K260295 · Motion Composites · IOR · May 28, 2026 · Physical Medicine
Device Facts
Record ID
K260295
Device Name
Helio Kids; Apex P
Applicant
Motion Composites
Product Code
IOR · Physical Medicine
Decision Date
May 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.3850
Device Class
Class 1
Attributes
Pediatric
Indications for Use
Helio Kids: The Helio Kids is indicated as a manual wheelchair to provide mobility for pediatric persons (23 kg to 75 kg) limited to a sitting position. Apex P: The Apex P is indicated as a manual wheelchair to provide mobility for pediatric persons (23 kg to 75 kg) limited to a sitting position.
Device Story
Helio Kids and Apex P are non-powered manual wheelchairs; pediatric versions of existing adult Helio and Apex families. Helio Kids is foldable; Apex P is non-foldable. Constructed from aluminum or carbon fiber. Components include frame, seat, backrest, wheels, handrims, casters, wheel locks, anti-tippers, armrests, and footrests. Operated by patient or caregiver; used in clinical or home environments. Provides mobility for pediatric users limited to sitting. No electronic or software components; purely mechanical operation.
Clinical Evidence
Bench testing only. No clinical data. Performance evaluated against ISO 7176 series (Parts 1, 3, 5, 7, 8, 11, 15, 19, 22) and RESNA WC-4 Section 19 for static stability, brake effectiveness, dimensions, fatigue strength, and transit safety.
Technological Characteristics
Manual, non-powered wheelchair. Materials: Aluminum or Hi-Modulus Carbon fiber with Epoxy Resin. Foldable (Helio Kids) or non-foldable (Apex P) frame. Adjustable seat depth, width, back height, and angle. Complies with ISO 7176 and RESNA WC-4 standards. No energy source, connectivity, or software.
Indications for Use
Indicated for pediatric patients (23 kg to 75 kg) limited to a sitting position requiring a manual wheelchair for mobility.
Regulatory Classification
Identification
A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
{1}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 28, 2026
Motion Composites
Catherine Marquis
Regulatory Affairs & Quality Assurance Manager
160 Armand-Majeau Sud
Saint-Roch-Del’Achigan, QC JOK 3H0
Canada
Re: K260295
Trade/Device Name: Helio Kids; Apex P
Regulation Number: 21 CFR 890.3850
Regulation Name: Mechanical Wheelchair
Regulatory Class: Class I, reserved
Product Code: IOR
Dated: April 28, 2026
Received: April 28, 2026
Dear Catherine Marquis:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260295 - Catherine Marquis
Page 2
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260295 - Catherine Marquis
Page 3
Sincerely,
Tushar Bansal -S
Tushar Bansal, PhD
Acting Assistant Director, Acute Injury Devices Team
DHT5B: Division of Neuromodulation and
Physical Medicine Devices
OHT5: Office of Neurological and
Physical Medicine Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{4}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260295 | ? |
| Please provide the device trade name(s). | | ? |
| Helio Kids ; Apex P | | |
| Please provide your Indications for Use below. | | ? |
| Helio Kids: The Helio Kids is indicated as a manual wheelchair to provide mobility for pediatric persons (23 kg to 75 kg) limited to a sitting position. Apex P: The Apex P is indicated as a manual wheelchair to provide mobility for pediatric persons (23 kg to 75 kg) limited to a sitting position. | | |
| Please select the types of uses (select one or both, as applicable). | ☐ Prescription Use (21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{5}
510(k) Summary
Devices: Helio Kids & Apex P
Applicant/Manufacturer: Motion Composites
Company Address: 160 Armand-Majeau Sud
Saint-Roch-de l'Achigan, Quebec
Canada JOK 3H0
Contact Person: Catherine Marquis
Date Summary Prepared: January 29, 2026
Trade Name: Helio Kids, Apex P
Classification Name: Manual Wheelchair
Product Code(s): IOR / wheelchair, mechanical
Device Class: Class I
Regulation Number: 21 CFR 890.3850
Predicate Devices:
| Trade name | Primary Predicate #1 Helio C2, Helio A7 | Primary Predicate #2 Apex A, Apex C | Secondary Predicate Ki Mobility Spark |
| --- | --- | --- | --- |
| 510(k) | K120628, K143101 | K161425 | K223527 |
| Manufacturer | Motion Composites | Motion Composites | Ki Mobility LLC |
| Class | I | I | I |
| Product Code / Name | IOR / wheelchair, mechanical | IOR / wheelchair, mechanical | IOR / wheelchair, mechanical |
| Classification | 890.3850 / Manual Wheelchair | 890.3850 / Manual Wheelchair | 890.3850 / ManualWheelchair |
Device Description
The Helio Kids and Apex P are mechanical non-powered wheelchairs respectively based on the current Helio and Apex families of wheelchairs. They are new pediatric sizes to extend the current adult sizes in each of the families. The Helio family are of the foldable type while the Apex type is non-foldable. They are provided in carbon or aluminum versions.
Comparison of intended use - indications for use
As compared in the table below to the primary and secondary predicates, the intended use as manual wheelchairs is unchanged. The indications in both the Helio Kids Apex P have been modified from that in the adult Helio and Apex adult versions to being for pediatric populations per their modified size. This indication is equivalent to the Ki Mobility Spark indication.
{6}
| Intended Use / Indications | | | |
| --- | --- | --- | --- |
| K260295 Helio Kids, Apex P | Primary Predicate #1 Helio C2, Helio A7 | Primary Predicate #2 Apex A, Apex C | Secondary Predicate Ki Mobility Spark |
| **Helio Kids:** The Helio Kids is indicated as a manual wheelchair to provide mobility for pediatric persons (23 kg to 75 kg) limited to a sitting position. **Apex P:** The Apex P is indicated as a manual wheelchair to provide mobility for pediatric persons (23 kg to 75 kg) limited to a sitting position. | **Helio C2:** The Helio wheelchair is a manually operated device intended to be used as a means of mobility for persons restricted to a sitting position. It is not indicated for the pediatric population. **Helio A7:** The indication for use of the Helio A7 manual wheelchair is to provide mobility to persons limited to a sitting position. | **Apex A & C:** The indications for use of the APEX manual wheelchair are to provide mobility to persons limited to a sitting position. | **Ki Mobility Spark:** The Ki Mobility Spark manual wheelchair is intended to provide mobility to pediatrics limited to a seating position. |
## Comparison of Technological Characteristics
As detailed in the comparison table shown below, the proposed pediatric Helio Kids and Apex P wheelchairs present the same main features as the predicates including: main frame upon which all components are assembled, seat with backrest, wheels with handrims and casters to each side, back canes / push handles, wheelocks, anti-tippers to the back, armrests, and footplates and leg rests.
As compared to each of their corresponding current Helio C2/A7 and Apex A/C adult predicate models, the base wheelchair components in the proposed pediatric Helio Kids and Apex P models are the same except that the frame/back width's and seat depth's ranges and weight limits have been reduced corresponding with compatible smaller pediatric users starting around 51 lbs (23 kg) up to 165 lbs (75kg).
As compared to the Ki Mobility Spark predicate for pediatric use, it presents the same main features with similar component and adjustment options. The differences are detailed engineering implementation differences or of a sizing nature without presenting any differences in the main technology or means of operation.
{7}
| | APEX P | Apex A or C | Helio Kids | Helio A7 or C2 | Ki Mobility Spark |
| --- | --- | --- | --- | --- | --- |
| Type | Non-foldable | Non-foldable | foldable | foldable | foldable |
| Weight Limit | 165 lbs. | 265 lbs | 165 lbs. | 265 lbs (also 350 lbs Heavy Duty option) | 165 lbs. |
| Product Weight | 24.8 lbs (standard configuration) | 25.9 lbs (A) / 25.0 lbs (C) (standard configuration) | 29.8 lbs (standard configuration) | 31.2 lbs (A7) / 27.5 lbs (C2) (standard configuration) | 28.7 lbs |
| Transit approved | yes | yes | yes | yes | yes |
| Primary Frame Materials | Aluminum (Upgrade of Carbon Fiber) | Aluminum (for A) / Carbon (for C) | Aluminum (Upgrade of Carbon Fiber) | Aluminum (for A7) / Hi-Modulus Carbon fiber + Epoxy Resin (for C2) | Aluminum |
| Frame Widths | 10" to 15" | 12" to 20" | 12" to 16" | 14" to 22" | 10"-16" |
| Seat Depths | 10" to 15" | 12" to 20" | 12" to 19" | 14" to 20" | 12"-18" |
| Back Heights | 9" to 21" (adjustable angle 80° to 101°) | 9" to 21" (adjustable angle 80° to 101°) | 9" to 21" (adjustable angle 90° to 105°) | 9" to 21" (adjustable angle 85° to 110°) | 13" to 24" (adjustable angle 85° to 115°) |
| Seat Height | 14" to 21" | 14" to 21" | 13" to 21.5" | 13" to 21.5"" | (13"-21") Front Seat-to-Floor (11"-18.5") Rear Seat-to-Floor |
| Armrests: Type | Optional T-Shaped armrests, removable, height adjustable (by user, no tool required) or tubular swing away armrests or flip-up armrests | Optional T-Shaped armrests, removable, height adjustable (by user, no tool required) or tubular swing away armrests flip-up armrests | Optional T-Shaped or flip-back armrests, removable, height adjustable (by user, no tool required) or tubular swing away armrests or flip-up armrests | Optional T-Shaped or flip-back armrests, removable, height adjustable (by user, no tool required) or tubular swing away armrests or flip-up armrests | Single Post Height Adjustable Height Adjustable T-Arm Angle Adjustable Locking Extendable Flip Up Pediatric Height Adjustable T-Arm Tubular Flip up |
| Back Type | Angle adjustable, straight back canes with optional push handles. Optional height adjustable. Optional Depth adjustable. | Angle adjustable, straight back canes with optional push handles. Optional height adjustable. Optional Depth adjustable. | Height adjustable, straight or bended back canes with optional push handles. Optional angle adjustable. Optional Depth adjustable | Height adjustable, straight or bended back canes with optional push handles. Optional angle adjustable Optional Depth adjustable | Stroller Handle Back Post Height Adjustable Straight with Push Handle Dynamic Rocker Back Angle adjustable |
| Footrests/L egress | Integrated to the frame Adjustable Height | Integrated to the frame Adjustable Height | Swing-in / Swing-out Flip in, flip out footrest; 60", 70",80", 90". Optional elevating legress or residual limb support | Swing-in / Swing-out Flip in, flip out; 60", 70",80", 90". Optional elevating legress or residual limb support | Swing Away Elevating Leg Rest PRO, PRO Pediatric) Residual Limb Support Adjustable Footplate angle adjustable |
| Footplates Type | Tubular or high mount platform Optional adjustable angle platform or Flip back platform | Tubular Optional adjustable angle platform or Flip back platform Optional high mount platform | Standard non angle adjustable Optional angle adjustable or one piece flip up platform or high mount footplates | Standard non angle adjustable Optional angle adjustable or one piece flip up platform or high mount footplates | |
| Wheel sizes | 20", 22", 24", 25", 26" | 20", 22", 24", 25", 26" | 20", 22", 24", 25", 26" | 20", 22", 24", 25", 26" | 18", 20", 22", 24" |
{8}
| | APEX P | Apex A or C | Helio Kids | Helio A7 or C2 | Ki Mobility Spark |
| --- | --- | --- | --- | --- | --- |
| Handrims | Aluminum Anodized Optional Plastic Coated (regular or high friction) Optional Spinergy Flexrim Optional Natural Fit Handrim Optional Surge Handrim | Aluminum Anodized Optional Plastic Coated (regular or high friction) Optional Spinergy Flexrim Optional Natural Fit Handrim Optional Surge Handrim | Aluminum Anodized Optional Plastic Coated (regular or high friction) Optional Spinergy Flexrim Optional Natural Fit Handrim Optional Surge Handrim | Aluminum Anodized Optional Plastic Coated (regular or high friction) Optional Spinergy Flexrim Optional Spinergy Flexrim Optional Natural Fit Handrim Optional Surge Handrim | Aluminum Anodized Aluminum with Plastic Coating Aluminum with Non-Slip Tape Projection Knob with Plastic Coating Ergonomic Thumb Grip: Flexible Polymer Ergonomic Thumb Grip: Alum with Grip Coating |
| Caster Sizes | 3" to 6" | 3" to 6" | 3" to 8" | 3" to 8" | 3" to 8" (depending on configuration/options) |
| Wheel Locks | Push to lock, pull to lock | Push to lock, pull to lock, | Push to lock, pull to lock, one arm lock | Push to lock, pull to lock, one arm lock | Push to lock, Pull to lock Push to Lock Flush Mount Hemi-Wheel Locks |
| Anti-Tipper | Various sizes, with extension and contact wheels, removable | Various sizes, with extension and contact wheels, removable | Various sizes, with extension and contact wheels, removable | Various sizes, with extension and contact wheels, removable | Removable/swing-up |
The Helio Kids and Apex P wheelchairs have been verified for their safety and effectiveness per the bench test and assessments as follows:
- Biocompatibility:
The materials that the user may contact and the applicable contact characteristics are the same and unchanged in the two proposed pediatric sizes as compared to the primary predicates (Helio C2 & A7, and Apex A & C). As such the proposed pediatric models present the same safe biocompatibility characteristics as is established in the current Helio and Apex adult models per their long history of safe use.
- Performance Evaluation:
As is detailed in the Performance Bench Test Section, the proposed pediatric sizes have been evaluated to meet the performance and safety requirements of the applicable parts of the ISO 7176 series of standards as well as RESNA WC-4 for manual wheelchairs. These include:
- ISO 7176-1 Wheelchairs - Part 1: Determination of static stability
- ISO 7176-3 Wheelchairs - Part 3: Determination of effectiveness of brakes
- ISO 7176-5 Wheelchairs - Part 5: Determination of overall dimensions, mass and manoeuvring space
- ISO 7176-7 Wheelchairs - Part 7: Measurement of seating and wheel dimensions
- ISO 7176-8 Wheelchairs - Part 8: Requirements and test methods for static, impact and fatigue strengths
- ISO 7176-11 Wheelchairs - Part 11: Test dummies
- ISO 7176-15 Wheelchairs - Part 15: Requirements for information disclosure, documentation and labeling
- ISO 7176-19 Wheelchairs - Part 19: Wheeled mobility devices for use as seats in motor vehicles
- ISO 7176-22 Wheelchairs - Part 22: Set-up procedures
- RESNA WC-04 - Section 19: Wheelchairs used as seats in motor vehicles
{9}
Conclusion on Substantial Equivalence:
The pediatric Helio Kids and Apex P manual wheelchairs are of the same design as their base adult version predicates, the Helio C2/A7 and Apec A/C version, except for given sizing differences to accommodate pediatric users. They also present equivalent features to the Ki Mobility Spark predicate as is indicated for pediatric use. The differences do not raise any different questions of safety and effectiveness as compared to either of the Helio and Apex adult sized predicates, or to the Ki Mobility Spark pediatric predicate. Safety and performance evaluations and testing demonstrated that the proposed Helio Kids and Apex P can be expected to be safe and effective equivalently as the predicates. In conclusion, the Helio Kids and Apex P as indicated for pediatric use are substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.