K260114 · CIVCO Medical Instruments Co., Inc. · ITX · Sep 24, 2026 · Radiology
Device Facts
Record ID
K260114
Device Name
Align® Cover and Pad Guidance Tool
Applicant
CIVCO Medical Instruments Co., Inc.
Product Code
ITX · Radiology
Decision Date
Sep 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1570
Device Class
Class 2
Indications for Use
The medical device is intended to provide a guide for interventional devices while also serving as a microbial barrier between the patient and the ultrasound transducer. - Abdominal - Diagnostic imaging and minimally invasive puncture procedures. - Small Parts - Diagnostic imaging and minimally invasive puncture procedures. - Regional Anesthesia - Diagnostic imaging and minimally invasive puncture procedures. - Vascular Access - Diagnostic imaging and minimally invasive puncture procedures.
Device Story
Align® Cover and Pad Guidance Tool functions as a sterile, single-use accessory for ultrasound transducers. Device consists of a probe cover, a pad, and an integrated instrument guide. During interventional procedures (e.g., biopsies, vascular access), clinician places cover/pad over transducer; guide positions interventional instrument within ultrasound beam's field of view. Device provides microbial barrier between patient and transducer; pad facilitates acoustic coupling. Used in clinical settings by physicians or trained healthcare professionals. Output is physical guidance for needle/instrument placement, enabling real-time visualization of interventional path. Benefits include improved procedural accuracy and maintenance of sterile field.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including biocompatibility (ISO 10993-1), microbial barrier performance (ASTM F1671-13), and simulated usability testing.
Technological Characteristics
Materials: Polycarbonate, silicone elastomer, polyether polyurethane, polypropylene (guide), and specialized coating (Hydromer or Aculon). Meets ASTM F1671-13 for microbial barrier and ISO 10993-1 for biocompatibility. Mechanical features include instrument holder retaining tab with spring force functionality. Sterile, single-use, non-electronic, passive mechanical guidance system.
Indications for Use
Indicated for patients undergoing abdominal, small parts, regional anesthesia, or vascular access procedures requiring diagnostic ultrasound imaging and minimally invasive puncture. Contraindicated for use outside of these clinical applications.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Ultrasound General Purpose Guidance System (K093713)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 24, 2026
CIVCO Medical Instruments Co., Inc.
James Leong
Regulatory Affairs
750 Lois Lane
Coralville, IA 52241
Re: K260114
Trade/Device Name: Align® Cover and Pad Guidance Tool
Regulation Number: 21 CFR 892.1570
Regulation Name: Diagnostic Ultrasonic Transducer
Regulatory Class: Class II
Product Code: ITX
Dated: August 26, 2026
Received: August 26, 2026
Dear James Leong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Before making any change significantly affecting the safety or effectiveness of the device, you must submit a new premarket notification in accordance with 21 CFR 807.81. Additional information about changes that
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260114 - James Leong
Page 2
may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260114 - James Leong
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**MARJAN NABILI -S**$^{for}$
Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological
Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260114
Please provide the device trade name(s).
Align® Cover and Pad Guidance Tool
Please provide your Indications for Use below.
The medical device is intended to provide a guide for interventional devices while also serving as a microbial barrier between the patient and the ultrasound transducer.
- Abdominal - Diagnostic imaging and minimally invasive puncture procedures.
- Small Parts - Diagnostic imaging and minimally invasive puncture procedures.
- Regional Anesthesia - Diagnostic imaging and minimally invasive puncture procedures.
- Vascular Access - Diagnostic imaging and minimally invasive puncture procedures.
Please select the types of uses (select one or both, as applicable).
☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
?
{4}
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being prepared in accordance with the requirements of 21 CFR 807.92.
The assigned 510(k) number is: K260114
1. Submitter's Identifications:
| Establishment: | CIVCO Medical Instruments Co., Inc. |
| --- | --- |
| Address: | 750 Lois Lane Coralville, IA 52241 |
| Registration Number: | 1937223 |
| Operations Manufacturer | CIVCO Medical Instruments Co., Inc. |
| Owner/Operator: | |
| Owner/Operator Number: | 1937223 |
| Contact Person: | Jim Leong |
| Phone: | 319-248-6502 |
| e-mail: | James.Leong@civco.com |
2. Date 510(k) Summary Prepared: January 12, 2026
3. Name of the Subject Device and Classification Information:
| Trade/Device Name | Align® Cover and Pad Guidance Tool |
| --- | --- |
| Regulation Number | 21 CFR 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
4. Information for the Primary Predicate Device:
| Tradename/Device Name | Envision™ Ultrasound Cover and Pad |
| --- | --- |
| Manufacturer | CIVCO Medical Instruments Co., Inc. |
| 510(k) Number | K190802 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
1
{5}
# Information for the Secondary Predicate Device:
| Tradename/Device Name | Ultrasound General Purpose Guidance System |
| --- | --- |
| Manufacturer | CIVCO Medical Instruments Co., Inc. |
| 510(k) Number | K093713 |
| Regulation Number | 892.1570 |
| Classification Name | Diagnostic ultrasonic transducer |
| Regulatory Class | Class II |
| Product Code | ITX |
# 5. Device Description:
The Align® Cover and Pad Guidance Tool offers a unique approach to ultrasound-guided procedures by providing a guide that positions the interventional instrument in the field of view of the ultrasound beam, as well as a microbial barrier and ultrasound conductive medium between the patient and the ultrasound transducer.
# 6. Intended Use / Indications for Use:
The medical device is intended to provide a guide for interventional devices while also serving as a microbial barrier between the patient and the ultrasound transducer.
- Abdominal - Diagnostic imaging and minimally invasive puncture procedures.
- Small Parts - Diagnostic imaging and minimally invasive puncture procedures.
- Regional Anesthesia - Diagnostic imaging and minimally invasive puncture procedures
- Vascular Access - Diagnostic imaging and minimally invasive puncture procedures
# 7. Comparison to Legally Marketed Device
| | Subject Device Align® Cover and Pad Guidance Tool | Primary Predicate Envision™ Ultrasound Cover and Pad-K190802 | Secondary Predicate Ultrasound General Purpose Guidance System (Infiniti)-K093713 |
| --- | --- | --- | --- |
| Materials | | | |
| • Pad | Polycarbonate, silicone elastomer, polyether polyurethane, coating formulation (Hydromer or Aculon) No-Tox Liquid Ink FGN-1636 NT21 Black | Polycarbonate, silicone elastomer, polyether polyurethane, coating formulation (Hydromer or Aculon) | N/A |
| • Cover | Polyurethane (and those in the pad) | Polyurethane (and those in the pad) | N/A |
2
{6}
| | Subject Device Align® Cover and Pad Guidance Tool | Primary Predicate Envision™ Ultrasound Cover and Pad-K190802 | Secondary Predicate Ultrasound General Purpose Guidance System (Infiniti)-K093713 |
| --- | --- | --- | --- |
| • Guide | Polypropylene Clariant Colorant OC5SAF (Pale Blue) | N/A | ABS |
| Microbial Barrier | Meets requirements of ASTM F1671-13 for prevention of blood-borne pathogens | Meets requirements of ASTM F1671-13 for prevention of blood-borne pathogens | N/A |
| Biocompatibility | Meets ISO 10993-1 Biocompatibility requirements for surface devices with limited (≤ 24 hours) contact with breached or compromised surface | Meets ISO 10993-1 biocompatibility requirements for limited (≤ 24 hours) contact duration: • surface devices in contact with breached or compromised surface • external communicating device with indirect contact with blood path/tissue | Meets ISO 10993-1 Biocompatibility requirements for limited (≤ 24 hours) contact duration for external communicating devices in contact with tissue/bone/dentin |
| Shelf Life | 15 Months | 3 years | 3 years |
| Included Accessories | Sterile Saline Wipe Bands with cover configurations | Sterile Saline Wipe Bands with cover configurations | Probe Cover with gel and bands |
### 8. Comparison of Indications to the Legally Marketed Device:
The proposed Align device has the same basic intended use as the predicate, which is to help direct interventional instruments into an anatomical structure using an ultrasound transducer. CIVCO has further combined the indications for use for the Envision Cover and Pad along with the Infiniti Needle Guide into bullet point format.
- Abdominal - Diagnostic imaging and minimally invasive puncture procedures.
- Small Parts - Diagnostic imaging and minimally invasive puncture procedures.
- Regional Anesthesia - Diagnostic imaging and minimally invasive puncture procedures
- Vascular Access - Diagnostic imaging and minimally invasive puncture procedures
### 9. Summary of Non-Clinical Tests Performed:
- Biocompatibility:
The Align® Cover and Pad Guidance Tool are manufactured with materials which meet ISO 10993-1 biocompatibility requirements for limited contact duration for surface devices of breached or compromised surface. These same materials are already utilized on CIVCO cleared medical devices for contact in the same regions.
3
{7}
• Design specification:
The Align® Cover and Pad Guidance Tool have certain design functionality which was tested internally to verify they meet the established requirements. These criteria are listed as follows:
○ Visual Inspection testing - This test confirms that the device is assembled correctly, there is no exposed adhesive.
○ Dimensional testing - This test confirms that the guide is properly attached to the cover and pad.
○ Adhesive Force Test - This test confirms that the device will remain attached with a minimum amount of force.
○ Spring Force and Deflection Test - This test confirms that the force to deflect the instrument holder retaining tab can deflect with a minimum amount of force and return within 10 degrees of the original position.
○ Functional Scanning Test - This test confirms the pad and cover are compatible with ultrasound imaging equipment.
○ Cover Leak Test – This test ensures manufacturing produces covers without holes.
○ Pin Slip Test - This test confirms that the device can be utilized with instruments up to 18GA in size.
○ Rehydration Test – This test ensures the coating on the Align Covers and Pads is capable of being rehydrated multiple time during a procedure.
○ Tensile Strength Test – This testing confirmed that the device demonstrates the same material tensile strength as the predicate device.
○ Tear Resistance Test - This testing confirmed that the device demonstrates the same material tear resistance as the predicate device.
○ Air Burst Pressure and Volume Testing – This test confirms mechanical integrity of the covers.
• Simulated Usability Testing
Simulated use evaluations were performed by customers to ensure the design of the Align device conforms to the user needs and intended use.
10. Clinical Test Performed:
Clinical tests were not required to demonstrate substantial equivalence.
11. Conclusions:
Given that the subject and predicate devices have the same intended use and their technological characteristics achieve the same goal even though the designs are different, there is not a different question of safety and effectiveness. Therefore, the subject device is substantially equivalent to the primary predicate device.
4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.