K260133 · Sep 12, 2026
CADDIE
Odin Medical Limited
Recorded colonoscopy videos from routine clinical procedures; Photo-documented static frames from routine colonoscopies
Retrospective clinical datasets were used to conduct standalone performance testing of the CADDIE software algorithms (polyp detection and Cecum AI) to demonstrate substantial equivalence.
The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients. The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only.
Device Story
CADDIE is a cloud-based AI software that interfaces with endoscopic video processors during colonoscopies. It analyzes real-time video feeds to identify visual characteristics of colorectal polyps; it superimposes green bounding boxes on detected lesions and emits low-volume audio alerts. It includes a 'Cecum AI' feature that identifies cecal landmarks (appendiceal orifice, ileocecal valve) in photo-documented frames to remind clinicians to verify polyp detection status. Used in hospitals/clinics by gastroenterologists; output serves as an aid to visual inspection. The clinician retains full responsibility for confirming polyp presence and clinical decision-making. The device is compatible with standard white-light endoscopy and the Arc Endocuff Vision accessory.
Clinical Evidence
No new clinical data provided; relies on baseline clinical performance evaluation reviewed in K240044. Standalone bench testing used 632 subjects (1,112/1,139 polyps detected; 97.63% object-level TPR; 1.72% frame-level FPR). Cecum AI performance evaluated on 464 procedures (89.65% frame-level TPR; 7.53% frame-level FPR).
Technological Characteristics
Standalone cloud-based software accessed via web browser; requires network connection. Utilizes AI-based algorithms for polyp detection and cecum identification. Compatible with standard white-light endoscopy and Arc Endocuff Vision. Software version 1.6.0.
Indications for Use
Indicated for gastroenterologists to assist in detecting suspected colorectal polyps in patients aged 45 and over undergoing screening and surveillance endoscopic mucosal evaluations. Not indicated for pregnant women.
Regulatory Classification
Identification
A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope.
Special Controls
In combination with the general controls of the FD&C Act, the gastrointestinal lesion software detection system is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
(4) Performance data must demonstrate electromagnetic compatibility and electrical safety, mechanical safety, and thermal safety testing for any hardware components of the device.
(5) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
(6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
(ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
(iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
(iv) A summary of the standalone performance testing and associated statistical analysis.
{1}

September 12, 2026
Odin Medical Limited
Luke Sampson
COO
74 Rivington St.
London, EC2A 3AY
United Kingdom
Re: K260133
Trade/Device Name: CADDIE
Regulation Number: 21 CFR 876.1520
Regulation Name: Gastrointestinal lesion software detection system
Regulatory Class: Class II
Product Code: QNP, SBX
Dated: August 14, 2026
Received: August 14, 2026
Dear Luke Sampson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{2}
K260133 - Luke Sampson
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{3}
K260133 - Luke Sampson
Page 3
Sincerely,
# SHANIL P. HAUGEN -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity, and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{4}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260133
Device Name
CADDIE
Indications for Use (Describe)
The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment.
CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients.
The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only.
Type of Use (Select one or both, as applicable)
☑
Prescription Use (Part 21 CFR 801 Subpart D)
☐
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
{5}
K260133
K260133 Page 1 of 14
Odin Vision
## 510(k) Summary
| **510(k) Number:** | K260133 |
| --- | --- |
| **510(k) Type:** | Special |
| **Date Prepared:** | 14-Aug-2026 |
| **510(k) Owner:** | Odin Medical Ltd, 74 Rivington Street, London EC2A 3AY United Kingdom Tel - +44 (0)7957 948411 |
| **Official Contact:** | Luke Sampson - COO |
| **Submission Correspondent:** | Luke Sampson - COO |
| **Proprietary or Trade Name:** | CADDIE |
| **Common/ Usual Name:** | Gastrointestinal Lesion Software Detection System |
| **Classification CFR:** | 21 CFR 876.1520 |
| **Classification Code:** | QNP, SBX |
| **Classification Name:** | Gastrointestinal Lesion Software Detection System |
| **Class:** | Class II |
| **Predicate Device:** | K252586 CADDIE – Odin Medical LTD |
Page 1 of 14
{6}
K260133
K260133 Page 2 of 14
Odin Vision
### Device Description:
CADDIE is cloud based artificial intelligence medical device software. CADDIE interfaces with the video feed generated by an endoscopic video processor during a colonoscopy procedure.
CADDIE also includes a Cecum AI feature as an additional convenience function that prompts the user to verify that the CADDIE detection mode is enabled before inspection begins.
The software is intended to be used by trained and qualified healthcare professionals as an accompaniment to video endoscopy for the purpose of drawing attention to regions with visual characteristics consistent with colonic mucosal lesions (such as polyps and adenomas).
CADDIE analyses the data from the endoscopic video processor in real-time and provides information to aid the endoscopist in detecting suspected colorectal polyps, if they are in the field of view of the endoscope.
The areas highlighted by CADDIE are not to be interpreted as definite polyps or adenomas. The responsibility to make a decision as to whether or not a highlighted region contains a polyp or is an adenoma lies with the user. The endoscopist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp and its classification based on their own medical judgement.
### Indications for Use:
The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment.
CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic procedures, medical diagnosis, or recommendations of treatment/course of action for patients.
The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only.
### Patient Population:
CADDIE is intended to be used on patients aged 45 and over referred for screening and surveillance endoscopic mucosal evaluations. This does not include pregnant women, for which no clinical evaluation has been carried out.
### Environments of use:
Hospitals and clinics or in other secure endoscopy units where colonoscopies are performed.
### Summary of Technological Characteristics:
At a high level we present the technological comparison of the subject device and the predicate in the Table below.
Page 2 of 14
{7}
K260133
Odin Vision
K260133 Page 3 of 14
# Summary of Technological Characteristics
| Characteristics | Subject Device: CADDIE (Software version 1.6.0) | Predicate Device: CADDIE K252586 (Software version 1.5.0) | Comparison |
| --- | --- | --- | --- |
| Manufacturer | Odin Medical Ltd | Odin Medical Ltd | Same |
| Regulation Number | 21 CFR 876.1520 | 21 CFR 876.1520 | Same |
| Regulation Title | Gastrointestinal lesion software detection system | Gastrointestinal lesion software detection system | Same |
| Classification | Class II | Class II | Same |
| Classification Product Code | QNP, SBX | QNP, SBX | Same |
| Intended Use - definition | A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope. | A gastrointestinal lesion software detection system is a computer-assisted detection device used in conjunction with endoscopy for the detection of abnormal lesions in the gastrointestinal tract. This device with advanced software algorithms brings attention to images to aid in the detection of lesions. The device may contain hardware to support interfacing with an endoscope. | Same |
| Indications for Use | The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic | The CADDIE computer-assisted detection device is intended to assist the gastroenterologist in detecting suspected colorectal polyps only. The gastroenterologist is responsible for reviewing CADDIE suspected polyp areas and confirming the presence or absence of a polyp based on their own medical judgment. CADDIE is not intended to replace a full patient evaluation, nor is it intended to be relied upon to make a primary interpretation of endoscopic | Same |
Page 3 of 14
{8}
K260133
Odin Vision
K260133 Page 4 of 14
| Characteristics | Subject Device: CADDIE (Software version 1.6.0) | Predicate Device: CADDIE K252586 (Software version 1.5.0) | Comparison |
| --- | --- | --- | --- |
| | procedures, medical diagnosis, or recommendations of treatment/course of action for patients. The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. | procedures, medical diagnosis, or recommendations of treatment/course of action for patients. The CADDIE computer-assisted detection device is limited for use with standard white-light endoscopy imaging only. | |
| **Patient Population** | CADDIE is intended to be used on patients aged 45 and over referred for screening and surveillance endoscopic mucosal evaluations. This does not include pregnant women, for which no clinical evaluation has been carried out. | CADDIE is intended to be used on patients aged 45 and over referred for screening and surveillance endoscopic mucosal evaluations. This does not include pregnant women, for which no clinical evaluation has been carried out. | Same |
| **Technological Characteristics** | CADDIE is standalone software that is deployed on the cloud and accessed via a web browser. The device is designed to highlight portions of the colon where the device detects potential colorectal polyps. CADDIE requires a network connection unless deployed locally. | CADDIE is standalone software that is deployed on the cloud and accessed via a web browser. The device is designed to highlight portions of the colon where the device detects potential colorectal polyps. CADDIE requires a network connection unless deployed locally. | Same |
| **Convenience Features** | When the clinician confirms the cecum and photo documents the cecal landmarks in standard clinical workflow, Cecum AI is triggered, and the image is sent for analysis. If the cecal landmarks are confirmed in the image, by the AI, the user will be given a reminder to check the status of Polyp detection. If polyp detection is turned off, the cross icon will flash three times along with the cecum icon. If detection is on, the cecum icon will flash three times next to the tick icon. This is a convenience feature that provides a check to the user that the CADDIE polyp detection function is on and in use. | When the clinician confirms the cecum and photo documents the cecal landmarks in standard clinical workflow, Cecum AI is triggered, and the image is sent for analysis. If the cecal landmarks are confirmed in the image, by the AI, the user will be given a reminder to check the status of Polyp detection. If polyp detection is turned off, the cross icon will flash three times along with the cecum icon. If detection is on, the cecum icon will flash three times next to the tick icon. This is a convenience feature that provides a check to the user that the CADDIE polyp detection function is on and in use. | Same |
Page 4 of 14
{9}
K260133
Odin Vision
K260133 Page 5 of 14
| Characteristics | Subject Device: CADDIE (Software version 1.6.0) | Predicate Device: CADDIE K252586 (Software version 1.5.0) | Comparison |
| --- | --- | --- | --- |
| **Software Algorithms** | CADDIE utilizes an artificial intelligence-based algorithms to perform the polyp detection and cecum AI functions. | CADDIE utilizes an artificial intelligence-based algorithms to perform the polyp detection and cecum AI functions. | Same |
| **Device Output** | Polyp Detection: During a colonoscopy, CADDIE generates markers, which look like green squares and are accompanied by a short, low-volume sound, and superimposes them on the video from the endoscope camera when it identifies a potential lesion. Cecum AI: When the clinician photo documents cecum, Cecum AI is triggered. If cecal landmarks are confirmed in the image, by the AI, the user will be given a reminder to check the on/ off status of Polyp detection. | Polyp Detection: During a colonoscopy, CADDIE generates markers, which look like green squares and are accompanied by a short, low-volume sound, and superimposes them on the video from the endoscope camera when it identifies a potential lesion. Cecum AI: When the clinician photo documents cecum, Cecum AI is triggered. If cecal landmarks are confirmed in the image, by the AI, the user will be given a reminder to check the on/ off status of Polyp detection. | Same |
| **Additional compatible accessory** | Arc Endocuff Vision (K162205) | N/A | Different; compatibility does not affect device design or performance as supported by performance testing |
Page 5 of 14
{10}
K260133
K260133 Page 6 of 14
Odin Vision
## Discussion of Differences
The Indications for Use for both devices are identical; the subject device is an upgrade to the predicate device, without altering its Indications for Use.
Both the subject device and CADDIE (K252586) are identical in polyp detection.
Both devices take a colonoscopy video as an input from an endoscopy image processor and provide an output of a green bounding box that highlights the detected polyps.
Both devices are used in real-time to aid the clinician in detecting abnormal lesions live during a colonoscopy.
There is an introduction to a new compatible medical device, Arc Endocuff Vision (K162205), as an optional tool that can be used with CADDIE during procedures.
The subject and predicate device (CADDIE) shared identical indications for use, principles of operation as well as technological characteristics. The introduction of new compatible device has been assessed through identical standalone non-clinical performance testing, and the results demonstrated that the performance of the subject device is non inferior to that of the predicate device. As a result, the subject device can be considered substantially equivalent to the predicate.
## Performance Testing
This Special 510(k) addresses the introduction of new compatible medical device, Arc Endocuff Vision. Performance testing was conducted to evaluate the additional compatible device and is summarized below.
The addition of compatible device does not alter the safety or performance characteristics previously cleared under K252586. Changes to other aspects of the device that are not the subject of this submission have been evaluated through Odin's risk management process and FDA's guidance on deciding when to submit a 510(k) and determined not to require a new 510(k).
For all changes, design and development activities were performed in compliance with Odin's Quality Management System and FDA Design Control requirements (21 CFR 820.30), using the same methods and acceptance criteria described in the cleared submission (K252586). All performance characteristics continue to meet their established requirements and do not raise any additional questions associated with the safety and effectiveness of the device.
Special controls per 21 CFR 876.1520 have been addressed as follows:
Page 6 of 14
{11}
K260133
K260133 Page 7 of 14
Odin Vision
### Software controls
Special control (4) Software verification, validation, and hazard analysis must be provided. Software description must include a detailed, technical description including the impact of any software and hardware on the device's functions, the associated capabilities and limitations of each part, the associated inputs and outputs, mapping of the software architecture, and a description of the video signal pipeline.
Relevant documented evidence was generated through adherence to Odin's Quality Management System and FDA Design Control requirements and form part of the documentation associated with this 510(k).
### Human Factors and Usability Testing
Special Control (3) Usability assessment must demonstrate that the intended user(s) can safely and correctly use the device.
Human Factors and Usability Testing was conducted and reviewed in K240044 and is still applicable to the version of the device that is the subject of this 510(k).
### Non-clinical Performance Testing
Special Control (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, which include:
(i) Standalone algorithm performance testing;
(ii) Pixel-level comparison of degradation of image quality due to the device;
(iii) Assessment of video delay due to marker annotation; and
(iv) Assessment of real-time endoscopic video delay due to the device.
(2)(ii)-(iv) have not been impacted by the change subject of this 510(k); testing that was conducted and reviewed in previous 510(k)s for this device is still applicable to the version of the device that is the subject of this 510(k).
Standalone algorithm performance testing is summarized below:
### Polyp Detection Data Description & Standalone Performance testing
Polyp Detection Standalone Bench Testing Dataset
Standalone performance testing was performed to assess the ability of CADDIE to discriminate between normal mucosa and polyp tissue on video frames from a standard colonoscopy procedure. A set of recorded colonoscopy videos was analyzed by CADDIE and the results were compared to the historical control (known polyp status per frame).
| | Test Set (No. of subjects=632) | |
| --- | --- | --- |
| | Mean (years) | SD (years) |
| Age | 59.43 | 9.58 |
| | Count | % |
Page 7 of 14
{12}
K260133
K260133 Page 8 of 14
Odin Vision
| Sex | | |
| --- | --- | --- |
| Male | 339 | 53.64 |
| Female | 276 | 43.67 |
| Unknown | 17 | 2.69 |
| Indication for Colonoscopy | | |
| Surveillance | 178 | 28.16 |
| Screening | 337 | 53.32 |
| Other | 31 | 4.91 |
| Unknown | 86 | 13.61 |
| Race/Ethnicity | | |
| African American | 110 | 17.41 |
| American Indian | 1 | 0.180.16 |
| Asian | 23 | 3.64 |
| Caucasian | 225 | 35.60 |
| Hispanic | 65 | 10.28 |
| Persian | 20 | 3.16 |
| North American | 36 | 5.70 |
| African | 9 | 1.42 |
| Latin American | 3 | 0.47 |
| Black or Black British | 2 | 0.32 |
| Mixed | 2 | 0.32 |
| Caribbean | 1 | 0.16 |
| Other | 23 | 0.47 |
| Unknown | 132 | 20.89 |
| Hospital's country | | |
| United States of America | 222 | 35.13 |
| Italy | 73 | 11.55 |
| Spain | 69 | 10.92 |
| Poland | 36 | 5.7 |
| Germany | 54 | 8.54 |
| Iran | 20 | 3.16 |
| Japan | 54 | 8.54 |
| Canada | 70 | 11.08 |
| United Kingdom | 34 | 5.38 |
The following table presents the standalone performance testing patients' demographic distribution, including the reason for colonoscopy:
| Histopathology | Count |
| --- | --- |
| Adenoma | 646 |
| Non-Adenoma | 493 |
| Polyp Location | |
| Cecum | 115 |
| Ascending Colon | 183 |
| Hepatic Flexure | 29 |
| Transverse Colon | 203 |
Page 8 of 14
{13}
K260133
K260133 Page 9 of 14
Odin Vision
| Splenic Flexure | 7 |
| --- | --- |
| Descending Colon | 113 |
| Sigmoid Colon | 278 |
| Rectum | 203 |
| Unknown | 8 |
| Polyp Morphology | |
| --- | --- |
| Polypoid | 813 |
| Non-Polypoid | 293 |
| Unknown | 33 |
| Polyp Size | |
| Diminutive | 817 |
| Small | 238 |
| Large | 84 |
Annotation was performed on a per-frame basis, where a team of trained clinical annotators labelled polyp structures with a bounding box, followed by an additional layer of review by a separate team of experts (with over 2000 endoscopic procedures experience). Each polyp was histologically confirmed. These polyp annotations were used as ground truth reference standards. The table above shows the distribution of polyp characteristics.
### Standalone Performance Testing Results
The table below shows a subgroup analysis. For video processors and endoscope family subgroup analyses, only models for which CADDIE compatibility is claimed are included. Overall performance results and all other subgroup analyses include all procedures in the evaluation dataset. These results are computed at a persistence of > 0.5 seconds and with an IoU > 20% (when applicable). The reported metrics correspond to the polyp detection endpoints and are summarized as follows:
• Frame-level TPR: proportion of frames with polyps detected by the device
• Object-level TPR: proportion of polyps detected by the device
• Frame-level FPR: proportion of non-polyp frames where the device detects a polyp.
| | TPR/Frame[95% CI] | Object-level TPR[95% CI] | Detected Polyps | Procedures |
| --- | --- | --- | --- | --- |
| Overall | 61.44% [59.80%, 63.07%] | 97.63% [96.75%, 98.51%] | 1112/1139 | 492 |
| Histology | | | | |
| Adenoma | 67.42% [65.43%, 69.40%] | 98.76% [97.91%, 99.61%] | 638/646 | 341 |
| Non-Adenoma | 53.60% [51.03%, 56.18%] | 96.15% [94.45%, 97.85%] | 474/493 | 268 |
| Lesion Size | | | | |
| Diminutive (0-5mm) | 58.89% [56.92%, 60.85%] | 97.31% [96.20%, 98.42%] | 795/817 | 402 |
Page 9 of 14
{14}
K260133
K260133 Page 10 of 14
Odin Vision
| Small(6-10mm) | 67.65% [64.35%,70.95%] | 98.32% [96.69%,99.95%] | 234/238 | 168 |
| --- | --- | --- | --- | --- |
| Large(>10mm) | 68.66% [63.47%,73.84%] | 98.81% [96.49%,100.00%] | 83/84 | 68 |
| Accessory | | | | |
| Endocuff | 77.96% [74.15%,81.77%] | 97.65% [95.37%,99.93%] | 166/170 | 69 |
| Non-Endocuff | 58.54% [56.80%,60.28%] | 97.63% [96.67%,98.58%] | 946/969 | 423 |
| | FPR/FRAME [95% CI] | Procedures |
| --- | --- | --- |
| Overall | 1.72% [1.49%, 1.95%] | 140 |
| Accessory | | |
| Endocuff | 1.80% [1.22%, 2.37%] | 12 |
| Non-Endocuff | 1.72% [1.47%, 1.96%] | 128 |
### Summary
Comprehensive standalone performance testing was conducted on a large dataset which was more than two times larger than the predicate device. The dataset comprised \( >30\% \) US patients and contained demographically diverse patients, representative of the US population. The standalone performance testing demonstrated the generalizability of the CADe device across race and ethnicity with \( >16\% \) African American patients and 10% Hispanic/Latino patients.
The standalone performance testing demonstrated the efficacy of the device in detecting polyps of all sizes and morphologies across all locations in the colon in the intended population and comparison of the performance with and without using the additional compatible device, Arc Endocuff Vision. The results were compared to the predicate device's and were found to be similar or better. No claims other than equivalence are made. The evidence generated in the performance testing supports the substantial equivalence of the CADDIE device to the predicate.
Page 10 of 14
{15}
K260133
K260133 Page 11 of 14
Odin Vision
### Cecum AI Data Description & Standalone Performance Testing
In addition to the standalone performance testing for the polyp-detection component of CADDIE, a standalone performance test for the convenience feature - Cecum AI was conducted to assess the ability of CADDIE to identify the cecum on photo-documented frames from a standard colonoscopy procedure. A set of recorded colonoscopy frames were analyzed by CADDIE and the results were compared to the historical control (known cecum status per frame).
CADDIE's cecum AI is set up to only be used on photo-documentation static frames during a colonoscopy. To reflect this use case, this evaluation was performed on static images captured as photo-documentation during colonoscopies.
#### Data Description
The dataset used to develop the Cecum AI algorithm consisted of a diverse sample of patients and images. The summary breakdown of the development data used across the training, tuning, and testing phases are presented in the following table.
| Split | Patients | Images | Positive images | AO images | ICV images | Negative images |
| --- | --- | --- | --- | --- | --- | --- |
| Training | 2,475 | 1,127,724 | 374,816 | 180,405 | 204,589 | 752,908 |
| Tuning | 808 | 46,315 | 15,107 | 7,381 | 8,342 | 31,208 |
| Testing | 109 | 1,656 | 211 | 129 | 83 | 1,445 |
The dataset was generated by including all cecum images containing the appendiceal orifice (AO) and/or the ileocecal valve (ICV) as positive samples, and all other non-cecum photo-documented images as negative samples. The table below shows the distribution of cecal structure characteristics in the dataset used for the standalone performance evaluation. SD denotes standard deviation.
| | Procedures | Frames | Frames/procedure (Mean [SD]) | Total (#frames) |
| --- | --- | --- | --- | --- |
| Positive | 400 | 921 | 2.3 [1.45] | 5,224 |
| AO | 317 | 512 | 1.62 [0.95] | |
| ICV | 290 | 451 | 1.56 [1.12] | |
| Negative | 449 | 4,303 | 9.58 [8.63] | |
#### Study Endpoints
The study endpoints are the same as for the predicate device. The primary endpoints for Cecum AI are:
- Frame-Level TPR: The proportion of all the frames annotated with cecum which CADDIE characterizes correctly.
- Frame-Level FPR: The proportion of all frames without cecum landmarks which CADDIE incorrectly detects cecum.
Page 11 of 14
{16}
K260133
K260133 Page 12 of 14
Odin Vision
The following table shows presents a subgroup analysis of the device performance, grouped by cecal structure, video processor, endoscope family, and accessories. For video processor and endoscope family subgroup analyses, only models for which CADDIE compatibility is claimed are included. Overall performance results and all other subgroup analyses include all procedures in the evaluation dataset. Frame-level performance metrics are summarized as the mean and 95% confidence intervals computed per procedures/patients.
## Results/Comparison to Predicate
The performance testing indicates that the Cecum AI convenience feature performs as expected and is non-inferior to the performance of the predicate device. No claims other than equivalence are made.
| | Accuracy/ FRAME (%) | TPR/ FRAME (%) | FPR/ FRAME (%) | Procedures |
| --- | --- | --- | --- | --- |
| Overall | 91.72 [90.58, 92.87] | 89.65 [87.28, 92.03] | 7.53 [6.20, 8.87] | 464 |
| Structure | | | | |
| AO | 94.38 [93.36, 95.39] | 85.28 [81.75, 88.82] | 3.88 [3.00, 4.76] | 317 |
| ICV | 95.28 [94.48, 96.08] | 83.03 [79.10, 86.96] | 3.52 [2.70, 4.34] | 290 |
| Accessory | | | | |
| Endocuff | 92.24 [89.36, 95.12] | 92.22 [86.89, 97.55] | 5.03 [2.71, 7.34] | 45 |
| Non-Endocuff | 91.67 [90.44, 92.90] | 89.33 [86.74, 91.92] | 7.81 [6.35, 9.27] | 419 |
## Summary
The standalone performance testing demonstrated the efficacy of the device in detecting cecal landmarks in the colon in the intended population. The results were compared to the predicate devices and were found to be non-inferior when using the additional compatible device, Arc Endocuff Vision. No claims other than equivalence are made. The evidence generated in the performance testing supports the substantial equivalence of the Cecum AI convenience feature to the predicate.
## Summary of Clinical Performance
Special control (1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including detection of gastrointestinal lesions and evaluation of all adverse events.
The baseline Clinical Performance Evaluation was conducted and reviewed in K240044 and is still applicable to the version of the device that is the subject of this 510(k).
## Labeling
Special Control (6) Labeling must include:
(i) Instructions for use, including a detailed description of the device and compatibility information;
Page 12 of 14
{17}
K260133
K260133 Page 13 of 14
Odin Vision
- (ii) Warnings to avoid overreliance on the device, that the device is not intended to be used for diagnosis or characterization of lesions, and that the device does not replace clinical decision making;
- (iii) A summary of the clinical performance testing conducted with the device, including detailed definitions of the study endpoints and statistical confidence intervals; and
- (iv) A summary of the standalone performance testing and associated statistical analysis.
Device labeling created as part of this 510(k) satisfies the special control requirements.
Page 13 of 14
{18}
K260133
K260133 Page 14 of 14
Odin^{}[] Vision
## **Substantial Equivalence Conclusion**
The subject device is similar to the predicate device with the main difference being the addition of compatibility with a legally marketed accessory to endoscopes (Endocuff), which does not raise different questions of safety and effectiveness. The two CADDIE versions (subject and predicate) share the same intended use, indications for use and principles of operation. The main difference and subject of this 510(k) has been addressed through non-clinical performance testing, using methods, protocols, and acceptance criteria that supported the previously cleared version of CADDIE (K252586). Performance data demonstrate that the subject device is substantially equivalent to the predicate device.
Page 14 of 14
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.