K260194 · Sheng Sheng Yi (Beijing) Technology Company Limited · MQG · Apr 28, 2026 · Obstetrics/Gynecology
Device Facts
Record ID
K260194
Device Name
Vacuum Pump (Model: AD-VPUM-500)
Applicant
Sheng Sheng Yi (Beijing) Technology Company Limited
Product Code
MQG · Obstetrics/Gynecology
Decision Date
Apr 28, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6120
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vacuum Pump (Model No.: AD-VPUM-500) is intended for the aspiration of oocytes, during assisted reproduction procedures using low flow, intermittent vacuum.
Device Story
Electrically powered vacuum pump; used for oocyte collection from ovarian follicles during IVF procedures. System generates negative pressure via diaphragm pump; pressure regulated by PID controller using integrated pressure sensor; air tank buffers fluctuations. Operated via foot pedal to control suction access; used with disposable vacuum line, hydrophobic filter, collection tube, and retrieval needle under transvaginal ultrasound guidance. Provides vacuum levels -10 to -500 mmHg. Benefits include controlled, stable aspiration of follicular fluid and oocytes, facilitating retrieval for assisted reproduction.
Clinical Evidence
Bench testing only. Performance verified against ISO 10079-1:2022 and ISO 10079-4:2021. Testing included visual inspection, vacuum pressure parameters (range/accuracy), and boost mode functionality after worst-case conditioning. Electrical safety and EMC testing performed per ANSI AAMI ES60601-1 and IEC 60601-1-2. Software validation conducted per 2023 FDA guidance.
Technological Characteristics
Electrically powered diaphragm pump; PID-controlled negative pressure system. Vacuum range: -10 to -500 mmHg; flow rate: 20-25 mL/min. Includes main control unit, power cable, foot pedal. Ingress protection: IP42. Standards: ISO 10079-1:2022, ISO 10079-4:2021, ANSI AAMI ES60601-1, IEC 60601-1-2.
Indications for Use
Indicated for the aspiration of oocytes during assisted reproduction procedures using low flow, intermittent vacuum.
Regulatory Classification
Identification
Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.
Special Controls
*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
April 28, 2026
Sheng Sheng Yi (Beijing) Technology Company Limited
Wei Zhao
Regulatory Engineer
Rm.235-238, Jianjin Ctr, # Jia 1, Yongtaiyuan, Qinghe
Haidian Dist, Rm.201, Bldg. 4, # 25 Yanqi 4th S St., Huairou
Beijing, 100085
China
Re: K260194
Trade/Device Name: Vacuum Pump (Model: AD-VPUM-500)
Regulation Number: 21 CFR§ 884.6120
Regulation Name: Assisted Reproduction Accessories
Regulatory Class: II
Product Code: MQG
Dated: January 22, 2026
Received: January 22, 2026
Dear Wei Zhao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260194 - Wei Zhao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260194 - Wei Zhao
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Monica D. Garcia -S
Monica D. Garcia, Ph.D.
Assistant Director
DHT3B: Division of Reproductive, Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260194 | |
| Device Name Vacuum Pump (Model: AD-VPUM-500) | |
| Indications for Use (Describe) The Vacuum Pump (Model: AD-VPUM-500) is intended for the aspiration of oocytes, during assisted reproduction procedures using low flow, intermittent vacuum. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K260194
Page 1 of 7
# 510(k) Summary
# K260194
## I. Submitter Information
Name: Sheng Sheng Yi (Beijing) Technology Company Limited
Address: Room 201, Building 4, No. 25 Yanqi 4th South Street, Huairou District, 101400 Beijing, China
Contact Person: Wei Zhao
Phone: +86 10 62927009
E-mail: zhaow@fertsy.com
## II. Date of Preparation: April 24, 2026
## III. Subject Device Information
Device Trade Name: Vacuum Pump (Model No.: AD-VPUM-500)
Common Name: Vacuum Pump
Regulation Number: 21 CFR 884.6120
Regulation Name: Assisted Reproduction Accessories
Regulation Class: II
Product code: MQG (Accessory, Assisted Reproduction)
## IV. Predicative Device
Submission Number: K160753
Device Trade Name: COOK Vacuum Pump
The predicate device has not been subject to a design-related recall.
## V. Device Description
The Vacuum Pump (Model: AD-VPUM-500) is an electrically powered vacuum pump that is used for the collection of oocytes from ovarian follicles for use during in vitro fertilization (IVF) procedures. The Vacuum Pump (Model: AD-VPUM-500) provides vacuum levels ranging from -10 mmHg to -500 mmHg. It is supplied non-sterile.
This device is used in conjunction with a disposable vacuum line with hydrophobic filter, collection tube, and oocyte retrieval needle to perform puncture under transvaginal ultrasound guidance. Through intermittent high vacuum and low flow provided by the Vacuum Pump (Model: AD-VPUM-500), follicular fluid and oocytes from the follicles are aspirated.
The product consists of the main control unit, the power connection cable, and the foot pedal. All other devices, including a disposable vacuum line with hydrophobic filter, collection tube, and oocyte retrieval needle, are not included in the Vacuum Pump
{5}
K260194
Page 2 of 7
(Model: AD-VPUM-500).
## VI. Indications for Use
The Vacuum Pump (Model No.: AD-VPUM-500) is intended for the aspiration of oocytes, during assisted reproduction procedures using low flow, intermittent vacuum.
## VII. Comparison of Intended Use and Technological Characteristics of the Subject and Predicate Devices
The table below compares the intended use and technological characteristics of the subject and predicate device.
| Attribute | Subject device (K260194) | Predicative device (K160753) | Comparison |
| --- | --- | --- | --- |
| Manufacturer | Sheng Sheng Yi (Beijing) Technology Company Limited | William A. Cook Australia Pty Ltd | / |
| Trade name | Vacuum Pump (Model No.: AD-VPUM-500) | COOK Vacuum Pump | / |
| Regulation name | Assisted Reproduction Accessories | Assisted Reproduction Accessories | Same |
| Regulatory Class | II | II | Same |
| Product code | MQG | MQG | Same |
| Clinical characteristics | | | |
| Indications for Use | The Vacuum Pump (Model No.: AD-VPUM-500) is intended for the aspiration of oocytes, during assisted reproduction procedures using low flow, intermittent vacuum. | The COOK Vacuum Pump is intended for the aspiration of eggs (ova), during assisted reproduction procedures using low flow, intermittent vacuum. | Same |
| General technological characteristics | | | |
| Device design | The Vacuum Pump (Model No.: AD-VPUM-500) is an electrically-powered vacuum pump that is used for the collection of oocytes from ovarian follicles for use during in vitro fertilization (IVF) procedures. The Vacuum Pump (Model | The Vacuum Pump is designed to maintain a vacuum accurately at a user specified setting with a range of -10 mmHg to -500 mmHg when configured to display mmHg and a range of -1.0 kPa to -67.0 kPa when configured to display kPa. The device van also boost the | Different |
{6}
K260194
Page 3 of 7
| Attribute | Subject device (K260194) | Predicative device (K160753) | Comparison |
| --- | --- | --- | --- |
| | No.: AD-VPUM-500) provides vacuum levels ranging from -10 mmHg to -500 mmHg (-1.0 kPa to -67.0 kPa). It is supplied non-sterile. This device is used, in conjunction with a disposable vacuum line with hydrophobic filter, collection tube, and oocyte retrieval needle to perform puncture under transvaginal ultrasound guidance. Through intermittent high vacuum and low flow provided by the Vacuum Pump (Model No.: AD-VPUM-500), follicular fluid and oocytes from the follicles are aspirated. The product consists of the main control unit, the power connection cable, and the foot pedal. All other devices, including a disposable vacuum line with hydrophobic filter, collection tube, and oocyte retrieval needle, are not included in the Vacuum Pump (Model No.: AD-VPUM-500). | vacuum to -500 mmHg (or -67.0 kPa in kPa display mode) from any setting. The disposable Vacuum Line and Filter (K-DVLF-240) consists of a one-way hydrophobic filter and 240 cm long low volume vacuum line. The Disposable Vacuum Line and Filter is used to connect ovum aspiration needles to the Cook Vacuum Pump to prevent contamination of the unit. It is supplied sterile in peel-open packages and is intended for single-use. | |
| Working principle | The Vacuum Pump (Model No.: AD-VPUM-500) is an integrated system based on the generation, transmission, and control of negative pressure. Its core power | A single diaphragm pump is attached to a motor. The suctioned air is collected in the pressure reservoir. A pressure gauge is connected to the pressure reservoir and the pressure is displayed. The | Same |
{7}
K260194
Page 4 of 7
| Attribute | Subject device (K260194) | Predicative device (K160753) | Comparison |
| --- | --- | --- | --- |
| | comes from a diaphragm pump, which actively expels air through the reciprocating motion of the diaphragm, creating a negative pressure environment inside the pump and providing basic power for suction. To maintain stable negative pressure, a pressure sensor is integrated into the pipeline. The system reads the pipeline pressure value, sends it together with the set pressure value to the PID (Proportional-Integral-Derivative) controller for calculation, and the result provides data support for the control of the diaphragm pump. Additionally, the system is equipped with an air tank to buffer pressure fluctuations and reduce sudden pressure changes during the suction process. A foot pedal-controlled stop valve is used to open or close the access of the oocyte retrieval needle, ensuring controllability of the suction process. For safety and sample purity, the system is equipped with multi-stage filters at key nodes to prevent microorganisms and | aspiration pressure sensor detects the set aspiration pressure and the aspiration pressure during the boost operation. A motor controlled by an aspiration pressure control circuit adjusts the detected pressure and maintains the set pressure. | |
{8}
K260194
Page 5 of 7
| Attribute | Subject device (K260194) | Predicative device (K160753) | Comparison |
| --- | --- | --- | --- |
| | particles from contaminating the sample or damaging the equipment. | | |
| Power supply | Power Supply and Frequency: 100 - 240 VAC, 50 - 60 Hz
Max Current: 500 mA (115 VAC); 250 mA (240 VAC)
Maximum power consumption: 60 VA | Power Supply and Frequency: 100 - 240 VAC, 50 - 60 Hz
Max Current: 500 mA (115 VAC); 250 mA (240 VAC)
Maximum power consumption: 60 VA | Same |
| Operating & storage environment | Environmental operating conditions: +5°C ~ +35°C
30% ~ 75% RH no condensing
79 kPa ~ 106 kPa
Storage and transportation directions: +5°C ~ +40°C
5% ~ 75% RH non-condensing
80 kPa ~ 106 kPa | Environmental operating conditions: +5°C ~ +35°C
10% ~ 75% RH no condensing
700 hPa ~ 1060 hPa
Storage and transportation directions: +5°C ~ +40°C
10% ~ 75% RH non-condensing | Similar |
| Ingress Protection degree | IP42 | IP41 | Different |
| Recommended flow rate | 20-25 mL/min | 20-25 mL/min | Same |
| Vacuum range accuracy | ± 5 mmHg (or ±0.7 kPa) | ± 5 mmHg (or ±0.7 kPa) | Same |
| Maximum Pressure in Boost Mode | -500 mmHg or -67 kPa | The vacuum can be boosted to -500 mmHg from any setting during operation | Same |
| Foot pedal | Yes | Yes | Same |
| Controller | Microprocessor electronic control | Microprocessor electronic control | Same |
| Software | Yes | Yes | Same |
The subject and predicate devices have the same indications for use and intended use. They also use the same fundamental technology to create a vacuum and have the same
{9}
K260194
Page 6 of 7
recommended flow rate for oocyte aspiration procedures. The subject device and predicate device differ in ingress protection degree and operating/storage environment. These differences do not raise different questions of safety and effectiveness.
## VIII. Summary of Non-clinical Performance Testing
The following performance tests were conducted on the Vacuum Pump (Model No.: AD-VPUM-500) to verify that the subject device met all design specifications, demonstrated performance, and to support substantial equivalence to the predicate device:
### Electric Safety and Electromagnetic Compatibility
- ANSI AAMI ES60601-1: 2005 + A1:2012 + A2: 2021, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2: 2014 + A1: 2020, Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance -Collateral Standard: Electromagnetic disturbances -Requirements and tests
- IEC TS 60601-4-2:2024, Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- 2022 FDA Guidance Document, Electromagnetic Compatibility (EMC) of Medical Devices
### Software validation
Software Verification and Validation was conducted and documented at a Basic Documentation Level according to the 2023 FDA Guidance Document, Content of Premarket Submissions for Device Software Functions
### Bench Performance
The following performance testing was conducted on the subject device after worst-case conditioning:
- Visual inspection
- Vacuum pressure parameters (e.g., range, accuracy, etc.).
- Boost mode
The subject device met the specifications for all the testing conducted. The testing was conducted in accordance with the following standards:
- ISO 10079-1: 2022, Medical suction equipment Part 1: Electrically powered suction equipment
- ISO 10079-4: 2021, Medical suction equipment Part 4: General requirements
## IX. Conclusion
The results of the performance testing described above demonstrate that the Vacuum
{10}
K260194
Page 7 of 7
Pump (Model No.: AD-VPUM-500) is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.