KLS Martin IPS Oral-Max Implants - MR Conditional; Individual Patient Solutions - Ti (IPS-Ti); Individual Patient Solutions - PEEK (IPS-PEEK); KLS Martin Individual Patient Solutions
Applicant
KLS-Martin L.P.
Product Code
JEY · Dental
Decision Date
Aug 17, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4760
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Individual Patient Solutions - Ti (IPS-Ti) is intended to replace bony voids in the mandibular, maxillofacial or craniofacial skeleton. Individual Patient Solutions - PEEK (IPS-PEEK) is intended to replace bony voids in the cranial and/or craniofacial skeleton. KLS Martin Individual Patient Solutions (IPS) is intended as a pre-operative software tool for simulating / evaluating surgical treatment options as a software and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS software and the result is an output data file that may then be provided as digital models or used as input in an additive manufacturing portion of the system that produces physical outputs including implants, anatomical models, guides, splints, and case reports for use in maxillofacial, midface, & mandibular surgery. KLS Martin Individual Patient Solutions (IPS) implant devices are intended for use in the stabilization, fixation, and reconstruction of the maxillofacial / midface and mandibular skeletal regions in children (2 years of age to < 12 years of age), adolescents (12 years of age - 21 years of age), and adults.
Device Story
System uses CT scan data to create patient-specific implants, anatomical models, guides, and splints; physician collaborates with KLS Martin engineers to manipulate digital models; virtual fitting performed on 3D anatomy models to verify dimensions; implants manufactured via subtractive or additive methods from CP Titanium (ASTM F67) or Ti-6Al4V (ASTM F136); devices fixated with titanium screws; used in maxillofacial, midface, and mandibular surgery; provides pre-operative simulation and surgical planning; benefits include patient-specific fit for skeletal reconstruction; subject device adds MR Conditional safety labeling.
Clinical Evidence
No clinical data. Bench testing only. Non-clinical testing included MR safety validation (displacement force, torque, image artifacts, RF-induced heating) and comparative static/dynamic mechanical testing for material grade change (Grade 2 vs Grade 4 titanium).
Technological Characteristics
Materials: CP Titanium (ASTM F67) or Ti-6Al4V (ASTM F136); PEEK. Manufacturing: Subtractive or additive. Connectivity: Software-based image processing system. Sterilization: Non-sterile. MR Conditional status supported by ASTM F2052-21, F2213-17, F2119-07, F2182-19e2.
Indications for Use
Indicated for patients (children 2-12, adolescents 12-21, and adults) requiring stabilization, fixation, and reconstruction of maxillofacial, midface, and mandibular skeletal regions, or replacement of bony voids in the mandibular, maxillofacial, craniofacial, or cranial skeleton.
Regulatory Classification
Identification
A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.
KLS Martin Oral-Max Implants MR Conditional (K241314)
Submission Summary (Full Text)
{0}
**FDA** U.S. FOOD & DRUG
ADMINISTRATION
August 17, 2026
KLS-Martin L.P.
Melissa Bachorski
Senior Director, RA/QMS North America
11201 Saint Johns Industrial Pkwy. S
Jacksonville, Florida 32246
Re: K260137
Trade/Device Name: KLS Martin IPS Oral-Max Implants - MR Conditional; Individual Patient Solutions - Ti (IPS-Ti); Individual Patient Solutions - PEEK (IPS-PEEK); KLS Martin Individual Patient Solutions
Regulation Number: 21 CFR 872.4760
Regulation Name: Bone Plate
Regulatory Class: Class II
Product Code: JEY, DZJ, LLZ, GXN, GXP
Dated: July 24, 2026
Received: July 24, 2026
Dear Melissa Bachorski:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260137 - Melissa Bachorski
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K260137 - Melissa Bachorski
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# ANDREW I. STEEN -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260137
Please provide the device trade name(s).
KLS Martin IPS Oral-Max Implants - MR Conditional;
K062570: Individual Patient Solutions - Ti (IPS-Ti);
K072707: Individual Patient Solutions - PEEK (IPS-PEEK) ;
K210731: KLS Martin Individual Patient Solutions
Please provide your Indications for Use below.
Individual Patient Solutions - Ti (IPS-Ti) (K062570)*:
Individual Patient Solutions - Ti (IPS-Ti) is intended to replace bony voids in the mandibular, maxillofacial or craniofacial skeleton.
Individual Patient Solutions - PEEK (IPS-PEEK) (K072707)*:
Individual Patient Solutions - PEEK (IPS-PEEK) is intended to replace bony voids in the cranial and/or craniofacial skeleton.
KLS Martin Individual Patient Solutions (K210731):
KLS Martin Individual Patient Solutions (IPS) is intended as a pre-operative software tool for simulating / evaluating surgical treatment options as a software and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS software and the result is an output data file that may then be provided as digital models or used as input in an additive manufacturing portion of the system that produces physical outputs including implants, anatomical models, guides, splints, and case reports for use in maxillofacial, midface, & mandibular surgery.
KLS Martin Individual Patient Solutions (IPS) implant devices are intended for use in the stabilization, fixation, and reconstruction of the maxillofacial / midface and mandibular skeletal regions in children (2 years of age to < 12 years of age), adolescents (12 years of age - 21 years of age), and adults.
*Formerly known as "Patient Contoured Mesh"
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
{4}
510(k) #: K260137
# 510(k) Summary
Prepared on: 2026-08-17
Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | KLS-Martin L.P. |
| --- | --- |
| Applicant Address | 11201 Saint Johns Industrial Pkwy S Jacksonville FL 32246 United States |
| Applicant Contact Telephone | 904-641-7746 |
| Applicant Contact | Ms. Melissa Bachorski |
| Applicant Contact Email | rapm_na@klsmartin.com |
| Correspondent Name | KLS-Martin L.P. |
| Correspondent Address | 11201 Saint Johns Industrial Pkwy S Jacksonville FL 32246 United States |
| Correspondent Contact Telephone | 904-641-7746 |
| Correspondent Contact | Ms. Tugce Kalin |
| Correspondent Contact Email | rapm_na@klsmartin.com |
Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | KLS Martin IPS Oral-Max Implants - MR Conditional; K062570: Individual Patient Solutions - Ti (IPS-Ti); K072707: Individual Patient Solutions - PEEK (IPS-PEEK) ; K210731: KLS Martin Individual Patient Solutions |
| --- | --- |
| Common Name | Bone plate |
| Classification Name | Plate, Bone |
| Regulation Number | 872.4760 |
| Product Code(s) | JEY, DZJ, LLZ, GXN, GXP |
Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K210731 | KLS Martin Individual Patient Solutions | JEY |
| K062570 | Individual Patient Solutions - Ti (IPS-Ti) f/k/a Patient Contoured Mesh | GXN |
| K072707 | Individual Patient Solutions - PEEK (IPS-PEEK) f/k/a Patient Contoured Mesh | GXN |
| K241314 | KLS Martin Oral-Max Implants MR Conditional (bundled) | JEY |
{5}
# Device Description Summary
21 CFR 807.92(a)(4)
Individual Patient Solutions - Ti (IPS-Ti) (K062570)*:
The Individual Patient Solutions - Ti (IPS-Ti) is a preformed implant manufactured from Titanium material. The implant is pre-shaped to fit the anatomy of the patient using a CT-based model of the patient's skull. The IPS-Ti is fixated to native bone using previously cleared KLS Martin titanium screws.
Individual Patient Solutions - PEEK (IPS-PEEK) (K072707)*:
The Individual Patient Solutions - PEEK (IPS-PEEK) is a preformed implant manufactured from PEEK material. The implant is pre-shaped to fit the anatomy of the patient using a CT-based model of the patient's skull. The IPS-PEEK is fixated to native bone using previously cleared KLS Martin titanium plates and screws.
KLS Martin Individual Patient Solutions (K210731):
KLS Martin Individual Patient Solutions (IPS) is comprised of a collection of software and associated additive manufacturing equipment intended to produce various outputs to support reconstructive and orthognathic surgeries. The system processes the medical images to produce various patient-specific physical and/or digital output devices which include implants, anatomical models, guides, splints, and case reports.
Patient-specific metallic bone plates are used in conjunction with metallic bone screws for internal fixation of maxillofacial, midface, and mandibular bones. The devices are manufactured based on medical imaging (CT scan) of the patient's anatomy with input from the physician during virtual planning and prior to finalization and production of the device. The physician provides input for model manipulation and interactive feedback by viewing digital models of planned outputs that are modified by trained KLS Martin engineers during the planning session. For each design iteration, verification is performed by virtually fitting the generated output device over a 3D model of the patient's anatomy to ensure its dimensional properties allow an adequate fit.
Implants are provided non-sterile and are manufactured using traditional (subtractive) or additive manufacturing methods from either CP Titanium (ASTM F67) or Ti-6Al4V (ASTM F136). These patient-specific devices are fixated with previously cleared KLS Martin screws.
*Formerly known as "Patient Contoured Mesh"
# Intended Use/Indications for Use
21 CFR 807.92(a)(5)
Individual Patient Solutions - Ti (IPS-Ti) (K062570)*:
Individual Patient Solutions - Ti (IPS-Ti) is intended to replace bony voids in the mandibular, maxillofacial or craniofacial skeleton.
Individual Patient Solutions - PEEK (IPS-PEEK) (K072707)*:
Individual Patient Solutions - PEEK (IPS-PEEK) is intended to replace bony voids in the cranial and/or craniofacial skeleton.
KLS Martin Individual Patient Solutions (K210731):
KLS Martin Individual Patient Solutions (IPS) is intended as a pre-operative software tool for simulating / evaluating surgical treatment options as a software and image segmentation system for the transfer of imaging information from a medical scanner such as a CT based system. The input data file is processed by the IPS software and the result is an output data file that may then be provided as digital models or used as input in an additive manufacturing portion of the system that produces physical outputs including implants, anatomical models, guides, splints, and case reports for use in maxillofacial, midface, & mandibular surgery.
KLS Martin Individual Patient Solutions (IPS) implant devices are intended for use in the stabilization, fixation, and reconstruction of the maxillofacial / midface and mandibular skeletal regions in children (2 years of age to < 12 years of age), adolescents (12 years of age - 21 years of age), and adults.
*Formerly known as "Patient Contoured Mesh"
# Indications for Use Comparison
21 CFR 807.92(a)(5)
The indications for use for the subject and predicate devices are identical. The basis of this submission is to support the conditional safety and labeling modification of the subject device, KLS Martin IPS Oral-Max Implants - MR Conditional, in the magnetic resonance environment.
# Technological Comparison
21 CFR 807.92(a)(6)
Similarities to the Predicate Devices
{6}
The subject and predicate devices have the same intended use, design, manufacturing methods, principles of operation, sterilization, packaging, and overall technological characteristics. The subject devices continue to be manufactured from commercially pure titanium and are intended for the same clinical applications as the legally marketed predicate devices.
Differences from the Predicate Devices
MR Environment Safety Information:
The subject device labeling differs from the predicate device labeling by adding safety information for use in the magnetic resonance environment. Non-clinical testing has been performed to support the conditional safety of the subject device in the MR environment. Hazards addressed include magnetically induced displacement force (ASTM F2052-21) and torque (ASTM F2213-17), image artifacts (ASTM F2119-07, R2013), and RF induced heating (ASTM F2182-19e2).
Material Grade Change and Mechanical Performance:
The applicable IPS Trauma acute mandible trauma (AMT) implants include a material grade change from ASTM F67 commercially pure titanium Grade 2 to ASTM F67 commercially pure titanium Grade 4. No changes have been made to the implant design, dimensions, manufacturing process, compatible components, or intended use. The material grade is the only technological difference between the subject and predicate devices.
Comparative static and dynamic mechanical testing was performed on the worst-case device configuration using devices that were identical in design, dimensions, and manufacturing procedures, with the material grade being the only variable. The testing demonstrated that the material grade change does not negatively affect device performance and supports equivalent or improved mechanical performance compared to the previously cleared Grade 2 implants. Therefore, the material grade change does not raise different questions of safety or effectiveness.
Conclusion
Based on the technological comparison and the mechanical performance testing provided in this submission, the subject devices remain substantially equivalent to the legally marketed predicate devices.
# Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Nonclinical Testing:
MR Environment Safety Testing:
Nonclinical testing has been provided to support the conditional safety of the subject devices in the MR environment. Hazards addressed include magnetically induced displacement force (ASTM F2052-21) and torque (ASTM F2213-17), image artifacts (ASTM F2119-07, R2013), and RF-induced heating (ASTM F2182-19e2).
Performance data demonstrate that KLS Martin IPS Oral-Max Implants, when used in the MR environment using specified MR parameters and instructions, are not adversely affected by magnetically-induced displacement force and torque, RF-heating, and image artifacts.
Material Grade Change and Mechanical Performance Testing:
The applicable IPS Trauma acute mandible trauma (AMT) implants include a material grade change from ASTM F67 commercially pure titanium Grade 2 to ASTM F67 commercially pure titanium Grade 4. No changes have been made to the implant design, dimensions, manufacturing process, compatible components, or intended use. The material grade is the only technological difference between the subject and predicate devices.
Comparative static and dynamic mechanical testing was performed on the worst-case device configuration using devices that were identical in design, dimensions, and manufacturing procedures, with the material grade being the only variable. The testing demonstrated that the material grade change does not negatively affect device performance and supports equivalent or improved mechanical performance compared to the previously cleared Grade 2 implants. Therefore, the material grade change does not raise different questions of safety or effectiveness.
Clinical Testing:
Not Applicable
Conclusions:
Based on the results of the nonclinical testing and other supporting information, KLS Martin IPS Oral-Max Implants can be safely scanned under the conditions presented in the labeling and has been demonstrated to be substantially equivalent to the predicate devices with respect to intended use, technological characteristics, and performance. Any differences in design or materials do not raise new questions of safety and effectiveness. Therefore, the subject device is as safe, effective, and performs as intended within the parameters of MR Conditional use as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.