Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System
Applicant
Stryker Corporation (Tornier, Inc.)
Product Code
PHX · Orthopedic
Decision Date
Jun 12, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) • Correction of functional deformity • Fractures of the humeral head • Traumatic arthritis • Revision of the devices if sufficient bone stock remains Notes: • All components are single use. • The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation. • All compatible humeral components are available for cemented or non-cemented fixation.
Device Story
Modular reverse shoulder arthroplasty system; replaces shoulder joint to reduce pain and restore mobility. Composed of glenoid component assembly and humeral component assembly. Glenoid sphere anchored to bone via screws (non-cemented); humeral components allow cemented or non-cemented fixation. Used by surgeons in clinical settings. System modifications include additional baseplate-glenosphere compatibility configurations and additive manufacturing updates. Device functions as a mechanical prosthesis; no software or algorithmic components.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by nonclinical bench testing, including range of motion analysis (ASTM F1378), glenoid loosening testing (ASTM F2028), fatigue testing, and biocompatibility evaluation (ISO 10993-1).
Technological Characteristics
Modular reverse shoulder prosthesis. Materials and design principles consistent with predicate. Fixation: non-cemented glenoid (screws), cemented/non-cemented humeral. Manufacturing: includes additive manufacturing components. Testing standards: ASTM F1378, ASTM F2028, ISO 10993-1. Sterilization: single-use components.
Indications for Use
Indicated for adults with functional deltoid muscle and massive, non-repairable rotator cuff tears causing pain due to rheumatoid arthritis, non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), functional deformity, humeral head fractures, traumatic arthritis, or revision surgery with sufficient bone stock.
Regulatory Classification
Identification
A shoulder joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a shoulder joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a humeral resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a glenoid resurfacing component made of ultra-high molecular weight polyethylene, and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Special Controls
*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” and
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-aluminum 4-vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-chromium-molybdenum casting alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-chromium-molybdenum alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra-high Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” and
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(v) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vi) F 1147-95 “Test Method for Tension Testing of Porous Metal,”
(vii) F 1378-97 “Standard Specification for Shoulder Prosthesis,” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”
Predicate Devices
Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System (K161742)
{0}
June 12, 2026
Stryker Corporation (Tornier, Inc.)
Jang Long
Staff Specialist, Regulatory Affairs
10801 Nesbitt Ave. S.
Bloomington, Minnesota 55437
Re: K260102
Trade/Device Name: Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System
Regulation Number: 21 CFR 888.3660
Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis
Regulatory Class: Class II
Product Code: PHX
Dated: April 15, 2026
Received: April 15, 2026
Dear Jang Long:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260102 - Jang Long
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260102 - Jang Long
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Farzana Sharmin, PhD
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260102 | ? |
| --- | --- | --- |
| Please provide the device trade name(s). | | ? |
| Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System | | |
| Please provide your Indications for Use below. | | ? |
| Indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by: • Rheumatoid arthritis • Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis) • Correction of functional deformity • Fractures of the humeral head • Traumatic arthritis • Revision of the devices if sufficient bone stock remains Notes: • All components are single use. • The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation. • All compatible humeral components are available for cemented or non-cemented fixation. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
{4}
K260102 (Page 1 of 2)
510(k) #:
# 510(k) Summary
Prepared on: 2026-06-12
## Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | Stryker Corporation (Tornier, Inc.) |
| --- | --- |
| Applicant Address | 10801 Nesbitt Avenue South Bloomington MN 55437 United States |
| Applicant Contact Telephone | 612-258-5639 |
| Applicant Contact | Ms. Jang Long |
| Applicant Contact Email | jang.long@stryker.com |
## Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System |
| --- | --- |
| Common Name | Shoulder joint metal/polymer semi-constrained cemented prosthesis |
| Classification Name | Shoulder Prosthesis, Reverse Configuration |
| Regulation Number | 888.3660 |
| Product Code(s) | PHX |
## Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K161742 | Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System (previously branded as Aequalis PerFORM Reversed Glenoid and Aequalis PerFORM+ Reversed Glenoid, respectively) | PHX |
## Device Description Summary
21 CFR 807.92(a)(4)
The Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System is a modular reverse shoulder arthroplasty system composed of a Glenoid Component Assembly and a Humeral Component Assembly. This system replaces the shoulder joint to help reduce pain and restore shoulder mobility compared to the patient's preoperative condition.
## Intended Use/Indications for Use
21 CFR 807.92(a)(5)
Indicated for use as a replacement of shoulder joints for patients with a functional deltoid muscle and with massive and non-repairable rotator cuff-tear with pain disabled by:
- Rheumatoid arthritis
- Non-inflammatory degenerative joint disease (i.e. osteoarthritis and avascular necrosis)
- Correction of functional deformity
- Fractures of the humeral head
- Traumatic arthritis
- Revision of the devices if sufficient bone stock remains
Notes:
- All components are single use.
- The glenoid sphere implant is anchored to the bone with screws and is for non-cemented fixation.
- All compatible humeral components are available for cemented or non-cemented fixation.
{5}
K260102 (Page 2 of 2)
## Indications for Use Comparison
21 CFR 807.92(a)(5)
The intended use of the subject device, Tornier Perform Reversed Glenoid and Tornier Perform Reversed Augmented Glenoid Shoulder System, is the same as that of the predicate device, the Aequalis PerFORM Reversed Glenoid and Aequalis PerFORM+ Reversed Glenoid (K161742, K182696 and K213124). Both devices are intended for replacement of the shoulder joint to reduce pain and improve shoulder mobility in comparison with preoperative status. There are no differences in intended use that raise new questions of safety or effectiveness.
## Technological Comparison
21 CFR 807.92(a)(6)
The Tornier Perform Reversed system included in this submission has the same core technological characteristics as the currently cleared Tornier Perform Reversed configurations, including intended use, fundamental scientific technology, materials, design principles, fixation method, and sterilization. The proposed changes introduce additional baseplate–glenosphere compatibility configurations and additive manufacturing updates without altering the core technology of the device.
Although additive manufacturing updates and expanded compatibility combinations were implemented, these changes do not alter device performance characteristics or introduce new risks. Any differences in technological characteristics were addressed through nonclinical bench testing conducted in accordance with applicable FDA-recognized consensus standards and previously established acceptance criteria.
Based on the comparison of technological characteristics and the supporting nonclinical performance and biocompatibility data, the Tornier Perform Reversed system is substantially equivalent to the currently cleared Tornier Perform Reversed configurations.
## Non-Clinical and/or Clinical Tests Summary & Conclusions
21 CFR 807.92(b)
Nonclinical performance testing was conducted to support a determination of substantial equivalence for the proposed additional baseplate–glenosphere compatibilities within the Tornier Perform Reversed System. Testing was performed using applicable FDA guidance documents and FDA recognized consensus standards, with test methods and acceptance criteria consistent with those previously established for currently cleared Tornier Perform Reversed configurations.
Submitted and referenced testing includes:
- Range of motion analysis conducted in accordance with ASTM F1378
- Glenoid loosening testing performed per ASTM F2028
- Fatigue testing using previously established protocols to evaluate worst case baseplate–glenosphere configurations
Additional material property and characterization testing was performed, as applicable, to evaluate additive manufacturing updates and to confirm that these updates do not adversely affect device performance.
The results of the nonclinical testing demonstrate that the Tornier Perform Reversed System with the proposed changes meets applicable performance requirements and does not introduce new risks or new questions of safety or effectiveness. Collectively, these results support a determination of substantial equivalence.
No clinical studies were performed.
The nonclinical testing submitted and referenced in this 510(k), including bench performance testing and biocompatibility evaluation, demonstrates that the Tornier Perform Reversed system with the proposed additional baseplate–glenosphere compatibilities and additive manufacturing updates is as safe and effective as the currently cleared Tornier Perform Reversed configurations. Bench testing conducted in accordance with applicable FDA guidance and FDA-recognized consensus standards confirmed that the device meets established performance requirements and does not introduce different risks or different questions of safety or effectiveness.
Biocompatibility was addressed through biological testing, chemical characterization with toxicological risk assessment, and risk-based evaluation in accordance with ISO 10993-1, demonstrating that the relevant biological endpoints are adequately addressed. No clinical testing was required for this submission, as the nonclinical data are sufficient to support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.