Dynamis Robotic Surgical System

K260369 · LEM Surgical AG · OLO · Jun 20, 2026 · Neurology

Device Facts

Record IDK260369
Device NameDynamis Robotic Surgical System
ApplicantLEM Surgical AG
Product CodeOLO · Neurology
Decision DateJun 20, 2026
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.4560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Dynamis Robotic Surgical System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding qualified surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on intraoperative CT scans. The Dynamis Robotic Surgical System is indicated for the placement of spinal pedicle screws.

Device Story

Integrated navigation-based robotic platform for spine surgery; utilizes intraoperative CT images (DICOM) for planning. System comprises two robotic arms for guidance and one arm for an optical navigation camera, all mounted on a single cart. Surgeons use a 'Dynacuff' (squeeze bracelet) and foot pedal for force-controlled robotic arm movement. System registers virtual patient images to physical anatomy using bone-mounted markers. Provides visual assistance for robotic alignment of instruments or freehand navigation. Supports individual vertebra tracking and stabilization workflows to manage vertebral motion. Used in OR by surgeons; output displayed on system screen to guide instrument placement. Benefits include precise anatomical localization and trajectory guidance for spinal instrumentation.

Clinical Evidence

Bench testing only. Includes navigation accuracy verification per ASTM F2554-22, quantitative system-level accuracy validation in a cadaveric model, software verification/validation (IEC 62304), cybersecurity testing, electromagnetic/electrical safety (IEC 60601 series), biocompatibility (ISO 10993), and cleaning/sterilization validation. Human factors/usability testing performed per ANSI AAMI IEC 62366-1.

Technological Characteristics

Multi-arm robotic platform on mobile cart; optical tracking camera; force-controlled movement via Dynacuff/foot pedal. Materials biocompatibility per ISO 10993. Connectivity via USB/LAN for DICOM data. Software-based planning, registration, and navigation. Accuracy: 1.5mm/2 degrees (robotic), 2mm/2 degrees (freehand). Sterilization: user-performed per instructions. Standards: ASTM F2554-22, IEC 62304, IEC 60601-1, ISO 13485.

Indications for Use

Indicated for patients requiring spinal pedicle screw placement in open or percutaneous procedures, including cervical, thoracic, lumbar, and sacral approaches.

Regulatory Classification

Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

Predicate Devices

Submission Summary (Full Text)

{0} **FDA** **U.S. FOOD & DRUG** ADMINISTRATION June 20, 2026 LEM Surgical AG % Alexia Haralambous Principal Consultant Harmonia Insights, LLC 6800 Wisconsin Ave. #1280 Chevy Chase, Maryland 20815 Re: K260369 Trade/Device Name: Dynamis Robotic Surgical System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: June 16, 2026 Received: June 16, 2026 Dear Alexia Haralambous: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260369 - Alexia Haralambous Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K260369 - Alexia Haralambous Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} # Indications for Use Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. K260369 ? Please provide the device trade name(s). ? Dynamis Robotic Surgical System Please provide your Indications for Use below. ? The Dynamis Robotic Surgical System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding qualified surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on intraoperative CT scans. The Dynamis Robotic Surgical System is indicated for the placement of spinal pedicle screws. Please select the types of uses (select one or both, as applicable). ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) ? Page 1 of 1. {4} Dynamis Robotic Surgical System # **510(K) SUMMARY - K260369** # **DATE PREPARED** February 4, 2026 # **MANUFACTURER AND 510(K) OWNER** LEM Surgical AG Morgenstrasse 136 3018 Bern, Switzerland Telephone: +41 31 382 30 00 Official Contact: Yossi Bar, Chief Executive Officer # **REPRESENTATIVE/CONSULTANT** Alexia Haralambous, MS, RAC Principal Consultant Email: alexia@harmoniainsights.com # **DEVICE INFORMATION** Proprietary Name/Trade Name: Dynamis Robotic Surgical System Common Name: Orthopedic Stereotaxic Instrument Regulation Number: 21 CFR 882.4560 Class: II Product Code: OLO Premarket Review: CDRH/OPEQ/OHT6/DHT6C Review Panel: Orthopedic # **PREDICATE DEVICE IDENTIFICATION** The Dynamis Robotic Surgical System is substantially equivalent to the following predicate device: | *510(k) Number* | *Predicate Device Name / Manufacturer* | *Primary Predicate* | | --- | --- | --- | | K243326 | Dynamis Robotic Surgical System / LEM Surgical AG | ☑ | # **DEVICE DESCRIPTION** The Dynamis System is an integrated navigation-based robotic platform with real-time tracking capability for spine surgical procedures that include cervical, thoracic, lumbar, and sacral approaches. The system was previously cleared under K243326. The purpose of this submission is to introduce modifications to the previously cleared Dynamis System to expand navigation capabilities, enhance vertebra tracking, and add workflow improvements, while maintaining the same intended use and fundamental scientific technology. The Dynamis System is comprised of two computer-controlled robotic arms to support surgical robotic guidance, while a third robotic arm holds and controls the Scout navigation camera. All components are integrated into one physical cart located partially underneath the surgical table. The cart employs bilateral elevator mechanisms to deploy and fold the surgical robotic arms and secure them. This system functions in conjunction with different intraoperative imaging systems by data transfer of DICOM format images by USB and/or Direct LAN transfer for intraoperative K260369, Page 1 of 5 {5} Dynamis Robotic Surgical System planning. The system software is responsible for all motion control, navigation, data storage, user management, case management, and safety functions. The system establishes registration between the virtual patient (points on the patient DICOM images) and the physical patient (corresponding to the patient’s anatomy). The information of the plan, coupled with the registration, provides the necessary information to give visual assistance to the surgeon during the robotic alignment of qualified instruments or during freehand navigation. Qualified instruments that are rigid, straight, round, and concentric are permitted for use through the Dynacan or used with the Dynatracker (i.e., navigation tracker attached to hand-held instrument), given they pass the Instrument Setup process. The system also supports individual vertebra tracking to maintain alignment to vertebra specific trajectories by monitoring relative vertebral motion, as well as a stabilization workflow which provides additional control of vertebral motion during instrumentation. The instruments included in the system include a registration marker, patient reference markers (Patient Marker and Vertebra Markers), surgical instruments, Dynacans (end effectors), and Dynatrackers (Universal Markers). The instruments are reusable and supplied non-sterile to the user. The user is responsible for cleaning and sterilization of the instruments per the provided instructions. ## INDICATIONS FOR USE The Dynamis Robotic Surgical System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding qualified surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on intraoperative CT scans. The Dynamis Robotic Surgical System is indicated for the placement of spinal pedicle screws. ## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS The subject Dynamis System includes modifications to the previously cleared Dynamis System, including expanded freehand navigation support, individual vertebra tracking using additional bone-mounted markers, stabilization workflow options, introduction of cervical indications, and user interface refinements to support these capabilities. These modifications do not introduce new fundamental scientific technology. The subject device continues to rely on the same core architecture and principles as the predicate Dynamis System, including a multi-arm robotic platform integrated into a single cart, an optical tracking camera, bone-mounted reference markers, end effector-based instrument tracking, software-based planning, registration, navigation, and robotic alignment, force-controlled arm movement via the Dynacuff and foot pedal, and the same underlying optical tracking, robotic guidance, and DICOM-based planning framework. The proposed updates extend and refine the predicate system’s established technological characteristics without altering its fundamental scientific basis. The modifications do not raise new or different questions of safety or effectiveness. The subject K260369, Page 2 of 5 {6} Dynamis Robotic Surgical System device continues to operate within the same safety framework and uses the same tracking principles, robotic arm architecture, force-controlled movement, and navigation algorithms as the predicate device, leveraging previously cleared mechanisms for tracking, instrument control, and force detection. | | **Subject Device: Dynamis Robotic Surgical System** | **Primary Predicate: Dynamis Robotic Surgical System** | | --- | --- | --- | | **Manufacturer** | LEM Surgical AG | LEM Surgical AG | | **Submission Number** | K260369 | K243326 | | **Regulation Number** | 21 CFR 882.4560 | 21 CFR 882.4560 | | **Regulation Name** | Stereotaxic Instrument | Stereotaxic Instrument | | **Regulatory Class** | Class II | Class II | | **Product Code** | OLO | OLO | | **Indications for Use** | The Dynamis Robotic Surgical System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding qualified surgical instruments in open or percutaneous procedures, provided that the required fiducial markers and rigid patient anatomy can be identified on intraoperative CT scans. The Dynamis Robotic Surgical System is indicated for the placement of spinal pedicle screws. | The Dynamis Robotic Surgical System is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of an instrument holder or guide tube to be used by surgeons for navigating and/or guiding qualified surgical instruments in open or percutaneous procedures provided that the required fiducial markers and rigid patient anatomy can be identified on intraoperative CT scans. The Dynamis Robotic Surgical System is indicated for the placement of non-cervical spinal pedicle screws. | | **Imaging Modality** | 3D intraoperative imaging | 3D intraoperative imaging | | **Integrated Planning & System Software** | Dynamis Robotic Surgical Software | Dynamis Robotic Surgical Software | | **DICOM Compatibility & Storage** | Yes | Yes | | **Navigation system** | Optical camera | Optical camera | | **Controller** | Force-controlled movement via Foot Pedal and Squeeze Bracelet ('Dynacuff') | Force-controlled movement via Foot Pedal and Squeeze Bracelet ('Dynacuff') | | **Surgical Workflow** | Freehand and Robotic-Guided; includes optional stabilization mode. | Freehand and Robotic-Guided | | **Patient Fixation & Reference Marker Options** | Bone-mounted Patient Marker; Bone-mounted Vertebra Marker. | Bone-mounted Patient Marker. | K260369, Page 3 of 5 {7} Dynamis Robotic Surgical System | | Subject Device: Dynamis Robotic Surgical System | Primary Predicate: Dynamis Robotic Surgical System | | --- | --- | --- | | Registration & Tracking Architecture | Patient marker on registration fixture; system can track patient reference frame and update trajectories based on individually assigned vertebrae. | Patient marker on registration fixture | | Instrument Tracking/Qualification | Dynacan end effector with markers + Instrument Setup step, including support for freehand navigation instruments via Dynatracker | Dynacan end effector with markers + Instrument Setup step | | Implant/Instrument System Compatibility | End effector tracking with qualified instruments | End effector tracking with qualified instruments | | Robotic Guidance Trajectory Accuracy | Up to 1.5mm and 2 degrees | Up to 1.5mm and 2 degrees | | System Navigation Accuracy | Robotic-Guided: Up to 1.5mm and 2 degrees Freehand: Up to 2mm and 2 degrees | Robotic-Guided: Up to 1.5mm and 2 degrees | | System configuration | 2 Robotic Arms, 1 Arm containing optical camera, control screen, instruments, system software, all integrated into a single cart located partially underneath surgical table | 2 Robotic Arms, 1 Arm containing optical camera, control screen, instruments, system software, all integrated into a single cart located partially underneath surgical table | | Safety Features | Emergency stop button & system Surface mapping Collaborative Robot | Emergency stop button & system Surface mapping Collaborative Robot | K260369, Page 4 of 5 {8} Dynamis Robotic Surgical System ## **SUMMARY OF NON-CLINICAL TESTING** Verification and validation testing was conducted on the subject Dynamis System to demonstrate substantial equivalence and in accordance with standards, where applicable. The following tests were conducted: - Non-clinical system and instrument verification and validation testing - Navigation accuracy verification per ASTM F2554-22 - Quantitative system level accuracy validation in a clinically relevant model (cadaver) - Software verification and validation per IEC 62304 & IEEE/ISO/IEC 29119-1-2-3-5-2021 - Cybersecurity testing per IEC 62304:2006/AMD 1:2015, IEC 81001-5-1:2021, AAMI TIR57:2016/(R)2023 - Electromagnetic and Electrical Safety testing per IEC 60601-1:2005, IEC 60601-1-2:2014, IEC 80601-2-77:2019, IEC 60601-1-6:2010, IEC 60601-1-8: IEC 60601-1-8:2006, and IEC 60825-1:2014 - Biocompatibility testing per ISO 10993-1:2018, ISO 10993-5:2009, ISO 10993-23:2021, ISO 10993-10:2021, ISO 10993-11:2017 - Cleaning and sterilization validation per ANSI/AAMI ISO 17665-1: 2006/(R) 2013, AAMI ST98:2022, AAMI TIR12:2020, AAMI TIR30:2011/(R)2016, EN ISO 17664-1:2021 - Human factors/usability testing per ANSI AAMI IEC 62366-1:2015 & AAMI/ANSI HE75:2009/(R)2018 ## **CONCLUSION** The data demonstrate substantial equivalence of the subject Dynamis Robotic Surgical System to the predicate Dynamis Robotic Surgical System. The subject Dynamis system is substantially equivalent to the predicate device with respect to intended use, principle of operation, technological characteristics, and performance. K260369, Page 5 of 5
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...