← Product Code [BWK](/productcode/BWK) · K260230

# Needle Stimulator (SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V) (K260230)

_Suzhou Medical Appliance Factory · BWK · Oct 22, 2026 · NE · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K260230

## Device Facts

- **Applicant:** Suzhou Medical Appliance Factory
- **Product Code:** [BWK](/productcode/BWK.md)
- **Decision Date:** Oct 22, 2026
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** NE
- **Attributes:** Therapeutic

## Indications for Use

Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

## Device Story

Needle Stimulator (models SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V) is an electro-acupuncture device used in clinical settings by qualified acupuncture practitioners. The device delivers controlled, low-frequency, charge-balanced asymmetric biphasic electrical pulses to acupuncture needles via lead wires and alligator connectors. Practitioners use the console to adjust stimulation mode (continuous, interrupted, dense-disperse), pulse frequency (1–100 Hz), treatment duration, and output intensity. The device includes a built-in timer with an audible alert at treatment completion. Power is provided by internal batteries or an external AC adapter. The device assists practitioners in delivering acupuncture stimulation, potentially benefiting patients through therapeutic electrical modulation. Output parameters are verified against predefined specifications to ensure safety and effectiveness.

## Clinical Evidence

Bench testing only. Testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2, IEC TS 60601-4-2), and software verification. Output waveform analysis was performed at 500 Ω, 2,000 Ω, and 10,000 Ω loads to verify output parameters including voltage, current, phase charge, and power density.

## Technological Characteristics

ABS plastic housing; asymmetric biphasic square wave output; transformer-based channel isolation; Type BF patient leakage current protection. Powered by 9V battery or 100-240 VAC adapter. Connectivity: standalone. Standards: IEC 60601-1, IEC 60601-2-10, IEC 60601-1-2. Microprocessor-controlled stimulation parameters.

## Predicate Devices

- Needle Stimulator (Model: CMNS6-1) ([K202861](/device/K202861.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}

FDA U.S. FOOD & DRUG
ADMINISTRATION

October 22, 2026

Suzhou Medical Appliance Factory
% Boyle Wang
General Manager
ABMED MEDICAL TECHNOLOGY PTY, Ltd.
48/6 Herbert St.
St Leonards, NSW 2065
Australia

Re: K260230

Trade/Device Name: Needle Stimulator (SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V)
Regulatory Class: Unclassified
Product Code: BWK
Dated: September 22, 2026
Received: September 22, 2026

Dear Boyle Wang:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{2}

K260230 - Boyle Wang

Page 2

When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{3}

K260230 - Boyle Wang

Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# AMBER T. BALLARD -S

Amber Ballard, PhD

Assistant Director

DHT5B: Division of Neuromodulation and

Physical Medicine Devices

OHT5: Office of Neurological and

Physical Medicine Devices

Office of Product Evaluation and Quality

Center for Devices and Radiological Health

Enclosure

{4}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

# **Indications for Use**

Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

510(k) Number (if known)
K260230

Device Name

Needle Stimulator (SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V)

Indications for Use (Describe)

Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)

☐ Over-The-Counter Use (21 CFR 801 Subpart C)

**CONTINUE ON A SEPARATE PAGE IF NEEDED.**

This section applies only to requirements of the Paperwork Reduction Act of 1995.

**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*

FORM FDA 3881 (8/23)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{5}

510(k) Summary

# 510(k) Summary
## K260230

In accordance with 21 C.F.R. §807.92(a) the following summary of information is provided:

### 1.0 Submitter's Information

Name: Suzhou Medical Appliance Factory.

Address: No.18 Huatuo Rd., SSTT, 215163 Suzhou New District, China

Tel: +86- 18451500721

Contact: Ms. Yang Zi

### Designated Submission Correspondent

Contact: Mr. Boyle Wang

Company Name: ABMED MEDICAL TECHNOLOGY PTY LTD.

Address: 48/6 Herbert Street, St Leonards NSW 2065, Australia

Telephone: + 61 432 758 222

Email: info@abmed.com.au

Date Prepared: October 16, 2026

### 2.0 Device Information

Trade name: Needle Stimulator(Model: SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V)

Common name: Electro-Acupuncture

Classification name: Stimulator, electro-acupuncture

Product code: BWK

Device Class: Unclassified

Panel: Neurology

### 3.0 Predicate Device Information

Manufacturer: Wuxi Jiajian Medical Instrument Co., Ltd.

Trade/Device Name: Needle Stimulator (Model: CMNS6-1)

510(k) number: K202861

### 4.0 Device Description

The Needle Stimulator is a family of electro-acupuncture stimulation devices

Page 1 of 20

{6}

510(k) Summary

comprising models SDZ-IIB, SDZ-IIC, SDZ-III, and SDZ-V. The devices deliver controlled low-frequency electrical pulses to acupuncture needles through lead wires and alligator-type connectors. All four models share the same intended use and fundamental operating principle, with differences in output channel configuration, user interface, and stimulation sequence timing.

Each device consists of a main console, lead wires, alligator-type connectors, and a power adapter. The console incorporates stimulation control circuitry and software that allows the practitioner to adjust the stimulation mode, pulse frequency, treatment duration, and output intensity. The housing is made of acrylonitrile butadiene styrene (ABS).

The principal model differences are summarized below:

|  Feature | SDZ-IIB | SDZ-IIC | SDZ-III | SDZ-V  |
| --- | --- | --- | --- | --- |
|  Number of output channels | 2 | 6 | 6 | 6  |
|  Display | LCD | No LCD | LCD | LCD  |
|  Mode adjustment | Buttons | Buttons | Rotary control | Buttons  |
|  Frequency/Treatment time/ Output intensity adjustment | Buttons | Rotary controls | Rotary controls | Buttons  |
|  Interrupted mode timing | 15 seconds on / 5 seconds off | 15 seconds on / 5 seconds off | 15 seconds on / 5 seconds off | 9 seconds on / 5 seconds off  |
|  Dense-disperse mode timing | 5 seconds disperse / 10 seconds dense | 5 seconds disperse / 10 seconds dense | 5 seconds disperse / 10 seconds dense | 5 seconds disperse / 9 seconds dense  |

SDZ-IIB permits simultaneous operation of up to two channels on a single patient. SDZ-IIC, SDZ-III, and SDZ-V each provide six output channels, with a maximum of three channels permitted to operate simultaneously on a single patient.

All models provide asymmetric biphasic pulse output with three selectable stimulation modes: continuous, interrupted, and dense-disperse. The selectable pulse frequency range is 1–100 Hz ±10%. In dense-disperse mode, the ratio of disperse frequency to dense frequency is 1:5.

Page 2 of 20

{7}

510(k) Summary

The devices include a built-in treatment timer. When the selected treatment time has elapsed, an audible alert indicates the end of treatment and the output intensity returns to zero. All models can be powered by an internal 9 V battery supply or the supplied power adapter, which has an input rating of 100–240 VAC, 50/60 Hz and an output rating of 9 VDC, 0.3 A.

Acupuncture needles are supplied separately and are not included with the devices that are within the scope of this submission. The manufacturer-recommended needle body dimensions are:

- Diameter of 0.22–0.24 mm with a length of 14–50 mm; or
- Diameter of 0.25–0.45 mm with a length of 14–75 mm.

The smallest needle body diameter is 0.22 mm, and the smallest needle body length is 14 mm.

The devices are intended for use in hospitals and clinics by qualified practitioners of acupuncture as determined by the states.

### 5.0 Indications for Use Statement

Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states.

### 6.0 Technological Characteristic Comparison Table

Table 1 Comparison of Technology Characteristics

Page 3 of 20

{8}

510(k) Summary

|  Item | Subject Device | Predicate Device | Comparison  |
| --- | --- | --- | --- |
|  510(k) No. | K260230 | K202861 | --  |
|  Manufacturer | Suzhou Medical Appliance Factory | Wuxi Jiajian Medical Instrument Co., Ltd. | --  |
|  Product Name | Needle Stimulator | Needle Stimulator | --  |
|  Model | SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V | CMNS6-1 | --  |
|  Product Code | BWK | BWK | Identical  |
|  Classification | Class II | Class II | Identical  |
|  Intended Use/Indication for Use | Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | Needle Stimulator is an electro-acupuncture stimulator device, which is indicated for use in the practice of acupuncture by qualified practitioners of acupuncture as determined by the states. | Identical  |
|  Patient Population | Adults aged 22 years and older | Not publicly available | The subject device is intended for use in adults aged 22 years and older, which does not raise increased safety and/or effectiveness concerns compared to the predicate device.  |
|  Power Source(s) | SDZ-V/ SDZ-III: DC 9V Type R14/UM2 ×6 or AC 100-240V; SDZ-IIC: DC 9V Type AA/UM3 or AC 100-240V; SDZ-IIB: DC 9V Type 6F22 or AC 100-240V | DC I.5Vx6TypeR14 or AC 100-240V | The power source between the subject and predicate device are similar and do not raise increased concerns of safety and/or effectiveness compared to the predicate device.  |
|  Method of Line Current Isolation | Type BF | Type BF | Identical  |

Page 4 of 20

{9}

510(k) Summary

|  - Patient Leakage Current - Normal Condition(μA) - Single Fault Condition(μA) | -- 2μA ≤50μA | -- 2μA ≤50μA | Identical  |
| --- | --- | --- | --- |
|  Average DC current through electrodes when device is on but no pulses are being applied(μA) | N/A | N/A | Identical  |
|  Number of Output Modes | 3 (continuous wave/ interrupted wave/ Dense-disperse wave) | 3 (continuous wave/ interrupted wave/ Dense-disperse wave) | Identical  |
|  Number of Output channels | SDZ-IIB: 2 (2 channels at most work together on single patient) SDZ-IIC/ SDZ-III/ SDZ-V: 6 (3 channels at most work together on single patient) | 6 (3 channels at most work together on single patient) | Different for SDZ-IIB; identical for SDZ-IIC, SDZ-III, and SDZ-V. SDZ-IIB provides two output channels, compared with six channels for the predicate device. The lower channel count limits the number of needle pairs that can be stimulated simultaneously, but does not alter the intended use, operating principle, or stimulation delivered through each active channel. SDZ-IIB does not claim the ability to provide the same simultaneous  |

Page 5 of 20

{10}

510(k) Summary

|   |  |  | treatment coverage as a three-channel configuration. Electrical performance testing supports that its output meets the predefined specifications. Therefore, the lower channel count does not raise different questions of effectiveness.  |
| --- | --- | --- | --- |
|  - Synchronous or Alternating? | Synchronous | Synchronous | Identical  |
|  - Method of Channel Isolation | Transformer | Transformer | Identical  |
|  Regulated Current or Regulated Voltage? | Voltage Control | Voltage Control | Identical  |
|  Software/Firmware/ Microprocessor Control? | Yes | Yes | Identical  |
|  Automatic Overload Trip? | No | No | Identical  |
|  Automatic No-Load Trip? | No | No | Identical  |
|  Automatic Shut Off? | Yes | Yes | Identical  |
|  User Override Control? | No | Yes | Different The user override control of the subject device is different from the  |

Page 6 of 20

{11}

510(k) Summary

|   |  |  |  | predicate device. However, both devices adopt the same fundamental stimulation output technology and provide similar therapeutic effects. In addition, the stimulation can be immediately stopped by the operator. Therefore, the difference do not raise new questions of safety or effectiveness.  |
| --- | --- | --- | --- | --- |
|  Indicator Display | On/Off Status? | Yes | Yes | Identical  |
|   |  Low Battery? | No | Yes | Different The subject devices do not have a dedicated low-battery indicator but use the same fundamental stimulation technology as the predicate device. All models also support operation with the supplied power adapter. All device models have been evaluated for electrical safety, EMC, and software verification. Therefore, this difference does not raise increased safety and/or effectiveness concerns compared to the predicate device.  |

Page 7 of 20

{12}

510(k) Summary

|   | Voltage/Current Level? | Yes | Yes | Identical  |
| --- | --- | --- | --- | --- |
|  Timer Range (minutes) |   | 1-60min | 0-60min | Similar  |
|  Compliance with Voluntary Standards? |   | IEC60601-1, IEC60601-2-10, IEC60601-1-2 | IEC60601-1, IEC60601-2-10, IEC60601-1-2 | Identical  |
|  Housing Materials& Construction |   | ABS; Injection molded | ABS;Injection molded | Identical  |
|  Waveform |   | Biphasic | Biphasic | Identical  |
|  Shape |   | Asymmetric biphasic square wave | Asymmetric biphasice square wave | Identical  |
|  Maximum Output Voltage(volts) | SDZ-IIB: Continuous: 24.4V @500Ω, 70.4V@2kΩ, 102V@10kΩ Interrupted: 24.4V @500 Ω ,70.4V@2kΩ, 102V@10kΩ Dense-Disperse: 24.4 V @500 Ω ,70.4V@2kΩ, 102V@10kΩ | 27V±10%@500Ω; 60.4V±10%@2kΩ; 75V±10%@10kΩ; | Similar The measured maximum output voltages of the subject device models were 24.4–24.8 V at 500 Ω , 69.6–70.4 V at 2 kΩ, and 102–104 V at 10 kΩ across all output modes, The predicate device specifies 27 V ±10%, 60.4 V ±10%, and 75 V ±10% at the corresponding loads. The subject device's output voltage is within the predicate specification at 500 Ω but is larger at 2 kΩ and 10 kΩ. However, each subject device model has comparable stimulation  |   |
|   |   |   |   |  SDZ-IIC: Continuous: 24.8V @500 Ω ;70.4V@2kΩ; 102V@10kΩ; Interrupted:24.4V @500 Ω ;70.4V@2kΩ; 104V@10kΩ; Dense-Disperse: 24.4V @500 Ω ;70.4V@2kΩ;  |

Page 8 of 20

{13}

510(k) Summary

|   | 104V@10kΩ. |  | output waveform, frequency, pulse duration, maximum output current, and maximum phase charge to the predicate device. Verification testing confirmed that all models and modes of the subject device met the predefined output specifications. Therefore, these difference does not raise different questions of safety or effectiveness.  |
| --- | --- | --- | --- |
|   |  SDZ-III: Continuous: 24.4V @500 Ω ;69.6V@2kΩ; 102V@10kΩ; Interrupted: 24.4V @500 Ω ;70.4V@2kΩ; 102V@10kΩ; Dense-Disperse: 24.4V @500 Ω ;69.6V@2kΩ; 102V@10kΩ;  |   |   |
|   |  SDZ-V: Continuous: 24.4V @500 Ω ;70.4V@2kΩ; 102V@10kΩ. Interrupted: 24.4V @500 Ω ;70.4V@2kΩ; 102V@10kΩ; Dense-Disperse: 24.4V @500 Ω ;70.4V@2kΩ; 102V@10kΩ;  |   |   |

Page 9 of 20

{14}

510(k) Summary

|  Maximum Output Current | SDZ-IIB: Continuous: 48.8mA @500 Ω ; 35.20mA @2kΩ; 10.2mA @10kΩ. Interrupted: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ. Dense-Disperse: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ. | 54mA±10%@500Ω; 30.2mA±10%@2kΩ; 7.5mA±10%@10 kΩ | Similar The measured maximum output currents across all subject device models and output modes were 48.8–49.6 mA at 500 Ω, 34.0–35.2 mA at 2 kΩ, and 10.2–10.4 mA at 10 kΩ. The predicate device specifies 54 mA ±10%, 30.2 mA ±10%, and 7.5 mA ±10% at the corresponding loads. The subject device's output current is within the predicate specification at 500 Ω but slightly larger at 2 kΩ and 10 kΩ. The waveform, pulse duration, frequency, maximum phase charge, and maximum current and maximum power densities were evaluated as part of the waveform output evaluation to demonstrate that each subject device model met its pre-specified output parameters. Therefore, this minor difference in maximum output current does not raise increased concerns of safety and effectiveness compared to the  |
| --- | --- | --- | --- |
|   |  SDZ-IIC: Continuous: 49.6mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ. Interrupted: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.4mA @10kΩ. Dense-Disperse: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ.  |   |   |
|   |  SDZ-III: Continuous: 48.8mA @500 Ω ;34.0mA @2kΩ; 10.4mA @10kΩ. Interrupted: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ. Dense-Disperse: 48.8mA @500 Ω ;34.80mA @2kΩ;  |   |   |
|   |  |   |   |

Page 10 of 20

{15}

510(k) Summary

|   | 10.2mA @10kΩ. |  | predicate device.  |
| --- | --- | --- | --- |
|   |  **SDZ-V:** Continuous: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ. Interrupted: 48.8mA @500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ. Dense-Disperse : 48.8mA@500 Ω ;35.20mA @2kΩ; 10.2mA @10kΩ.  |   |   |
|  Maximum Phase Charge | **SDZ-IIB:** Continuous: 9.42μC @500 Ω Interrupted: 9.41 μC @500 Ω Dense-Disperse: 9.42 μC @500 Ω | 9.4μC@500Ω | Different The maximum phase charge of the subject device ranges from 9.41 μC to 9.56 μC per phase at a 500 Ω load across all models and operating modes, compared with 9.4μC per phase for the predicate device. This small difference results from minor variations in the measured output current and pulse width and does not change the biphasic stimulation waveform or the charge-balancing characteristics of the output. The subject device uses a  |
|   |  **SDZ-IIC:** Continuous: 9.56μC @500 Ω Interrupted: 9.42μC @500 Ω Dense-Disperse: 9.42μC @500 Ω  |   |   |
|   |  **SDZ-III:** Continuous: 9.418μC @500 Ω  |   |   |

Page 11 of 20

{16}

510(k) Summary

|   | Interrupted: 9.418μC @500 Ω Dense-Disperse: 9.42μC @500 Ω |  | charge-balanced biphasic waveform consistent with its design specification. This difference does not raise different questions of safety or effectiveness.  |
| --- | --- | --- | --- |
|   |  **SDZ-V:** Continuous: 9.42μC @500 Ω Interrupted: 9.42 μC @500 Ω Dense-Disperse: 9.41μC @500 Ω  |   |   |
|  Maximum Average Current | **SDZ-IIB:** Continuous: 1.033 mA @500 Ω Interrupted: 1.071 mA @500 Ω Dense-Disperse: 1.068 mA @500 Ω | 1.08mA@500Ω | Different The maximum average current of the subject device ranges from 0.757 to 1.107 mA at 500 Ω, compared with 1.08 mA for the predicate device. The variation results from differences in waveform timing among models and modes. The subject device maintains comparable maximum output current and phase charge, and output waveform bench testing verified that the subject device delivers output with the predefined output specifications. Therefore, the difference does not  |
|   |  **SDZ-IIC:** Continuous: 1.067mA @500 Ω Interrupted: 1.054mA @500 Ω Dense-Disperse: 1.0505mA @500 Ω  |   |   |
|   |  **SDZ-III:** Continuous: 0.764mA @500 Ω Interrupted: 0.769mA @500 Ω  |   |   |

Page 12 of 20

{17}

510(k) Summary

|   | Dense-Disperse: 0.757mA @500 Ω |  | raise different questions of safety or effectiveness compared to the predicate device.  |
| --- | --- | --- | --- |
|   |  **SDZ-V:** Continuous: 1.082 mA @500 Ω Interrupted: 1.050mA @500 Ω Dense-Disperse: 1.107mA @500 Ω  |   |   |
|  Maximum Current Density, (mA/cm²) | **SDZ-IIB:** Continuous: 10.598 mA/cm² @500 Ω Interrupted: 10.988 mA/cm² @500 Ω Dense-Disperse: 10.957 mA/cm² @500 Ω | 9.4mA/cm²@500Ω | Different The maximum average current density of the subject device ranges from 7.766 to 11.353 mA/cm² at 500 Ω, compared with 9.4 mA/cm² for the predicate device, CMNS6-1. The subject and predicate devices employ asymmetric biphasic waveforms. FDA has cleared other electro-acupuncture devices, such as K122812, with current density of 17.6 mA/cm², which is larger than the maximum current densities reported for the subject device. Therefore, these differences will not  |
|   |  **SDZ-IIC:** Continuous: 10.947mA/cm² @500 Ω Interrupted: 10.855 mA/cm² @500 Ω Dense-Disperse: 10.778 mA/cm² @500 Ω  |   |   |
|   |  **SDZ-III:** Continuous: 7.891mA/cm² @500 Ω Interrupted: 7.945 mA/cm² @500 Ω  |   |   |

Page 13 of 20

{18}

510(k) Summary

|   | Dense-Disperse: 7.766mA/cm² @500 Ω |  | raise any increased safety and effectiveness concerns compared to the predicate device.  |
| --- | --- | --- | --- |
|   |  **SDZ-V:** Continuous: 11.101mA/cm² @500 Ω Interrupted: 10.947mA/cm² @500 Ω Dense-Disperse :11.353mA/cm² @500 Ω  |   |   |
|  Maximum Average Power Density | **SDZ-IIB:** Continuous: 0.229 W/cm² @500 Ω Interrupted: 0.237 W/cm² @500 Ω Dense-Disperse: 0.229 W/cm² @500 Ω | 0.141W/cm²@500Ω | Different The maximum average power density of the subject devices ranges from 0.229 to 0.239 W/cm² at a 500 Ω load, compared with 0.141 W/cm² for the predicate device. Although higher than the predicate value, all subject device values remain below the established safety threshold. So the difference in maximum average power density does not raise increased thermal safety concerns compared to the predicate device.  |
|   |  **SDZ-IIC:** Continuous: 0.239 W/cm² @500 Ω Interrupted: 0.237 W/cm² @500 Ω Dense-Disperse: 0.237 W/cm² @500 Ω  |   |   |
|   |  **SDZ-III:** Continuous: 0.237 W/cm² @500 Ω Interrupted: 0.237 W/cm² @500 Ω  |   |   |

Page 14 of 20

{19}

510(k) Summary

|   | Dense-Disperse: 0.237 W/cm² @500 Ω |  |   |   |
| --- | --- | --- | --- | --- |
|   |   |   |   |  **SDZ-V:** Continuous: 0.237 W/cm² @500 Ω Interrupted: 0.237 W/cm² @500 Ω Dense-Disperse: 0.237 W/cm² @500 Ω  |
|  Pulse width (μsec) | Positive | **SDZ-IIB:** Continuous: 193.00 μs @500 Ω Interrupted: 192.80 μs @500 Ω Dense-Disperse: 193.80 μs @500 Ω | 200μs±10% | Similar The positive phase duration of the subject device ranges from 192.8 to 193.8 μs at 500 Ω, which is within the predicate device specification of 200 μs ±10%. Therefore, the minor variation does not affect the stimulation waveform or raise different questions of safety or effectiveness.  |
|   |   |  **SDZ-IIC:** Continuous: 193.60us @500 Ω Interrupted: 193.75us @500 Ω Dense-Disperse: 193.65us @500 Ω  |   |   |
|   |   |  **SDZ-III:** Continuous: 193us @500 Ω Interrupted: 193us @500 Ω  |   |   |

Page 15 of 20

{20}

510(k) Summary

|   |  | Dense-Disperse: 192.9us @500 Ω |  |   |
| --- | --- | --- | --- | --- |
|   |   |  SDZ-V: Continuous: 193.70us @500 Ω Interrupted: 193.00us @500 Ω Dense-Disperse: 193.70us @500 Ω  |   |   |
|   |  Negative | SDZ-IIB: Continuous: 1034 μs @500 Ω Interrupted: 1034 μs @500 Ω Dense-Disperse: 1024 μs @500 Ω | 1030μs(5.15x(+Phase)) | Similar. The negative phase duration of the subject device ranges from 1024 to 1054 μs at 500 Ω, compared with 1030 μs for the predicate device. The minor variation maintains the intended asymmetric biphasic waveform and does not adversely affect charge balance. Therefore, it does not raise different questions of safety or effectiveness.  |
|   |   |  SDZ-IIC: Continuous: 1034us @500 Ω Interrupted: 1036us @500 Ω Dense-Disperse: 1030us @500 Ω  |   |   |
|   |   |  SDZ-III: Continuous: 1038us @500 Ω Interrupted: 1036us @500 Ω Dense-Disperse: 1036us @500 Ω  |   |   |

Page 16 of 20

{21}

510(k) Summary

|   |  | SDZ-V:Continuous: 1040us @500 ΩInterrupted: 1040us @500 ΩDense-Disperse: 1054us @500 Ω |  |   |
| --- | --- | --- | --- | --- |
|  Pulse Period(msec) |   | 10~1000ms | 10~1000ms | Identical  |
|  Max. pulse frequency(Hz) [or Rate(pps)] |   | 1~100Hz±10% | 1~100Hz±10% | Identical  |
|  Net Charge (μC per pulse) |   | 0μC@500Ω, + and- pulses cancel | 0μC@500Ω, + and- pulses cancel | Identical  |
|  Burst Mode |   | N/A | N/A | Identical  |
|  Accessories |   | Lead wires, Alligator type connectors | Lead wires, Alligator type connectors | Identical  |
|  OTC or Rx |   | Rx | Rx | Identical  |
|  Compliance with 21 CFR Part 898 |   | Yes | Yes | Identical  |
|  Weight (g) |   | SDZ-IIB: Approx. 190 (Without battery)SDZ-IIC: Approx. 650 (Without battery)SDZ-III: Approx. 700 (Without battery)SDZ-V: Approx. 500 (Without battery) | approx.657 | DifferentDifferences in weight do not alter the intended use or fundamental stimulation principle. Therefore, these physical differences do not adversely affect safety or  |

Page 17 of 20

{22}

510(k) Summary

|   |  |  | effectiveness.  |
| --- | --- | --- | --- |
|  Dimensions (mm) | SDZ-IIB: 157 × 79 × 33; SDZ-IIC: 275 × 160 × 54; SDZ-III: 286 × 180 × 64; SDZ-V: 258 × 169 × 45 | 230*155*55 | Different The differences in external dimensions reflect differences in enclosure design among the subject device models. These dimensional differences do not alter the intended use or fundamental stimulation principle. Therefore, the differences in external dimensions do not raise different questions of safety or effectiveness.  |

Page 18 of 20

{23}

510(k) Summary

# **Substantial Equivalence Discussion:**

The subject device (Needle Stimulator, models SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V) and the K202861 predicate device have the same intended use. Both the subject and predicate devices employ charge-balanced asymmetric biphasic waveforms and operate within comparable pulse-duration and frequency ranges. Minor differences in channel configuration, maximum output voltage and maximum output current, maximum phase charge, maximum average current, maximum average current density, maximum average power density, and phase durations do not raise increased concerns of safety and/or effectiveness compared to the predicate device.

# **7.0 Summary of Non-Clinical Testing**

Non-clinical tests were conducted to verify that the subject devices met all design specifications to be considered substantially equivalent to the predicate device:

Electrical Safety and Electromagnetic Compatibility:

- IEC 60601-1: 2020, Medical electrical equipment – Part 1: General requirements for basic safety, and essential performance
- IEC 60601-1-2: 2020, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic Compatibility
- IEC TS 60601-4-2:2024, Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems

Software Verification

- Software was evaluated at basic documentation level in accordance with 2023 FDA Guidance: Content of Premarket Submissions for Device Software Functions.

The following non-clinical tests were conducted:

- Oscilloscope tracings of the output waveforms at loads of 500 Ω, 2,000 Ω, and 10,000 Ω for each output mode (continuous, interrupted, and dense-disperse) of each subject device model (SDZ-IIB, SDZ-IIC, SDZ-III, and SDZ-V).
- Values and calculation methodologies for the output parameters of Maximum Output Voltage, Maximum Output Current, Maximum Phase Charge (μC), Maximum Average Current (mA), Maximum Current Density (mA/cm²); Maximum Average Power Density (W/cm²) and Maximum Charge Density (μC/cm²).

The test results demonstrated that the subject devices (SDZ-IIB, SDZ-IIC, SDZ-III, SDZ-V) met the predefined output specifications. The results demonstrate that the SDZ-IIB, SDZ-IIC, SDZ-III, and SDZ-V subject device models do not raise increased safety and/or effectiveness concerns compared

Page 19 of 20

{24}

510(k) Summary

with the predicate device.

## **8.0 Conclusion**

The comparison and analysis above demonstrate that the proposed device is as safe and effective as the K202861 predicate device. Thus, the subject device is substantially equivalent to the predicate device.

Page 20 of 20

---

**Source:** [https://fda.innolitics.com/device/K260230](https://fda.innolitics.com/device/K260230)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
