The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
Device Story
Pelvic Muscle Trainer (KM510, KM516B) is a hand-held, home-use electrical stimulation device for pelvic floor muscle rehabilitation. User inserts a reusable, dual-electrode vaginal probe connected to a control unit. Device delivers electrical stimulation via stainless steel electrodes to induce pelvic floor muscle contractions; intended to strengthen muscles and improve contraction duration. User controls stimulation intensity via manual push-button interface. KM510 offers 9 preset programs; KM516B offers 6. Device operates on rechargeable lithium (KM510) or AAA batteries (KM516B). Output parameters (frequency, pulse width, current) are pre-programmed to treat specific incontinence types. Clinical benefit includes improved muscle function and urinary continence management.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), nerve/muscle stimulator performance (IEC 60601-2-10), and biocompatibility (ISO 10993-1/5/10).
Technological Characteristics
Hand-held electrical stimulator; stainless steel dual-electrode vaginal probe. KM510: 3.7V rechargeable Li-ion; KM516B: 2xAAA batteries. Output: 10-50Hz, 200-300μs pulse width, max 80mA (KM510) or 60mA (KM516B) at 500Ω. Connectivity: wired probe-to-controller. Software: moderate level of concern. Sterilization: reusable single-patient use.
Indications for Use
Indicated for adult women (22+ years) with weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence and to maintain urinary continence.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
Shenzhen Konmed Technology Co., Ltd.
% Cassie Lee
Manager
Guangzhou GLOMED Biological Technology Co., Ltd.
2231, Bldg. 1, Rui Feng Center
Kaichuang Rd., Huangpu District
Guangzhou, 510000
CHINA
Re: K260178
Trade/Device Name: Pelvic Muscle Trainer (KM510, KM516B)
Regulation Number: 21 CFR 876.5320
Regulation Name: Nonimplanted Electrical Continence Device
Regulatory Class: II
Product Code: KPI
Dated: May 19, 2026
Received: May 19, 2026
Dear Cassie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: The Center for Devices and Radiological Health (CDRH) does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260178 - Cassie Lee
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled 'Deciding When to Submit a 510(k) for a Change to an Existing Device' (https://www.fda.gov/media/99812/download) and 'Deciding When to Submit a 510(k) for a Software Change to an Existing Device' (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ('UDI Rule'). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260178 - Cassie Lee
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
JESSICA K. NGUYEN
-S
Jessica K. Nguyen, Ph.D.
Assistant Director
DHT3B: Division of Reproductive,
Gynecology, and Urology Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260178
?
Please provide the device trade name(s).
?
Pelvic Muscle Trainer (KM510, KM516B)
Please provide your Indications for Use below.
?
The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
Please select the types of uses (select one or both, as applicable).
☐ Prescription Use (21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
?
Please select the age group(s) for which the device(s) is to be used.
☐ Neonates/Newborns (Birth to < 29 days old)
☐ Infants (29 days old to < 2 years old)
☐ Children (2 years old to < 12 years old)
☐ Adolescents (12 years old to < 22 years old)
☑ Adults (22 years old and greater)
?
{4}
K260178
Page 1 of 11
# 510(k) Summary
*For*
Pelvic Muscle Trainer Model: KM510, KM516B**
Date of the summary prepared: June 16, 2026
{5}
K260178
Page 2 of 11
## 510(k) Summary
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
### 1. Submitter's Information
Company Name: Shenzhen Konmed Technology Co.,Ltd.
Establishment Registration Number: 3011905669
Address: 601, Building B4, Shenchengtou Creative Factory Life Science Park, Julongshan A Road, Xiuxin Block, Kengzi Street, Pingshan District, Shenzhen, Guangdong, CHINA
Postal Code:518118
Tel: +86-755-86704558
Phone: 13603046428
Fax: +86-755-86704558
Contact Person (including title): Tony (Project Manager)
mail: 78003390@qq.com
### Application Correspondent:
Contact Person: Ms. Cassie Lee
Guangzhou GLOMED Biological Technology Co., Ltd.
Address: 2231, Building 1, Rui Feng Center, Kaichuang Road, Huangpu District, Guangzhou, Guangdong, China
Tel: +86 20 8266 2446
Email: regulatory@glomed-info.com
### 2. Subject Device Information
Trade Name: Pelvic Muscle Trainer
Model: KM510, KM516B
Regulatory Class: II
Product Code: KPI
Regulation Number: 21 CFR Part 876.5320
### 3. Predicate Device Information
#### Primary Predicate Device:
510(K) Number: K191312
Company Name: Tenscare Ltd
Trade Name: Perfect PFE
Regulatory Class: Class II
Product Code: KPI
Regulation Number: 21 CFR Part 876.5320
The predicate device has not been subject to a design-related recall.
#### Reference Device 1:
510(K) Number: K141643
Company Name: Internation Trade Group, Inc.
Trade Name: Yarlap
Regulatory Class: Class II
Product Code: KPI
Regulation Number: 21 CFR Part 876.5320
Page 2 / 11
{6}
K260178
Page 3 of 11
# **Reference device 2:**
510(K) Number: K220161
Company Name: Shenzhen Konmed Technology Co., Ltd.
Trade Name: Biofeedback Nerve and Muscle Stimulator
Regulatory Class: Class II
Product Code: KPI
Regulation Number: 21 CFR Part 876.5320
# **4. Device Description**
The Pelvic Muscle Trainer is a hand-held, home-use device designed to treat stress, urge and mixed urinary continence in women and maintain urinary continence in women. The device is supplied with a reusable (single-patient use) vaginal, dual-electrode, stimulation probe. The probe connects to the control unit by cable and plug and is inserted into the vagina by the end user.
Electrical stimulation is delivered via stainless steel electrodes on the probe to induce a contraction of the pelvic floor muscles. Muscle stimulation is used to train and strengthen the pelvic floor muscles in a controlled manner. Muscle stimulation is used to improve the ability of muscles to hold a contraction for an extended period of time and is a treatment for urinary incontinence. During a session, stimulation is delivered to specific muscles to encourage their contraction. The level of electrical stimulation is controlled by the end user using manual, push-button controls.
The device has 2 models, KM510 and KM516B. KM510 has 9 preset treatment programs, KM516B has 6 preset treatment programs.
The device consists of a control unit, a probe, one USB wire and one user manual.
# **5. Intended Use / Indications for Use**
The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women.
# **6. Test Summary**
# **Summary of Non-clinical Performance Testing**
# 1) Performance testing summary
The safety and performance of the Pelvic Muscle Trainer (model: KM510, KM516B) has been evaluated with the following lab bench testing:
| Title of the test | Device description/sample size | Test methods/applicable standards | Acceptance criteria | Unexpected results/significant deviations | Test results |
| --- | --- | --- | --- | --- | --- |
| General requirements for basic safety and essential performance | The test sample is the final finished product: KM510, KM516B | IEC 60601-1: 2005/AMD1: 2012/AMD2: 2020 | The test is carried out under the test method specified in the standard, and the test result is within the acceptance range of the standard. | N/A | Pass |
| Electromagnetic disturbances | The test sample is the final finished product: KM510, KM516B | IEC 60601-1-2: 2015/AMD1: 2020 | No degradation of performance was found during test or lower than limits or measurement. | N/A | Pass |
{7}
K260178
Page 4 of 11
| Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. | The test sample is the final finished product: KM510, KM516B | IEC 60601-1-11: 2015/AMD1: 2020 | The device operates normally, and can provide basic safety and essential performance. | N/A | Pass |
| --- | --- | --- | --- | --- | --- |
| Particular requirements for the basic safety and essential performance of nerve and muscle stimulators. | The test sample is the final finished product: KM510, KM516B | IEC 60601-2-10:2012 AMD 1:2016 | The test is carried out under the test method specified in the standard, and the test results is within the test acceptance range of the standard. | N/A | Pass |
| Medicalelectrical equipment -Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems | The test sample is the final finished product: KM510, KM516B | IEC TR 60601-4-2: Edition 1.0 2016- 05 | No degradation of performance was found during test or lower than limits or measurement. | N/A | Pass |
| IEC 62133-2 Edition1.0 2017-02 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems | The test sample is the final finished product: KM510 | IEC 62133-2 Edition1.0 2017-02 | The test is carried out under the test method specified in the standard, and the test result is within the acceptance range of the standard. | N/A | Pass |
| Battery life test | The test sample is the final finished product: KM510 | N/A | The battery capacity should retain >80% initial operating time after 3 years | N/A | Pass |
| Output waveform testing | The test sample is the final finished product: KM510, KM516B | “Guidance Document for Powered Muscle Stimulator 510(k)s, June 9, 1999”, “Class II Special Controls Guidance Document: Transcutaneous Electrical Nerve Stimulator for Pain Relief” “IEC 60601-2-10, Particular requirements for the safety of | The output current limit r.m.s. with 500Ω resistance load is 80mA For pulse durations of less than 0.1s, the pulse energy with 500Ω resistance load shall not exceed 300mJ per pulse. The output voltage shall not exceed a peak value of 500V, when measured under open-circuit condition. Otherwise, we also set the following two Acceptance Criteria for | N/A | Pass |
{8}
K260178
Page 5 of 11
| | | nerve and muscle stimulators". | our device according to FDA Guidance. Maximum Power Density < 0.25 (W/cm²) | | |
| --- | --- | --- | --- | --- | --- |
# 1) Biocompatibility testing
The vaginal probe (KM-503) is identical to the vaginal probe (KM-503) cleared in K163288. Accordingly, biocompatibility testing per requirements of the ISO 10993-1 and the FDA Guidance "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." from K163288 was leveraged to support biocompatibility of the subject device. This includes the testing to the following standards:
- Cytotoxicity testing per ISO 10993-5:2009/(R)2014 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- Sensitization testing per ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- Irritation or intracutaneous reactivity testing per ISO 10993-10:2010/(R)2014 Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
# 2) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by "Content of Premarket Submissions for Device Software Functions. Guidance for Industry and Food and Drug Administration Staff" The software for this device was considered as a "moderate" level concern, since a malfunction of, or a latent design flaw in, the Software Device leads to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to Minor Injury.
# 3) Testing on patient leads or patient cables
The lead wire of the Konmed Incontinence Stimulation Electrode is the same lead wire use in the K220161, which complies with regulation 21 CFR 898 and IEC 60601-1: 2006:+A1: 2012+A2: 2020.
# 7. Comparison to predicate device and conclusion
Compared with the predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise different questions of safety or effectiveness.
{9}
K260178
Page 6 of 11
| Elements of Comparison | Subject Device | Primary Predicate Device (K191312) | Reference Device 1 (K141643) | Reference device 2 (K220161) | Remark |
| --- | --- | --- | --- | --- | --- |
| Device Name | Pelvic Muscle Trainer | Perfect PFE | Yarlap | Pelvifine Pelvic Muscle Trainer | -- |
| 510(k) Number | K260178 | K191312 | K141643 | K220161 | -- |
| Product Code | KPI | KPI | KPI | KPI | Same |
| Regulation No | 21 CFR Part 876.5320 | 21 CFR Part 876.5320 | 21 CFR Part 876.5320 | 21 CFR Part 876.5320 | Same |
| Indications for use | The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women. | The Perfect PFE is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge, and mixed urinary incontinence in women and to maintain urinary continence in women. The Perfect PFE is intended for OTC use. | The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women. | The device is intended to provide electrical stimulation and neuromuscular re-education for the purpose of rehabilitation of weak pelvic floor muscles for the treatment of stress, urge and mixed urinary incontinence in women and to maintain urinary continence in women. | Same |
Page 6 / 11
{10}
K260178
Page 7 of 11
| Prescriptive or OTC | OTC | OTC | Rx and OTC | OTC | Same |
| --- | --- | --- | --- | --- | --- |
| Number of output modes | KM510: 9 KM516B: 6 | 4 | 6 | Not publicly available | Similar |
| Number of output channel | KM510: 1 KM516B: 1 | 1 | 2 | Not publicly available | Same |
| Electrode Surface Area | KM510: 7.85 cm² KM516B: 7.85 cm² | 4.24 cm² | 6.4 cm² | 15.7 cm² | Similar |
| Frequency (Hz) | KM510: 10 - 50, ±10% Stress: 12/35/50 Mixed: 10/12/20 Urge: 10/12 Tone: 35 KM516B: 10-35, ±10% Stress: 12/35 Mixed: 12/20 Urge: 10/12 | Stress: 50Hz Tone: 35Hz Mixed: 10Hz/50 Hz Urge: 10Hz | 10 Hz - 35Hz, program Dependent Stress: 12/35 Mixed: 12/20 Urge: 10/12 | 2-100Hz | Same |
Page 7 / 11
{11}
K260178
Page 8 of 11
| Pulse width (us) | KM510: 200 μS – 300 μS, ±10% Stress: 200/250/300 Mixed: 200/250/300 Urge: 200 Tone: 250 KM516B: 200 μS – 250 μS, ±10% Stress:200/250 Mixed: 200/250 Urge: 200 | Stress: 300μs Tone: 250μs Mixed: 200μs / 300μs Urge: 200μs | 200 μS – 250 μS, program dependent Stress: 200/250 Mixed: 200/250 Urge:200/250 | 50-450μs | Same |
| --- | --- | --- | --- | --- | --- |
| Maximum Output Voltage (V) | KM510: 40V@ 500Ω KM516B: 30V@ 500Ω | 45V @ 500Ω | Not publicly available | Not publicly available | Similar |
| Maximum Output Current (A) | KM510: 80mA @ 500Ω KM516B: 60mA @ 500Ω | 0.09A @ 500Ω | 80mA +0/- 8% | 94.4mA @ 500 Ω 54mA @ 2k Ω 15 mA @ 10k Ω | Similar |
| Maximum Phase Charge (μC) | KM510: 24 μC KM516B: 15 μC This corresponds to the longest pulse at the highest current. | 24.29 @ 500Ω | 80mA x 250 μS = 20 μC This corresponds to the longest pulse at the highest current. | 51.4μC@500 Ω | Similar |
| Maximum Current Density, (mA/cm²) | KM510: 11.21 mA/ cm² (area=7.85 cm²) KM516B: 8.41 mA/ cm² (area=7.85 cm²) | 21.2 @ 500Ω (Area=4.24 cm²) | Not publicly available | 6.01mA/ cm²@ 500 Ω Surface = 15.7 cm² | Similar |
Page 8 / 11
{12}
K260178
Page 9 of 11
| Maximum Average Power Density, (mW/cm²) | KM510: 14.80 mW/cm² @ 500Ω KM516B: 3.88 mW/cm² @ 500Ω | 11.6 mW/cm² @ 500Ω (Area=4.24 cm²) | 3.5 mW/sqcm At maximum frequency of 35Hz pulse width 200μS and current of 80mA. (P06) PC Electrode area: 6.4 cm² | 0.2814mW/cm²@ 500 Ω | Similar |
| --- | --- | --- | --- | --- | --- |
| Power source | KM510: DC 3.7V/900mAh rechargeable lithium battery KM516B: DC 3.0V 2×AAA batteries | 2 Alkaline AA 1.5V Batteries | Not publicly available | 7.4V DC/1200mAh rechargeable lithium battery | Similar |
| Electrode lead wires and patient cable compliance with 21 CFR 898 | Yes | Yes | Not publicly available | Not publicly available | Same |
The below tables compare the subject device treatment programs to the predicate and reference devices.
KM510:
Page 9 / 11
{13}
K260178
Page 10 of 11
| Utility | KM510 Max. 80mA | K141643 Max. 80mA | K191312 Max. 90mA | Rate (Hz) | Pulse Duration(μs) | work (s) | rest (s) | Time (min) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Mixed | P01 | | | 12 | 200 | 5 | 5 | 15 |
| | | 1 | | 12 | 200 | 5 | 5 | 15 |
| Mixed | P02 | | | 20 | 250 | 8 | 8 | 20 |
| | | 2 | | 20 | 250 | 8 | 8 | 20 |
| Mixed | P03.1 | | | 10 | 200 | 6 | 12 | 10 |
| | | | mixed | 10 | 200 | / | / | 20 |
| | P03.2 | | | 50 | 300 | 5 | 15 | 10 |
| | | | mixed | 50 | 300 | / | / | 20 |
| Urge | P04 | | | 12 | 200 | 5 | 10 | 15 |
| | | 3 | | 12 | 200 | 5 | 10 | 15 |
| Urge | P05 | | | 10 | 200 | 6 | 12 | 20 |
| | | 4 | | 10 | 200 | 6 | 12 | 20 |
| Stress | P06 | | | 12 | 250 | 5 | 15 | 15 |
| | | 5 | | 12 | 250 | 5 | 15 | 15 |
| Stress | P07 | | | 35 | 200 | 6 | 18 | 20 |
| | | 6 | | 35 | 200 | 6 | 18 | 20 |
| Stress | P08 | | | 50 | 300 | 5 | 15 | 15 |
| | | | stress | 50 | 300 | 6 | 10 | 20 |
| Tone | P09 | | | 35 | 250 | 5 | 5 | 20 |
| | | | tone | 35 | 250 | 5 | 6 | 20 |
# **KM516B:**
| Utility | KM516B Max. 60mA | K141643 Max. 80mA | Rate (Hz) | Pulse Duration(μs) | Work (s) | Rest (s) | Phase Time(min) |
| --- | --- | --- | --- | --- | --- | --- | --- |
Page 10 / 11
{14}
K260178
Page 11 of 11
| Mixed | P01 | | 12 | 200 | 5 | 5 | 15 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | 1 | 12 | 200 | 5 | 5 | 15 |
| | P02 | | 20 | 250 | 8 | 8 | 20 |
| | | 2 | 20 | 250 | 8 | 8 | 20 |
| Urge | P03 | | 12 | 200 | 5 | 10 | 15 |
| | | 3 | 12 | 200 | 5 | 10 | 15 |
| | P04 | | 10 | 200 | 6 | 12 | 20 |
| | | 4 | 10 | 200 | 6 | 12 | 20 |
| Stress | P05 | | 12 | 250 | 5 | 15 | 15 |
| | | 5 | 12 | 250 | 5 | 15 | 15 |
| | P06 | | 35 | 200 | 6 | 18 | 20 |
| | | 6 | 35 | 200 | 6 | 18 | 20 |
The main differences between the subject and predicate device are the stimulation parameters (maximum current density and maximum average power density) and the power source. These differences were addressed with selection of reference devices and the testing summarized above.
# **Final Conclusion:**
The performance testing provided support a substantial equivalence determination. The subject device Pelvic Muscle Trainer (model: KM510, KM516B) is substantially equivalent to the predicate device.
Page 11 / 11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.