airways. These Device software functions assist users to navigate of the
3
{8}
510(k) Summary – K260328
bronchoscope to the desired anatomy, as part of its fulfillment of the devices, intended use, to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedure
Refer to table below for a Comparison of Technological Characteristics with the Predicate device.
4
{9}
510(k) Summary – K260328
| Key Attributes | Predicate Device Auris Health, Inc. MONARCH™ Platform K243219 | Subject Device Auris Health, Inc. MONARCH™ Platform | Comparison |
| --- | --- | --- | --- |
| Regulatory Classification Information | Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ | Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOQ | Same. The Predicate device and Subject device have the same regulatory classification information. |
| Device Description | The MONARCH™ Platform is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The MONARCH™ Platform enables robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician and allows for precise access of the lung anatomy and continuous visualization using the bronchoscope distal tip camera, under the control of a physician operator. Additionally, the MONARCH™ Platform can integrate imaging data directly from a Cone Beam Computed Tomography system to update the scope and target position to overcome any changes in anatomy not reflected in the pre-op CT scan. | The MONARCH™ Platform is intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. The MONARCH™ Platform enables robotic electro-mechanical articulation and precise control of a flexible bronchoscope under continuous and direct control by a physician and allows for precise access of the lung anatomy and continuous visualization using the bronchoscope distal tip camera, under the control of a physician operator. Additionally, the MONARCH™ Platform can integrate imaging data directly from a Cone Beam Computed Tomography system to update the scope and target position to overcome any changes in anatomy not reflected in the pre-op CT scan. | Equivalent. Compared to the Predicate Device, the Subject Device has been equipped with updated software to optionally enable receipt of imaging data directly from Siemens Cone Beam CT Systems via NaviLink interface, to update the scope and target position to overcome changes in anatomy not reflected in the pre-op CT scan. The predicate device incorporated a similar interface with GE OEC 3D systems. Additional changes were made to update software relating to navigation, nodule segmentation, and to provide cybersecurity updates. The difference do not raise new questions of safety or efficacy. |
| Intended use / Indications for use | The MONARCH Bronchoscope and the MONARCH™ Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. | The MONARCH Bronchoscope and the MONARCH™ Platform and its accessories are intended to provide bronchoscopic visualization of and access to patient airways for diagnostic and therapeutic procedures. | The predicate and subject device have identical Intended use / Indications for Use statements. |
5
{10}
510(k) Summary – K260328
| Key Attributes | Predicate Device Auris Health, Inc. MONARCH™ Platform K243219 | Subject Device Auris Health, Inc. MONARCH™ Platform | Comparison |
| --- | --- | --- | --- |
| Fundamental scientific technology | Electromechanical system using direct (camera and light source) supplemented by indirect (electromagnetic navigation) visualization of the flexible bronchoscope position by the physician operator. Steering, as well as insertion and withdrawal is under continuous and direct control by the physician operator. | Electromechanical system using direct (camera and light source) supplemented by indirect (electromagnetic navigation) visualization of the flexible bronchoscope position by the physician operator. Steering, as well as insertion and withdrawal is under continuous and direct control by the physician operator. | Same. The Predicate and Subject devices have the same fundamental scientific technology. |
| Primary mode of action | Physician input (movement of endoscopic controller) to move the endoscope. | Physician input (movement of endoscopic controller) to move the endoscope. | Same. The primary mode of action is unchanged between the Predicate and Subject device. |
| Method of distal tip movement | Pull wires | Pull wires | Same. The method of distal tip movement is unchanged between the Predicate and Subject device. |
| Major components | Auris Cart Auris Tower Bronchoscope and Sheath Accessories (working channel instruments, patient introducer adaptor, fluidics tubing, navigation field generators, field generator mount, USB stick, patient sensors, patient patches, etc.). | Auris Cart Auris Tower Bronchoscope and Sheath Accessories (working channel instruments, patient introducer adaptor, fluidics tubing, navigation field generators, field generator mount, USB stick, patient sensors, patient patches, etc.). | Same. The Predicate and Subject devices are comprised as the same major components. |
| Control/physician interfaces | Controller with joystick and buttons. | Controller with joystick and buttons. | Same. The Predicate and Subject devices feature the same Control/physician interfaces. |
| Cart base elements | Touch screen monitor and embedded PC. | Touch screen monitor and embedded PC. | Equivalent. The Predicate and Subject devices have the same cart base elements; however, the modified device includes updated software. |
6
{11}
510(k) Summary – K260328
| Key Attributes | Predicate Device Auris Health, Inc. MONARCH™ Platform K243219 | Subject Device Auris Health, Inc. MONARCH™ Platform | Comparison |
| --- | --- | --- | --- |
| Robotic arms | Two identical arms (2). | Two identical arms (2). | Same. The Predicate and Subject device have the same number of arms. |
| Tower key characteristics | Touch screen monitor, on-board computers, fluidics pump, power supply. | Touch screen monitor, on-board computers, fluidics pump, power supply. | Equivalent. The tower in the modified device includes updated software. |
| Electromagnetic Navigation | Yes | Yes | Same. The Predicate and Subject Devices both include Electromagnetic Navigation. |
| Working channel Instruments | Aspirating biopsy needles, Biopsy Forceps, Cytology Brush. Third party compatible (meet working channel length and diameter requirements) instruments and tools (e.g., RBUS probe) | Aspirating biopsy needles, Biopsy Forceps, Cytology Brush. Third party compatible (meet working channel length and diameter requirements) instruments and tools (e.g., RBUS probe). | Same. The Predicate and Subject Devices are both compatible with the same working channel accessories. |
| Navigation and segmentation | - Bronchoscope tip localization - 3D Airway Model Segmentation - Nodule Segmentation | - Bronchoscope tip localization - 3D Airway Model Segmentation - Nodule Segmentation | Equivalent. Both the subject device and predicate device provide nodule segmentation, however the subject device allows the user to identify a nodule via a single-click, or optionally to select the target by dragging across the displayed image. The revised functionality employs a static algorithm based on annotated images. Both the subject device and predicate device have a physician defined navigation path, however the subject device has a modification that is limited to path smoothing and does not affect the safety or performance of the feature. The change does not raise new questions of safety and efficacy. |
7
{12}
510(k) Summary – K260328
| Key Attributes | Predicate Device Auris Health, Inc. MONARCH™ Platform K243219 | Subject Device Auris Health, Inc. MONARCH™ Platform | Comparison |
| --- | --- | --- | --- |
| Anatomical position indicator | Stylized humanoid figure | Stylized humanoid figure & optional “compass” overlay | Equivalent. The subject device offers an optional compass overly for assisting users in confirming anatomical position and direction of view. The change does not raise new questions of safety and efficacy as both methods confirm orientation. |
| Direct Interface with CBCT system to reduce potential body divergence. | Implemented. | Implemented. | Equivalent. Compared to the predicate, the subject device can integrate imaging data directly from additional models of Cone Beam Computed Tomography system during a procedure to update the scope and target position to overcome changes in anatomy not reflected in the pre-op CT scan. |
8
{13}
510(k) Summary – K260328
**Non-clinical Performance Data:** The MONARCH™ Platform was tested to ensure that it functions in accordance with the system design specifications related to substantial equivalence in terms of device safety and effectiveness.
The following nonclinical tests were performed:
1. Proof of Design electrical tests, to verify all hardware modules perform within specifications.
2. Location Accuracy tests
3. Software functional tests, covering the complete system functionality, and including error handling, usability and time
4. Simulated use accuracy test, in which a complete CT image registration and instrument navigation workflow was performed, to verify the overall accuracy of the system.
5. Pre-clinical (cadaver) tests were designed to mimic surgical procedures using the MONARCH™ Platform in a simulated clinical environment, to assess the execution of a complete robotically assisted bronchoscopy procedure workflow and to qualitatively estimate the system clinical accuracy.
6. The proposed MONARCH™ Platform passed all tests in accordance with appropriate test criteria and standards, and the modified device did not raise new questions of safety or effectiveness.
**Clinical Performance Data:**
Clinical data was not necessary to determine that the MONARCH™ Platform was substantially equivalent to the Predicate device. The performance data demonstrated that the device performs as intended.
**Conclusion:**
The modified MONARCH™ Platform is substantially equivalent to the currently cleared MONARCH™ Platform based on the completion of non-clinical bench testing as well as similar principles of design, operation and indications for use.
9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.