HALYARD* CHEMO360 Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL); HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
K260310 · O&M Halyard, Inc. · QSO · Oct 28, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260310
Device Name
HALYARD* CHEMO360 Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL); HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
Applicant
O&M Halyard, Inc.
Product Code
QSO · General, Plastic Surgery
Decision Date
Oct 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use, AAMI PB70 Level 2 Procedure Gowns worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients. Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs, using ASTM F739, Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted.
Device Story
Disposable, non-sterile procedure gowns (CHEMO360 full-back or standard open-back) designed for healthcare worker protection. Constructed from polypropylene fabric laminated with polypropylene-polyethylene film. Used in clinical settings for general infection control and chemotherapy drug administration. Provides AAMI PB70 Level 2 fluid barrier protection. Evaluated for chemotherapy drug permeation resistance per ASTM F3267 and ASTM F739. Output is a physical barrier; healthcare providers use the gown to mitigate exposure to hazardous drugs and fluids. Benefits include reduced risk of skin contact with chemotherapy agents and bodily fluids during patient care.
Clinical Evidence
No clinical data. Safety and performance demonstrated through bench testing, including biocompatibility (ISO 10993-5, 10993-10), flammability (16 CFR 1610), physical strength (ASTM D5034, D5587, D1683), liquid barrier performance (AAMI PB70 Level 2), and chemotherapy drug permeation (ASTM F739/F3267).
Technological Characteristics
Materials: 1.02-1.42 osy polypropylene fabric laminated with polypropylene-polyethylene film. Design: Full-back (CHEMO360) or open-back (standard) with knit cuffs. Barrier: AAMI PB70 Level 2. Standards: ASTM F3267, ASTM F739, ISO 10993, ASTM D5034, ASTM D5587, ASTM D1683, 16 CFR 1610.
Indications for Use
Indicated for healthcare workers requiring AAMI PB70 Level 2 barrier protection against minimal to low fluid contact (water, blood, bodily fluids) and for protection during the administration of chemotherapy drugs.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{1}
**U.S. FOOD & DRUG**
ADMINISTRATION
October 28, 2026
O&M Halyard, Inc.
Natrisya Nur Abdul Rahman
RA Specialist
1220 Old Alpharetta Rd.
Suite 320
Alpharetta, Georgia 30005
Re: K260310
Trade/Device Name: HALYARD* CHEMO360 Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL); HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
Regulation Number: 21 CFR 878.4040
Regulation Name: Surgical Apparel
Regulatory Class: Class II
Product Code: QSO
Dated: September 29, 2026
Received: September 29, 2026
Dear Natrisya Nur Abdul Rahman:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{2}
K260310 - Natrisya Nur Abdul Rahman
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{3}
K260310 - Natrisya Nur Abdul Rahman
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Bifeng Qian, M.D., Ph.D.
Assistant Director
DHT4C: Division of Infection
Control Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260310
Device Name
HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL);
HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
Indications for Use (Describe)
HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use AAMI PB70 Level 2 Procedure Gown worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients.
Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs, using ASTM F739, Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted. Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period:
Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm)
Azacytidine (Vidaza), 25 mg/mL (25,000 ppm)
Bendamustine HCl, 25 mg/mL (25,000 ppm)
Bleomycin Sulfate, 15 mg/mL (15,000 ppm)
Bortezomib, 1 mg/mL (1,000 ppm)
Busulfan, 6 mg/mL (6,000 ppm)
Carboplatin, 10 mg/mL (10,000 ppm)
Car Izomib, 2 mg/mL (2,000 ppm)
Carmustine, 3.3 mg/mL (3,300 ppm)
Cladribine, 1 mg/mL (1,000 ppm)
Cyclosporin A, 100 mg/mL (100,000 ppm)
Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm)
Dacarbazine (DTIC), 10 mg/mL (10,000 ppm)
Dactinomycin, 0.5 mg/mL (500 ppm)
Daunorubicin HCl, 5 mg/mL (5,000 ppm)
Decitabine, 5 mg/mL (5,000 ppm)
Docetaxel, 10 mg/mL (10,000 ppm)
Doxorubicin HCl, 2 mg/mL (2,000 ppm)
Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm)
Eribulin Mesylate, 0.5 mg/mL (500 ppm)
Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm)
Gemcitabine HCL, 38 mg/mL (38,000 ppm)
Idarubicin HCL, 1 mg/mL (1,000 ppm)
Lenvatinib, 20 mg/mL (20,000 ppm)
Leuprolide acetate salt, 5 mg/mL (5,000 ppm)
Mechlorethamine HCl, 1 mg/mL (1,000 ppm)
Melphalan HCl, 5 mg/mL (5,000 ppm)
Methotrexate, 25 mg/mL (25,000 ppm)
FORM FDA 3881 (8/23)
Page 1 of 3
PSC Publishing Services (301) 443-6740
EF
{5}
Mitomycin C, 0.5 mg/mL (500 ppm)
Mitoxantrone HCl, 2 mg/mL (2,000 ppm)
Oxaliplatin, 5 mg/mL (5,000 ppm)
Pemetrexed, 25 mg/mL (25,000 ppm)
Raltitrexed, 0.5 mg/mL (500 ppm)
Retrovir, 10 mg/mL (10,000 ppm)
Rituximab, 10 mg/mL (10,000 ppm)
Sorafenib Tosylate, 200 mg/mL (200,000 ppm)
Tamoxifen, 2 mg/mL (2,000 ppm)
Thiotepa, 10 mg/mL (10,000 ppm)
Topotecan HCL, 1 mg/mL (1,000 ppm)
Vinblastine Sulfate, 1 mg/mL (1,000 ppm)
Vinorelbine, 10 mg/mL (10,000 ppm)
# Caution:
The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:
Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutes
Etoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutes
Fludarabine, 25 mg/ml (25,000 ppm) 120 minutes
Ifosfamide, 50 mg/ml (50,000 ppm) 30 minutes
Irinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutes
Paclitaxel, 6 mg/ml (6,000 ppm) 120 minutes
Vincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutes
Per ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of 0.1μg/cm2/minute:
Cyclophosphamide, at a concentration of 20 mg/mL (20,000 ppm), demonstrated a breakthrough detection time of 120 minutes. The cumulative permeation was 0.00 μg/cm² at 30 minutes, 0.00 μg/cm² at 60 minutes, and 516.00 μg/cm² at 240 minutes.
Cytarabine HCl (Cytosine), at a concentration of 100 mg/mL (100,000 ppm), demonstrated a breakthrough detection time of 60 minutes. The cumulative permeation was 0.26 μg/cm² at 30 minutes, 0.46 μg/cm² at 60 minutes, and 68.35 μg/cm² at 240 minutes.
Fludarabine, at a concentration of 25 mg/mL (25,000 ppm), demonstrated a breakthrough detection time of 120 minutes. The cumulative permeation was 0.08 μg/cm² at 30 minutes, 0.27 μg/cm² at 60 minutes, and 7.28 μg/cm² at 240 minutes.
Irinotecan HCl, at a concentration of 20 mg/mL (20,000 ppm), demonstrated a breakthrough detection time of 240 minutes. The cumulative permeation was 0.03 μg/cm² at 30 minutes, 0.10 μg/cm² at 60 minutes, and 0.30 μg/cm² at 240 minutes.
Paclitaxel (Taxol), at a concentration of 6 mg/mL (6,000 ppm), demonstrated a breakthrough detection time of 120 minutes. The cumulative permeation was 0.00 μg/cm² at 30 minutes, 0.00 μg/cm² at 60 minutes, and 1.64 μg/cm² at 240 minutes.
The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes:
Fentanyl Citrate Injection, 100 mcg/2 ml
Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 2 of 3
PSC Publishing Services (301) 443-6740
EF
{6}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 3 of 3
{7}
Owens & Minor
HALYARD
# 510(k) Summary
510(k)
Number: K260310
Date
Prepared: October 27, 2026
Submitter:
O&M Halyard Inc.
1220 Old Alpharetta Road, Suite 320
Alpharetta, GA 30005
Regulatory
Contact: Natrisya Nur
natrisya.nur@owens-minor.com
Device Trade
Names: HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs (L, XXL)
Device
Common Name: Procedure Gown
FDA Device Product Code: QSO
FDA Device Classification: Class II
FDA
Regulation Number: 21 CFR 878.4040
Predicate
Device: K230135 Copioumed Chemotherapy Isolation Gown
Page 1
{8}
Owens & Minor
HALYARD
# **Device**
# **Descriptions:**
The HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs is a blue, full-back, non-sterile, single-use disposable gown available in sizes L and XXL. The gown is constructed of polypropylene fabric laminated with a polypropylene-polyethylene film and features a neck tape closure, front tie, and knit cuffs.
The HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs is a blue, open-back, non-sterile, single-use disposable gown available in sizes L and XXL. The gown is constructed of polypropylene fabric laminated with a polypropylene-polyethylene film and features an over-the-head design with knit cuffs.
# **Indications for Use**
HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use, AAMI PB70 Level 2 Procedure Gowns worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients.
Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs, using ASTM F739, Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted.
Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period:
Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm)
Azacytidine (Vidaza), 25 mg/mL (25,000 ppm)
Bendamustine HCl, 25 mg/mL (25,000 ppm)
Bleomycin Sulfate, 15 mg/mL (15,000 ppm)
Bortezomib, 1 mg/mL (1,000 ppm)
Busulfan, 6 mg/mL (6,000 ppm)
Carboplatin, 10 mg/mL (10,000 ppm)
Car Izzomib, 2 mg/mL (2,000 ppm)
Carmustine, 3.3 mg/mL (3,300 ppm)
Cladribine, 1 mg/mL (1,000 ppm)
Cyclosporin A, 100 mg/mL (100,000 ppm)
Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm)
Dacarbazine (DTIC), 10 mg/mL (10,000 ppm)
Dactinomycin, 0.5 mg/mL (500 ppm)
Daunorubicin HCl, 5 mg/mL (5,000 ppm)
Decitabine, 5 mg/mL (5,000 ppm)
Docetaxel, 10 mg/mL (10,000 ppm)
Doxorubicin HCl, 2 mg/mL (2,000 ppm)
Page 2
{9}
[Non-Text]
Owens & Minor
[Non-Text]
HALYARD
Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm)
Eribulin Mesylate, 0.5 mg/mL (500 ppm)
Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm)
Gemcitabine HCL, 38 mg/mL (38,000 ppm)
Idarubicin HCL, 1 mg/mL (1,000 ppm)
Lenvatinib, 20 mg/mL (20,000 ppm)
Leuprolide acetate salt, 5 mg/mL (5,000 ppm)
Mechlorethamine HCl, 1 mg/mL (1,000 ppm)
Melphalan HCl, 5 mg/mL (5,000 ppm)
Methotrexate, 25 mg/mL (25,000 ppm)
Mitomycin C, 0.5 mg/mL (500 ppm)
Mitoxantrone HCl, 2 mg/mL (2,000 ppm)
Oxaliplatin, 5 mg/mL (5,000 ppm)
Pemetrexed, 25 mg/mL (25,000 ppm)
Raltitrexed, 0.5 mg/mL (500 ppm)
Retrovir, 10 mg/mL (10,000 ppm)
Rituximab, 10 mg/mL (10,000 ppm)
Sorafenib Tosylate, 200 mg/mL (200,000 ppm)
Tamoxifen, 2 mg/mL (2,000 ppm)
Thiotepa, 10 mg/mL (10,000 ppm)
Topotecan HCL, 1 mg/mL (1,000 ppm)
Vinblastine Sulfate, 1 mg/mL (1,000 ppm)
Vinorelbine, 10 mg/mL (10,000 ppm)
### Caution:
The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:
Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutes
Etoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutes
Fludarabine, 25 mg/ml (25,000 ppm) 120 minutes
Ifosfamide, 50 mg/ml (50,000 ppm) 30 minutes
Irinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutes
Paclitaxel, 6 mg/ml (6,000 ppm) 120 minutes
Vincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutes
Per ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of \( 0.1 \mu g/cm^{2}/minute \)
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) | | |
| --- | --- | --- | --- | --- |
| | | 30 min | 60 min | 240 min |
| Cyclophosphamide, 20 mg/mL (20,000 ppm) | 120 minutes | 0.00 | 0.00 | 516.00 |
| Cytarabine HCl (Cytosine), 100 mg/mL (100,000 ppm) | 60 minutes | 0.26 | 0.46 | 68.35 |
| Fludarabine, 25 mg/mL (25,000 ppm) | 120 minutes | 0.08 | 0.27 | 7.28 |
Page 3
{10}
[LOGO]
Owens & Minor
[LOGO]
HALYARD
| Irinotecan HCL, 20 mg/mL (20,000 ppm) | 240 minutes | 0.03 | 0.10 | 0.30 |
| --- | --- | --- | --- | --- |
| Paclitaxel (Taxol), 6 mg/mL (6,000 ppm) | 120 minutes | 0.00 | 0.00 | 1.64 |
The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes:
Fentanyl Citrate Injection, 100 mcg/2 ml
Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution
Page 4
{11}
Owens & Minor HALYARD
### Comparison of Technological Characteristics
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| Product Name | HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs. | Copioumed Chemotherapy Isolation Gown | Different |
| Intended Use/ Indication for Use | HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs are non-sterile, disposable, single-use AAMI PB70 Level 2 Procedure Gown worn by healthcare workers for minimal to low fluid contact situations involving water, blood, bodily fluids, or other fluids for general infection control and/or staff protection. Also, can be worn by healthcare workers during the administration of chemotherapy drugs to patients. Both the HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, *Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs*, using ASTM F739, *Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact*. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted. Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period: Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm) Azacytidine (Vidaza), 25 mg/mL (25,000 ppm) Bendamustine HCl, 25 mg/mL (25,000 ppm) Bleomycin Sulfate, 15 mg/mL (15,000 ppm) Bortezomib, 1 mg/mL (1,000 ppm) Busulfan, 6 mg/mL (6,000 ppm) | Copioumed Chemotherapy Isolation Gown is intended to be worn by healthcare personnel to protect health care patients and health care personnel from the transfer of microorganisms, body fluids and particulate. Copioumed Chemotherapy Isolation Gown is non-sterile and for single use only. Copioumed Chemotherapy Isolation Gown meets the requirements of an AAMI Level 3 barrier protection for an isolation gown per ANSI/AAMI PB70:2012 Liquid Barrier Performance Classification of Protective Apparel Drapes Intended for Use in Health Care Facilities (ANSI/AAMI PB70). Copioumed Chemotherapy Isolation Gown is a disposable device intended to protect health care personal from exposure to chemotherapy drugs during preparation, handling, and administration. The Copioumed Chemotherapy Isolation Gown has been evaluated for resistance to permeation of various chemotherapy drugs | Similar |
Page 5
{12}
Owens & Minor
HALYARD
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| | Carboplatin, 10 mg/mL (10,000 ppm) Car Ixomib, 2 mg/mL (2,000 ppm) Carmustine, 3.3 mg/mL (3,300 ppm) Cladribine, 1 mg/mL (1,000 ppm) Cyclosporin A, 100 mg/mL (100,000 ppm) Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm) Dacarbazine (DTIC), 10 mg/mL (10,000 ppm) Dactinomycin, 0.5 mg/mL (500 ppm) Daunorubicin HCl, 5 mg/mL (5,000 ppm) Decitabine, 5 mg/mL (5,000 ppm) Docetaxel, 10 mg/mL (10,000 ppm) Doxorubicin HCl, 2 mg/mL (2,000 ppm) Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm) Eribulin Mesylate, 0.5 mg/mL (500 ppm) Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm) Gemcitabine HCL, 38 mg/mL (38,000 ppm) Idarubicin HCL, 1 mg/mL (1,000 ppm) Lenvatinib, 20 mg/mL (20,000 ppm) Leuprolide acetate salt, 5 mg/mL (5,000 ppm) Mechlorethamine HCl, 1 mg/mL (1,000 ppm) Melphalan HCl, 5 mg/mL (5,000 ppm) Methotrexate, 25 mg/mL (25,000 ppm) Mitomycin C, 0.5 mg/mL (500 ppm) Mitoxantrone HCl, 2 mg/mL (2,000 ppm) Oxaliplatin, 5 mg/mL (5,000 ppm) Pemetrexed, 25 mg/mL (25,000 ppm) Raltitrexed, 0.5 mg/mL (500 ppm) Retrovir, 10 mg/mL (10,000 ppm) Rituximab, 10 mg/mL (10,000 ppm) Sorafenib Tosylate, 200 mg/mL (200,000 ppm) Tamoxifen, 2 mg/mL (2,000 ppm) Thiotepa, 10 mg/mL (10,000 ppm) Topotecan HCL, 1 mg/mL (1,000 ppm) Vinblastine Sulfate, 1 mg/mL (1,000 ppm) | per ASTM F739-20 "Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact." Chemotherapy drug permeation resistance, average standardized breakthrough time in minutes.*: Carmustine [3.3 mg/ml] >480 Cisplatin [1.0 mg/ml] >480 Cyclophosphamide [20 mg/ml] >480 Cytarabine Hydrochloride [100 mg/ml] >480 Dacarbazine [10 mg/ml] >480 Daunorubicin Hydrochloride [5 mg/ml] >480 Doxorubicin Hydrochloride [2 mg/ml] > 480 Etoposide [20 mg/ml] >480 Fluorouracil [50 mg/ml] >480 Ifosfamide [50 mg/ml] > 480 Methotrexate [25 mg/ml] >480 Mitomycin C [0.5 mg/ml] >480 Mitoxantrone [2.0 mg/ml] >480 Paclitaxel [6 mg/ml] >480 Thiotepa [10 mg/ml] >480 Vincristine Sulfate [1 mg/ml] >480 *No permeation was detected at either the minimum detectable permeation or 0.1 μg/cm²/min | |
Page 6
{13}
Owens & Minor
O
HALYARD
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| | Vinorelbine, 10 mg/mL (10,000 ppm)Caution:The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutesEtoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutesFludarabine, 25 mg/ml (25,000 ppm) 120 minutesIfosfamide, 50 mg/ml (50,000 ppm) 30 minutesIrinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutesPaclitaxel, 6 mg/ml (6,000 ppm) 120 minutesVincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutesPer ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of \( 0.1\mu g/cm^{2}/minute \) | | |
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) |
| 30 min | 60 min | 240 min |
| Cyclophosphamide, 20 mg/mL (20,000 ppm) | 120 minutes | 0.00 | 0.00 | 516.00 |
| Cytarabine HCl (Cytosine), 100 mg/mL (100,000 ppm) | 60 minutes | 0.26 | 0.46 | 68.35 |
| Fludarabine, 25 mg/mL (25,000 ppm) | 120 minutes | 0.08 | 0.27 | 7.28 |
| Irinotecan HCL, 20 mg/mL (20,000 ppm) | 240 minutes | 0.03 | 0.10 | 0.30 |
| Paclitaxel (Taxol), 6 mg/mL (6,000 ppm) | 120 minutes | 0.00 | 0.00 | 1.64 |
Page 7
{14}
[LOGO]
Owens & Minor
[LOGO]
HALYARD
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| | The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes: Fentanyl Citrate Injection, 100 mcg/2 ml Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution | | |
| Color | Blue | Blue | Same |
| Design Features | HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs is a full back gown with a neck tape closure, front tie, and knit cuffs. HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs is an open back, over the head gown with knit cuffs. | Knit cuff in three style(model) Tie neck closure, Velcro neck closure and Sticker neck closure | Similar |
| Sizes | Large, XX-Large | Large, X-Large | Similar |
| Materials | 1.02-1.42 osy Polypropylene Gown laminated with polypropylene-polyethylene film. | Polypropylene Gown laminated with polyethylene | Similar |
| Performance per AAMI PB70 | AAMI Level 2 | AAMI Level 3 | Different^{1} |
| Resistance to permeation of chemotherapy drugs testing | HALYARD* CHEMO360* Procedure Gown Tested for Use with Chemotherapy Drugs and the HALYARD* Procedure Gown Tested for Use with Chemotherapy Drugs were evaluated for resistance to permeation by selected chemotherapy drugs in accordance with ASTM F3267, *Standard Specification for Protective Clothing for Use Against Liquid Chemotherapy Drugs*, using ASTM F739, *Standard Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact*. The gowns meet the requirements for Part A: Selective Protection under ASTM F3267 based on the chemotherapy drug permeation testing conducted. Chemotherapy drug permeation resistance was reported as the average standardized breakthrough time over a 480-minute test period: Arsenic Trioxide (Trisenox), 1 mg/mL (1,000 ppm) Azacytidine (Vidaza), 25 mg/mL (25,000 ppm) Bendamustine HCl, 25 mg/mL (25,000 ppm) Bleomycin Sulfate, 15 mg/mL (15,000 ppm) Bortezomib, 1 mg/mL (1,000 ppm) Busulfan, 6 mg/mL (6,000 ppm) Carboplatin, 10 mg/mL (10,000 ppm) | ASTM F739-20 Results showed no average Standardized breakthrough for up to 480 minutes for claimed 16 drugs. Chemotherapy drug permeation resistance, average standardized breakthrough time in minutes.*: Carmustine [3.3 mg/ml] >480 Cisplatin [1.0 mg/ml] >480 Cyclophosphamide [20 mg/ml]>480 Cytarabine Hydrochloride [100 mg/ml] >480 Dacarbazine [10 mg/ml] >480 Daunorubicin Hydrochloride [5 mg/ml] >480 Doxorubicin Hydrochloride [2 mg/ml] >480 | Similar^{2} |
Page 8
{15}
Owens & Minor
HALYARD
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| | Car Izomib, 2 mg/mL (2,000 ppm) Carmustine, 3.3 mg/mL (3,300 ppm) Cladribine, 1 mg/mL (1,000 ppm) Cyclosporin A, 100 mg/mL (100,000 ppm) Cytovene (Ganciclovir), 10 mg/mL (10,000 ppm) Dacarbazine (DTIC), 10 mg/mL (10,000 ppm) Dactinomycin, 0.5 mg/mL (500 ppm) Daunorubicin HCl, 5 mg/mL (5,000 ppm) Decitabine, 5 mg/mL (5,000 ppm) Docetaxel, 10 mg/mL (10,000 ppm) Doxorubicin HCl, 2 mg/mL (2,000 ppm) Epirubicin HCl (Ellence), 2 mg/mL (2,000 ppm) Eribulin Mesylate, 0.5 mg/mL (500 ppm) Fluorouracil (Adrucil), 50 mg/mL (50,000 ppm) Gemcitabine HCL, 38 mg/mL (38,000 ppm) Idarubicin HCL, 1 mg/mL (1,000 ppm) Lenvatinib, 20 mg/mL (20,000 ppm) Leuprolide acetate salt, 5 mg/mL (5,000 ppm) Mechlorethamine HCl, 1 mg/mL (1,000 ppm) Melphalan HCl, 5 mg/mL (5,000 ppm) Methotrexate, 25 mg/mL (25,000 ppm) Mitomycin C, 0.5 mg/mL (500 ppm) Mitoxantrone HCl, 2 mg/mL (2,000 ppm) Oxaliplatin, 5 mg/mL (5,000 ppm) Pemetrexed, 25 mg/mL (25,000 ppm) Raltitrexed, 0.5 mg/mL (500 ppm) Retrovir, 10 mg/mL (10,000 ppm) Rituximab, 10 mg/mL (10,000 ppm) Sorafenib Tosylate, 200 mg/mL (200,000 ppm) Tamoxifen, 2 mg/mL (2,000 ppm) Thiotepa, 10 mg/mL (10,000 ppm) Topotecan HCL, 1 mg/mL (1,000 ppm) Vinblastine Sulfate, 1 mg/mL (1,000 ppm) Vinorelbine, 10 mg/mL (10,000 ppm) | Etoposide [20 mg/ml] >480 Fluorouracil [50 mg/ml] >480 Ifosfamide [50 mg/ml] > 480 Methotrexate [25 mg/ml] >480 Mitomycin C [0.5 mg/ml] >480 Mitoxantrone [2.0 mg/ml] >480 Paclitaxel [6 mg/ml] >480 Thiotepa [10 mg/ml] >480 Vincristine Sulfate [1 mg/ml]>480 *No permeation was detected at either the minimum detectable permeation or 0.1 μg/cm2/min | |
Page 9
{16}
Owens & Minor
O
HALYARD
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| | Caution:The following chemotherapy drugs and concentration showed breakthrough detected in less than 480 minutes:Cytarabine (Cytosine), 100 mg/ml (100,000 ppm) 60 minutesEtoposide (Toposar), 20 mg/ml (20,000 ppm) <30 minutesFludarabine, 25 mg/ml (25,000 ppm) 120 minutesIfosfamide, 50 mg/ml (50,000 ppm) 30 minutesIrinotecan HCl, 20 mg/ml (20,000 ppm) 240 minutesPaclitaxel, 6 mg/ml (6,000 ppm) 120 minutesVincristine sulfate, 1 mg/ml (1,000 ppm) <30 minutesPer ASTM F3267-22, the following chemotherapy drugs had the following breakthrough at the Standard Breakthrough Rate of \( 0.1\mu g/cm^{2}/minute \) | | |
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) |
| 30 min | 60 min | 240 min |
| Cyclophosphamide, 20 mg/mL (20,000 ppm) | 120 minutes | 0.00 | 0.00 | 516.00 |
| Cytarabine HCl (Cytosine), 100 mg/mL (100,000 ppm) | 60 minutes | 0.26 | 0.46 | 68.35 |
| Chemotherapy Drugs | Breakthrough Detection Time (min) | Cumulative Permeation (\( \mu g/cm^{2} \)) |
| 30 min | 60 min | 240 min |
| Fludarabine, 25 mg/mL (25,000 ppm) | 120 minutes | 0.08 | 0.27 | 7.28 |
| Irinotecan HCL, 20 mg/mL (20,000 ppm) | 240 minutes | 0.03 | 0.10 | 0.30 |
| Paclitaxel (Taxol), 6 mg/mL (6,000 ppm) | 120 minutes | 0.00 | 0.00 | 1.64 |
Page 10
{17}
Owens & Minor
HALYARD
| Device Characteristic | Proposed Device | Predicate Device | Comparison Analysis |
| --- | --- | --- | --- |
| | The following hazardous drugs (opioids) and concentration had NO breakthrough detected within 480 minutes: Fentanyl Citrate Injection, 100 mcg/2 ml Fentanyl Citrate Injection / Simulated Gastric Acid Fluid Mix, 50/50 solution | | |
| Biocompatibility | Under the test conditions, the subject device was shown to be non- cytotoxic, non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10 | Under the test conditions, the subject device was shown to be non- cytotoxic, non-irritating and non-sensitizing per ISO 10993-5 & ISO 10993-10 | Same |
| Flammability | Meets requirements of Flame Resistant CPSC 16 CFR 1610 Class 1 | Meets requirements of Flame Resistant CPSC 16 CFR 1610 Class 1 | Same |
| Tensile Strength | Meets the requirement of ASTM D5034-9 | Meets the requirement of ASTM D5034-9 | Same |
| Tear Resistance | Meets the requirement of ASTM D5587-15 | Meets the requirement of ASTM D5587-15 | Same |
| Seam Strength | Meets the requirement of ASTM 1683-17 | Meets the requirement of ASTM 1683-17 | Same |
Different¹ While the proposed device and the predicate device differ in AAMI levels of protection, the predicate device was selected for the purpose of demonstrating evaluation of resistance to permeation against various chemotherapy drugs. The difference in the AAMI protection level is addressed through labeling.
Similar² The predicate device was selected solely to demonstrate evaluation for resistance to permeation by various chemotherapy drugs. Both the proposed device and the predicate device were tested for resistance to permeation of chemotherapy drugs in accordance with the standard ASTM F739, Test Method for Permeation of Liquids and Gases Through Protective Clothing Materials Under Conditions of Continuous Contact. The proposed device was evaluated against a broader range of chemotherapy drugs than the predicate device. All chemotherapy drugs tested for the proposed device are clearly identified in the product labeling
Page 11
{18}
Owens & Minor
HALYARD
### Summary of Non-Clinical Testing
The subject device was evaluated through applicable non-clinical testing, and the results support the safety and performance of the device for its intended use.
| Standard | Name | Acceptance Criteria | Result |
| --- | --- | --- | --- |
| ISO 10993-5 | Tests for In Vitro Cytotoxicity | Non-Cytotoxicity | Pass |
| ISO 10993-10 | Tests for Skin Irritation and Sensitization | Non-Irritation and Non-Sensitizing | Pass |
| AATCC 42-17 | Water Resistance | AATCC 42 ≤ 1.0g | Pass |
| AATCC 127-18 | Hydrostatic Pressure | AATCC 127 ≥ 50 cm H_{2}O | Pass |
| ASTM D5034-09 | Breaking Strength and Elongation of Textile Fabrics (Grab Test) | ≥ 30N | Pass |
| ASTM D5587-15 | Tearing Strength of Fabrics by Trapezoid Procedure | ≥ 10N | Pass |
| ASTM D1683 | Seam strength | ≥ 30N | Pass |
| 16 CFR 1610 | Flammability of Clothing Textiles | Class 1 (Normal flammability) | Pass |
| ANSI AAMI PB70:2022 | Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Healthcare Facilities | Level 2 | Pass |
| ASTM F739:2020 | Permeation of Liquids and Gasses Through Protective Clothing | No permeation above detection limit during test period or breakthrough time ≥ specified time | Pass |
### Summary of Clinical Testing
No clinical testing was conducted for the subject device. Clinical testing was not considered necessary to support the safety and effectiveness of the device, as the relevant safety and performance characteristics can be adequately demonstrated through non-clinical testing
### Conclusion
The conclusions drawn from the non-clinical tests demonstrate that the subject device, the HALYARD* CHEMO360* Procedure Gown Tested for use with Chemotherapy Drugs and HALYARD* Procedure Gown Tested for use with Chemotherapy Drugs, is as safe, as effective, and performs as well as or better than the legally marketed predicate device in K230135, with respects to resistance to permeation of chemotherapy drugs and Fentanyl Citrate.
Page 12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.