SpaceOAR Vue System (SV-2101)

K260119 · Boston Scientific Corporation · OVB · Feb 10, 2026 · Radiology

Device Facts

Record IDK260119
Device NameSpaceOAR Vue System (SV-2101)
ApplicantBoston Scientific Corporation
Product CodeOVB · Radiology
Decision DateFeb 10, 2026
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

SpaceOAR Vue System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space it is the intent of SpaceOAR Vue System to reduce the radiation dose delivered to the anterior rectum. The SpaceOAR Vue System is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient’s body over time.

Device Story

System consists of single-use delivery mechanism and precursors for absorbable hydrogel spacer; used by clinicians to inject hydrogel between anterior rectal wall and prostate. Precursors (trilysine buffer and PEG-SG-TIB) mix in Y-connector to initiate cross-linking; reaction produces no measurable heat. Hydrogel acts as physical barrier to displace rectal wall during radiotherapy; degrades via hydrolysis and excreted via renal filtration. Device maintains space for duration of treatment course. Benefits include reduced radiation exposure to anterior rectum, potentially minimizing side effects. No software or algorithm involved.

Clinical Evidence

No clinical data. Substantial equivalence supported by bench verification and validation testing, including hydrogel swell, gelation time, persistence, injection force, and usability testing.

Technological Characteristics

Absorbable hydrogel spacer composed of polyethylene glycol (PEG), succinimidyl glutarate, triiodobenzoate, and trilysine. System includes delivery mechanism with Y-connector. Principle of operation: in-situ cross-linking of hydrogel precursors upon injection. Biodegradable via hydrolysis; renal excretion. No energy source. Mechanical device.

Indications for Use

Indicated for prostate cancer patients undergoing radiotherapy to temporarily position the anterior rectal wall away from the prostate to reduce radiation dose to the rectum.

Regulatory Classification

Identification

An absorbable perirectal spacer is composed of biodegradable material that temporarily positions the anterior rectal wall away from the prostate during radiotherapy for prostate cancer with the intent to reduce the radiation dose delivered to the anterior rectum. The absorbable spacer maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time.

Special Controls

An absorbable perirectal spacer must comply with the following special controls: (1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing. For all clinical investigations used to support premarket notification submissions for this type of device, line listings of the study data must be provided. Performance bench testing must demonstrate appropriate perirectal space (i) creation and maintenance for the duration of prostate radiotherapy; Performance bench testing must demonstrate that therapeutic radiation levels (ii) do not alter the performance of the device; (iii) Performance in vivo testing must demonstrate appropriate deployment of spacer as indicated in the accompanying labeling, and demonstrate appropriate expansion and absorption characteristics in a clinically relevant environment; (iv) Clinical study must demonstrate appropriate spacer stability and lack of migration for the entire course of radiotherapy, complete absorption, and lack of long term toxicity: Sterility testing must demonstrate the sterility of the device and the effects of (v) the sterilization process on the physical characteristics of the spacer; (vi) Shelf-life testing must demonstrate the stability of the physical characteristics of the spacer throughout the shelf-life as indicated in the accompanying labeling: and. (vii) The device must be demonstrated to be biocompatible. (2) The risk management activities performed as part of the manufacturer's 21 CFR 820.30 design controls must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique. (3) The device labeling must include the following: A detailed summary of reported or observed complications related to the use (i) of the device; (ii) Appropriate warnings; Detailed instructions for system preparations and detailed implant procedure (iii) instructions; and, (iv) An expiration date that is supported by performance data as specified in subparagraph (b)(i)(vi).

*Classification.* Class II (special controls). The special controls for this device are:(1) The premarket notification submission must include methodology and results of the following non-clinical and clinical performance testing. For all clinical investigations used to support premarket notification submissions for this type of device, line listings of the study data must be provided. (i) Performance bench testing must demonstrate appropriate perirectal space creation and maintenance for the duration of prostate radiotherapy. (ii) Performance bench testing must demonstrate that therapeutic radiation levels do not alter the performance of the device. (iii) Performance in vivo testing must demonstrate appropriate deployment of spacer as indicated in the accompanying labeling, and demonstrate appropriate expansion and absorption characteristics in a clinically relevant environment. (iv) Clinical study must demonstrate appropriate spacer stability and lack of migration for the entire course of radiotherapy, complete absorption, and lack of long term toxicity. (v) Sterility testing must demonstrate the sterility of the device and the effects of the sterilization process on the physical characteristics of the spacer. (vi) Shelf-life testing must demonstrate the stability of the physical characteristics of the spacer throughout the shelf-life as indicated in the accompanying labeling. (vii) The device must be demonstrated to be biocompatible. (2) The risk management activities performed as part of the manufacturer's § 820.10(c) of this chapter design and development activities must document an appropriate end user initial training program which will be offered as part of efforts to mitigate the risk of failure to correctly operate the device, including, but not limited to, documentation of an appropriate end user initial training program on the proper spacer deployment technique. (3) The device labeling must include the following: (i) A detailed summary of reported or observed complications related to the use of the device; (ii) Appropriate warnings; (iii) Detailed instructions for system preparations and detailed implant procedure instructions; and (iv) An expiration date that is supported by performance data as specified in paragraph (b)(1)(vi) of this section.

Predicate Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION February 10, 2026 Boston Scientific Corporation Cassidy Yeomans Regulatory Affairs Specialist 300 Boston Scientific Way Marlborough, Massachusetts 01752 Re: K260119 Trade/Device Name: SpaceOAR Vue System (SV-2101) Regulation Number: 21 CFR 892.5725 Regulation Name: Absorbable Perirectal Spacer Regulatory Class: Class II Product Code: OVB Dated: January 14, 2026 Received: January 14, 2026 Dear Cassidy Yeomans: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260119 - Cassidy Yeomans Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K260119 - Cassidy Yeomans Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ![img-0.jpeg](img-0.jpeg) Lora D. Weidner, Ph.D. Assistant Director Radiation Therapy Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260119 | ? | | Please provide the device trade name(s). | | ? | | SpaceOAR Vue System (SV-2101) | | | | Please provide your Indications for Use below. | | ? | | SpaceOAR Vue System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space it is the intent of SpaceOAR Vue System to reduce the radiation dose delivered to the anterior rectum. The SpaceOAR Vue System is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient’s body over time. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} K260119 # 510(k) Summary for SpaceOAR Vue System ## A Submitter Boston Scientific Corporation Urology and Women’s Health Division 100 Boston Scientific Way Marlborough, MA 01752 ## B Contact Cassidy Yeomans Regulatory Affairs Specialist 508-382-9137 Cassidy.Yeomans@bsci.com Alternate Contact: Julie Boylston Sr. Regulatory Affairs Manager 508-683-5927 Julie.Boylston@bsci.com ## C Subject Device Trade Name: SpaceOAR Vue System (SV-2101) Device Type: Absorbable Perirectal Spacer Device Classification Name: Hydrogel Spacer Device Classification: Class II per 21 CFR §892.5725 Product Code: OVB Review Panel: Radiology Identification of Subject Device: SpaceOAR Vue System, K260119 ## D Predicate Device Trade Name: SpaceOAR Vue System (SV-2101) Device Type: Absorbable Perirectal Spacer Device Classification Name: Hydrogel Spacer Device Classification: Class II per 21 CFR §892.5725 Product Code: OVB 510(k) Submitter/Holder: Boston Scientific Corporation Identification of Predicate Device: SpaceOAR Vue System, K182971 SpaceOAR Vue System Special 510(k) Page 1 of 3 {5} SpaceOAR Vue System Special 510(k) Page 2 of 3 # E Device Description The SpaceOAR Vue System consists of components for the preparation of an absorbable hydrogel spacer and a delivery mechanism packaged in a single use system. Upon injection, between the anterior rectal wall and prostate, the precursor (trilysine buffer solution with PEG-SG-TIB) and accelerator (a buffered solution) mix in the device Y-connector where the formation of SpaceOAR Vue hydrogel is initiated. The SpaceOAR Vue hydrogel is formed through cross-linking of the polyethylene glycol with succinimidyl glutarate and triiodobenzoate (PEG-SG-TIB) and Trilysine. This reaction gives off no measurable heat. The hydrogel temporarily positions the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space it is the intent of SpaceOAR Vue System to reduce the radiation dose delivered to the anterior rectum. The hydrogel is biodegradable and maintains space for the entire course of prostate radiotherapy treatment. The hydrogel degrades via hydrolysis and is excreted via renal filtration. # F Indications for Use SpaceOAR Vue System is intended to temporarily position the anterior rectal wall away from the prostate during radiotherapy for prostate cancer and in creating this space it is the intent of SpaceOAR Vue System to reduce the radiation dose delivered to the anterior rectum. The SpaceOAR Vue System is composed of biodegradable material and maintains space for the entire course of prostate radiotherapy treatment and is completely absorbed by the patient's body over time. The Indications for Use Statement for the proposed SpaceOAR Vue System is identical to the predicate device. # G Technological Characteristics Compared to Predicate The principles of operation and underlying technological characteristics of the proposed SpaceOAR Vue System and the predicate device are identical. Both devices are systems that facilitate implantation of 10cc PEG biodegradable hydrogel between the anterior rectal wall and prostate prior to radiotherapy. This temporarily creates space between the anterior rectal wall and prostate during radiotherapy. The PEG hydrogel is broken down through hydrolysis and excreted from the body through renal filtration. The differences between the subject device and the predicate are minor. Differences include changes to: - The nominal pH of the accelerator solution - The inner diameter of the hypotubes within the Y-connector - The luer ring color of the Y-connector - The shelf carton color These differences have been assessed and do not raise any new questions of safety or effectiveness. This evaluation demonstrates that the SpaceOAR Vue System is substantially equivalent to the predicate device. {6} SpaceOAR Vue System Special 510(k) Page 3 of 3 # H Performance Data Applicable verification and validation testing was performed as required by the Boston Scientific design control process. The following performance data evaluated over the specified shelf life were provided in support of the substantial equivalence determination: - Percent swell of hydrogel - Hydrogel gelation time - Hydrogel persistence - Maximum force required to inject hydrogel - Hydrogel performance after 60 minutes post precursor solution preparation - Usability Testing The predicate and subject device are predominantly identical in device design with limited changes to device components. A risk-based biocompatibility assessment of these limited changes indicates that no new biocompatibility testing is required to assess the proposed device change. The proposed SpaceOAR Vue System is biologically safe for its intended use. The principles of operation and intended use are identical for both the predicate and proposed device. Verification and usability testing of the proposed device were completed and confirmed that the proposed SpaceOAR Vue System meets the same functional and performance specifications as the predicate SpaceOAR Vue System. The clinical workflow for device implantation is identical for both the predicate and proposed device. Therefore, clinical data were not required to support a determination of substantial equivalence. The conclusion of the assessments demonstrates that the modified device continues to function as intended in a manner equivalent to the predicate device. The modified device raises no new issues of safety or effectiveness compared to the predicate. # I Conclusion Based on the performance data, the same principles of operation the same intended use, and same indications for use, the modified SpaceOAR Vue System is substantially equivalent to its predicate device, K182971.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...