← Product Code [QSO](/productcode/QSO) · K260310

# HALYARD* CHEMO360 Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL); HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL) (K260310)

_O&M Halyard, Inc. · QSO · Oct 28, 2026 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K260310

## Device Facts

- **Applicant:** O&M Halyard, Inc.
- **Product Code:** [QSO](/productcode/QSO.md)
- **Decision Date:** Oct 28, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use, AAMI PB70 Level 2 Procedure Gowns worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients. Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs, using ASTM F739, Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted.

## Device Story

Disposable, non-sterile procedure gowns (CHEMO360 full-back or standard open-back) designed for healthcare worker protection. Constructed from polypropylene fabric laminated with polypropylene-polyethylene film. Used in clinical settings for general infection control and chemotherapy drug administration. Provides AAMI PB70 Level 2 fluid barrier protection. Evaluated for chemotherapy drug permeation resistance per ASTM F3267 and ASTM F739. Output is a physical barrier; healthcare providers use the gown to mitigate exposure to hazardous drugs and fluids. Benefits include reduced risk of skin contact with chemotherapy agents and bodily fluids during patient care.

## Clinical Evidence

No clinical data. Safety and performance demonstrated through bench testing, including biocompatibility (ISO 10993-5, 10993-10), flammability (16 CFR 1610), physical strength (ASTM D5034, D5587, D1683), liquid barrier performance (AAMI PB70 Level 2), and chemotherapy drug permeation (ASTM F739/F3267).

## Technological Characteristics

Materials: 1.02-1.42 osy polypropylene fabric laminated with polypropylene-polyethylene film. Design: Full-back (CHEMO360) or open-back (standard) with knit cuffs. Barrier: AAMI PB70 Level 2. Standards: ASTM F3267, ASTM F739, ISO 10993, ASTM D5034, ASTM D5587, ASTM D1683, 16 CFR 1610.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Copioumed Chemotherapy Isolation Gown ([K230135](/device/K230135.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

**U.S. FOOD & DRUG**
ADMINISTRATION

October 28, 2026

O&M Halyard, Inc.
Natrisya Nur Abdul Rahman
RA Specialist
1220 Old Alpharetta Rd.
Suite 320
Alpharetta, Georgia 30005

Re: K260310

Trade/Device Name: HALYARD* CHEMO360 Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL); HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)

Regulation Number: 21 CFR 878.4040

Regulation Name: Surgical Apparel

Regulatory Class: Class II

Product Code: QSO

Dated: September 29, 2026

Received: September 29, 2026

Dear Natrisya Nur Abdul Rahman:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of

{2}

K260310 - Natrisya Nur Abdul Rahman

Page 2

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{3}

K260310 - Natrisya Nur Abdul Rahman

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

![img-0.jpeg](img-0.jpeg)

Bifeng Qian, M.D., Ph.D.

Assistant Director

DHT4C: Division of Infection

Control Devices

OHT4: Office of Surgical and

Infection Control Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{4}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120

Expiration Date: 07/31/2026

See PRA Statement below.

510(k) Number (if known)

K260310

Device Name

HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL);

HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)

Indications for Use (Describe)

HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use AAMI PB70 Level 2 Procedure Gown worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients.

Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs, using ASTM F739, Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted. Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period:

Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm)

Azacytidine (Vidaza), 25 mg/mL (25,000 ppm)

Bendamustine HCl, 25 mg/mL (25,000 ppm)

Bleomycin Sulfate, 15 mg/mL (15,000 ppm)

Bortezomib, 1 mg/mL (1,000 ppm)

Busulfan, 6 mg/mL (6,000 ppm)

Carboplatin, 10 mg/mL (10,000 ppm)

Car Izomib, 2 mg/mL (2,000 ppm)

Carmustine, 3.3 mg/mL (3,300 ppm)

Cladribine, 1 mg/mL (1,000 ppm)

Cyclosporin A, 100 mg/mL (100,000 ppm)

Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm)

Dacarbazine (DTIC), 10 mg/mL (10,000 ppm)

Dactinomycin, 0.5 mg/mL (500 ppm)

Daunorubicin HCl, 5 mg/mL (5,000 ppm)

Decitabine, 5 mg/mL (5,000 ppm)

Docetaxel, 10 mg/mL (10,000 ppm)

Doxorubicin HCl, 2 mg/mL (2,000 ppm)

Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm)

Eribulin Mesylate, 0.5 mg/mL (500 ppm)

Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm)

Gemcitabine HCL, 38 mg/mL (38,000 ppm)

Idarubicin HCL, 1 mg/mL (1,000 ppm)

Lenvatinib, 20 mg/mL (20,000 ppm)

Leuprolide acetate salt, 5 mg/mL (5,000 ppm)

Mechlorethamine HCl, 1 mg/mL (1,000 ppm)

Melphalan HCl, 5 mg/mL (5,000 ppm)

Methotrexate, 25 mg/mL (25,000 ppm)

FORM FDA 3881 (8/23)

Page 1 of 3

PSC Publishing Services (301) 443-6740

EF

{5}

Mitomycin C, 0.5 mg/mL (500 ppm)

Mitoxantrone HCl, 2 mg/mL (2,000 ppm)

Oxaliplatin, 5 mg/mL (5,000 ppm)

Pemetrexed, 25 mg/mL (25,000 ppm)

Raltitrexed, 0.5 mg/mL (500 ppm)

Retrovir, 10 mg/mL (10,000 ppm)

Rituximab, 10 mg/mL (10,000 ppm)

Sorafenib Tosylate, 200 mg/mL (200,000 ppm)

Tamoxifen, 2 mg/mL (2,000 ppm)

Thiotepa, 10 mg/mL (10,000 ppm)

Topotecan HCL, 1 mg/mL (1,000 ppm)

Vinblastine Sulfate, 1 mg/mL (1,000 ppm)

Vinorelbine, 10 mg/mL (10,000 ppm)

# Caution:

The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:

Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutes

Etoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutes

Fludarabine, 25 mg/ml (25,000 ppm) 120 minutes

Ifosfamide, 50 mg/ml (50,000 ppm) 30 minutes

Irinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutes

Paclitaxel, 6 mg/ml (6,000 ppm) 120 minutes

Vincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutes

Per ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of 0.1μg/cm2/minute:

Cyclophosphamide, at a concentration of 20 mg/mL (20,000 ppm), demonstrated a breakthrough detection time of 120 minutes. The cumulative permeation was 0.00 μg/cm² at 30 minutes, 0.00 μg/cm² at 60 minutes, and 516.00 μg/cm² at 240 minutes.

Cytarabine HCl (Cytosine), at a concentration of 100 mg/mL (100,000 ppm), demonstrated a breakthrough detection time of 60 minutes. The cumulative permeation was 0.26 μg/cm² at 30 minutes, 0.46 μg/cm² at 60 minutes, and 68.35 μg/cm² at 240 minutes.

Fludarabine, at a concentration of 25 mg/mL (25,000 ppm), demonstrated a breakthrough detection time of 120 minutes. The cumulative permeation was 0.08 μg/cm² at 30 minutes, 0.27 μg/cm² at 60 minutes, and 7.28 μg/cm² at 240 minutes.

Irinotecan HCl, at a concentration of 20 mg/mL (20,000 ppm), demonstrated a breakthrough detection time of 240 minutes. The cumulative permeation was 0.03 μg/cm² at 30 minutes, 0.10 μg/cm² at 60 minutes, and 0.30 μg/cm² at 240 minutes.

Paclitaxel (Taxol), at a concentration of 6 mg/mL (6,000 ppm), demonstrated a breakthrough detection time of 120 minutes. The cumulative permeation was 0.00 μg/cm² at 30 minutes, 0.00 μg/cm² at 60 minutes, and 1.64 μg/cm² at 240 minutes.

The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes:

Fentanyl Citrate Injection, 100 mcg/2 ml

Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution

Type of Use (Select one or both, as applicable)

☐ Prescription Use (Part 21 CFR 801 Subpart D)

☑ Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

FORM FDA 3881 (8/23)

Page 2 of 3

PSC Publishing Services (301) 443-6740

EF

{6}

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/23)

Page 3 of 3

{7}

Owens & Minor

HALYARD

# 510(k) Summary

510(k)

Number: K260310

Date

Prepared: October 27, 2026

Submitter:

O&M Halyard Inc.
1220 Old Alpharetta Road, Suite 320
Alpharetta, GA 30005

Regulatory

Contact: Natrisya Nur
natrisya.nur@owens-minor.com

Device Trade

Names: HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)

Device

Common Name: Procedure Gown

FDA Device Product Code: QSO

FDA Device Classification: Class II

FDA

Regulation Number: 21 CFR 878.4040

Predicate

Device: K230135 Copioumed Chemotherapy Isolation Gown

Page 1

{8}

Owens & Minor

HALYARD

# **Device**

# **Descriptions:**

The HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs is a blue, full-back, non-sterile, single-use disposable gown available in sizes L and XXL. The gown is constructed of polypropylene fabric laminated with a polypropylene-polyethylene film and features a neck tape closure, front tie, and knit cuffs.

The HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs is a blue, open-back, non-sterile, single-use disposable gown available in sizes L and XXL. The gown is constructed of polypropylene fabric laminated with a polypropylene-polyethylene film and features an over-the-head design with knit cuffs.

# **Indications for Use**

HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use, AAMI PB70 Level 2 Procedure Gowns worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients.

Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs, using ASTM F739, Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted.

Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period:

Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm)

Azacytidine (Vidaza), 25 mg/mL (25,000 ppm)

Bendamustine HCl, 25 mg/mL (25,000 ppm)

Bleomycin Sulfate, 15 mg/mL (15,000 ppm)

Bortezomib, 1 mg/mL (1,000 ppm)

Busulfan, 6 mg/mL (6,000 ppm)

Carboplatin, 10 mg/mL (10,000 ppm)

Car Izzomib, 2 mg/mL (2,000 ppm)

Carmustine, 3.3 mg/mL (3,300 ppm)

Cladribine, 1 mg/mL (1,000 ppm)

Cyclosporin A, 100 mg/mL (100,000 ppm)

Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm)

Dacarbazine (DTIC), 10 mg/mL (10,000 ppm)

Dactinomycin, 0.5 mg/mL (500 ppm)

Daunorubicin HCl, 5 mg/mL (5,000 ppm)

Decitabine, 5 mg/mL (5,000 ppm)

Docetaxel, 10 mg/mL (10,000 ppm)

Doxorubicin HCl, 2 mg/mL (2,000 ppm)

Page 2

{9}

[Non-Text]

Owens & Minor

[Non-Text]

HALYARD

Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm)

Eribulin Mesylate, 0.5 mg/mL (500 ppm)

Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm)

Gemcitabine HCL, 38 mg/mL (38,000 ppm)

Idarubicin HCL, 1 mg/mL (1,000 ppm)

Lenvatinib, 20 mg/mL (20,000 ppm)

Leuprolide acetate salt, 5 mg/mL (5,000 ppm)

Mechlorethamine HCl, 1 mg/mL (1,000 ppm)

Melphalan HCl, 5 mg/mL (5,000 ppm)

Methotrexate, 25 mg/mL (25,000 ppm)

Mitomycin C, 0.5 mg/mL (500 ppm)

Mitoxantrone HCl, 2 mg/mL (2,000 ppm)

Oxaliplatin, 5 mg/mL (5,000 ppm)

Pemetrexed, 25 mg/mL (25,000 ppm)

Raltitrexed, 0.5 mg/mL (500 ppm)

Retrovir, 10 mg/mL (10,000 ppm)

Rituximab, 10 mg/mL (10,000 ppm)

Sorafenib Tosylate, 200 mg/mL (200,000 ppm)

Tamoxifen, 2 mg/mL (2,000 ppm)

Thiotepa, 10 mg/mL (10,000 ppm)

Topotecan HCL, 1 mg/mL (1,000 ppm)

Vinblastine Sulfate, 1 mg/mL (1,000 ppm)

Vinorelbine, 10 mg/mL (10,000 ppm)

### Caution:

The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:

Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutes

Etoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutes

Fludarabine, 25 mg/ml (25,000 ppm) 120 minutes

Ifosfamide, 50 mg/ml (50,000 ppm) 30 minutes

Irinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutes

Paclitaxel, 6 mg/ml (6,000 ppm) 120 minutes

Vincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutes

Per ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of  \( 0.1 \mu g/cm^{2}/minute \)

|  Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \))  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  30 min | 60 min | 240 min  |
|  Cyclophosphamide, 20 mg/mL (20,000 ppm) | 120 minutes | 0.00 | 0.00 | 516.00  |
|  Cytarabine HCl (Cytosine), 100 mg/mL (100,000 ppm) | 60 minutes | 0.26 | 0.46 | 68.35  |
|  Fludarabine, 25 mg/mL (25,000 ppm) | 120 minutes | 0.08 | 0.27 | 7.28  |

Page 3

{10}

[LOGO]

Owens & Minor

[LOGO]

HALYARD

|  Irinotecan HCL, 20 mg/mL (20,000 ppm) | 240 minutes | 0.03 | 0.10 | 0.30  |
| --- | --- | --- | --- | --- |
|  Paclitaxel (Taxol), 6 mg/mL (6,000 ppm) | 120 minutes | 0.00 | 0.00 | 1.64  |

The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes:

Fentanyl Citrate Injection, 100 mcg/2 ml

Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution

Page 4

{11}

Owens & Minor HALYARD

### Comparison of Technological Characteristics

|  Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis  |
| --- | --- | --- | --- |
|  Product Name | HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs. | Copioumed Chemotherapy Isolation Gown | Different  |
|  Intended Use/ Indication for Use | HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use AAMI PB70 Level 2 Procedure Gown worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients. Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, *Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs*, using ASTM F739, *Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact*. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted. Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period: Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm) Azacytidine (Vidaza), 25 mg/mL (25,000 ppm) Bendamustine HCl, 25 mg/mL (25,000 ppm) Bleomycin Sulfate, 15 mg/mL (15,000 ppm) Bortezomib, 1 mg/mL (1,000 ppm) Busulfan, 6 mg/mL (6,000 ppm) | Copioumed Chemotherapy Isolation Gown is intended to be worn by healthcare personnel to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate. Copioumed Chemotherapy Isolation Gown is non-sterile and for single use only. Copioumed Chemotherapy Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). Copioumed Chemotherapy Isolation Gown is a disposable device intended to protect health care personal from exposure to chemotherapy drugs during preparation, handling, and administration. The Copioumed Chemotherapy Isolation Gown has been evaluated for resistance to permeation of various chemotherapy drugs | Similar  |

Page 5

{12}

Owens & Minor

HALYARD

|  Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis  |
| --- | --- | --- | --- |
|   | Carboplatin, 10 mg/mL (10,000 ppm) Car Ixomib, 2 mg/mL (2,000 ppm) Carmustine, 3.3 mg/mL (3,300 ppm) Cladribine, 1 mg/mL (1,000 ppm) Cyclosporin A, 100 mg/mL (100,000 ppm) Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm) Dacarbazine (DTIC), 10 mg/mL (10,000 ppm) Dactinomycin, 0.5 mg/mL (500 ppm) Daunorubicin HCl, 5 mg/mL (5,000 ppm) Decitabine, 5 mg/mL (5,000 ppm) Docetaxel, 10 mg/mL (10,000 ppm) Doxorubicin HCl, 2 mg/mL (2,000 ppm) Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm) Eribulin Mesylate, 0.5 mg/mL (500 ppm) Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm) Gemcitabine HCL, 38 mg/mL (38,000 ppm) Idarubicin HCL, 1 mg/mL (1,000 ppm) Lenvatinib, 20 mg/mL (20,000 ppm) Leuprolide acetate salt, 5 mg/mL (5,000 ppm) Mechlorethamine HCl, 1 mg/mL (1,000 ppm) Melphalan HCl, 5 mg/mL (5,000 ppm) Methotrexate, 25 mg/mL (25,000 ppm) Mitomycin C, 0.5 mg/mL (500 ppm) Mitoxantrone HCl, 2 mg/mL (2,000 ppm) Oxaliplatin, 5 mg/mL (5,000 ppm) Pemetrexed, 25 mg/mL (25,000 ppm) Raltitrexed, 0.5 mg/mL (500 ppm) Retrovir, 10 mg/mL (10,000 ppm) Rituximab, 10 mg/mL (10,000 ppm) Sorafenib Tosylate, 200 mg/mL (200,000 ppm) Tamoxifen, 2 mg/mL (2,000 ppm) Thiotepa, 10 mg/mL (10,000 ppm) Topotecan HCL, 1 mg/mL (1,000 ppm) Vinblastine Sulfate, 1 mg/mL (1,000 ppm) | per ASTM F739-20 "Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact." Chemotherapy drug permeation resistance, average standardized breakthrough time in minutes.*: Carmustine [3.3 mg/ml] >480 Cisplatin [1.0 mg/ml] >480 Cyclophosphamide [20 mg/ml] >480 Cytarabine Hydrochloride [100 mg/ml] >480 Dacarbazine [10 mg/ml] >480 Daunorubicin Hydrochloride [5 mg/ml] >480 Doxorubicin Hydrochloride [2 mg/ml] > 480 Etoposide [20 mg/ml] >480 Fluorouracil [50 mg/ml] >480 Ifosfamide [50 mg/ml] > 480 Methotrexate [25 mg/ml] >480 Mitomycin C [0.5 mg/ml] >480 Mitoxantrone [2.0 mg/ml] >480 Paclitaxel [6 mg/ml] >480 Thiotepa [10 mg/ml] >480 Vincristine Sulfate [1 mg/ml] >480 *No permeation was detected at either the minimum detectable permeation or 0.1 μg/cm²/min |   |

Page 6

{13}

Owens & Minor

O

HALYARD

| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
|  | Vinorelbine, 10 mg/mL (10,000 ppm)Caution:The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutesEtoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutesFludarabine, 25 mg/ml (25,000 ppm) 120 minutesIfosfamide, 50 mg/ml (50,000 ppm) 30 minutesIrinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutesPaclitaxel, 6 mg/ml (6,000 ppm) 120 minutesVincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutesPer ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of \( 0.1\mu g/cm^{2}/minute \) |  |  |
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) |
| 30 min | 60 min | 240 min |
| Cyclophosphamide, 20 mg/mL (20,000 ppm) | 120 minutes | 0.00 | 0.00 | 516.00 |
| Cytarabine HCl (Cytosine), 100 mg/mL (100,000 ppm) | 60 minutes | 0.26 | 0.46 | 68.35 |
| Fludarabine, 25 mg/mL (25,000 ppm) | 120 minutes | 0.08 | 0.27 | 7.28 |
| Irinotecan HCL, 20 mg/mL (20,000 ppm) | 240 minutes | 0.03 | 0.10 | 0.30 |
| Paclitaxel (Taxol), 6 mg/mL (6,000 ppm) | 120 minutes | 0.00 | 0.00 | 1.64 |

Page 7

{14}

[LOGO]

Owens & Minor

[LOGO]

HALYARD

|  Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis  |
| --- | --- | --- | --- |
|   | The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes: Fentanyl Citrate Injection, 100 mcg/2 ml Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution |  |   |
|  Color | Blue | Blue | Same  |
|  Design Features | HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs is a full back gown with a neck tape closure, front tie, and knit cuffs. HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs is an open back, over the head gown with knit cuffs. | Knit cuff in three style(model) Tie neck closure, Velcro neck closure and Sticker neck closure | Similar  |
|  Sizes | Large, XX-Large | Large, X-Large | Similar  |
|  Materials | 1.02-1.42 osy Polypropylene Gown laminated with polypropylene-polyethylene film. | Polypropylene Gown laminated with polyethylene | Similar  |
|  Performance per AAMI PB70 | AAMI Level 2 | AAMI Level 3 | Different^{1}  |
|  Resistance to permeation of chemotherapy drugs testing | HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, *Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs*, using ASTM F739, *Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact*. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted. Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period: Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm) Azacytidine (Vidaza), 25 mg/mL (25,000 ppm) Bendamustine HCl, 25 mg/mL (25,000 ppm) Bleomycin Sulfate, 15 mg/mL (15,000 ppm) Bortezomib, 1 mg/mL (1,000 ppm) Busulfan, 6 mg/mL (6,000 ppm) Carboplatin, 10 mg/mL (10,000 ppm) | ASTM F739-20 Results showed no average Standardized breakthrough for up to 480 minutes for claimed 16 drugs. Chemotherapy drug permeation resistance, average standardized breakthrough time in minutes.*: Carmustine [3.3 mg/ml] >480 Cisplatin [1.0 mg/ml] >480 Cyclophosphamide [20 mg/ml]>480 Cytarabine Hydrochloride [100 mg/ml] >480 Dacarbazine [10 mg/ml] >480 Daunorubicin Hydrochloride [5 mg/ml] >480 Doxorubicin Hydrochloride [2 mg/ml] >480 | Similar^{2}  |

Page 8

{15}

Owens & Minor

HALYARD

|  Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis  |
| --- | --- | --- | --- |
|   | Car Izomib, 2 mg/mL (2,000 ppm) Carmustine, 3.3 mg/mL (3,300 ppm) Cladribine, 1 mg/mL (1,000 ppm) Cyclosporin A, 100 mg/mL (100,000 ppm) Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm) Dacarbazine (DTIC), 10 mg/mL (10,000 ppm) Dactinomycin, 0.5 mg/mL (500 ppm) Daunorubicin HCl, 5 mg/mL (5,000 ppm) Decitabine, 5 mg/mL (5,000 ppm) Docetaxel, 10 mg/mL (10,000 ppm) Doxorubicin HCl, 2 mg/mL (2,000 ppm) Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm) Eribulin Mesylate, 0.5 mg/mL (500 ppm) Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm) Gemcitabine HCL, 38 mg/mL (38,000 ppm) Idarubicin HCL, 1 mg/mL (1,000 ppm) Lenvatinib, 20 mg/mL (20,000 ppm) Leuprolide acetate salt, 5 mg/mL (5,000 ppm) Mechlorethamine HCl, 1 mg/mL (1,000 ppm) Melphalan HCl, 5 mg/mL (5,000 ppm) Methotrexate, 25 mg/mL (25,000 ppm) Mitomycin C, 0.5 mg/mL (500 ppm) Mitoxantrone HCl, 2 mg/mL (2,000 ppm) Oxaliplatin, 5 mg/mL (5,000 ppm) Pemetrexed, 25 mg/mL (25,000 ppm) Raltitrexed, 0.5 mg/mL (500 ppm) Retrovir, 10 mg/mL (10,000 ppm) Rituximab, 10 mg/mL (10,000 ppm) Sorafenib Tosylate, 200 mg/mL (200,000 ppm) Tamoxifen, 2 mg/mL (2,000 ppm) Thiotepa, 10 mg/mL (10,000 ppm) Topotecan HCL, 1 mg/mL (1,000 ppm) Vinblastine Sulfate, 1 mg/mL (1,000 ppm) Vinorelbine, 10 mg/mL (10,000 ppm) | Etoposide [20 mg/ml] >480 Fluorouracil [50 mg/ml] >480 Ifosfamide [50 mg/ml] > 480 Methotrexate [25 mg/ml] >480 Mitomycin C [0.5 mg/ml] >480 Mitoxantrone [2.0 mg/ml] >480 Paclitaxel [6 mg/ml] >480 Thiotepa [10 mg/ml] >480 Vincristine Sulfate [1 mg/ml]>480 *No permeation was detected at either the minimum detectable permeation or 0.1 μg/cm2/min |   |

Page 9

{16}

Owens & Minor

O

HALYARD

| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
|  | Caution:The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutesEtoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutesFludarabine, 25 mg/ml (25,000 ppm) 120 minutesIfosfamide, 50 mg/ml (50,000 ppm) 30 minutesIrinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutesPaclitaxel, 6 mg/ml (6,000 ppm) 120 minutesVincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutesPer ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of \( 0.1\mu g/cm^{2}/minute \) |  |  |
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) |
| 30 min | 60 min | 240 min |
| Cyclophosphamide, 20 mg/mL (20,000 ppm) | 120 minutes | 0.00 | 0.00 | 516.00 |
| Cytarabine HCl (Cytosine), 100 mg/mL (100,000 ppm) | 60 minutes | 0.26 | 0.46 | 68.35 |
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) |
| 30 min | 60 min | 240 min |
| Fludarabine, 25 mg/mL (25,000 ppm) | 120 minutes | 0.08 | 0.27 | 7.28 |
| Irinotecan HCL, 20 mg/mL (20,000 ppm) | 240 minutes | 0.03 | 0.10 | 0.30 |
| Paclitaxel (Taxol), 6 mg/mL (6,000 ppm) | 120 minutes | 0.00 | 0.00 | 1.64 |

Page 10

{17}

Owens & Minor

HALYARD

|  Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis  |
| --- | --- | --- | --- |
|   | The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes: Fentanyl Citrate Injection, 100 mcg/2 ml Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution |  |   |
|  Biocompatibility | Under the test conditions, the subject device was shown to be non- cytotoxic, non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10 | Under the test conditions, the subject device was shown to be non- cytotoxic, non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10 | Same  |
|  Flammability | Meets requirements of Flame Resistant CPSC 16 CFR 1610 Class 1 | Meets requirements of Flame Resistant CPSC 16 CFR 1610 Class 1 | Same  |
|  Tensile Strength | Meets the requirement of ASTM D5034-9 | Meets the requirement of ASTM D5034-9 | Same  |
|  Tear Resistance | Meets the requirement of ASTM D5587-15 | Meets the requirement of ASTM D5587-15 | Same  |
|  Seam Strength | Meets the requirement of ASTM 1683-17 | Meets the requirement of ASTM 1683-17 | Same  |

Different¹ While the proposed device and the predicate device differ in AAMI levels of protection, the predicate device was selected for the purpose of demonstrating evaluation of resistance to permeation against various chemotherapy drugs. The difference in the AAMI protection level is addressed through labeling.

Similar² The predicate device was selected solely to demonstrate evaluation for resistance to permeation by various chemotherapy drugs. Both the proposed device and the predicate device were tested for resistance to permeation of chemotherapy drugs in accordance with the standard ASTM F739, Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The proposed device was evaluated against a broader range of chemotherapy drugs than the predicate device. All chemotherapy drugs tested for the proposed device are clearly identified in the product labeling

Page 11

{18}

Owens & Minor

HALYARD

### Summary of Non-Clinical Testing

The subject device was evaluated through applicable non-clinical testing, and the results support the safety and performance of the device for its intended use.

|  Standard | Name | Acceptance Criteria | Result  |
| --- | --- | --- | --- |
|  ISO 10993-5 | Tests for In Vitro Cytotoxicity | Non-Cytotoxicity | Pass  |
|  ISO 10993-10 | Tests for Skin Irritation and Sensitization | Non-Irritation and Non-Sensitizing | Pass  |
|  AATCC 42-17 | Water Resistance | AATCC 42 ≤ 1.0g | Pass  |
|  AATCC 127-18 | Hydrostatic Pressure | AATCC 127 ≥ 50 cm H_{2}O | Pass  |
|  ASTM D5034-09 | Breaking Strength and Elongation of Textile Fabrics (Grab Test) | ≥ 30N | Pass  |
|  ASTM D5587-15 | Tearing Strength of Fabrics by Trapezoid Procedure | ≥ 10N | Pass  |
|  ASTM D1683 | Seam strength | ≥ 30N | Pass  |
|  16 CFR 1610 | Flammability of Clothing Textiles | Class 1 (Normal flammability) | Pass  |
|  ANSI AAMI PB70:2022 | Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Healthcare Facilities | Level 2 | Pass  |
|  ASTM F739:2020 | Permeation of Liquids and Gasses Through Protective Clothing | No permeation above detection limit during test period or breakthrough time ≥ specified time | Pass  |

### Summary of Clinical Testing

No clinical testing was conducted for the subject device. Clinical testing was not considered necessary to support the safety and effectiveness of the device, as the relevant safety and performance characteristics can be adequately demonstrated through non-clinical testing

### Conclusion

The conclusions drawn from the non-clinical tests demonstrate that the subject device, the HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs, is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K230135, with respects to resistance to permeation of chemotherapy drugs and Fentanyl Citrate.

Page 12

---

**Source:** [https://fda.innolitics.com/device/K260310](https://fda.innolitics.com/device/K260310)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com