The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). The Giraffe Incubator Carestation is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature-controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Device Story
Giraffe OmniBed/Incubator Carestations provide thermal support for neonates; operate as enclosed incubators or open radiant warmers (OmniBed only); utilize servo-controlled heating based on air or skin temperature sensors; include optional servo-controlled oxygen delivery. Used in NICU, labor/delivery, or nurseries by professional staff. Device maintains infant temperature via heated air circulation or infrared radiant heat. Output includes thermal regulation and oxygen concentration; healthcare providers monitor via touchscreen UI. Benefits include stable microenvironment for vulnerable infants. Subject device updates material components (polyester vs polyacetal) in breathing gas pathways.
Clinical Evidence
Bench testing only. No clinical data. Safety/effectiveness demonstrated via functional verification, material stress testing, and biocompatibility testing (ISO 10993, ISO 18562).
Indicated for neonates unable to thermoregulate; provides controlled thermal support via enclosed incubator or open radiant warmer; facilitates transition from uterus to external environment; provides servo-controlled oxygen (21-65%) in infant compartment.
Regulatory Classification
Identification
A neonatal incubator is a device consisting of a rigid boxlike enclosure in which an infant may be kept in a controlled environment for medical care. The device may include an AC-powered heater, a fan to circulate the warmed air, a container for water to add humidity, a control valve through which oxygen may be added, and access ports for nursing care.
{0}
**FDA U.S. FOOD & DRUG**
ADMINISTRATION
July 28, 2026
Datex-Ohmeda, Inc.
Isabel McGann
Director, Regulatory Affairs, Maternal Infant Care
3114 N Grandview Blvd.
Waukesha, Wisconsin 53188
Re: K260326
Trade/Device Name: Giraffe™ OmniBed™ Carestation (CS1); Giraffe™ Incubator Carestation (CS1)
Regulation Number: 21 CFR 880.5400
Regulation Name: Neonatal incubator
Regulatory Class: Class II
Product Code: FMZ, FMT
Dated: June 30, 2026
Received: June 30, 2026
Dear Isabel McGann:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260326 - Isabel McGann
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K260326 - Isabel McGann
Page 3
Sincerely,
DAVID WOLLOSCHECK -S
David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260326
Device Name
Giraffe™ OmniBed™ Carestation (CS1);
Giraffe™ Incubator Carestation (CS1)
Indications for Use (Describe)
The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology.
Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment.
This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
The Giraffe Incubator Carestation is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature-controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
# 510(k) Summary – K260326
In accordance with 21 CFR 807.92 the following summary information is provided:
| Date: | 23 July 2026 |
| --- | --- |
| Owner/Submitter: | GE HealthCare (dba Datex- Ohmeda, Inc.) 3114 N. Grandview Blvd Waukesha, WI 53188 |
| Primary Contact Person: | Isabel McGann Regulatory Affairs Director GE HealthCare Phone: (720) 447-0747 Email: Isabel.mcgann@gehealthcare.com |
| Secondary Contact Person: | Jordan Baum Regulatory Affairs Program Manager GE HealthCare Phone: (574) 529-1811 Email: Jordan.baum@gehealthcare.com |
| Device Trade Name(s): | Giraffe™ OmniBed™ Carestation (CS1) Giraffe™ Incubator Carestation (CS1) |
| Common/Usual Name: | Giraffe OmniBed Carestation (CS1): Neonatal Incubator/Infant Radiant Warmer Giraffe Incubator Carestation (CS1): Neonatal Incubator |
| Regulation: | Neonatal Incubator |
| Classification: | 21 CFR 880.5400 |
| Product Code: | FMZ, FMT |
| Predicate Device: | Giraffe OmniBed Carestation (CS1), K251663 Giraffe Incubator Carestation (CS1), K251663 |
| Device Description: | The Giraffe OmniBed Carestation (CS1) is a device that can function as an incubator (in the closed mode) or as an infant radiant warmer (in the open mode) based on the user's selection. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermoregulate based on their own physiology. In the closed bed mode of operation, the bed functions as an incubator, maintaining the infant's temperature by circulating heated air within the enclosed bed compartment. Warm air is circulated through the closed patient compartment. The operator may select either the air or skin temperature control method. Depending on the control method selected, heat is regulated based on either the air temperature or the infant's skin temperature compared to the operated selected control temperature. Physical access to the patient is obtained through the side portholes or by opening one of the side doors. In the open bed mode, this bed operates like a conventional open, radiantly heated infant bed. Radiant heat from an infrared heat source is focused onto the bed to warm the patient. Accessories for the Giraffe OmniBed include an optional weighing scale, optional Servo Oxygen, Uninterruptible Power Supply (UPS), Giraffe Shuttle, Mounting Accessories Rail, |
Page 1 of 9
{5}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
| | Shelves, and Patient Probes. |
| --- | --- |
| | The Giraffe Incubator Carestation (CS1) is an enclosed infant bed, which provides thermal support for infants who are unable to provide for their own heat requirements. The device maintains the infant's temperature by circulating heated air within the closed bed compartment. The operator may select either the air or skin temperature control method. Depending on the control method selected, heat is regulated based on either the air temperature or the infant's skin temperature compared to the operator selected control temperature. Physical access to the patient is obtained through the side portholes or by opening one of the side doors. The Giraffe Incubator Carestation has a color touchscreen user interface (UI) and includes a Hands-Free Alarm Silence (HFAS) feature. The incubator includes a mattress for patient comfort. Accessories for the Giraffe Incubator include an optional weighing scale, optional Servo Oxygen, Uninterruptible Power Supply (UPS), Giraffe Shuttle, Mounting Accessories Rail, Shelves and Patient Probes. |
| Indications for Use: | The Giraffe OmniBed Carestation is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). The Giraffe Incubator Carestation is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature-controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). |
Page 2 of 9
{6}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
A comparison of the indications for use and technological features of the subject and predicate device is provided in Table 1 below.
Table 1: High-level Comparison of Subject Device, Giraffe Omnibed CareStation (CS1), to Predicate
| Specification: | Predicate Device: Giraffe Omnibed Carestation CS1 (K251663) | Subject Device: Giraffe Omnibed Carestation CS1 | Discussion of Differences: |
| --- | --- | --- | --- |
| Indications for Use | **The Giraffe OmniBed Carestation** is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). | **The Giraffe OmniBed Carestation** is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user's demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate's transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%). | Identical |
Page 3 of 9
{7}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
| Specification: | Predicate Device: Giraffe Omnibed Carestation CS1 (K251663) | Subject Device: Giraffe Omnibed Carestation CS1 | Discussion of Differences: |
| --- | --- | --- | --- |
| Product Code | FMZ, FMT | FMZ, FMT | Identical |
| Contraindications | None | None | Identical |
| User Population | Professional Use Only | Professional Use Only | Identical |
| Environment of Use | Labor and Delivery, NICU, or Newborn Nursery | Labor and Delivery, NICU, or Newborn Nursery | Identical |
| Fundamental Principle of Operation | Enclosed infant bed for thermal support. The device maintains the infant's temperature by circulating heated air within the closed bed compartment. Controller-based, open care radiant warmer that facilitates thermoregulation and emergency resuscitation of infants. | Enclosed infant bed for thermal support. The device maintains the infant's temperature by circulating heated air within the closed bed compartment. Controller-based, open care radiant warmer that facilitates thermoregulation and emergency resuscitation of infants. | Identical |
| Sterility | Non-sterile device | Non-sterile device | Identical |
| Alarm Silence | Two Options: • Touch Screen Silence • Hands free Alarms Silence (HFAS) | Two Options: • Touch Screen Silence • Hands free Alarms Silence (HFAS) | Identical |
| Dimensions | Weight: 149 kg Height: 152 cm (bed lowered)/ 178 cm (bed raised) Width: 68 cm Depth: 114 cm | Weight: 149 kg Height: 152 cm (bed lowered)/ 178 cm (bed raised) Width: 68 cm Depth: 114 cm | Identical |
| Bed Tilt | Mattress tilt angle: 12° | Mattress tilt angle: 12° | Identical |
| Electrical Power Ratings | 11.5@ 100V ~ 50/60 Hz 9.5A @115V ~ 50/60 Hz 5.5A @ 220/230/240V ~ 50/50 Hz | 11.5@ 100V ~ 50/60 Hz 9.5A @115V ~ 50/60 Hz 5.5A @ 220/230/240V ~ 50/50 Hz | Identical |
| Operating Environment | Temperature: 20° to 30° C Humidity: 10 to 95% RH (non-condensing) Air Velocity: Up to 0.3 m/sec | Temperature: 20° to 30° C Humidity: 10 to 95% RH (non-condensing) Air Velocity: Up to 0.3 m/sec | Identical |
| User Control Settings | • Patient control temperature 35-37.5°C in 0.1° increments | • Patient control temperature 35-37.5°C in 0.1° increments | Identical |
Page 4 of 9
{8}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
| Specification: | Predicate Device: Giraffe Omnibed Carestation CS1 (K251663) | Subject Device: Giraffe Omnibed Carestation CS1 | Discussion of Differences: |
| --- | --- | --- | --- |
| | • Air control temperature 20-39°C in 0.1 increments • Radiant heat power 0-100% in 5% increments • Humidity 30- 95 % RH in 5% increments | • Air control temperature 20-39°C in 0.1 increments • Radiant heat power 0-100% in 5% increments • Humidity 30- 95 % RH in 5% increments | |
| Materials | Use of polyacetal components in some assemblies | Contains polyester components in some assemblies with equivalent performance and specifications | Different Material changes were adequately evaluated in consideration of the requirements of ISO 18562 and ISO 10993. All testing passed. |
| Software Cybersecurity | Meets current cybersecurity expectations | Meets current cybersecurity expectations | Identical |
Table 2: High-level Comparison of Subject Device, Giraffe Incubator CareStation (CS1), to Predicate
| Specification | Predicate Device: Giraffe Incubator Carestation CS1 (K251663) | Subject Devices: Giraffe Incubator Carestation CS1 | Discussion of Differences |
| --- | --- | --- | --- |
| Indications for Use | **The Giraffe Incubator Carestation** is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature-controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the | **The Giraffe Incubator Carestation** is an Infant Incubator. Incubators provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. They achieve this by providing an enclosed temperature-controlled environment to the infant. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide a stable oxygen concentration within the infant compartment at the | Identical |
Page 5 of 9
{9}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
| Specification | Predicate Device: Giraffe Incubator Carestation CS1 (K251663) | Subject Devices: Giraffe Incubator Carestation CS1 | Discussion of Differences |
| --- | --- | --- | --- |
| | value set by the operator (21-65%). | value set by the operator (21-65%). | |
| **Product Code** | FMZ | FMZ | Identical |
| **Contraindications** | None | None | Identical |
| **User Population** | Professional Use Only | Professional Use Only | Identical |
| **Environment of Use** | Labor and Delivery, NICU, or Newborn Nursery | Labor and Delivery, NICU, or Newborn Nursery | Identical |
| **Fundamental Principle of Operation** | Enclosed infant bed for thermal support. The device maintains the infant's temperature by circulating heated air within the closed bed compartment. | Enclosed infant bed for thermal support. The device maintains the infant's temperature by circulating heated air within the closed bed compartment. | Identical |
| **Sterility** | Non-sterile device | Non-sterile device | Identical |
| **Alarm Silence** | Two Options: - Touch Screen Silence - Hands free Alarms Silence (HFAS) | Two Options: - Touch Screen Silence - Hands free Alarms Silence (HFAS) | Identical |
| **Dimensions** | Weight: 138 kg Height: 152 cm (bed lowered)/ 178 cm (bed raised) Width: 66 cm Depth: 114 cm | Weight: 138 kg Height: 152 cm (bed lowered)/ 178 cm (bed raised) Width: 66 cm Depth: 114 cm | Identical |
| **Bed Tilt** | Mattress tilt angle: 12° | Mattress tilt angle: 12° | Identical |
| **Electrical Power Ratings** | 11.5@ 100V ~ 50/60 Hz 9.5A @115V ~ 50/60 Hz 5.5A @ 220/230/240V ~ 50/50 Hz | 11.5@ 100V ~ 50/60 Hz 9.5A @115V ~ 50/60 Hz 5.5A @ 220/230/240V ~ 50/50 Hz | Identical |
| **Operating Environment** | Temperature: 20° to 30° C Humidity: 10 to 95% RH (non-condensing) Air Velocity: Up to 0.3 m/sec | Temperature: 20° to 30° C Humidity: 10 to 95% RH (non-condensing) Air Velocity: Up to 0.3 m/sec | Identical |
| **User Control Settings** | - Patient control temperature 35-37.5°C in 0.1° increments - Air control temperature 20-39°C in 0.1 increments - Humidity 30- 95 % RH in 5% increments | - Patient control temperature 35-37.5°C in 0.1° increments - Air control temperature 20-39°C in 0.1 increments - Humidity 30- 95 % RH in 5% increments | Identical |
Page 6 of 9
{10}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
| Specification | Predicate Device: Giraffe Incubator Carestation CS1 (K251663) | Subject Devices: Giraffe Incubator Carestation CS1 | Discussion of Differences |
| --- | --- | --- | --- |
| Materials | Use of polyacetal components in some assemblies | Contains polyester components in some assemblies with equivalent performance and specifications | Different Material changes were adequately evaluated in consideration of the requirements of ISO 18562 and ISO 10993. All testing passed. |
| Software Cybersecurity | Meets current cybersecurity expectations | Meets current cybersecurity expectations | Identical |
### PERFORMANCE DATA: Determination of Substantial Equivalence
#### Summary of Non-Clinical Tests:
The safety and effectiveness of the replacement materials used at different locations in the breathing gas pathway was evaluated per ISO 18562, and the results met the acceptance criteria without any deviations. In addition, bench verification testing included Device/System Functional Verification Testing, Non-Operating and Device Packaging Testing, Verification by Analysis, Material Stress Testing, and Reliability Testing. All verification results were determined to be PASS, with any deviations assessed and determined not to impact testing results. No additional non-clinical testing beyond the testing described above was required in support of the material changes.
#### Compliance with Voluntary Standards:
The following FDA recognized consensus standards were used to demonstrate substantial equivalence:
- ANSI AAMI ISO 10993-1:2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- ANSI AAMI ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 Fourth edition 2021-11 Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- ISO 10993-23 First edition 2021-01 Biological evaluation of medical devices - Part 23: Tests for irritation
- ISO 18562-1:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 1: Evaluation and testing within a risk management process
- ISO 18562-2:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 2: Tests for emissions of particulate matter
- ISO 18562-3:2017 Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds
Page 7 of 9
{11}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
- ANSI AAMI IEC 62304:2006/A1:2016 Medical device software - Software life cycle processes
- ANSI AAMI ISO 14971: 2019 Medical devices - Applications of risk management to medical devices
- ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)]
- IEC 60601-2-19: 2020 (3rd ed) Medical electrical equipment –Part 2-19: Particular requirements for the basic safety and essential performance of infant incubators
- IEC 60601-2-21:2020 (3rd ed) Medical electrical equipment - Part 2-21: Particular requirements for the basic safety and essential performance of infant radiant warmers
- ISTA 3B 2017 - Packaged-Products for Less-Than-Truckload (LTL) Shipment
- ISO 17664-1:2021, Processing of health care products Information to be provided by the medical device manufacturer for the processing of medical devices
- ISO 17664-2: 2021, Processing of health care products Information to be provided by the medical device manufacturer for the processing of medical devices
### Biocompatibility
Evaluations of safety and effectiveness of the device with new plastic materials was performed to substantiate the performance of the updated device in accordance with ISO 10993 and ISO 18562. All results passed.
### Reprocessing
Note that the material changes have not affected the cleaning or disinfection instructions or validated state of the device. The changes are in material selection only and do not affect the features, geometry, or surface texture of the impacted components. However, changes to material require additional evaluation of chemical compatibility for the chemicals listed in the product labeling. Chemical compatibility evaluation demonstrates that the stability of the device over the duration of the expected service life is not affected by the changes in materials.
### Human Factors
The material changes do not impact the usability of the devices therefore no new human factors testing was completed to support substantial equivalence.
### Software and Cybersecurity
The material changes have no impact on software. However, in accordance with Section 524B of the Food Drug and Cosmetic Act (FD&C Act) and “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions Guidance for Industry and Food and Drug Administration Staff, Document issued on June 27, 2025, current cybersecurity content has been included.
Page 8 of 9
{12}
Traditional
510(k) Summary
Giraffe™ OmniBed™ CareStation (CS1)
Giraffe™ Incubator CareStation (CS1)
Material Component Updates
GE HealthCare
# **CONCLUSION:**
The results of the testing described above demonstrate that the Giraffe Incubator Carestation CS1 and Giraffe OmniBed Carestation CS1 are as safe and effective as the predicate devices and support a determination of substantial equivalence.
Page 9 of 9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.