The intended use of EPIQ Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: - Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. - Modes of operation include: B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging. - The clinical environments where EPIQ Series Diagnostic Ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. - When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Device Story
EPIQ Series Diagnostic Ultrasound System is a software-controlled diagnostic imaging platform. This submission adds the eV14-2v endocavitary transducer, a broadband PureWave curved linear array with multi-row technology (436 elements, 3 rows). The transducer uses a high-voltage multiplexer to switch between inner and full elevation apertures, optimizing near-field clutter and far-field penetration. It supports 2D, PW/Color Doppler, M-mode, 3D/4D, and Elastography. Used in clinics, hospitals, and point-of-care settings by trained healthcare professionals, the system acquires ultrasound data via the transducer, processes it through the console, and displays images for diagnostic assessment. The system integrates with Philips EchoNavigator for live X-ray/echo guidance during cardiovascular procedures. Output assists clinicians in visualizing anatomy and fluid flow to inform clinical decision-making and treatment guidance.
Clinical Evidence
No clinical data was required for this submission. Substantial equivalence was demonstrated through design features, indications for use, fundamental scientific technology, and non-clinical performance testing (bench testing).
Technological Characteristics
Broadband endocavitary curved linear array transducer (eV14-2v); 6 MHz center frequency; 436 elements (3 rows of 218); high-voltage multiplexer for aperture switching (inner vs. full); 3D/4D capable; biocompatible materials (ISO 10993-1); IEC 60601-1 and IEC 60601-2-37 compliant; cart-based system connectivity.
Indications for Use
Indicated for diagnostic ultrasound imaging and fluid flow analysis in patients requiring abdominal, cardiac (adult/pediatric/fetal), vascular, cephalic, obstetric, gynecological, musculoskeletal, ophthalmic, urological, small organ, transesophageal, transrectal, transvaginal, or lung examinations. Specific indications for the eV14-2v transducer include Fetal/OB, Fetal Echo, GYN, Urology, and Transvaginal imaging.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
3D9-3V with EPIQ Series Diagnostic Ultrasound System (K160807)
Reference Devices
V9-2 with EPIQ Series Diagnostic Ultrasound System (K182857)
C9-4ec with Lumify Diagnostic Ultrasound System (K242519)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 27, 2026
Philips Ultrasound, LLC
Aditi Chaubal
Senior RA Manager
22100 Bothell Everett Hwy
Bothell, Washington 98021
Re: K260123
Trade/Device Name: EPIQ Series Diagnostic Ultrasound System
Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX, OBJ, QIH
Dated: April 24, 2026
Received: April 27, 2026
Dear Aditi Chaubal:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260123 - Aditi Chaubal
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260123 - Aditi Chaubal
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**YANNA S. KANG -S**
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K260123
Device Name
EPIQ Series Diagnostic Ultrasound System
Indications for Use (Describe)
The intended use of EPIQ Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
- Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
- Modes of operation include: B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
- The clinical environments where EPIQ Series Diagnostic Ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
- When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed.
However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
Page 1 of 6
# TRADITIONAL 510(k)
Philips Ultrasound
EPIQ Series Diagnostic Ultrasound System
510(k) Summary
Section 10.02 510(k) Summary
510(k) Number: K260123
This summary of safety and effectiveness information is submitted in accordance with 21 CFR § 807.92.
1. Submitter's name, address, telephone number, contact person(s)
Manufacturer: Philips Ultrasound LLC
22100 Bothell Everett Hwy
Bothell, WA 98021-8431
Contact Person: Aditi Chaubal
Senior RA Manager
aditi.chaubal@philips.com
1-425-482-8243
Secondary Contact: Irma Sandoval-Watt
Senior RA Manager
Irma.Sandoval-Watt@philips.com
1-303-453-3421
Date Prepared: May 26, 2026
2. Name of the device, including the trade of proprietary name if applicable, the common or usual name, and the classification name, if known:
Proprietary Name: EPIQ Series Diagnostic Ultrasound System
Common Name: Diagnostic ultrasound system and transducers
Regulation Description:
| Classification Name | 21 CFR § | Product Code |
| --- | --- | --- |
| Primary | | |
| System, imaging, pulsed doppler, ultrasonic | 892.1550 | IYN |
| Secondary | | |
| System, imaging, pulsed echo, ultrasonic | 892.1560 | IYO |
| Transducer, ultrasonic, diagnostic | 892.1570 | ITX |
| Automated radiological image processing software | 892.2050 | QIH |
| Catheter, ultrasound, intravascular | 870.1200 | OBJ |
Device Class: Class II
Classification Panel: Radiology
{5}
Page 2 of 6
# 3. Device Description Summary
The Philips EPIQ Series Diagnostic Ultrasound Systems are general purpose, software controlled, diagnostic ultrasound systems. Their function is to acquire ultrasound data and to display the data in various modes of operation. The systems consist of two parts: the system console and the transducers. The system console contains the user interface, a display, system electronics, and the transducer connection points. In addition to the physical knobs and buttons of the main control panel, the user interface consists of a touch screen with soft key controls. The transducers connect to the system console and facilitate ultrasound scanning of patients.
There are multiple marketing-named versions of the EPIQ Series Diagnostic Ultrasound systems. Each version of the EPIQ systems can have multiple feature configurations that are based on different software feature options and hardware options.
This submission introduces the eV14-2v transducer on the Philips EPIQ Series Diagnostic Ultrasound Systems.
# eV14-2v Transducer
The eV14-2v is a broadband, endocavitary PureWave transducer with elevation focus. It is 3D/4D capable and features multi-row array technology. It is designed to support obstetric and gynecological applications. The transducer offers a variety of imaging modes, including 2D, PW Doppler, Color Flow, M-mode, 3D/4D, and Elastography.
eV14-2v includes a curved linear array with 6 MHz center frequency and 10.5 mm azimuth radius of curvature. The array has a total of 436 elements arranged in 3 rows of 218 elements. The inner row of 218 elements provide an "inner" elevation aperture 3 mm in height. This row is connected to 128 system channels via a high-voltage multiplexer. The two outer rows of 218 elements are electrically connected together and, also via a high-voltage multiplexer, can be selectively connected to the same 128 system channels, resulting in the "full" aperture height of 6 mm. The ability to switch between inner and full apertures provides the means to achieve both low clutter in the near field and penetration in the far field.
The same multiplexer provides the ability to translate these apertures in the azimuth direction provides the field of view. The multiplexer is controlled by data signals provided by the system to coordinate azimuth aperture position with the system beamformer. The linear curved array can be mechanically rotated and acquire planar tomographic slices for reconstructing 3D/4D images.
The eV14-2v consists of a linear curved array, rotational mechanism and drive train, motor, and interconnect circuitry. A transducer housing assembly encloses the lens module assembly (LMA) and an integral connectorized cable assembly provides interface to the system. The connectorized cable assembly contains a cable and a connector with PCBs providing cable and system interconnect, electrical tuning, multiplexer control, and system interface features.
# 4. Intended Use and Indications for Use
The intended use of EPIQ Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
- Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
- Modes of operation include: B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
- The clinical environments where EPIQ Series Diagnostic Ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
- When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures
{6}
and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
## Indications for Use for eV14-2v Transducer:
The indications for use for the eV14-2v transducer are Fetal/OB, Other: Fetal Echo, Other: GYN, Other: Urology, and Transvaginal.
## 5. Substantially Equivalent Devices
- Predicate Device (Transducer): 3D9-3V with EPIQ Series Diagnostic Ultrasound System (K160807)
- Reference Device 1 (Transducer): V9-2 with EPIQ Series Diagnostic Ultrasound System (K182857)
- Reference Device 2 (Transducer): C9-4ec with Lumify Diagnostic Ultrasound System (K242519)
## 6. Technological Comparison to Predicate & Reference Devices
| Feature/ Characteristic | EPIQ Series Diagnostic Ultrasound System with eV14-2v Transducer Subject | EPIQ Series Diagnostic Ultrasound System with 3D9-3V Transducer (K160807) Predicate | Comparison/ Discussion/ Comments |
| --- | --- | --- | --- |
| System level Feature/Characteristics | | | |
| Regulation Number | 892.155 | 892.155 | Identical |
| Device Classification Name | II | II | Identical |
| Primary Product Code | IYN | IYN | Identical |
| Secondary Product Codes | IYO, ITX, OBJ, QIH | IYO, ITX, OBJ, QIH | Identical |
| Intended Use | Intended for diagnostic ultrasound imaging and fluid flow analysis | Intended for diagnostic ultrasound imaging and fluid flow analysis | Identical |
| Indications for Use | Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. | Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. | Identical |
| Modes of Operation | B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, Harmonic Imaging | B Mode (3D/4D), M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, Harmonic Imaging | Identical |
| Scientific Technology | Ultrasound Imaging | Ultrasound Imaging | Identical |
| Users | Trained healthcare professionals | Trained healthcare professionals | Identical |
Page 3 of 6
{7}
| Feature/ Characteristic | EPIQ Series Diagnostic Ultrasound System with eV14-2v Transducer Subject | EPIQ Series Diagnostic Ultrasound System with 3D9-3V Transducer (K160807) Predicate | Comparison/Discussion/Comments |
| --- | --- | --- | --- |
| | | | |
| Use Environment | Clinics, hospitals, and clinical point-of-care for diagnosis of patients. | Clinics, hospitals, and clinical point-of-care for diagnosis of patients. | Identical |
| Transducer Features | eV14-2v | 3D9-3V | |
| Transducer Model | Endocavitary Curved Linear Transducer | Endocavitary Curved Linear Transducer | Identical |
| Indications for Use | Fetal/OB; Fetal Echo; Trans-vaginal; GYN; Urology | Fetal/OB; Fetal Echo; Trans-vaginal; GYN; Urology | Identical |
| Acoustic Output | Ispta.3 ≤ 720 (mW/cm2) TI ≤ 6.0 MI ≤ 1.9 | Ispta.3 ≤ 720 (mW/cm2) TI ≤ 6.0 MI ≤ 1.9 | Identical |
| Acoustic Working Frequency | Center Frequency 6.1MHz | Center Frequency 5.7 MHz | Similar |
| Geometrical Configuration | Curved Linear Array | Curved Linear Array | Identical |
| Total Number of elements | 436 addressable elements connected to form dual aperture array with 218 elements per aperture | 128 | Different |
| Field of View (Degrees) | 180 degrees | 155.5 degrees | Different |
| Transducer Modes of Operation | B (2D); M-mode; PW; Doppler; 3D/4D, Color; MPR | B, M, PWD, Color; 3D/4D; Doppler, Combined | Similar |
| Tip Dimensions | Height: 25.3mm Width: 25.3mm Tip Diameter: 25.3 mm | Height: 26.3mm Width:26.3mm Tip Diameter: 26.3 mm | Different |
| Connector | Cart-based system transducer connector | Cart-based system transducer connector | Identical |
| Weight (Including connector and cable) | < 0.9 kg (including connector and cable) | < 1.2 kg (including connector and cable) | Similar |
| Radiated Field Operating Controls | The transducer is controlled via the Ultrasound System's control panel. | The transducer is controlled via the Ultrasound System's control panel. | Identical |
| Patient Contact Materials | Biocompatible ISO 10993-1 | Biocompatible ISO 10993-1 | Identical |
Page 4 of 6
{8}
| Feature/ Characteristic | EPIQ Series Diagnostic Ultrasound System with eV14-2v Transducer Subject | EPIQ Series Diagnostic Ultrasound System with 3D9-3V Transducer (K160807) Predicate | Comparison/ Discussion/ Comments |
| --- | --- | --- | --- |
| | | | |
| Transducer Maximum Surface Temperature (per IEC 60601-2-37, clause 201.11) | Temperatures in °C Still Air: 42.6 Simulated Use: 42.5 | Temperatures in °C Still Air: 41.8 Simulated Use: 42.5 | Similar |
| Transducer Element Check Incorporated | Yes | Yes | Identical |
| Transducer Provided Sterile | No | No | Identical |
| Transducer Recommended use of covers | Yes | Yes | Identical |
| Transducer reusable | Yes | Yes | Identical |
| Maximum Insertion Portion Width | 25.3 mm | 27.3mm | Similar |
| Working Length (Intended Insertion Length) | 150mm | 165mm | Similar |
## 7. Safety Considerations
The proposed eV14-2v transducer with EPIQ Series Diagnostic Ultrasound System are categorized as Track 3 devices.
## 8. Non-Clinical Performance Data
Philips Ultrasound performed the following testing to ensure the safety and effectiveness of the proposed product:
- IEC 62304 Medical device software - Software life cycle processes, 2006 + A 2015
- IEC62366-1 Medical devices – Part 1: Application of usability engineering to medical devices, 2015+ AMD1: 2020
- ISO 14971 Medical devices- Application of risk management to medical devices, 2019
- IEC 60601-1 - Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
- IEC 60601-2-37- Medical electrical equipment – Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- IEC 60601-1-2 4.1 ed - Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral Standard: Electromagnetic disturbances – Requirements and tests; 2020
- IEC 62359, Ultrasonics – Field characterization – Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields, 2017-09
- ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process; 2018
Non-Clinical performance testing has been conducted addressing the system level requirements with the transducer according to the system and design specifications and risk control measures. The activities to assure safe and effective performance of the EPIQ Series Diagnostic Ultrasound System include but are not limited to the following:
- Requirements Review
- Risk Analysis and Management
- Product Specifications
- Design Reviews
Page 5 of 6
{9}
Page 6 of 6
9. Clinical Performance Data
The proposed EPIQ Series Diagnostic Ultrasound System with eV14-2v transducer did not require clinical data for determination of substantial equivalence since it demonstrated substantial equivalence based on the following attributes:
- Design Features
- Indications for Use
- Fundamental scientific technology
- Non-clinical performance testing
- Safety and Effectiveness
10. Conclusion
For testing, all pre-determined acceptance criteria were met. Results of these tests show that the proposed transducer eV14-2v for EPIQ Series Diagnostic Ultrasound System, does not affect the use of the device, nor does it introduce any new or significantly modified risks.
There is no change in the indications to EPIQ Series Diagnostic Ultrasound System. All comparison points between the subject eV14-2v transducer and the predicate, 3D9-3v transducer (compatible with the predicate EPIQ Series Diagnostic Ultrasound System) and reference devices (V9-2 with EPIQ Series Diagnostic Ultrasound System; C9-4ec with Lumify Diagnostic Ultrasound System) are either identical or equivalent. The differences between the proposed device and the predicate device do not raise new questions of safety and/or effectiveness. Therefore, the proposed EPIQ Series Diagnostic Ultrasound System with eV14-2v transducer is substantially equivalent to the predicate device in terms of indications for use, design, technological characteristics, modes of operations, safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.