LumiCera is used for fabricating permanent single crowns, permanent inlays and onlays, veneers, bridges.
Device Story
LumiCera is a liquid photo-curable methacrylate polymer resin used for additive manufacturing of dental restorations. Used in conjunction with a SHINING3D printer and post-curing device, the resin undergoes free radical polymerization triggered by photoinitiators. The printer builds restorations in layers, with each layer light-cured (365-405 nm) before the next is added, followed by a final post-curing process. Intended exclusively for professional dental use, the device allows clinicians to fabricate permanent crowns, inlays, onlays, veneers, and bridges. The output is a finished dental restoration that replaces missing or damaged tooth structure, potentially improving patient oral function and aesthetics. The process requires a CAD/CAM workflow including scanning, design software, and validated printing/curing equipment.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility evaluated per ISO 10993-1 and ISO 7405 (cytotoxicity, sensitization, irritation, reactivity, systemic toxicity, implantation, pyrogen, genotoxicity). Physical/mechanical properties tested per ISO 4049 and ISO 10477, meeting requirements for flexural strength (≥100 MPa), water sorption (≤40 μg/mm3), and water solubility (≤7.5 μg/mm3).
Technological Characteristics
Liquid photo-curable methacrylate polymer resin with photoinitiators and pigments. Curing via UV light (365-405 nm) and post-curing. Non-sterile. Shelf-life 2 years. Complies with ISO 4049:2019 and ISO 10477:2020. Compatible with validated 3D printers and post-curing units.
Indications for Use
Indicated for fabrication of permanent single crowns, permanent inlays, onlays, veneers, and bridges in professional dental settings.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 21, 2026
Hangzhou SHINING3D Dental Technology Co., Ltd.
% Dave Yungvirt
CEO
Third Party Review Group, LLC
7 Giralda Farms, Suite 120a
Madison, New Jersey 07940
Re: K260170
Trade/Device Name: LumiCera
Regulation Number: 21 CFR 872.3690
Regulation Name: Tooth shade resin material
Regulatory Class: Class II
Product Code: EBF
Dated: January 20, 2026
Received: January 20, 2026
Dear Dave Yungvirt:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260170 - Dave Yungvirt
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K260170 - Dave Yungvirt
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K260170 | |
| Device Name LumiCera | |
| Indications for Use (Describe) LumiCera is used for fabricating permanent single crowns, permanent inlays and onlays, veneers, bridges. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K260170
# 510(k) Summary
## I Submitter
Submitter Name: Hangzhou SHINING3D Dental Technology Co., Ltd.
Establishment
Registration Number: 3026312357
Submitter Address:
9-5-2, Tri-River Valley, Wenyan Street, Xiaoshan, Hangzhou, Zhejiang, 311258, China
Contact Person: Yuzhuo Wang
Title: Regulatory Affairs Specialist
Phone: +86 15005173276
Email: wangyuzhuo@shining3d.com
Date Prepared: January 16th, 2026
## II Device
Device Name: LumiCera
Model: CB200-A1, CB200-A2, CB200-B1, CB200-BL, CB200-C2
Regulation Name: Tooth shade resin material
Regulation Number: 21 CFR 872.3690
Regulatory Class: Class II
Product Code: EBF
Review Panel: Dental
## III Predicate Device
Trade/Device Name: FREEPRINT® crown
1 / 10
{5}
Regulation Name: Tooth shade resin material
Regulation Number: 21 CFR 872.3690
Regulatory Class: Class II
Product Code: EBF
Submitter Name: DETAX GmbH
510(k) Number: K222877
## IV Device description
The product should be used in combination with SHINING3D printer. The product is a liquid photo-curable material, which is produced by free radical polymerization of oligomers triggered by photoinitiator contained in the resin. The resin is printed in multiple layers automatically, with each layer being subjected to light curing before the addition of the next layer, and then undergoes post-curing in a curing device. The product is intended exclusively for professional dental work.
## V Intended Use /Indications for use
LumiCera is used for fabricating permanent single crowns, permanent inlays and onlays, veneers, bridges.
## VI Available model and Specification
| Model | Color |
| --- | --- |
| CB200-A1 | A1 |
| CB200-A2 | A2 |
| CB200-B1 | B1 |
| CB200-BL | 0M2 |
| CB200-C2 | C2 |
{6}
Specification: 0.5 kg/bottle, Nano
## VII Comparison to predicate devices
The following table shows comparison and discussion between the subject device and the predicate device.
3 / 10
{7}
| Item | Subject device | Predicate device | Remark |
| --- | --- | --- | --- |
| Product Name | LumiCera | FREEPRINT® crown | -- |
| Regulation Name | Tooth shade resin material | Tooth shade resin material | Same |
| Product Code | EBF | EBF | Same |
| Regulation Number | 21 CFR 872.3690 | 21 CFR 872.3690 | Same |
| Regulatory Class | Class II | Class II | Same |
{8}
| Item | Subject device | Predicate device | Remark |
| --- | --- | --- | --- |
| Indication for Use /Intended Use | LumiCera material is used for fabricating permanent single crowns, permanent inlays and onlays, veneers, bridges. | FREEPRINT® crown is indicated as an indirect restorative for both anterior and posterior restorations, including occlusal surfaces.The FREEPRINT® crown material is used for fabricating temporary or permanent restorations such as crowns and bridges, inlays, onlays, veneers and full crown restorations.Fabrication of FREEPRINT® crown requires a computer-aided and manufacturing (CAD/CAM) system that includes the following: scanner, design software, additive printer, and post-cure unit. | Similar (Analysis 1) |
{9}
| Item | Subject device | Predicate device | Remark |
| --- | --- | --- | --- |
| Technology | 3D liquid (light-cured) print resin for dental CAD/CAM | 3D liquid (light-cured) print resin for dental CAD/CAM | Same |
| Materials | LumiCera is composed of methacrylate polymer resin with photo initiator, and pigments. | Methacrylate polymer resin with photo initiator, and pigments. | Same |
| Materials Shade | Common VITA-shades: A1, A2, B1, BL, C2 | Common VITA-shades: A1, A2, A3, B1, B3, BL, C3, D3 | Similar (Analysis 2) |
| Curing Method | UV Light, 365~405 nm w/post curing | Visible light, 385 nm w/post curing | Similar (Analysis 3) |
| Equipment | Validated 3D-Printer and post curing devices. | Validated 3D-Printer and post curing devices | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Shelf-Life | 2 years | 2 years | Same |
{10}
| Item | Subject device | Predicate device | Remark |
| --- | --- | --- | --- |
| Performance Testing | ISO 4049:2019ISO 10477:2020 | ISO 4049:2019ISO 10477:2020 | Same |
| Depth of Cure | Hardness of bottom surface ≥70% top surface | Hardness of bottom surface ≥70% top surface | Same |
| Surface Finish | Glossy surface after polishing | Glossy surface after polishing | Same |
| Flexural Strength | ≥ 100 MPa | ≥ 100 MPa | Same |
| Water Sorption | ≤ 40 μg/mm3 | ≤ 40 μg/mm3 | Same |
| Water Solubility | ≤ 7.5 μg/mm3 | ≤ 7.5 μg/mm3 | Same |
| Biocompatibility Testing | Comply with ISO 10993-1:2018, and ISO 7405:2018 | Comply with ISO 10993-1:2018, and ISO 7405:2018 | Same |
{11}
VIII Discussion for similarities
**Analysis 1. Indication for Use/Intended Use**
Both subject devices and predicate devices can be used to manufacture permanent single crowns, permanent inlays and onlays, veneers, and bridges with same manufacturing processes. Predicate devices have a wider range of intended use compared to subject devices. The minor difference does not influence substantial equivalence.
**Analysis 2. Materials Shade**
The different color due to the composition in the content of pigments. These color variations do not raise new questions of safety or effectiveness, as both devices meet the required performance and biocompatibility standards. Therefore, the differences in color do not impact the substantial equivalence of the subject device to the predicate device.
**Analysis 3. Curing Method**
The Subject and Predicate devices are both photo-curable polymer resins. Photo-curable polymer resins exhibit similar free radical polymerization efficiency among 365 nm ~405 nm light regions. Slight differences in the curing light wavelength do not influence substantial equivalence.
IX Summary of Testing (Performance Data):
**Non-Clinical Performance Test Conclusion**
**Biocompatibility testing**
Based on ISO 10993-1 and ISO 7405, the subject device is categorized as a surface device in contact with mucosal membrane with Long-term contact (>30d). The subject device was evaluated for:
8 / 10
{12}
- Cytotoxicity Test
- Sensitization Test
- Oral Mucosa Irritation Test
- Intracutaneous Reactivity Test
- Acute Systemic Toxicity Test
- Subchronic Systemic Toxicity Test
- Implantation Test
- Pyrogen Test
- Genotoxicity Test
## Bench Testing:
Physical and mechanical properties of the subject device were evaluated according to FDA-recognized version ISO 4049 Dentistry - Polymer-based restorative materials and ISO 10477 Dentistry - Polymer-based crown and veneering materials.
The test results demonstrated the Subject device meets the property requirements of the referenced standards.
Validation of the manufacturing process and compatible equipment was performed demonstrating consistency of the process output with that of the process input.
## Sterility and Shelf-Life Testing:
The device is provided non-sterile.
From the Shelf-life testing, LumiCera has a shelf life of 2 years.
## Clinical Test Conclusion:
No clinical study is included in this submission.
9 / 10
{13}
X Conclusion
The subject device is as safe and effective as its predicate device. The subject device has similar intended use, and same material, technology and performance characteristics.
The minor differences among the subject device and predicate devices have not raised extra safety and performance concerns, based on the relevant tests and evaluations provided in this submission. Performance data confirm that the subject device demonstrates equivalent safety and effectiveness to the predicate device. Thus, LumiCera is substantially equivalent.
10 / 10
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.