K260165 · Spectrum Vascular · FOZ · Aug 10, 2026 · General Hospital
Device Facts
Record ID
K260165
Device Name
Unimpregnated Central Venous Catheter
Applicant
Spectrum Vascular
Product Code
FOZ · General Hospital
Decision Date
Aug 10, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only The device is a short-term use catheter, intended for less than 30 days. The Cook dilator is used for dilating puncture sites or catheter tracts. The Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations. The Unimpregnated Central Venous Catheter with Power Injection is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only 7. Power injection of contrast media* *The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0 and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0 and 10.0 French catheters. The device is a short-term use catheter, intended for less than 30 days. The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter. The Cook dilator is used for dilating puncture sites or catheter tracts. The Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.
Device Story
Central venous catheter (CVC) for short-term vascular access; inserted via Seldinger technique; tip positioned at SVC-RA junction. Available in single, double, triple, or five-lumen configurations; sizes 2.5-10 French. Used for drug infusion, blood monitoring, sampling, and hyperalimentation. Power-injectable models support contrast media delivery at defined flow/pressure limits. Includes wire guide, access needle, and optional accessories like dilators. Used in clinical settings by healthcare professionals. Benefits include reliable central venous access for critical care and diagnostic imaging. Subject device differs from predicate only by use of TPC 980 black ink for shaft markings.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility testing per ISO 10993 and ink legibility performance testing.
Technological Characteristics
Materials: Polyethylene, ethylene-vinyl acetate, polyurethane (shaft/hub); stainless steel (needle/wire guide). Dimensions: 2.5-10 Fr, 5-26.5 cm length. Sterilization: EtO (SAL 10^-6). Non-powered, non-electrical, no software.
Indications for Use
Indicated for adult and pediatric patients requiring short-term (<30 days) central venous access for drug infusion, CVP monitoring, hyperalimentation, blood sampling, blood product delivery, or power injection of contrast media (specific models).
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
Predicate Devices
Cook Unimpregnated Central Venous Catheter (K182252)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
8/10/2026
Spectrum Vascular
Sharon Klugewicz
Sr VP, Quality & Regulatory Affairs
50 Main St.
Suite 1000
White Plains, New York 10606
Re: K260165
Trade/Device Name: Unimpregnated Central Venous Catheter
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular Catheter
Regulatory Class: Class II
Product Code: FOZ
Dated: July 9, 2026
Received: July 9, 2026
Dear Sharon Klugewicz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260165 - Sharon Klugewicz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260165 - Sharon Klugewicz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
GANG PENG -S
Gang Peng for
David Wolloscheck
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260165
Please provide the device trade name(s).
Unimpregnated Central Venous Catheter
Please provide your Indications for Use below.
The Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:
1. Continuous or intermittent drug infusions
2. Central venous blood pressure monitoring (CVP)
3. Acute hyperalimentation
4. Blood sampling
5. Delivery of whole blood or blood products
6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
The device is a short-term use catheter, intended for less than 30 days.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations.
The Unimpregnated Central Venous Catheter with Power Injection is used for:
1. Continuous or intermittent drug infusions
2. Central venous blood pressure monitoring (CVP)
3. Acute hyperalimentation
4. Blood sampling
5. Delivery of whole blood or blood products
6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
7. Power injection of contrast media*
*The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0 and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0 and 10.0 French catheters.
The device is a short-term use catheter, intended for less than 30 days.
The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
510(k) Summary
Spectrum Vascular
K260165
# **1. SUBMITTER INFORMATION**
Applicant: Spectrum Vascular
Contact: Ms. Sharon Klugewicz
Phone: 516-425-4446
Email: sklugewicz@spectrumvascular.com
Address: 50 Main Street, Suite 1000
White Plains, NY 10606
# **2. CORRESPONDENT INFORMATION**
Contact: Ms. Sharon Klugewicz
Title: SVP, Quality and Regulatory Affairs
Firm: Spectrum Vascular
50 Main Street, Suite 1000
White Plains, NY 10606
Phone: 516-425-4446
Email: sklugewicz@spectrumvascular.com
# **3. DATE PREPARED: AUGUST 6, 2026**
# **4. DEVICE INFORMATION**
Device Name: Unimpregnated Central Venous Catheter
Common Name: Intravascular catheter
Regulation Number: 21 CFR 880.5200
Regulation Name: Intravascular catheter
Product Code: FOZ
Regulatory Class: Class II
# **5. PREDICATE DEVICE INFORMATION**
Predicate Device Name: Cook Unimpregnated Central Venous Catheter
510(k) Number: K182252
Manufacturer: Cook Incorporated
Page 1 of 6
{5}
510(k) Summary
Spectrum Vascular
K260165
## 6. DEVICE DESCRIPTION
The Unimpregnated Central Venous Catheter set includes a central venous catheter (CVC), wire guide, access needle, and syringe. The CVC is inserted into the vasculature using the Seldinger technique. The tip of the catheter is then advanced until it is above the superior vena cava-right atrium (SVC-RA) junction. The CVC is a single, double, triple, or five-lumen shaft manufactured from polyethylene, ethylene-vinyl acetate, or polyurethane tubing. The CVC is offered with and without power injection, and supplied in multiple size configurations, ranging from 2.5 to 10 French, to support a variety of vascular access needs. Single lumen CVCs are designed with a pre-molded winged hub on the proximal end. Double, triple, and five lumen CVCs are manufactured with a manifold assembly, which is designed as a winged manifold connected to extension tubes. Each extension tube has a slide clamp and is manufactured with a proximal winged hub. The wire guide is manufactured with stainless steel and may be coated with polytetrafluoroethylene (PTFE). The access needle is manufactured from stainless steel. Additional set configurations may also contain a dilator, injection caps, clamps, catheter securement device, connecting tube, and other convenience accessories.
## 7. INDICATIONS FOR USE
The Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:
1. Continuous or intermittent drug infusions
2. Central venous blood pressure monitoring (CVP)
3. Acute hyperalimentation
4. Blood sampling
5. Delivery of whole blood or blood products
6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
The device is a short-term use catheter, intended for less than 30 days.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations.
The Unimpregnated Central Venous Catheter with Power Injection is used for:
1. Continuous or intermittent drug infusions
2. Central venous blood pressure monitoring (CVP)
3. Acute hyperalimentation
4. Blood sampling
5. Delivery of whole blood or blood products
6. Simultaneous, separate infusion of drugs for multi-lumen catheters only
Page 2 of 6
{6}
510(k) Summary
Spectrum Vascular
K260165
### 7. Power injection of contrast media*
*The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0 and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0 and 10.0 French catheters.
The device is a short-term use catheter, intended for less than 30 days.
The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter.
The Cook dilator is used for dilating puncture sites or catheter tracts.
The Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations.
## 8. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Other than the change in ink as noted in Table 1, the subject device is identical to the predicate device.
Table 1: Technological Comparison to Predicate Device
| | Predicate Device (K182252) | Subject Device (K260165) | Comparison |
| --- | --- | --- | --- |
| Regulation | 21 CFR 880.5200 | 21 CFR 880.5200 | Same |
| Product Code | FOZ | FOZ | Same |
| Classification | Class II | Class II | Same |
| Indications for Use | The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:1. Continuous or intermittent drug infusions2. Central venous blood pressure monitoring (CVP)3. Acute hyperalimentation4. Blood sampling5. Delivery of whole blood or blood products6. Simultaneous, separate infusion of drugs for multi-lumen catheters only.The device is a short-term use catheter, intended for less than 30 days.The Cook dilator is used for dilating puncture sites or catheter tracts. | The Unimpregnated Central Venous Catheter (Non-Power Injectable) is used for:1. Continuous or intermittent drug infusions2. Central venous blood pressure monitoring (CVP)3. Acute hyperalimentation4. Blood sampling5. Delivery of whole blood or blood products6. Simultaneous, separate infusion of drugs for multi-lumen catheters onlyThe device is a short-term use catheter, intended for less than 30 days.The Cook dilator is used for dilating puncture sites or catheter tracts. | The subject device Indications for Use is the same as the predicate device with the exception of the “Cook” name being removed in the device name. |
Page 3 of 6
{7}
510(k) Summary
Spectrum Vascular
K260165
| | Predicate Device (K182252) | Subject Device (K260165) | Comparison |
| --- | --- | --- | --- |
| | The Cook Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations. The Cook Unimpregnated Central Venous Catheter with power injection is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only 7. Power injection of contrast media* *The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0, and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0, and 10.0 French catheters. The device is a short-term use catheter, intended for less than 30 days. The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter. The Cook dilator is used for dilating puncture sites or catheter tracts. The Cook Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations. | The Unimpregnated Central Venous Catheter (Non-Power Injectable) is intended for adult and pediatric populations. The Unimpregnated Central Venous Catheter with Power Injection is used for: 1. Continuous or intermittent drug infusions 2. Central venous blood pressure monitoring (CVP) 3. Acute hyperalimentation 4. Blood sampling 5. Delivery of whole blood or blood products 6. Simultaneous, separate infusion of drugs for multi-lumen catheters only 7. Power injection of contrast media* *The flow rate may not exceed 3 mL/sec for 4.0 and 5.0 French catheters and 10 mL/sec for 7.0, 8.0, 9.0 and 10.0 French catheters. Verify prior to use that the maximum safety cut-off pressure limit is set at or below 250 psi for 4.0 and 5.0 French catheters and 325 psi for 7.0, 8.0, 9.0 and 10.0 French catheters. The device is a short-term use catheter, intended for less than 30 days. The 9.0 and 10.0 French catheters include an inner catheter to facilitate insertion of the main catheter. The Cook dilator is used for dilating puncture sites or catheter tracts. The Unimpregnated Central Venous Catheter with power injection is intended for adult and pediatric populations. | |
| Patient Population | Neonates/Newborns; Infants; Children; Adolescents; Adults | Neonates/Newborns; Infants; Children; Adolescents; Adults | Same |
| Catheter Placement Method | Percutaneously via Seldinger technique | Percutaneously via Seldinger technique | Same |
| Catheter Tip Target Anatomy | SVC-RA junction | SVC-RA junction | Same |
Page 4 of 6
{8}
510(k) Summary
Spectrum Vascular
K260165
| | Predicate Device (K182252) | Subject Device (K260165) | Comparison |
| --- | --- | --- | --- |
| Catheter tip Location Confirmation Method | ECG/Ultrasound/Fluoroscopy | ECG/Ultrasound/Fluoroscopy | Same |
| Power Injectable Catheter Configurations (7-10Fr) | Dual Lumen: 8 FrTriple Lumen: 7, 9 FrFive-Lumen: 10 Fr | Dual Lumen: 8 FrTriple Lumen: 7, 9 FrFive-Lumen: 10 Fr | Same |
| Upper Pressure Limit for Power Injection (7-10Fr) | 325 psi | 325 psi | Same |
| Power Injectable Catheter Configurations (4-5 Fr) | Dual Lumen: 4, 5 FrTriple Lumen: 5 Fr | Dual Lumen: 4, 5 FrTriple Lumen: 5 Fr | Same |
| Upper Pressure Limit for Power Injection (4-5 Fr) | 250 psi | 250 psi | Same |
| Catheter Shaft Material | PolyethyleneEthylene-Vinyl AcetatePolyurethane | PolyethyleneEthylene-Vinyl AcetatePolyurethane | Same |
| Catheter Shaft Marking Ink | UVA Ink (Color #1 Black) | UVA Ink (TPC 980 Black Ink) | Different: Biocompatibility testing was conducted in accordance with ISO 10993 and supports a determination that the new ink is biocompatible and thus the catheters are substantially equivalent.An additional ink legibility performance test was conducted to demonstrate that the device meets its design requirements and performs as intended. |
| Catheter Hub Material | PolyethylenePolyurethane | PolyethylenePolyurethane | Same |
| Number of Catheter Lumens | 1, 2, 3, 5 | 1, 2, 3, 5 | Same |
| Catheter Outer Diameter (Fr) | Single – 2.5, 3, 4, 5, 6, 6.3 FrDual – 4, 5, 7.5, 8 FrTriple – 5, 7, 9 FrFive – 10 Fr | Single – 2.5, 3, 4, 5, 6, 6.3 FrDual – 4, 5, 7.5, 8 FrTriple – 5, 7, 9 FrFive – 10 Fr | Same |
| Catheter Length | 5 to 26.5 cm | 5 to 26.5 cm | Same |
| Packaging | PETG Tray w/ Tyvek Lidstock | PETG Tray w/ Tyvek Lidstock | Same |
| Single Use? | Yes | Yes | Same |
| Sterilization Method | EtO | EtO | Same |
| Sterility Assurance Level | \( 10^{-6} \) | \( 10^{-6} \) | Same |
Page 5 of 6
{9}
510(k) Summary
Spectrum Vascular
K260165
# 9. SUMMARY OF NON-CLINICAL PERFORMANCE TESTING
# Biocompatibility Testing
The catheter of the subject device is identical to the Unimpregnated Central Venous Catheter cleared in K182252 with the exclusion of the marker ink. Additional biocompatibility testing on the subject device was conducted in accordance with the ISO 10993 series.
# Electrical Safety
Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type.
# Electromagnetic Compatibility (EMC)
Not applicable. The subject device contains no electrical components, generates no electrical emissions, and uses no electrical energy of any type.
# Software
Not applicable. The subject device contains no software.
# Performance Testing
An additional ink legibility performance test was conducted to demonstrate that the device meets its design requirements and performs as intended.
# 10. CONCLUSION
The results of the performance testing described above demonstrate that the Unimpregnated Central Venous Catheter performs equivalently as compared to the predicate device and supports a determination of substantial equivalence.
Page 6 of 6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.